UNI-T UT51-55, UT51, UT52, UT53, UT54 Operating Manual

...
Page 1
Contents
A Your Meters Feature B Specifications C Making Measurements
D Maintenance E Accessories F Using Holster
Introduction
One kind of UNI-T brand new UT50 series Multimeters is a 3 1/2 digit with steady function and highly reliable hand-held measuring instrument. The Meter uses large scale of integrated circuit with double integrated A/D converter as its core and has full range overload protection. The Meter can measure DC current, AC current, DC Voltage, AC Voltage, Resistance, Capacitance, Diode, Temperature, Frequency and Continuity, which is an ideal tool for users.
Model UT51-55: OPERATING MANUAL
1
Page 2
Safety Rules
l Do not operate the Meter before the cabinet
has been closed and screwed safely as
terminal can carry voltage.
l Make sure before each measurement the
Meter is set to the suitable range.
l Before using the Meter, please inspect the
cabinet and test leads for damaged insulation
or exposed metal.
l Connect the red and black test lead to the
correct measuring input jack properly.
l Do not input values over the maximum
range of each measurement to avoid damages
of the Meter.
l The UNI-T UT50 series comply with
IEC 61010: in pollution degree 2,
overvoltage category (CAT I 1000V and
CAT II 600V) and double insulation.
Use the Meter only as specified in this
manual, otherwise the protection provided
by the Meter may be impaired.
l CAT I-For signal level, telecommunication,
electronic with small transient over voltage.
l CAT II -For local level, appliances, main
wall outlets, portable equipment.
l The meter is designed to withstand the
stated Max. voltages. If it’s not possible
to exclude without doubts that impulses,
transients, disturbance or for other reasons,
these voltages are exceeded a suitable
prescale (10:1) must be used.
l Do not turn the rotary function switch
during Voltage and Current measurement,
otherwise the Meter could be destroyed.
Model UT51-55: OPERATING MANUAL
2
Page 3
l Replace the battery as soon as the battery indicator " " appears. With a low battery, the Meter might produce false readings that can lead to electric shock and personal injury. l T urn off the Meter once finished measuring, fetch out the battery, when the meter will not be used for long period. l Do not operate the Meter under adverse environmental condition especially humid area. l To avoid damages and dangerous, do not change the circuit. l Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
l International Electrical Symbols:
Low Battery Earth Ground Safety Rules
Double Insulated
AC
Diode
Dangerous Voltages
DC
Buzzer
Fuse
l Make sure to use new fuses with proper rating in stead of bad fuses . l To avoid electric shock or damages, do not apply more than 1000V between the "COM" terminals and" "earth ground. l Use caution when working with Voltages above 60V (DC) or 30Vrms (AC). These Voltages pose shock hazard.
l The Meter is suitable for indoor use only.
Model UT51-55: OPERATING MANUAL
3
Page 4
A. Your Meters Feature
l 32 ranges. l Liquid Crystal Display , digits height is 27mm. l Overload display l. l Maximum display 1999 (3 1/2 digit). l Full range overload protection. l Auto-Power Off (For UT53, UT54 and UT55
ONLY). l Temperature: Operating:0
o
C to 40oC (32
0
F to 104 0F).
Storing:-10
o
C to 50oC (14
0
F to 122 0F ).
l Altitude: Operating: 2000m Storing: 10000m. l Relative Humidity:Max. relative humidity 80% for temperature up to 31
o
C decreasing
linearly to 50% relative humidity at 40
o
C.
l Low Battery display " " . l Strap for easy carry.
l Tilt stand design, three observation angles,
is in favor of reading display.
l Dimension: 190mm x 88mm x 34mm. l Weight: Meter only(excluding test leads)
about 270g .Meter + holster + tilt stand about 550g.
l Battery T ype: 9V NEDA 1604 or 6F22 or 006P
Model UT51-55: OPERATING MANUAL
4
Page 5
B. Specifications
Accuracy is specified for one year after calibration,
at operating temperatures 23
o
C +/- 5oC, with relative
humidity at < 75%.Accuracy specifications take
the form of:+/- (a% readings +digits )
B-1Direct Current Voltage (DC Voltage)
Input impedance: All ranges are 10MΩ . Overload protection: 200mV is 250VDC or AC RMS. All other ranges is 750Vrms or 1000Vp-p.
B-2 Alternating Current Voltage (AC Voltage)
Input impedance: All ranges are 10MΩ. Frequency: 40Hz-400Hz . Overload protection: 200mV is 250VDC or AC RMS. All other ranges is 750Vrms or 1000Vp-p. Display: Average Value (RMS of Sine Wave).
200mV
2V
20V
200V
1000V
100µV
1mV
10mV
100mV
1V
Range Resolution
(0.5% +1)
UT51 UT 52 UT53 UT54 UT55
Accuracy
(0.8% +2)
200mV
2V
20V 200V 750V
100µV
1mV
10mV
100mV
1V
Range Resolution
UT51 UT 52 UT53 UT54 UT55
Accuracy
(1.2% +3)
(0.8% +3) (1.2% +3)
Model UT51-55: OPERATING MANUAL
5
Page 6
B-3 Direct Current Current (DC Current)
Overload protection:
20µA
200µA
2mA 20mA
200mA
2A 10A 20A
0.01µA
0.1 µA 1 µA
10µA
100 µA
1mA
Range Resolution
UT51 UT 52 UT53 UT54 UT55
Accuracy
(2%+5)
(0.8%+1)
(1.5%+1 )
(0.8%+1)
10mA
(1.5%+1)
(2%+5)
(2%+5 )
For UT51:
2A, 250V fast acting fuse, φ5x20mm(below 2A range)
10A, 250V fast acting fuse,φ5x20mm(at 10A range) .
For UT52/53/54/55:
315mA, 250V fast acting fuse, φ5x20mm (No fuse at 20A range). Max current input : For UT51: 10A (The measurement time for high current should be less than 10 seconds maximum, and the interval time between two measurements should be greater than 15 minutes. For UT52/53/54/55: 20A (The measurement time for high current should be less than 15 seconds maximum, and the interval time between two measurements should be greater than 15 minutes.
Model UT51-55: OPERATING MANUAL
6
Page 7
7
B-4 Alternating Current Current (AC Current)
200µA
2mA
20mA
200mA
2A 10A 20A
0.1µA 1µA
10µA
100 µA
1mA
Range
Resolution
UT51 UT 52 UT53 UT54 UT55
Accuracy
(1%+3)
(1%+3)
10mA
(1.8%+3)
(3%+7)
(3%+7 )
(1.8%+3)
(1.8%+3)
Overload protection:
For UT51:
2A, 250V fast acting fuse, φ5x20mm(below 2A range) 10A, 250V fast acting fuse,φ5x20mm(at 10A range) .
For UT52/53/54/55:
315mA, 250V fast acting fuse,φ5x20mm (No fuse at 20A range). Max current input : For UT51: 10A (The measurement time for high current should be less than 10 seconds maximum, and the interval time between two measurements should be greater than 15 minutes. For UT52/53/54/55: 20A (The measurement time for high current should be less than 15 seconds maximum, and the interval time between two measurements should be greater than 15 minutes. Measuring voltage drop: Full range is 200mV. Display: Average Value (RMS of Sine Wave ) .
Model UT51-55: OPERATING MANUAL
Page 8
B- 6 Capacitance
Testing signal: around 400Hz 40mVrms
Range Resolution
UT51 UT 52 UT53 UT54 UT55
Accuracy
(4% +3)
2nF
20nF
200nF
2µF
20µF
1pF
10pF
100pF
1nF
10nF
B-5 Resistance
200
2K
20K
200K
2M
20M
200M
0.1 1
10
100
1K
10K
100K
Range Resolution
UT51 UT 52 UT53 UT54 UT55
Accuracy
(0.8% +1)
(0.8% +3)
(1% +2)
[5%(-10) +10]
B-7 Frequency
Range Resolution
UT51
Accuracy
UT52 UT54
UT55
UT53
2kHz
20kHz
1Hz
10Hz
(2%+5)
(1.5%+5)
Voltage at open circuit: 700mV (200M range, open circuit voltage around 3V) . Overload protection: All ranges 250VDC or AC RMS . Caution: At 200M range, as test lead is short circuit the LCD display 10 digits is normal, deduct the 10 digits from the measured reading during measuring.
Model UT51-55: OPERATING MANUAL
8
Page 9
Overload protection: 250V DC or AC RMS.
B-10 Transistor hFE test
Comment
Measuring Condition
Range
hFE
Can measure NPN or PNP transistor hFE.
Range: 0-1000β
Basic polarity current abt 10µA,Vce abt 2.8V
B-9 Diode T est and Continuity Beeper
Comment
Measuring Condition
Range
Display diode forward­ voltage near value, Unit mV
Forward DC current abt 1mABackward DC voltage abt 2.8V
Voltage at open circuit abt 2.8V
“ ”
“ ”
B-8 T emperature
Range
Resolution
Accuracy
UT51,52,53
UT53
UT55
(5%+3) (1%+3)
2%
-20
o
C
to
1000
o
C
1
o
C
-20
o
C to 0oC
0
o
C to 400oC
400
o
Cto
1000
o
C
Input sensitivity: 100mVrms.
Overload protection: 250Vrms.
Model UT51-55: OPERATING MANUAL
9
Page 10
C. Making Measurements
“ ”
1 On/Off Switch 2 Capacitance jack 3 Liquid Crystal Display 4 Temperature jack 5 Rotary switch 6 Transistor jack 7 Input jack
Model UT51-55: OPERATING MANUAL
10
Page 11
Caution
“ ”
“ ”
“ ”
“ ”
“ ”
“ ”
Caution
“ ”
“ ”
“ ”
“ ”
Model UT51-55: OPERATING MANUAL
11
Page 12
“ ”
“ ”
Caution
“ ”
“ ”
“ ”
“ ”
“ ”
1) Please refer to DC Current Caution 1, 2, 3.
Caution
“ ”
Model UT51-55: OPERATING MANUAL
12
Page 13
“ ”
C-5 Measuring Resistance
“ ”
“ ”
Caution
“ ”
“ ”
C-6 Measuring Capacitance
Before measuring capacitance, remember it takes time for zeroing when changing ranges. Floating reading does not effect accuracy.
1) To avoid damage of the Meter or the equipment under testing, disconnect circuit powers and discharge capacitors before measuring capacitance.
2) Connect capacitor to the capacitance jack.
3) Stabilizing reading takes some time when measuring high capacitance.
4) Unit: 1pF=10-6µF, 1nF = 10-3µF
Model UT51-55: OPERATING MANUAL
13
Page 14
C-7 Measuring Frequency
“ ”
“ ”
“ ”
“ ”
“ ”
“ ”
Model UT51-55: OPERATING MANUAL
14
Page 15
15
D. Maintenance I. General Service
“ ”
“ ”
“ ” “ ”
Model UT51-55: OPERATING MANUAL
Page 16
2. Turn the Meter power OFF
3. Remove the holster from the Meter.
4. Remove rubber feet and screws from the case bottom, and separate the case bottom from the case top.
5. Remove the battery from the battery compartment.
6. Replace the battery with a new 9V battery (NEDA 1604 or 6F22 or 006P).
7. Rejoin the case bottom and case top, and install the screws and rubber feet.
III.Replacing the Fuses
T o avoid electrical shock or arc blast, or personal injury or damage to the Meter, use specified fuses ONLY in accordance with the following procedure.
To replace the Meter’s fuse:
1. Disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.
2. Turn the Meter power OFF.
3. Remove the rubber feet and screws from the case bottom, and separate the case bottom from the case top.
4. Remove the fuse by gently prying one end loose, and then take out the fuse from its bracket.
5. Install ONLY replacement fuses with the identical type and specification as follows and make sure the fuse is fixed firmly in the bracket.
UT51: 2A, 250V fast acting fuse,φ5x20mm (below 2A range) 10A, 250V fast acting fuse,φ5x20mm (at 10A range)
Caution
Model UT51-55: OPERATING MANUAL
16
Page 17
F. Using Holster
Three different ways to use holster:
1. Set holster parallel on the table, do not open the tilt stand (see diagram 1).
2. Set holster in a small angle on the table, tilt it up by the first part of tilt stand (see diagram 2)
3. Set holster in a large angle on the table, tilt it up by all two parts of tilt stand (see diagram 3).
figure1 figure2 figure3
E. Accessories
1. A book of users manual
2. A pair of test lead
3. A pair of WRN-01B bread temperature probe (for UT53 and UT55 ONLY)
4. A piece of holster (if selected)
17
UT52/53/54/55: 315mA, 250V fast acting fuse,
φ 5x20mm
6. Rejoin the case bottom and the case top, and install the screw and rubber feet Replacement of fuses is seldom required. Burning of a fuse always results from the improper operation.
Model UT51-55: OPERATING MANUAL
Page 18
~ END ~
* The manual is subject to changes without separate notice. *
G. Using Strap
1. Put the front end of the strap through the round metal of the Meter, see part 1 of the below diagram.
2. Put the bottom end of the strap through the front part and tide it up, see part 2 of the below diagram.
Model UT51-55: OPERATING MANUAL
18
Page 19
19
Model UT51-55: OPERATING MANUAL
Page 20
Manufacturer: UNI-TREND TECHNOLOGY(DONG GUAN)LIMITED Address: Dong Fang Da Dao, Bei Shan Dong Fang Industrial Development District, Hu Men Town, Dong Guan City, Guang Dong Province, China Headquarters: Uni-Trend International Limited Address: Rm901, 9/F, Nanyang Plaza 57 Hung To Road Kwun Tong Kowloon, Hong Kong Tel: (852) 2950 9168 Fax: (852) 2950 9303 Email: info@uni-trend.com http://www.uni-trend.com
Model UT51-55: OPERATING MANUAL
20