UV/Air Circulator
For the UV PCR Workstation
User Instructions
UVP, LLC Ultra-Violet Products Ltd
2066 W 11th Street, Upland, CA 91786 Unit 1, Trinity Hall Farm Estate
Tel: (800) 452-6788 / (909) 946-3197 Nuffield Road, Cambridge CB4 1TG UK
Fax: (909) 946-3597 Tel: +44(0)1223-420022 / Fax: +44(0)1223-420561
Web Site: www.uvp.com
81-0228-02 Rev C
Introduction
The UV/Air Circulator Module is designed to mount into the UV PCR Workstation to eliminate
reduce contamination within the workstation during experiment processing without exposing
samples. Connections for mounting and powering the module are built into the workstation.
Features of the UV/Air Circulator:
Easy to install, lightweight design
High intensity 15 watt shortwave UV
Recirculating decontaminating airflow
Antimicrobial coated stainless steel to reduce bacterial growth
Access port for measuring the UV intensity with a UV sensor
Easy access for UV tube replacement
Safety Information
Caution: The unit includes shortwave UV, which is a powerful source of UV radiation that will
cause damage to unprotected eyes and skin. Before operation, ensure person nel in the area are
protected and that instructions for use of this equipment are followed.
Specifications
Dimensions (W x H x D):
Light Source:
Construction:
Electrical:
26” x 7” x 4” (660 x 177.8 x 101.6)
One 25 watt, 254nm shortwave tube
Formed antimicrobial coated stainless steel
115V / 60 Hz 230V / 50 Hz 100V / 50/60 Hz
Installing the UV/Air Circulator Module
Carefully remove the unit from the packaging
materials. Turn off and unplug the PCR
workstation from the wall power supply prior
to installing the UV/Air Circulator.
Locate the two mounting brackets on the
PCR workstation. Open the door of the PCR
workstation to a fully open position.
Lift the module into the workstation and set it
onto the brackets.
Secure the module to the workstation with
the thumbscrews.
Connect the jumper power cable from the
circulator to the PCR Workstation.
Reconnect the main power source.