This manual and the equipment it describes are for use only by qualified
medical professionals trained in the particular technique and surgical
procedure to be performed. It is intended as a guide for servicing the
Valleylab Force FX™-8C Electrosurgical Generator only. Additional
information about using the generator is available in the Force FX™-8C
Electrosurgical Generator User’s Guide.
Caution
Federal (USA) law restricts this device to sale by or on the order of a physician.
Equipment covered in this manual:
Valleylab Force FX™-8C Electrosurgical Generator with
Instant Response™ Technology—
100–120 V ~ Nominal, 220–240 V ~ Nominal (auto selected)
The Force FX™-8C Electrosurgical Generator Service Manual consists of
two parts—the text (part 1 of 2) and a Schematics Supplement (part 2 of 2)
which contains the schematics.
Valleylab Part Number:
Effective Date:
Trademark Acknowledgments:
September 2000
945 103 068
Force GSU™ Argon System, PolyHesive™ Adhesive Conductor, REM™
Contact Quality Monitoring System, Force EZ™ Electrosurgical
Generator, Force Argon™ II-20 System, Instant Response™ Technology,
The EDGE™ Coated Electrode, ACCUVAC™ Smoke Evacuation
Attachment, and CUSA™ or CUSA EXcel™ Ultrasonic Surgical Aspirator
are trademarks of Valleylab.
Teflon is a registered trademark of E.I du Pont de Nemours and Company.
Patent Information:
Protected by U.S. Pat. Nos. 4,416,276; 4,416,277; 4,658,820; 5,599,344; and
5,628,745.
Manufactured by:
Valleylab,
a division of Tyco Healthcare Group LP
Boulder, Colorado 80301-3299 USA
Instant Response Technology
Ultrasonic Electrosurgery
Simultaneous Coag
REM Contact Quality Monitoring System
Control Board
Microcontrollers
Main Microcontroller
Feedback Microcontroller
Shared RAM
I/0 Expansion
Keyboard Interface and Activation Inputs
Power Supply Supervisor Circuit
A/D and D/A Conversion
Waveform Generation (T_ON ASIC)
T_ON Average Check
Audio Alarm
4-2
4-3
4-3
4-3
4-3
4-5
4-5
4-5
4-6
4-7
4-7
4-8
4-8
4-9
4-9
4-4
4-7
4-8
Force FX-8C Service Manual
v
vi
Serial Interface
Dosage Error Algorithm
Instant Response Algorithm
Front Panel
4-12
Membrane Keyboard
Power Switch
REM Connector/Switch
CEM Mechanism Switch
Display Board
RF Indicator Lamps
REM Indicators
4-9
4-10
4-11
4-12
4-12
4-12
4-12
4-13
4-13
4-13
LED and Seven-Segment Display Drivers
CEM Switch Circuit
4-14
Mode Selection and Power Control Switches
Footswitch Board
Footswitch Decode Circuit
Audio Circuit
Power Supply/RF Board
4-15
4-15
4-16
4-17
Power Supply/RF Board Interfaces
High Voltage Power Supply 4-18
Low Voltage Power Supply 4-21
RF Output Stage 4-21
Spark Control Circuit 4-24
RF Leakage Reduction Circuit 4-25
REM Circuit 4-25
IsoBloc Circuit 4-25
Temperature Sense Circuits 4-26
4-13
4-14
4-18
Section 5. Setup, Tests, and Adjustments
Setting Up the Generator 5-2
Connections for Bipolar or Macrobipolar Surgery 5-4
Setting the Bipolar Output 5-5
Connections for Monopolar Surgery 5-6
Selecting Cut and Coag Modes 5-8
Simultaneous Coag 5-8
Using Two Generators Simultaneously 5-9
Connecting the CUSA Handpiece with CEM Nosecone 5-10
Setting the Output Power 5-11
Simultaneous Coag with a CUSA System 5-11
Changing the Mode 5-11
Changing the Power Setting 5-11
Activating the Surgical Instrument 5-12
Force FX-8C Service Manual
Periodic Safety Check 5-13
Recommended Test Equipment 5-13
Inspecting the Generator and Accessories 5-14
Inspecting the Internal Components 5-15
Testing the Generator 5-16
Verifying REM Function 5-17
Confirming Outputs 5-17
Check the Output for the Cut Modes 5-19
Check the Output for the Coag Modes 5-20
Checking Low Frequency Leakage Current and Ground Resistance 5-21
Checking High Frequency Leakage Current 5-23
Calibrating the Generator 5-24
Preparing for Calibration 5-25
Entering Calibration Mode 5-25
Exiting Calibration Mode 5-26
Verify the Generator Data 5-26
Adjust the Calendar 5-27
Adjust the Clock 5-28
Check and Adjust the REM Oscillator Frequency and Impedance 5-29
Check and Adjust the Current Sense Gain 5-30
Check and Adjust the Voltage Sense Gain 5-32
Check and Adjust the Reactance Gain 5-34
Check and Adjust the ECON Factor 5-36
Using the RS-232 Serial Port 5-40
Establish the Communications Link 5-40
Enter the Commands 5-41
Disconnect the Computer from the Generator 5-44
Section 6. Troubleshooting
Inspecting the Generator 6-1
Inspecting the Receptacles 6-2
Inspecting the Internal Components 6-3
Correcting Malfunctions 6-4
Responding to System Alarms 6-10
Correcting IC U3 Malfunctions 6-19
Correcting IC U6 Malfunctions 6-21
Correcting T_ON ASIC Malfunctions 6-23
Correcting Battery-Backed RAM Malfunctions 6-25
Section 7. Replacement Procedures
Interconnect Diagram 7-2
Battery Replacement 7-3
Force FX-8C Service Manualvii
Control Board Replacement 7-4
Display Board Replacement 7-5
Remove the Display Board 7-5
Install the Display Board 7-6
Display Board Seven-Segment LED Replacement 7-7
Fan Replacement 7-8
Footswitch Board Replacement 7-8
Front Panel Replacement 7-9
Remove the Front Panel Assembly 7-9
Remove and Reinstall the Front Panel Components 7-9
Install the Front Panel Assembly 7-10
Front Panel REM Module Replacement 7-11
Front Panel Power Switch Replacement 7-12
Fuse Replacement 7-13
Replacing Fuses in the Fuse Drawer 7-13
Replacing the Fuse on the Power Supply/RF Board 7-14
Left Front Heat Sink and Component Replacement 7-15
Remove the Left Front Heat Sink 7-15
Replace Left Front Heat Sink Components 7-16
Install the Left Front Heat Sink 7-16
Left Rear Heat Sink and Component Replacement 7-17
Remove the Left Rear Heat Sink 7-17
Replace Left Rear Heat Sink Components 7-18
Install the Left Rear Heat Sink 7-19
Right Heat Sink and Component Replacement 7-20
Remove the Right Heat Sink 7-20
Replace Right Heat Sink Components 7-21
Install the Right Heat Sink 7-22
Low Voltage Power Supply Replacement 7-23
Remove the Low Voltage Power Supply 7-23
Install the Low Voltage Power Supply 7-24
Power Entry Module Replacement 7-25
Remove the Power Entry Module 7-25
Install the Power Entry Module 7-26
Power Supply/RF Board Replacement 7-27
Remove the Power Supply/RF Board Assembly 7-27
Remove Components from the Old Board 7-28
Install Components on the New Board 7-29
Install the Power Supply/RF Board Assembly 7-30
viiiForce FX-8C Service Manual
Section 8. Repair Policy and Procedures
Responsibility of the Manufacturer 8-1
Returning the Generator for Service 8-2
Obtain a Return Authorization Number 8-2
Clean the Generator 8-2
Ship the Generator 8-3
Returning Circuit Boards 8-3
Service Centers 8-4
Section 9. Service Parts
Ordering Replacement Parts 9-1
Generator Assembly 9-2
Parts List 9-4
Front Panel Assembly 9-6
Parts List 9-8
Control Board Components 9-9
Display Board Components 9-10
Footswitch Board Components 9-11
Power Supply/RF Board Assembly 9-14
Parts List 9-16
Power Supply/RF Board Components 9-18
Appendix A. Warranty
Schematics Supplement. Board Drawings and Schematics
Force FX-8C Service Manualix
List of Figures
Figure 2-1.Layout of controls and indicators on the front panel 2-2
Figure 2-2.Buttons and indicators for bipolar controls 2-3
Figure 2-3.Buttons and indicators for cut controls (monopolar) 2-5
Figure 2-4.Buttons and indicators for coag controls (monopolar) 2-6
Figure 2-5.Controls and receptacles on the rear panel 2-8
Figure 2-6.Components in the power entry module 2-10
Figure 2-7.The three ports behind the option panel 2-11
Figure 3-1.Output power versus impedance for Precise bipolar mode 3-14
Figure 3-2.Precise bipolar mode—output power vs. peak voltage 3-14
Figure 3-3.Output power versus impedance for Standard bipolar mode 3-15
Figure 3-4.Standard bipolar mode—output power vs. peak voltage 3-15
Figure 3-5.Output power versus impedance for Macrobipolar mode 3-16
Figure 3-6.Macrobipolar mode—output power vs. peak voltage 3-16
Figure 3-7.Output power versus impedance for Low cut mode 3-17
Figure 3-8.Low cut mode—output power vs. peak voltage 3-17
Figure 3-9.Output power versus impedance for Pure cut mode 3-18
Figure 3-10.Pure cut mode—output power vs. peak voltage 3-18
Figure 3-11.Output power versus impedance for Blend cut mode 3-19
Figure 3-12.Blend cut mode—output power vs. peak voltage 3-19
Figure 3-13.Output power versus impedance for Desiccate 1 coag mode 3-20
Figure 3-14.Desiccate 1 coag mode—output power vs. peak voltage 3-20
Figure 3-15.Desiccate 2 coag mode—load resistance vs. output power 3-21
Figure 3-16.Desiccate 2 coag mode—output power vs. peak voltage 3-21
Figure 3-17.Desiccate 3 coag mode—load resistance vs. output power 3-22
Figure 3-18.Desiccate 3 coag mode—output power vs. peak voltage 3-22
Figure 3-19.Output power versus impedance for Fulgurate coag mode 3-23
Figure 3-20.Fulgurate coag mode—output power vs. peak voltage 3-23
Figure 3-21.Output power versus impedance for LCF Fulgurate mode 3-24
Figure 3-22.LCF Fulgurate mode—output power vs. peak voltage 3-24
Figure 3-23.Output power versus impedance for Spray coag mode 3-25
Figure 3-24.Spray coag mode—output power vs. peak voltage 3-25
xForce FX-8C Service Manual
Figure 5-1.Bipolar or macrobipolar connections—footswitch activation and a
handswitching or footswitching instrument 5-4
Figure 5-2.Bipolar or macrobipolar connection—handswitching
instrument 5-5
Figure 5-3.Monopolar connection—footswitch activation and a footswitching or
handswitching instrument using Monopolar 1 Footswitch receptacle
and Monopolar 1/CEM Instrument receptacle 5-6
Figure 5-4.Monopolar connection—footswitch activation and a footswitching or
handswitching instrument using Monopolar 2 Footswitch receptacle
and Monopolar 2 Instrument receptacle 5-7
Figure 5-5.Monopolar connection—handswitch activation and a monopolar
handswitching instrument using either Monopolar Instrument
receptacle
Figure 5-6.Connections for simultaneous coag—two handswitching
instruments 5-9
Figure 5-7.Connection for simultaneous coag—two footswitching
instruments 5-9
Figure 5-8.Connections for combined monopolar/ultrasonic surgery 5-10
Figure 5-9.Leakage current test circuit per IEC 60601-1 5-21
5-7
Figure 7-1.Removing and disconnecting the power switch 7-12
Figure 7-2.Replacing fuses in the fuse drawer 7-13
Figure 7-3.Replacing left front heat sink components 7-16
Figure 7-4.Components of the left rear heat sink 7-18
Figure 7-5.Components of the right heat sink 7-21
Figure 7-6.Low voltage power supply 7-24
Figure 7-7.Cable connections on the power entry module 7-26
Figure 9-1.Generator assembly 9-2
Figure 9-2.Generator assembly – continued 9-3
Figure 9-3.Front panel assembly 9-6
Figure 9-4.Front panel assembly – continued 9-7
Figure 9-5.Power Supply/RF board assembly 9-14
Figure 9-6.Power Supply/RF board assembly – continued 9-15
Schematics Supplement
Schematic 1.Control board layout S-1
Schematic 2.Control board schematic, page 1 of 3 S-2
Schematic 3.Control board schematic, page 2 of 3 S-3
Schematic 10. Footswitch board schematic, page 1 of 2 S-10
Schematic 11. Footswitch board schematic, page 2 of 2 S-11
Schematic 12. Power supply/RF board layout S-12
Schematic 13. Power supply/RF board schematic, page 1 of 7 S-13
Schematic 14. Power supply/RF board schematic, page 2 of 7 S-14
Schematic 15. Power supply/RF board schematic, page 3 of 7 S-15
Schematic 16. Power supply/RF board schematic, page 4 of 7 S-16
Schematic 17. Power supply/RF board schematic, page 5 of 7 S-17
Schematic 18. Power supply/RF board schematic, page 6 of 7 S-18
Schematic 19. Power supply/RF board schematic, page 7 of 7 S-19
xiiForce FX-8C Service Manual
1Introduction
SECTION
This manual provides instructions for servicing the Valleylab Force FX-8C
Electrosurgical Generator with Instant Response Technology. This section
1
introduces the features and components of the generator and reviews the
precautions associated with generator repair.
Force FX-8C Service Manual1-1
General Description
General Description
The Valleylab Force FX-8C is an isolated output electrosurgical generator
that provides the appropriate power for cutting, desiccating, and
fulgurating tissue during bipolar and monopolar surgery.
It includes the following features:
• Instant Response Technology
• Three bipolar modes: precise (low), standard (medium), and macro
(macrobipolar)
• Three monopolar cut modes: low, pure, and blend
• Three monopolar coag modes: desiccate (low), fulgurate (medium),
and spray (high)
• Support for simultaneous coagulation
• The Valleylab REM Contact Quality Monitoring System
• Support for ultrasonic electrosurgery using the Valleylab CUSA
System 200 or CUSA EXcel system and a CUSA handpiece with a
CUSA electrosurgical module (CEM) nosecone
• Handswitch or footswitch activation
• Recall of most recently used mode and power settings
• Adjustable activation tone volume
• An RF activation port, RS-232 serial port, and expansion port
• Force GSU system and Force Argon system compatibility.
1-2Force FX-8C Service Manual
List of Components
List of Components
The Force FX-8C generator is a self-contained unit, consisting of a main
enclosure (cover and base) and power cord. The main components of the
generator are the following:
• Front panel components—power switch; controls for setting the modes
and output power; a button for recalling the power settings and modes
that were used last; receptacles for connecting electrosurgical
accessories; and indicators that alert you to the current settings and
patient return electrode status.
• Rear panel components—volume control; three footswitch receptacles;
power entry module; equipotential grounding lug; and three ports
(serial port, RF activation port, and expansion port).
• Internal components—control (microcontroller) board; display board;
footswitch board; power supply/radio frequency (RF) board; low
voltage power supply; fan; and heat sinks.
A handle is located on the underside of the chassis.
For details about the interaction of the main components and circuit board
descriptions, refer to Section 4, Principles of Operation.
Introduction
Service Personnel Safety
Before servicing the generator, it is important that you read, understand,
and follow the instructions supplied with it and with any other equipment
used to install, test, adjust, or repair the generator.
General
Use the generator only if the self-test has been completed as described.
Otherwise, inaccurate power outputs may result.
The instrument receptacles on this generator are designed to accept only one
instrument at a time. Do not attempt to connect more than one instrument at a
time into a given receptacle. Doing so will cause simultaneous activation of the
instruments.
Do not stack equipment on top of the generator or place the generator on top of
electrical equipment (except a Force GSU unit, a Force Argon unit, a CUSA
System 200, or a CUSA EXcel unit). These configurations are unstable and/or do
not allow for adequate cooling.
Provide as much distance as possible between the electrosurgical generator and
other electronic equipment (such as monitors). An activated electrosurgical
generator may cause interference with them.
Warning
Caution
Force FX-8C Service Manual1-3
Service Personnel Safety
Caution
Do not turn the activation tone down to an inaudible level. The activation tone
alerts the surgical team when an accessory is active.
Notice
If required by local codes, connect the generator to the hospital equalization
connector with an equipotential cable.
Connect the power cord to a wall receptacle having the correct voltage.
Otherwise, product damage may result.
Active Accessories
Warning
Electric Shock Hazard Do not connect wet accessories to the generator.
Electric Shock Hazard Ensure that all accessories and adapters are correctly
connected and that no metal is exposed.
Caution
Accessories must be connected to the proper receptacle type. In particular,
bipolar accessories must be connected to the Bipolar Instrument receptacle only.
Improper connection may result in inadvertent generator activation or a REM
Contact Quality Monitor alarm.
Set power levels to the lowest setting before testing an accessory.
Notice
During bipolar electrosurgery do not activate the generator until the forceps have
made contact with the patient. Product damage may occur.
Patient Return Electrodes
Warning
Using a patient return electrode without the REM safety feature will not activate
the Valleylab REM Contact Quality Monitoring System.
Fire/Explosion Hazards
Warning
Danger: Explosion Hazard Do not install the generator in the presence of
flammable anesthetics, gases, liquids, or objects.
Fire Hazard Do not place active accessories near or in contact with flammable
materials (such as gauze or surgical drapes). Electrosurgical accessories that
are activated or hot from use can cause a fire. Use a holster to hold
electrosurgical accessories safely away from personnel and flammable materials.
Fire Hazard Do not use extension cords.
1-4Force FX-8C Service Manual
Service Personnel Safety
Warning
Fire Hazard For continued protection against fire hazard, replace fuses only with
fuses of the same type and rating as the original fuse.
Electric Shock Hazards
Warning
Connect the generator power cord to a properly grounded receptacle. Do not use
power plug adapters.
Do not connect a wet power cord to the generator or to the wall receptacle.
To allow stored energy to dissipate after power is disconnected, wait at least five
minutes before replacing parts.
Always turn off and unplug the generator before cleaning.
Do not touch any exposed wiring or conductive surfaces while the generator is
disassembled and energized. Never wear a grounding strap when working on an
energized generator.
Introduction
When taking measurements or troubleshooting the generator, take appropriate
precautions, such as using isolated tools and equipment, using the “one hand
rule,” etc.
Potentially lethal AC and DC voltages are present in the AC line circuitry, high
voltage DC circuitry, and associated mounting and heat sink hardware described
in this manual. They are not isolated from the AC line. Take appropriate
precautions when testing and troubleshooting this area of the generator.
High frequency, high voltage signals that can cause severe burns are present in
the RF output stage and in the associated mounting and heat sink hardware
described in this manual. Take appropriate precautions when testing and
troubleshooting this area of the generator.
Servicing
Caution
Read all warnings, cautions, and instructions provided with this generator before
servicing.
The generator contains electrostatic-sensitive components. When repairing the
generator, work at a static-control workstation. Wear a grounding strap when
handling electrostatic-sensitive components, except when working on an
energized generator. Handle circuit boards by their nonconductive edges. Use an
antistatic container for transport of electrostatic-sensitive components and circuit
boards.
Notice
After installing a new low voltage power supply, verify that the voltages are
correct.
Force FX-8C Service Manual1-5
Service Personnel Safety
Calibration
Caution
To avoid inadvertent coupling and/or shunting of RF currents around the resistor
elements, keep the resistors at least 10.2 cm (4 in.) away from any metal surface
including tabletops and other resistors. This is especially true if several resistors
are connected in series or parallel to obtain a specified value. Do not allow the
resistor bodies to touch each other.
Notice
After completing any calibration step, proceed to the next step to save the values
from the completed calibration step.
Do not activate the generator with any load resistor higher than 10 ohms while
calibrating the current sense gain. Otherwise, product damage will result.
Do not activate the generator with any load resistor lower than 1000 ohms while
calibrating the voltage sense gain for bipolar output. Otherwise, product damage
will result.
Do not activate the generator with any load resistor lower than 3000 ohms while
calibrating the voltage sense gain for the Low and Pure cut modes. Do not
activate the generator with any load resistor lower than 2000 ohms while
calibrating the voltage sense gain for the Blend mode. Otherwise, product
damage will result.
Do not adjust the current sense gain (I factor), the voltage sense gain (V factor),
or the reactance (Z factor) gain while the generator is activated.
After calibration, the generator will be ready to use only after you initiate the
internal self-test by turning the generator off, then on.
Calibrate the generator after you install a new battery. Calibration values are lost
when the battery is replaced.
Calibrate the generator after you install a new control board. Otherwise, the
default calibration values are used.
Calibrate the generator after you install a new heat sink or replace components
on the heat sink. Component differences may affect output waveforms.
Calibrate the generator after you install a new power supply/RF board.
Component differences may affect output waveforms.
Cleaning
Notice
Do not clean the generator with abrasive cleaning or disinfectant compounds,
solvents, or other materials that could scratch the panels or damage the
generator.
1-6Force FX-8C Service Manual
SECTION
2Controls, Indicators, and Receptacles
This section describes the front and rear panels, including all controls,
indicators, receptacles, the fuse drawer, and ports.
2
Force FX-8C Service Manual2-1
Front Panel
Recall button
Pressing this button sets the
generator to the most recently
used mode and power settings.
Bipolar controlsCut controlsCoag controls
CEM
indicator
Bipolar instrument
receptacle
Monopolar 1/CEM
instrument receptacle
Monopolar 2
instrument receptacle
Power switch
This switch supplies power
to the generator.
Patient return electrode
receptacle
For monopolar electrosurgery,
connect a patient return
electrode to this receptacle.
To turn on the generator, press (|).
To turn off the generator, press (
O).
REM alarm
indicator
Front Panel
Figure 2-1.
Layout of controls and indicators on
the front panel
2-2Force FX-8C Service Manual
Bipolar Controls
Bipolar Controls
Figure 2-2.
Buttons and indicators for bipolar
controls
Bipolar display
Shows the power setting, in watts,
for the selected mode.
Mode indicators
Illuminate green when you press
the corresponding mode button.
Bipolar indicator
When you activate bipolar, this
indicator illuminates blue and an
activation tone sounds.
Power buttons
Press
∆ to increase the power.
∇ to decrease the power.
Press
Controls, Indicators,
and Receptacles
Precise mode button
Select for fine bipolar tissue
desiccation.
Macro mode button
Select for macrobipolar output.
Standard mode button
Select for standard bipolar
tissue desiccation.
This is the default bipolar mode.
Force FX-8C Service Manual2-3
Bipolar Instrument Receptacle
Bipolar Instrument Receptacle
Caution
Accessories must be connected to the proper receptacle type. In particular,
bipolar accessories must be connected to the Bipolar Instrument receptacle only.
Improper connection may result in inadvertent generator activation or a REM
Contact Quality Monitor alarm.
You can connect either a footswitching or handswitching bipolar
instrument to the Bipolar instrument receptacle.
Connect a footswitching instrument with a two-pin connector.
or
Connect a handswitching instrument with a three-pin connector.
2-4Force FX-8C Service Manual
Monopolar Cut Controls
Figure 2-3.
Buttons and indicators for cut controls
(monopolar)
Cut display
Shows the power setting, in
watts, for the selected mode.
Monopolar Cut Controls
Cut indicator
When you activate cut, this
indicator illuminates yellow and
an activation tone sounds.
Mode indicators
Illuminate green when you press
the corresponding mode button.
Low mode button
Select for a cut with little or
no sparking.
Power buttons
Press
∆ to increase the power.
Press
∇ to decrease the power.
Blend mode button
Select for slower cutting and
additional hemostasis.
Pure mode button
Select for an even cut with little or no hemostasis.
This is the default monopolar cut mode.
Controls, Indicators,
and Receptacles
Force FX-8C Service Manual2-5
Monopolar Coag Controls
Monopolar Coag Controls
Figure 2-4.
Buttons and indicators for coag
controls (monopolar)
Coag display
Shows the power setting, in
watts, for the selected mode.
Mode indicators
Illuminate green when you press
the corresponding mode button.
Coag indicator
When you activate coag, this
indicator illuminates blue and an
activation tone sounds.
Power buttons
Press
Press
∆ to increase the power.
∇ to decrease the power.
Desiccate mode button
Select to desiccate the area of
tissue that is in direct contact
with the active electrode.
Spray mode button
Select to evenly coagulate a wide area
of tissue with a spray of sparks;
penetration is shallower and tissue area
is larger than in fulgurate mode.
Fulgurate mode button
Select to fulgurate an area of tissue with a
spray of sparks.
This unit is equipped with an additional
fulgurate mode which incorporates a
lower crest factor (LCF) than the factory
default fulgurate mode. For details about
this additional fulgurate mode, LCF
Fulgurate, refer to Section 4, Before
Surgery, in the Force FX-8C
Electrosurgical Generator User’s Guide.
Fulgurate is the default monopolar coag
mode. However, the default coag mode
can be changed to either Desiccate or
Spray through the serial port on the rear
panel.
2-6Force FX-8C Service Manual
Monopolar Instrument Receptacles
Warning
The instrument receptacles on this generator are designed to accept only one
instrument at a time. Do not attempt to connect more than one instrument at a
time into a given receptacle. Doing so will cause simultaneous activation of the
instruments.
You can connect a footswitching or handswitching monopolar instrument
to the monopolar receptacles. Some footswitching instruments may
require a single-pin adapter (E0502 Series) or E0017, available from
Valleylab.
Connect one monopolar instrument to the Monopolar 1/CEM Instrument
receptacle:
Monopolar Instrument Receptacles
REM Alarm Indicator
• A single-pin footswitching instrument or a three-pin handswitching
instrument
or
• A four-pin CUSA handpiece with CEM nosecone. (The CEM indicator
in the upper right of the front panel illuminates green. Refer
to Section 4, Connecting the CUSA Handpiece with CEM Nosecone, in the
Force FX-8C Electrosurgical Generator User’s Guide.
Connect one monopolar instrument to the Monopolar 2 Instrument
receptacle:
• A single-pin footswitching instrument or a three-pin handswitching
instrument
This indicator illuminates red until you properly apply a REM patient
return electrode to the patient and connect it to the generator. Then the
indicator illuminates green. (When you connect an electrode without the
REM safety feature, the indicator does not illuminate.)
If the REM system senses an alarm condition, the indicator flashes red
until you correct the alarm condition—then the indicator illuminates
green. (If you are using a return electrode without the REM safety feature,
the red indicator light is extinguished when you correct the alarm
condition.)
Controls, Indicators,
and Receptacles
Force FX-8C Service Manual2-7
Rear Panel
Volume control
Power entry module
Monopolar Footswitch
receptacles
Option panelBipolar Footswitch
receptacle
Equipotential grounding lug
Use to connect the generator
to earth ground.
Rear Panel
Figure 2-5.
Controls and receptacles on the rear
panel
2-8Force FX-8C Service Manual
Footswitch Receptacles
The rear panel contains three footswitch receptacles: two for monopolar
and one for bipolar.
Monopolar Footswitch Receptacles
You must connect a monopolar footswitch if you connect a monopolar
footswitching instrument to the generator.
Footswitch Receptacles
1
2
Connect a two-pedal monopolar footswitch to the Monopolar 1
Footswitch receptacle.
The connected footswitch activates monopolar output for the instrument
that is connected to the Monopolar 1/CEM Instrument receptacle on the
front panel.
Connect a two-pedal monopolar footswitch to the Monopolar 2
Footswitch receptacle.
The connected footswitch activates monopolar output for the instrument
that is connected to the Monopolar 2 Instrument receptacle on the front
panel
Controls, Indicators,
and Receptacles
Bipolar Footswitch Receptacle
You must connect a bipolar footswitch if you connect a bipolar
footswitching instrument to the generator.
Connect a single-pedal bipolar footswitch to the Bipolar Footswitch
receptacle.
The connected footswitch activates bipolar output for the instrument that
is connected to the Bipolar Instrument receptacle on the front panel.
Force FX-8C Service Manual2-9
Power Entry Module
Power Entry Module
The power entry module consists of a power cord receptacle and a fuse
drawer.
Figure 2-6.
Components in the power entry
module
Activation Tone Volume Control
Fuse drawer
The fuse drawer contains two fuses. Refer to
the Force FX-8C Electrosurgical Generator Service Manual for instructions on changing
the fuses.
Power cord receptacle
Turn to adjust the volume of the tones that sound when the generator is
activated (activation tone). To ensure that the surgical team is alerted to
inadvertent activation, these tones cannot be silenced.
To increase the volume of activation tones, turn the knob clockwise.
To decrease the volume, turn the knob counterclockwise.
2-10Force FX-8C Service Manual
Option Panel
Option Panel
A removable plate on the rear panel covers a serial port, an expansion
port, and an RF (radio frequency) activation port. Remove this plate to
obtain information through the RS-232 port or to install a peripheral
device such as a Bipolar Current Monitor, but retain the original cover
plate. After obtaining information or removing a peripheral device,
reinstall the original cover plate.
Caution
To avoid product damage, do not operate the Force FX-8C without an appropriate
cover plate in place.
To review the technical specifications for each port, refer to Section 3,
Technical Specifications.
Figure 2-7.
The three ports behind the option
panel
Serial port
Allows connection of a computer to the generator.
You can obtain information about the generator
using RS-232 communications protocol or change
the default coag mode from Fulgurate to Desiccate
or Spray. Refer to Section 5, Using the RS-232 Serial Port, for instructions.
Controls, Indicators,
and Receptacles
Expansion port
Allows a connected device to receive
information about the RF voltage and current
being generated as well as signal the generator
to halt RF output.
RF Activation port
Allows a connected device to receive
information during RF activation of the
generator, which will then generate a
response in the device.
Force FX-8C Service Manual2-11
Notes
2-12Force FX-8C Service Manual
Loading...
+ 188 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.