Valleylab Force GSU Service manual [gb]

Page 1
Qj^^ ]
We
Valley
lab
Force
Service
Manual
Unit
Page 2
Force
GSU
Unit
Service
Manual
Valleylab Inc
Pfizer
Hospital
5920 Longbow Drive
Boulder, Colorado
Forinformationcall:
Copyright © Valleylab
All
rights
permission of Valleylab Inc.
ProductsGroup
reserved.
80301
1-800-255-8522/1-303-530-2300/
Inc,
Contents
USA
1991
ofthis
publication
TWX
910-940-2514
may
notbereproducedinany
form
without
the
written
Page 3
FOREWORD
The
following
manual.
This
Service
covering
warnings
Manual
the
concerning
suppliedwiththe Unit
service
instructions
Manual
covers
Technical
electrosurgery
are
for
the
installation
Specifications.
Circuit
are
use
onlybypersonnel
and
basic
service
Descriptions
beyond
the
scopeofthis
qualifiedtorepair
instructions
and
the
Testingofthe
manual.
and
service
for
the
Force
GSU
Unit
Instructions
The
readerisdirectedtothe
the
equipment
Unit
Also
for
describedinthis
included
use
and
cautions
Force
GSU
are
sections
and
Instruction
Caution:
Effective
Valleylab PartNo.:
Equipment
THEVALLEYLAB FORCEGSU SYSTEM
LIMITED
Valleylab
service optton.ofany
after defective. factoryina
accident
Federal
(USA)
law
Date:
September
945100141
coveredinthis
Force
GSU
Unit:
Electronics Module/Cart
Instruction Service Interface
E0502-12ActiveAdapter
Electro-surgical
(2
Argon
Manual
Cable
fender
GasTanks
WARRANTY
Inc
warrants
for
the
penod(s)
Manual
Generator
washers.2
(shipped
each
set
product,orpart
deliveryofthe
This
producttothe
warranty
does
waysoas,inValleylab's
restricts
1991
A
manual:
Mounting
nylon
screws)
separately)
product
forth
below.
thereof,
original
not
applytoany
this
devicetosalebyoronthe
Valleylab
110-120
220
-240
50/60
Hz
Hardware
manufacturedbyittobe
Valleylab's
which
has
purchaser,
judgmenttoaffect
Force
GSU
Unit
Vac
Nominal
Vac
Nominal
INCLUDES
and
obligation
been
returnedtoitorits
and
which
productorpart
its
stabilityorreliability,orwhich
orderofa
Force
Force GSU™ Handset
REM™
free
from
defectsinmaterial
under
this
warrantyislimitedtothe
Distributor
examination
thereof,
which
physician.
2or
Force4Series
PatientReturn Electrode
Electro-surgical
and
workmanship
repairorreplacementatits
within
the
applicable
discloses,toValleylab's
has
been
repairedoraltered
has
been
satisfaction,
subjectedtomisuse,
under
time
period
outside
Generator
normal
shown
that
the
Valleylab's
neglect
use
and
sole
below
product
or
is
The
warranty
FORCE
ELECTROSURGICAL
MOUNTING FOOTSWITCHES
PATIENT
STERILE DISPOSABLES
THIS
WARRANTYISIN
THE
WARRANTIESOFMERCHANTABIUTY
OBLIGATIONSORLIABILITIESONTHE
OTHER
VALLEYLAB'S COMMUNICATION,
SHALLBELIMITEDTOTHE THERE
HEREUNDERORELSEWHEREINCONNECTION
CONSEQUENTIAL
This
warranty
USA.
Boulder,
Valleylab
without
periods
for
Valleylab's
GSU™
UNIT
GENERATORS
FIXTURES
RETURN
(ALL
(ALL
MODELS)
ELECTRODES
LIEUOFALL
PERSONTOASSUME
PRODUCTS.
VALLEYLAB'S
ARENOWARRANTIES
DAMAGES.
and
the
The
sole
forum
StateofColorado.
Inc.
its
dealers
incurring
any
obligationtomake
NOTWITHSTANDING
AGGREGATE
rights
and
for
resolving
and
representatives,
products
MODELS)
OTHER
FORITANY
LIABILITY
WHICH
obligations
disputes
the
areas
follows:
OneYear
OneYear
OneYear
OneYear
Shelf
Sterility
WARRANTIES,
AND
FITNESS
PARTOFVALLEYLAB.
OTHER
ANY
WITH
PURCHASE
EXTEND
BEYOND
WITH
hereunder
arising
underorrelatinginany
reserve
sameorsimilar
from
dateofshipment
from
dateofshipment
from
dateof shipment
from
dateof shipment
life
only,asstatedonpackaging.
only,asstatedonpackaging.
EXPRESSEDORIMPUED,
FORAPARTICULAR
VALLEYLAB
LIABILITYINCONNECTION
OTHER
PROVISION
RESPECTTOTHIS
PRICE
FOR
THE
THE
TERMS
THE
SALEOFTHIS
shallbeconstrued
the
righttomake
changesinequipment
changesonequipment
NEITHER
WITH
HEREINORIN
AGREEMENT
GOODS
HEREOF.
PRODUCT,
under
and
waytothis
previously
INCLUDING,
PURPOSE,
ASSUMES
THE
SALEORUSEOFANY
ANY
AND
PRODUCTS
ANDOFALL
WITHOUT
NOR
AUTHORIZES
OTHER
DOCUMENT
SOLD
LIMITATION
OTHER
ANY
OF
OR
HEREUNDER
SOLDBYVALLEYLABTOCUSTOMER
VALLEYLAB
governedbythe
warrantyisthe
FOR
INDIRECT
built
built
and/or
DISCLAIMS
OR
lawsofthe
District
Court
and/or
soldbythematany
soldbythem.
ANY
LIABILITY
StateofColorado
for
the
Countv
of
time
Page 4
Service
Centers
Valleylab Inc
Boulder, Colorado,
800-255-8522
USA
Valleylab Australia
Prospect, AUSTRALIA
61-2-688-4888
Valleylab Benelux
Nieuwegein, HOLLAND
31-3402-32456
Valleylab France
c/o
CLIMO
Lyon, FRANCE
33-78-096241
Valleylab Germany
Wichmannstrasse
Postfach
D-2000
GERMANY
040-89
520452
Hamburg 52
68
84
4
Valleylab UK
London,
44-81-993-7808
UNITED KINGDOM
Valleylab Europe, Middle East and Africa
London,
44-81-993-7808
UNITED
KINGDOM
t
ATTENTION:
REFERTOTHE
ACCOMPANYING
DOCUMENTS
DRIP-PROOF
TYPECFEQUIPMENT
LOW
LEAKAGE
SUITABLE
FOR
CARDIAC
USE
Page 5
Table
of
Contents
FORCE
SECTION
VALLEYLAB
Handset
Alarms
GSU
WARNINGS
1
GENERAL
CLINICAL
FORCE
SYSTEM
AND
SERVICE
CAUTIONS
DESCRIPTION
USE
GSU
SYSTEM
OF
FEATURES
Dosage Error 4
Unit
Self
Test
Argon
Gas
Tanks
Storage 5
Adjustments 5
Other Accessories and Adapters 5
SECTION
UNPACKAGING
UNPACKING
2
THE
FORCE
AND
INSTALLATION
GSU
UNIT
...
Unpacking Instructions 7
Valleylab Force GSU UnitFeatures 7
INSTALLING
Responsibility of Attach
Force
Interface
THE
FORCE
the
the
Electrosurgical Generatorto
GSU
Cart
the
Force
GSU
Manufacturer 8
GSU
Unittothe
SYSTEM
the
Electrosurgical Generator 9
Install
the
Argon
Gas
Tanks
Check
the
Power
Connector
Ensure Proper Grounding 11
Perform a Preoperational Checkout 11
1
3
4
4
4
4
4
7
7
8
8
9
11
SECTION
DESCRIPTION
INDICATORS
CONTROLS
INDICATORS
ALARMS
RECEPTACLES
3
OF
CONTROLS,
AND
RECEPTACLES
12
12
12
13
13
Page 6
SECTION
TECHNICAL
OPERATING
STORAGE INPUTPOWER LOW GAS
AUDIO
APPROXIMATE APPROXIMATE
AND
FREQUENCY
FLOW
VOLUME
4
SPECIFICATIONS
PARAMETERS
SHIPPING
SOURCE
LEAKAGE
RANGE
WEIGHT
DIMENSIONS
14
14 14 14 14 14
14
15 15
SECTION
THEORY
FUNCTIONAL
CIRCUIT
SECTION
CHECKOUT
GENERAL
RECOMMENDED
MECHANICAL
Control
Rear
5
OF
DESCRIPTIVE
DESCRIPTIONS
6
TEST
INFORMATION
TEST
INSPECTION
Panel
Panel
Connectors
OPERATION
OVERVIEW
PROCEDURES
EQUIPMENT
Adjustable Platform 22
Power
POWER-ON
FUNCTIONAL
Checking Leakage Current and Ground Resistance
Cord
SELFTEST
TEST
WITH
GAS
..
Equipment Required 26
SECTION
MAINTENANCE
ROUTINE CLEANING STORING
ARGON
INTERNAL CALIBRATING
PRECALIBRATION
CALIBRATION
7
MAINTENANCE
INSTRUCTIONS
THE
FORCE
GAS
TANK
FILTER
THE
PROCEDURE
PROCEDURES
AND
GSU
SYSTEM
REPLACEMENT
REPLACEMENT
FORCE
CHECK
GSU
PROCEDURE
INSPECTIONS
UNIT
16
16
17
21
21 22
22 22
22
23 23
23
26
27
27
27 27
28
30
31
31
36
SECTION
8
OVERALL
TROUBLESHOOTING
SECTION
9
TROUBLESHOOTING
ELECTRONICS
SECTION
10
TROUBLESHOOTING
PRESSURE
SYSTEM
CHASSIS
REGULATOR
in
THE
THE
MODULE
HIGH
MODULE
39
41
57
Page 7
SECTION
REPLACEMENT
RECOMMENDED
ELECTRONICS
11
REPLACEMENT
CHASSIS
PROCEDURES
EQUIPMENT
MODULE
REPLACEMENT
..
Equipment Required 62
Removing
Installing
HIGH
REPLACEMENT
the
the
PRESSURE
Electronics
Electronics
REGULATOR
Chassis
Chassis
Module 62
Module 63
MODULE
Equipment Required 64
Removing the High Pressure Regulator Module 64 Installing the High Pressure Regulator Module 65
PURGING
CONTROL
THE
GAS
DISPLAY
LINES BOARD
REPLACEMENT
Equipment Required 66
Removing
the
Control Display Board 66
Installing the Control Display Board 67
POWER
SUPPLY
REPLACEMENT
Equipment Required 68
Removing the Power Supply 68
Installing
FUSE
the Power Supply 69
REPLACEMENT
Equipment Required 70
COMPONENT
REPLACEMENT
GUIDELINES
61
61
62
64
66 66
68
70
72
SECTION
FACTORY
RETURNING
Returning
Returning CircuitBoards and Other
12
the
THE
Force
SERVICE
EQUIPMENT
GSU
FOR
SERVICE
Unit 73
Subassemblies
..
OrderingReplacement Parts 73
SECTION
SERVICE
FORCE
ELECTRONICS
REAR
CONTROL
SECTION
ASSEMBLIES
GSU
PANEL
DISPLAY
13
PARTS
UNIT
CHASSIS
ASSEMBLY
14
LIST
MODULE
BOARD
AND
SCHEMATICS
73
73
73
74
74
75
76 77
79
iv
Page 8
List
of
Figures
Figure 1.1
Figure 2.1 ESU Interface Cable
Figure
Figure
Figure 5.1 FunctionalBlock Diagram 19
2.2
3.1
Figure
Figure
Figure
Figure
Figure
Rgure 7.3 Low Pressure Regulator 35
System
Connection
Argon
Force
5.2
6.1 Connecting the ForceGSU Handset
6.2
7.1
7.2
Force
to the Force GSU
Leakage
Argon
Internal
Features 6
and
Active Adapter
Gas
Tank
Installation
GSU
Control
GSU
Interconnect
Current
Gas
Tank
Filter
Panel
Diagram
System
Test
Circuit
Replacement
and Proper
Flow
Orientation . 30
10
12
20
24
26
28
9
Figure
Figure
Figure
Rgure
Figure
Figure
Rgure
Figure
Figure
Figure
7.4
8.1
9.1
9.2
9.3
9.4
9.5
9.6 NoGas
9.7
Figure
9.9
9.10
Control
Force
Display
Display
Display
System
No
No
Flow
Cannot
Valve
GSU
Shows
Shows
Shows
Resets
Display
Flow
Gas
Purge
Rate
Achieve0SPLM
Orifice
Unit
Overall
All
All
All
from
Errors
Screw
1's
2's
3's
Handset 47
Location
Troubleshooting
Error
Code 42
Error
Code
Error
Code 44
..
37
40
43
Flow
45
46
48
50
51
Rgure 9.11 CannotAchieve 16 SPLM
v
Flow
52
Page 9
Figure 9.12 Unstable Gas Flow 53
Figure
9.13
Obstruction
Problem
54 ^
Rgure 9.14 IncorrectTank Pressures Displayed 55
Figure 9.15 No/Weak ESU Output
Figure 10.1
Rgure
Rgure
10.2
10.3
System
Gas
Gas
Gas
Leak-Electronics
Leak-High
Leak
Pressure Equipment — 60
Chassis
Module
..
56
58
59
Figure 11.1 Power Supply Replacement 69
Figure
14.1
Force GSU Unit 79
Figure 14.2 Electronics Chassis Module 81
Figure 14.4 Rear Panel
Figure 14.5 High Pressure
Assembly
Assembly
83
84
Rgure 14.6 Control Display Board 85
Figure 14.7 Control Display Board Circuit
Schematics
. 86
VI
Page 10
FORCE
GSU
WARNINGS
SYSTEM
SERVICE
WARNINGS
AND
CAUTIONS
DANGER:
anesthetics,
Use
only
compromised.
High Pressure Hazard: Ensurethat Argon gas have
In filling
overpressurlze
The
gas
Explosion Hazard: DoNOT
gases,
high
been
pressure
the
tanks,
line fitting In
liquids,orobjects.
purity
grade
tested,
ensure
the
tanks.
this
use
the
(99.998%
pure
Force GSU
or better)
tanks
SystemInthe
Argon
gas.
and fittings
Patient
used
presence of flammable
safety
with
may
the
Force GSU
be
certified, and filled accordingtoapplicable standards and local
your
Unit is
supplier
compatible
meets
all local and national
with
many
common
gas
codes.
tanks
Do NOT
(e.g. freon
[tetrafluromethane], helium, krypton, neon, nitrogen, and xenon). ONLY CONNECT ARGON
TO
THE
FORCE
Take appropriate precautions (suchasuse
rule,"
etc.)
Electric
outlet.
Electric
comeincontact
ModuleorTP-1asa
Shock
Shock
GSU
SYSTEM.
when
taking
Hazard: Do NOT Install a
Hazard:
with
measurements
The
lower
it. Do NOT
measurement
of isolated
or
troubleshooting
wet
power
heatsinkonthe
use
It for
ground.
ground
reference.
tools
cord
power
Use
and equipment;
the
Unit.
assembly
supply
the
Is at an
chassisofthe
into
useofthe
the
Unitorinto a wall
elevated
Electronics
voltage. Do NOT
"one
Chassis
System
codes.
GAS
hand
Electrical
Fuse
and
Shock
Replacement:
ratingasthe
Hazard:
For
original
continued
Explosion Hazard: Do NOT
(99.998%
Electric
disassembled
Keep
safely.
pureorbetter)
Shock
the
Handset
Argon
Hazard: Do NOT
and
energized.
electrode
Disconnect
fuse.
use
oxygen
gas
touch
away
from
the
power
protection
for
any
for all
adjustments.
any
exposed
personnel.
cord
before
against
fire hazard,
adjustments.
wiring or
Use
replacing
the
Accessory
parts.
replace
Always
conductive
use
only
with
the
high
purity grade
surface
Holstertohold
while
same
the
the
type
Unit is
Handset
Page 11
CAUTIONS
Improperly Properly
Ensure that the Active Adapter is
secured gas tanks and/orimproper connectionof highpressure gas lines may resultin
connect the ESU interfacecable
fully
for
correct operation ofthe
Force
GSU System.
inserted intothe ACCESSORY receptacle of the electrosurgical
injury.
generatorandthat no metalisexposed. When usingthe ActiveAdapterwiththe Handset, ensure the RF
link
has
no metal
Do not
and
This Unit contains electrostatic sensitivecomponents.
use
cause
workstation.
circuit
boards bytheir nonconductiveedges. Use an antistatic containerfortransportof electrostatic
sensitive components and circuitboards.
exposed.
a knife or similar metal object to remove the
the
connection
Wear
a grounding strap
to
leak.
when
handling electrostatic sensitive components. Handle
filter
assembly hose. This will
When
repairing
the
Unit,
damage
work
at a static control
the
barbs
the
Page 12
SECTION
1
GENERAL
The Valleylab the active electrode of a special electrosurgical handset which is powered by a Valleylab
Series
as standardelectrosurgery. No footswitch is required to operate the
When the user activates the Handset inthe gas-enhanced modes, a
between the electrode and the tissue. The presence of inertgas at the surgical site reduces combustion
with
surgical site, thereby improving visibility.
The Gas-Enhanced Coagulation mode permits the userto coagulatetissue in a more
than with standard figuration and reduces coagulated blood on the tissue surface.
The Gas Shrouded Cut mode enhances the
Electrosurgical
othergases
charring
Pure Cut, Blend 1,
NOTE:
the
Four
and
Using
electrode.
major
Valleylab
DESCRIPTION
Force
GSU System provides a controlled
Generator. The handset
(e.g.,
oxygen,
smoke.
theGas
components comprisethe
Force
This
Blend2,and
Shrouded
GSU
nitrogen,
mode
primarily
Blend3modes.
Cut
Electronics
OF
CLINICAL
provides
etc.)and pushes
ability
enhances
modeat
Valleylab
higher
Module/Cart
USE
flow
gas-enhanced cutting andcoagulation as well
residual
to achieve
cuttingatlower
powerlevelsmay
Force
GSUSystem:
Unit
(hereafter
of Argon(an
inert,
non-reactive gas) over
Force
GSU System.
flow
of Argon gas is delivered
smoke orwatervaporaway
rapid
cutting and hemostasis while
power
levels
increase
referred
to as
"Unit")
Force
2 or4
from
controlled
manner
reducing
andisavailableinthe
arcing
and
sparking
the
from
Valleylab ForceGSU Sterile, Disposable Handset
Valleylab
Valleylab REM Patient Return Electrode
Force
2 or
Force4Electrosurgical
Generator
Page 13
VALLEYLAB
The Valleylab Force GSU System includes the following features:
Handset
FORCE
GSU
SYSTEM
FEATURES
Handswitching Technologyfor Electrosurgery
The Force GSU Sterile, Disposable Handset allows the
electrosurgical or positive tactile
2-Position
Place the
nozzle
Alarms
Designed for
Quality Monitor.
Gas
This alarm alerts the
modes
may
for
REM
Line
are
use
gas
better
Safety
the
Two-Stage
gas-enhanced
response
when
modes.
pressed.
The
Nozzle
flow nozzle in a forward position for more effective
visibilityorfor
use
with a Valleylab Force 2 or 4 Series Electrosurgical Generatorwith
Obstruction
user
rendered nonfunctional untilthe obstruction is removed. Duringthis alarm condition, the
System
inthe Standard Electrosurgery mode.
Low
Tank Pressure
cleaning
the
Alarm
of an obstruction suchasa compressed or kinked
Alarm
This alarm alerts the userwhen either gas tank pressure drops below10% and 5% ofthe volume of
remaining. m
and
gas
control buttons
blade.
Gas
user
Control
to activate the Force GSU
are
color
coded,
gas
coagulation, or retract
gas
System
recessed,
the
line. The
in normal
and
have
the
gas
flow
REM Contact
gas-enhanced
user
gas
a
Dosage
If
the
two seconds.
clear
Unit
The
displays flash 222. When the
Argon
Automatic
Whenone
alternate manually.
Two Argon
along withanadjustable retaining strap. The storage baydoors
Store
Error
gas
flow
rate deliveredexceeds a specified
All
displays flash 111, and the
the
dosage
Self
Test
Unit
performsa hardware and softwareselftest. Whenthe
Gas
gas
Gas
Tank
Gas
Tank
the
safety
error.
Unit
detects a software problem,alldisplays flash 333.
Tanks
Tank
Switching
gas
tank becomes empty(less than
tank, ifavailable, with no
Storage
gas
Safety
Bays
tanks
are
Cap
capsinthe
enclosed withinthe Force GSU cart. Eachstorage bay contains a tank cradle
Holders
rear
baseofthe
gas
limit,
Unit
immediatelyshuts down. Power the Unitoffand on to
5%),
flow interruption to
cart until needed.
the
Unit
the ForceGSUSystem automatically selects the
detects a dosage error inapproximately
Unit
detects a hardware problem, all
the
Handset.
are
self latching.
This
canbeswitched
Page 14
Storage
Storage
The
tilt-out
other
Compartment storage
accessories
compartment holds
and
adapters.
the
Instruction Manual, Handsets, patient return electrodes,
and
Footswitch
These
hooks
Storage
are
located
Hooks
on
the
rearofthe
carttostore
the
Cord Wrap Brackets
Store
the
electrosurgical generator power cord, the Force GSU Unitpower cord,
on
the
cord
wrap
brackets.
Adjustments
Adjustable the
Using
15°.
Locking
Lock
the
Other
Accessories
-
ESU
Generator
tiltadjustment, you
Casters
cart
casters
for safety
and
Interface
Cable
Platform
can
when
Adapters
adjust
the
mounting platform for
installing, removing
- Active Adapter
- Electrosurgical Generator Mounting Hardware (2 fender
footswitch
the
gas
washers,
when
the
electrosurgical
tanks,
and
2 nylon screws)
notinuse.
and
normal
use.
the footswitch cord
generator
for 0° to
-
Instruction
-
Instruction
- Multi-Language
Manual
Card
Labels
Available separately:
- Argon
Gas
Tanks
- Force 2 or Force 4
-
Force
GSU
Handset
-
REM
Patient
Return
Series
Electrode
Electrosurgical Generator
Page 15
FORCE2OR
ELECTROSURGICAL
FORCE4SERIES
(availableseparately)
GENERATOR
CONTROL POWER
STORAGE COMPARTMENT
PANEL/
SWITCH
ESU
LOCKING
INTERFACE
CABLE
REMOVABLE
ELECTRONICS
GAS
GAS
CASTERS
TANK
SAFETY
MODULE
STORAGE
CAP
HOLDERS
BAY
ESU
(2)
POWER
CORD
CORD
TILT
CORD
Figure 1.1
WRAP
BRACKET
ADJUSTMENT
FOOTSWITCH
STORAGE
WRAP
BRACKET
HOOKS
System
Features
D D
BACK
VIEW
Page 16
SECTION
2
UNPACKAGING
UNPACKING
Unpacking
Notify
Valleylab to arrange for repair or replacement of any parts
must have the approval from
number must be displayed on the package label.
information.
THE
Instructions
Carefullyremove the Force GSU and packing materials to
If
you have any questions concerningthe contents, contact
1-800-255-8522
AND
FORCE
use
INSTALLATION
GSU
UNIT
the
Valleylab CustomerService Department. The return authorization
Unit
and allaccessories fromthe shipping packages. Save allcartons
when transporting the
or your Valleylab representative.
See
Unit
Valleylab Force GSU Unit Features
The
Valleylab Force GSU Unit includes
Cart
Electronics
Chassis
Module
the
followingcomponents:
damaged
from shipping.
All
Section 12, "Factory Service," for further
or when returning itfor service.
Valleylab
Customer Service at
returns
The
Electronics
following
-
components:
Touch
Membrane
The control panel is arranged inthree distinctsections for
-
sections
*
*
*
Power
are
Gas
flow
Handset/gas
Gas
tank
Switch
The power switch contains a circuit breaker to protectthe
- PowerSupply
- Control Display Board
- Back
* *
*
Panel
Mass Control Relief
* Filter Assembly
*
Electrical
High
Pressure
Chassis
the
Moduleisremovabletofacilitate
Control
following:
controls
line receptacles
controls
Assembly
Flow
Meter
Valve
Valve
and
Pneumatic
Regulator Module
Panel
Connections
and
system purge
service.
ease
Unit.
This
module
contains
the
of set up and use. These
The HighPressure Regulator Moduleis removable to facilitateservice. This module contains the
following
-
- Lowpressure regulator and tank control assemblieswith
The Force GSU
Hz with no
components:
Sensors
that
Unit
is a universal inputdevice. It iscapable of accepting from85 to 264 Vac at 47 to 63
mainsorpower
sense
tank
supply switching
pressures
individually
necessary.
gas
supply
hoses
Page 17
INSTALLING
THE
FORCE
GSU
SYSTEM
Thisproceduredescribes
howtoinstall
the Force
GSU
System.
Installation
includes the
1. Attach the electrosurgical generator to the Force GSU Cart.
2.
Connect
3. Install
4.
Check
5.
Ensure
the
the
the
proper
Argon
power
ESU interface
gas
tanks.
connector.
grounding.
cable
and
Active Adapter.
6. Perform a Preoperational Checkout.
Responsibilityofthe
Valleylab Incis responsible for the safety,
installation procedures in this manual
Manufacturer
reliability,
are
followed.
and performance of the equipment only
assemblyoperations, extensions, re-adjustments, modifications,or repairs
persons authorized by Valleylab Inc.
the
electrical installation of
such
asIEC
and
BSI.
the
relevant room complies with local
codes
and
the equipment is used in accordance withthe Valleylab instructionsfor use.
following
steps:
if:
are
carried out by
regulatory requirements
Attach
Explosion
anesthetics,
the
Hazard. Do NOT install
gases,
Electrosurgical
liquids,orobjects.
Generatortothe
Valleylab recommends attaching
To attach
1. Removetwo of
2.
3. Attach
4. After mounting
the
electrosurgical generatorto the Force GSU cart:
the
The
removed feet
Center removed rubber feet.
underneath
removed. Tighten
the
the
a. Hold
mechanism.
two rubber collars
generatortothe
the
b. Slightly
mechanism.
c. Tilt
the
slowly lower
are
the
cart
the
mounting platform securely before pressing
lift
the
top mounting
the
bottom feet from diagonally opposite
not
Place
top.
Use
the
screws
generator, adjust
backof
mounting surface.
DANGER
the
Force GSU
Force GSU Cart
the
Electrosurgical Generator to
used
for mounting. Retain
over
the
grommets on
the
generator
cart
using
the
existing
finger tight only.
the
mounting surface and
surfacetothe
on the carttop.
the
two fender
threadsinthe
the
platformasdesired.
desired angle, slowly
SystemInthe
the
comersofthe
the
rubbercollars.
the
cart
top which
washers
press
and
generator
the
button on
the
release
presenceofflammable
Force GSU Cart.
electrosurgical generator.
match
two nylon finger
chassis
button on
the
the
where
the
tilt adjustment
the
tiltadjustment
button to
locations of
screws
the
feet
were
secure,
the
and
Page 18
Interface
the
Force
GSU
Unittothe
Electrosurgical
Generator
An ESU interface cable connects the Force GSU Unitto the electrosurgical generator and and Coag signals through this
This
cableisstoredinthe
Properly Ensure that
electrosurgical
Handset,
To
connect
connect
the
ensure
the
ESU
the
Active Adapter is fully inserted into the ACCESSORY
generator
the
interface
1. Open the Tank 2 side door to
2. Attach
3. Slide
4. Attach
the
ESU interface
the
ESU
the
ESU interface
interface
Monopolar Footswitch receptacle on
link.
This cable also couples the ground of the two pieces of equipment.
Unit
storage
ESU interface
and
that
RF link
has
cable:
access
cabletothe
cable
through
cable
compartment.
CAUTION
cable
no metal is
no metal
for correct operation of
exposed.
When
exposed.
the
using
the rear panel of the Force GSU Electronics
4-pin
from
the
the
the
connector on
rearofthe
rear
panel of
generator
Unit.
rear
the
the
Electronics
panel.
Electronics
Force GSU
receptacleofthe
the
Active Adapter with
Chassis
Chassis
Module to
sends
System.
Chassis
Module
the Cut
the
Module.
rear
the
5. Insert the Active Adapter into the Monopolar Active Footswitch (ACCESSORY) receptacle on the generatorfront panel.
panel.
the
Install
%
Argon
Use
only
compromised.
High
Pressure
GSU
System
standards
national
The
gas
[tetrafluromethene],
%
GAS
TO
\i&^&Zjii£iiiiSSi86&
ESU
INTERFACE
CABLE
^
Figure 2.1 ESU Interface Cable
Gas
Tanks
High
and
codes.
line
THE
Purity
Hazard:
have
local
fitting in
FORCE
Do
Grade
been
codes.
NOT
this
helium,
GSU
(99.998%
Ensure
pressure
In filling
overpressurize
Unitiscompatible
krypton,
SYSTEM.
that
tested,
and
WARNING
pure,orbetter)
Argon
gas
tanks
certified,
the
tanks,
the
neon,
ensure
tanks.
with
nitrogen,
ACTIVE
ADAPTER
\
Active Adapter Connection
Argon
and
and
many
and
gas.
Patient
fittings
filled your
common
xenon).
used
with
accordingtoapplicable
supplier
gas
meets
tanks
ONLY
::x:.':'::':
safety
the
may
Force
all local
(e.g.
freon,
CONNECT
.': 7
..,,,,,,:
V:....
.:.:„.,:::.:,:,::::::::,:.:,...
be
and
ARGON
Page 19
To install
the
Argon
gas
tanksinthe
Force GSU Unit:
1. The
2.
Argon
wish to
storage
Follow
Gas Installation Instruction labels
use
an instruction label which istranslated in another language, retrieve the label from the
compartment
and
the appropriate safety precautions for handlinghigh pressure
3. Makesure the Force GSU
mode.
4. Lock
the
cart
casters.
TANK
Open
BUCKLE
(English
place
it inside
Unit
is OFF,and the electrosurgical generator is OFFor inStandby
the
side
VALVE
door.
the
side
language) are inside the
doorofTank
1.
GAS COUPLING
LARGE
gas
tanks.
LINE
KNOB
RETAINING
Unit
STRAP
side door.Ifyou
5. Unbuckle
Figure 2.2 Argon
the
retaining strap and ensure that allpacking materials
Gas
Tank Installation
are
removed.
6. Carefully place the tank intothe tank cradle (tanksweighapproximately 35 lbs (16
7. Loosely attach the retaining strap by snapping the buckle together.
8. Remove the
remove
9. Orientthe tank
valve fitting, remove it.
the
gas
tank safety
gas
tank identificationtag(s).
with
the valve outlettowards the backof the cart.Ifthere is a safety plugon the tank
cap
and place it inthe
gas
cap
holder in the
base
of thecart. Do NOT
10. Secure the tank using the retaining strap.
11. Verifythat
- Looking at
12.
Align
large
Improperly secured
result in injury.
the
knob
the
correct
the
gas
linecouplingto the tank valve outlet. Carefullyconnect the coupling bytightening the
hand
front of
tight.
hose
is connected to
the
Unit,Tank 1 is on
gas
tanks and/or improper connectionof high
the
correct tank (Hose 1 to Tank 1
the
leftside.).
CAUTION
pressure
and
gas
kg)).
Hose 2 to Tank 2
lines may
10
Page 20
13.
Open
the
a. b.
tank valve
Press
the
Connect
the
and
Unit
power
Force GSU
check
for leaks.
switch in
Handsettothe
the
ON
(I)
position.
System.
c. Power on the electrosurgical generator and place itinthe Ready mode. d.
Press
the green (+) button on the Handset to place the
Unit
e. Press the yellowCUTor blue COAGbutton on the Handset. While
pinch offthe
*
The
* An audible alarm
*
The
f.
Release
gas
lineto obstruct
OBSTRUCTION
electrosurgical
the
pinched
gas
indicator
sounds
generator
line.
gas
flow.
Verify
the following:
illuminates.
for approximately three
deactivates.
seconds.
inthe
Gas
On mode.
gas
flowsto the Handset,
14. Closethe side door and repeat this procedure forthe other tank
1-2
buttontoswitchtothe
Check
The Force GSU
the
Power
Connector
Unit
is supplied witha hospital-grade power connector. Your
withthe proper connectorforyour operating
other
tanktobe
room,ifdifferent fromthat supplied.
tested.
(if
available). Press
Unit
TANK
SELECT
should be equipped
The powerconnector meets all requirementsforsafe grounding. Itspurpose should not be defeated by
using extensioncords or 3-prongto 2-prong adapters. The powercord assembly should be periodically
checked fordamaged insulation or connectors. Cords shouldalways be grasped bythe connector. Do
NOT pullonthe
Ensure
Proper
cord
itself.
Grounding
To ensure patient safety, the Force GSU System must be properlygrounded.
Unit
The ground wire inthe power cord connects to the
flowfrom or
2-prong
Undesirable leakage currents
responsibility of the
the
Unit.
the
Unit
electrical
cabinetinthe
adapters.
user
to ensure proper polarity and grounding ofthe power outlets supplyingpower to
event
of internal electrical failure. Do not
are
affected by the polarization of
chassis and ensures that no dangerous currents
use
extension
the
50/60 Hz input power. Itis the
cordsor3-prong
NOTE:Itis the user'sresponsibility to ensure thattheelectrical
with
local
codes
PerformaPreoperational
Before placing the Force GSU
6,
"Checkout
and regulatoryrequirementssuchasIEC
Checkout
System
Procedures."
intoclinical use, perform a system
and BSI.
11
installation
of therelevantroomcomplies
checkasdescribed in Section
Page 21
SECTION
3
DESCRIPTION OF
A
ON
1
0
OFF
POWER
CONTROLS
CONTROLS,
80
m-ca
\OBSTRUCTtOt\
©
1
QASON
I
GAS
®
OFF
| G
4SRCWI
1
GAS
|
INDICATORS
A
FLOW
V
Figure 3.1 Force
4
GSU
~&>Ar
m ®
PURGE
LINE
GSU
AND
RECEPTACLES
II
HANDSET
GAS
Control Panel
pop
% PRESSURE % PRESSURE
I 7VWJCJ I I
LOPRES
/K
B8B
WMCg
I
LOPRES
QD
TANK
SELECT
1-2
^
|
I
ON
J_
O
OFF
GAS
FLOW
TANK
SELECT
1-2
<2>
PURGE
INDICATORS
GAS
UNE
FLOW
Power On/Off Switch - This switch supplies power to
GSU
Unit
and
containsacircuit
Flow
Control - Allowsthe user to adjust the Argon gas flow setting.
Each press of a green
one
change
increment. Continuously pressing
GAS
flow setting to increase or
minimum
limit.
breaker.
FLOW button
causes
decreaseinstepstothe
the
the
button
maximum/
the
Force
Flowsetting to
causes
the
Tank Select Button- Select eitherTank 1 or Tank 2 manually using
this
blue
button.
Gas Line Purge Button- Pressing this button discharge of Argon gas through the gas tubing at
gas
setting. This is used to purge the
Gas
Flow Indicator- This indicatorilluminates
or
Gas
Shrouded
Cut
featureisactivatedorwhen
tubing.
button is pressed.
causes
the
when
the
a brief
current
the
Gas
PURGE
gas
Coag
flow
LINE
TANK1
TANK
2
Tank Indicators -
the
selected
gas
The
appropriate indicator illuminates to indicate
tank.
12
Page 22
GAS
GAS
0
0
ON
OFF
Gas
On
Mode
Indicator-
green
(+)
button is
indicates that
pressedonthe
gas
is available to
Coag or Gas Shrouded Cut
activated.
Gas
Off
Mode
Indicator-This
gray (-) button is
indicates that Handset
canbeusedinthe
pressedonthe
gas
is not available to
This
indicator
Force
the
mode
activates
indicator
Force
standard Cut
illuminates
GSU
when
Handset. This
Handset, therefore
when
the
illuminates
GSU
the
Handset, therefore
and
when
Handset. This
Coag
modes.
the
the
Gas
Handset is
the
the
%
ALARMS
88
<2>
eee
PRESSURE
LOPRES
OBSTRUCTION
Flow Display- Gas flow is indicated on a scale of 1 to 10 which
corresponds to a range of approximately 2 to 16 standard liters
per minute.
% Pressure Display - Indicates tank pressure
0%.
Low
Pressure
illuminates
flashes
5%.
Gas
Line
the
indicator
The
Alarms-The
when
and a
tone
Obstruction
illuminates
initial
gas
flow setting is 1.
Tank1or
the
tank pressure is below 10%.
sounds
when
Alarm-When
andanalarm
both
Tank2LO
tank
a line
tone
status
PRES
The
pressures
obstruction
sounds.
from
indicator
are
occurs,
100%
indicator
less
to
than
RECEPTACLES
*?J
GSU
l°A^
HANDSET
-^>
Ar
®
GAS
Handset Receptacle -
Force
GSU
Handset.
Gas
Line Receptacle - This luer receptacle
filter of
the
Handset
13
Accepts
gas
the
line.
3-pin
connector of
accepts
the
the
particle
Page 23
SECTION
4
TECHNICAL
OPERATING
SPECIFICATIONS
PARAMETERS
Temperature Range: 10° to 40° C
Humidity Range:
Atmospheric
STORAGE
Pressure:
AND
Temperature Range:
30%
to 75%,
500to1060
SHIPPING
-18°
to 70° C
non-condensing
millibar
Humidity Range: 10% to 100%, non-condensing
500
Atmospheric Pressure:
INPUT
POWER
SOURCE
No switching required.
Current: < 1
ampere
to 2500 millibar
The
line frequency may vary between 47 and 63 Hz.
Maximum Operating Range: 85 - 264 Vac Nominal Operating Range:
LOW
All
FREQUENCY
patient-connected
LEAKAGE
terminals tied together:
(50/60 Hz)
Source current normal polarity,intact chassis ground Source current normal polarity,ground open Source current reverse polarity, ground open Sink current, 110V applied, ailinputs
110V
<10uA
<100uA <100uA
<100uA
Sink current, 220V applied, allinputs
Chassis source current, ground
GAS
2-16
AUDIO
FLOW
standard
VOLUME
RANGE
liters/minute±10%offull
open
<100uA
scale.
110-120
220-240
220/240V
<2.0uA
<50uA <50uA
<50uA
<100uA
Vac
Vac
The
alarm
tones
are
set
to a level of >65 dBA at 1 meterand are not adjustable.
14
Page 24
APPROXIMATE WEIGHT (two full
166
pounds;
75,3
kilograms
tanks,
no generator)
APPROXIMATE
21 x39x28inches;
Use
Argon
gas
DIMENSIONS
53x99
tanks
with
x 71
the
centimeters
following
specifications:
Outside Diameter: 7 inches (18 centimeters), max
Height (with valve):
Pressure:
Max
Specifications
subjecttochange
24
inches
3000
(60 centimeters), max
psi (20600 kPa) (200 bar, nominal)
without notice.
15
Page 25
SECTION
5
THEORY
This
section
Block
Diagram
found inthe Control DisplayBoard schematic inSection 14 of this manual.
FUNCTIONAL
The
electronicsofthe
following:
Direct
• ControlArgon
Monitor
-
gas
-
low
-
dosage
- power supply voltage ]
OF
OPERATION
provides
the
appearsatthe
DESCRIPTIVE
interface
error
line obstructions
tank
signals
gas
deliveryat digitally preset
conditions:
pressures
error
circuit
Force
from
for
theory
endofthis
GSU
Unit
the
Force
each
tank
and
basic
section.
information
The
regarding
reference
designators
OVERVIEW
performavarietyoffunctions.
GSU
Unittothe
flow
associated
rates
the
Unit
circuitry.AFunctional
usedinthis
These
electrosurgical
functions
generator
section
are
consistofthe
(ESU)
-
proper
- software errors T
• A special error routinefor lineobstructions to exhaust the pressurized
• Automatic selection or switching of Argon
• Allows
Force
• Allowsfor
• Allowsfront panel control
• Digitally displays the % pressure available in
Has
Handset
• Allows
Provides
clock signals
gas
tank
lines
the
surgeontoselect
GSU
Handset
the
user
to purge
isolation electronics to prevent dielectric breakdown between
button
the
safe
unit to
shut
functions
operate
downincaseofpower
gas-enhanced
the
attached
and
monitoringof the Argon
from 85 to
264
gas
Vac
failure.
operation or normal electrosurgical operationthrough
tubing
gas
tanks
each
Argon
gas
tank
and
between47and
gas
inthe line
the
RF electrical signal
63 Hz
and
the
the
16
Page 26
CIRCUIT
The
Valleylab Force
microprocessor is
functions
DESCRIPTIONS
the
Unit requires.
GSU
Unit is controlled by an Intel 8749H single-chip microprocessor (uP). This
the
main control device, and with support hardware, performs all
the
necessary
An Isobloc circuit isolates
the
keying leads
maybeelevated
leads cannotbecoupled to ground. U18 forms a 60 kHz oscillator
The
and R29.
The
switches on
oscillator output drives a flyback Power Supply Circuitthat provides a floating DC supply.
the
signals are optically isolated and
determines which
key
octal buffer (U4) interfaces
An octal latch (U3) performs external
Enable Not (ALE/) line is
microprocessor first places
ALE/ in
that
the
octal latch. The latchoutputs are used as chip selects forthe various external circuitmodules
the
microcontroller
U10 is a Pulse Delay Circuitthat delays cycle to
(DAC). This The
meet
timing requirements
ensures
microprocessor
the microprocessor ifit is not toggled at a predetermined rate. This
and
Handset
detects
setupcoded
the
to relatively high RF voltages. To maintain an isolated generator,
then
senttothe
is closed and performs
the
microprocessorto the control panel and
address
the
latching signal. In ailexternal
the
address on the address/data
accesses.
setbythe
that data is stable at
system
algorithm interfaces with a hardware
keying signals from
keying signals through
microprocessor fordecoding.
the
necessary actions in
selection
the
microprocessor's READ and WRITE signals by
analog to digitalcircuit (ADC) and
the
time that it is read or written.
and
the
Handset. This is
whose
the
frequency is
multiple comparator (U17).
The
responsetothe
the
keying signals.
chip
select
generation.
address
bus
read
and
write
line, and this information latches using
the
Watchdog
ensures
Timer Circuit (U2) that
that
the
necessary
set
because
by C19, R28,
microprocessor
key
closure. An
The
Address
accesses,
the
one
digitalto analog circuit
microprocessor is
runningits control program and has not failed. Another function of the Watchdog Timer Circuit is
inhibiting
period by shutting off
the
operation of
the
the
tank valves and the electrosurgical generator keying during the initialization
switched
+12VSW
during this time.
these
These
Latch
machine
resets
The
thermal
that
hasa0-4
output is DC amplified by U20D to give a
mass
flow
The
Dosage
of range. It is controlledbythe
comparator level on U21A. This is compared to a feedback signal of
flow if
(DOSERR/),
DOSERR is considered a machine error and requires pressing
and on to clear
the dosage errorcomparator.
seconds.
A
second
an LM350 (U22)
mass
flow meter
Vdc
output at J5 corresponding to
meter
is 21
degreesC,760
Error
Circuit (U8D, U21A-B) determines if
the
two signals don't agree.
sounds
an audible alarm, and flashes all 111's on
the
error condition. U8 in this circuit is an integrator to eliminate
D/A converter(U5)
which
then
measures
the actual gas
0-16
0-8
Vdc signal for
flow.
SLPM (standard liters
torr, and calibrated with Argon gas.
the
D/A converter(U7).
The
microprocessor also
The
time
needed
sets
the
gas
flow rate. This analog signal
delivers a controlled voltage to
The
fora DOSERR tobedetected is approximately two
This component is a true
the
rest of
gas
flow rate delivered is within limits or is out
microprocessorwrites a word to
the
the
checks
the
the
the
displays if an error is detected.
the
Unit power switch to turn
control
valve.
basically a microprocessor-controlled digital/analog servo loop regulating
the
flow immediately
presentinthe
rate,
and
microcontrollerto allow
the
second
A display multiplexerchip (U15)
of an
8x8
RAM that contains
shuts
off if a Dosage ErrorSignal is detected.
the
gas
flow.
The
signal is
the
GASEN/
addresses
the
BCD format for
signal that allows
the
7-segment
the
display digitsaswellasthe
Two
first signal is
the
the
LM350 to
LED displays. This read-only device consists
mass
flow device
per
minute).
circuitry.
This
The
standard forthe
the
0-4
D/A to
actual flow rate and inhibits
output of this circuitry
the
noise
from
the
output of
connectstothe
This
the
Argon
actuating signals
digital
word
operate.
LED lamps.
adjustterminal of
entire circuit is
gas
flow. Note that
must
that
sets
the
The
be
Vdc
set
gas
Unitoff
flow
SEG
a
H
17
Page 27
line multiplexes the LEDlamps on and offthrough Q13 and Q14. U16 amplifiesthe output signals for the
LEDs
(the
digit
drive
lines).
U23
makesupthe
and amplifiesthem to be digitizedfor control algorithm use. The offset and gain adjustment pots allowfor H
differencesintransducer
differential amplifiersthat receive the pressure signals from
and
amplifier
performance
and
balance.
the
two tank transducers
A bridge-type pressure transducer (U19)
delivery
and has
An
digitizes them. The microprocessor
wellasfor inputsto control algorithms. Inaddition to controlvoltages, the converter monitorsthe +12V H
power
microcontrollergenerates thetank valve control, generatorkeying signals, etc.
using eithersmallsignal FETs, relaysor bipolar transistorsto ensure enough drivecurrent and/orvoltage ^ to the device beingcontrolled. These devices also isolatethe microcontroller outputs fromthe devices
under
The
generates
for. _
Anumber ofimportant internal test signals have been broughtoutto TP1-12 to
troubleshooting and finaltesting. These signals are the
Test
TP-1
TP-2
TP-3
TP-4
TP-5
TP-6
TP-7
TP-8
TP-9
TP-10
TP-11
TP-12
tube.
U20Aisa
R60
sets
the
gain of this circuit. This level ofthe signal (DLPRES) determines ifa line obstruction
occurred causing
A/D
converter
supply
controlincaseoffailure.
Audio
Point
voltage
Annunciator
the
oscillator
the
(U6)
for
differential
pressureto increase beyond allowable limits.
takes
the
analog
internal
Circuit
enable
Name
AGND +5
+12FIL
-12V
DLFLOW
DLPRES
T1PRES
T2PRES
FLOCON
ERRLIM
DOSERR
3MHZ
calibration
consistsofan
signal (TONEN/)
senses
amplifier
uses
and
signals
these
and
(tank
signals to ensure properoperation of control blocks
ensures
oscillator
and
the delivered pressure at
U20Dagain
pressures,
and
inhibits
following:
Function
analog ground
+5 volt supply
+12 voltfiltered supply
-12
volt supply
0-8V = 0-10V=0-30
T1 pressure 10V=3000 psig
T2 pressure 10V=3000 psig
0-10V
0-10V
5V CMOSerror signal
microprocessor clock oscillator signal
signal to flow valve
signal to error limitcircuit
the
proper
FET
the
0-16
SLPM output
stage.
flow
speaker
oscillatorfunction until a
psig output
the
proximal
R67
provides
control,
error
limit
operationofvoltage
and
then buffers them
driver.
The
simplify
endofthe
offset
adjustment,
voltage,
references.
etc.)
and
as
The
microcontroller
tone
is called
Unit
A45 watt switchingpower supplypowers the Force GSU
0.5 amps, and +5 Vdc at 3 amps. Itis a universal inputdevice, meaning it is capableof accepting from 85 to 264 Vac at 47 to 63 Hzwith no mains or power supply switching necessary.
The pneumatic portion of the Force GSU Unitconsists ofthe high pressure plate,
two cylinders of
into a compartment in the
tank's supply line
pressure
The
90 psi lines from
reducedto35
compressed
goes
is reduced to 90 psi. Each
each
psi
and
Argon
base
of the Force GSU cart to the High Pressure Regulator Module. Here
to a pressuretransducer and then through
tank
join together
controlled by
gas
pressurized to a maximum of
gas
the
line then
servo
and
go through
system
passes
Unit.
Itprovides +12 Vdcat 2 amps,
the
3000
psi.
The
the
first
stage
of regulation where its
through the electrical shutoff valve for
another
described
18
above.
regulator
where
flowcontrol,
gas
is piped down
the
pressure
-12
that
Vdcat
and
each
tank.
is
Page 28
AC
85-264
47-63
GAS
LINE
Vac
Hz
OUTPUT
HANDSET
- • - i
ISOBLOC
CIRCUIT
CONTROL
PANEL
RELIEF
VALVE
CIRCUIT
BREAKER
2
AMP
8749H
MICROPROCESSOR
SWITCHING
POWER
SUPPLY
W/2AFUSE
AUDIBLE
ALARM
DISPLAYS
+12
VDC
-12VDC
+
5VDC
INTERFACE
CABLE
ESU
GAS
TANK
INPUT
GAS
ELECTRICAL
LINE
1.2
micron
FILTER
PRESSURE
XDUCERS
FOR
TANKS
1AND2
1st
STAGE
REGULATOR
A/D
PRESSURE
XDUCER
TANK
SWITCHING
I
D/A
90
PSI
D/A
2nd
REGULATOR
STAGE
FLOW
METER
35
PSI
0.1
FILTER
micron
Figure 5.1 Functional Block Diagram
19
Page 29
AC
RECEPTACLE
£
o
AC
INPUT
CART
ASSEMBLY
HANDSET
GASIN
r-oa-
0}
I
-LD-
I I
|
L
MODULE
FILTER
CONTROL
PANEL
BOX
ji
J2
J3
J4
J6
A
FLOW
VALVE
PRESSURE
UNE
FILTER
POWER
SUPPLY
s
J5
PCB
J7
:>-<:
TRANSDUCERS
H
J8
J9
CIRCUIT
BREAKER
PRESSURE
XDUCER
>i
FLOW
METER
2A
D-i
1
y
RELIEF
VALVE
V
GAS
.OUT
'--
-JOQ--IFILTER
P
n 1
.j
GAS
LINE
ELECTRICAL
PRESSURE
XDUCER
1
1st
STAGE
REGULATOR
HIGH
2nd
STAGE
REGULATOR
Figure 5.2 Force GSU Interconnect Diagram
20
PRESSURE
1st
STAGE
REGULATOR
ASSEMBLY
?n
PRESSURE
XDUCER
2
/
ARGON
A
TANK2
|
J
Page 30
SECTION
6
CHECKOUT
This section provides checkout procedures to
the Unitfails to meet any of
PROCEDURES
the
checks, refer to Section 8, "Overall SystemTroubleshooting."
verify
that the Force GSU System is functioningproperly.
Some checkout procedures require partial disassemblyof the
Procedures,"for
GENERAL
Electric
Do NOT
Electronics
Take appropriate precautions (such as
hand
Electric
Unit Is
disassembly
TEST
Shock
INFORMATION
Hazard:
comeIncontact
Chassis
rule," etc.)
Shock
disassembled
when
Hazard: Do NOT
instructions.
The
lower heatsinkonthe
with
it. Do NOT
ModuleorTP-1asa
taking
and
measurementsortroubleshooting
touch
energized.
any
WARNING
use
it for ground. Use
measurement
use
of Isolated
exposed
power
Unit.
Refer to Section 11, "Replacement
supply
ground
tools
wiring or
is at an elevated voltage.
the
chassisofthe
reference.
and equipment;
the
Unit.
conductive
useofthe
surface
while
If
"one
the
CAUTION
The
Unit
contains electrostatic sensitive components. When repairingthe
Unit,
work at a static controlworkstation. Wear a grounding strap when handling electrostatic sensitive components. Handle the circuitboards by their nonconductive edges. Use an antistatic container fortransport
of electrostatic sensitive components and circuitboards.
Whentesting,
follow
propertesting procedures inorderto duplicate manufacturertest data. Keeptest
leads as short as possible. Lead inductance and straycapacitance can affect meter readings adversely.
Useof uncompensated scope probes maycause largeerrors inmeasurements. When measuring
microampere leakage currents, accidental capacitive or inductivecoupling may
order-of-magnitude errorin
the
observed values.
Performthe leakage current and ground resistance tests before returningthe
cause
Unit
to clinical use.
21
Page 31
RECOMMENDED
You
will
need
the
TEST
following equipment to perform
EQUIPMENT
equipment isused, itmustmeetor exceedthe
the
checks
specifications
described in this section. Ifsubstitute
ofthe recommended equipment. ^
Description
Specifications
DigitalMultimeter Fluke 77 or 87, or equivalent
Oscilloscope Tektronix 2445, or equivalent Leakage Current
Leakage Table
Row
Meter 0
Tester
UseULload device or commercially available leakage Per
IEC specifications
-20
SLPM, 21
- Sierra Instruments
degreesC,760
821-S,
Force Electrosurgical Generator Series 2 or 4
2 bottles of Argon Withknown
gas
pressures
(within 100 psig)
CylinderPressure Gauge/Regulator 0 - 3000 psi cylinderpressure,
Gas
Pressure
Force
GSU
MECHANICAL
This inspection
Unitis NOT
Control
Handset
connected
Panel
Gauge
INSPECTION
assumes
familiaritywith
0-35
the
psig
Unit's controls, indicators,
to powerwhen performing this inspection.
torr, calibrated with Argon
or equivalent
0-50
and
psi delivery pressure
receptacles.
Ensure
tester
that
gas
the
1. Check
the
followingcontrol panel receptacles for
damage
or corrosion:
a. GAS (luer lock)
b.
GSU
HANDSET
2. Using a
3. Check
Rear
Verifythat all connectors
the
Panel
banana
plug,
check
ON/OFF switch for
Connectors
are
each
connection for obstruction
damage
or corrosion.
properly and securely tightened.
and
secure
fit.
Adjustable Platform '
Check that
1. Simultaneously
2. While pressing
ranges.
3. Return the platform to the desired position,
the
Unitplatform easily adjusts.
lift
the
platform slightly,
the
release button, raise and lower
and
press
and
the
the
release
release
button.
platform to
the
release
the
button.
limitsof
the
high
and
low
22 J
Page 32
Power
1.
Remove
Cord
the
power
cord
from
the
Unit.
2. Disassemble the power cord connector and inspect for
screws
3. Assemble
4. Clean
install
POWER-ON
This
1. Plugthe
2.
Press
3.
The
the
the
test
verifies that
the
Unit
are
tight.
the
power
powercord with a
powercord
SELF
Unit
and electrosurgicalgenerator powercords
power switch in
conducts
connector
assembly.
TEST
the
Force GSU Unitpowers on and properly performs the self test.
a self
the
test.
and
inspect
damp
cloth and milddetergent. Let
ON
(I)
position.
Check
the
following:
the
power cord for
WARNING
damage
into
or corrosion. Ensure that connector
damage.
the
power cord completely dry and
a hospital-grade grounded receptacle.
a. Bothtank indicators illuminate for approximately0.5 seconds at the start of the test.
b.
All
other
LED indicators illuminate all
seconds).
888s
for the remainder of
the
self
test
(approximately 10
4. After
FUNCTIONAL
Thistest provides a checkoutprocedureto
itis
the Force GSU
Force Electrosurgical Generator, referto the appropriate Instruction
as
1.
the
self
test,
verify that:
a. Tank 1 is selected.
b. A single
c.
GAS
powered
tone
FLOWissetto1
TEST
on and used
Unit,
(If
Tank 1 is empty or closed, Tank 2 is selected.)
sounds.
WITH
with
and that the
and
gas.
the
Unit is in
GAS
This
Argon
the
Gas
Off
mode.
verify
thatthe ForceGSUSystem is
functioning
test assumes thatthe electrosurgical generatoris connectedto
tanks are installed. Forcomplete instructions regarding usingthe
Manual.Ifthe
Unit
does not perform
properly
described inthis test, refer to Section 8, "OverallSystem Troubleshooting," forfurther information.
Connect
a. Attach the filterend of the Handset
the
Force
GSU
Handsettothe
System.
gas
tubingto the
Gas
Line Receptacle on the
Unit
control
panel. Rotate clockwise untilfinger tight.
b. Plugthe 3-pin Handset connector
into
the
HANDSET
receptacle onthe Force GSU
Unit
panel. Do NOTplug the 3-pin Handset connector intothe electrosurgical generator.
link)
to the
Active
c. Connect the Handset banana lead (RF
CAUTION
Adapter on the electrosurgical generator.
when
control
Ensure that the
Active
Adapteris
fully
inserted
into
the
ACCESSORY
generator and that no metalisexposed. When usingthe
RF link
has
no metal
exposed.
23
Active
receptacle ofthe electrosurgical
Adapter
with
the Handset, ensure the
Page 33
GAS
LINE
FILTER
Figure
6.1
Connecting the Force
GSU
Handsetto the Force
2. Power on the electrosurgical generator and placeitin
a.
Press
the generator rear panel power switch ON
(I).
b. Afterthe generator performs a self test, short the generator
c.
Press
the
generator
d.
Ensure
Keep
Handset
3.
Verify
a. Press the
Set
b.
Ensure
that all
the
Handset electrode away from personnel. Usethe AccessoryHolster to
safely.
that the control panel controlsoperate correctly.
GAS
the
gas
flow to 10.
that
both
tanks have sufficient pressure.
c.
Press
and release the PURGE
approximately
current
gas
setting.
READY button.
generator
FLOW
gas
power levels
UP/DOWN
tank
valves
buttons.
are
Verify
LINE
are
open.
that the
button.
threeseconds, and thatthe
set
at one.
WARNING
Verify
Pressthe
Unit
Verify
Unit
discharges
that thedisplay varies
GSU
System
the Ready mode.
REM
connector.
hold
the
from
1 to 10 and 10 to 1.
TANK
SELECT
toggles between
that the GAS FLOW indicator illuminates for
Argon
gas
1-2
TANK
through
button.
1 and
the gas
Verify
TANK
that
2.
tubing
both
atthe
4. Verify
gas
flow
and
ESU control.
a.
Press
the green (+) button on the Handset.
tone
sounds.
Verify
b. Press the yellow CUTand blue COAGbuttons, and
that the GAS ON indicator illuminates
verify
that the ESU activates after
approximately a 0.2 second delay, and that gas flowsto the Handset when activating CUT and
COAG.
24
and
a
Page 34
c.
Press
sounds.
Handset.
the
gray (-) button on
Press
the
yellow
the
Handset. Verify that
CUT
and blue COAG buttons,
the
GAS
OFF indicator illuminates, a
and
verify that no
gas
flows to
tone
the
5. Verify that
a.
Press Press
b.
tone
c. Press
Handset, pinch off
*
* An audible alarm
* *
d. With the gas line
the
Gas
Line Obstruction Alarm operates properly.
the
GAS
FLOW UP/DOWN buttons to
the
green
(+)
button on
sounds.
the
yellow CUT button orthe blue COAG button on the Handset. While
the
gas
The
OBSTRUCTION
sounds
released.
The
gas
line
depressurizes.
The
electrosurgical generator deactivates.
still
pinched off, press the yellowCUT button orthe blue COAG button on the
the
Handset. Verify that
line to obstruct
indicator
for approximately three
the
illuminates.
set
gas
the
gas
flow to 5.
the
GAS
flow. Verify
seconds
ON indicator illuminates
gas
the
following:
or untilthe CUT or COAG button is
Handset. Verify that the alarm recurs. Release the pinched gas line.
6. Verify that
a. Turn off both
yellow CUT button on
the
Low Pressure Alarms operates properly.
tank
valves.
the
Press
the
Handset.
GAS
FLOW UP button to
set
the
gas
flow to 10. Press
b. Observe a decrease in the pressure of the tank being used. Verifythat LO PRES indicator
illuminates
for
the
tank
below
10%.
and
flows to the
that a
the
c. Observe a tank switch when the pressure is below 5% and a flashing LO PRES indicator.
d. Continue to press the yellowCUT buttonon the handset. Verifythat the LO PRES indicator
illuminates
for
the
second
tank
below
10%.
e. Verifythat both LO PRES indicators flash and that an audio alarmsounds when both tanks are
below
5%.
f. Press the gray (-) button on the Handset. Verifythat the audible alarmdoes not sound.
g. Pressthe
mode
h. Press
mode
7. Press
8. Press
9.
Remove
the the
yellow
illuminates.
the
blue COAG button on the Handset. Verifythat the generator's blue lamp in the Coag
illuminates.
Unit power switch in
CUTbuttononthe Handset.
the
OFF (O) position.
Verify
thatthe generator's
yellow
lampinthe Cut
generator rear panel power switch OFF (O).
the
REM
connector
short.
25
Page 35
Checking
Leakage
Current
and
Ground
Resistance
Check
Third wire leakage current is
the1kohm
the
leakage current
resistorfrom
and
ground resistance before returning
measured
chassis
to ground. The maximum voltage
microamps leakage would be 100
for
this
test.
Line cords longer
than
15 feet or extension cords increase the third wire leakage
recommended. Withthe Force GSU
microamps.
Equipment
DVM
Leakage
Required
current
tester
by opening the
millivolts.
Unit
0.15
the
Force GSU Unitto clinical
green
grounding wire at
across
Commercially available leakage current
the
plug
the1kohm
and
and
resistorfor
testers
are
not
connecting
maybeused
turned off,the thirdwire leakage should be less than 10
10.2
uf
ohms
-VVV-n
VNAA
1000
ohms
use.
100
Figure 6.2 Leakage Current
1.
With
the powercord assemblyconnected
cord ground prong to all
0.1
ohms
at 25
amps.
NOTE: To eliminate
exposed
effectsofcontactresistance, this
2. Measure the leakage currentofthe
metal on
Unit
following
only
the
tothe
chassis.
shouldbea
the instructions supplied
Test
Circuit
Unit,
measure the resistance throughthe power
The
measured
4-wire
resistance
resistance
with
yourleakage current
must
not
measurement.
tester. This leakage currentmust notexceed 100 microamps at 120voltsor 50 microampsat 220
volts.
3.
Record
an
the resultsforreferencein
impending failure.
future
resistance/leakage tests. A
significant
change mayindicate
exceed
26
Page 36
SECTION
7
MAINTENANCE
ROUTINE
Valleylab recommendsthat
months. Check
Valleylab recommends replacing
Calibrate
maintenance
This Service Manual describes
major repairs,
available from Valleylab. Refer to Section 13, "Service Parts List"
Schematics"
CLEANING
Clean clean the cart surfaces, control panel,
MAINTENANCE
the
the
Unitevery six months ifnecessary. The procedures describing how to perform this
are
within
the
for
additional
INSTRUCTIONS
the
Force GSU Unitusing standard hospital procedures. Use a milddetergent
PROCEDURES
AND
the
Force GSU Unitbe inspected by qualified service personnel every six
powercord assembly periodically for
the
this
manual.
the
Unitor modules
information.
INSPECTIONS
internal filterand functionally checking
recommended inspection, testing,
can
be returned to Valleylab. Repair parts
and
cords. Do not allowfluids to
damaged
insulation or connectors.
the
Unitevery six months.
and
calibration procedures. For
and
information
and
Section 14, "Assemblies
and
enter
the
chassis. Do not
caustic, corrosive, or abrasive cleaning materials. The Unit, cart surfaces, control panel,
be
completely dry before
use.
damp
and
cords must
are
and
cloth to
use
The
Force
GSU
Unit
Electric
wall
STORING
When
the
Shock
outlet.
THE
Force GSU
cannot
Hazard: Do NOT install a
FORCE
System
be
sterilized.
GSU
is not in
SYSTEM
use,
WARNING
wet
power
disconnect the Force GSU Unit powercord assembly from
the wall receptacle. The ESU interface cable connecting
remain
connected
unless
the
electrosurgical
generatorisused
cord
assembly
the
Unit
and
electrosurgical generatorshould
by itself with a footswitch.
into
the
Unit or
into
a
27
Page 37
ARGON
Use
GAS
only
compromised.
High
Pressure
GSU
System
standards
national
The
gas
[tetrafluromethene],
GAS
TO
TANK
REPLACEMENT
WARNING
High Purity Grade (99.998% pure,orbetter)
Hazard:
have
and
local
codes.
Do NOT
line fitting in
THE
FORCE
been
codes.
this
helium,
GSU
Ensure
pressure
that
tested,
In filling
overpressurize
Unit Is
compatible
krypton,
SYSTEM.
Argon
the
neon,
gas
tanks
certified,
tanks,
the
tanks.
with
nitrogen,
ensure
and
many
and
Argon
and
filled
your
common
gas.
fittings
Patient
used
with
safety
the
accordingtoapplicable
supplier
xenon).
gas
ONLY
meets
tanks
CONNECT
all local
(e.g. freon
may
Force
be
and
ARGON
To replace
1. Follow
the
Argon
gas
tanks inthe Force GSU
the
appropriate safety precautions for handling high pressure
cap(s) before proceeding.
2. Make sure the Force GSU
the
valves on both Argon
3. Lock
the
cart
casters.
TANK
BUCKLE
Open
Unit:
gas
tanks. Obtain tank safety
Unit
is OFF, the electrosurgical generatoris offor in Standby mode, and
gas
tanks
are
completely closed.
the
side
door.
VALVE.
GAS
LINE
COUPLING
LARGE
KNOB
RETAINING
STRAP
Figure 7.1 Argon
4. Itis normal for a small amount of Argonto
fromthe high pressure
turning
the
large knob. Use a wrench to assist loosening this connection.
5. Loosen but do not remove
6. Attach
7. Unbuckle
the
safety
the
retaining
lbs (16 kg).) Refillor replace
8. Carefully place
the
gas
line, holdthe
the
retaining strap.
captothe
strap
top of
and
the
the
carefully
tank.
replacement tankinto
tank.
Gas
Tank Replacement
escape
gas
lift
the
during this procedure. To disconnect
the
linecoupling stationary and loosen the coupling by
the tank from the cart. (Tanks weigh approximately 35
tank cradle.
28
tank
Page 38
9. Loosely attach the retaining strap by snapping the buckle together.
10. Remove the
remove
11. Orient
the
the
tank
valve fitting, remove it.
12.
Secure
13. Verify that
- Lookingatthe
14. Align large
the
the knob
the
gas
hand
Improperly result in injury.
15.
Open
the
valve
Press
a.
b.
c.
d. e.
the
Connect
Poweronthe
Press
the green (+) button on the Handset to place the
Press
the yellow CUTor blue COAG button on the Handset. While
pinch off the
*
The
OBSTRUCTION
* An audible alarm
*
The
electrosurgical
f.
Release
gas
tank safety
gas
tank identification tag(s).
with
the
tank using
correct
front of
line coupling to
tight.
secured
"™"_
and
check
Unit
power
the
Force
cap
valve outlet towards
the
retaining strap.
hoseisconnectedtothe
the
Unit,
the
gas
tanks and/or improper connection of high
'""""".„
for leaks. switch in
GSU
electrosurgical
gas
line to obstruct
indicator
sounds
generator
the
pinched
gas
and place it inthe
Tank
1 is on
tank
valve outlet. Carefully
CAUTION
the
ON
Handsettothe
generator
gas
flow.
illuminates.
for approximately
deactivates.
line.
gas
cap
holder in the
the
backofthe
cart.Ifthere
correct tank. (Hose 1 to
the
left side.)
connect
„.„
,...,
(I)
position.
System.
and
place it in
the
Unit
Verify
the following:
three
seconds.
base
of the cart. Do NOT
is a safety plug on
Tank1and
the
coupling by tightening
Ready
pressure
mode.
gas
inthe Gas On mode.
gas
flowsto the Handset,
Hose
2 to Tank 2
lines may
the
tank
the
16.
Close
the
side
door.
29
Page 39
INTERNAL
FILTER
REPLACEMENT
Thisprocedure describes howto replacethe
internal
filter
and hose assembly.
Valleylab
recommends
internalfilterreplacement every six months. Obtain the internalfilterpart number from Section 13,
"Service Parts List," and orderit from Valleylab CustomerService at
1-800-255-8522
or your Valleylab
representative.
Equipment
Slotted
Pliers
1. Turn off both Argon
2. Purge both
3.
Press
4. Unplug
5. Remove
Procedures,"
6. Remove
Required
Screwdriver
the
Unitpower switch inthe OFF (O) position.
the
the
the
1/8"
gas
tanks.
gas
lines. (Refer to Section 11, "Replacement Procedures," for further information.)
Unitpowercord from
Electronics
Chassis
the
wall receptacle.
Modulefrom the
Unit.
(Refer to Section 11, "Replacement
for further information.)
cover
from
the
module by loosening
the
two
thumb
screws
located on
the
module
panel.
7. Locate
the
internal filter,
and
note
the
properflow orientation.
rear
FLOWCONTROL-
VALVE
HIGH
VALVE
PRESSURE
MASS
FLOW
METER
ESU INTERFACE
RELIEF
VALVE
INTERNAL
FILTER
HIGH
PRESSURE
TRANSDUCER
HOSE
CLAMPS
7. Using
8. Gently
the
pliers, remove
loosen
and
Rgure 7.2 Internal Filter
the
black
hose
remove
the
internal filter
and
clamps from
and
hose.
30
Proper FlowOrientation
each
sideofthe
filter
assembly.
Page 40
CAUTION
Do not
barbs
9. Connect
use
a knife or similar metal object to remove the filterassembly hose. This
and
cause
the
connection
the
new internal filter and
to
leak.
hose
and replace the hardware. Verifythe properflow orientation.
10. Secure the Electronics Chassis Module inthe
for further information.)
11. Verify
12.
13. Unplug
14.
15.
that
there
are
no leaks.
a. Reconnect the Argon b. Plug
c.
d. Turn on e. Verify
Press
Place
module
Secure
for further information.)
the
Unitpower cord into a hospital-grade grounded receptacle.
Press
the
power switch in
the
gas
that
there
connections.
the
Unitpower switch in
the
Unit
power
the
coveronthe
rear
panel.
the
Electronics
tanks. arenoleaksatthe
gas
hose
to the Electronics Chassis Module rearpanel.
the
ON (|) position.
the
OFF (O) position.
cord
from the wall receptacle.
module. Attach
Chassis
Module in the Unit. (Refer to Section 11, "Replacement Procedures,"
the
Argon
cover
will
damage the
Unit.
(Refer to Section 11, "Replacement Procedures,"
gas
hose
connections
by tightening
the
two
and
thumb
the
internal filter
screws
located on
assembly
the
CALIBRATING
This
procedure
Precalibration
the
it. If
Unit
Procedure. If
according to
continue performing
PRECALIBRATION
The
Precalibration
performed in This
procedure
the Argon tanks
THE
FORCE
contains
Check
passes
the
Unit
the
appropriate corresponding
order
assumes
are
a Precalibration Check Procedure
Procedure to verify that
the Precalibration Check Procedure, itis not
does
not
passastepofthe Precalibration Procedure,
the
Precalibration Check Procedure.
CHECK
Check
to continue
Procedure contains six steps. Each
the
that
the
installed.
The
GSU
PROCEDURE
next Precalibration
electrosurgical
test
UNIT
and
the
Force GSU Unitis functioning properly before calibrating
stepofthe
Calibration Procedure. Afterthe adjustment,
step.
generatorisconnectedtothe
points referenced in this procedure
a Calibration
necessary
step
contains
Procedure.
to perform
then
adjust the Unit
substeps
Force
are
inthe Control Display
Perform
the
that must be
GSU Unit,
Calibration
and
Board schematic in Section 14. For complete instructions regarding using the Force Electrosurgical
Generator, refer to
Electric Do NOT
Chassis
the
appropriate Instruction Manual.
Shock
comeIncontact
Hazard:
The
with
ModuleorTP-1asa
lower
heatsinkonthe
it. Do NOT
measurement
WARNING
use
it for
ground
power
ground.
reference.
supplyisatanelevated
Use
the
chassis
of
voltage.
Electronics
the
that
31
Page 41
Equipment
DMM
Flow meter (21
Oscilloscope
Force
Mass
Regulated
Pressure
Slotted
Small, slotted, plastic or nonconductive adjusting tools
Allen
wrenches
Required
GSU
Handset
flow
meter
power
gauge
screwdriver
degrees
(capable
supply
and
transducer
1/8"
C, 760 torr, calibrated with Argon gas), Sierra Instruments
of accurately reading 20 seem)
821-S,
or equivalent
The Precalibration Check Procedure includes the
PRECALSTEP 1.0 Check the power supply.
PRECAL
PRECAL
PRECAL STEP4.0 Check
PRECAL STEP5.0 Check
PRECALSTEP 6.0 Check
Perform
PRECAL STEP 1.0
1.1
1.2 Use a
these
Access
Procedures,"for further information).
switchingpower supply voltages at TP1-4 on the Control DisplayBoard.
STEP
2.0 Check
STEP
3.0
stepsinthe
the
Electronics
DVM
withfine pointtest leads or clipleads and make the
Test
Point Voltage
TP-1
Check
order
Check
Chassis
the
the
the
the
the
listed.
the
Module
Ground
tank pressure.
flow
obstruction pressure.
dosage
lowpressure regulator.
power supply. H
rate.
from
following
error.
the
Unit.
steps:
(RefertoSection
following
11,
"Replacement
measurements ofthe
TP-2 TP-3
TP-4
+5.0 +12
Vdc±0.6V
-12
Vdc±0.6V
Vdc±0.25V
1.3Ifthe power supplyvoltages are notcorrect, adjustthem per the Calibration Procedure.
PRECAL
2.1
2.2
2.3 Purge both
2.4 Verifythat
2.5Ifthe
Plug
Press
STEP
2.0
the
Unit
and
electrosurgical
the
Unitpower switch in
gas
lines. (Refer to Section 11, "Replacement Procedures," forfurther information.)
the
Tank 1
known
tank
pressures
Check
and
the
tank
pressure.
generator
the
ON
Tank2 % PRESSURE indicators display 0% (zero %).
are
not correct, adjust
(I)
power
position.
32
cords
intoahospital-grade
Verify
the Unitperforms a self test.
them
per
the
Calibration
grounded
Procedure
receptacle.
in this section.
Page 42
PRECAL
3.1
3.2
STEP
3.0
Connectanew
control
Connect
Ensure
panel.
the
that
the
Force GSU
Handset
Active
Check
the
Handsettothe
nozzletothe
Adapter
is fully inserted into
flow
mass
rate.
Gas
Line Receptacle
flow
meter.
CAUTION
the
ACCESSORY
and
3-pin
receptacleofthe
connectoronthe
electrosurgical generator and that no metal is exposed. When using the Active Adapter with the
Handset,
3.3
Poweronthe
a.
ensure
Press
the
RF link
electrosurgical
hasnometal
generator
and
the generator rear panel power switch ON
exposed.
place it in
Ready
(I).
mode.
b. After the self test, connect a patient return electrode to the generator. c. Ensure that all generator power levels are
Keep
the
Handset
Handset
safely.
electrode
away
from
WARNING
personnel.
set
at one.
Press
the generator
Use the Accessory Holster to hold the
READY
Unit
button.
3.4
Press
the green
3.5
Press
and
these
flow
pressure.
If
these settings are NOTwithinthe acceptable range, check all
Verifythat the flow rates for
Setting
1
2
3
4
5
6
7
8
9
10
3.6
Release
the
(+)
button on the Handset to place the
hold
the
Handset
rate settings are
each
Nominal
2.0
3.6
5.1
6.7
8.2
9.8
11.3
12.9
14.4
16.0
CUTorCOAG
CUT or COAG button,
within
the acceptable range, proceed to Step 4.0. Check the obstruction
setting
Flow
button
are
withinthe specified range.
Acceptable
0.4-3.6
2.0-5.2
3.5-6.7
5.1-8.3
6.6-9.8
8.2-11.4
9.7-12.0
11.3-14.5
12.8-16.0
14.4-17.6
to turn off
the
and
gas
Unit
check
flow.
inthe Gas On mode.
the
flow rate at settings 1
flow
rates at settings from 1 through 10.
Range
and
10.
If
3.7Ifany flow rate is not withinthe specified range, adjust the
this
section.
33
flow
rate perthe Calibrate Procedure in
Page 43
PRECAL
4.1
4.2
STEP
4.0
Press
the
yellow CUT or blue COAG button on the Handset. While
off
the
gas
line to obstruct the
a.
The
OBSTRUCTION
b.
Pressureisreleasedbyan
is
heard
from
the
c.
The
control valve is disabled
d. An audible alarm
released
e.
The
line
depressurizes.
f.
The
electrosurgical
Release
the
pinched
Check
indicator
Electronics
sounds
generator
gas
line.Ifthe
the
obstruction
gas
flow.
illuminates.
automatic
Chassis
andnogas
Verify
pressure.
the
following:
actuation ofaninternal relief valve,
Module.
flows.
for approximately three
deactivates.
Unit
does
not respondasdescribed
seconds
gas
or until
flows to
the
CUT or COAG button is
when
pinched off, adjust the obstruction pressure per the Calibration Procedure inthis section.
the
Handset, pinch
and
the
the
gas
releaseofgas
line is
PRECAL
As a safety feature,
dosage
the flowwith a
5.1 If
5.2 Disconnectthe
5.3
5.4
5.5
5.6 Activatethe
STEP
5.0
the
error condition must occurfor a minimum oftwo
mass
flowmeter. Makesure the
not
connected,
flow
Connecta0-12
Connect
Press
dosage
the
the
error (DOSERROR) condition occurs. (The
Record the
Vdc regulated power supplyto the
+ to pin 1
green
(+) button on
Unit
inCUT orCOAGat settings 1 and 9.
flow
rate when DOSERRORoccurs (approximatelytwo seconds). Return the power
supply voltageto zero.
error condition
and
acceptable range, proceedto
If
the
limits
are not
indicatedinthe
Flow
1
2
3
4
5
within
table
below:
Setting
6
7 8 9
10
Check
the
dosage
error.
Unittriggers an error condition if
Unit
is powered on and the
connect
the
Handset
controlvalve
and
the - to pin 2. Turn on the power supply,
the
Press
the
repeat
the
test
Step
nozzle
from
connectorJ7 onthe
into
the
Handsetto place the Unitin
Unitpower switch to turn the Unitoff
for
the
next setting.Ifthe
6.0. Check the low pressure regulator.
the acceptable range, then activate the
DOSERROR
3.6-6.8
5.2-8.4
6.7-9.9
8.3-11.5
9.8-13.0
11.4-14.6
12.9-16.1
14.5-17.7
16.0-19.2
17.6-20.8
Limits
the
delivered flow
seconds
mass
to alarm. To
flow
Control
endofJ7
Slowly
Unit
that
the
increase power supply voltage untilthe
flashes all 111sand a tone sounds.)
DOSERROR limits
Unit
exceeds
gas
meter.
Display
goestothe
and
Gas
On mode.
and
a specified
check
this feature, monitor
tank valves
Board.
valve.
set
itto zero to begin. *i
on again to
are
inCUTor COAGat each setting
are
clear
within
limit.
open.
the
the
The
5.7
Ifa
dosage
error
does
not
occur
Troubleshooting," for further information.
with
the
specified limits, referto Section 10, "Overall
34
System
Page 44
PRECAL
6.1
Disconnect the
middle
STEP
6.0
connector).
Check
high
pressure plate cable
the
low
pressure
from
regulator.
the Electronics Chassis
Module
rear panel
(bottom
6.2 Connect tothe highpressure cable to a pressure gauge and
psig.
6.3Ifthe pressure is not
within
this specification, adjustthe
verify
low
pressure per the
6.4 Reconnect the highpressure plate cable to the Electronics Chassis
6.5 Replace the Electronics Chassis
Procedures,"
for further information.)
Module
inthe
Unit.
(Referto Section 11, "Replacement
LOW
PRESSURE
that the pressure is 35.0 ±0.5
Module
REGULATOR
Calibration
rear panel.
Procedure.
This
concludes
the
Precalibration
Figure 7.3 Low Pressure Regulator
Check
Procedure.
35
Page 45
CALIBRATION
PROCEDURE
Perform the Precalibration Check Procedure before calibratingthe Force GSU
Unit unless it is
perform the appropriate step ofthe Calibration Procedure. This procedure
that
the
that
that
the
the
CAL CAL CAL
CAL CAL
that a Force GSU
The Calibration Procedure includes the following steps:
necessary.Ifthe
Unitis
disassembledasthe
electrosurgical
Handset
Argon
STEP STEP STEP
STEP STEP
gas
tanks
1.0
Adjust
2.0 Adjust
3.0
Adjust
4.0 Adjust
5.0 Adjust
Unit
does
not
passastepofthe
Precalibration
generatorisconnectedtothe
is connected to the Force GSU
are
installed.
the
power supply.
the
tank
pressure.
the
flow rate.
the
obstruction pressure.
the
low pressure regulator.
Check
Unit
Precalibration
Procedure
System
Unit.
assumes
requires
Do NOTadjust the
Check
Procedure,
the
following:
then
CAL STEP1.0 Adjustthe power supply. I
Use the small slotted adjusting tool to adjust R26 on
TP-2.
all
three
R26
controls all
voltages
three
into
specification.
output voltages,soit maybenecessarytomake
the
switching powersupply module for +5.1 Vdc at
several
adjustmentstoget
Test
Point
TP-1
TP-2 TP-3
TP-4
Aftermaking
CAL
STEP
2.1 Observe
%
display 0% continuously.
2.2 Observe
%
fiOA
0% continuously.
NOTE:
accuratelymeasuring the highpressure
2.3 Tum on the Tank 1 valve. Adjust R56 the tank. Valleylab
2.4 Turn on
the tank. Valleylab
these
adjustments, continue the Precalibration Check Procedure.
2.0
the
Tank 1 % Pressure. Adjust R63 (on
PRESSURE
PRESSURE
ortntint
Use
a fulltankforthis adjustment Theaccuracy of the Unitdigitalreadout depends upon
the
indicator displays 0 (zero) %. Do not adjust R63
the
Tank 2 % Pressure. Adjust R65 (on
indicator
irti
ielw
Tank 2 valve. Adjust R51 (T2 span) to read the appropriate % for
Adjust
sets
sets
Voltage
Ground
+5.0 +12
-12
the
displays
the
the
0%.
Unitwith
Unit
with
from
Vdc±0.25V
Vdc±0.6V
Vdc±0.6V
tank
pressure.
the
Control Display Board)sothat
the
Control Display Board)sothat
Do
not
adjust
R65
beyond
the
tanks.
(T1
span) to read
full
tanks for 100% equals 2000 psig.
full
tanks for 100% equals 2000 psig.
the
appropriate % for
beyond
the
point
the
the
point
whereitbegins
the
to
whereitbeginstodisplay
the
known pressure in
the
known
pressure
in
Aftermaking
these
adjustments, continue the Precalibration Check Procedure.
36
Page 46
CAL
STEP
3.0
NOTE:
Various
pressureof 760
3.1
Calibrate
a.
Connect
b.
Press
c.
Press
d.
Press
potentiometer located on the rear panel)
flow.
flow rate is in specification.
e.
Press
f. Adjust
meterindicates 2.0 ± 0.2 SLPM of flow.
seconds
g.
Press
h.
Verify
interactiveand steps 3.1b-3.1g may need to be repeated several times
are
attained.
Adjust
flow
meter
manufacturers
torratsea
the
initial
the
the
GAS
levelas a
flow
Handset
FLOW
rate.
the green (+) button on the Handset to place the
and
hold the Handset CUTor COAG button. Adjust
Allow
the
flowto stabilize for approximately 10
the
GAS
FLOW
the
ZERO (tower adjustment potentiometer located on
the
tube
into
UP
buttontoset
DOWN
flow rate.
use
standard.
the
mass
buttontoset
different
flow
the
until
Allow
standards.
meter
flow
rateto10.
Valleylab
inlet.
Unit
the
the
mass
flow
seconds
the
flow
rateto1.
the flowto stabilize for approximately 10
between adjustments untilthe flow rate iswithin specification.
the
GAS FLOW UP button to
the
flow
rate is 16.0 ± 0.2 SLPMfor
set
the
flow rate to 10.
flow.
The SPAN and ZERO adjustments are
uses21 degrees Canda
inthe
Gas
On mode.
SPAN (upper adjustment
meter indicates 16.0 ± 0.2 SLPMof
between adjustments untilthe
the
rear
panel) until
until
both
the
flow
mass
settings
flow
3.2 Connect a
3.3
Press
the GAS FLOW
Press
3.4
the green (+) button on the Handset to place the
3.5 Pressthe
0.25 Vdc.
panel)asfollows:
a.
If
the voltage is too high, turn the screw counterclockwise.Ifthe voltage is too
screw clockwise. Press and holdthe yellowCUTor blue COAGbutton on the Handset while
making this adjustment.
Release the CUTor
stops.
c.
Pressthe
± 0.25
attain this specification. Release the CUTor COAGbuttonto stop
DMM
across the J7 terminals. Pin 1 ofJ7 is positive, and pin2 of J7 is negative.
DOWN
yellow
CUTor blue COAGbutton on the Handset.
If
the voltage is outside ofthis range, adjust the control valve orificescrew (on the rear
yellow
CUTor blue
COAG
button to
buttonand
COAG
set
the
gas
flowto 1.
Unit
allow
the gas
flow
buttonon the Handset.
inthe
Gas
On mode.
Verify
the voltage across J7 is 6.0 ±
to cease. Repeat
Verify
the voltage acrossJ7 is 6.0
until
low,
the gas
turn the
Vdc.Ifthe voltage is outside ofthis range, repeat steps 3.5a and 3.5b as necessary to
gas
flow.
<s>
o
0
30C
A
9
9
0
"'ft'
- CONTROL VALVE
ORIFICE
O
SCREW
flow
o
0
Figure
s
f
7.4
Control Valve Orifice
37
Screw
1
Location
O
Page 47
3.6
Check
accurately reading 20 seem, measure the leakage rate.Ifa mass graduated cylinder
thegas leakage rate
filled
withwater and invertedintoa largercontainer of water. Insertthe free end
(20
ccper
minute,
maximum).
Using
a mass
flow
flow
metercapableof
meter is not available,
ofthe Handset tubing intothe open (submersed) end ofthe cylinder and measure the time to
accumulate 20 cc of
gas.
An acceptable time is
one
minute.
use
a
Aftermaking
CAL
STEP
EXPLOSION HAZARD: Do NOT
Grade
4.1 Connect a
4.2
Set
4.0
R36
these
adjustments, continue with
Adjust
(99.998%
DMM
fully
pure,orbetter)
across
TP-1
counterclockwise. Observe
the
obstruction
use
Argon
(ground)
the
Precalibration Check Procedure.
pressure.
WARNING
oxygen
and
for
any
adjustments.
gas
for all
adjustments.
TP-6.
TP-6
(DLPRES)with an oscilloscope
Always
use
High Purity
set
to 50 mV div.
Adjustpotentiometer R67 untilthe DCvoltage is zero. Do notover-rotate the pot beyond the point
whereitzeroes
4.3 Apply 27 ± 1 psig from a regulated Argon
transducer (U19)on
damaging
4.4 Adjustpotentiometer R36
sounds.
± 1 psig.
the
transducer.
gas
supply to the luer
the
Control Display Board. Do not apply pressure greater
until
the
OBSTRUCTION
indicatorilluminates and an audible alarm
fitting
going to the pressure
than
30 psig to avoid
Aftermakingthis adjustment, continue withthe Precalibration Check Procedure.
CAL
STEP
5.0
5.1 Loosen the
locking
Adjust
the
nut on top ofthe
low
pressure
Low
Pressure Regulator, and
regulator.
use
output pressure to 35.0 ± 0.5 psig.
5.2 Retighten the locking nut on top of the lowpressure regulator.
5.3 Couplethe hose to the eibow. Connect to a pressure gauge and
0.5 psig.
verify
Aftermaking this adjustment, continue the Precalibration Check Procedure.
an alien wrench to adjust the
that the pressure is 35.0 ±
38
Page 48
SECTION
8
OVERALL
Referto the ForceGSU Instruction
determine whatthe
manual to help locate
Information
SYSTEM
Unit
and
TROUBLESHOOTING
Manual
is not
doing
correctly.
correct a malfunction.
to learn howa properly operating
Refer
tothischapterandthe
about troubleshooting and repairing majorcomponents inthe Electronics Chassis
inSection 9, and the High Pressure Regulator Moduleis in Section 10.
If
youare not
Operation," to better understand the
for
the
proper disassemblyof
Electrical
the
parts.
contact
ModuleorTP-1asa
This Unitcontains electrostatic sensitive components. When repairing
control workstation.
Handle
electrostatic sensitive
familiar
Unitisdisassembled
The
with
the
with
the
information
the
Unit.
Shock
circuit boards bytheir nonconductive edges. Use an antistatic containerfor transport of
Hazard: Do NOT
lower
heatsinkonthe
It;doNOT
Wear
components
and
use
measurement
a grounding
inthese troubleshooting sections, referto Section5, Theory of
Unit
functions.
touch
energized.
power
it for
ground.
and
ground
strap
circuit boards.
Referto the Section 11,"ReplacementProcedures,"
WARNING
any
exposed
Disconnect
supply
Use
the
reference.
CAUTION
when handling electrostatic sensitive components.
wiringorconductive
the
power
Is at an
elevated
chassisofthe
Unit
should
following
cord
chaptersinthis
surfaces
before
replacing
voltage;doNOT
Electronics
the
Unit,work at a static
Chassis
function
Module
come
and to
is
while
in
the
Let
Perform a visual inspection of
Ifonly a single function is not working, concentrate on
•Ifseveral symptoms
Look for
symptoms
solution to
or burnt (discolored)
and
the
test
are
the
problem is not obvious,
any
obvious conditions
components.
resultsbeyour guide.
the
Unitfirst.
present, locate a circuitthat is common to all
check
the
power supply circuits first.
that
maybethe
causeofthe
the
circuit
that
controls that function.
the
affected functions.Ifthe
malfunction: loose connections, broken
39
Page 49
/
FORCE
GSU
UNIT
I
OVERALL
V
TROUBLESHOOTING
\
J
J
SEE
ELECTRONICS
MODULE
TROUBLESHOOTING
SECTION
ELECTRONIC
CHASSIS
9
YES
PNEUMATIC
SEE
HIGH
PRESSURE
REGULATOR
TROUBLESHOOTING
SECTION
MODULE
10
Figure
8.1 ForceGSU
Unit
OverallTroubleshooting
40
Page 50
SECTION
9
TROUBLESHOOTING
This
section
describes
howtotroubleshoot
THE
ELECTRONICS
the
Electronics
Chassis
CHASSIS
Module.
MODULE
Refer to the Force GSU Instruction Manual to learn how a properly operating Unitshould function and to
determine what the
help locate
If
you
and
are
not familiar with
Unit
is not doing correctly. Refer to this chapter and the
correct a malfunction.
the
information in
following
these
troubleshooting sections, refer to Section 5, "Theory of
chapters inthis manual to
Operation," to better understand the Unitfunctions. Refer to the Section 11, "Replacement Procedures," for the
proper
disassemblyofthe
Electrical
the
parts.
contact
ModuleorTP-1asa
Unit is
The
with
Shock
Hazard: Do NOT
disassembled
lower
heatsinkonthe
it;doNOT
Unit.
and
use
it for
measurement
WARNING
touch
any
exposed
energized. Disconnect
power
ground.
supply
ground
CAUTION
is at an
Use
the
reference.
chassisofthe
wiring or
the
power cord
elevated
conductive
before
surfaces
replacing
voltage; do NOT
Electronics
Chassis
while
come
in
This Unitcontains electrostatic sensitive components. When repairing the Unit, work at a static
control workstation. Wear a grounding strap when handling electrostatic sensitive components. Handle the circuit boards by their nonconductive edges. Use an antistatic container fortransport of
electrostatic sensitive components and circuit boards.
41
Page 51
COtaPUT
all
BBBOR COM J
IIO¥! ^
i*a
1
Bapalz Or
Raplaca
Coatrol/Dlaplay
Board
Baoallbtata
Syataa
taa
flaotloa
l'l
IHDXCAtl
actual
d0ba8i
bxcbxds
dbbibxd
bi
3.2
fL»
Dlacoaaaet
flow
latar
At
Baeoaaaet
Olaeoaaaet
UaaAtBaar
7
Natar
Kodala
Aad
Bos
View
eaa
Of
Maaaara
Voltaga At J7
Via
1
(Coattol
Valra)
rata
Oilfaca
Adlnitlag
Scrav
Ob
Coattol
7al»a
Cloekvlaa
Baplaea
Coattol
Valra
Oa
Baar
Baaal
Of
Hodala
Bapalr Or
Baplaea
Coatiol/Dltplay
Board
Bos
Baeoaaaet
UaaAtBaar
Of
Hodala
Saa
Box
Figure 9.1 Display
Adjsat Coattol
Valva
Shows
42
lor
C.OV
At2SLBB
Saa
Sactloa
7
All 1's Error
Code
Page 52
DISPLAY
ALL
BBBOB
2'S
XHCOBBSCf
BSBOLt
AM A10D
cemuxoi
Verify
Syataa
Coaaaotloaa
sacata
Vrooaadlaf
880W8
2* S
CODS
XVDXCAIB
7X01
that
Oroaad
Bafora
All
Ara
Irror
Oparatloa
Xadleataa
lateralttaat
Problea
Adjaat
Baplaea
Bopply Aa
Veeeaaary
Barlag
Pover
Or
If
Problea
Perelata
Bapalr
Or
Baplaea
Coatrol/Dleplay
Board
Brobabla
Xa 05 Bapalr Or
Baplaea
Coatrol/Dleplay
Board
Figure
Caaaa
9.2
Display
Probable
Caaaa
Xa Of Bapalr Or
Baplaea
Coatrol/Dleplay
Board
Brobabla
Caaaa
Xa 07 Bapalr Or
Baplaea
Coatrol/Olaplay
Board
Shows
All
2's
Error
Code
43
Page 53
(DISPLAY
ALL
IBBOB
Softvaze
Baplace Dl Oa
Coatrol/Dlaplay
Board
::
3'S
CODB
SSOOS
Btror
\
1
/
Figure 9.3 Display Shows All3's
44
Error
Code
Page 54
1
SlSIXIf
ST8TIB 18AIOKCIXOV
OP
BAICEDOS
CXBCDXI
BB8BIS
::
BB8BT
111
I
02
H0«-BBSSt kobial
Biapoaas
to
povxb
DBOPOOta,
conxcta,
Dlaeoaaaet
Coatrol/Dleplay
Board
Lias
DX8-
itc
The
Pollevlag
teata
Shoald
Be
Perforaad
If
Problaa
Paralata
Proa
Power Sapply
Aad
Beaoaaare
the
the
18
Software
or
•leroprocaaaor Hot
Baaattlag
Watchdog Ilaer
Bapalr or
Baplaea
Coatrol/Dlaplay
Board
Hardware
Balfaaetloa
Baplaea 01
Bapalr
Or
Baplaee Power
Sapply
Figure
9.4
System
45
Resets
Page 55
fBODISPLAY
J
Connect
Cotd
*
Secure
Bepalr Or
Beplaee
Contiol/Dliplay
Board
Power
Hiring
Replace Fuaa
Cheek Filter
Internal
External
Power
Line
And
And
Corda
Beplaee Power
twitch
Dlieonneet
Control/Display
Control/Diaplay
On
Board
Repair Or
Replace
Board
J5
Replace Power
Supply
Dlsconneet
Control/Display
ReconnectJSOn
Control/Diaplay
Board
Board
Rateat
JS
And
Rgure 9.5 No Display
46
Page 56
Preaa
LIBI
Oa
Replace
Baadaat
Bapaat
Proat
Paael
P0B01
Battea
teat
Aad
Praaa
Battoa
Saadaet
Coaflra
OB
Problea
•oa-Faaetloaal
Baadaet
0A8
Bode
Oa
Aad
OAS
OB
Was
Figure
BO
0A8
(BODOLB
9.6
PLOW
BOX)
No
Gas
Flow from
47
Handset
Page 57
Correct
Aad
Bepeat teat
Preae
Select
to
Switch
laaka
teak
Battoa
Beplaee Oaa
Or
Both
faake
Check
Xeyboard Aad
Coaaaetor
Paeaaatlca
Beplaee
Baadaet
Bad
task
Select
Valve
Or
Coaaeetloa
Dleeoaaeet
laadaat
Aad
Praaa
POBflB
LXBB
Battoa
Proa
Bodala
Depreae
Coaaaetor
Paael
Baar
Oa
Oaa
Box
Proat
Jl
OB
Dleeoaaeet
Boee
Of
Aad
Kaaeare
Veltagea At
Coaaaetor
Valwe
Cable/Coaaeotor
•aaaara
Voltagae At
Bad
J3
J3
Of
HO
0A8
PLOW
KODDLB
BOX
Bepals Or
Beplaee
Cabla/Ceaaeetor
JS
•0
BODOLB
Yee
<
Problea
Xa
Both
OAS
PLOW
BOB
Figure 9.7 No
Praei
legal
Gas
ralvaa
ere
ator
Purge
Beplaee Blgh
Preaaare
Bagalater
Bodala
See
Seetlea
11
48
Page 58
Praia
POBOB
LIBS
Battoa
Paaal
At
Ixamine
Por
eaa
J7
Box
Llaea
Oa
Aad
Proat
Beaeara Voltage
Modale
Obatrecloae
la
Seeare
Wiring
Aad
Coaaeetleaa
Coaaeet
All
Oaa
Llaaa
Problealala
Coatrol/Dleplay
Beard
Baplaea
Coatrol Oa
Bear
Of
Bodale
Figure 9.8 No
Bapalr Aad
Bateet
Valve Paaal
Box
Gas
Flow (Electronics Module)
49
Page 59
Coataet
Valleylab Pos
Partner
Aaalataaca
Calibrate
ayetea
See
Sactioa
Problea
Incorrect
feedback
Plow
CABBOC
ACRIBVB
SLPM
Beter
PLOW
Xe
Proa
If
7
Secare
Cable
Aad
Bateet
Correct
Or
Repair Leake
Probable
Caaaa
Xa
Boa-Paactloaal
Plow
Beter
Figure 9.9 Flow Rate Errors
50
Page 60
(CAKBOt
0 SLPM PLOW I
•OtB-IBIS
CRABS
TSAt
PLOW
CALIXEAtlOS
BXBB
ACRXBVB \
PBBSOHB8
RATI
BAS
DOBS
•eaeare
Voltage At J7
Pla
1
(Coatrol
Valve)
Bapalr
Or
Baplaea
Coatrol/Dleplay
Board
Recalibrate
Syatea
See
Seetloa
fara
Orifice
Adjsetlag
Screw
Oa
Coatrol
Valve
Clockwlae
Beplaee
Coatrol
Valve
Aeeesbly
Adject
Coatrol
valve
Pot
f.OV
7
At2SLBB
See
Seetloa
7
Figure
9.10
Cannot
Achieve
51
0 SPLM Flow
Page 61
(CAXBOt
If
KOtB•(IIS
CBABf
IBAC CALXBBATXOB
ACSIXVX \
SLPB
A
PBBSUBBS
PLOW
BBBV
DOBB
PLOW
BAtB
I
IAS
Repair
Or
Beplaee
Coatrol/Dleplay
Board
Repair Leak
Aad
Beteat
Coaflra/Ad]aat
Coatrol
Valve
For
f.OV
At2SLPB
See
Seetloa
Beaeara Voltage
AtJ7Pla
(Coatrol Valve)
At
Coatrol
Setting 10
Look Caretally
Por
Oaa
Leake
Bctweea
Plow
Beter
Aad
Laer
fitting
Beaeare
Preeaare
of
Impact
Xa
lotoraal filter
Bodale
See
Low
Oaa
Beplaee
Seetloa
Llaee
Bos
At
Bad
Llae
Por
Xa
Beplaee XIgh
Preeeare
Regulator
Bodale
See
Section
Correct
Beteet
11
Aad
Coaaaetor
7
1
Praeeare
Reatrletloaa
Figure 9.11 CannotAchieve 16SPLM Flow
52
Page 62
SECTION
10
TROUBLESHOOTING
This
section
Refer to
describes
the
Force GSU Instruction Manual to learn how a properly operating Unit should function
howtotroubleshoot
THE
HIGH
the
PRESSURE
High
Pressure
REGULATOR
Regulator
Module.
MODULE
and
to
determine what the Unit is not doing correctly. Refer to this chapter and the following chapters in this manual to
help
locate
and
correct
Ifyou
are
not familiar with
Operation," to
proper
disassemblyofthe
Electrical the
parts.
contact
ModuleorTP-1
This control workstation.
Handle
better
Shock
Unitisdisassembled
The
with it;doNOT
Unit
contains
the
circuit
electrostatic sensitive
wmmm
a malfunction.
the
understand
Unit.
Hazard: Do NOT
lower
heatsinkonthe
asameasurement
electrostatic
Wearagrounding
boardsbytheir
components
information in
the
Unitfunctions. Refer to
these
touch
and
energized.
power
use
it for
ground.
ground
sensitive
nonconductive
components.
strap
and
circuit boards.
troubleshooting sections, refer to Section 5, "Theory of
the
WARNING
any
exposed
Disconnect
supply
Use
reference.
CAUTION
when
handling
edges.
Section 11,
wiring or
the
is at an
the
chassisofthe
When
electrostatic
Useanantistatic
power
elevated
repairing
"Replacement
conductive
cord
before
surfaces
replacing
voltage;doNOT
Electronics
the
Unit, work at a
sensitive
container
components.
for
Procedures," for
while
come
Chassis
static
transport
the
in
of
mm
57
Page 63
Oacoaple Qalck
Coaaeet
oa
Blactroalce
Chaaele
Bodale
Bear
Paaal
81gb Preeaare
Bqalpaeat
Leak
Chart
Slectroalci
Chaaele
Leak
Bleetroslce
Chaaele
Leak
Bodale
Chart
Bodale
Chart
Figure 10.1
58
System
Gas
Leak
Page 64
I
BLBCTB0BIC8
V K0D0LB
Blectroalca
Bodale
Opaa Bodale Bos lop
See
Attach
Connect
Seetloa
Bear
CBAS8I8
Qalck
to
Chaaele
Paaal
11
Check Aad tlghtea
All Plttlaga Aad
Claapa After flow
Coatrol
Valve
Bepleee Leaking
Cospoaaat Or
Contact Valleylab
Service
Por
Aeelataaca
Check
All
Plow
Aad
Plttlaga
Coatrol
tlghtea
Op to
Valve
Figure 10.2
Gas
Leak- Electronics
59
Chassis
Module
Page 65
tlghtea
Oaa
tank Plttlaga
(0A8
SXOI
LXAB
PBBS80BB
XQOIPBBBt
\
j
J
Check
Oaa
tank Pitting
Aad
O-Blag
Beplaeed
Daaaged
Coapoaeat
Oaa
taak
Beplaee Blgh
Preeeare
loae
Aeeeably
tlghtea Blgh
Preeeare
Beee
O-Rlag
Preeeare
to
at
llgh Preeeare
Begalator Bodale
Or
Beplaee
llgh
Bodale or Replace
llgh Preeeare
Begalator Bodale
Check And tighten
Plttlaga aad Claapa
oa llgh
Preeeare
Regulator Bodale
Replace Blgh
Preeeare
Begalator
Bodale
Figure
10.3 Gas
Leak-High
60
Pressure Equipment
Page 66
SECTION
11
REPLACEMENT
PROCEDURES
This section describes how to replace most major subassemblies in the Force GSU
leakage
current
Electric
Do
NOT
Electronics
Take
hand
Electric
Unitisdisassembled
The
Unit
test
and
Shock
comeincontact
Hazard:
Chassis
appropriate
rule,"
etc.)
Shock
Hazard: Do NOT
contains
when
electrostatic sensitive
control workstation.
Handle of
the
circuit
electrostatic
boardsbytheir
sensitive
ground
precautions
resistance
The
lower
with
It. Do NOT
ModuleorTP-1asa
(suchasuseofisolated
taking
measurements
touch
and
energized.
Wear
a grounding
nonconductive
components
tests
(Section 6) before returning
WARNING
heatsinkonthe
use
it for
measurement
or
any
exposed
CAUTION
components.
strap
when
handling electrostatic sensitive
edges.
and
circuit
boards.
power
ground.
ground
tools
supply
Use
and
troubleshooting
wiringorconductive
When
repairing
Useanantistatic
the
Unitto clinical
is atanelevated
the
chassis
reference.
equipment;
the
Unit.
the
Unit, work at a static
container
Unit.
Perform the
of
the
useofthe
surface
components.
for
use.
voltage.
while
the
transport
"one
RECOMMENDED
You will
need
the
Description
Slotted
Hex
Nut
DriverorCrescent
REPLACEMENT
following
Screwdriver
Wrench
equipment
Wrench
EQUIPMENT
to perform
the
repairs
Specifications
1
/8"
and1/4"
7/64"
3/8"
described
in this section.
61
Page 67
ELECTRONICS
CHASSIS
MODULE
REPLACEMENT
This procedure describes how to remove the Electronics Chassis Module from the Unitfor further
disassembly
access
or troubleshooting,
the
module
for
some
and
repairs
how to install
and
checks.
the
WARNING
module into
the
Unit. Italso
describes
how to
Electric
Unitisdisassembled
The
Unit
Shock
contains
Hazard: Do NOT
and
touch
energized.
any
CAUTION
exposed
electrostatic sensitive components.
wiring or
mmmmmmmm^^^^MMMmm^mmmmM
When
conductive
repairing
the
Unit, work at a static
surface
control workstation. Wear a grounding strap when handling electrostatic sensitive components.
Handle the circuit boards by their nonconductive
of electrostatic sensitive
Equipment
Slotted
Removing
To
access
1.
Verify
2. Unplug
3. Lock
Required
Screwdriver
the
Electronics
the
Electronics
the
Unit
the
the
cart
Unit power cordfrom
1/8"
power switch is inthe OFF
casters.
components
Chassis
Chassis
Open
the
Module
Module, perform
the
Unit
side
and
circuit boards.
wall receptacle.
door.
(O)
edges.
steps
position.
Use an antistatic container for transport
1-5.
while
the
4. Loosen (turn 1/4 counterclockwise) the four wing screws located on the Electronics rear panel. Gently wiggle the screws to unseat them.
5. Carefully slide the Electronics Chassis Module out from the front of the Unit
table top to work on it. Do not pull itout so far that it strains
6. Disconnect
High
High
Argon
7.
Remove
ESU
Power
the
the
following from
Pressure Pressure
Extension
Cable
Gas
Hose.
module
the
module
Sensor Valve
Cable
(use
from
Cable
Cable
a slotted screwdriver to remove
the
cart.
rear
panel:
the
the
strain relief)
cables
and
and
rest it on the Unitor
gas
line.
Chassis
Module
62
Page 68
Installing
the
Electronics
Chassis
Module
WARNING
Electrical
the
parts. The lower heatsink on
contact with it;doNOT
ModuleorTP-1asa
1.
Place
2.
Reconnect
High
High
Argon
3. Ensure
Unit is
the
ESU
Power
the
Shock
Hazard: Do NOT
disassembled
and
use
measurement
Electronics
the
Pressure
Pressure
Extension
Cable
Gas
Chassis
following to
Sensor
Valve Cable
Cable
(use
a slotted screwdriver to
Hose
the
Cable
Unitfunctions properly.
touch
energized. Disconnect
the
power supply is at an elevated voltage; do NOTcome
it for ground. Use
Moduleinthe
module
ground
rear
panel:
any
exposed
reference.
cart.
attach
wiring or
the
the
chassisofthe
the strain relief)
a. Plug the Unit power cord into a hospital-grade grounded receptacle. b.
Press
the
power switch in
c.
The
Unitconducts a self test. Check
the
ON
(I)
position.
the
following:
* Bothtank indicators illuminate for approximately0.5 seconds.
*
All
other LEDindicators illuminateforthe remainder ofthe self test (approximately 10
seconds).
d. After
the
self
test,
make
sure
that:
* Tank 1 is selected. (IfTank 1 is empty or closed, Tank 2 is selected.)
* A single
*
GAS
tone
sounds.
FLOWissetto1
and
the
Unitisin
the
Gas
Off
* The % PRESSURE indicator displays the appropriate pressure.
conductive
power cord
Electronics
mode.
before
surfaces
replacing
Chassis
while
In
4.
Press
the
Unitpower switch in
the
OFF (O) position.
5. Unplugthe Unit powercord from the wall receptacle.
6. Carefully slide
NOTE:
When
7. Pressthe module towards the rear ofthe
the
module into
inserting the module,
the
orient
front of
the
the
Unit.
wing
screws
Unit.
vertically.
Fasten the module into place by tightening (1/4 turn
clockwise) the four wing screws on the module rear panel untilsnug.
8.
Close
the
Unit
side
door.
63
Page 69
HIGH
PRESSURE
REGULATOR
MODULE
REPLACEMENT
Thisprocedure describes howto removethe
disassembly or troubleshooting,
Electric
Shock
Hazard: Do NOT
Unitisdisassembled
The Unitcontains electrostatic sensitive components. When repairing the
and
and
how to installthe module into the Unit.
energized.
touch
High
Pressure Regulator
WARNING
any
exposed
CAUTION
wiring or
Module
from
conductive
Unit,
the
Unit
surface
work at a static
control workstation. Weara grounding strapwhen handling electrostatic sensitive components.
Handle
of electrostatic sensitive
Equipment
Hex
Wrench
Nut
DriverorCrescent
Removing
1. Purge
the
circuit
Required
7/64"
the
High
the
Unitof Argon
boards
Wrench
Pressure
by their nonconductive
components
3/8"
Regulator
gas
before disconnecting
and
circuit
Module
edges.
boards.
the
gas
Use an antistatic containerfor transport
lines. (Referto "Purging
the
procedure withinthis section forfurther information).
forfurther
while
the
Gas
Lines"
Press
the
2.
Unitpower switch in
the
OFF (O) position.
3. Unplug the Unitpower cord from the wall receptacle.
4. Disconnect
a. Follow
the
gas
tank
lines from
the
appropriate safety precautionsfor handling of high
the
Unit.
safety cap(s) before proceeding.
b. Lock the cart casters. Open c. A small amount of Argon may
high
pressure
the
large knob. Use a wrench to
gas
d. Loosen but do not remove
5.
e. Attach
Remove
a.
Open
the
the
the
safety
Unit
storage
storage
b. Remove the plastic
c. Using a hex wrench,
d.
Remove clears
Pressure
6. Disconnect
the
storage
the
screwsatthe
Regulator Module is on bottom of
the
following from
line, hold the
captothe
compartment.
compartment.
caps
tum
compartment by grasping
the
side door.
escape
the
retaining strap.
top of
during this procedure. To disconnect
gas
line coupling stationary
assist
loosening this connection.
the
tank.
from the screws in the bottom corners of
the
two hex
bottom. Slide
the
High
screws
the
Pressure
clockwise
compartment bottomforward
the
Regulator Module:
and
(screws
the
handle
cart. Removing
and
pressure
loosen
the
inward).
lifting
the
the
Module
gas
tanks. Obtain tank
the
tankfrom
the
coupling by turning
compartment.
compartment and
lower it.
until it
The
the
High
Tank1and
Tank1and
Tank1and
Tank
2 High
Tank2select
Tank2pressure
Pressure
cables
transducer
Hoses
cables
64
Page 70
Some
7.
residual
Using
Save
a nut
the
gas
may
driver,
hardware.
escape.
loosenthe fourhex nutsand lockwashers locatedon the
8. Carefullyremove the plate and hose assembly fromthe bottomof the cart.
high
pressure plate.
Installing
1.
Place
the
the
High
High
Pressure
Pressure
Regulator
Module
Regulator Module in
the
cart.
2. Using a nut driver, attach the module to the cart using the four hex nuts and lockwashers.
3.
Reconnect
Tank 1
4.
Connect
a. Remove
b.
c. d. Align
e.
5. Plug
6.
Press
7. After
the
following from
Tank1and
Tank1and
and
the
gas
the
NOT remove
Orient
the
tank
Secure
the
Close
the
the
the tank with
the
gas
large knob
the
side
Unit
power
the
power switch in
Unit self
the
High
Tank
2 High
Tank2select
Pressure
cables
Pressure
Hoses
Regulator Module:
Tank 2 pressure transducercables.
tank
lines to
gas
tank
the
gas
valve
line coupling
hand
door.
cord
test,
open
the
Unit.
safety
cap
tank
identification tag(s).
outlet
toward
the
retaining strap.
the
tank valve outlet. Carefully
tight.
into a
hospital-grade
the
ON
(I)
both Argon
and
place
the
backofthe
position.
gas
itin
grounded
tanks.
the
gas
cart.
cap
holder in
connect
receptacle.
the
baseofthe
the
coupling by tightening
cart. Do
8. Listen for
gas
Troubleshooting
9. If
there
arenogas
a. Catch
b.
Lift
and
c. Lower
the
the
d. Using a hex wrench,
storage
e.
Place
the
leaks. If
the
there
High
is a
gas
Pressure
leaks, replace
storage
tilt
the
storage
compartmenttosecure
plastic
compartmentflange under
lower part
the
compartment over
back
out
caps
over
the
leak, check all
the
fittings
and
refer to Section 10,
Regulator Module,"for further information.
the
storage
storage
the
the
screws
compartment in
the
compartment into
the
screws
two hex
screwsinthe
compartment in
and
close
the
Unit.
top frame of
the
the
Unitopening.
protruding from
side
frames
the
Unit.
the
storage
compartment.
the
Unitopening.
Unit
side
from
the
frames.
inside of
the
65
Page 71
PURGING
THE
GAS
LINES
Thisprocedure describes howto purge both
Argon
gas lines.
disconnecting them.
1. Ensure that both Argon
2.
Turn
off
both
tanks.
3. Press the power switch inthe ON
Tankl.
4.
Press
the
GAS
FLOW
5.
Press
the
PURGE
gas
tanks have a pressure greaterthan 5%.
(I)
UP
buttontoset
LINE
button to purge Tank 1
position. Afterthe self test, the
the
flow
rateto10.
gas
line.
6. Open Tank 2. The pressure should rise above 5% and the
off
Tank
2.
7. WithTank 2 selected,
8.
The
Tank 1
CONTROL
This procedure
and
DISPLAY
describes
press
the PURGE
LINE
button to purge the Tank 2
Tank 2 % PRESSURE indicators should display 0%.
BOARD
how to remove
REPLACEMENT
and
installthe Control Display Board.
calibrated after replacing the Control Display Board.
Always
purge both gas linesbefore
Unit
automatically selects
Unit
automatically selects Tank 2. Tum
gas
linecompletely.
The
Unitmust
be
WARNING
Electric
Unitisdisassembled
The
Shock
Hazard: Do NOT
and
energized.
touch
any
exposed
CAUTION
wiringorconductive
Unitcontains electrostatic sensitive components. When repairing
surface
the
Unit,work at a static
control workstation. Wear a grounding strap when handling electrostatic sensitive components.
Handle the circuitboards bytheir nonconductive edges. Use an antistatic container fortransport
of electrostaticsensitive components and circuitboards.
Equipment
Slotted
Nut
DriverorCrescent
Removing
1.
Verify
2. Unplug
3. Remove
Required
Screwdriver
the
1/8"
Control
Wrench
Display
1/4"
Board
the Unitpower switch is inthe OFF (O) position.
the
Unitpower cord from the wall receptacle.
the
Electronics
Chassis
Module from
the
Unit. (Refer to
the
"Electronics
Chassis
Replacement" procedure withinthis section for further information.)
while
Module
the
66
Page 72
4.
Locate
the
Control
Display
Board.
Disconnect
the
following
cables
from
the
board:
Keyboard
JackstripCable (3 pinconnector)
Tank
Internal
DC Pressure Sensor Cable (8pinconnector)
Flow
Cable(7 pinconnector)
Valve
PowerCable (5 pinconnector)
Interface
Cable
from(4pin
Power Cable (6pinconnector)
Control
Valve
Wires(2 pinconnector)
from
J1
from
J2
connector)
from
J5
from
from
from
J3
from
J6
J7
J4
ReliefValve Wires (3 pinconnector) from J8
Flow
MeterCable (4 pinconnector) fromJ9
5.
Remove
hardware.
the six
nuts,
two
screws,and
lockwashers
from
the
Control
Display
Board.
Savethe
6. Carefullyremove the Control DisplayBoard fromthe module.
Installing
the
Control Display Board
1. Carefullyplace the Control DisplayBoard inthe module.
2. Secure the Control DisplayBoard to the module using six nuts, two screws, and lockwashers.
3. Reconnect
the
following
Keyboard Cable (7 pin connector) to J1 Jackstrip Cable (3 pin connector) to
Tank Valve Power Cable (5 pin connector) to
Internal Interface Cable from (4 pin connector) to DC Power Cable (6 pin connector) to
Pressure
Flow Control Valve Wires (2 pin connector) to
Relief Valve Wires (3 pin connector) to
Flow Meter Cable (4 pin connector) to
Sensor
cables
Cable
to the Control DisplayBoard:
(8 pin connector) to
J2
J3
J5
J6
J8
J9
J4
J7
NOTE:
4.
Place
5.
Reconnect
6.
Ensure
Make
sure the
the
Electronics
High
Pressure
High Pressure Valve Cable
ESU
Extension
Power
Argon
the
a. Plug the
b.
Press
c.
The
Unit
Keyboard
the
following to
Sensor
Cable
Cable
Gas
Hose.
Unit functions properly.
Unit
power cord into a hospital-grade grounded receptacle.
the
power switch in
conducts
Cabledoes notget
Chassis
the
Cable
a self
Moduleinthe
module
the
test.
rear
ON
(I)
Check
trapped
cart.
panel:
position.
the
following:
behindthe
* Bothtank indicators illuminate for approximately0.5 seconds.
*
All
other LEDindicators illuminatefor the remainder
seconds).
67
the
Control
self
test
Display
(approximately 10
Board.
Page 73
d. After
the
self
test,
make
sure
that:
* Tank 1 is selected.
* A single tone sounds.
*
GAS
FLOWissetto1
* The % PRESSURE indicator displays the appropriate pressure.
(If
Tank 1 is emptyor closed, Tank 2 is selected.)
and
the
Unit is in
the
Gas
Off
mode.
Open both Argon
7.
Press
the
8.
Unplug the Unit power cord from the wall receptacle.
9.
10. Calibrate
11. Install
Replacement" procedure withinthis section forfurther information.)
POWER
This procedure
Electric
Unit Is
The
control
Handle
of electrostatic sensitivecomponents and circuit boards.
Unit
the
the
Electronics
SUPPLY
describes
Shock
disassembled
Unit
contains electrostaticsensitivecomponents. When repairing the
workstation.
the
gas
tanks, and check forleaks.
power switch inthe OFF
Unit. (Referto
REPLACEMENT
how to remove
Hazard: Do NOT
Weara
circuit
boardsbytheir
Section
Chassis
and
energized.
grounding
(0)
position.
7, "Maintenance
Module in the Unit. (Referto the "Electronics
and
replace the power supply.
WARNING
touch
any
exposed
CAUTION
strapwhen
nonconductive
Procedures,"
handling
edges.
for further information.)
Chassis
wiring or
conductive
Unit,
surface
workat a static
electrostatic sensitivecomponents.
Useanantistatic
container
for
Module
while
transport
the
Equipment
Slotted
DVM
Removing
1.
Verify
2.
Unplug
Remove Replacement" procedure
Required
Screwdriver
the
Power
the
Unit
the
Unit
the
1/8"
Supply
power
switch
powercord
Electronics
is inthe OFF
from
Chassis
within
(O)
the
wall
receptacle.
Module
from
thissectionfor
position.
the
Unit.
further
(Referto the "Electronics Chassis
information.)
Usinga slotted screwdriver, remove the four screws and lockwashers from the module rear cover.
Carefully
the
Locate
Power Supply Cable from J1
Power Supply Ground Cable from P1
module.
the
DC
Power
remove
power
Cable
the
rear
supply.
from
cover.Donot
Disconnect
J2.
the
pullitoutsofar
following
cables
68
thatitstrains
from
the
power
the
cables
supply:
andgas
Module
line
inside
Page 74
6.
Using
a slotted
screwdriver,
remove
supply. Save the hardware and cable clamps.
POWER
SWITCH
the three
metal
screwsand one
DC
CABLE
POWER
nylon
screw
CABLE
from
the
CLAMP
ACTOPOWER
SWITCH.
power
ACTO
SUPPLY
POWER
CABLE &
NYLON
CLAMP
SCREW
POWER
FUSE
SUPPLY
Figure 11.1 PowerSupply Replacement
7. Remove
Installing
the
the
power
Power
supply from
Supply
the
module.
1. Orient the power supply so that fuse is at the bottomof the module, and the J2 connector is at the
top ofthe module. Using a slotted screwdriver,attach the power supplyto the ElectronicsChassis
Moduleusing three metal screws and one nylonscrewasshown in Figure 11.1.
screw
and
cable clamps placementbefore attaching them to the module.
Verify
the nylon
2.
Reconnect
the
following
cablestothe
power supply:
Power Supply Cable to J1
Power Supply Ground Cableto P1
DC
Power
3.
Place
the
CabletoJ2
Electronics
Chassis
Moduleinthe
cart.
4. Using a slotted screwdriver, attach the module rear cover using four screws and lockwashers.
5. Reconnect
6. Use a
switching power supply voltages at
High
High
ESU
Power
Argon
DVM
the
followingfrom
Pressure
Pressure
Extension
Cable
Gas
Hose
Sensor
Valve Cable
Cable
the
Cable
module
rear
panel:
withfine point test leads orclip leads and make the
TP1-4
on the Control Display Board.
69
following
measurements ofthe
Page 75
Test
Point
TP-1
TP-2 TP-3
TP-4
If
the power supplyvoltages are notcorrect, adjustthem per the Calibration Procedure inSection 7.
7.
Ensure
the
Unitfunctions properly.
Voltage
Ground
+5.0
Vdc±0.25V
+12
Vdc±0.6V
-12
Vdc±0.6V
a. Plug the Unitpower cord into a hospital-grade grounded receptacle.
b. Press the power switch inthe ON
c.
The
Unit
* Both
*
All
seconds).
conducts
tank
other LEDindicators illuminate for
a self test.
indicators illuminate for approximately
(I)
Check
position.
the
following:
the
0.5
remainderof
seconds.
the
self
test
(approximately 10
* The % PRESSURE indicator displays the appropriate pressure.
d. After
the
self
test,
make
sure
that:
* Tank 1 is selected. (IfTank 1 is empty or closed, Tank 2 is selected.) * A single
*
GAS
Press
the
8.
Unit powerswitch in
tone
sounds.
FLOWissetto1
and
the
the
OFF
Unitisin
the
(O) position.
Gas
Off
mode.
Unplug
9.
10.
Installthe Electronics Chassis Module inthe
the
Unitpower cordfrom the wall receptacle.
Unit.
(Refer to the "Electronics Chassis Module
Replacement" procedure withinthis section for further information.)
FUSE
REPLACEMENT
This procedure describes howto replace the fuse on the power supply.
WARNING
Electric
Unit Is
Shock
Hazard: Do NOT
disassembled
and
touch
energized.
any
exposed
wiring or
conductive
Fuse Replacement: Forcontinued protection against fire hazard, replace only with
type
and
ratingasthe
The
Unit
contains electrostatic sensitive components. When repairing the
control workstation. Wear a grounding strap when
Handlethe
of electrostatic sensitive
circuit
original
fuse.
CAUTION
Unit,
handling
electrostaticsensitive components.
boards bytheir nonconductive edges. Use an antistaticcontainer fortransport
components
and
circuit boards.
surface
while
the
work at a static
the
same
Equipment
Slotted
DVM
Screwdriver
Required
1/8"
70
Page 76
I.
Verify
the Unitpower switch is in the OFF (O)position.
Unplug
3.
Access
Replacement"
4.
Using
Save
cables
5. Locate the fuse on the power supply. Remove the fuse
Use a
switching power supplyvoltages at
If
the power supply voltages are not correct, adjust them per the Calibration Procedure in Section 7.
7.
Usingaslotted
8. Ensure
the
Unit
power
cord
from
the
Electronics
procedure
slotted
the
DVM
screwdriver,
hardware. Carefully remove
and
tubes
withfine point
Test
Point
Chassis
inside
the
within
remove
module.
test
TP-1
TP-2
TP-3 TP-4
screwdriver,
the
Unitfunctions properly.
the
wall
receptacle.
Module from
this
section
the
four
the
leads or clip leads and make the following measurements of the
TP1-4
Voltage
the
Unit. (Refer to
for
further
screws
rear cover. Do not pull itoutsofar that it disconnectsthe
on the Control Display Board.
information.)
and
lockwashers
and
the
"Electronics
from
replace it with
the
the
Chassis
module
same
rear
type
Ground
+5.0
Vdc±0.25V
+12
Vdc±0.6V
-12
Vdc±0.6V
attach
the
rear
covertothe
module
with
four
screws
and
lockwashers.
Module
cover.
and
rating.
a. Plug the
b.
Press
c.
The
* Both tank indicators illuminatefor approximately 0.5 seconds.
*
d. After
* Tank 1 is selected.
* A single
*
*
9. Press the
10. Unplugthe Unitpower cord from the wall receptacle. II.
Installthe Electronics Chassis Module inthe
Replacement"procedure
Unit
power cord intoa hospital-grade grounded receptacle.
the
power switch in
Unit
conducts
All
other
seconds).
the
self
GAS
FLOWissetto1
The
% PRESSURE indicator displays
Unit
power switch inthe OFF (O) position.
a self
LED indicators illuminate forthe remainder of the self
test,
make
tone
sounds.
the
ON
test.
Check
sure
that:
(If
Tank 1 is empty or closed, Tank 2 is selected.)
and
the
within
this sectionforfurther
(I)
position.
the
Unit is in
following:
the
the
Unit.
test
(approximately 10
Gas
Off
mode.
appropriate pressure.
(Refer to the "Electronics Chassis Module
information.)
71
Page 77
COMPONENT
REPLACEMENT
GUIDELINES
WARNING
Electric
Unit Is
The control workstation.
Shock
disassembled
Unit
contains
Hazard: Do NOT
and
electrostatic sensitive components. When repairing
Wear
a grounding
touch
energized.
any
strap
exposed
CAUTION
when
wiringorconductive
handling electrostatic sensitive
surface
the
Unit, work at a static
components.
while
Handle the circuit boards by their nonconductive edges. Use an antistatic container for transport
of electrostatic sensitive
The following
are
guidelines that relate to installingand mounting components on
1. Install all non-polarized componentsso that the value, tolerance,
2. Install polarized components so that polarization markings
3. Components with coating on
lead enters the component body. For example,
4. Center the component body between
components
are
visible andthe correct polarity.
the
and
circuit boards.
the
positive symbol (+), negative symbol (-), or other
leads may have the coating removed only to
bends
(except for miniature diodes, it is desirable to leave the
the
circuit boards.
and
part number
use
a soldering ironto remove coating on the leads.
are
visible.
the
point where that
cathode lead longer than the anode lead for identification). Lineup the leads, where required, with
the
mounting holes prior to installation.
the
5. Position lead-mounted components so that the major axis of the component is parallel to any two of
the three major planes (sides) ofthe
excess
of 1/2
ouncebymeans
other
Unit.
Mechanically support any component having a weight in
than
the
leads.
6. Properly insulate ailcomponent leads whichare, or could be, forced into a shorting conditionwith
another component lead or circuit.
7.
Do
notstress component leads between
damagetothe
component and solderpins.
mounting
pins.
Provide
an adequate strain
relief
to prevent
8. When installing components on circuitboards withthe circuitry on the component side of the board,
insulate
9.
Mount
10.
Mount
circuit
11. Radial lead capacitors
the
potting
the
sleeving of
resistors rated at 2watts or less
the
components withmetallic bodies.
flush
tothe
circuit
resistorsofgreater than 2 wattsor more0.25± 0.06 inch
board.
(dip):ifinstallation requires bending the lead, the bend should be no closer to
body
than
0.06 inch. Mounting height shouldbe0.032 - 0.25 inch from the board surfaceto the
material.
board (0.0- 0.006 inchclearance).
from
the surface ofthe printed
12. Soldered surfaces should be clean and free ofcontaminants that would result inpoor soldering.
13. Controlsoldering temperature to prevent damage to components or circuitry.
14. Remove flux residue on the board surface. For example,
remove
the
flux
residue.
use
isopropyl alcohol
and
a brush to
72
Page 78
SECTION
12
FACTORY
SERVICE
RETURNING THE EQUIPMENT FOR
Before Return
information ready:
Attach
you
return
the
UnittoValleylab,
Authorization
Hospital/clinic
Technician's
Telephone number
Department/address
City,state, and ZIP code
Model
Serial
Description of problem
Type of repair tobedone
a tag
number
number
with
Number,orcall
name/customer
name
thissame
information
call
your
Valleylab
number
tothe
SERVICE
the
Valleylab
Unit
Service
representative
when
shipping
Department
for
assistance.
itforservice.
(1-800-255-8522)
Have
the
following
for
a
DoNOTretum
Retuming
Packagethe
available, use the
Place
Use a packing container with double wall construction.
Placeaplastic
Use non-abrasive packing materialthat
Use
Retuming
When
packaging
Place each
Providea separate packingcontainer for each
Whenreturning the Electronics Chassis
leastfourinchesof
including
Ordering
gas
tanks to Valleylab.
the
Force
Unit
the
GSU
inthe
following
interface
bag
four
crossed
Circuit
Replacement
circuit
PCBA
the
Boards
bottom.
Unit
original
strapstosecure
boardsor othersubassemblies,use the
or subassembly inan ESD(Electrostatic Discharge) bag or container.
Parts
packaging
guidelinesforpackaging:
cableinthe
over
the
Unit.
and
Other
tightly
packed
container,ifavailable.Ifthe
Unit
storage
will
the
box.
Subassemblies
Module
industry
compartment.
notdamage the surface of the
approved shock absorbing
PCBA
and/orthe
original
following
or subassembly.
High
guidelines:
Pressure Regulator
packagingisnot
Unit.
packing
materialonall
Module,
use at
sides
When ordering replacement parts forthe
Model Number (located on
Serial Number (located on the rear ofthe unit)
Part Number (referto Section 13, "Service Parts
the
Unit,
rear ofthe unit)
includethis information:
List")
73
Page 79
SECTION
13
SERVICE
When ordering replacement parts
All
componentsmust be
and
tolerances
PARTS
Model
Serial
mustbe
Number:
Number
replaced
equalorbetter
LIST
for
locatedonthe
locatedonthe
with
the
Unit,
include the
rearofthe
rearofthe
partsofidentical
than
original.
following
Unit
Unit
construction
Unit
performance
information:
and
value.
canbe
Replacement
adversely
substitution of Iower grade parts.
AllItems referenced inthis Section are foundin Section 14, Assemblies and Schematics.
FORCE
REFERENCE
DESIGNATOR
Iteml
Item
Item
Item Item
Item Item
Item
Item
Item
Item Item
Item
Item
Not
Not Not Not
2
3
4 5
6 7
8
9
10
11 12
13
14
Shown
Shown Shown Shown
GSU
UNIT
DESCRIPTION
Gas Coag Cart
Electronic
Chassis
Module
High Pressure Regulator Module
High
Pressure
TankIDLabel
Inside
Door
Gas
LineIDLabel
High Pressure
ESU
Interface
Hose
Label
Sense
Cable
Cable
Power Receptacle Cable
Cord
Retainer
Bracket
High Pressure Valve Cable
4-pin Plug Connector
#18 Cable Clamp
Power
Cord
ArgonGas Tank (not supplied)
Active Adapter
Force
GSU
Force
Unit Instruction
GSU
Unit
Service
Manual
Manual
Mufti-Language Instruction Cards
VALLEYLAB
PART
207
202
202 202 216101384 216101385
216101
202
202
202
223
202
208 213150
202
GT100
E0502-12
945110
945100141
225
part
affected by
NUMBER
500122
701
594
701
593
701
522
389
701
586
400
844
400
891
301
108
701
591
300
648 049
400
846
060
500 528
ratings
74
Page 80
ELECTRONICS
CHASSIS
MODULE
202
701
594
REFERENCE
DESIGNATOR
Item
1
Item
2
Item
3
Item
4
Item
5
Item
6
Item
7
Item
8
Item
9
Item
10
Item
11
Item
12
Item
13
Item
14
Item
15
Item
16
Item
17
Item
18
Item
19
Item
20
Item
21
Item
22
Item
23
Item
24
DESCRIPTION
Module
Box
Rear Panel Assembly
Top
Cover
0.25
Cross
Barb
Tube
Fit
0.25
ID,
75 psi, PVC Tube
Female
3/16
Barb
Luer
Lock
0.43 OD Nylon Hose Clamp
Power
ID
Switch
Tube Bulkhead
Fitting
Power Supply Cable
Power Supply Assembly
DC
Power
Cable
Cable Clamp
Power Supply Ground Cable
Fuse 2 amp,
250
Vac
Control/ Display Board
Front
Panel Assembly
Internal
Chassis
Luer
Nylon
Standoff, Standoff,
Jackstrip
Banana
Power
Ground
Lock,
Spacer
Cable
Cable
Male
3/16
M/F
6-32SS
M/F
6-32SS
Assembly
Jack,
white
barb
5/16 7/16
Jackstrip
VALLEYLAB
PART
NUMBER
223
200
202
701
223
400
243
500122
249
101
223
100
213150 243
075 000
243
500
202
400
207
000151
202
400
213150
202
400
215
005
201
285
207
500135
202
701
202
400
223
100
213
110
213
110513
213
110514
202
701511
208
400
.223
100
489 587 528
494
717
055
075
886
895
015
900 017
002
497
901
716
448
063
695
75
Page 81
REAR
PANEL
ASSEMBLY
202
701
587
REFERENCE DESIGNATOR
Iteml
Item
2
Item
3
Item
4
Item
5
Item
6
Item
7
Item
8
Item
9
Item
10
Item
11
Item
12
Item
13
Item
14
Item
15
DESCRIPTION
Hose Clamp, 0.43 OD, Nylon HPSensor Cable Assembly
Filter
Assembly .
Solenoid
Valve
ESU Cable Assembly
Circular
Chassis
Flow
Mass
Flow
Tubing 0.25
HP
Chassis
1/4
Rear
Connector, 4 pin
Ground
Control
Flow
Meter
Meter
Cable
ID,
Valve
Internal
Connector, 4 wire
NPT
Hose
Panel
Cable
Valve
75 psi, PVC
Cable
Barb
Hose Panel Coupling
Internal
Power
Cable
Power Receptacle
VALLEYLAB
PART
213150
202
.^202
202
202
208
202
202
245
202
249
202 208
243
223 200
243
202 208
NUMBER
055
701
585
701556
701
547
701
545
300
646
400
901
701
525
500
087
400
912
101
494
701
592
071
000
500
077
488
500
081
701
497
300
669
(MCp-S^
76
Page 82
CONTROL
DISPLAY
BOARD
201
285
002
REFERENCE DESIGNATOR
Resistors
R1.15,16,19,22,23,58,59,86
R25-27, 37,39,40, 41. 72-76. 80, 81
R7,
8,9,10,11,12
R13,14
R17
R18,85
R20.77
R28
R29 R30
R31.33, 35, 43, 45, 47, 49. 61, 70
R32.
34,
60
R36,51,56
R38
R44,
46,
48,
50
R53
R54 R57
R63,
65,
67
R62, 71
R64.66 R68.79
R82
R83.84
R87,
88
RA1.4
RA2 RA3
RA5
DESCRIPTION
1
kohm±5%,
10kohm
51
ohm
±5%,
10kohm
4.3kohm 3 k
ohm
±5%,
5.1
kohm±5%,
51
kohm±5%,
11k
ohm
240
ohm
10kohm±1%,
100kohm
Trimpot 50 k
1 Meg
340k±1%,
32.4kohm
5.11
68
Pot10k
±5%.
ohm±1%,
ohm
±5%,
±20%
71.5kohm
100kohm
330kohm
±5%,
±5%,
±5%,
±5%,
±5%,
±1%,
1/8W
±1%,
±1%,
±5%,
±5%,
1/4W
1/4W
1/2W
1/2W
1/2W
1/4W
1/4W
1/4W
1/4W
1/4W
1/8W
1/8W
ohm
1/4W
1/8W
1/8W
1/2W
1/8W
1/4W
1/4W
15ohm±5%,3W
49.9kohm
33kohm
±1%
±5%,
1/4W
Resistor Array. 1 k ohm, 9x Resistor Array, 10 k ohm, 8x
Resistor Array, 10 k ohm, 7x
Resistor Array, 1 k ohm, 8x
VALLEYLAB
PART
234 234
234
234
234
234
234
234 234
234
234
236
234
234 234
234
234 236 234 234 234
234
234 234
234
234100148 234 234100147
234
NUMBER
024 024
014
014 068
024 078
024
024 080
024104
024 024
201
201
200
024135
201 201
201
014 200 201 024111 024123
400
201 024
100134
100
063 087
083
074
088 048
385
481
038
532 434
357
032 036 467
252
452 099
098
Capacitors
C1 -
9,14-18,
C12
26, 35 - 47. 50, 51
C13, 22, 23, 27, 28, 30
C20,31,32,33
C21
C29
C34
C48
C49
CA1
0.1
ui±
20%,
50V
100pf±5%,
100V
1.0uf±20%,50V
10ut
±20%,
0.0047
0.027
ui±
22ui±20%,
2200
pf±
4.7uf±20%,
uf±
10%,
5%,
10%,
35V
100V
50V
25-35V
100
10V
Capacitor Array0.01
77
204
118
007
204
200
013
204118
204104
204
204
204102
V
u,f,
8x
204 200 204 204
014
048
200
033
400101
030
029
074
001
200
241
Page 83
Integrated
Circuits
U2 U3
U4
U5
U6,7
U8.20
U9 U10
U11 U14
U15 U16
U17, 21
U18 U19
U22 U23 U24
Diodes
CR1.3-9
CR2,10-13
DS1232 74LS374 74LS240
ADC0809
DAC0832 LM324N 74HCT04
4013B
74HCT02
LM336-5.0V
ICM7218
2803A
LM339A
40106B
Pressure
LM350
TL084
74HCT32
1N4148 1N5240B
XDCR30PSIG
210
760
210
520
210
520
210
740
210
750
210
022 000
210
240
210
027
210
240
210
300
210
700
210
800
210
300
210
211
209
001
210
300
210
200
210 240
239
014 000
239
600
015
374
240
000
001
004 001
002
018 001 002 015 106
000 086 040
032
001
Transistors
Q4-7,10,11,13,15-18,
Q8.9,12
Q14 Q17
Transformers
T1
LEDS
D1-8
D9,10
D11
D12,13,14
Miscellaneous
OP1.2.3
Y1
L1
K1.2
VN10KM
MPS-U45
2N2907A MTD
2955-1
HV
Toroid
Isobloc
LEDDisplay HDSP7501 LEDArray HLMP2450 LEDArray HLMP2550
LEDArray HLMP2350
Photo
Isolator
XTAL
Osc.
RF
Choke
Relay Dip
3.0
55
Reed
uH
OPI1264B
MHz
SPSTNO
Speaker8ohm Speaker,8ohm
12V
239 200
239100 239100
239 200
251
300
239
750
239
750
239
750
239
750
239
750
250
100
251 100
230
010
241
003
241
003
012
002 012
036
013
054
059
057
058
019
013
098
000
001
000
78
Page 84
SECTION
14
ASSEMBLIES
See
Section13"Service
AND
Parts
SCHEMATICS
List"toreference
corresponding
Item
Numbers
and
Reference
Designators.
Figure
14.1
Force GSU Unit
79
(not supplied with
Unit)
Page 85
Rgure 14.1 (continued)
80
Page 86
Figure 14.2 Electronics Chassis Module
81
Page 87
Figure 14.2 (continued)
82
Page 88
Figure 14.4 Rear Panel
83
Assembly
Page 89
Figure 14.5 High Pressure
84
Assembly
Page 90
SB aa
ffl
SJ]
rV
cmC
$ 1
4 1
cN
3]
123.
s&
£
0
UU
UiU
fl ol
SB]
' P
i=7
O
"s
s>
%
3U
eaa
XlC3§
U
Figure 14.6 Control Display Board
^B—I
j\e
3 a
rjo!
«0!
J]
oi
TJ
cS?
85
Page 91
&1
H 2
.4
«o
<OOQUJU.OI
oooooooo
Ul
UJUJU
tutu
III
O
no
3w
<4<QQQQQ
QQQMMMMH
OOOGOQOQ
<<<<<<<<
toinr*
0)0)
0X1)0)0}
*N
0)0)
>0)(0
tu
FROM
MICROCONTROLLER
r—
CIRCUIT
|
Figure 14.7 Control Display Board Circuit
86
Schematics
DISPLAY
CIRCUIT
Page 92
DPI
\
/
+
12VO—f-
ULN2803
dps
^
•kJIBEJO*2
op49U1g
_BQ2_
8AS
09A 08B
lLJg
*vr
D10A
*vi^SvH5vH**r%
011A
1 LJ 2 3 LJ 4 S y 6 7
i
*vr^SvH5vll
"sS>S*-
012A
*vl!
013A
Mg
*v
014A
W2ay4sy6
3 y 4 S LJ
HLMP24S0
8AS
0108
HLMP2450
8AS
01IB
HLMP25B0
OCCLUSION
012B
5vTJ*vr*5vIT
KLMP23S0
TANK1LOW
013B
3 y 4 s y 6 7 y
«vH5vri*vn
KLMP2350
TANK2LOW
014B
3vn
HLMP23B0
ON
OFF
FLOW
010C
011C
012C
013C
014C
?fH
D9C
67
7y
09D
01.00
0110
5vN
01.20
0130
01.40
5*rT
LJ 8
li
R78
10K
B
ii
TNK2EN
ltd
jn
R13
10
Qi6
vni<
VN10KM
Rgure
013
TANK1SEL
<a
016
TANK2SEL
(a
14.7 (continued)
4ll2V
DISPLAY
9
CIRCUIT
(cont'd)
—'
87
Page 93
Figure
14.7 (continued)
MICROCONTROLLER
CIRCUIT
J
f^
^
(S:,
fS
Page 94
Hii-
UI
MO«V{0««<0
JI
000000000)0.
HXXQQQQQOQQUX
ml«
am*noim
««<«<<«•«<
3
a
inn<ios
tor- * n
ui
bl
N(oio«n(u«40
JE
00000000(1)0.
HXXQQOQQQOQUX
u
3.3
Hi"
<<<<<*
Figure
14.7
89
(continued)
MICROCONTROLLER
CIRCUIT
(cont'd)
—'
Page 95
(CW)
R67
10K
R32
100K
-J7
R3S
10.
OK
(CW)
R80
100K
+12VFILn36S0K
0/
"X
-—
L_£
_L_ •»
J20A>—i-
LH324
C50
1 I
JC
,
DLPRES
CR12
1N3240B
+12VSW
TNK1EN
TNKPBN
YAWVEN
BSIA.
OC
POWER
US.
1
2 5-
3
5 >
6 >
O-
CRS
MPSU4S
R74
^r-LM
.
>-—+•
P-
X7
•^•1£
CRS
MPSU45
CR9.
KPSU4S
rl00rl
R72
TANK
I
MTD29S5-1
I
MTD2
VN10KM
+12V
C32
*%=•
R73
10K
VALVE/RELIEF VALVE —•
-12V
12V
♦sr
•P6NC
ASNC
-18V
C33
10
-S
10K
-i8TT±rol7-°+12Vaw
10K
10K
It,018
)P]
+SV
C3lU
012
TANK
RELIEF
R82
IS
LI
SSuH
VALVES
JJL
< S
ja.
-e
1
-c
2
< 3
VALVE
C34 |»
22
RS8
+12VFIL
IK
5f
1 1
U14
LM338-S.0V
K?
12VE1L.
♦8VREF
^n7
DELIVERY
PRESSURE
TO
ESU
GENERATOR KEYING -1
+SV
MT8
HOLE
ti.
C3sf
-a:
036?
-in
C37I
.3:
POWER
SUPPLIES —•
Figure 14.7 (continued)
90
-12VFIL
C29
.027
RS2
1
10.
TQNEfV
£
OK
IE06
AS.
$?£c
/
10.OK
4
RS3
—<\/S/*
RS4
use;
32.4K
LM324
S.11K
•n
•12V
O
RS7
68
*—
r=0
C30
AUDIO —J
SPl
Page 96
(CW)
10K
R44
340K
R61
r—•^•V—f
r$?
R70
I
"Vv*
T^
10.OKR62
_10_
10.
OK
R71
»
I 8
>W*
"VV*
71.SK
U20
71.
J20B
LM324
SK
LM324
*
RB6SOK
8
1NS240B
"
RSI
SOK
• I
JS
K7
CR10
(CW)
T1PBE3
!3B
\7
TgPRES
CR11
1NS240E
PENCIL
TANK
PRESSURE
♦BV
Rgure
HANDSWITCHING
14.7
(continued)
91
ISOBLOC
CIRCUIT
Page 97
♦12VFIL
4
P V FLOW
3
P-
2
1
F^n-iLf^
•=•
VCS3
FLOWMETER
.1
LM339
R41
10K
+SV
i
POSERR
SASgN/
12V
x—2_Jvi
LM3S0
VO
Iaqi°uaa
74HCT04
ELQ&QD.
10K
^g)H"-l^°S
74HCT32
CR4
iJZ_
C 1 < 2
CONTROL
VALVE
Figure
14.7 (continued)
92
GAS
FLOW&ERROR
LIMITS
Page 98
Loading...