Valleylab Argon Gas Delivery Unit II Service Manuali
Preface
This manual and the equipment it describes are for use only by qualified medical
professionals trained in the particular technique and surgical procedure to be performed.
It is intended as a guide for servicing the Force FX Electrosurgical Generator C only.
Additional users information is available in the Valleylab Argon Gas Delivery Unit II User’s Guide.
Equipment covered in this manual:
Valleylab Argon Gas Delivery Unit II (Force Argon II-20)
120 VAC nominal, 50–60 Hz
The Valleylab Argon Gas Delivery Unit II Service Manual consists of two parts: the text
(part 1 of 2) and a Schematics Supplement (part 2 of 2) which contains the schematics.
Conventions Used in this Guide
Warning
Indicates a potentially hazardous situation, which if not avoided, could result in death or serious
injury.
Caution
Indicates a hazardous situation, which if not avoided, may result in minor or moderate injury.
Notice
Indicates a hazard which may result in product damage.
Important
Indicates an operating tip or maintenance suggestion.
iiValleylab Argon Gas Delivery Unit II Service Manual
Limited Warranty
Covidien warrants each covered product listed below to be free from defects in material
and workmanship for normal use and service for the period(s) set forth below. Covidien’s
obligation under this warranty is limited to the repair or replacement, at its sole option,
of any product, or part thereof, which has been returned to it (or its authorized
distributor) within the applicable time period shown below after delivery of the product
to the original purchaser, and which examination discloses, to Covidien’s satisfaction, that
the product is defective. This limited warranty does not apply to any product, or part
thereof, which has been repaired or altered in a way so as, in Covidien’s judgment, to
affect its stability or reliability, or which has been subjected to misuse, neglect, or
accident.
The warranty periods for Covidien products are as follows:
ForceTriadTM Energy PlatformOne year from date of shipment
Electrosurgical GeneratorsOne year from date of shipment
TM
Cool-tip
EvidentTM MWA GeneratorOne year from date of shipment
RFG-3C
LigaSure
RFA GeneratorOne year from date of shipment
TM
Plus Lesion GeneratorOne year from date of shipment
TM
Vessel Sealing SystemOne year from date of shipment
LigaSureTM Reusable InstrumentsOne year from date of shipment
Mounting Fixtures (all models)One year from date of shipment
Footswitches (all models)One year from date of shipment
TM
Valleylab
RapidVac
LigaSure
Argon Gas Delivery Unit IIOne year from date of shipment
TM
Smoke EvacuatorOne year from date of shipment
TM
Sterile Single Use ItemsSterility only as stated on packaging
Cool-tipTM Sterile Single Use ItemsSterility only as stated on packaging
Sterile Single Use ItemsSterility only as stated on packaging
Patient Return ElectrodesShelf life only as stated on packaging
Notwithstanding any other provision herein or in any other document or communication,
Covidien’s liability with respect to this limited warranty and the products sold hereunder
shall be limited to the aggregate purchase price for the products sold to the customer.
This limited warranty is non-transferable and runs only to the original purchaser of the
Valleylab Argon Gas Delivery Unit II Service Manualiii
covered product(s). There are no warranties which extend beyond the terms hereof.
Covidien disclaims any liability hereunder or elsewhere in connection with the sale of
products and for any form of indirect, tort, or consequential damages.
This limited warranty and the rights and obligations hereunder shall be construed under
and governed by the laws of the State of Colorado, USA. The sole forum for resolving
disputes arising under or relating in any way to this limited warranty is the District Court
of the County of Boulder, State of Colorado, USA.
Covidien reserves the right to make changes in covered products built or sold by it at any
time without incurring any obligation to make the same or similar changes to equipment
previously built or sold by it.
THE OBLICATION TO REPAIR OR REPLACE A DEFECTIVE OR NONPERFORMING PRODUCT
IS THE SOLE REMEDY OF THE CUSTOMER UNDER THIS LIMITED WARRANTY. EXCEPT AS
EXPRESSLY PROVIDED HEREIN, COVIDIEN DISCLAIMS ALL OTHER WARRANTIES,
WHETHER EXPRESS OR IMPLIED, ORAL OR WRITTEN, WITH RESPECT TO
PRODUCTS, INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES,
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
ivValleylab Argon Gas Delivery Unit II Service Manual
viiiValleylab Argon Gas Delivery Unit II Service Manual
Chapter 1
Introduction
This manual provides instructions for servicing the Valleylab Argon Gas Delivery
Unit II. This chapter introduces the features and components of the unit and
reviews the precautions associated with repair.
Introduction
Valleylab Argon Gas Delivery Unit II Service Manual1-1
General Description
General Description
The Valleylab Argon Gas Delivery Unit II System is designed to provide a controlled flow
of argon (an inert, nonreactive gas) over the active electrode of an electrosurgical
accessory which is powered by a Force series electrosurgical generator. A variety of
handsets provide argon-enhanced electrosurgery as well as standard electrosurgery.
Argon-enhanced coag permits the coagulation of certain tissue in a more controlled
manner than with standard electrosurgical coagulation.
In argon-enhanced coag, activating the handset creates an argon gas plasma between
the electrode and the tissue. The presence of inert gas at the surgical site inhibits
combustion with other gases (e.g., oxygen, nitrogen, etc.) and pushes residual smoke,
water vapor, or fluids away from the surgical site, thereby improving visibility.
The Valleylab Argon Gas Delivery Unit II has an overpressure system that continuously
measures peritoneal cavity pressure in real-time. This system operates when the unit is in
the low flow range for argon gas (0.5 - 4 LPM), i.e. the gas flow range for laparoscopic
procedures. Use the overpressure system in conjunction with a carbon dioxide insufflator
monitor.
The overpressure system senses when the peritoneal cavity pressure meets or exceeds the
user selectable setpoint (between 5 and 30 mm Hg). As the peritoneal cavity pressure
approaches the setpoint, the Overpressure indicator on the front panel illuminates, and
an intermittent alarm sounds. When the peritoneal cavity pressure meets or exceeds the
setpoint, the Overpressure indicator on the front panel illuminates, and a continuous
alarm sounds.
List of Components
The Valleylab Argon Gas Delivery Unit II is shipped with the following components:
• Valleylab Argon Gas Delivery Unit II
• ESU Interface Cable
• E0502-12 Active Adapter
• Argon Gas Tank Regulator Assembly for Tank 1
•Power Cord
• Power cord retaining clip
• User’s Guide, Service Manual, Reference Card
1-2Valleylab Argon Gas Delivery Unit II Service Manual
Service Personnel Safety
Service Personnel Safety
Before servicing the Valleylab Argon Gas Delivery Unit II, it is important that you read,
understand, and follow the instructions supplied with the unit and with any other
equipment used to install, test, adjust, or repair the unit.
General Service Precautions
Caution
The unit contains electrostatic sensitive components. When repairing the unit, work at a static
control workstation. Wear a grounding strap when handling electrostatic sensitive components.
Handle the circuit boards by their nonconductive edges. Use an antistatic container for transport
of electrostatic sensitive components and circuit boards.
Introduction
Notice
Connect the power cord to a wall receptacle having the correct voltage. Otherwise, product
damage may result.
If required by local codes, connect the unit to the hospital equalization connector with an
equipotential cable.
Fire/Explosion
Warning
Do not install the Valleylab Argon Gas Delivery Unit II System in the presence of flammable
anesthetics, gases, liquids, or objects.
Always use high purity grade (99.998% pure or better) argon gas. Do not use oxygen for any
adjustments. Oxygen regulation values differ from argon regulation values, and incorrect
adjustments result.
For continued protection against fire hazard, replace fuses only with those having the same type
and rating as the original fuse.
Fire Hazard Do not place active accessories near or in contact with flammable materials (such as
gauze or surgical drapes). Electrosurgical accessories that are activated or hot from use can cause
a fire. Use a holster to hold electrosurgical accessories safely away from personnel and flammable
materials.
Valleylab Argon Gas Delivery Unit II Service Manual1-3
Service Personnel Safety
Electric Shock
Warning
Disconnect the power cord before replacing parts.
Always unplug the Valleylab Argon Gas Delivery Unit II before cleaning.
Do not connect a wet power cord assembly to the unit or a wall receptacle.
Ensure that all accessories and adapters are correctly connected and that no metal is exposed.
The interface cable must be connected to the Valleylab Argon Gas Delivery Unit II and the
electrosurgical generator for operation. If the Valleylab Argon Gas Delivery Unit II is not being
used, disconnect the power cord from the wall receptacle.
Do not touch any exposed wiring or conductive surface while the unit is disassembled and
energized. Never wear a grounding strap when working on an energized unit. Otherwise,
damage to circuit board components may result.
Take appropriate precautions (such as use of isolated tools and equipment; use of the “one hand
rule,” etc.) when taking measurements or troubleshooting the unit.
The front heatsink on the power supply is at an elevated voltage. Do not come in contact with it.
Do not use it for ground. Otherwise, burns to service personnel may result. Damage to circuit
board components may result.
Caution
When attaching the Argon Handset, ensure that the RF jumper cable is disconnected from the RF
Input receptacle on the Valleylab Argon Gas Delivery Unit II front panel.
High Pressure
Warning
Ensure that argon gas tanks and fittings used with the Valleylab Argon Gas Delivery Unit II have
been pressure tested, certified, and filled according to applicable standards and local codes. In
filling the tanks, ensure your supplier meets all local and national codes. Do not over pressurize
the tanks. Do not exceed 3,000 psig.
Use only the high pressure regulator supplied with the Valleylab Argon Gas Delivery Unit II. Order
replacement regulators from Valleylab. Argon gas pressure should not exceed 434 KPa (63 psig) ±
5% at the regulator hose connection to the receptacle on the rear panel.
Always ensure that the regulator tank fitting is properly connected to the tank valve outlet.
Caution
Improperly secured gas tanks and/or improper connection of gas lines may result in injury.
1-4Valleylab Argon Gas Delivery Unit II Service Manual
Service Personnel Safety
Argon Gas
Warning
Use only high purity grade (99.998% pure or better) argon gas. Patient safety may be
compromised by using low purity gas.
The regulator coupling in this unit is compatible with many common gas tanks (e.g., Freon,
helium, krypton, neon, nitrogen, and xenon). Only connect argon gas to the Valleylab Argon Gas
Delivery Unit II. Using other gases may affect gas flow, create phosgene or other harmful gases,
or create an explosion hazard.
Cleaning
Notice
Do not clean the unit with abrasive cleaning or disinfectant compounds, solvents, or other
materials that could scratch the panels or damage the generator.
Introduction
Valleylab Argon Gas Delivery Unit II Service Manual1-5
Chapter 2
Controls, Indicators, and Receptacles
The controls, indicators and receptacles for accessories are located on the front
and rear panels of the Valleylab Argon Gas Delivery Unit II. This chapter describes
each component of the unit and its function.
Controls, Indicators, and
Detailed specifications for the unit are in Chapter 3, Technical Specifications.
Receptacles
Valleylab Argon Gas Delivery Unit II Service Manual2-1
Valleylab Argon Gas Delivery Unit II Front Panel Descriptions
Valleylab Argon Gas Delivery Unit II
Front Panel Descriptions
Overpressure System
This system operates only when the Valleylab Argon Gas Delivery Unit II System is in the
low gas flow range. Use only this range for laparoscopic procedures.
Overpressure Setpoint Display - The display shows the overpressure setpoint (available
only when using the low flow range).
Overpressure Indicator - As the peritoneal cavity pressure approaches the overpressure
setpoint, this indicator illuminates and an intermittent alarm sounds. When the
peritoneal cavity pressure meets or exceeds the setpoint, this indicator remains
illuminated and a continuous alarm sounds.
Peritoneal Pressure Display - The display indicates the peritoneal cavity pressure
(available only when using the low flow range). The display flashes when the pressure is
below 1 mm Hg.
Overpressure On/Off Switch - Press this button to turn peritoneal cavity pressure
monitoring on or off (available only when using the low flow range).
Overpressure Control Knob - To increase (+) the alarm setting, turn the knob
clockwise. To decrease (-) the alarm setting, turn the knob counterclockwise.
Setpoint Symbol
System Purge Control
Handset Tubing or Peritoneal Pressure Tubing Obstruction Indicator - When a gas
tubing obstruction occurs, the indicator illuminates and an alarm sounds. The Gas Flow
indicator or the Overpressure indicator will also be flashing.
Gas Line Purge Button - Pressing this button removes air and causes a brief discharge
of argon gas through the handset tubing at a gas flow of 3.0 LPM.
Gas Flow Controls
Gas On/Gas Off Button - Press this button to select gas on or gas off.
Gas On Indicator - When you turn on the gas, the indicator illuminates and gas is
available to the handset.
Gas Off Indicator - When you turn off the gas, the indicator illuminates and gas is not
available to the handset.
Gas Flow Indicator - This indicator illuminates when you activate argon-enhanced coag
or argon-shrouded cut or when you press the Gas Line Purge button.
Flow Rate Display - The display shows gas flow on a scale from 0.5 to 12 LPM. The
2-2Valleylab Argon Gas Delivery Unit II Service Manual
Valleylab Argon Gas Delivery Unit II Front Panel Descriptions
default gas flow setting is 0.5 LPM.
Standard Gas Flow Indicator - This green LED illuminates when you select the standard
gas flow range.
Gas Flow Range Selector - Press this switch to the top position for standard flow range
(0.5 - 12 LPM) or bottom position for low gas flow (0.5 - 4 LPM).
Low Gas Flow Indicator - This green LED illuminates when you select the low gas flow
range.
Flow Control Knob - To increase (+) the argon gas flow, turn the knob clockwise. To
decrease (-) the argon gas flow, turn the knob counterclockwise.
Controls, Indicators, and
Gas Tank Indicators and Controls
Tank Select Button, Tank 1 - Press this button to select tank 1.
Tan k I n di cat or, Tan k 1 - The indicator illuminates when you select tank 1.
Low Pressure Alarm, Tank 1 - The indicator illuminates when the tank pressure is
below 10% of full tank pressure. The indicator flashes and a brief alarm sounds when all
available tanks are low.
% Argon in Tank 1 - The bar graph indicates percent of argon gas remaining in tank 1.
Tank Select Button, Tank 2 - Press this button to select tank 2.
Tan k I n di cat or, Tan k 2 - The indicator illuminates when you select tank 2.
Low Pressure Alarm, Tank 2 - The indicator illuminates when the tank pressure is
below 10% of full tank pressure. The indicator flashes and a brief alarm sounds when all
available tanks are low.
% Argon in Tank 2 - The bar graph indicates percent of argon gas remaining in tank 2.
Gas Lines and Receptacles
On/Off Switch - This switch supplies power to the unit.
Gas Line Receptacle - Connect the handset particle filter to this luer receptacle.
RF Input Receptacle - Connect the RF jumper cable to this receptacle when using the
Valleylab Argon Accessory.
Handset Receptacle - Connect the handset three-pin connector to this receptacle.
Valleylab Argon Gas Delivery Unit II Service Manual2-3
Valleylab Argon Gas Delivery Unit II Front Panel Descriptions
Front panel
Overpressure Setpoint Display
Obstruction Indicator (Handset
Tubing or Peritoneal Pressure Tubing)
Overpressure
Indicator
Gas On/Gas Off Button
Standard Gas
Flow Indicator
Tan k S el ec t
Button, Tank 1
Peritoneal
Pressure
Display
Overpressure
On/Off Switch
On/Off
Switch
Peritoneal Pressure
Tubing Receptacle
Overpressure Control
Knob
Gas On Indicator
Gas Line
Purge Button
Flow Rate
Display
Gas Flow
Indicator
Flow Control
Knob
Gas Off
Indicator
% Argon in Tank 1
Gas Flow
Range
Selector
Low Gas
Flow
Indicator
Gas Line
Receptacle
Ta nk
Indicator,
Tan k 1
Low Pressure
Alarm, Tank 2
RF Input
Receptacle
Tank Indicator,
Tank 2
% Argon in
Tank 2
Handset
Receptacle
Tank Select
Button, Tank 2
Low Pressure
Alarm, Tank 1
2-4Valleylab Argon Gas Delivery Unit II Service Manual
Rear panel
Tank 1 Receptacle
Connect the tank 1
regulator hose to this
receptacle.
Connect the tank 2 regulator
hose to this receptacle.
The Valleylab Argon Gas Delivery Unit II may
be adjusted through this serial port to stop
argon gas delivery when the peritoneal
cavity pressure meets or exceeds the
selected setpoint. Contact your Valleylab
representative for further information.
Tank 2 Receptacle
Valleylab Argon Gas Delivery Unit II Front Panel Descriptions
RS-232 Serial Port
Power Entry Module
Connect a power cord with a
three-prong hospital grade
connector to this receptacle.
There are two entry module fuses
in this receptacle.
Controls, Indicators, and
Receptacles
Tank 2 Transducer Receptacle
Connect the tank 2 transducer cable
to this receptacle.
Tank 1 Transducer Receptacle
Connect the tank 1 transducer cable to
this receptacle.
Connect a Valleylab electrosurgical footswitch to this receptacle. Use
only a Valleylab footswitch with the Valleylab Argon Gas Delivery Unit
II. Use of an incompatible footswitch may case unexpected output.
Do not plug the footswitch into the electrosurgical generator.
Equipotential Grounding Lug
Use an equipotential grounding
cable to connect the generator to
protective earth ground.
ESU Interface Cable Receptacle
Connect a Valleylab electrosurgical
generator to this receptacle using the ESU
interface cable provided by Valleylab.
Footswitch Receptacle
Valleylab Argon Gas Delivery Unit II Service Manual2-5
Chapter 3
Technical Specifications
Specifications subject to change without notice.
Technical Specifications
Valleylab Argon Gas Delivery Unit II Service Manual3-1
Performance Characteristics
Performance Characteristics
Standard Conditions of Measurement
Unless otherwise specified, all measurements are specified at an ambient temperature of
70° F (21° C) and 760 torr.
Source current normal polarity, intact chassis ground< 10 μA
Source current normal polarity, ground open< 50 μA
Source current reverse polarity, ground open< 50 μA
Sink current, all inputs < 20 μA
3-2Valleylab Argon Gas Delivery Unit II Service Manual
Performance Characteristics
Gas Flow Ranges
Low Gas Flow: 0.5 - 4 standard liters/minute ± 15% of the flow rate shown on the Flow
display.
Standard Gas Flow: 0.5 - 12 standard liters/minute ± 15% of the flow rate shown on the
Flow display.
Audio Volume
The alarm tones are set to a level of > 65 dBA at 1 m and are not adjustable.
Approximate Weight
24 lb. (10.9 kg)
Size
17.8 in. x 16.0 in. x 5.6 in. (45.2 cm x 40.6 cm x 14.2 cm)
Gas Tanks
Maximum Pressure: 2000 psi (13790 kPa) (138 bar)
Power Cord Specification
This unit was equipped from the factory with a 110 NEMA 5-15 power cord. Should the
AC power cord need to be replaced to match another plug configuration, the
replacement plug/cable/receptacle configuration must meet or exceed the following
specifications:
100 - 120 VAC
Cable - SJT16/3, IEC color code, maximum length 15 ft. (5 m)
Plug - minimum 10 A - 125 VAC
Unit receptacle - IEC female, minimum 10 A - 125 VAC
Technical Specifications
Valleylab Argon Gas Delivery Unit II Service Manual3-3
Standards and IEC Classifications
Standards and IEC Classifications
The Valleylab Argon Gas Delivery Unit II meets all pertinent clauses of the IEC 60601-1
second edition and IEC 60601-2-2 third edition.
ATTENTION: Consult accompanying documents.
Danger: Explosion risk if used with flammable anesthetics.
To reduce the risk of electrical shock, do not remove the cover. Refer servicing to qualified
service personnel.
Classified with respect to electrical shock, fire, mechanical, and other specified hazards
only in accordance with UL60601-1 and CAN/CSA C22.2 No. 601.1.
Type CF Equipment (IEC 60601-1) /Defibrillator Proof
The Valleylab Argon Gas Delivery Unit II unit provides a high degree of protection against
electric shock, particularly regarding allowable leakage currents. It is type CF isolated
(floating) output and may be used for procedures involving the heart.
This generator complies with the ANSI/AAMI HF18 specifications for “defibrillator proof”
designation and IEC 60601-2-2.
Liquid Spillage (IEC 60601-2-2 Clause 44.3)
The Valleylab Argon Gas Delivery Unit II is constructed so that liquid spillage in normal
use des not wet electrical insulation or other components which when wetted are likely
to adversely affect the safety of the equipment.
3-4Valleylab Argon Gas Delivery Unit II Service Manual
EMC Compatibility
Electromagnetic Compatibility (IEC 60601-1-2 and IEC
60601-2-2)
The Valleylab Argon Gas Delivery Unit II unit complies with the appropriate IEC 60601-12 and IEC 60601-2-2 specifications regarding electromagnetic compatibility.
Notice
The Valleylab Argon Gas Delivery Unit II should not be used adjacent to or stacked with operating
room equipment, unless otherwise specified in the Valleylab Argon Gas Delivery Unit II User’s Guide and Service Manual. The Valleylab Argon Gas Delivery Unit II should be used only in the
intended configuration.
The electrosurgical generator used with the Valleylab Argon Gas Delivery Unit II delivers RF energy
during activation. Observe other electronic medical equipment during activation of the Valleylab
Argon Gas Delivery Unit II for any possible adverse electromagnetic effects. Provide adequate
separation of this equipment to minimize the risk of adverse effects.
EMC Compatibility
The use of accessories, other than specified in the Valleylab Argon Gas Delivery Unit II User’s
Guide and Service Manual, may increase emissions or decrease immunity of the Valleylab Argon
Gas Delivery Unit II.
Technical Specifications
Valleylab Argon Gas Delivery Unit II Service Manual3-5
EMC Compatibility
Guidance and manufacturer's declaration - electromagnetic emissions
The Valleylab Argon Gas Delivery Unit II is intended for use in the electromagnetic environment
specified below. The customer or the user of the Valleylab Argon Gas Delivery Unit II should
ensure that it is used in such an environment.
Emissions test ComplianceElectromagnetic environment -
guidance
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Group 1The Valleylab Argon Gas Delivery Unit II
uses RF energy only for its internal
function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic
equipment.
Class AThe Valleylab Argon Gas Delivery Unit II
is suitable for use in all establishments
other than domestic and those directly
Class A
Complies
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
3-6Valleylab Argon Gas Delivery Unit II Service Manual
Guidance and manufacturer's declaration - electromagnetic immunity
The Valleylab Argon Gas Delivery Unit II is intended for use in the electromagnetic environment
specified below. The customer or the user of the Valleylab Argon Gas Delivery Unit II should
assure that it is used in such an environment.
EMC Compatibility
Immunity testIEC 60601 test
level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst IEC
61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
+/-6 kV contact
+/-8 kV air
+/-2 kV for power
supply lines
+/-1 kV for input/
output lines
+/-1 kV
differential mode
+/-2 kV common
mode
< 5% U t
(>95% dip in Ut)
for 0,5 cycle
4 0% Ut
(>60% dip in Ut)
for 5 cycles
7 0% Ut
(>30% dip in Ut)
for 25 cycles
< 5% U t
(>95% dip in Ut)
for 5 sec
Compliance
level
+/-6 kV contact
+/-8 kV air
+/-2 kV for power
supply lines
+/-1 kV for input/
output lines
+/-1 kV
differential mode
+/-2 kV common
mode
< 5% U t
( > 9 5 % d i p i n U t )
for 0,5 cycle
4 0% U t
( > 6 0 % d i p i n U t )
for 5 cycles
7 0% U t
( > 3 0 % d i p i n U t )
for 25 cycles
< 5% U t
( > 9 5 % d i p i n U t )
for 5 sec
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the Valleylab Argon Gas
Delivery Unit II requires
continued operation during
power mains interruptions,
it is recommended that the
Valleylab Argon Gas
Delivery Unit II be powered
from an uninterruptible
power supply or a battery.
Technical Specifications
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
NOTE: Ut is the a.c. mains voltage prior to the application of the test level.
3 A/m3 A/mPower frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Valleylab Argon Gas Delivery Unit II Service Manual3-7
EMC Compatibility
Guidance and manufacturer's declaration - electromagnetic immunity
The Valleylab Argon Gas Delivery Unit II is intended for use in the electromagnetic environment
specified below. The customer or the user of the Valleylab Argon Gas Delivery Unit II should
assure that it is used in such an environment.
Immunity
test
Conducted
RF IEC
61000-4-6
Radiated RF
IEC 61000-
4-3
IEC 60601
test level
3 Vrms
150KHz to
80MHz
3 V/m
80MHz to
2.5GHz
Compliance
level
3 V
7 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the Valleylab Argon Gas Delivery Unit II,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
√P
d=1.2
d=0.5
√P 80MHz to 800MHz
d=√P 800MHz to 2.5GHz
Where P is the maximum output power
rating of the transmitter in watts (W) according
to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than the
compliance level in each frequency range
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At a 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the Valleylab Argon Gas Delivery Unit II is used exceeds the
applicable RF compliance level above, the Valleylab Argon Gas Delivery Unit II should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the
Valleylab Argon Gas Delivery Unit II.
b. Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
3-8Valleylab Argon Gas Delivery Unit II Service Manual
Recommended separation distances between portable and mobile RF communication
equipment and the Valleylab Argon Gas Delivery Unit II
The Valleylab Argon Gas Delivery Unit II is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The Customer or the user of the Valleylab Argon
Gas Delivery Unit II can help prevent electromagnetic interferences by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the
Valleylab Argon Gas Delivery Unit II as recommended below, according to the maximum output
power of the communications equipment.
Separation distance according to frequency of transmitter (m)
Rated maximum
output power of
transmitter (W)
0.01.12 m0.05 m0.1 m
150 kHz to 80MHz
√P
d=1.2
80MHz to 800MHz
d=0.5√P
800MHz to 2.5GHz
d=√P
EMC Compatibility
0.1.38 m0.16 m0.32 m
11.2 m0.5 m1 m
103.8 m1.6 m3.2 m
10012 m5 m10 m
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At a 80MHz and 800MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Technical Specifications
Valleylab Argon Gas Delivery Unit II Service Manual3-9
Loading...
+ 78 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.