UVP Ultraviolet PCR User Manual

Page 1
Ultraviolet PCR Workstations and Cabinets
Installation and User Instructions
UVP, LLC Ultra-Violet Products Ltd. 2066 W 11th Street, Upland, CA 91786 Unit 1, Trinity Hall Farm Estate, Tel: (800) 452-6788 / (909) 946-3197 Nuffield Road Cambridge CB4 1TG UK Fax: (909) 946-3597 Tel: +44(0)1223-420022 / Fax: +44(0)1223-420561
Web site: www.uvp.com

81-0228-01 Rev Y

Page 2

Introduction

95-0367-01 115V
95-0367-03 100V
Light Sources:
Chamber
Features include:
block below 400nm
Features include:
below 400nm
Features include:
block below 400nm
Features include:
below 400nm
Dimensions:
mm)
The Ultraviolet PCR Systems create an ideal environment for preparing PCR and other samples by reducing the chance of contamination. The built-in high intensity UV provides a source for inactivation of DNA between experiments. Additional contamination control is provided by the uniquely coated stainless steel and aluminum design that maintains antimicrobial efficacy. The surface resists growth of destructive bacteria, molds and fungi.
An optional UV/Air Circulator module connects to the interior of the workstation (95-0367-xx only) . The module eliminates amplicons of DNA or genomic DNA from a previously dispensed experiment. Connections for mounting and powering the module are built into the workstation. For ordering information, refer to the “Maintenance, Replacement Parts/Accessories” section.

Safety Information

The Ultraviolet PCR Systems are designed with function, reliability, and safety in mind. Caution: The unit includes shortwave UV, which is a powerful source of UV radiation that will cause damage to unprotected eyes and skin. Before operating any unit, ensure all personnel in the area are properly protected and that instructions for use of this equipment are followed. A safety shut-off switch automatically turns the UV light off when the door is open, protecting users from UV exposure. The Makrolon 32P labeled compounds.
®

Specifications

enclosure is UV and beta-blocking and effec tive for protection against
Ultraviolet PCR
Cabinet
95-0437-01 115V 95-0437-02 230V UK 95-0437-04 230V Euro 95-0437-03 100V
Light Sources:
254nm 8-watt UV:
Chamber
White 8-watt
Chamber
Two power outlets Two small shelves UV Timer Antimicrobial coated
stainless steel (work surface) & aluminum
®
Makrolon
panels
STANDARD UV PCR MODELS
UV2 PCR Cabinet
95-0436-01 115V 95-0436-02 230V UK 95-0436-04 230V Euro 95-0436-03 100V
Light Sources:
254nm 8-watt UV:
Chamber UV/air
circulator
White 8-watt:
Chamber
Two power outlets Two small shelves UV Timer Antimicrobial coated
stainless steel (work surface) & aluminum
Makrolon® panels block
Ultraviolet PCR
Workstation
95-0367-02 230V UK 95-0367-04 230V Euro
Light Sources:
254nm 25-watt UV:
Chamber
White 15-watt:
Chamber
Four power outlets Two shelves UV Timer Antimicrobial coated
stainless steel (work surface) & aluminum
Makrolon® panels
UV2 PCR
Workstation
95-0439-01 115V 95-0439-02 230V UK 95-0439-04 230V Euro 95-0439-03 100V
254nm 25-watt UV:
Chamber
254nm 8-watt UV:
UV/air
circulator
White 15-watt:
Four power outlets Two shelves UV Timer Antimicrobial coated
stainless steel (work surface) & aluminum
Makrolon® panels block
Dimensions:
Exterior: 28.7H x
21.4W x 24D in. (729 x 544 x 610 mm) Interior: 19.7W x
21.4D in. (500 x 544 mm)
Exterior: 28.7H x
21.4W x 24D in. (729 x 544 x 610 mm) Interior: 19.7W x
21.4D in. (500 x 544
Dimensions:
Exterior: 28.7H x 29W x 24D in. (729 x 737 x 610 mm) Interior: 27.8W x
21.4D in. (706 x 544 mm)
Dimensions:
Exterior: 28.7H x 29W x 24D in. (729 x 737 x 610 mm) Interior: 27.8W x
21.4D in. (706 x 544 mm)
Page 3
Ultraviolet PCR Cabinets and Workstations Page 3
Design
Formed antimicrobial coated aluminum, qty 2
Antimicrobial coated stainless steel (horizontal work surface) and aluminum
Door and Side Panels
Two panel assembly, Makrolon® 115V
60 Hz
230V
50 Hz
100V
50/60 Hz
Part
Voltage
Max
95-0437-01
95-0439-05
115V
230V
9.1
8.7
Shelves
System Design
Interior Exterior Aluminum powder coated
Electrical

Operational Ratings

Unit intended to be used indoors
Altitude must not exceed 2000 m
Ambient temperature not to exceed 5°C to 40°C (41°F to 104°F)
Relative humidity not to exceed 80% for temperatures up to 31°C (88°F) decreasing linearly to
50% relative humidity at 40°C (104°F)
Mains supply voltage fluctuations up to ±10% of the nominal voltage
Pollution rating 2 or better for laboratory areas
blocks wavelengths below 400nm
Number
95-0437-02 95-0437-04 95-0437-03 95-0437-05 95-0436-01 95-0436-02 95-0436-04 95-0436-03 95-0436-05 95-0367-01 95-0367-02 95-0367-04 95-0367-03 95-0367-05 95-0439-01 95-0439-02 95-0439-04 95-0439-03
230V 230V 100V 230V 115V 230V 230V 100V 230V 115V 230V 230V 100V 230V 115V 230V 230V 100V
Amps
8.8
8.8
9.1
8.8
9.1
8.7
8.7
9.1
8.7
8.8
8.6
8.6
8.8
8.6
9.1
8.7
8.7
9.1
Page 4
Ultraviolet PCR Cabinets and Workstations Page 4
Screw
Screw
Base
Outer back panel. Note the notch in

Assembling the PCR System

Carefully remove the unit from the packaging materials. Follow the assembly instructions provided in this manual. NOTE: Place the PCR base on a level surface large enough to accommodate the entire base. The surface must be level to ensure proper ali gn ment of the Makrolon® panels.
Tools Required
Philips head screwdriver
Flathead screwdriver
Gloves, to reduce transfer of fingerprints to PCR components
Two people are recommended for assembly or lifting of this equipment. NOTE: Assembly drawings show the 95-0436-01 UV2 PCR Cabinet. These instructions apply for all PCR
models.
Outer Back Panel
NOTE: The outer and inner back panels and shelf bracket are assembled at the factory for shipping purposes. The metal Makrolon® frames (2) are taped to the panels. Disassemble the back panels/shelf bracket and remove the protective film prior to completing the assembly instructions below.
1. Attach the outer back panel to the base with two screws as shown below.
Do not tighten the screws at this time.
the top of the panel allows room for
the wires.
Page 5
Ultraviolet PCR Cabinets and Workstations Page 5
Inner back panel
Bracket
Position the inner
Cutout for wires
Inner Back Panel
1. Position the wires so that they run between the inner back panel and the outer back panel up to the top for later connection to the top assembly.
2. Position the inner back panel so that the screw holes are aligned with the outer back panel. Attach the inner back panel to the outer back panel with four (4) screws. Make sure the wires come out of the cutout in the back panel, as shown by the red circle below.
3. Secure the shelf bracket in front of the inner back panel using:
a. Eight (8) screws in the Workstation or b. Six (6) screws in the Cabinet
NOTE: Two people are required to hold the panels in place and insert the screws. Do not fully tighten screws at this time.
panel so that the
edge of the frame
rests against the
stainless steel b ase
Page 6
Ultraviolet PCR Cabinets and Workstations Page 6
Frames for
Makrolon® Panels
Shelf Bracket
Makrolon® Panel Assembly
1. Loosen the screws (4 per side) attaching the Makrolon® panel frames to the inner back panel.
2. Slide the frames for the Makrolon frames are under the shelf brackets.
3. Remove the protective film from the Makrolon brackets and the inner back panel.
4. Press the bracket s tightly against the Makrolon
5. Tighten all frame screws to secure the Makrolon
Makrolon® Panels
®
panels under the screws on the inner back assembl y s o th a t the
®
panels. Slide the panels into position between the
®
panels.
®
panels in place.
Page 7
Ultraviolet PCR Cabinets and Workstations Page 7
WARNING: Ensure that the unit is not plugged in.
Flange
Screw locations
Shelf bracket
Top Assembly
Top Assembly and Shelf Assembly
1. Rest the top assembly on top of the unit.
2. Connect the bottom and top assembly wires together according to their labels, matching wire “A” with wire “A,” wire “B” with wire “B” and so forth, until all labeled wires are connected. Connect the wires by joining the male and female red connectors located on the ends of each wire. Then, connect the two unlabeled green/yellow ground wires together using the red connectors.
3. Slide the top assembly down so that the Makrolon Make sure that the back flange of the top assembly is positioned between the inner back and the outer back panel. Slide the wires into the space between the inner and outer back panels. Secure and tighten the top assembly to the back panel with screws, pushing from the back to ensure the screws are secure.
4. Slide the shelves into the shelf bracket.
®
panels are inside the top assembly channels.
Page 8
Ultraviolet PCR Cabinets and Workstations Page 8
Thumb Knobs
Door Assembly
Side Brackets
Magnet
Top Assembly
Door Hinges
Door Assembly
1. Slide the upper panel of the door assembly into slotted area in the top assembly. Holding the door in position, insert the thumb screws (Qty. 3) inside the top assembly.
2. Secure the upper panel of the door assembly to the left and right side panels with the two side brackets.
3. Create the door assembly by connecting the upper and lower door panels using the two hinges and 8 nuts.
4. A magnet is attached to the bottom right bracket on the door. The purpose of the magnet is to shut off the overhead UV light when the door is open.
Note: Some systems may have a single large bracket running along the width of the entire lower door, rather than the two small brackets shown below.
WARNING: The PCR is for laboratory use and cannot be connected to a public power supply.
When the PCR is assembled, plug the power cord into the back of the unit and then into a wall outlet in the laboratory.

Operating the PCR Systems

The Ultraviolet PCR Systems feature:
Overhead UV for decontamination of the chamber or PCR equipment
UV safety switch and keylock for preventing exposure to UV
Antimicrobial coated stainles s stee l and aluminum for reducing bacterial growth
Easy-access door
Overhead white light
Shelves for placement of tools
Built-in power outlets built into the chamber interior
Page 9
Ultraviolet PCR Cabinets and Workstations Page 9
Power switch
Overhead white light
UV keylock
115V
230V UK
230V Euro
30-minute
UV timer
Timer

Use of PCR Equipment inside the Cabinet or Workstation

Shelves are provided for placement of small tools for decontamination. Power outlets allow use of shakers, rockers, and other equipment for PCR experiments inside the
PCR chamber. The equipment can be decontaminated between experiments. Power plugs built into the chamber allow use of shakers, rockers, and other equipment for PCR
experiments. Equipment can be decontaminated between uses. To operate equipment within the chamber, plug the equipment into the power plug. Specific power plugs are installed as shown in the illustrations below.
NOTE: When operating equipment plugged into the outlets, combined Amps total is not to exceed 8 Amps.
NOTE: Crazing is a normal process for Makrolon
over a period of time. Reduce crazing by keeping exposure to UV to a minimum. Crazing may occur within the warranty period and is regarded as normal wear and tear not covered by the warranty. Makrolon® panels can be replaced. Refer to Replacement Parks for ordering information.

Operating the UV and Fluorescent White Lights

UV Keylock
The Ultraviolet PCR Cabinets and Workstations provide an automated process for eliminating contamination by utilizing the germicidal properties of shortwave 254nm ultraviolet light. The UV light is typically operated when no samples are inside the chamber. The chamber is decontaminated, eliminating viable fungi, bacteria, and yeast.
for the Optional
UV/Air
Circulator
To use the UV decontamination, turn the UV keylock ON. Then press the timer to operate the germicidal lamp housed within the chamber. The UV timer allows selection of UV exposure time. The timer may take a few seconds to initialize once the timer has power. The timer will only operate when the door is shut and the magnetic safety switch on the door is engaged. When door is open the timer will not operate.
®
panels exposed to UV light. Crazing will occur
calibration
button hole
Page 10
Ultraviolet PCR Cabinets and Workstations Page 10
UV2 PCR Cabinet and Workstation. Airflow
The default timer setting is 30 minutes, adjustable at 5 minute increments from 5 to 60 minutes, then at 15 minute increments up to 12 hours. To adjust the timer settings:
1. Turn the system on and wait until the timer reads “OFF”.
2. Use a small paper clip to press the Timer calibration button loc ated in t he small hole to the left of the timer for one second.
3. Press the main timer button to adjust the timer settings. Each press will increase the amount of time by 5 minutes (eg 35:00, 40:00, etc.) until one hour is reached. After one hour is reached, the time will be adjusted in 15-minute increments and will read without a colon. For example, one hour will read “1 00,” one hour and 15 minutes will read “1 15” and so forth.
4. Once the desired time setting is reached, do not press any additional buttons. The timer will then cycle through the remainder of its settings. When the timer reads “OFF” for 5 seconds without blinking, the timer is ready to use with the new time setting.
Turn the key to the off position after UV irradiation to prevent exposure of samples to the UV. Ensure the key is completely inserted into the lock.
[Note] Do not attempt to perform PCR procedures with the germicidal lamp on. The germicidal lamp is used to decontaminate the internal chamber of the between experiments.
[Warning] Exposure to UV light is harmful. The Makrolon
germicidal lamp will shut off if the panel door is opened. The PCR Workstation contains a powerful source of UV radiation that will cause damage to unprotected eyes and skin. Before operating any unit, be sure all personnel in the area are properly protected. Even though the unit shuts the UV off when the door is open, UV Blocking Eyewear should be worn as well. UVP has a complete line of UV Blocking Eyewear: Spectacles, Goggles, and Faceshield designed for this purpose.
Extended periods of time UV exposure can be used to disinfect the interior chamber by the destruction of organisms. Refer to the Maintenance section for information on measuring the intensity of the UV tubes with a UV meter.
Overhead White Light
The overhead white light can remain lit at any time throughout experiment procedures or the decontamination process. The fluorescent white light provides a bright light within the PCR Workstation. To use the white light, press the ON button located on the top panel.
UV Air C irculator
The UV2 PCR Cabinet and UV2 PCR Workstation inclu de a built-in UV Air Circulator.
To operate the circulator, press the UV/Air Circulator switch to the ON position. The circulator will pull air from inside the chamber through the circulator, decontaminating the air with shortwave UV and blowing the air back into the chamber.
NOTE: Do not place equipment in front of the air vents to block air flow.
An optional UV/Air Circulator module can be mounted into the top of workstation chamber (95-00367-XX only) for elimination of amplicons of DNA or genomic DNA from a previously dispensed experiment. The power switch, outlet, and mounting brackets for the circulator are pre­installed at the factory if users purchase a module at a later time.
®
door is UV blocking and will not allow UV radiation to pass through the door. The
into and out of the UV Air Circulator
Page 11
Ultraviolet PCR Cabinets and Workstations Page 11
Microwatt
seconds/cm
Microwatt
seconds/cm2
Saccharomyces ellipsoideus
B. parathyphosus
6100
Brewer’s yeast
6600
B. subtilis
11000
Baker’s yeast
8800
Corynebacterium diptheriae
6500
MOLD SPORES
Escherichlia coli
6600
Penicillium expansum
22000
Micrococcus cadidus
12300
Penicillium digitatus
88000
Micrococcus sphaeroides
15400
Aspergillus glaucus
88000
Proteus vulgaris
6600
Mucor racemosus A
35200
Pseudomonas aeruginosa
10500
Mucor racemosus B
35200
Salmonella
10000
VIRUS
Sarratia marcescens
6160
Tobacco mosaic
44000
Dysentery bacilli
4200
Influenza
6600
Shigella paradyseneriae
3200
PROTOZOA
Staphylococcus aereus
6600
Nematode eggs
9200
Streptococcus lactis
8800

Measuring the UV Tube Intensity

As the ultraviolet tubes age, the intensity and germicidal destruction rate decreases. It is important to monitor the efficiency of the tubes to ensure the germicidal requirements are met. The germicidal destruction rate is a function of the UV intensity at wavelength 254nm and exposure time. The lower the tube intensity, the longer the lamps must be on to accomplish the same objective.
The UVX Radiometer using a 254nm sensor will allow the user to measure the 254nm emissions from the tubes. When it is time to measure the UV intensity inside the PCR workstation (recommended weekly or as needed), the sensors can be placed on the floor of the PCR. Close the hood and turn on the UV tubes. Wait three minutes to record the measurement on the UV meter. It is recommended that the user record an initial value upon receipt and assembly of the PCR unit.
The germicidal destruction rate calculation:
Microwatt seconds/cm2 = microwatts/cm2 x seconds of exposure
For ordering information on the UVX Radiometer, J-225 meter and sensors, refer to the Replacement Parts and Accessories section in this manual.
The bacterial destruction chart below indicates the amount of shortwave (254nm) UV energy required for complete destruction of various organisms.
Bacterial Organisms
Bacillus anthracis 8700 S. enteritidis 7600 B. Megatherium sp. (veg.) 2500 Saccharomyces sp. 17600
B. Megatherium sp. (spores) 5200 Saccharomyces cerevisiae 13200
Additional Organisms
2
YEAST
13200
B. subtilis spores 22000 Common yeast cake 13200 Clostridium tetani 22000
Eberthella typosa 4100 Penicillium roqueforti 26400
Mycobacterium tuberculosis 1000 Aspergillus flavus 99000 Neisseria catarrhalis 8500 Aspergillus niger 330000 Phytomonas tumefaciens 8500 Rhisopus nigricans 220000
Pseudomonas fluorescens 6600 Oospora lactis 11000 S. typhimusium 15200
Sarcina lutea 26400 Bacteriophasge (E. coli) 6600
Spirillum rubrum 6160 Staphlococcus albus 5720 Paramecium 200000
Streptococcus hemolyticus 5500 Chlorella vulgaris (algae) 22000
Streptococcus viridans 3800
Page 12
Ultraviolet PCR Cabinets and Workstations Page 12
System/Part Number
Replacement Tube
Tube Description
Qty.
UV PCR Cabinet
34-0007-01
8W 254nm UV
4
95-0437-xx
34-0056-01
8W White
2
UV2 PCR Cabinet
34-0007-01
8W 254nm UV
5
95-0436-xx
34-0056-01
8W White
2

Maintenance, Replacement Parts/Accessories

NOTE: Crazing is a normal process for Makrolon® panels exposed to UV light. Crazing will occur over a period of time. Reduce crazing by keeping exposure to UV to a minimum. Crazing may occur within the warranty period and is regarded as normal wear and not covered by the warranty. Makrolon

Care and Cleaning

The door and side panels are made of Makrolon® and are subject to scuffing and scratches if improperly cleaned. The stainless steel and aluminum surfaces are manufactured with an antimicrobial coating to reduce bacterial growth. Care in cleaning and use of the equipment is recommended to reduce wear of the coating. To clean the unit:
®
panels can be replaced. Refer to Replacement Parks for ordering information.
Wipe excess water from inside the unit and outside the unit with an absorbent soft cloth or sponge.
Use mild soap and a damp soft cloth or damp sponge to clean the exterior of the unit.
Use a mild detergent on the interior surfaces.
Warning: Unplug the unit before cleaning around the UV tube contacts and internal plug
outlets.
Clean the door with a mild detergent; never use organic based compounds, Alcohol, or Ammonia containing cleaners.
Do not use abrasive pads or cleansers.
A plastic cleaner solution is recommended to clean the door and is available from local
plastic supply distributors.

Replacement Parts and Accessories

Replacement parts are listed below. Only authorized UVP service personnel should perform repairs or replacements other than specified in these procedures.
Replacement Part Part Number
Key, replacement 53-0178-04 Shelf (for workstations) 20-0970-02 Shelf (for cabinets) 20-1209-01 Fuse (2 required) 56-0049-01
Makrolon
Side Panels (2 Required) 10-0618-01 Upper Door Panel (Workstation) 10-0619-01
Lower Door Panel (Workstation) 10-0620-01 Upper Door Panel (Cabinet) 10-0621-01
Lower Door Panel (Cabinet) 10-0622-01
Replacement Tubes
®
Panels
Page 13
Ultraviolet PCR Cabinets and Workstations Page 13
UV PCR Workstation
34-0073-01
25W 254nm UV
2
95-0367-xx
34-0087-01
15w White
1
UV2 PCR Workstation
34-0007-01
8W 254nm UV
1
95-0439-xx
34-0073-01
25W 254nm UV 2
34-0087-01
15w White
1
Accessories Part Number
Table, stainless steel top 98-0077-01 Shelf, long 27” 20-0970-01 UV/Air Circulator Module (for 95-0367-xx) 97-0157-01 UVX Radiometer 97-0015-02 UVX Sensor (UVX-25) 97-0016-01 J-225 Meter 97-0004-01 Spectacles, UV blocking (UVC-303) 98-0002-01 Goggles, UV blocking (UVC-503) 98-0002-02 Face Shield, UV blocking (UVC-803) 98-0002-04
PCR Table

Replacing the UV Germicidal or White Light Tubes

Turn the unit power off and unplug the unit from the outlet. Lift up the panel door and locate the tubes. The tube has two prongs on each end that first fit into a socket then rotate into place. Carefully hold the tube on the end and insert the tube into the socket.
Twist the tube 90° until the tube locks into place.
Cabinet:
Outer (White Bulb) – White light (8 watt) Inner (Clear Bulb) – UV (8 watt)
Workstation:
Outer (Clear Bulb) – UV Inner (White Bulb) – White light

Replacing the Fuses

Fuses are located on lower left side of the system. To remove the fuse, use a flathead screwdriver to turn the fuse counter clockwise. Pull the fuse out. Insert the new fuse and using the screwdriver, turn the fuse clockwise to lock.
Page 14
Ultraviolet PCR Cabinets and Workstations Page 14
p.m. PST.
5:30 p.m.
uvp@uvp.co.uk
Cambridge CB4 1TG UK
Circulator

Replacing the Tube in UV/Air Circulator

1. Locate and remove the two thumb knobs on the circulator cap on the outer back panel of the system.
2. Pull the circulator cap away from the system.
3. The tube, attached to the cap, has two prongs on each end that fit into a socket. Carefully hold the tube on the ends, rotate the tube and remove from the cap. Insert a new tube into the socket. Twist the tube 90° until the tube locks into place.
4. Replace the place cap back into position and reattach the two thumb knobs.
cap with
UV tube

Technical Assistance

UVP offers technical support for all of its products. If you have any questions about the product’s use, operation or repair, please contact our offices at the locations below.
Note: A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before returning any product.
You may contact UVP’s Customer Service based on your location:
If you are in North America, South America, East Asia or Australia:
If you are in Europe, Africa, the Middle East or Western Asia:
ask for Customer Service during regular business days, between 7:00 a.m. and 5:00
E-Mail your message to: info@uvp.com
Write to: UVP, LLC 2066 W. 11th Street,
Upland, CA 91786 USA
Warranty
UVP, LLC warrants its products to be free of detects in materials and workmanship for a period of two (2) years from the date of purchase. The foregoing warranty of UVP shall be of no force and effect if buyer has modified or damaged the product. Tube s and filters are warranted for 90 days.
NOTE: “Crazing” (cracking and/or yellowing) of the Makrolon warranty period. This is regarded as normal wear and tear and is not covered by the warranty.
All warranties or merchantability and fitness for any purpose and all other warranties, expressed or implied, except those expressly set forth herein, are deemed waived and excluded.
UVP’s duty under the warranty is limited to replacement and/or repair of the defective part at the option of UVP, LLC. UVP shall not be liable for any expenses or damages incurred by the purchaser except as expressly set forth herein, and in no event shall UVP be liable for any special, incidental or consequential damages of any kind. This warranty does not supersede any statutory rights that may be available in certain countries.
Makrolon is a registered trademark of Bayer AG.
Customer Service during regular business days, between 8:30 a.m. and
Unit 1, Trinity Hall, Nuffield Road,
®
panels may occur within the
Loading...