This manual and the equipment it describes are for use only by qualified medical
professionals trained in the particular technique and surgical procedure to be performed.
It is intended as a guide for servicing the Frame EZ
TM
Electrosurgical Generator C only.
Additional information is available in the Frame EZ Electrosurgical Generator C Service Manual.
Equipment covered in this manual:
Frame EZ Electrosurgical Generator C
100–120 V ~ (110 V ~ nominal), 220–240 V ~ (230 V ~ nominal) – (user selectable)
Conventions Used in this Guide
Warning
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious
injury.
Caution
Indicates a hazardous situation which, if not avoided, may result in minor or moderate injury.
Notice
Indicates a hazard which may result in product damage.
Important
Indicates an operating tip or maintenance suggestion.
iiForce EZ Electrosurgical Generator C User’s Guide
Limited Warranty
Covidien warrants each covered product listed below to be free from defects in material
and workmanship for normal use and service for the period(s) set forth below. Covidien’s
obligation under this warranty is limited to the repair or replacement, at its sole option,
of any product, or part thereof, which has been returned to it (or its authorized
distributor) within the applicable time period shown below after delivery of the product
to the original purchaser, and which examination discloses, to Covidien’s satisfaction, that
the product is defective. This limited warranty does not apply to any product, or part
thereof, which has been repaired or altered in a way so as, in Covidien’s judgment, to
affect its stability or reliability, or which has been subjected to misuse, neglect, or
accident.
The warranty periods for Covidien products are as follows:
ForceTriadTM Energy PlatformOne year from date of shipment
Electrosurgical GeneratorsOne year from date of shipment
TM
Cool-tip
RFG-3CTM Plus Lesion GeneratorOne year from date of shipment
LigaSure
LigaSure
RFA GeneratorOne year from date of shipment
TM
Vessel Sealing SystemOne year from date of shipment
TM
Reusable InstrumentsOne year from date of shipment
Mounting Fixtures (all models)One year from date of shipment
Footswitches (all models)One year from date of shipment
TM
Valleylab
RapidVac
LigaSure
Cool-tip
Argon Gas Delivery Unit IIOne year from date of shipment
TM
Smoke EvacuatorOne year from date of shipment
TM
Sterile Single Use ItemsSterility only as stated on packaging
TM
Sterile Single Use ItemsSterility only as stated on packaging
Sterile Single Use ItemsSterility only as stated on packaging
Patient Return ElectrodesShelf life only as stated on packaging
Notwithstanding any other provision herein or in any other document or communication,
Covidien’s liability with respect to this limited warranty and the products sold hereunder
shall be limited to the aggregate purchase price for the products sold to the customer.
This limited warranty is non-transferable and runs only to the original purchaser of the
covered product(s). There are no warranties which extend beyond the terms hereof.
Covidien disclaims any liability hereunder or elsewhere in connection with the sale of
products and for any form of indirect, tort, or consequential damages.
Force EZ Electrosurgical Generator C User’s Guideiii
This limited warranty and the rights and obligations hereunder shall be construed under
and governed by the laws of the State of Colorado, USA. The sole forum for resolving
disputes arising under or relating in any way to this limited warranty is the District Court
of the County of Boulder, State of Colorado, USA.
Covidien reserves the right to make changes in covered products built or sold by it at any
time without incurring any obligation to make the same or similar changes to equipment
previously built or sold by it.
THE OBLIGATION TO REPAIR OR REPLACE A DEFECTIVE OR NONPERFORMING PRODUCT
IS THE SOLE REMEDY OF THE CUSTOMER UNDER THIS LIMITED WARRANTY. EXCEPT AS
EXPRESSLY PROVIDED HEREIN, COVIDIEN DISCLAIMS ALL OTHER WARRANTIES,
WHETHER EXPRESS OR IMPLIED, ORAL OR WRITTEN, WITH RESPECT TO
PRODUCTS, INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES,
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
ivForce EZ Electrosurgical Generator C User’s Guide
viiiForce EZ Electrosurgical Generator C User’s Guide
Chapter 1
Introduction
This chapter includes the following information:
• Instant Response technology
•Bipolar modes
• Monopolar cut and coag modes
• REM Contact Quality Monitoring System
• Special features of the Force EZ-C generator
Caution
Read all warnings, cautions, and instructions provided with this generator before using.
Read the instructions, warnings, and cautions provided with electrosurgical accessories before
using. Specific instructions are not included in this manual.
Force EZ Electrosurgical Generator C User’s Guide1-1
Overview
Overview
The Covidien Force EZ Electrosurgical Generator C—also called the Force EZ-C generator
in this manual—is an isolated output electrosurgical generator that provides power for
cutting, desiccating, and fulgurating tissue during bipolar and monopolar surgery.
It includes the following features:
• Instant Response
• Standard bipolar mode
• Two monopolar cut modes: pure and blend
• Two monopolar coag modes: low (desiccate) and high (fulgurate)
• The REM Contact Quality Monitoring System
• Handswitch or footswitch activation
• User selectable coag settings
TM
technology
• User selectable default settings
• Adjustable activation tone volume
• An RF activation port and RS-232 serial port
• Argon Gas Delivery Unit II system compatibility
• Tilt support provided by the bottom mounted handle or use with Covidien carts
Covidien electrosurgical generators, patient return electrodes, and active accessories are
designed to work as a system. Covidien offers a selection of patient return electrodes and
electrosurgical instruments that are fully compatible with this generator. When
considering other manufacturer’s patient return electrodes and/or active accessories,
customers should seek detailed user instructions and warning information from the
manufacturer.
Instant Response Technology
The Force EZ-C generator automatically senses resistance and adjusts the output voltage,
current, and power to maintain a consistent effect across different tissue density. This
adjustment is based on the selected mode, power setting, and level of tissue resistance.
The maximum output voltage is controlled to reduce capacitive coupling and video
interference and to minimize sparking. This technology applies to the standard bipolar
mode, the cut modes, and the low 2 and low 3 coag settings. It does not apply to the
low 1, high 1, and high 2 coag settings.
1-2Force EZ Electrosurgical Generator C User’s Guide
Bipolar Modes
The Force EZ-C generator provides a standard bipolar mode usable for most bipolar
applications.
Delicate tissue requires less heat to desiccate quickly. The generator provides low voltage,
continuous current for faster desiccation without sparking.
The possibility of sparking increases as desiccated tissue dries and becomes more
resistant. The generator protects against sparking by limiting the bipolar voltage at
relatively high levels of tissue resistance.
For details about the bipolar output characteristics, refer to Chapter 9, Technical Specifications.
Monopolar Cut and Coag Modes
Bipolar Modes
Introduction
Cut Modes
Two cut modes—pure and blend—allow a wide range of power settings necessary to
perform diverse surgical procedures.
• Pure provides an even cut with little or no hemostasis. It offers good cutting
performance over a wide range of tissue resistance
• Blend provides cutting ability with additional hemostasis
Coag Modes
Two coagulation modes help control the size of the area and the depth of penetration
during tissue coagulation. The low (desiccate) mode has three settings; the high
(fulgurate) mode, two settings. You can select, as default settings, one low setting and
one high setting. For a description of each setting, refer to Special Features on page 1-5.
• Low (desiccate) dehydrates and destroys tissue without sparking or cutting. Because
the active electrode directly touches the tissue, more current reaches the patient.
Desiccation places the greatest demand on the patient return electrode.
• High (fulgurate) coagulates tissue by sparking from the active electrode, through air,
to the patient tissue. Since sparks may spray unpredictably from the electrode during
fulguration, using fulguration for delicate tissue or in confined areas can complicate
surgery. Accidental sparking to adjacent areas can occur as tissue at the surgical site
dries and becomes more resistant to current flow.
For details about the output characteristics, refer to Chapter 9, Technical Specifications.
Force EZ Electrosurgical Generator C User’s Guide1-3
REM Contact Quality Monitoring System
REM Contact Quality Monitoring System
During monopolar electrosurgery, a patient return electrode is always required to safely
recover the current that flows through the patient’s body and return it to the generator.
A reduction in surface area contact or poor conductivity between the patient and the
return electrode can cause the current to become concentrated, potentially resulting in
burns at the return electrode site.
The Force EZ-C generator uses the REM Contact Quality Monitoring System to monitor
the quality of electrical contact between the patient return electrode and the patient. The
REM system is designed to minimize the risk of burns at the return electrode site due to a
reduction in patient contact area during monopolar electrosurgery.
Use of any return electrode other than a REM Polyhesive
compromise the REM safety feature. This could result in a patient burn.
How the REM System Works
TM
patient return electrode may
The REMTM system continuously measures the resistance at the return electrode site and
compares it to a standard range of safe resistance (between 5 and 135 ohms), thus
eliminating intermittent false alarms that could result from small changes in resistance.
The REM system also adapts to individual patients by measuring the initial contact
resistance between the patient and the patient return electrode and lowering the
baseline resistance if the contact resistance drops.
A REM alarm sounds and the generator stops producing output power when either of
the following occurs:
• The measured resistance is below 5 ohms or above 135 ohms, the limits of the
standard range of safe resistance
• An increase in contact resistance is greater than 40% from the baseline measurement
Electrodes Without the REM Safety Feature
Warning
Using a patient return electrode without the REM safety feature will not activate the REM Contact
Quality Monitoring System.
When you use a patient return electrode that does not have the REM safety feature, the
REM system cannot monitor the patient contact area as previously described. The REM
system can monitor only the pin-to-pin resistance at the connector and can detect
broken wires or connectors in the return electrode cord.
1-4Force EZ Electrosurgical Generator C User’s Guide
Special Features
Five special features allow customizing the Force EZ-C generator. You can access these
features only in the setup mode. For details on selecting these features, refer to Setting Up the Special Features on page 4-13.
Low (Desiccate) Coag Settings
The low (desiccate) coag mode provides three settings with subtle differences in output
characteristics:
• Low 1 is appropriate for the majority of surgical procedures. It provides tissue
desiccation with a minimal tendency to cut or spark.
• Low 2 produces tissue desiccation and further reduces the chance of cutting or
sparking by using the lowest possible voltage (200 V
• Low 3 uses a slightly higher voltage (300 V
comparable desiccation.
rms
) than the low 2 setting, but offers
rms
).
Special Features
Introduction
High (Fulgurate) Coag Settings
The high (fulgurate) coag mode provides two settings:
• High 1 produces coagulation of smaller areas without touching the electrode tip to
the tissue.
• High 2 produces coagulation of larger areas without touching the electrode tip to the
tissue.
Recall of Most Recently Used Modes and Power Settings
When you activate this feature, the generator will, when turned on, revert to the most
recently used modes and power settings.
Default Coag Mode
Important
The default coag mode feature is available only when the most recently used modes and power
settings feature is turned off
You can select either low (desiccate) or high (fulgurate) as the default coag mode for the
Force EZ-C generator. Each time you turn on the generator, it automatically selects the
default coag mode.
Force EZ Electrosurgical Generator C User’s Guide1-5
Special Features
Original Default Settings
This feature resets the generator to the original default setting for each mode, power
setting, and special feature. The next time you turn on the generator, it automatically
selects the original default settings.
Mode or FeatureOriginal Default Setting
Monopolar Pure
Coag High (fulgurate)
Low (desiccate) coag setting 1 (low 1)
High (fulgurate) coag setting 2 (high 2)
Last used modes and power settings1 (on)
1-6Force EZ Electrosurgical Generator C User’s Guide
Chapter 2
Controls, Indicators and Receptacles
This chapter describes the front and rear panels, including all controls, indicators,
receptacles, the fuse drawer, and ports.
Force EZ Electrosurgical Generator C User’s Guide2-1
Front Panel
Front Panel
Bipolar controlsCut controls
Bipolar
instrument
receptacle
Power switch
This switch supplies
power to the
generator.
Footswitch
selector
button
Footswitch
receptacle
Footswitch
indicators
Monopolar
footswitching
accessory
receptacle
Coag controls
Monopolar
footswitching or
handswitching
instrument
receptacle
REM alarm
indicator
Patient return
electrode receptacle
For monopolar
electrosurgery, connect
a patient return
electrode to this
receptacle.
To turn on the generator, press (|).
To turn off the generator, press (
O).
2-2Force EZ Electrosurgical Generator C User’s Guide
Bipolar Controls
Bipolar display
Shows the power setting, in
watts, for standard bipolar
mode.
Bipolar Controls
Bipolar indicator
When you activate bipolar, this
bar illuminates blue and an
activation tone sounds.
Controls, Indicators and
Receptacles
Bipolar Power Control Knob
To increase (+) the power, turn
the knob clockwise.
To decrease (–) the power, turn
the knob counterclockwise.
Force EZ Electrosurgical Generator C User’s Guide2-3
Bipolar Instrument Receptacle
Bipolar Instrument Receptacle
Caution
Accessories must be connected to the proper receptacle type. In particular, bipolar accessories
must be connected to the Bipolar Instrument receptacle only. Improper connection may result in
inadvertent generator activation or a REM Contact Quality Monitor alarm.
You can connect either a footswitching or handswitching bipolar instrument to the
Bipolar instrument receptacle.
Connect a footswitching instrument with a two-pin connector.
or
Connect a handswitching instrument with a three-pin connector.
Footswitch Receptacle, Button, and Indicators
Connect a two-pedal Valleylab monopolar footswitch to this receptacle. Press the
Footswitch Selector button to select bipolar or accessory output.
Use only a Valleylab monopolar footswitch with the Force EZ-C generator. Use of an
incompatible footswitch may cause unexpected output.
When the left arrow indicator illuminates green, the footswitch activates the instrument
connected to the Bipolar Instrument receptacle.
When the right arrow indicator illuminates green, the footswitch activates the instrument
connected to the Monopolar Footswitching Accessory receptacle.
The footswitch will not activate an instrument connected to the Monopolar
Footswitching or Handswitching Instrument receptacle.
2-4Force EZ Electrosurgical Generator C User’s Guide
Monopolar Cut Controls
Monopolar Cut Controls
Cut display
Shows the power setting, in
watts, for cut output.
Cut indicator
When you activate cut, this bar
illuminates yellow and an
activation tone sounds.
Cut Power Control Knob
To increase (+) the power, turn the
knob clockwise.
To decrease (–) the power, turn the
knob counterclockwise.
Pure Button
Select for an even
cut with little or no
hemostasis.
Blend Button
Select for slower
cutting and
additional
hemostasis.
Controls, Indicators and
Receptacles
Force EZ Electrosurgical Generator C User’s Guide2-5
Monopolar Coag Controls
Monopolar Coag Controls
Coag display
Shows the power
setting, in watts, for
coag output.
Coag indicator
When you activate the generator in
coag mode, this bar illuminates blue
and an activation tone sounds.
Low (Desiccate)
Button
Select to desiccate
the area of tissue that
is in direct contact
with the active
electrode.
High (Fulgurate)
Button
Select to fulgurate
an area of tissue
with a spray of
sparks.
Coag Power Control Knob
To increase (+) the power, turn the
knob clockwise.
To decrease (–) the power, turn the
knob counterclockwise.
2-6Force EZ Electrosurgical Generator C User’s Guide
Monopolar Instrument Receptacles
Monopolar Instrument Receptacles
Warning
The instrument receptacles on this generator are designed to accept only one instrument at a
time. Do not attempt to connect more than one instrument at a time into a given receptacle.
Doing so will cause simultaneous activation of the instruments.
Monopolar Footswitching Accessory Receptacle
Connect a monopolar footswitching instrument with a single-pin connector to this
receptacle.
To activate the instrument, connect a monopolar footswitch to the front panel.
Monopolar Footswitching or Handswitching Instrument
Receptacle
You can connect either a handswitching instrument (three-pin connector) or a
footswitching instrument (single-pin connector) to this receptacle.
Controls, Indicators and
Receptacles
To activate a footswitching instrument, connect a monopolar footswitch to the rear
panel.
Some footswitching instruments may require a single-pin adapter (E0502 Series) or
E0017, available from Covidien.
Force EZ Electrosurgical Generator C User’s Guide2-7
REM Alarm Indicator
REM Alarm Indicator
This indicator illuminates red until you properly apply a REM Polyhesive patient return
electrode to the patient and connect it to the generator. Then the indicator illuminates
green. (When you connect an electrode without the REM safety feature, the indicator
does not illuminate.)
If the REM system senses an alarm condition, the indicator flashes red until you correct
the alarm condition—then the indicator illuminates green. (If you are using a return
electrode without the REM safety feature, the red indicator light is extinguished when
you correct the alarm condition.)
2-8Force EZ Electrosurgical Generator C User’s Guide
Rear Panel
Rear Panel
Vol ume control
Bipolar Footswitch
receptacle
Monopolar Footswitch
receptacle
Power entry module
Controls, Indicators and
Receptacles
Option panel
Equipotential grounding lug
Use to connect the generator to
earth ground.
Force EZ Electrosurgical Generator C User’s Guide2-9
Footswitch Receptacles
Footswitch Receptacles
Monopolar Footswitch Receptacle
Connect a two-pedal Valleylab monopolar footswitch to this receptacle if you connect an
instrument to the Monopolar Footswitching or Handswitching Instrument receptacle on
the front panel.
Use only a Valleylab monopolar footswitch with the Force EZ-C generator. Use of an
incompatible footswitch may cause unexpected output.
The footswitch will not activate instruments connected to the Bipolar Instrument or
Monopolar Footswitching Accessory receptacles on the front panel.
Bipolar Footswitch Receptacle
Connect a single-pedal bipolar footswitch to this receptacle if you connect an instrument
to the Bipolar Instrument receptacle on the front panel.
The footswitch will not activate instruments connected to the Monopolar Instrument or
Monopolar Footswitching Accessory receptacles on the front panel.
2-10Force EZ Electrosurgical Generator C User’s Guide
Power Entry Module
The power entry module consists of a power cord receptacle and a fuse drawer.
Fuse drawer
The fuse drawer contains two fuses. Refer
to the Force EZ Electrosurgical Generator C/8C Service Manual for instructions on
changing the fuses.
Power cord receptacle
Activation Tone Volume Control
Power Entry Module
Controls, Indicators and
Receptacles
Turn to adjust the volume of the tones that sound when the generator is activated
(activation tone). To ensure that the surgical team is alerted to inadvertent activation,
these tones cannot be silenced.
To increase the volume of activation tones, turn the knob clockwise.
To decrease the volume, turn the knob counterclockwise.
Force EZ Electrosurgical Generator C User’s Guide2-11
Option Panel
Option Panel
A removable plate on the rear panel covers a serial port and an RF (radio frequency)
activation port. Remove this plate to obtain information through the RS-232 port or to
install a peripheral device such as a Bipolar Current Monitor, but retain the original cover
plate. After obtaining information or removing a peripheral device, reinstall the original
cover plate.
To review the technical specifications for each port, refer to Chapter 9, Technical Specifications.
RF Activation port
Allows a connected device to receive
information during RF activation of the
generator, which will then generate a
response in the device.
Serial port
Allows connection of a computer to the
generator. You can obtain information about
the generator using RS-232 communications
protocol. Refer to the Force EZ Electrosurgical Generator C/8C Service Manual.
2-12Force EZ Electrosurgical Generator C User’s Guide
Chapter 3
Patient and Operating Room Safety
The safe and effective use of electrosurgery depends to a large degree upon
factors solely under the control of the operator. There is no substitute for a
properly trained and vigilant surgical team. It is important that the operating
instructions supplied with this or any electrosurgical equipment be read,
understood, and followed.
Electrosurgery has been used safely in numerous procedures. Before starting any
surgical procedure, the surgeon should be trained in the particular technique and
surgical procedure to be performed, should be familiar with the medical
literature related to the procedure and potential complications, and should be
familiar with the risks versus the benefits of utilizing electrosurgery in the
procedure.
Force EZ Electrosurgical Generator C User’s Guide3-1
General
General
Warning
Do not use electrosurgical equipment unless properly trained to use it in the specific procedure
being undertaken. Use by physicians without such training has resulted in serious, unintended
patient injury, including bowel perforation and unintended, irreversible tissue necrosis.
Hazardous Electrical Output This equipment is for use only by trained, licensed physician
Always use the lowest output setting necessary that achieves the desired surgical effect. The
active electrode should be utilized only for the minimum time necessary in order to lessen the
possibility of unintended burn injury. Pediatric applications and/or procedures performed on small
anatomic structures may require reduced power settings. The higher the current flow and the
longer the current is applied, the greater the possibility of unintended thermal damage to tissue,
especially during use on small structures.
Use electrosurgery with caution in the presence of internal or external pacemakers. Interference
produced by the use of electrosurgical devices can cause devices such as a pacemaker to enter an
asynchronous mode or can block the pacemaker effect entirely. Consult the pacemaker
manufacturer or hospital Cardiology Department for further information when use of
electrosurgical appliances is planned in patients with cardiac pacemakers.
If the patient has an internal cardiac defibrillator (ICD), contact the ICD manufacturer for
instructions before performing an electrosurgical procedure. Electrosurgery may cause multiple
activations of ICDs.
Covidien recommends against performing laparoscopic surgery on pregnant patients.
Caution
Read all warnings, cautions, and instructions provided with this generator before using.
For surgical procedures where the current could flow through delicate parts of the body, the use
of bipolar techniques may be desirable in order to avoid unwanted coagulation.
Always use the lowest output setting that achieves the desired surgical effect. The active
electrode should be utilized only for the minimum time necessary in order to lessen the possibility
of unintended burn injury. Pediatric applications and/or procedures performed on small anatomic
structures may require reduced power settings. The higher the current flow and the longer the
current is applied, the greater the possibility of unintended thermal damage to tissue, especially
during use on small appendages.
3-2Force EZ Electrosurgical Generator C User’s Guide
Fire/Explosion
Warning
Danger: Explosion Hazard Do not use electrosurgery in the presence of flammable
anesthetics.
Fire/Explosion Hazard The following substances will contribute to increased fire and explosion
hazards in the operating room:
• Flammable substances (such as alcohol based skin prepping agents and tinctures)
• Naturally occurring flammable gases which may accumulate in body cavities such as the bowel
• Oxygen enriched atmospheres
• Oxidizing agents (such as nitrous oxide [
The sparking and heating associated with electrosurgery can provide an ignition source. Observe
fire precautions at all times. When using electrosurgery in the same room with any of these
substances or gases, prevent their accumulation or pooling under surgical drapes, or within the
area where electrosurgery is performed.
N2O] atmospheres).
General
Fire Hazard with Oxygen Circuit Connections
Warning
Fire/Explosion Hazard Verify that all oxygen circuit connections are leak free before and during
the use of electrosurgery. Verify that endotracheal tubes are leak free, and that the cuff is properly
sealed to prevent oxygen leaks. Enriched oxygen atmospheres may result in fires and burns to
patients or the surgical team.
Electrosurgical Smoke
Caution
Studies have shown that smoke generated during electrosurgical procedures can be potentially
harmful to patients and the surgical team. These studies recommend adequately ventilating the
smoke by using a surgical smoke evacuator or other means.
a. U.S. Department of Health and Human Services. National Institute for Occupational Safety and
Health (NIOSH). Control of Smoke from Laser/Electric Surgical Procedures. HAZARD
CONTROLS, Publication No. 96-128, September, 1996.
a
Patient and Operating Room
Safety
Force EZ Electrosurgical Generator C User’s Guide3-3
General
Inadvertent Radio Frequency Burns
Warning
Electrodes and probes used with monitoring, stimulation, and imaging devices (or similar
equipment) can provide a path for high frequency current even if the electrodes or probes are
isolated at 50-60 Hz, insulated, and/or battery operated.
To reduce the risk of an inadvertent electrosurgical burn at the electrode or probe site, place the
electrode and/or probe as far away as possible from the electrosurgical site and/or patient return
electrode. Protective impedances (resistors or RF inductors) installed in the monitoring leads may
reduce the risk of such burns. Consult the hospital biomedical engineer for further information.
Do not use needles as monitoring electrodes during electrosurgical procedures. Inadvertent
electrosurgical burns may result.
In some circumstances, potential exists for alternate site burns at points of skin contact (e.g.,
between the arm and the side of the body). This occurs when electrosurgical current seeks a path
to the patient return electrode that includes the skin to skin contact point. Current passing
through small skin to skin contact points is concentrated and may cause a burn. This is true for
grounded, ground referenced, and isolated output generators.
To reduce the potential for alternate site burns, do one or more of the following:
• Avoid skin-to-skin contact points, such as fingers touching leg, when positioning the patient.
• Place five to eight centimeters (two to three inches) of dry gauze between contact points to
ensure that contact does not occur.
• Position the patient return electrode to provide a direct current route between the surgical site
and the return electrode which avoids skin-to-skin contact areas.
• In addition, place patient return electrodes according to the manufacturer’s instructions.
Potential for alternate site burns increases if the return electrode is compromised. Covidien
recommends the use of REM Polyhesive patient return electrodes and Covidien generators with
the REM system.
Ensure Proper Connections
Caution
Examine all accessories and connections to the electrosurgical generator before using. Ensure that
the accessories function as intended. Improper connection may result in arcs, sparks, accessory
malfunction, or unintended surgical effects.
3-4Force EZ Electrosurgical Generator C User’s Guide
Loading...
+ 84 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.