UVP TW-26 User Manual

Warranty
UVP’s quality transilluminators are guaranteed to be free of defects in materials, work­manship and manufacture for two (2) years from the date of purchase. Consumable and disposable parts including, but not limited to bottles, tubes and filters, are guaran­teed to be free from defects in manufacture and materials for ninety (90) days from date of purchase. If equipment failure or malfunction occurs during the warranty period, UVP shall examine the inoperative equipment and have the option of repairing or replacing any part(s) which, in the judgment of UV P, were originally defective or be­came so under conditions of normal usage and service.
No warranty shall apply to any instrument, or part thereof, that has been subject to accident, negligence, alteration, abuse or misuse by the end-user. Moreover, UVP makes no warranties whatsoever with respect to parts not supplied by UVP or that have been installed, used and/or serviced other than in strict compliance with the instruc­tions appearing in the operational manual supplied to the end-user.
In no event shall UVP be responsible to the end-user for any incidental or consequen­tial damages, whether foreseeable or not, including, but not limited to property dam­age, inability to use equipment, lost business, lost profits, or inconvenience arising out of or connected with the use of instruments produced by UV P. Nor is UVP liable or responsible for any personal injuries occurring as a result of the use, installation and/ or servicing of equipment. This warranty does not supersede any statutory rights that may be available in certain countries.
White Light Transilluminators
Models TW-43 and TW-26
Operating Instructions
White Light Transilluminators offer the researcher cool white light through an evenly diffused glass surface. The plexiglass and glass combination is permanently sealed into the unit - double sealed safety fused for durability. Excellent for viewing coomassie blue stained gels as well as autoradiographs, X-Rays, and microtiter plates. The TW-26 is designed for use with 8 x 10" film sizes; the TW-43 with 14 x 17" film sizes.
For additional products, go to UVP.com for more information.
UVP, LLC 2066 W. 11th Street, Upland, California 91786
Tel: (800)452-6788; (909)946-3197 / Fax: (909)946-3597
Ultra-Violet Products Ltd. Unit 1, Trinity Hall Farm Estate,
Tel: +44(0)1223 420022 Fax: +44(0)1223 420561
81-0125-01 Rev C
Nuffield Road, Cambridge CB4 1TG UK
Operation
Place the Transilluminator on a level work surface. Be sure that adequate air space exists around the unit. This space allows for proper air circulation through the unit.
Plug the female end of the power cord into the Transilluminator. For 230 volt mod­els, or those requiring special power-cord connectors, ensure that the proper con­figuration of male connector or plug has been properly connected to the power cord. Plug the male end of the power cord into a properly grounded electrical outlet. The proper voltage of the Transilluminator is found on the product information label.
Place sample on the filter area. Press the ON/OFF switch to ON. The white lights should be glowing beneath the filter after a momentary flickering during the start period. After viewing/ photographing your sample, turn the unit off.
Schematic for TW-43
MODEL TW-43
TW-43 95-0214-03 100V/50-60Hz/1.3A 36x43cm 4 TW-43 95-0214-02 230V/50Hz/.8A 36x43cm 4 TW-43 95-0214-01 115V/60Hz/1.2A 36x43cm 4 Model Part No. Nom. Volt/Hz/Amp Filter Sz # of Tubes
Page 2 Page 3
Tube, White, 14 Watt 34-0059-01 Tube, White, 8 Watt 34-0056-01 Filter Assembly 38-0161-04 38-0165-01
For TW-26 For TW-43
Replacement Parts
POWER PRIOR TO CLEANING. ALWAYS DISCONNECT THE TRANSILLUMINATOR FROM ELECTRICAL sponge. Never use abrasive cleaners as this will damage the filter surface. The painted surface and filter areas should be cleaned with a damp soft cloth or
Cleaning the Transilluminator
locks into place.
4. Fit the replacement tube into the socket and rotate the tube 1/4 turn until it
MODEL TW-26
Schematic for TW-26
3. Carefully twist the tubes from lamp holder.
2. Remove filter cover by unscrewing retainer screws.
1. Disconnect Transilluminator from electrical supply.
To replace white light tubes: The TW-26 contains two 8-watt tubes; the TW-43 contains four 14-watt tubes.
Replacing the Tubes
+44(0)1223-420561 (fax)
In Europe/UK: +44(0)1223-420022 (telephone)
(909) 946-3597 (fax) (909) 946-3197 (telephone)
In the USA: (800) 452-6788 (toll free US & Canada)
TW-26 95-0208-03 100V/50-60Hz/.4A 21x26cm 2 TW-26 95-0208-02 230V/50Hz/.37A 21x26cm 2 TW-26 95-0208-01 115V/60Hz/.375A 21x26cm 2 Model Part No. Nom. Volt/Hz/Amp Filter Sz # of Tubes
from UVP prior to returning any product.
note that a Returned Goods Authorization (RGA) number must be obtained
Technical support is provided by UVP at the telephone numbers below. Please
Maintenance