Vaisala PTB210 Technical data

PTB210 Digital Barometer
www.vaisala.com
The PTB210 paired with the SPH10 Static Pressure Head.
The Vaisala BAROCAP® Digital Barometer PTB210 is a reliable outdoor barometer that withstands harsh conditions.
For harsh environments
The Vaisala BAROCAP® Digital Barometer PTB210 is ideal for outdoor installations and harsh environments. The barometers are designed to operate in a wide temperature range, and the electronics housing provides IP65 (NEMA 4) standardized protection against sprayed water.
Features/Benefits
500 ... 1100 hPa or
50 ... 1100 hPa pressure ranges with serial output
Dierent scalings between
500 ... 1100 hPa with analog output
Electronics housing IP65
protected against sprayed water
Accurate and stable
measurement
NIST traceable
(certificate included)
The PTB210 barometers are ideal for use in applications such as weather stations, data buoys and ships, airports, and agrology. They are also an excellent solution for monitoring barometric pressure in industrial equipment such as laser interferometers and engine test benches.
Several pressure ranges
The PTB210 barometers are designed for various pressure ranges. They are available in two basic congurations: serial output for 500 ... 1100 hPa and 50 ... 1100 hPa and analog output with different scalings between 500 ... 1100 hPa.
Accurate and stable measurement
All the PTB210 barometers are digitally adjusted and calibrated by using electronic working standards. A higher accuracy barometer, that is ne-tuned and calibrated against a
High Precision Pressure Calibrator, is available for the 500 ... 1100 hPa pressure range.
In addition, the PTB210 integrates directly with Vaisala Static Pressure Head Series SPH10/20. This pairing offers accurate measurement in all wind conditions.
Vaisala BAROCAP® technology
The PTB210 barometers use the Vaisala BAROCAP capacitive absolute pressure sensor developed by Vaisala for barometric pressure applications. The Vaisala BAROCAP excellent hysteresis and repeatability characteristics and outstanding temperature and long-term stability. All PTB210 barometers are delivered with a factory calibration certicate which is NIST traceable.
®
®
Sensor, a silicon
Sensor provides
Technical data
Operating range (1hPa=1mbar)
Pressure range (order specified)
serial output 500 ... 1100 hPa
50 ... 1100 hPa
analog output 500 ... 1100 hPa
600 ... 1060 hPa
800 ... 1060 hPa
900 ... 1100 hPa Operating temperature range -40 ... +60 °C (-40 ... +140 °F) Humidity range non-condensing
Accuracy
SERIAL OUTPUT (units in hPa)
Pressure range 500 ... 1100 50 ... 1100
Class A Class B Non linearity* ± 0.10 ± 0.15 ± 0.20 Hysteresis* ± 0.05 ± 0.05 ± 0.10 Repeatability* ± 0.05 ± 0.05 ± 0.10 Calibration uncertainty** ± 0.07 ± 0.15 ± 0.20 Accuracy at +20 °C (+68 °F)*** ± 0.15 ± 0.20 ± 0.35 Temperature dependence**** ± 0.20 ± 0.20 ± 0.40 Total accuracy***
-40 ... +60 °C (-40 ... +140 °F) Long term stability (hPa/year) ± 0.10 ± 0.10 ± 0.20
ANALOG OUTPUT
Non linearity* ± 0.20 hPa Hysteresis* ± 0.05 hPa Repeatability* ± 0.05 hPa Calibration uncertainty** ± 0.15 hPa Accuracy at +20 °C (+68 °F)*** ± 0.30 hPa Temperature dependence**** ± 0.50 hPa Total accuracy*** -40 ... +60 °C (-40 ... +140 °F) ± 0.60 hPa Long term stability ± 0.10 hPa/year
* Defined as the ±2 standard deviation limits of end point
non-linearity, hysteresis error or repeatability error.
** Defined as ±2 standard deviation limits of inaccuracy of
the working standard including traceability to NIST.
*** Defined as the root sum of the squares (RSS) of end
point non-linearity, hysteresis error, repeatability error and calibration uncertainty at room temperature.
**** Defined as ±2 standard deviation limits of temperature
dependence over the operating temperature range.
± 0.25 ± 0.30 ± 0.50
General
(• Factory setting)
SERIAL OUTPUT
Current consumption
normal mode < 15 mA• power down mode < 0.8 mA
shutdown mode 0.2 mA Shutdown ON/OFF Settling time at power up 2 s Serial I/O (factory setting•) RS232C
RS232C /TTL (optional)
RS485, non isolated (optional) parity none, even•, odd data bits 7•, 8 stop bits 1•, 2
Baud rate 1200, 2400, 4800, 9600•, 19200 Response time 1 s• Resolution 0.01 hPa (1 measurement/s)
0.03 hPa (10 measurements/s)
ANALOG OUTPUT
Outputs 0 ... 5 VDC, 0 ... 2.5 VDC (order specified) Current consumption
normal mode < 8 mA shutdown mode 0.2 mA
Shutdown ON/OFF Response time 500 ms Resolution 300 µV Measurement rate 3 measurements/s
ALL MODELS
Supply voltage (reverse polarity protected)
with RS232/TTL output 5 ... 28 VDC with RS485 or analog output 8 ... 18 VDC
Max. pressure 5 000 hPa abs. Pressure connector M5 (10-32) internal thread Pressure fitting barbed fitting for 1/8" I.D. tubing Housing
electronics IP65 (NEMA 4) sensor IP53
Housing material PC plastic Supply/output cable length 1, 2, 3, 5 or 10 m
Instrument 110 g Cable 28 g/m
Electromagnetic compatibility Complies with EMC standard
EN61326-1, Generic Environment
For more information, visit www.vaisala.com or contact us at sales@vaisala.com
Ref. B210942EN-A ©Vaisala 2010
This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. The reproduction, transfer, distribution or storage of information contained in this brochure in any form without the prior written consent of Vaisala is strictly prohibited. All specifications — technical included — are subject to change without notice.
Loading...