UNI-T UT50D Operating Manual

Page 1
1
Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter Structure Functional Buttons Display Symbols Measurement Operation A. DC Voltage Measurement B. AC Voltage Measurement C. DC Current Measurement D. AC Current Measurement E. Measuring Resistance F. Inductance Measurement G. Capacitance Measurement H. Temperature Measurement I. Measuring Diodes & Continuity Sleep Mode Turning on the Auto Display Backlight General Specifications Accuracy Specifications A. DC Voltage B. AC Voltage C. DC Current D. AC Current E. Resistance Test F. Inductance G. Capacitance H. Temperature I. Diodes and Continuity Test
Table of Contents
3
10
4
11
13 13
15 16
17 19
20 21 23
27 27
28 29
29 29
30 30 31 31 31
5 6 8 9
24
32 32
Model UT50D: OPERATING MANUAL
Page 2
33 33
33 34
Maintenance A. General Service B. Replacing the Battery C. Replacing the Fuses
Model UT50D: OPERATING MANUAL
2
Page 3
Digital Multimeters Model UT50D (hereafter referred to as “the Meter”) is a 3 1/2 digits with steady operations, fashionable structure and highly reliable hand-held measuring instrument. The Meter uses large scale of integrated circuit with double integrated A/D converter as its core and has full range overload protection. The Meter not only can measure AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Temperature, Inductance, Diodes and Continuity, but also has Data Hold, Full Icon Display and Sleep Mode features.
The Meter adopted advanced “co-injection” technique in order to provide sufficient insulation and anti-shaking. In addition, the Automatic Display Backlight feature enables user to work in a dim condition.
Warning To avoid electric shock or personal injury, read the “Safety Information” and “Rules for Safety Operation” carefully before using the Meter.
Model UT50D: OPERATING MANUAL
3
Page 4
1 English Operating Manual 1 piece 2 Test Lead 1 pair 3 Test Clip 1 piece 4 Point Contact Temperature Probe 5 9V Battery (NEDA 1604, 6F22 or 009P) (installed inside the Meter)
1 piece 1 piece
Item
Description
Qty
In the event you find any missing or damage, please contact your dealer immediately.
4
Model UT50D: OPERATING MANUAL
Page 5
Safety Information
This Meter complies with the standards IEC61010: in pollution degree 2, overvoltage category (CAT. II 1000V, CAT. III 600V) and double insulation.
CA T. II: Local level, appliance, POR TABLE EQUIPMENT etc., with smaller transient voltage overvoltages than CAT. III CAT. III: Distribution level, fixed installation, with smaller transient overvoltages than CAT. IV
Use the Meter only as specified in this operating manual, otherwise the protection provided by the Meter may be impaired.
In this manual, a Warning identifies conditions and actions that pose hazards to the user, or may damage the Meter or the equipment under test.
A Note identifies the information that user should pay attention on.
International electrical symbols used on the Meter and in this Operating Manual are explained on page 8.
5
Model UT50D: OPERATING MANUAL
Page 6
Rules For Safe Operation (1)
Before using the Meter inspect the case. Do not use the Meter if it is damaged or the case (or part of the case) is removed. Look for cracks or missing plastic. Pay attention to the insulation around the connectors. Inspect the test leads for damaged insulation or exposed metal. Check the test leads for continuity . Replace damaged test leads with identical model number or electrical specifications before using the Meter. Do not apply more than the rated voltage, as marked on the Meter, between the terminals or between any terminal and grounding. The rotary switch should be placed in the right position and no any changeover of range shall be made during measurement is conducted to prevent damage of the Meter. When the Meter working at an effective voltage over 60V in DC or 30V rms in AC, special care should be taken for there is danger of electric shock. Use the proper terminals, function, and range for your measurements. Do not use or store the Meter in an environment of high temperature, humidity, explosive, inflammable and strong magnetic field. The performance of the Meter may deteriorate after dampened. When using the test leads, keep your fingers behind the finger guards.
6
Warning To avoid possible electric shock or personal injury, and to avoid possible damage to the Meter or to the equipment under test, adhere to the following rules:
l
l
l
l
l
l l
l
Model UT50D: OPERATING MANUAL
Page 7
Rules For Safe Operation (2)
Disconnect circuit power and discharge all high
-voltage capacitors before testing resistance, continuity, diodes, capacitance or current. Before measuring current, check the Meter’s fuses and turn off power to the circuit before connecting the Meter to the circuit. Replace the battery as soon as the battery indicator appears. With a low battery, the Meter might produce false readings that can lead to electric shock and personal injury. Remove test leads, test clips and temperature probe from the Meter and turn the Meter power off before opening the Meter case. When servicing the Meter, use only the same model number or identical electrical specifications replacement parts. The internal circuit of the Meter shall not be altered at will to avoid damage of the Meter and any accident. Soft cloth and mild detergent should be used to clean the surface of the Meter when servicing. No abrasive and solvent should be used to prevent the surface of the Meter from corrosion, damage and accident. The Meter is suitable for indoor use. Turn the Meter power off when it is not in use and take out the battery when not using for a long time. Constantly check the battery as it may leak when it has been using for some time, replace the battery as soon as leaking appears. A leaking battery will damage the Meter.
l
l
l
l
l
l
l
l l
l
7
Model UT50D: OPERATING MANUAL
Page 8
International Electrical Symbols
AC (Alternating Current). DC (Direct Current). Grounding. Double Insulated. Deficiency of Built-In Battery. Warning. Refer to the Operating Manual.
Conforms to Standards of European Union.
Diode. AC or DC.
Fuse. Continuity Test.
Model UT50D: OPERATING MANUAL
8
Page 9
9
(figure 1)
The Meter Structure (see figure 1)
1. LCD Display
2. Data Hold Button.
3. Rotary Switch
4. Other Input Terminals
5. COM Input Terminal
6. 20A Input Terminal
7. mA Input terminal
8. Power
Model UT50D: OPERATING MANUAL
Page 10
Button Operation Performed
Turn the Meter on and off. l Press down the POWER to turn on the Meter. l Press up the POWER to turn off the Meter.
HOLD
(Blue Button)
l Press HOLD once to enter hold mode. l Press HOLD again to exit hold mode. l In Hold mode, is displayed and the present value is shown.
POWER
(Yellow Button)
Functional Buttons
Below table indicated for information about the functional button operations
Model UT50D: OPERATING MANUAL
10
Page 11
(figure 2)
Display Symbols(1) (see figure 2)
No. Symbol Meaning
1
Dangerous Voltages.
2
The battery is low.
Warning: To avoid false readings, which could lead to possible electric shock or personal injury , replace the battery as soon as the battery
indicator appears.
3
Indicator for AC voltage or current.The displayed value is the mean value.
4
Indicates negative reading.
5
Test of diode.
6
Data hold is active.
7
The continuity buzzer is on.
8
Centigrade temperature
o
C
o
F
9
Fahrenheit temperature
10
V: Volts. The unit of voltage. mV: Millivolt. 1 x 10-3 or 0.001 volts.
mV, V
11
Model UT50D: OPERATING MANUAL
11
Page 12
No. Symbol Meaning
12
mH,H
13
14
H: Henry. The unit of Inductance. mH: Millihenry. 1 x 10
-3
or 0.001
henry.
Display Symbols(2) (see figure 2)
12
Model UT50D: OPERATING MANUAL
Page 13
Make sure the Sleep Mode is not on if you found there is no display on the LCD after turning on the Meter. Make sure the Low Battery Display is not on, otherwise false readings may be provided.
Pay extra attention to the
symbol which is located besides the input terminals of the Meter before carrying out measeurement.
l
l l
(figure 3)
The DC Voltage ranges are: 200mV, 2V, 20V, 200V and 1000V. To measure DC voltage, connect the Meter as follows:
13
Model UT50D: OPERATING MANUAL
Page 14
14
Model UT50D: OPERATING MANUAL
l If the value of voltage to be measured is unknown, use the maximum measurement position (1000V) and reduce the range step by step until a satisfactory reading is obtained.
l The LCD displays “1” indicating the existing selected range is overloaded, it is required to select a higher range in order to obtain a correct reading. l In each range, the Meter has an input impedance of approx. 10M . This loading effect can cause measurement errors in high impedance circuits. If the circuit impedance is less than or equal to 10k , the error is negligible (0.1% or less). l When DC voltage measurement has been completed, disconnect the connection between the testing leads and the circuit under test.
1. Insert the red test lead into thei V nput terminal and the black test lead into the COM input terminal.
2. Set the rotary switch to an appropriate measurement position in V range.
3. Connect the test leads across with the object being measured. The measured value shows on the display.
Page 15
The AC voltage measurement has 4 measurement positions on the rotary switch: 2V, 20V, 200V and 750V To measure AC Voltage, connect the Meter as follows:
l If the value of voltage to be measured is unknown, use the maximum measurement position (750V) and reduce the range step by step until a satisfactory reading is obtained. l The LCD displays “1” indicating the existing selected range is overloaded, it is required to select a higher range in order to obtain a correct reading. l In each range, the Meter has an input impedance of approx. 10M . This loading effect can cause measurement errors in high impedance circuits. If the circuit impedance is less than or equal to 10k , the error is negligible (0.1% or less). l When AC voltage measurement has been completed, disconnect the connection between the testing leads and the circuit under test.
1. Insert the red test lead into the V terminal and the black test lead into the COM terminal.
2. Set the rotary switch to an appropriate measurement position in V range.
3. Connect the test leads across with the object being measured. The measured value shows on the display, which is effective value of sine wave (mean value response).
15
Model UT50D: OPERATING MANUAL
Page 16
( figure 4)
The DC current measurement has 3 measurement positions on the rotary switch: 2mA, 200mA and 20A.
To measure current, do the following:
1. Turn off power to the circuit. Discharge all high­ voltage capacitors.
2. Insert the red test lead into the mA or 20A terminal and the black test lead into the COM terminal.
3. Set the rotary switch to an appropriate measurement position in A range.
Model UT50D: OPERATING MANUAL
16
Page 17
17
Measurement Operation(5)
4. Break the current path to be tested. Connect the red test lead to the more positive side of the break and the black test lead to the more negative side of the break.
5. Turn on power to the circuit. The measured value shows on the display.
l If the value of current to be measured is unknown, use the maximum measurement position (20A) and 20A terminal, and reduce the range step by step until a satisfactory reading is obtained. l When current measurement has been completed, disconnect the connection between the testing leads and the circuit under test.
( figure 4)
Model UT50D: OPERATING MANUAL
Page 18
Measurement Operation(6)
The AC current measurement has 3 measurement positions on the rotary switch: 20mA, 200mA and 20A. To measure current, do the following:
1. Turn off power to the circuit. Discharge all high-voltage capacitors.
2. Insert the red test lead into the mA or 20A terminal and the black test lead into the COM terminal.
3. Set the rotary switch to an appropriate measurement position in A range.
4. Break the current path to be tested. Connect the red test lead to the more positive side of the break and the black test lead to the more negative side of the break.
5. Turn on power to the circuit. The measured value shows on the display.
l If the value of current to be measured is unknown, use the maximum measurement position (20A) and 20A terminal and reduce the range step by step until a satisfactory reading is obtained. l When current measurement has been completed, disconnect the connection between the testing leads and the circuit under test.
Model UT50D: OPERATING MANUAL
18
Page 19
Measurement Operation(7)
( figure 5)
The resistance ranges are:200 ,2k ,200k ,2M ,and 20M . To measure resistance, connect the Meter as follows:
1. Insert the red test lead into the V terminal and the black test lead into the COM terminal.
2. Set the rotary switch to an appropriate measurement position in range.
3. Connect the test leads across with the object being measured. The measured value shows on the display.
l The test leads can add 0.1 to0.3 of error when measuring low resistance, that is the range of 200 .
To obtain precision readings in low-resistance, use
the following equation:
Model UT50D: OPERATING MANUAL
19
Page 20
Precision readings of resistance = Measurement displayed reading minus the short circuit value of red and black test lead. l For high resistance (>1M ), it is normal taking several seconds to obtain a stable reading. l When there is no input, for example in open circuit condition, the Meter displays “1”. l When resistance measurement has been completed, disconnect the connection between the testing leads and the circuit under test.
Measurement Operation(8)
( figure 6)
Inductance measurement has 4 measurement positions on the rotary switch: 2mH, 20mH, 200mH and 20H.
To test for Inductance, connect the Meter as below:
20
Model UT50D: OPERATING MANUAL
Page 21
1. According to the size of the tested object’s leads, insert multi-purpose socket or test clip into the mA and V terminal.
2. Set the rotary switch to an appropriate measurement position in HLx range.
3. Insert the tested object into the corresponding jack of the multi-purpose socket or connect the test clip to the object being measured. The measured value shows on the display.
Measurement Operation(9)
l If the value of inductance to be measured is unknown, use the maximum measurement position, and reduce the range step by step until a satisfactory reading is obtained. l When inductance testing has been completed, remove the multi-purpose socket or test clip from the input terminal, and remove multi-purpose socket or test clip away from the input terminal of the Meter.
( figure 7)
21
Model UT50D: OPERATING MANUAL
Page 22
22
Model UT50D: OPERATING MANUAL
Measurement Operation(10)
l For testing the capacitor with polarity, connect the red test lead or test clip to anode & black test lead or test clip to cathode l When the tested capacitor is shorted or the capacitor value is overloaded, the LCD display “1”. l To minimize the measurement error caused by the distributed capacitor, the testing lead or testing clip should be as short as possible. l To ensure precision especially when measuring small capacitance 200nF range, the correct reading is as follows: Measurement displayed reading minus the Meter’s open circuit displayed value. l It is normal to take a while for zeroing when changing over the measurement range. This process will not affect the accuracy of the final readings obtained.
30V
Page 23
Measurement Operation(11)
( figure 8)
The temperature measurement range is from -40oC ~1000oC. To measure temperature, connect the Meter as follows:
1. Insert the red temperature into the V terminal and the black temperature probe into the COM terminal.
2. Set the rotary switch to .
3. Place the temperature probe to the object being measured. The measured value shows on the display.
l The Meter automatically displays the temperature value inside the Meter when there is no temperature probe connection. l The included temperature probe can only be measured up to 250 . For any mesaurement higher than that, the rod type temperature probe must be used instead.
23
Model UT50D: OPERATING MANUAL
Page 24
l When temperature measeuremnet has been completed, disconnect the connection between the testing leads and te circuit under test.
Measurement Operation(12)
( figure 9)
Testing Diodes Use the diode test to check diodes, transistors, and other
semiconductor devices. The diode test sends a current through the semiconductor junction, and then measures the voltage drop across the junction. A good silicon junction drops between 0.5V and 0.8V.
To test out a diode out of a circuit, connect the Meter as follows:
1. Insert the red test lead into the V terminal and the black test lead into the COM terminal.
Model UT50D: OPERATING MANUAL
24
Page 25
25
2. Set the rotary switch to
.
3. For forward voltage drop readings on any semiconductor component, place the red test lead on the component’s anode and place the black test lead on the component’s cathode. The measured value shows on the display.
Measurement Operation(13)
l In a circuit, a good diode should still produce a forward voltage drop reading of 0.5V to 0.8V; however; the reverse voltage drop reading can vary depending on the resistance of other pathways between the probe tips. l Connect the test leads to the proper terminals as said above to avoid error display . The LCD will display “1” indicating open-circuit for wrong connection.The unit of diode is Volt (V), displaying the positive­ connection voltage-drop value. l When diode testing has been completed, disconnect the connection between the testing leads and the circuit under test.
To test for continuity, connect the Meter as below:
1. Insert the red test lead into V terminal and the black test lead into the COM terminal.
2. Set the rotary switch to
.
3. Connect the test leads across with the object being measured. The buzzer sounds if the resistance of a circuit under test is less than 70 The LCD displays the resistance value of a circuit under test.
Model UT50D: OPERATING MANUAL
Page 26
l The LCD display “1” indicating the circuit being tested is open. l When continuity testing has been completed, disconnect the connection between the testing leads and the circuit under test.
Measurement Operation(14)
Model UT50D: OPERATING MANUAL
26
Page 27
Model UT50D: OPERATING MANUAL
27
Sleep Mode
To preserve battery life, the Meter automatically turns off if you do not turn the rotary switch or press any button for around 8 minutes. At that time, the Meter consumes around 10µA current. The Meter can be activated by pressing the POWER two times.
Turning on the Auto Display Backlight
The Meter has a built-in sensor. Therefore the Display Backlight turns on and off automatically depending on the brightness of the environment. In a dim condition, the Display Backlight turns on automatically while in a bright condition, the Display Backlight turns off automatically.
Page 28
28
Model UT50D: OPERATING MANUAL
l Maximum voltage between any Terminals and Grounding :1000V rms.
l
Fused Protection for mA Input Terminal :0.5A, 250V fast type, 5x20mm.
l
Fused Protection for 20A Input Terminal :Un-fused.
l Range :Manual ranging. l Maximum Display :1999. l Mesaurement Speed :Updates 2~3 times/second. l Temperature:
Operating:0oC~40oC(32oF~104oF)
Storage: -10oC~50oC(14oF~122oF). l Relative Humidity : 75% @ 0oC~30oC; 50%@31oC~40oC.
l Altitude: Operating :2000m; Storage: 10000m. l Battery Type :9V NEDA1604 or 6F22 or
006P.
l Battery Deficiency :Display “ ”. l Data Holding :Display “ H l Negative reading :Display “
l Overloading :Display “1”. l Dimensions (HxWxL) :165x80x38.3mm. l Weight : Approx.275g (battery
included). l Safety/Compliances :IEC61010 CAT II 1000V, CAT III 600V overvoltage and double insulation standard. l Certificate:
Page 29
A. DC Voltage
200mV
2V
20V
200V
1000V
Range Resolution
(0.5%+1)
Overload Protection
Accuracy
250V DC or AC rms.
(0.8%+2)
0.1mV
0.001V
0.01V
0.1V 1V
1000V DC /
750V ACrms.
B. AC Voltage
Range Resolution
(0.8%+3)
Overload ProtectionAccuracy
2V
20V 200V 750V
0.001V
0.01V
0.1V 1V
(1.2%+3)
250V DC or AC rms.
1000V DC /
750V ACrms.
29
Model UT50D: OPERATING MANUAL
Accuracy: (a% reading + b digits) Operating temperature:23oC
5oC.
Relative humidity:<75%. Temperature coefficient: 0.1 x (specified accuracy) / 1oC.
guarantee for 1 year.
Page 30
30
Model UT50D: OPERATING MANUAL
Remarks:
l At 20A Range: For continuous measurement
10
seconds and interval not less than 15 minutes.
l Measurement voltage drop: Full range at 200mV.
D. AC Current
Remarks:
l At 20A Range: For continuous measurement
10
seconds and interval not less than 15 minutes.
l Measurement voltage drop: Full range at 200mV. l Frequency reaponse: 40Hz~400Hz l Display effective value of sine wave (mean value
response).
Range Resolution Overload ProtectionAccuracy
20mA
200mA
20A
0.01mA
0.1mA 10mA
0.5A. 250V fast type fuse, 5x20mm
(1%+3)
Un-Fused
(1.8%+3)
(3%+5)
C. DC Current
Range Resolution Overload ProtectionAccuracy
2mA
200mA
20A
0.001mA
0.1mA 10mA
0.5A. 250V fast type fuse, 5x20mm
(0.8%+1)
(2%+5)
Un-Fused
(1.5%+1)
Page 31
E. Resistance Test
Remarks:
Range Resolution
Overload Protection
Accuracy
(0.8%+3)
250V rms
(0.8%+1)
(1%+5)
Range Resolution Overload ProtectionAccuracy
2mH
20mH
200mH
20H
250V rms
10pF
100pF
1nF
100nF
Range Resolution
Overload Protection
Accuracy
250V rms
31
Model UT50D: OPERATING MANUAL
l At 200 range, short circuit the test lead first to display the test lead resistance value. During measurement, subtract the test lead resistance value from the displayed reading to obtain the accurate reading.
(2%+10)
(3%+10)
(2.5%+5)
(5%+4)
Page 32
H. Temperature
Range
Resolution
(3%+3)
Overload Protection
Accuracy
-40oC~0oC
0oC~400oC
400oC~1000oC
(1%+3)
1oC
2.5%
250V rms
Open circuit voltage approx.
2.8V
250V DC
or AC
<70 buzzer beeps continuously
Diodes
Continuity
Test
1mV
Model UT50D: OPERATING MANUAL
32
I. Diodes and Continuity Test
Page 33
33
Model UT50D: OPERATING MANUAL
This section provides basic maintenance information including battery and fuse replacement instruction.
Warning Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service information. To avoid electrical shock or damage to the Meter, do not get water inside the case.
A. General Service
l Periodically wipe the case with damp cloth and mild detergent. Do not use chemical solvent. l To clean the terminals with cotton bar with detergent, as dirt or moisture in the terminals can affect readings. l Turn the Meter OFF when it is not in use and take out the battery when not using for a long time. l Do not store the Meter in place of humidity, high temperature, explosive, inflammable and strong magnetic field
Screw
Fuse
Fuse
Rubber foot
( figure 10)
Page 34
Model UT50D: OPERATING MANUAL
34
Warning To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator appears.
To replace battery:
1. Disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.
2. Turn the Meter off
3. Remove the rubber feet and screws from the battery compartment, and separate the battery compartment from the case bottom
4. Remove the battery from the battery compartment.
5. Replace the battery with a new 9V battery (NEDA 1604 or 6F22 or 006P).
6. Rejoin the battery compartment and the case bottom, and install the screw and rubber feet.
Screw
Fuse
Fuse
Rubber foot
( figure 10)
Page 35
~ END ~
This operating manual is subject to change without notice.
35
Model UT50D: OPERATING MANUAL
To replace the Meter’s fuse:
1. Disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.
2. Turn the Meter off.
3. Remove the rubber feet and screws from the case bottom, and separate the case bottom from the case top.
4. Remove the fuse by gently prying one end loose, and then take out the fuse from its bracket.
5. Install ONL Y replacement fuses with the identical type and specification as follows and make sure the fuse is fixed firmly in the bracket.
0.5A. 250V fast type fuse, 5x20mm
6. Rejoin the case bottom and the case top, and install the screw and rubber feet
Replacement of fuses is seldom required. Burning of a fuse always results from the improper operation.
Page 36
Copyright 2003 Uni-Trend Group Limited.
All rights reserved. Manufacturer:
Uni-Trend Technology (Dongguan) Limited Dong Fang Da Dao Bei Shan Dong Fang Industrial Development District Hu Men Town, Dongguan City Guang Dong Province China Postal Code: 523 925
Headquarters: Uni-Trend Group Limited Rm901, 9/F, Nanyang Plaza 57 Hung To Road Kwun Tong Kowloon, Hong Kong Tel: (852) 2950 9168 Fax: (852) 2950 9303 Email: info@uni-trend.com http://www.uni-trend.com
Model UT50D: OPERATING MANUAL
36