UNI-T UT51-55, UT51, UT52, UT53, UT54 Operating Manual

...
Contents
A Your Meters Feature B Specifications C Making Measurements
D Maintenance E Accessories F Using Holster
Introduction
One kind of UNI-T brand new UT50 series Multimeters is a 3 1/2 digit with steady function and highly reliable hand-held measuring instrument. The Meter uses large scale of integrated circuit with double integrated A/D converter as its core and has full range overload protection. The Meter can measure DC current, AC current, DC Voltage, AC Voltage, Resistance, Capacitance, Diode, Temperature, Frequency and Continuity, which is an ideal tool for users.
Model UT51-55: OPERATING MANUAL
1
Safety Rules
l Do not operate the Meter before the cabinet
has been closed and screwed safely as
terminal can carry voltage.
l Make sure before each measurement the
Meter is set to the suitable range.
l Before using the Meter, please inspect the
cabinet and test leads for damaged insulation
or exposed metal.
l Connect the red and black test lead to the
correct measuring input jack properly.
l Do not input values over the maximum
range of each measurement to avoid damages
of the Meter.
l The UNI-T UT50 series comply with
IEC 61010: in pollution degree 2,
overvoltage category (CAT I 1000V and
CAT II 600V) and double insulation.
Use the Meter only as specified in this
manual, otherwise the protection provided
by the Meter may be impaired.
l CAT I-For signal level, telecommunication,
electronic with small transient over voltage.
l CAT II -For local level, appliances, main
wall outlets, portable equipment.
l The meter is designed to withstand the
stated Max. voltages. If it’s not possible
to exclude without doubts that impulses,
transients, disturbance or for other reasons,
these voltages are exceeded a suitable
prescale (10:1) must be used.
l Do not turn the rotary function switch
during Voltage and Current measurement,
otherwise the Meter could be destroyed.
Model UT51-55: OPERATING MANUAL
2
l Replace the battery as soon as the battery indicator " " appears. With a low battery, the Meter might produce false readings that can lead to electric shock and personal injury. l T urn off the Meter once finished measuring, fetch out the battery, when the meter will not be used for long period. l Do not operate the Meter under adverse environmental condition especially humid area. l To avoid damages and dangerous, do not change the circuit. l Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
l International Electrical Symbols:
Low Battery Earth Ground Safety Rules
Double Insulated
AC
Diode
Dangerous Voltages
DC
Buzzer
Fuse
l Make sure to use new fuses with proper rating in stead of bad fuses . l To avoid electric shock or damages, do not apply more than 1000V between the "COM" terminals and" "earth ground. l Use caution when working with Voltages above 60V (DC) or 30Vrms (AC). These Voltages pose shock hazard.
l The Meter is suitable for indoor use only.
Model UT51-55: OPERATING MANUAL
3
A. Your Meters Feature
l 32 ranges. l Liquid Crystal Display , digits height is 27mm. l Overload display l. l Maximum display 1999 (3 1/2 digit). l Full range overload protection. l Auto-Power Off (For UT53, UT54 and UT55
ONLY). l Temperature: Operating:0
o
C to 40oC (32
0
F to 104 0F).
Storing:-10
o
C to 50oC (14
0
F to 122 0F ).
l Altitude: Operating: 2000m Storing: 10000m. l Relative Humidity:Max. relative humidity 80% for temperature up to 31
o
C decreasing
linearly to 50% relative humidity at 40
o
C.
l Low Battery display " " . l Strap for easy carry.
l Tilt stand design, three observation angles,
is in favor of reading display.
l Dimension: 190mm x 88mm x 34mm. l Weight: Meter only(excluding test leads)
about 270g .Meter + holster + tilt stand about 550g.
l Battery T ype: 9V NEDA 1604 or 6F22 or 006P
Model UT51-55: OPERATING MANUAL
4
B. Specifications
Accuracy is specified for one year after calibration,
at operating temperatures 23
o
C +/- 5oC, with relative
humidity at < 75%.Accuracy specifications take
the form of:+/- (a% readings +digits )
B-1Direct Current Voltage (DC Voltage)
Input impedance: All ranges are 10MΩ . Overload protection: 200mV is 250VDC or AC RMS. All other ranges is 750Vrms or 1000Vp-p.
B-2 Alternating Current Voltage (AC Voltage)
Input impedance: All ranges are 10MΩ. Frequency: 40Hz-400Hz . Overload protection: 200mV is 250VDC or AC RMS. All other ranges is 750Vrms or 1000Vp-p. Display: Average Value (RMS of Sine Wave).
200mV
2V
20V
200V
1000V
100µV
1mV
10mV
100mV
1V
Range Resolution
(0.5% +1)
UT51 UT 52 UT53 UT54 UT55
Accuracy
(0.8% +2)
200mV
2V
20V 200V 750V
100µV
1mV
10mV
100mV
1V
Range Resolution
UT51 UT 52 UT53 UT54 UT55
Accuracy
(1.2% +3)
(0.8% +3) (1.2% +3)
Model UT51-55: OPERATING MANUAL
5
B-3 Direct Current Current (DC Current)
Overload protection:
20µA
200µA
2mA 20mA
200mA
2A 10A 20A
0.01µA
0.1 µA 1 µA
10µA
100 µA
1mA
Range Resolution
UT51 UT 52 UT53 UT54 UT55
Accuracy
(2%+5)
(0.8%+1)
(1.5%+1 )
(0.8%+1)
10mA
(1.5%+1)
(2%+5)
(2%+5 )
For UT51:
2A, 250V fast acting fuse, φ5x20mm(below 2A range)
10A, 250V fast acting fuse,φ5x20mm(at 10A range) .
For UT52/53/54/55:
315mA, 250V fast acting fuse, φ5x20mm (No fuse at 20A range). Max current input : For UT51: 10A (The measurement time for high current should be less than 10 seconds maximum, and the interval time between two measurements should be greater than 15 minutes. For UT52/53/54/55: 20A (The measurement time for high current should be less than 15 seconds maximum, and the interval time between two measurements should be greater than 15 minutes.
Model UT51-55: OPERATING MANUAL
6