
This unit is designed and manufactured strictly in
accordance with GB4793 safety requirements for
electronic testing meters and IEC61010-1 safety
standards. It fully meets CAT II 600V insulation and
overvoltage requirements and Grade II anti-pollution
safety standards. To prevent personal injuries and
damage of this unit or any other devices connected to
it please take note of the following safety precautions
To avoid potential hazards use this unit strictly as
instructed by this User Manual
power cable which is authorized in the country of use.
or testing cable when they are connected to power.
the ground wire of the power cable. To avoid electric
shock, the grounding conductor must touch the
ground. Before connecting the input or output
terminal, ensure the unit is properly grounded.
The probe ground cable is the same
as ground potential. Do not connect the ground cable
to non ground voltage or high voltage.
fire and excessive current shock please check all
rated values and label data Read the manual
carefully and check the rated values before
Do not operate this unit when t
Only use specified fuse types
and rated specifications.
make contact with exposed adaptor or components.
Maintenance should only be carried out by a
To avoid fire and personal injury
Use a correct power cable
Remove the plug correctly
Connect the probe of the digital storage
Check the rated values of all terminals
Do not operate the unit with the chassis cover

When fault is suspected, stop operation
Maintain good ventilation.
Do not operate in humid condition.
Do not operate in combustible and explosive
Keep the product surface clean and dry.
suspect a fault, ask a qualified maintenance
professional to carry out inspection.
Safety terminology used in this manual. The following
messages may appear in this manual
messages may appear on the product
means potential damage that is
means potential damage that is not
means possible damage to this product
Safety Messages and Symbols
Warning statements identify conditions
or practices that could result in injury or loss of life
Caution statements identify conditions
or practices that could result in damage to this unit

This manual provides information on the operation of the
UTD4000 four-channel digital storage oscilloscope series.
Guidance is given in several chapters as follows :
Simple guide to oscilloscope
functions and installation
the UTD4000 four channel digital storage oscilloscope
illustrations are provided to solve various testing
Chapter 2 Instrument Setups
Chapter 3 Practical Illustrations
Chapter 4 System Prompts and Trouble shooting
Chapter 5 Servicing and Support
Appendix A : Technical Indicators
Appendix B : Accessories for UTD4000 Four-channel
Digital Storage Oscilloscopes
Appendix C : Maintenance and Cleaning
Appendix D : Factory Setup

The UTD4000 Four-channel Digital Storage Oscilloscope Series
UTD4000 four-channel digital storage oscilloscopes offer user-friendliness, outstanding technical indicators
and a host of advanced features. They are your perfect tools to complete testing tasks swiftly and efficiently.
This manual is a user guide for 3 models of this digital storage oscilloscope series :

UTD4000 four-channel digital storage oscilloscopes
offer a user-friendly front panel with clear indications
to allow access to all basic functions for easy
operation. The scaling and position buttons for all
channels are optimally arranged for intuitive
operation As design is based on the familiar
practices of traditional instruments users can use
the new units without spending considerable time
in learning and familiarizing with operation For
faster adjustment to ease testing, there is an [ ]
key to instantly display the appropriate waveform and
atures listed below will explain
why the UTD4000 series can fully
satisfy your testing and measurement requirements
2GS/s real-time sampling rate and 50GS/s
Dual time base function unrivaled waveform
detail observation and analysis capabilities
24k storage depth; 60M equivalent storage
depth; 1024k recording length
Unique envelop sampling feature with direct
visual display of carrier wave details after
Scroll display in scan mode for continuous
monitoring of signal variations
Unique XY mode that displays the waveform and
USB drive system software upgrade
Supports plug-and-play USB storage device.
Communication with computer through the USB
Storage of waveforms setups and bit maps
waveforms and setups reproduction
Multiple waveform mathematics functions
(including add, subtract, multiply and divide)

Edge, video, pulse, slope and alternate trigger
Automatic measurement of 24 waveform parameters
parameters testing and customization
Multiple setups for extra flexibility
Visual system help messages
4 x 1.2m, 1:1/10:1 probe. For details refer to the
probe instructions. These accessories conform
with EN61010-031: 2008 standards
Power line conforming to international standards
applicable in the country of use
USB connecting cable : UT-D06
UTD4000 four-channel oscilloscope
communication control software (USB-device)
2x multimeter test lead; 2x c -to-vol
converter module: UT M03/UT-M04
UTD4000 Four-channel accessories :

AutomaticSetupforWaveformDisplay
GettingtoKnow the VerticalSystem
GettingtoKnow the HorizontalSystem
GettingtoKnow theTriggerSystem
Chapter2 InstrumentSetups
SettingtheHorizontalSystem
Setting the Display System
UserGuide------------------------------------------------------------------------------
GeneralInspection----------------------------------------------------------------------
-----------------------------------------------------------------------
------------------------------------------------------------------
----------------------------------------------
--------------------------------------------------
-----------------------------------------------
----------------------------------------------
---------------------------------------------------------------------
------------------------------------------------------------
---------------------------------------------------------
------------------------------------------------------------
----------------------------------------------------
------------------------------------------------------------------
--------------------------------------------------------
Setting the Sampling System
166101111131517183540525659

Chapter3 PracticalExampleIllustrations
Illustration 1 : Measuring simple signals
Illustration 2 : Observing the delay caused by a
sine wave signal passes through the circuit
Illustration 3 : Acquiring single signals
Illustration 4 : Reducing random noise of signals
Illustration 5 : Using the cursors for measurement
Illustration 6 : Using the X-Y function
----------------------------------------------------------------
-----------------------------------------------------------
-----------------------------------------------------------------
-----------------------------------------------------------------------
-------------------------------------------------------------------------------
-------------------------------------------------------------
-----------------------------------------------------------------------------
-----------------------------------
-------------------------------------------
---------------------------------------
---------------------------------------------
---------------------------------
-------------------------------
-----------------------------------------------
-----------------------------------------------
------------------------------
-------------------------------------------------
Using the dual time base function
Illustration 9 : Using the multimeter
707476777878808081838486909294

Chapter 4 System Prompts and Trouble-shooting -------------------------------------------
Definitions of System Prompts -----------------------------------------------------
Troubleshooting-----------------------------------------------------------------------
Chapter 5 ----------------------------------------------
Chapter 6 Technical Indicators ------------------------------------------------------------------
Appendix A : Technical indicators--------------------------------------------------
--------------------------------------------------
Appendix C : Maintenance and Cleaning-----------------------------------------
Appendix D -------------------------------------------------------
Upgrading System Software on USB
Accessories for UTD4000 Four-channel Digital Storage

Your UTD4000 Four-channel digital storage
oscilloscope is a small and compact benchtop device.
The user-friendly front panel enables easy operation.
This chapter will guide you through basic testing
This chapter provides notes on the following :
Automatic setups for waveform display
Getting to know the vertical system
Getting to know the horizontal system
Getting to know the trigger system
When beginning to use your UTD4000 Four-channel
oscilloscope, first familiarize yourself with the
operation front panel. This ch
the operation and functions of the front panel, so you
can get started with your UTD4000 four-channel
digital storage oscilloscope as quickly as possible.
Your UTD4000 Four-channel oscilloscope comes with
a front panel with at-a-glance functions for easy
operation There are knobs and function keys on
the front panel The functions of knobs are similar
to other oscilloscopes On the right you will find 5
menu operation keys designated as to
from top down With these keys you can set up
different options of the current menu. The other keys
are function keys. You can use the
function menus or access particular functions directly.

Figure1-1 Series Digital Storage Oscilloscope

Figure 1-2 Back cover of UTD4000 four-channel digital storage oscilloscope

DIGITA LSTO RAGE OSCILLO SCOPE
Figure1-3 Series Digital Storage Oscilloscopes

Figure1-4 Schematic diagram ofthedisplayinterface
Displayingthehorizontal triggerposition
Trigger frequency counter

We suggest checking your new UTD4000 Four-
channel oscilloscope in the following steps.
If the package carton or foam plastic protective lining
is seriously damaged, please arrange for exchange
A checklist of accessories that come with your
UTD4000 Four-channel oscilloscope is provided in
the section Accessories for UTD4000 Four-channel
Digital Storage Oscilloscopes of this user manual
Please check any missing items against this list
If any item is missing or damaged please contact
your NI-T dealer or our local office.
If the exterior of the unit is damaged, or it is not
erating normally, or it fails to pass the performance
test, please contact your UNI-T dealer or our local
In the event of any shipping damages, please retain
the packaging and notify our shipping department or
your UNI-T dealer. We will be glad to arrange
Carry out a quick functional check in the following
steps to make sure your oscilloscope is operating
1. Check the unit for possible shipping damages
3. Thorough inspection of the entire unit

Power on the unit. AC power supply voltage range is
100V to 240V, frequency 45Hz-440Hz. After
connecting to power, start the self calibration process
on the optimal oscilloscope signal path at greatest
measurement accuracy. Press the [ ] button
and [ ] twice, then press the
control knob to perform the function Press
[ ] and [ ], then press [ ] to go to the next
page. There, press [ ] then the
control knob To recall default setup, see
At the end of the above process, press [ ] to enter
To avoid danger, ensure the digital
storage oscilloscope is safely grounded.

Your UTD4000 Four-channel digital storage
oscilloscope has four input channels and an external
trigger input channel, as shown in Figure 1-6. Please
access signals in the following steps :
Connect the probe of the digital storage
oscilloscope to the CH input terminal and set
the attenuation switch of the probe to X
Four-channel input and external
EXT external trigger channel
Figure 1-7 Setting the attenuation switch of

You have to set the probe attenuation factor of
the oscilloscope. This factor changes the vertical
range multiple to ensure the measurement result
correctly reflects the amplitude of the signal being
tested Set the attenuation factor of the probe as
follows Press then to show X on
Connect the probe tip and ground clamp to the
connection terminal for the probe compensation
signal. Press [ ] and you will see a square wave
in the display (1kHz, approximately 3V, peak-to-peak
value) in a few seconds, as shown in Figure 1-
Press [ ] twice to close CH1, then press [ ]
to activate CH2 and repeat steps 2 and 3. Use the
same method for CH3 and CH4.
Figure1-8 Settingthedeflectionfactor
Figure1-9 Probecompensationsignal

When connectingthe probeto any input channel for thefirst
time, perform this adjustment to match the probe to the
channel. Skipping the compensation calibration step will
result in measurement error or fault. Please adjust probe
1 Setthe probeattenuationfactorto10X.Movethe switch
on the probe to 10X and connect the probe to CH1.
When using a hook-tip, ensure it is well connected to
the probe. Connect the probe tip to the output terminal
of the probe compensator s signal connector, and the
ground clamp to the ground cable connector of the
3. Ifyouseean Undercompensation or
Overcompensation waveform display adjust the
adjustable capacitance tab of the probe with a
screwdriver with non metal handle until a Correct
Compensation waveform shown in the above figure is
2. Observe the displayed waveform.
Figure1-10 Probecompensationcalibration
To avoid electric shock when measuring high
voltage with the probe ensure the probe s
insulation lead is in good condition. Do not
touch the metal part of the probe when

Automatic Setup for Waveform Display Getting to Know the Vertical System
Your UTD4000 Four-channel digital storage
oscilloscope features an auto setup function. It can
automatically adjust the vertical graticule factor,
scanning time base and trigger mode based on the
input signal, until the most appropriate waveform is
displayed. The auto setup function can only be
operated when the signal to be measured is 50Hz or
above and the duty ratio is larger than 1%.
1. Connect the signal to be tested to the signal input
2. P r es s [ ]. Th e o s ci l l os c o p e w i l l
automatically set the vertical graticule factor
scanning time base and trigger mode Should
ake more detailed check you
can adjust manually after the auto setup
process until you get the optimal waveform
As shown in the figure below, there are a group of
buttons and knobs in the vertical control zone. The
following exercise will guide you through vertical
Using the Auto Setup Function :
Vertical control zone on the front panel

The knob can move the waveform
vertically. Press this knob to quickly return to the
Press the [ ], [ ], [ ], [ ], [ ] and
[ ] keys for the vertical channel operation menu,
or to open or close the waveform display channel.
Use the [ ](CH1, CH2, CH3, CH4) key to
set the vertical graticule factor.
Displacemen t and v er tical gr aticule facto r
adjustments of [ ], [ ] channels by
1. Press the vertical knob to display
the waveform signal in the centre of the window.
The vertical knob controls the
vertical display position of the signal When
you turn the vertical knob the
level will move up and down with
2. Change the vertical setup and observe changes
of status information. You can identify changes of
any vertical range by reading the status display
column at the lower corner of the waveform
window. Turn the vertical knob to
change the vertical range You will
find that the range in the current status
column has changed accordingly Press
and the screen will show the corresponding
operation menu, sign, waveform and range status
If the channel coupling is DC, you can measure the
signal s DC component quickly by checking the
distance between the waveform and signal
In the case of AC coupling, the DC of the
signal will be blocked. With this coupling mode you
can display the AC of the signal with

Getting to Know the System
As shown in the figure below, there are one button
and two knobs in the horizontal control zone. The
following steps will get you familiar with horizontal
The knob can move all channels MATH
waveforms and REF waveforms horizontally Press
this knob to quickly return to the centre point
[ ] horizontal menu, to display al
Use the knob to set the SEC DIV
graticule factor for horizontal scan If the window
is expanded you can adjust graticule of the
1. Use the horizontal knob to change the
horizontal time base setup and check any
changes in status information. Turn the
time base range You will find that
the time base range in the current status
column has changed accordingly Range of
horizontal scanning rate is 5ns/div~50s/div
(UTD4104C), in steps of 1-2-5
Note Horizontal scanning time base
Series varies from model to model
Horizontal control zone on the front panel

2. Use the horizontal knob to adjust
the horizontal position of the waveform
window. When the horizontal knob is
turned you can see that the waveform moves
horizontally with the knob
3. Press [ ] to activate the display window and
dual time base menu. In this menu press [ ] to
activate window expansion Then press
again to quit window expansion and return to
the main display screen For dual time base
setup press You can also set the
holdoff time with this menu by turning the
Shortcut key for resetting the trigger point to
When the trigger point has shifted significantly away
from the horizontal centre point, use the [POSITION]
knob to quickly reset the trigger point to the
horizontal centre point. You can also use the
horizontal knob for adjustments
means the actual trigger point relative
to the centre point of the storage device. By turning
the horizontal knob you can move the
trigger point horizontally
means time before another
trigger to be accepted. Turn the
control knob to set the holdoff time By adjusting
holdoff time you can observe complex or

Getting to Know the Trigger System
As shown in Figure there are one knob and
one button in the trigger menu control zone. The
following steps will get you familiar with trigger setup.
When operating edge, pulse width and slew rate
triggers, set the amplitude to be crossed by the
waveform upon signal occurrence by turning the
trigger knob To quickly set the trigger
level as the vertical centre point of the trigger
signal press the trigger knob
To display trigger menu contents
1. Use the trigger knob to change the
trigger level You will see a trigger sign on the
screen that indicates the trigger level The sign
nd down with the knob While you
move the trigger level you will find the trigger
level value on the screen changing accordingly.
2. Open the trigger [ ] key (see the figure 1 )
Press [ ] twice and select for

Press [ ] and select for [
knob to select and then press that key to
Press [ ] then [ ]. Set for
Press [ ] then [ ]. Set for [ ].
Press [ ] then [ ]. Set for [ ].
Ic o n uti l i ty fu n c tion o f the
quickly return to the centre point.
Icon utility function of the trigger
quickly return to horizontal ground
level, i.e. trigger zero level
Ic o n uti l i ty fu n c tion o f the
this key to confirm selection

You should be familiar with basic operation of the
vertical controls, horizontal controls and trigger
system menu of your UTD4000 Series
oscilloscope by now. After reading the last chapter,
you should be able to use the menus to set up your
digital storage oscilloscope. If you are still unfamiliar
with these basic operation steps and methods, please
This chapter will guide you through the following :
Settin g up the vertical system ([ ],
Setting up the horizontal system ([ ],
Setting up the Trigger system ( ,
Setting up the sampling method ([ ])
Setting up the display mode
Setting up the help system ([ ])
Automatic measurement ([ ])
Auto setup, run/stop key ([ ], [ ])
Multipurpose control knob (
It is recommended that you read this chapter carefully
to understand the various measurement functions
and system operation steps of your UTD4000 four-
channel digital storage oscilloscope.

Setting the Vertical System
CH1, CH2, CH3, CH4 and setups
Each channel has its own vertical menu. You should set up each item for each channel individually. Press the
[ ], [ ], [ ] or [ ] function button and the system will display the operation menu for CH1, CH2, CH3
or CH4. For explanatory notes please see Table 2 below
Table 2-1 Explanatory notes for channel menu
Pass AC and DC quantities of input signal. Intercept DC quantities of
Display reference ground level (without disconnecting the input signal).
Limit bandwidth to 20MHz to reduce noise display
Coarse tune in steps of 1-2-5 to set up the vertical graticulefactor of
the vertical system Fine tune means further tuning within the coarse
tune setup range to raise the vertical resolution
Select either one value based on the probe attenuation factor to keep
the vertical deflection factor reading correct. There are four values :

1. Setting up channel coupling :
Take an example of applying a signal to CH1. The
signal being tested is a sine signal that contains
DC%. Press [ ] to select AC then press [ ] to select
AC 1M It is now set up as AC coupling DC
quantities of the signal being tested will be
intercepted The waveform display is as follows
Press [ ] twice to select DC 1M Both DC and AC
quantities of the testing signal being inputted to
CH1 can pass through The waveform display is
Figure 2-1 DC quantities of the signal are intercepted
Figure 2-2 Both DC and AC quantities of the signal are

Press [ ] then [ ] to select ground. It is now set up
as ground. The waveform display is as follows :
Note In this mode although waveform is
not displayed the signal remains connected to
2. Setting the channel bandwidth limit
Take applying a signal to Ch1 as an example, the
signal to be tested contains high frequency
Press [ ] to turn CH1 on, then press [ ] and [F1].
Bandwidth is now set to full bandwidth The signal
being measured can pass through even if it
contains high frequency quantities. The waveform
Figure 2-3 Channel is set to ground mode
Figure 2-4 Waveform display at full

Press [ ] then [ ]. The noise and high frequency
quantities over 20MHz of the signal being tested are
now restricted. Waveform display is as follows.
3. Setting up the probe rate
To match the probe attenuation factor setup, it is
necessary to set up the probe attenuation factor in
the channel operation menu accordingly For
example when the probe attenuation factor is
10:1 set the probe attenuation factor at 10X in
the channel menu Apply this principle to other
values to ensure the voltage reading is correct
The figure below shows the setup and vertical
range display when the probe is set at 10:1.
Figure 2-5 Waveform display when bandwidth
Figure 2-6 Setting the probe attenuation factor in the

4. Vertical VOLTS/DIV adjustment setup 5. Waveform inversion setup
You can adjust the VOLTS/DIV range of the vertical
deflection factor either in the coarse tune mode or
fine tune mode. In coarse tune mode the
VOLTS DIV range is 2mV/div~5V/div Tuning is in
steps of 1-2-5 In fine tune mode you can change
the deflection factor in even smaller steps within
the current vertical range so as to continuously
adjust the vertical deflection factor within the range
of 2mV/div~5V/div without interruption.
Waveform inversion : The displayed signal is inverted
180 degrees. Figure 2-8 shows the non inverted
waveform Figure 2-9 shows the inverted waveform
Figure 2-7 Coarse tuning and fine tuning the vertical
Vertical graticulefactor VOLTS/DIV movement
Figure Inversion setup for vertical channel
non inverted graticule factor

Math functions are displays of and FFT
mathematical results of waveform channels CH1,
CH2, CH3 and CH4, and the digitally filtered
Figure 2-9 Inversion setup for vertical

Table 2-2a Explanatory notes for the Math menu (1)

Table 2-2b Explanatory notes for the Math menu (2)
Scale the waveform by ratio. There are four ratios to choose
from : 1/1, 1/10, 1/100, 1/1000
Use the control knob to move the waveform
Use the control knob to adjust the vertical

By using the FFT (Fast Fourier Transform algorithm
you can convert time domain signals YT into
frequency domain signals With FFT you can
conveniently observe the following types of
Measure the harmonic wave composition and
Demonstrate the noise characteristics of the DC

Table 2-3a Explanatory notes for the FFT menu (1)
Table 2-3b Explanatory notes for the FFT menu (2)
To carry out FFT algorithm functions
Set Hamming window function
Set Blackman window function
Set Rectangle window function
Set Hanning window function
Set the vertical coordinate unit to linear or V dB
Use the control knob to move the waveform
Use the control knob to adjust the vertical

Assuming the YT waveform is constantly repeating
itself, the oscilloscope will carry out FFT conversion
of time record of a limited length. When this cycle is a
whole number, the YT waveform will have the same
amplitude at the start and finish. There is no
waveform interruption. However, if the YT waveform
cycle is not a whole number, there will be different
amplitudes at the start and finish, resulting in
transient interruption of high frequency at the
connection point. In frequency domain, this is known
as leakage. To avoid leakage, multiply the original
waveform by one window function to s
for start and finish compulsively. For application of the
window function, please see the table below :
Signals with DC quantities or DC offset will cause
error or offset of FFT waveform quantities. To reduce
DC quantities, select AC coupling To reduce
random noise and frequency aliasing resulted by
repeated or single pulse event set the acquiring
mode of your oscilloscope to average acquisition.

Most Suitable Measurement Item
The best frequency recognition rate
the worst amplitude recognition rate
Basically similar to a status without
Temporary or fast pulse. Signal level is generally
the same before and after. Equal sine wave of very
similar frequency. There is broad-band random
noise with slow moving wave spectrum.
Frequency recognition rate is better
than the rectangle window, but
amplitude recognition rate is poorer.
Sine, cyclical and narrow band random noise
Frequency recognition rate is
marginally better than Hanning
Temporary or fast pulse. Signal level varies greatly
The best amplitude recognition rate
and the poorest frequency
Mainly for single-frequency signals to search for
higher-order harmonic wave.
means the quotient of the sampling and math points When math point value is
fixed the sampling rate should be as low as possible relative to the FFT recognition rate.
To rebuild the original waveform, at least 2f sampling rate should be used for waveform
with a maximum frequency of f. This is known as Nyquist stability criterion where f is the Nyquist
frequency and 2f is the Nyquist sampling rate.

Digital Filtering Function
Table 2-5a Explanatory notes for the digital filtering menu (1)

Table 2-5b Explanatory notes for the digital filtering menu (2)
Effective only during low-pass filtering or band-pass filtering.
Use the control knob to set the maximum
Effective only during high-pass filtering or band-pass filtering.
Use the control knob to set the minimum
Use the control knob to move the waveform
Use the control knob to adjust the vertical

Display of the saved reference waveforms can be set
on or off in the REF menu. The waveforms are saved
in the non volatile memory of the oscilloscope or an
external USB device and are identified with the
following names : RefA, RefB. To display (Load) or
hide (off) the reference waveforms, take the
2. Press [ ] for Load and select the signal
source by turning the control
knob You can choose from 1 to 10 After
selecting a numeral for saved waveform e g
1 press the control knob to
confirm and the waveform originally stored in that
position can be recalled. For instructio
storing or recalling reference waveforms on the
USB device, read Storage and Recall
3. Press [F1] for RefB to select the second
signal source for the math function by repeating
4. To close the reference waveform, press [ ].
In actual application, when using your UTD4000
to measure and observe such
waveforms, you can compare the current
waveform with the reference waveform for
analysis. Press to display the
reference waveform menu For setup please
: When [ ] is pressed after a waveform
is recalled or imported, that waveform will

Table 2-6a Explanatory notes for the REF menu (1)
Table 2-6b Explanatory notes for the REF menu (2)
Select as the reference waveform
Select as the reference waveform
Recall waveforms stored in 10 positions then select one with the
control knob Press the control knob to
To display the amplitude and time base contents of the current
Use the control knob to move the waveform
Use the control knob to adjust the vertical
Close the reference waveform
Enter the USB menu (see Table 2-7), recall the reference
waveform stored on the USB device

Table 2-7 Explanatory notes for the USB menu
: To select an internal storage position, choose
between 1 and 10. In the case of external storage
device, plug in the USB device. A message saying
USB installation complete appears Press
on the next page for the import menu and then
In the USB menu use the key and
control knob to set the document
name Press to select the character
positions that need to be changed Use the
control knob to change the
selected characters or numerals
: Y Offset Y level OFF are
operative only after the reference waveform has
been recalled or imported
Use the control knob and F key to set
the document name to be imported from the USB device.
For specific operation instruction see note 2
1OKAfter confirming, go back to the REF menu. If there is such a
document on the USB device, it will be imported. Otherwise a
I/O failure message will appear

Setting the Horizontal System
You can use the horizontal control knobs to change
the horizontal graticule (time base) and trigger the
horizontal position of the memory (triggering
position). Changing the horizontal graticule will cause
the waveform to increase or decrease in size relative
to the screen centre. When the horizontal position
changes, the position with respect to the waveform
triggering point is also changed.
Horizontal position : Adjust the horizontal positions of
channel waveforms (including math waveforms).
Resolution of this control button changes with the
Horizontal scaling : Adju
st the main time base, i.e
SEC/DIV. When time base extension is on, you can
use the horizontal scaling knob to change the delay
scanning time base and change the window width.
Two horizontal control knobs : Use the knob
to change horizontal time base graticule and use
the horizontal knob to change the
relative position if the triggering point on the
screen For instructions on how to display the
horizontal menu see Table
Horizontal controls menu : Horizontal menu display

Table 2-8 Explanatory notes for the horizontal menu
Represents the signal frequency currently
selected as trigger source.
Represents the triggering point position of the
Represents the trigger level of the current
Distance between the triggering position and the
The time base value of main time base M1, i.e
Horizontal parameter interface definitions :
Press [ ] to switch between the main screen and
Enter the dual time base menu. See Table 2-9
Use the control knob to adjust holdoff time
Figure 2-13 Horizontal parameter interface

Window expansion can be used to zoom in a band of
waveform to check image details. Please refer to
The window expansion setting cannot be slower than
the main time base setting. Maximum magnification
In the window extension mode, the display is divided
into two zones as shown above. The upper part
displays the original waveform. You can move this
zone left and right by turning the horizontal
knob or increase and decrease the
selected zone in size by turning the horizontal
knob The lower part is the horizontally
expanded waveform Please note that the
te of expanded time base relative to the
main time base is now higher (as shown in the above
figure). Since the waveform shown in the entire lower
part corresponds to the selected zone in the upper
part, you can increase the extended time base by
turning the horizontal knob to decrease
the size of the selected zone In other words you
can increase the multiple of waveform expansion
Horizontally expanded section

The dual time base function is similar to window
extension but there is a fundamental difference In
the window extension mode you can magnify the
waveform times whereas in the dual time
base mode you can magnify details of the
waveform being observed by thousands of times
In effect the main time base storage depth is
increased by thousands of times.
The dual time base menu and its operation are as
: The Math function is disabled in the dual time
Figure 2-15 Dual time base

Table 2-9 Explanatory notes for the dual time base menu
Switching between main time base and dual time base
For dual time base mode instructions see Figure
Switch to the waveform displayed on a different channel. Only
one channel can be displayed in the dual time base mode
When M1 is the main time base, use the horizontal and
knobs to adjust into main time base parameters
When M2 is the main time base, use the horizontal and
knobs to adjust into main time base parameters
Return to the horizontal menu
Set delayed time base M2 to move horizontally in escalating scale
Set delayed time base M2 to move horizontally in de-escalating scale

Setting the Trigger System
Triggering decides when the oscilloscope collects
data and display waveforms. Once the trigger is
correctly set up, it can transform unstable displays
into meaningful waveforms. When beginning to
acquire data the digital storage oscilloscope first
collects sufficient data required for drawing a
waveform on the left side of the trigger point
When trigger is detected it continuously acquires
sufficient data to draw a waveform on the right
side of the trigger point
The trigger control zone on the operation panel of
your oscilloscope comprises a trigger knob
se , Video and Slope rate.
: Trigger is set to occur when the signal
is at the rising or falling edge. You can use the trigger
knob to change the trigger point s vertical
position on the trigger edge, i.e. the intersection point
of the trigger level line and the signal edge on the
When the pulse width of the
trigger signal reaches a preset trigger condition
Carry out field or line trigger to
Trigger condition is the signal rising
Below are notes for various trigger menus.
For edge trigger menu setups please see Table

Allow AC and DC quantities of the input signal to pass
Intercept DC quantities of the input signal
Reject low frequency quantities below 80kHz of the signal
Reject high frequency quantities above 80kHz of the signal
Set CH1, CH2, CH3 or CH4 as the signal source trigger signal
Set to external trigger or divide the external trigger source by 5
CH1 and CH2 trigger their respective signals alternately
CH3 and CH4 trigger their respective signals alternately
The system automatically acquires waveform data when there is
no trigger signal . The scan baseline is shown on the display.
When the trigger signal is generated, it automatically turns to
The system stops acquiring data when there is no trigger signal.
When the trigger signal is generated, trigger scan occurs
One trigger will occur when there is an input trigger signal. Then

Table 2-10 Edge trigger Connected to the table
Pulse trigger means determining the triggering time
based on the pulse width. You can acquire abnormal
pulse by setting the pulse width condition.
Table 2-11 Pulse width Trigger
Set to trigger at the signal s rising edge
Set to trigger at the signal s falling edge
Set to trigger at the signal s rising and falling edges
Set CH1, CH2, CH3 or CH4 as the signal source trigger signal
Set to external trigger or divide the external trigger source by 5
CH1 and CH2 trigger their respective signals alternately
CH3 and CH4 trigger their respective signals alternately

Table 2-11 Pulse width Trigger
The system automatically acquires waveform data when there is
no trigger signal. The scan baseline is shown on the
display. When the trigger signal is generated, it automatically
The system stops acquiring data when there is no trigger signal.
When the trigger signal is generated, trigger scan occurs
One trigger will occur when there is an input trigger signal. Then
Set the positive pulse width as the trigger signal
Set the negative pulse width as the trigger signal
Trigger occurs when pulse width of the input signal is smaller
Trigger occurs when pulse width of the input signal is larger than
Trigger occurs when pulse width of the input signal equals to the

Table 2-1 Pulse width setup
By selecting video trigger, you can carry out field or
line trigger with NTSC or PAL standard video signals.
See Table 2-13 for the trigger menu :
Set pulse width trigger to 20.0ns-10s with the
Return to the pulse width trigger menu
Set CH1, CH2, CH3 or CH4 as the signal source trigger signal
Set to external trigger or divide the external trigger source by 5
CH1 and CH2 trigger their respective signals alternately
CH3 and CH4 trigger their respective signals alternately

When PAL is selected for video and synchronization
mode is line, you will see a screen display as shown in
Figure 2-16. When synchronization mode is field, you
will see a screen display as shown in Figure 2-17.
Suitable for PAL video signals
Suitable for NTSC video signals
Set the video odd field to synchronized trigger
Set the video even field to synchronized trigger
Set the line signal to synchronize with trigger
Set synchronized trigger on the specified line and adjust by
turning the control knob lines for PAL
Return to video trigger menu
Figure 2-16 Video trigger : Line synchronization

If slope trigger is selected, trigger occurs when the
signal s rising or falling rate meets the set condition.
For the trigger menu see Table 2-15 below
Figure 2-17 Video trigger : Field synchronization

Allow AC and DC quantities of the input signal to pass
Intercept DC quantities of the input signal
Reject low frequency quantities below 80kHz of the signal
Reject high frequency quantities above 80kHz of the signal
Set CH1, CH2, CH3 or CH4 as the signal source trigger signal
Set to external trigger or divide the external trigger source by 5
CH1 and CH2 trigger their respective signals alternately
CH3 and CH4 trigger their respective signals alternately
The system automatically acquires waveform data when there is
no trigger signal. The scan baseline is shown on the
display. When the trigger signal is generated, it automatically
The system stops acquiring data when there is no trigger signal.
When the trigger signal is generated, trigger scan occurs
One trigger will occur when there is an input trigger signal. Then

Table 2-16 Slew rate setup
Trigger occurs when the signal slew rate within the threshold is
greater than the set slew rate
Trigger occurs when the signal slew rate within the threshold is
smaller than the set slew rate
Trigger occurs when the signal slew rate within the threshold
equals to the set slew rate
Select the rising edge within the threshold for trigger
Select the falling edge within the threshold for trigger
Set the slew rate value with the control knob
Change the low level value with the control knob
Change the high level value with the control knob
Change the high and low level value with the
Return to the slew rate trigger menu
When alternate trigger is selected, the trigger signal will be present in two vertical channels. This triggering
mode is suitable for observing two signals of unrelated signal frequencies Alternate trigger can also be
used to compare pulse widths

Adjusting the Holdoff Time
You can adjust the holdoff time to observe
complicated waveforms (e.g. pulse string series).
Holdoff time means the waiting time for the trigger
circuit to be ready for use again when the
oscilloscope is restarted. During this time the
oscilloscope will not trigger until the holdoff is
complete. For example, if you wish to trigger one
group of pulse series at the first pulse, set the holdoff
time to the pulse string width as shown in Figure 2-18.
Table 2-1 Trigger holdoff menu
Figure 2-18 Use the holdoff function to synchronize
Press [ ] to switch between the Main and
Enter the dual time base menu. See Table 2-9
Use the control knob to adjust holdoff time

1. Follow the normal signal synchronization
procedure and select the edge and trigger source in
trigger [ ]. Adjust the trigger level to make the
waveform display as stable as possible.
2. Press the horizontal [ ] key to display the
3. Adjust the control knob in the
upper front panel The holdoff time will change
accordingly until the waveform display is stable
: Holdoff time is usually slightly shorter
than the Large cycle When observing a Rs232
communication signal waveform it is easier to
observe if holdoff time is slightly longer than the
starting edge time of every data fram
1. The signal used for trigger.
The signal for trigger can be obtained from
various sources : input channel (CH1, CH2, CH3,
CH4), external trigger EXT EXT/5 and LINE
source is to select one of the four input channels.
The channel selected as trigger source can
operate normally whether the input waveform
This trigger signal can be input
directly through the external trigger input terminal. For
example, you can use an external clock or the signal
from a circuit to be tested as the trigger source. The
trigger source uses the input terminal to
ternal trigger signals setup is enabled
when signal trigger level range is 0.8V to +0.8V. To
allow input of larger signal through external trigger, the
trigger signal is divided by 5 in the EXT/5 mode, so that
the trigger level range is extended to 4V to +4V.

means the AC power source This
trigger mode is suitable for observing signals
related to the AC power e g the correlation
between lighting equipment and power source
equipment to achieve stable synchronization
2. Determine the action of your
oscilloscope at trigger. This oscilloscope offers three
trigger modes for selection : auto, normal and single.
The system will acquire and
display waveform data automatically when
there is no trigger signal input When the
trigger signal is generated, it automatically turns
to trigger scan for signal synchronization.
Time base of the scan range can b
50ms/div or slower to generate a roll waveform.
oscilloscope samples waveforms only when
triggering conditions are met. The system stops
acquiring data and waits when there is no trigger
signal. When the trigger signal is generated,
In this mode you only have
to press the RUN button once and the
oscilloscope will wait for trigger. One sampling
will occur and the acquired waveform will be
displayed when the digital storage oscilloscope
detects a trigger. Then trigger will stop.
3. Trigger coupling determines
which quantities of the signal are tr
the trigger circuit. Coupling modes are DC, AC,
low frequency suppression and high frequency
:Allowing all quantities to pass
:Intercepting DC quantities and attenuating
:Intercepting DC quantities and
attenuating low frequency quantities under
:Attenuating high frequency

The trigger position is typically set at the
horizontal center of the screen. In this case, you are
able to view 6 divisions of pretrigger and delay
information. Adjusting the horizontal displacement of
the waveform with the knob allows you to
see more pretrigger information. By observing
pretrigger data, you can see the waveform before
trigger occurs. For example, you can detect the glitch
that occurs when the circuitry starts. Observation and
analysis of trigger data can help you identify the
As shown below, [ ] button in the control
zone is the function key for the sampling system.
Press the [ ] key to pop out the sampling
setup menu. You can use this menu to adjust the
Pretrigger/Delayed Trigger
Setting the Sampling System
Figure 2-19 Function key for the sampling

By changing the acquisition setup, you can observe
the consequent changes in waveform display. If the
signal contains considerable noise, you will see the
following displays when average sampling is not
selected and when 32-time average sampling is
selected. For sampling waveform display please see
Figure 2-20 and Figure 2-21.
Set to average sampling with display of the average number of
Set the average number of times in multiples of 2, i.e. 2, 4, 8, 16,
32, 64, 128, 256. To change the average number of times, use
the control knob shown on the left in
Turn the equivalent sampling mode on or off. In this mode, selecting
alternate trigger (e.g. CH1 and CH2) as the trigger source is

1. Use Real time sampling to observe single
2. Use Equivalent sampling to observe high
frequency cyclical signals.
3. To avoid mixed envelop when observing a signal,
select Peak Detect To reduce random noise
of the displayed signal select average
sampling and increase the average number of
times in multiples of 2 i e selecting from 2 to
Figure 2-20 Waveform without average
Figure 2-21 Waveform when 32-time

Acquiring the data required in
one go. Maximum sampling rate is 2GS/s.
This is a repeated sampling
mode that allows detailed observation of repeated
cyclical signals In the equivalent sampling mode,
the horizontal pixel aspect ratio is 20ps higher than
the real time mode, i.e. 50GS/s equivalent.
Your oscilloscope acquires signal
samples at equal time intervals to reconstruct
In this acquisition mode, the
oscilloscope identifies the biggest and smallest
values of the input signals at each sampling interval
and use these values to display the waveform. In
ope can acquire and display
narrow pulse which would otherwise be omitted in the
sampling mode. Noise seems to be more significant in
The oscilloscope acquires several
waveforms and take the average value to display
the final waveform You can use this mode to
The oscilloscope acquires multi-
amplitude waveforms and calculate all sampling
points that synchronize with the triggering point. The
maximum and minimum values are then displayed

Setting the Display System
As shown below, the [ ] key in the control
zone is the function key for the display system.
Press the [ ] button to pop out the setup
menu shown below. You can use this menu to adjust
Figure 2-22 Function key for the sampling
Voltage relative to time (horizontal graticule) is displayed
There are two groups in X-Y display mode. The first group is CH1
for X input and CH2 for Y input. The second group is CH for X
input and CH for Y input.
Sampling points are linked for display
Sampling points are directly displayed

Set the grid display mode of the waveform zone to full, grid,
Waveform on the screen is refreshed at normal rate
Waveform on the screen is maintained for 1 second and then
Waveform on the screen is maintained for 2 seconds and then
Waveform on the screen is maintained for 5 seconds and then
Waveform on the screen remains on display. New data will be
added continuously until this function is disabled
In Fast ACQ mode intensity of the waveform can
be adjusted with the control knob

This mode offers two separate displays. The first
group is CH1 and CH2, i.e. CH1 signal input on the
horizontal axis (X axis) and CH2 signal input on the
vertical axis (Y axis). The other group is CH3 and
CH4, i.e. CH3 signal input on the horizontal axis (X
axis) and CH4 signal input on the vertical axis Y
axis as shown in Figure 2 23
In the normal mode, you can move figures
horizontally with the CH1 or CH3 knob
and move figures vertically with the CH2 or CH4
knob Adjust the dimension and shape
of XY figures with of the four channels
and turn the knob to improve the display
quality of the lissajous figure
ks to the unique display feature, your
oscilloscope can display channel waveform and
lissajous figure concurrently. The UTD4000
series also offers the following functions :
Automatic measurement mode
Reference or math waveform
The following functions are disabled under X-Y mode:
Figure 2-23 Waveform display in X-Y

Vector display fills the space
between adjacent sampling points in display. The dot
mode displays only the sampling points
Refresh rate is the number of
times the digital storage oscilloscope refreshes the
waveform display per second. The refreshing speed
affects the capability to observe signal movements
As shown below, the [ ] button in the control
zone is the function key for the storage system
Press the [ ] button to display the storage
setup menu. You can use this menu to save
waveforms or setup status of the oscilloscope in the
internal memory or the USB device, and recall any
ored waveform through RefA (or RefB), or press
[ ] to recall the setup status. When the USB
device is inserted, you can store the oscilloscope s
waveform display in bitmap format on the USB device.
Figure 2-24 Function key for the sampling

Press [ to go to the type menu. There
are three types to choose from : Setting, Wave,
1. Select Wave to go to the waveform storage
menu shown below (see Table 2-2 ). A stored
reference waveform can be recalled with the REF
menu (reference waveform, P 32). For specific
steps see the REF instructions.
About Internal and CSV formats : When
exporting a reference waveform to the USB
device you can select internal or CSV.
Internal waveforms can only be imported to the
oscilloscope in REF mode and cannot be displayed
on any other interface. CSV waveforms can be
L on a PC. It is a group of
voltage and time correlation 2D data.
Select a signal source waveform displayed on the screen
To save the reference waveform on the internal memory of the
oscilloscope, turn the control knob to select
the storage position and press the control

2. Select setup to enter the save setting menu.
Table 2-22 Save setup menu
Set the document name to be imported from the USB device with
the control knob and key For specific
----OKAfter confirming, go back to the reference waveform storage
menu. If there is a USB drive it will be exported to the USB,
otherwise the prompt I/O operation failed will appear
Save the current front panel setup status
To save on the internal memory of the oscilloscope, select a save
position with the control knob and press the

Table 2-22 Save setup menu
Enter the document name to be imported from the USB device
with the control knob and key For
----OKAfter confirming, go back to the save setup menu. If there is a
USB drive it will be exported to the USB, otherwise the prompt
I/O failure will appear when exporting a waveform
After confirming, go back to the save setup menu. If the document
exists on the USB it will be imported, otherwise the prompt I/O
will appear when importing a waveform
Recall 10 internal save positions and select one of them with the
MULTIPURPOSE MULTIPURPOSE

3. Select bitmap to enter the bitmap export menu
: A bitmap can only be exported to a USB device.
Table 2-24 Bitmap export menu
Export the display waveform to the USB device in bmp format
Enter the document name to be imported from the USB device
with the control knob and F key For
specific steps see note P Default is UNIT bmp
After confirming, go back to the save setup menu. If there is a
USB drive it will be exported to the USB, otherwise the prompt

As shown below, the [ ] key in the control zone
is the function key for utility functions.
Press the [ ] key to pop out the setup menu for
Figure 2-25 Function key for the sampling

Table 2-26 Utility menu 1
Run auto system calibration
Display information like oscilloscope model, version and serial
Clear all saved reference waveforms and setups
Enter date and time setup. See Table 2-
Restore factory setup. See Appendix D
Set the auto function. See Table 2-
Return to the Utility menu
Select your desired languages with the
control knob and then press the control
Classical, Traditional, Modern
Set the waiting time for automatic menu shutoff s s s
and manual Manual means automatic shutoff is disabled
Press to turn it on or off
Adjust the brightness of the display grid with the
Return to the Utility menu
Set waveform recording operation

Table 2-27 Utility menu 2
The Cymometer counts the frequency of trigger
events of the trigger channel. It works on edge or
pulse trigger but not alternate trigger. The
Cymometer is also disabled for video trigger
Displayed at the top right corner of the screen when turned on.
Set GPIB and LAN for communication.Applicable for optional
Return to the previous page

Table 2-28 Waveform record menu
1. When you press this button the system plays back and
displays the number of screens to be played back. By turning
the control knob you can suspend
playback Turn the knob to select display of waveforms in
2. If you wish to continue full playback, press [F4] to stop and
Record button. Press this button then the [MENU ON/OFF] key
to record. The number of screens recorded is displayed. You
can record up to 100 screens of data Recorded data will be
lost when the oscilloscope is powered off This function
is disabled in dual time base roll average and envelop

RTC setup Real Time Clock
Turn on the time display. Time display is disabled when the frequency
Press [ ] to select minute or hour, then adjust with the
Press [ ] to select date or month, then adjust with the
Press [ ] to select and adjust with the control knob
Confirm your setting and return to the Utility menu
In the mode, default status is restored. The default status is as follows :
Bandwidth limit : full bandwidth VOLTS DIV coarse tune Inversion Off
In the mode, default status is restored. The default status is as follows :
Sampling mode : normal sampling equivalent sampling Off
In the mode, default status is restored. The default status is as follows :
Type edge signal source adjusted see note trigger coupling DC
Channel is open/closed depending on the presence of input
signal Locked : Channel is maintained open or closed
Return to the Utility menu

: In Auto mode the setup will go back to
: In Auto mode the current status is locked for
For unloc or locked status see the table below 2-30.
The adjustment basis for the trigger signal source in
1. If there is input signal in only one channel, that
channel will be the trigger source.
2. When more than one channels have input signal,
the system will search for trigger signal from CH1.
The first channel where signal is detected will be
3. If no signal is detected in any channel, CH1 will be
You can correct measurement
rs caused by environmental changes with the
auto calibration function. This process can be run
selectively when necessary. To make calibration more
accurate, power on your oscilloscope and allow 20
minutes to warm up. Then press the [ ] button
(help function) and follow on-screen instructions.

As shown below, the [ ] key is the function
key for auto measurement. Read the following
instructions to familiarize with all the powerful
automatic measurement functions of your UTD4000
Four channel digital storage oscilloscope
Operation : Press [ to display 4 function
options, as shown in Table 2-
Table 2- Measurement menu
Figure 2-26 Function key for the sampling
system (auto measurement)
Display all parameters. To quit all parameters press any key or knob
Press [ ] then use the control knob to select the
desired parameter Press the control knob to confirm
Up to 4 parameters can be displayed on screen simultaneously i e
parameter to parameter from left to right
Visual indication of the measurement value s physical properties by lines.
Turn on the indicator and select between parameters 1 and 4
Clear all customized parameters

Table 2- Measurement menu
The rising edge or falling edge time interval of a random channel
Confirm your setting and return to the Utility menu
Select one channel waveform as the delayed reference waveform
Take the centre point of 10%-90% of the waveform rising or falling edge
Select one channel waveform as the delayed measurement waveform
Take the centre point of 10%-90% of the waveform rising or falling edge
Selection of the measurement point of delayed measurement
waveform is complete Return to the measurement menu

Note Automatic measurement of voltage parameters
Customised parameters are for quick parameter
measurement. The measurement menu of your
oscilloscope is capable of measuring 26 waveform
parameters. Two of them (delay and phase for
advance measurement) are generally used when you
do not require to measure all parameters but only a
few. You can set these as customized parameters.
Customized parameters are displayed on the screen.
Your UTD4000 Series Oscilloscope
can automatically measure the following voltage
Peak-to-peak value (Vpp) : The voltage value from the
highest point to lowest point of the waveform.
lue (Vmax) : The voltage value from the
highest point to ground (GND) of the waveform.
Minimum value Vmin The voltage value from
the lowest point to ground GND of the waveform
Amplitude value (Vamp) : The voltage value from top
Mean (Vmid) Half of the amplitude value.
Top value (Vtop) : The voltage value from the level top
to ground (GND) of the waveform.
Base value (Vbase) : The voltage value from the level
base to ground (GND) of the waveform.
Overshoot : The ratio value of the difference between
maximum value and top value to the amplitude value.
Figure 2-27 Measurement parameters

Preshoot : The ratio value of the difference between
minimum value and base value to the amplitude
Average : Average amplitude of signals within 1 cycle.
Root mean square value (Vrms) : The effective value.
Energy generated by AC signal conversion during 1
cycle with respect to the DC voltage that produces
equivalent energy, i.e. root mean square value.
Your UTD4000 Series Oscilloscope
can automatically measure the frequency, cycle,
rising time, falling time, positive pulse width, negative
pulse width delay positive duty ratio negative
duty ratio and time parameters Definition
these parameters are as follows
RiseTime : The time taken by the waveform amplitude
FallTime : The time taken by the waveform amplitude
Positive Pulse Width (+Width) : The pulse width when
positive pulse is at 50% amplitude.
Negative Pulse Width (-Width) : The pulse width when
negative pulse is at 50% amplitude.
Delay (Rising or falling edge) : Delayed time from
wa v e f o r m t o w a ve f o r m (d u r i n g ad v a n c e
Positive duty ratio (+Duty) : Ratio of positive pulse
Negative duty ratio (-Duty) : Ratio of negative pulse
Automatic measurement of time parameters

Press [CURSOR] to display the measurement cursor
and cursor menu, then adjust the cursor position by
turning the control knob As
shown in the figure below in the
control zone is the function key for cursor
You can move the cursor to carry out measurement in
the [ ] mode. There are two modes to choose
from : voltage and time. When measuring voltage,
select voltage for type and adjust the positions of
the two cursors with the control
knob to measure V. Likewise, by selecting time you
can measure T. The [ ] menu has two other
modes : independent and tracking In the
independent mode you can adju
the two cursors independently In the tracking
mode you press the control knob
to move two cursors simultaneously
1. In the independent or tracking mode, select one of
the cursors with the control knob
The selected cursor is a solid line. Adjust its
position with the control knob
2. There are two percentage options. The current
status is 100%. The value between the two
cursors is set to 100% i e one cursor is 0%
and the other is 100% When 100% is set to 6
DIV the 3DIV area on the screen is 100%.
3. When the cursor function is enabled, measurement
value is automatically displayed in the
corner. @ is the selected cursor value.
Figure 2-27 Function key for the sampling

Table 2- ime cursor measurement menu
Table 2- Voltage cursor measurement menu
Measure time with the cursor
Select time as the measurement parameter
Select frequency as the measurement parameter
Select percent as the measurement parameter
Select d as the measurement parameter
Move any one of the two cursors
Move two cursors simultaneously without changing T
Measure voltage with the cursor
Select voltage as the measurement parameter
Select percent as the measurement parameter
Move any one of the two cursors
Move two cursors simultaneously without changing V

There is a button on the top right corner on the front
panel : [ ]. When this button is pressed and
a green indicator lights up, your oscilloscope is in a
running status. If a red light comes on after pressing
this key, it indicates the unit has stopped operation.
Acquire waveform continuously or stop acquisition.
If you want the oscilloscope to acquire waveform
continuously, press [ ] once. Press the
button again to stop acquisition. You can use this
button to switch between acquiring and stop acquiring
waveform. In the Run mode, a green light comes on
and AUTO appears on the screen. In the STOP mode
a red light comes on and STOP appears on the

As shown above, Auto Setup can simplify operation.
Press [ ] and the oscilloscope can automatically
adjust the vertical graticule factor and horizontal time
base range according to the amplitude and frequency
of the waveform, and ensure stable display of the
waveform. When the oscilloscope is in auto setup
mode, the system setup is as follows :
Adjust according to signal frequency
Automatically adjusted. See notes to Table 2-30 for details
Adjust according to amplitude of the signal

Vertical Position Control
To display or hide the current menu
With the menu ON/OFF key you can display or
hide the menu currently on screen but help
messages are not displayed
Operation instructions : Open the multimeter function
menu by pressing the [ ] (Digital Multimeter)
If it is necessary to test electric current, follow these
Insert the current/voltage converter module (UT-M03 or
UT-M04) that comes with the oscilloscope into the
digital multimeter port, and move the switch of the
current/voltage converter so that it is consistent with the
selected DMM current. Proceed with testing.
. Please select a suitable current/voltage converter
module that fits the measurement range, and
select an appropriate range. If you are not certain
about the current being tested, make an
estimation, use the UT-M04(4A) which has the
largest range, then select a suitable module range
2. The current-voltage converter module of the
multimeter cannot be used to measure AC current.
3. When measuring large current with the UT-M04
module, to ensure precision, carry out power on
measurement every 15 minutes Do not power
on for more than 15 seconds each time
elp system with topics that
cover all the features of the oscilloscope. The
oscilloscope displays information about the last menu
displayed on the screen when you push the HELP
.Information about specific menus and controls, such
Push button to display or exit help

Activate the voltage measurement function. Maximum measurement
value for DC voltage is 400V. Maximum measurement value for AC
Activate the resistance measurement function. Maximum measurement
value for resistance is 40M
Activate the on/off measurement function. When on/off measuring is
lower than 70 the beeper will sound
Activate the diode measurement function. Connect the positive pole to
the red , and the negative pole to the black test lead
Current range : 4mA, 40mA, 400mA, 4A
Set the horizontal shift or vertical shift of the window with the

Chapter 3 Practical Example
Illustration 1 : Measuring simple signals
1. To quickly display this signal, follow the
2. Voltage and time parameters for automatic
To observe and measure an unknown signal, and to
quickly display and measure the signal s frequency
In the probe menu, set the attenuation factor to
10X and set the switch on the probe to 10X.
Connect the CH3 probe to the circuitry point to be
The oscilloscope will carry out auto setup to
optimise waveform display. In this status, you can
further adjust the vertical and horizontal range
until you get the desired waveform display.
Your oscilloscope can automatically measure
most display signals. To measure signal
frequency and peak-to-peak value, f
Press [ ] to display the automatic
Press [ ] to enter the measurement type
Use the control knob to select
peak to peak value then press it down to
confirm Next use the control
knob to select frequency then press it down to
Press [ ] to exit the selection

Peak-to-peak and frequency measurements will now
appear at the bottom of the screen, as shown in
As in the previous scenario, set the probe attenuation
factor of the probe and oscilloscope channel to 10X.
Connect the CH1 to the circuit signal input terminal
Connect CH2 to the output terminal
Continue to adjust the horizontal and vertical
range until you get the desired waveform display.
Press [ ] to select CH1. Adjust vertical
position of the CH1 waveform by turning the
vertical position control knob.
Press [ ] to select CH2. In the same way
described above, adjust vertica
CH2 waveform so that the waveforms of CH1 and
CH2 do not overlap. This will make observation
Illustration 2 : Observing the delay caused by a
sine wave signal passes through the circuit
1. To display CH1 and CH2 signals :
Figure 3-1 Automatic measurement

2. Observing the delay caused by a sine wave
signal passes through the circuit and observing
When measuring channel delay automatically :
Press [ ] to display the automatic
measurement menu. Press [ ] to enter next
measurement menu. Press [ ] for advance
m e a s u r e m e n t , the n [ ] f o r d e l a y e d
In the delayed measurement menu, set
measurement from CH1, i.e. from waveform
edge "rising”, to Ch2, i e to waveform edge
"rising” Press OK to confirm.
"Delay” appears on the screen.
Figure 3-2 Waveform delay

Illustration 3 : Acquiring single signals
The advantage and special feature of your digital
storage oscilloscopes lie in its ability to acquire non
cyclical signals like pulse and glitch. To acquire a
single signal, you must have transcendental
knowledge of that signal to set the trigger level and
trigger edge. For example, if the pulse is a logical
signal of TTL level, the trigger level should be set at
about 2V and the trigger edge should be set to rising
edge trigger. If you are not certain about the signal,
you can observe it by automatic or normal trigger to
determine the trigger level and trigger edge.
1. As in the previous illu
stration, set the attenuation
factor of the probe and CH1.
2. Carry out trigger setup.
Press [ ] in the trigger control zone to
display the trigger setup menu.
In this menu, use F1~F5 menus to set the trigger
type to edge set trigger source to Ch1 set
trigger coupling to DC set trigger mode to
single and set slew rate type to rising
Adjust horizontal time base and vertical range to
Turn the knob to get the desired level.
Press [ ] and wait for a signal that
meets the trigger condition. If any signal reaches
the set trigger level, the system will sample once
and display it on the screen
function you can easily acquire any occasional
event. For example, when a sudden glitch of
relatively big amplitude is acquired : set the
trigger level to just higher than the normal signal
level. Press [ ] and begin waiting.
When a glitch occurs, the machine will
automatically trigger and record the waveform
immediately before and after triggering. By
turning the horizontal knob in the
horizontal control zone on the front panel, you
can change the trigger position horizontally to
achieve negative delay trigger of various lengths
for easy observation of waveform occurring

Illustration 4 Reducing random noise of signals
If the signal being measured is stacked with random
noise, you can adjust the setups of your oscilloscope
to filter or reduce the noise, so it will not cause
interference to the signal during measurement.
(Waveform is shown below)
Figure 3-4 Reducing random noise of signals

As in the previous illustration, set the
attenuation factor of the probe and Ch1.
Connect the signal to ensure stable display of
Improving trigger by setting trigger coupling
Reducing display noise by setting the
1.2.See the previous illustration for
operation instruction. See the previous chapter
for guidance on adjusting the horizontal time
Press [ ] in the trigger zone to display the
Set trigger coupling to low frequency
suppression or high frequency suppression
By selecting low frequency suppression a
high pass filter is set up It filters low
frequency signal quantities under kHz and
allows high frequency signal quantities to
pass through If you select high frequency
suppression a low-pass filter is set up. It filters
high frequency signal quan
80kHz and allows low frequency signal quantities
to pass through. By setting low frequency
suppression or high frequency suppression
you can suppress low frequency or high
frequency noise respectively and achieve a
If the signal being measured is stacked with
random noise and the waveform is too coarse as
a result, you can use the average sampling mode
to eliminate random noise display and reduce the
size of waveform for easy observation and
measurement. After getting the mean, random
noise is reduced and details of the signal are
Press [ ] in the menu zone of the front
panel to display the sampling setup menu. Press
menu operation key [ ] then [ ] to set
sampling mode to AVERAGE, then press the
control knob to adjust the
average number of times in multiples of 2

i e 2 to 256 until you get the desired
waveform display that meets observation and
measurement requirements See the figure
Your oscilloscope can measure 28 waveform
parameters automatically. All auto parameters can be
measured with the cursors. By using the cursors, you
can quickly measure the time and voltage of a
To measure the peak frequency of a Sinc signal,
1. Pr es s [ ] to display the cursor
2. Press [ ] to activate the cursor measurement
3. Press [ ] again to set cursor type to time
4. Press [ ] to set the vertical cursor unit t
5. Turn the control knob to set
cursor 1 at the Sinc first peak
Illustration 5 : Using the cursors for measurement
I. Measuring the frequency of a Sinc peak
Figure 3-5 Signal noise suppressed
Caution : In the average sampling mode the
waveform display will update at a slower rate.

6. Press the control knob to
select cursor 2 then turn the
control knob again to set cursor 2 at the Sinc
second peak See the figure below.
1. Pr es s [ ] to display the cursor
2. Press [ ] to activate the cursor measurement
3. Press [ ] again to set cursor type to time
4. Press [ ] to set the vertical cursor unit to ratio
5. Turn the control knob to set
cursor 1 at the first rising edge peak of the
6. Press the control knob to
select cursor 2 then turn the
control knob again to set cursor 2 at the
second rising edge peak of the pulse.
7. Press [ ]. Current position is 100%. Set the
nt value between cursor 1 and cursor
II. Measuring the negative duty ratio of a pulse
Note : When using the cursor to measure voltage,
follow step 2 only and set the cursor type to
Figure 3-6 Using the cursors to test signal

8. Use the control knob to set
cursor 1 at the first falling edge of the pulse
This value is the negative duty ratio.
Measure the phase difference that occurs when a sine
signal passes through the circuit. Connect CH1 to the
input signal of the circuit and connect CH2 to the
output signal of the circuit. To make testing easier,
CH1 and CH2 displacements are set in the centre
point, as shown in Figure 3-8. Follow the steps below :
1. Pr es s [ ] to display the cursor
2. Press [ ] to activate the cursor measurement
3. Press [ ] again to set cursor type to time
ress [ ] to set the vertical cursor unit to Degree
5. Turn the control knob to set
cursor 1 at the center point of the first sine
rising edge where is crosses with channel
III. Measuring the phase difference between two
Figure 3-7 Using the cursors to test pulse

6. Press the control knob to
select cursor 2 then turn the
control knob again to set cursor 2 at the
center point of the second rising edge
7. Press [ ]. Current position is 100%. Set the
current value between cursor 1 and cursor
8. Press [ ], then [ ] to set the cursor at
the CH2 signal being tested. Keep cursor 1
stationary and move cursor 2.
9. Use the control knob to set
cursor 2 at the center point of the first rising edge
of the CH2 signal. This value is the phase
difference between the two signals
Figure 3-8 Using the cursors to test the
phase difference between two

Illustration 6 : Using the X-Y function
To check the phase difference between two channels.
Example : To measure the phase variation when a
signal passes through the circuit, connect your
oscilloscope to the circuit and monitor the input and
output signals. To check the circuit s input and output
in the X-Y coordinate mode, follow the steps below :
1. Set the attenuation factor of the probe to 10X. Set
the switch of the probe to 10X.
2. Connect the CH1 probe to the input terminal of
the network. Connect the CH2 probe to the
output terminal of the network
3. If the channel is not displayed, press the [CH1]
and [ ] menu button to enable the two
5. Adjust the vertical knob to make the
displayed amplitude of the two channels to
6. Press the [ ] menu key to recall the
7. Press [ ] to select X-Y The oscilloscope will
display the circuit s input and output
characteristics in a Lissa jous figure.
8. Adjust the vertical scale and vertical position
knob to achieve the best result of waveform.
9. Using the elliptic oscilloscope display graph to
observe, measure and calculate the phase

10. X-Y Phase difference table :
θ is the angle of disparity between the channels For
definitions of A B C D see the above figure
Calculating with this formula the angle of disparity is
θ arcsin( ) or θ arcsin ( )
If the ellipticmain axis is within quadrants I and III, the angle
ofdisparityshouldbewithinquadrantsIandIV,i.e.inside
If the main axis is within quadrants II and IV, the angle of
disparityshouldbewithinquadrantsIIandIII,i.e. inside