Valleylab SSE2-3 Test procedure

Title: Electrosurgical Generator Date: 2-6-2012 By: Valley Lab File = esu-SSE2-3.doc
DISCLAIMER: THIS PROCEDURE PROVIDED "AS IS" AND WITH POSSIBLE FAULTS. USER MUST VERIFY BEFORE USE. NEITHER PROVIDER NOR WEBSITE ASSUMES ANY RESPONSIBILITY FOR ITS USE.
1. General Applies to Valleylab models SSE2K, SSE3 models.
2. Reference Documents Valley Lab Maintenance Notes - see Appendix. Valley Lab SSE3 “Complete Instruction Manual”
3. Tools / Fixtures / Labels
Electrosurgical Analyzer: ex: Dempsey 443. Insulated wire set terminated with banana plugs. Hand Control Pencil, footswitch. Safety Tester
4. Basic PM Procedure
4.1. Inspection
Check unit for any damage, especially the front connectors, the power cord and the footswitch with its cable.
SSE2 ESU
SSE3 ESU
4.2. With no accessories plugged in, turn on unit. For the
SSE2, either the monopolar or bipolar lamp will illuminate; press in the light switch that is off; it should come on. For the SSE3, include the extra mode lamps.
4.3. Setup
Connect analyzer to ESU using a handcontrol pencil; wire the pencil chuck to the analyzer active (this simultaneously will check handcontrol sense circuits.) Connect a banana type lead between the generator’s patient jack over to the same on the analyzer. Set analyzer load to 300 ohms if available. Press in the monopolar switch-lamp. Set the Blend switch to pure cut.
4.4. Cut Mode
With both dials at zero, key the unit in Cut mode, and note audible tone. The cut lamp will not illuminate until power is turned up past approximately Dial = 1. Gradually turn power to max dial and note power into analyzer. Refer to table for acceptable results. For reference, wattages at other loads are listed.
DEMPSEY 443 ANALYZER
BC GROUP ESU-2000A ANALYZER
4.5. Blend Mode
For each of three blend modes (SSE3 only), repeat the protocol for Cut, and note the max dial output power in the table below.
4.6. Coag Mode
With both dials at zero, key the unit in Coag mode, and note distinct audible tone. The coag lamp not illuminate until power is turned past approximately Dial = 1. Gradually turn power to max dial and note power into analyzer. Refer to table for acceptable results. For reference, wattages at other loads are listed.
4.7. Bipolar Mode
Press in the bipolar switch/lamp. Put Cut mode selector in Pure Cut. Connect the analyzer to the two outside socket pins of the generator’s bipolar output connector. Set analyzer load to 100 or 125 ohms. With both dials at zero, use the footswitch to key the unit in Cut mode, and note distinct audible tone. Gradually turn power to max dial and note power into analyzer. Refer to table for acceptable results.
Repeat, but using the coag side of footswitch. Refer to table for acceptable results.
POWER OUTPUTS FOR VARIOUS LOAD @ Max Dial* MODE/LOAD R=500 R=300 R = 125 R=100 Cut 210 300 +/- 10% 270 Coag 110 115 +/- 10% 100
Blend 1 Blend 2 200 Blend 3 200
Listed blend wattages are from the SSE3 manual.
250
Bipolar Cut NA NA 55 60 Bipolar Coag NA NA 22 25
* Note: The wattage levels at 300 ohms are listed in SSE3 manual; powers at other loads were derived from the published impedance curves of the Force series generators.
4.8. Verify volume adjust on back of unit.
4.9. Using the safety tester, check that line leakage with ground open is less than 50 uA.
4.10. Using the safety tester, check that ground lead resistance is less than .15 ohms.
4.11. REM – these units do not have the REM pad sensing feature.
5. Extended PM Procedure
Loading...
+ 1 hidden pages