2066 W. 11th Street Unit 1, Trinity Hall Farm Estate
Upland, CA 91786 Nuffield Road, Cambridge CB4 1TG UK
Phone: (800) 452-6788 Phone: +44(0)1223-420022
Fax: (909) 946-3597 Fax: +44(0)1223-420561
Web Site: www.uvp.com
81-0064-01 Rev M
UVX Radiometer 2
Table of Contents
General Instructions ...................................................................................................................................... 5
Calibration Information ............................................................................................................................ 15
General Precautions ................................................................................................................................ 19
Component List, Replacement Parts, and Options ..................................................................................... 29
Component List ....................................................................................................................................... 29
Replacement Parts .................................................................................................................................. 30
Your new UVX Digital Radiometer from UVP allows measurement of UV intensities quickly and easily.
The digital readout is in radiometric units with a broad dynamic range from 0.1 µW/cm
accuracy and precision traceable to the National Institute of Standards and Technology (NIST). The
sensors are cosine corrected for measuring UV radiation at 254, 310 and 365nm. Each sensor is also
pre-calibrated and quickly interchangeable with any other sensor without affecting the overall system
accuracy.
Designed for simplicity and durability, the UVX is built with the latest in solid state electronics, has a
rugged housing, over-ranging, and low battery indicato r s . By using circuitry that is powered by a single 9V
transistor battery, UVX is completely portable. It can be used in a wide variety of research and industrial
applications. Some typical applications using UV light sources are: genetic experiments, photochemical
reactions, sterilization procedures, optical lab measurements, experimental biology, dermatology
research, NDT techniques and graphics procedures.
The UVX Series Sensor supplied with this unit, as well as all other UVX Series Sensors, is designed and
calibrated for measurement of radiant incidences from line type and phosphor coated mercury arc
sources. There are several parameters including calibration, spectral response and sensor field of view
which users should understand to ensure accurate measurements. Refer to the Applications Techniques
section of this manual, which addresses these subjects.
Features
The following is a list of features provided by the UVX Radiometer and Sensor:
• Three sensitivity ranges covering the span from 200 µW/cm
• 3-1/2 digit liquid crystal display readout
• Sensitivity down to 0.1 µW/cm
2
• Outstanding ruggedness and reliability
• Negligible sens it i v it y to inf r a re d
• Linearity of
± 1% over the total range, ± one digit
•Readout and sensor are stored in their own storage and carrying case with room for
one additional sensor
• Light weight
• Powered by a single 9V transistor battery
• Long battery life (more than 120 hours of continuous operation with a single alkaline
battery)
• Automatic low voltage battery indication (”LOBAT” or “:” displayed on the LCD screen)
• Not sensitive to capacitive or AC pick up
• Fully interchangeable sensors for full UV band coverage
• Automatic circuit that tests the display each time the unit is turned on
• Read/Hold switch that allows holding any reading indefinitely
• Calibration of sensors is traceable to NIST through a standard lamp and proven
laboratory techniques
•Independent calibration of all sensors and digital Radiometers allowing complete
interchangeability without reference to individual calibration factors
•Remote sensor of small size permitting measurements to be taken in constricted
areas
•Three foot long electrically shielded connecting cable
1/4-20 UNC-2B threaded mounting hole compatible with standard tripod mount screws
•
2
to 20 mW/cm2
2
to 20 mW/cm2 with
UVX Radiometer 6
• Excellent cosine response with typical curve supplied in the manual
• A filter system which significantly reduces shortwave solarization phenomenon
• Internal temperature correction retains the accuracy of the sensors at both high and
low temperatures
• Externally accessible zero adjus t
• Connector port for an external chart recorder (recorder not included)
Optional Accessories
•A 10-to-1 attenuator allowing readings to be made up to 200mW/cm
2
.
Specifications
Radiometer
Conversion Rate: 2.8 readings/sec
Display: 3-1/2 digit LCG
Accuracy: ±2.5%
Linearity: ±0.2%
Resolution: 1 part in 1999
Sensitivity Ranges: 0 to 199.9 µW/cm
0 to 1999 µW/cm
0 to 19.99 mW/cm0 to 199.9 mW/cm
Temperature Coefficient: ±0.025%/°C ± 1 digit, 0 to 50°C
Zero Drift: ±0.02 uW/cm
2
2
2
2
w/10:1 Attenuator
2
/°C nominal 0 to 50°C
Power: 9V transistor battery
Battery Life (Alkaline): 120 Hrs (without recorder output option)
Sensor
Spectral Response: See Figures 6, 7, 8
Accuracy: ±5% (includes NIST standard)
Linearity: ±1.0%
Cosine Response: See Figure 9
Temperature Coefficient: ±0.04%/°C nominal, 0 to 40°C
Zero Drift: ±0.35 uW/cm
2
/°C nominal 0 to 50°C
Operating Environment
Temperature: 0 to 50°C
Humidity: 5% to 90% RH
Operating Procedures
The UVX Radiometer provides versatilit y and simplicity of operation. The num ber of front panel controls
has been held to a minimum. Range changes are made by means of a three position range switch.
Access to the zero adjust trimpot is available through the hole in the left hand side of the case. This
simplifies the operation of the instrument and also minimizes operator error.
The procedure that follows will help in understanding the operation of this instrument.
•Turn the Radiometer over and open the battery cover to determine if the battery has been
installed. If it is has not, s n ap the b at tery clip in place, insert the battery into the holder, then clos e
the battery cover.
•Plug the UVX sensor into the mating connector at the left rear of the UVX Radiometer. This
receptacle is identified in Figure 1.
Note: The connector goes t hrough two "snap " positions before it is fully inserte d. Failure to ins ert the
connector all the way will result in erroneous readings.
UVX Radiometer 7
•Place the three position range s witch in the 20 mW/cm 2 position. Place the READ/HOLD switch
in the READ position.
•Turn the unit on by placing the OFF-ON/TEST switch in the ON/TEST position.
Note: For the first second of operation, the entire display will be turned on demonstrating that all
segments, decim al points and “LOBAT” (the low battery indicator sho wn on the upper left c orner, for
2014 and newer models) on the display are f unctioning properly. T he readout you should see during
that one second period, if ever ythi ng is f unct ioning c orrectly, is “LOBAT” (for 2014 and newer models )
and “-18:8.8”. If any one of these s egments is not on during this time, refer to the Troubleshooting tips
in this manual.
NOTE: A low batter y will be indicat ed by “LOBAT ” text on the display on 2014 and newer models, or
a colon (“:”) for 2013 and older models.
After the one second test mode is completed, t he circuit will autom atically revert to the nor mal read
status.
•At this time, the meter should be zeroed to assure the best accuracy. This is done by first
covering the sensor s o that no UV light strik es the sens or windo w. (Genera lly, tur ning the sensor
upside down on a be nch will s uff ice.) T he range s witc h is the n plac ed in the 20 0 µW/cm
and the zero adjustment trimpot for a display reading of 00.0 µW/cm
2
.
2
position
•After setting the zero, ultraviolet readings may be taken by placing the sensor in the UV
environment. The resultant intensities will be shown on the display.
UVX Radiometer 8
CAUTION
Refer to the Safety Precautions section of this manual before making ultraviolet
measurements with thi s Radiometer. If all the digits are blank except for the
leading “1” when the sensor has been placed in the UV radiant incident field, an
overrange condition is indicated. This may be corrected by switching to a less
sensitive range or by moving the sensor away from the UV source. If this is
impractical, then the optional 10:1 Attenuator must be used.
If, however, the reading contains less than three significant digits, use a more
sensitive scale or move the sensor closer to the source.
To hold a reading after it has been obtained, change the position of the
HOLD/READ switch to the HOLD position. This will hold the reading until the
switch is returned to the READ position.
In the event the battery voltage falls below that necessary to ensure accurate
operation, “LOBAT” text will be shown on the display on 2014 and newer models,
or a colon (“:”) for 2013 and older models. When this happens, turn off the
Radiometer and replace the battery with a new one.
UVX Radiometer 9
Radiometer Details
Circuit Description
This section describes how the UVX Radiometer operates. Three illustrations are included for use in
conjunction with this section. Figure 2 is a Functional Block Diagram.
Current to Voltage Converter
The silicon detector used in the sensor may be modeled as a current source. For maximum linearity, the
detector load should represent a short circuit. The input amplifier on the board meets this requirement
and converts the input current to a voltage that is then applied to the analog to digital
switchable 4, 40, or 400:1 voltage divider in the amplifier feedback loop allows the current gain selection
required to produce a readout directly in the appropriate units of measure.
A/D Converter
A/D chip used in this circuit is a high performance, low power 3-1/2 digit converter, complete with
The
seven segment decoder, display drives, internal reference and a clock. This chip retains the high
accuracy of much more expensive units by providing an auto-zero to less than 10 uV, zero drift of less
than 1uV/°C, and roll over error of less than one count. The output of the chip drives the LCD directly.
Decimal Point Circuitry
The decimal point circuitry operates through a C-MOS chip to drive the appropriate decimal point as the
range is changed.
Low Battery Voltage Circuit
A low battery is detected by means of comparing a diode reference voltage to the trip-point of an
exclusive OR gate. When the battery voltage decreases to less than 7.0v, “LOBAT” text will be shown on
the display on 2014 and newer models, or a colon (“:”) for 2013 and older models.
HOLD/READ Circuit
The A/D chip is equipped with "hold" function pin which, when driven high (by the READ/HOLD switch),
holds the currently displayed reading.
ON/TEST Circuit
The ON/TEST circuit applies a one second DC signal to the back-plane of the LCD, forcing all the
segments to come on. After one second, the TEST portion of the circuit returns the unit to normal
operation.
Radiometer Circuit Calibration
This section refers to the calibration of the Radiometer, not of the sensor. UVP recommends recalibration
of the Radiometer ever y 6 months and provides calibration services for a nominal fee. Contact the factory
for details.
Note: Should the user attempt to perform a calibration on the Radiometer prior to the end of the warranty
period, then the warranty on the Radiometer becomes void.
Required test equipment for those who w is h to calibrate their own Radiometer:
• Variable current with range from .005uA to 1.0uA. Note: See Fig 3 for a calibration current source
• Precision microammeter measures current source output
• 500 K ohm resistor (±0.1 %)
• 5 to 9 V DC adjustable power supply (10mA or more)
• 0 to 10V DC digital voltmeter
• Small Phillips head screwdriver
• Small flat head screwdriver for trimpot adjustment
(A/D) converter. A
UVX Radiometer 10
Procedure
Calibration
•Remove the battery. Remove the two screws from the bottom of the case and carefully lift off the
bottom. As you do this, feed the battery clip through the aperture in the side of the battery holder.
Place the bottom of the case aside and lift the PC board out of the case top. Set the PC board on
the test bench.
• Be sure the OFF-ON/TEST switch is in the OFF position.
• Connect the DC power source to the plus and minus terminals of the battery clip. Note: The plus
side of the battery snap is the larger of the two. Set the voltage of the power supply at 9V DC.
•Connect the DC voltmeter ground connection to pin 40 of the A/D converter chip. Connect the
plus side of the DC voltmeter to pin 44 of the A/D converter chip.
•Turn the 9V DC power on and depress the ON/TEST switch. This will apply power to the circuit.
Measure the voltage between pin 44 a nd pin 40.
• If the voltage is not 1.000 volts DC, adjust RV3 until this indication is obtained.
• Disconnect the voltmeter from the circuit.
• Reduce the power supply voltage to 7.00V DC.
• While viewing the display, adjust trimpot RV2 until “LOBAT” (2014 and newer models) or a colon
“:” (2013 and older models) begins to appear on the display. Achieving this sets the low battery
voltage indication circuit.
•Check for proper low battery voltage indication by reducing the supply voltage to 6.8 volts and
see that “LOBAT” or the colon (“:”) is fully visible. Then increase the supply voltage to 7.20V DC
and see that “LOBAT” or the colon (“:”) is completely off.
• Return the supply voltage to 9.0 volts DC.
• Place the range selector switch in the 200 uW/cm
2
position.
• Adjust trimpot RV1 until the display shows 00.0
• This completes the calibration procedure.
Ranging
• Connect the current source to the input connector as shown on Figure 3.
• Set the current source for an output of 0.00500 uA.
• Set the range switch to the 200 uW/cm
2
position. The display should now read 100.00 ±2.5
uW/cm2
• Change the current to 0.0500 uA.
• Set the range switch to the 2000 uW/cm
uW/cm2.
•Change the current to 0.500 uA. Set the range switch to 20 uW/cm
10.00 ±25 uW/cm
2.
2
position. The display should now read 1000 ±25
2
. The display should now read
•This completes the test of the ranges of the un it.
Turn off the unit by depressing the OFF switch, disconnect the power supply and test equipment and
reassemble the circuit board back into the case. Test the equipment and reassemble the circuit board
back into the case.
Radiometer Maintenance
Care of Radiometer Case
The case of the UVX Radiometer is fabricated from a durable ABS plastic. As with all plastics, solvents
should not be used for cleaning the case. A high quality plastic cleaner or mild soap and water should be
used to clean fingerprints, dust or dirt from the case.
Note: Do not use abrasive cleaners on the Radiometer. The contacts for the sensor input and the
recorder output are recessed and should not require attention.
UVX Radiometer 11
Troubleshooting Procedures
In the event the Radiometer malfunctions, the following procedures will enable the user to determine the
source of error and affect repairs.
If any of the active circuit elements are replaced in the following proced ur es , then the calibration
procedure must be performed prior to putting the Radiometer back into service.
Troubleshooting Tips
The following paragraphs detail some of the more common malfunctions of the circuit and give the basic
procedures by which the cause of the circuit malfunction can be determined and ultimately corrected.
These procedures are limited to those active elements that can be easily removed and replaced.
Remove the circuit board from the case. Check the battery voltage to make sure it is still above 7.0 volts.
If it is, connect the battery back into the circuit. If it is not, replace it with a new battery.
Note: In the event any component is removed and replaced as a result of any work done on this unit,
then the full calibration procedure must be performed.
•Display segment does not turn on:
Turn the unit on and observe the display during the first second of operation: i.e. d ur ing test
mode. In this case the display should indicate “LOBAT” (for 2014 and newer models) and -18:8.8.
In the event any of the segments of the display is not on, it may indicate that a contact is not
making a firm connection at the liquid crystal display.
To correct this, use a pencil tip and place it on the pin contact of the malfunctioning segment.
(Refer to Figure 4 for the display pin-outs) Carefully work the contact pin back and forth with the
pencil tip and watch the segment. Performing this operation will generally correct all display
segment malfunctions.
•“LOBAT” (2014 and newer models) or c o lon (“:” on 2013 and older models) does not
appear on the display:
If “LOBAT” / “:” does not appear during the test mode, refer to the previous troubleshooting topic.
If “LOBAT” / “:” does not operate as a result of low battery voltage conditions and a new battery
has already been installed, then proceed as follows.
Disconnect the battery from the circuit and connect the circuit to a variable 0 to 10 volt DC power
supply. Set the power supply voltage at 9 volts. Turn the power supply and the unit on.
Slowly reduce the output of the DC power supply to 7.0 volts. If “LOBAT” / “:” does not appear,
then the trimpot RV2 is out of adjustment or the integrated circuit U2 is malfunctioning.
With the power supply set at 7.0 volts, adjust trimpot RV2 fully counterclockwise then fully
clockwise. If “LOBAT” / “:” does not come on at any time during this adjustment, then the
integrated circuit U2 is malfunctioning. Remove U2 and replace it with a new integrated circuit.
•Decimal point does not turn on:
Turn the unit on and place the range switch in the position in which the decimal point should be
on. Connect the negative input of the voltmeter to the negative side of the battery.
If the decimal point that should be on is the one corresponding to the 20 uW/cm
connect the positive side of the voltmeter to pin 9 of integrated circuit U2. If the decimal point that
is not on corresponds to the 200 uW/cm
2
range, connect the positive input of the voltmeter to pin
12 of integrated circuit U2.
When the range switch is in the appropriate position for the decimal point to be on, the voltmeter
should read the battery supply voltage; i.e., 9.0 volts. When the range switch is in some other
2
range, then
UVX Radiometer 12
position, then the voltmeter should read less than 5 volts DC.
If the voltmeter indicates correctly according to the above requirements and the decimal point
fails to come on, then the integrated circuit U2 is malfunctioning and must be replaced. If the
voltmeter does not indicate as described above, then the malfunction is in the range selector
switch and the UVX must be returned to UVP.
•Display does not turn on:
If the display fails to turn on immediately check the lead wires from the battery to make sure they
are not broken, also check the battery for an output of at least 7.0 volts. Also check the snaps on
the battery connectors. Be sure the snap fingers are pinched together for good contact when
snapped onto the battery terminals. If all these check out correctly, then integrated circuit U4 is
malfunctioning and must be replaced.
•Test mode does not operate:
Turn the unit on and momentarily connect a jumper between the plus side of the power supply
and pin 1 of integrated circuit U2. NOTE: Do not leave the jumper connected longer than a few
seconds. To do so may cause permanent damage to the Liquid Crystal Display.
During that time the test mode condition should be displayed on the readout. If the display shows
the test mode correctly, then the malfunction is in the ON/TEST switch and the unit must be
returned for repair. If it is not, then integrated circuit U2 is malfunctioning and must be replaced.
•Hold mode does not operate:
Turn the unit on and connect the DC voltmeter between the negative side of the battery and pin 8
of integrated circuit U4. Switch from READ to HOLD. The voltmeter should indicate the battery
voltage while the switch is in the HOLD position.
If the voltmeter indicates the battery voltage, and the display does not remain steady, then the
malfunction is in the integrated circuit U4 and it must be replaced. If the voltmeter does not
indicate the battery voltage, then the malfunction is in the switch and the unit must be returned to
UVP for repair.
•No display change with UV intensity changes:
Disconnect the sensor from the sensor input jack. Connect a 500K resistor between pin 2 of
integrated circuit U1 and the junction between resistors R1and R3. Connect a variable source
between pin 3 (-) and 2 (+) of integrated circuit U1. Set the current source for an output of .00500
micro amps.
Turn the unit on and set the range switch to the 200 uW/cm2 position.
Connect the negative side of a DC digital voltmeter to pin 3 of integrated circuit U1 and the
positive side to pin 6. Turn the unit on.
For normal operation, the display should read 100.0 uW/cm
volts. If the display is not indicating correctly or not at all and the voltmeter is indicating 1.00 volts,
then integrated circuit U4 is malfunctioning and must be replaced.
If the voltmeter is not indicating 1.00 Volts, then the input amplifier U1 is malfunctioning and must
be replaced.
Change the range switch to the 2000 uW/cm
volts and the display should read 100 uW/cm
2
position. The voltmeter indication should be 0.100
2
.lf the voltmeter did not change from 1.00 volts,
down to 0.100 volts, then the malfunction is in the range changing switch and the unit must be
returned to UVP for repair.
2
and the voltmeter should read 1.00
UVX Radiometer 13
Change the range switch to the 20 uW/cm2 position. The display should now indicate 0.10
uW/cm2 and the voltmeter should read .0100 volts Again, if the voltmeter indication did not
decrease by a factor of 10, then the malfunction is in the range changing switch and the unit must
be returned to UVP for repair.
In the event the input amplifier, U1, the range changing switch, and integrated circuit, U4, are
found to be functioning correctly, then the malfunction is in the UVX sensor.
Note: For direct traceability to NIST standards, the sensor must be returned to UVP for repair and
recalibration. Refer to the section in the manual on Sensor Details.
Figure 2: Radiometer Functional B lock Diagram
UVX Radiometer 14
Figure 3: Calibration Current Source
Figure 4: LCD Display Pin-outs
UVX Radiometer 15
Sensor Details
Sensor Characteristics
Solarization of Filters
Shortwave ultraviolet radiation causes a decrease in sensor output that is proportional to total exposure.
This is due to changes in the transmission properties of the filters; a phenomenon called solarization.
Solarization is especially noticeable in the 254nm sensors. Only a small amount is seen in the midrange
(310nm) sensors and virtually none in the longwave (365nm) sensors. In typical shortwave sensors
operating at 254nm, solarization begins immediately and reduces the sensor output at a rate of about
0.4% per mW-h/cm
that significantly reduces the short term solarization effect. This is shown in Figure 5 where it can be seen
that solarization does not begin to take effect until after 50mW-h/cm
rate of 0.1% per mW-hcm
Spectral Response Curves
Figures 6, 7 and 8
Sensors (respectively). In each case the absolute response curve is typical. Individual sensors may
exhibit variations in response at wavelengths distant from the calibration wavelength, but all are
accurately calibrated to a response of 1.00 at the specified calibration wavelength.
Sensor Temperature Coefficient
An added feature of the UVX Series Sensors is that special circuitry has been incorporated to reduce the
temperature induced errors present in all solid state detectors. The result is a sensor with a nominal
overall temperature coefficient of only ±0.04%/°C from 0 to 40°C.
Sensor Response Versus Incidence Angle
The UVX Series Sensors incorporate a unique diffuser at the incident surface. This diffuser enables the
sensor to detect radiation to incidence angles of ±81 from the normal. Figure 9 shows the real response
of the UVX Series Sensors, as compared to the perfect cosine response curve, with sufficient detail that
analytical corrections can be made where accurate measurements of large UV sources are required.
Calibration Information
Authorized Calibration Service
Please note this section refers to the calibration of sensors, not the Radiometer. UVP maintains a
calibration laboratory to recalibrate UVX Series Sensors. A nominal fee is charged for this service. To
verify such a recalibration, the date of the most recent calibration is placed on the label attached to the
sensor and a certificate of calibration is included with the returned sensor. It is suggested that the first
recalibration take place 6 months after the initial unpacking and use of the use of the unit, not 6 months
from the initial calibration date on the sensor. Periods of time spent sealed on the dealer's shelves have
negligible effect on the accuracy of the calibration.
2
. To correct this problem, the UVX shortwave (254nm) filters incorporate a new design
2
2
.
total dosage and then at the reduced
show the typical spectral response characteristics of the UVX 25, 31 and 36 Series
UVX Radiometer 16
Figure 5: UVX-25 Sensor Solarization Curv e
Figure 6: Typical UVX-25 Sensor Spectral Response
UVX Radiometer 17
Figure 7: Typical UVX-31 Sensor Spectral Response
Figure 8: Typical UVX-36 Sensor Spectral Response
UVX Radiometer 18
Figure 9: Typical Sensor Response Versus Inciden ce Angle
Calibration Traceability
For direct traceability to NIST standards, UVX sensors must be returned to the UVP calibrations
laboratory for calibration once every 6 months. UVP does not stand behind calibrations performed by
unauthorized companies.
UVX Series Sensors are calibrated by adjusting their response to match that of an UVP Transfer
Standard Detector when exposed to radiant incidence of the appropriate calibration wavelength from a
stable, effectively monochromatic source.
The calibration of UVP Transfer Standard Detectors is accomplished by measuring their response to a
stable, effectively monochromatic radiant source of the appropriate wavelength. The absolute radiant
incidence level from this source is set by simultaneously matching it with that of National Institute of
Standards and Technology (NIST) standard of spectral irradiance. This calibration is performed at UVP
calibrations laboratory a minimum of once every six months.
The NIST standard of spectral irradiance used by UVP is a 1000 watt tungsten lamp. This lamp is
replaced after every 20 hours of operation.
Period of Calibration
UVP recommends recalibration of the UVX Series Sensors every 6 months, especially when used by the
governmental agencies and in other areas of critical application.
NOTE: All sensors must be calibrated by the UVP calibrations laboratory for direct traceability to NIST
standards.
In non-critical applications, the user may choose a longer period of time between calibrations. However,
UVX Radiometer 19
the actual time that a sensor will remain within tolerance is dependent on such factors as care in handling
and total dosage of radiation on the sensor filters (refer to “Solarization of Filters” in this manual).
Sensor Maintenance
The user should avoid performing corrective maintenance on the sensor. If required, the UVX Sensor
should be returned to the factory.
Care of Sensor Case
CAUTION: Do not use abrasive cleaners on any part of the UVX Series Sensors or allow water to get
inside the sensor.
The case of the UVX sensors is fabricated from a durable ABS plastic. As with all plastics, solvents
should not be used in cleaning of the case. A high quality plastic cleaner or mild soap and water may be
clean fingerprints, dust or dirt from the case. The contacts on the cable connector should not require
attention.
Care of the Sensor Window
Dirt, dust or fingerprints on the sensor window will cause inaccurate readings. To ensure accurate
readings, it is essential that the sensor window be kept clean. There are several ways of ensuring
cleanliness. To remove dust or dirt, use a lens tissue or a cotton swab. To remove fingerprints, use a lens
tissue or lint-free cloth m oistene d with alc oh ol. A high pur it y alcoho l will ensure that no residue remains
after the cleaning procedure.
General Precautions
Handling
The UVX Series Sensors, although rugged, should be protected from physical shocks, moisture and
extremes of temperature.
Sensor Exposure
The sensors have been designed for minimum solarization effect as described elsewhere in this manual.
However, it is suggested that the sensor be exposed to ultraviolet radiation only for the amount of time
necessary to actually make a measurement. Excessive exposure may reduce the time between
recalibrations.
Sensor Continuity Measurements
When an ohmmeter is used to check continuity of the sensor, be careful not to use a meter that applies
more than 2 volts to the circuit. This is especially important when measuring continuity of the radiation
sensor itself; i.e., between the ring and sleeve of the connector. (See schematic, Figure 10).
UVX Radiometer 20
Figure 10: Schematic of UVX Series Sensor
Figure 1: Measurement of Irradian ce
UVX Radiometer 21
Figure 12: Typical Output From Low Pressure Mercury Lamp
UVX Radiometer 22
Application Techniques
Definition of Radiant Incidence
The UVX Radiometer is an instrument for measuring radiant incidence, where radiant incidence is defined
as the power per unit area incident on a surface at a location irradiated by a light source. See Figure 14.
Using the metric system of units, radiant incidence is dimensionally given as power per unit area. The
UVX Radiometer measures irradiance in milliwatts per square centimeter (mW/cm
square centimeter (µW/cm
power is incident on an area of one square centimeter located at the surface on the sensor.
Depending on the sensor used, the irradiance readings will only have "meaning" at a particular
wavelength or over a wavelength range and over a specified field of view.
2
). A reading of 10.0 mW/cm2 displayed on the UVX means that 0.010 watt of
Wavelength Considerations
Consider the wavelength of the light falling on the sensor. In the definition or irradiance, no restriction was
placed on the wavelength of light generating the irradiance. In practice, howev er, most sensor systems
are limited in sensitivity to a particular wavelength interval.
Several of the UVX sensors are designed to make irradiance measurements at particular wavelengths
from sources of special interest to the user. Figure 6 provides an example of the wavelength sensitivity for
a typical UVX-25 sensor. This sensor is calibrated at and designed to measure the 254nm wavelength
output from low pressure mercury lamps. Figure 12 shows a typical output from such a lamp. These light
sources are the most common type of shortwave emitters in use today. Although this source emits light at
several different wavelengths, the major output occurs at 254nm. In fact, this wavelength accounts for
approximately 90% of the total lamp output. Note that there are many spectral regions which yield no
output. An output of the type shown in Figure 13 is commonly referred to as a "line source".
In many applications, the irradiance produced by this 254nm radiation is important and needs to be
measured. The UVX Radiometer, using the UVX-25 sensor, is the answer to this measurement problem
because of two important factors. One is that negligible radiation, other than the 254nm wavelength,
occurs within the sensitivity wavelength region of the UVX-25 sensor. The second is that the sensor is
calibrated to yield a correct measurement for 254nm light. Thus, the Radiometer system yields an
accurate measurement of irradiance.
Consider now that the UVX-25 sensor is used to measure the output from the hypothetical light source
given in Figure 13. Here, the only output within the sensitive wavelength range of the sensor occurs at
270nm wavelength. The sensor, however, is more sensitive to radiation at 270nm than at 254nm
wavelength (see Figure 6). Therefore, the Radiometer system will yield a larger output. This output can
be corrected to yield a more accurate answer by multiplying the displayed output by the factor 0.54. The
correction factor was obtained by using Figure 6, and noting the ratio of the sensor sensitivity at 254nm to
the sensitivity at 270nm.
The UVX-36 is calibrated to accurately read the irradiance from 365nm line source (B-100 type) lamps.
When it is used to measure the irradiance from a phosphor coated long wave lamp it will read
approximately 65% of the true value so that the meter reading should be multiplied by 1/.65 = 1.54 to
ascertain the true irradiance.
Figure 14 illustrates how UVX-25 sensors are calibrated. Each of the curves in the family of curves
plotted represents slight differences in spectral sensitivity for different sensors. These small differences
result from variations in filter transmission and detector response. The important point to note, however, is
that all the curves are forced, as a result of calibration, to yield the same response at 254nm wavelength.
Sensors calibrated in this manner are said to be calibrated to a line source; in this case to a line source at
254nm wavelength.
2
) and microwatts per
UVX Radiometer 23
UVP manufactures other sensors which are also calibrated using line sources at other wavelengths. The
user must take special precautions when using "line source calibrated sensors" to measure irradiance
from lamps radiating at wavelengths other than those for which the sensor was calibrated. A common
application of this type occurs when measuring irradiance from phosphor coated lamps which emit
continuously over a broad wavelength region. Figure 15 shows an emitter.
When measuring a phosphor coated lamp a true absolute reading will only be obtained if the
measurement taken is multiplied by a correction factor. This correction factor can be determined by using
the following equation.
Field of View and Distance
Figure 16 shows two different sensors measuring the output from a lamp of extended length. The sensor
1
has a larger field of view or acceptance angle than the sensor S2. Since it is radiant incidence that is
S
being measured, sensor S
1
will yield a larger output than S2 because the emission from the total lamp
length is being sensed.
Remember it is power incident upon the sensor surface that is being measured. Obviously, some of the
power comes from the ends of the lamp as well as the middle and the entire lamp length. Therefore, it is
best to use sensors having a large field of view when measuring radiant incidence from extended
sources.
If this is not possible, then the sensor must be backed away until the entire source is within the field of
view of the sensor. Of course, this will reduce the overall radiant incidence on the sensor and
correspondingly the sensor output.
UVX Radiometer 24
Figure 13: Hypothetical Light Sourc e Having an Output at 270nm Wavelength
Figure 14: Response of Three Sensors Cali b ra ted to Yield the Same Output at 254nm Wavelength
For a sensor to accurately measure irradiance, it is necessary that it preserve a cosine relationship which
depends on the angle of incident radiation. This relationship is a function of the projected area of the
sensor, as seen in a plane normal to the incident light. See Figure 17.
Since the projected area is proportional to the power that the sensor intercepts, the irradiance measured
for radiation incident at an angle O will be given by 1
the sensor were pointed directly toward the incident region. Sensors that "weigh" their response in this
cos O. 1
0
is the irradiance that would be measured if
0
manner are said to be cosine corrected.
The Polar Plot in Figure 18 has been drawn from the curve of Figure 10 to show how the UVX sensor
compares to the perfect cosine response. The sensor is identified by the broken curve in the figure. The
solid curve, also shown, identifies a perfect cosine response. The UVX Series Sensors offer this same
high degree of cosine corrected response.
Summary
Many parameters affect radiant incidence measurements. These parameters can, however, be divided
into two classes; those which are a function of the sensor and those which are not. The most important
specifications of a sensor design have been discussed above. They are: spectral response, field of view,
distance and cosine response.
UVX Radiometer 26
Safety Precautions
It is essential that adequate precautions be taken in applications using ultra-violet light sources.
Prolonged exposures to high intensities can cause painful inflammation of the conjunctiva (the outer
membrane of the eye) as well as histological effects in the cornea, iris, and lens of the eye. Reddening or
even burning of the skin (erythema), similar to sunburn, will be caused by excessive exposure to ultraviolet energy. In extreme cases, permanent harmful effects can occur.
Individuals vary greatl y in their sensitivity to ultra-violet radiation. Children, f or example, are much more
sensitive to ultra-violet than are adults. Prolonged exposure of the unprotected skin and eyes should
always be avoided, and people with fair skins should avoid even oc casional exposures. The effects of
ultra-violet exposure may not be felt immediately. Normally, several hours will elapse before over
exposure is detected.
Recommended reading: "Criteria for a Recommended Standar d -Occupational Exposure to Ultra-Violet
Radiation", published by U.S. Department of Health, Education and Welfare, Public Health Service
Document HSM-73-11009, and "Practice for Occupational and Educational Eye and Face Protection",
published by the United States of America Standards Institute, Document USAS Z 87.1-1968.
These publications provide guidelines for maximum exposure to ultra-violet radiation in terms of time
intensity, and wavelen gth. The type of protecti on required to prevent over exposure to the eyes a nd skin
will vary depending on each of these parameters and each individual's susceptibility.
The wearing of protective eyewear, which great ly attenuates the transmission of ultra-violet radiation, is
always recommended when working in such environm ents. For low exposures, this prevent ive method
may be the only protection necessary.
The UVX Radiometer system is capable of measuring radiant incidence levels which are extremely
hazardous to unprotected personnel. In thes e cases (see rec ommended reading above), full face s hields
and protective clothing over all exposed areas of the body are required.
UVX Radiometer 27
Figure 16: Example, Illustrating Why Sensor Field of View is Important When Measurin g Irradiance
Figure 17: Projected Area of Sensor
UVX Radiometer 28
Figure 18: Cosine Response of Typical UVX Sensor (Polar Plot)
UVX Radiometer 29
PART NO.
DESCRIPTION
COMPONENT DESIGNATION
57-0005-01
Printed Circuit Board, UVX Radiometer
57-0041-10
Resistor, 4.99K, 1/8W, 1%
R1
50-0041-21
Resistor, 536 ohms, 1/8W, 1%
R2
50-0041-22
Resistor, 191 K, 1/8W, 1%
R3
50-0041-12
Resistor, 16.2K, 1/8W, 1%
R4
50-0025-05
Resistor, 1M, 1/4W, 5%
R5, R10, R11
50-0041-13
Resistor, 475K, 1/8W, 1%
R6
50-0041-14
Resistor, 10K, 1/8W, 1%
R7
50-0041-15
Resistor, 113K, 1/8W, 1%
R8
50-0041-16
Resistor, 51.1 K, 1/8W, 1%
R9
50-0025-07
Resistor, 270K, 1/4W, 5%
R14
50-0025-04
Resistor, 10K, 1/4W, 5%
R13
50-0041-19
Resistor, 1.2M, 1/4W, 5%
R12
50-0041-20
Resistor, 2.4M, 1/4W, 5%
R15
52-0011-01
Capacitor, .002uf 1000V, 20%
C2
52-0004-06
Capacitor, .1 uf, 12V, 20%
C3, C5, C8, C11, C12
52-0002-01
Capacitor, .22uf, 100V, 10% Mylar
C4
52-0002-02
Capacitor, .22uf, 100V, 10% polypropylene
C6
52-0012-01
Capacitor, .047uf, 250V 10%
C7
52-0004-07
Capacitor, .1 uf, 100V 10%
C9
52-0013-01
Capacitor, 100pf, 500V, 5%
C10
52-0002-03
Capacitor, .22uf, 50V, ± 80% to -20%
C13
52-0001-02
Capacitor, 10uf, 16V, -10% to +50%
C15, C16
51-0007-02
Diode, 1N914
CR1, CR2
51-0002-02
Rectifier, 1 Amp, 1N4001
CR3
51-0023-02
Transistor, 2N3904
Q1
50-0012-08
Potentiometer, Trimpot, 20K, 15 Turns
RV1
50-0012-09
Potentiometer, Trimpot, l00K, 15 Turns
RV2
50-0012-05
Potentiometer, Trimpot, 10K, 15 Turns
RV3
53-0014-05
Switch, Rocker, PCB, 5A. 125V
S1
53-0052-02
Switch, Rocker, PCB 28 VDC DPDT
S2
53-0052-01
Switch, Rocker, PCB 28 VDC SPOT
S3
51-0029-01
Integrated Circuit, OP97
U1
51-0028-01
Integrated Circuit, 4030
U2
51-0030-01
Converter, A/D, 3-1/2 Digit Sin gl e Chip LCD
U4
55-0064-01
Terminal, Circuit Socket, 20 Pin
51-0031-01
Display, Liquid Crystal, 3-1/2 Digit, .5" High
D1
55-0063-02
Connector, Female, Printed Circuit Board
J1
55-0066-01
Battery Snap, I Type, 6” Leads
Component List, Replacement Parts, and Options
Component List
The following list is the components used on the printed circuit board assembly P/N 57-0005-04.
UVX Radiometer 30
PART NO.
DESCRIPTION
COMPONENT DESIGNATION
52-0004-06
Capacitor, .1 µ, f, 12V, 20%
(C14)
55-0063-01
Phone Jack, Enclosed, Subminiature
(J2)
51-0029-01
Integrated Circuit, OP97
(U3)
57-0005-02
PC Board Assembly
PART NO.
DESCRIPTION
COMPONENT DESIGNATION
51-0023-02
Transistor, 2N3904
(Q1)
51-0029-01
Integrated Circuit. OP97
(U1, U3)
51-0028-01
Integrated Circuit, 4030
(U2)
51-0030-01
A/D Converter, 3-1/2 Digit Single Chip
(U4)
51-0031-01
Liquid Crystal Display
(D1 )
In addition to the components listed above, the UVX that includes the Recorder Output Option (57-0005-
05) also has the following components.
Replacement Parts
Replacement parts may be ordered from UVP or your electronics component supplier. It is important
when ordering parts to include in your order the part number, instrument type, and serial number.
Following is a list of replacement parts required for the UVX Radiometer.
Note: There are no replacement parts for the UVX sensors. In case of damage or malfunction, the sensor
must be returned to UVP for repair and recalibration.
Options
A number of options are available for use with the UVX Radiometer. These options expand the functions
of the UVX as well as provide gre ater user app lica tio n and ada ptab il ity to specific requirements.
Any of these options may be ordered separately or at the time of purchase. They are:
UVP's quality instruments are guaranteed to be free of defects in materials, workmanship, and
manufacture for one (1) year from date of purchase. Consumable and disposable products, including, but
not limited to lamps or light tubes, filters, or rechargeable batteries, are guaranteed to be free from
detects in manufacture and materials for ninety (90) days from date of purchase. If equipment failure or
malfunction occurs during the warranty period, UVP shall examine the inoperative equipment and have
the option of repairing or replacing any part(s) which, in the judgment of UVP, were originally defective or
became so under conditions of normal usage and service.
No warranty shall apply to any instrument, or part thereof, that has been subject to accident, negligence,
alteration, abuse or misuse by the end-user. Moreover, UVP makes no warranties whatsoever with
respect to parts not supplied by UVP or that have been installed, used and/or serviced other than in strict
compliance with the instructions appearing in the operational manual supplied to the end user.
In no event shall UVP be responsible to the end-user for any incidental or consequential damages,
whether foreseeable or not, including, but not limited to property damage, inability to use equipment, lost
business, lost profits, or inconvenience arising out of or connected with the ·use of instruments produced
by UVP. Nor is UVP liable or responsible for any personal injuries occurring as a result of the use,
installation and/or servicing of equipment.
Warranty is void if calibration is performed by any company other than UVP prior to the end of warranty
period.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.