UNI-T UT50A, UT50B, UT50C Operating Manual

Page 1
Page
3 4 5 6 8 9
10
1 1 13 13
14 16 17 19
20 21
22 23
24 25 25 26 27
27 27
28 28
Model UT50A/B/C: OPERATING MANUAL
1
Title
Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The multimeter Structure Functional Buttons
Display Symbols Measurement Operation
A. DC Voltage Measurement B. AC Voltage Measurement C. DC Current Measurement
D. AC Current Measurement
E. Measuring Resistance F. The Model UT50C:
Frequency Measurement G. The Model UT50B/UT50C:
Temperature Measurement
H. Capacitance Measurement I. Measuring Diodes & Continuity
General Specifications
Testing for Continuity Sleep Mode Turning on the Auto Display Backlight
Accuracy Specifications
A. DC Voltage B. AC Voltage C. DC Current D. AC Current
Page 2
Page
29 29
29
I. Diodes and Continuity Test
30 30
C. Replacing the Fuses
31 32
32 32
Model UT50A/B/C: OPERATING MANUAL
2
Title
E. Resistance Test F. The Model UT50C: Frequency G. The Model UT50B/UT50C:
Temperature H. Capacitance
Maintenance
A. General Service B. Replacing the Battery
Page 3
Overview
Warning To avoid electric shock or personal injury, read the "Safety Information" and "Rules for Safe Operation" carefully before using the Meter.
Digital Multimeters Model UT50A, UT50B and UT50C (hereafter referred to as “the Meter”) are 3 1/2 digits with steady operations, fashionable structure and highly reliable hand-held measuring instrument. The Meter uses large scale of integrated circuit with double integrated A/D converter as its core and has full range overload protection. The Meter not only can measure AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Temperature, Frequency , Diodes and Continuity, but also has Data Hold, Full Icon Display and Sleep Mode features.
The Meter adopted advanced “co-injection” technique in order to provide sufficient insulation and anti-shaking. In addition, the Automatic Display Backlight feature enables user to work in a dim condition.
Model UT50A/B/C: OPERATING MANUAL
3
Page 4
Unpacking Inspection
Open the package case and take out the Meter. Check the following items carefully to see any missing or damaged part:
In the event you find any missing or damage, please contact your dealer immediately.
Model UT50A/B/C: OPERATING MANUAL
4
Item
1 2
3 4
Description
English Operating Manual Test Lead Test Clip
Point Contact Temperature Probe (UT50B/UT50C only)
Qty
1 piece
1 pair 1 pair
1 piece
Page 5
Safety Information
This Meter complies with the standards IEC61010: in pollution degree 2, overvoltage category (CA T. II 1000V, CAT. III 600V) and double insulation.
CA T. II: Local level, appliance, PORTABLE EQUIPMENT etc., with smaller transient voltage overvoltages than CAT. III CA T. III: Distribution level, fixed installation, with smaller transient overvoltages than CAT. IV
Use the Meter only as specified in this operating manual, otherwise the protection provided by the Meter may be impaired.
In this manual, a Warning identifies conditions and actions that pose hazards to the user, or may damage the Meter or the equipment under test.
A Note identifies the information that user should pay attention on. International electrical symbols used on the Meter and in this Operating Manual are explained on page 8.
Model UT50A/B/C: OPERATING MANUAL
5
Page 6
Rules For Safe Operation (1)
Warning To avoid possible electric shock or personal injury, and to avoid possible damage to the Meter or to the equipment under test, adhere to the following rules:
l Before using the Meter inspect the case. Do not use the Meter if it is damaged or the case (or part of the case) is removed. Look for cracks or missing plastic. Pay attention to the insulation around the connectors. l Inspect the test leads for damaged insulation or exposed metal. Check the test leads for continuity. Replace damaged test leads with identical model number or electrical specifications before using the Meter. l Do not apply more than the rated voltage, as marked on the Meter, between the terminals or between any terminal and grounding. l The rotary switch should be placed in the right position and no any changeover of range shall be made during measurement is conducted to prevent damage of the Meter. l When the Meter working at an effective voltage over 60V in DC or 30V rms in AC, special care should be taken for there is danger of electric shock. l Use the proper terminals, function, and range for your measurements. l Do not use or store the Meter in an environment of high temperature, humidity, explosive, inflammable and strong magnetic field. The performance of the Meter may deteriorate after dampened.
l When using the test leads, keep your fingers behind the finger guards.
Model UT50A/B/C: OPERATING MANUAL
6
Page 7
Rules For Safe Operation (2)
l Disconnect circuit power and discharge all high­ voltage capacitors before testing resistance, continuity, diodes, capacitance or current. l Before measuring current, check the Meter’s fuses and turn off power to the circuit before connecting the Meter to the circuit. l Replace the battery as soon as the battery indicator appears. With a low battery , the Meter might produce false readings that can lead to electric shock and personal injury. l Remove test leads, test clips and temperature probe from the Meter and turn the Meter power off before opening the Meter case. l When servicing the Meter, use only the same model number or identical electrical specifications replacement parts. l The internal circuit of the Meter shall not be altered at will to avoid damage of the Meter and any accident. l Soft cloth and mild detergent should be used to clean the surface of the Meter when servicing. No abrasive and solvent should be used to prevent the surface of the Meter from corrosion, damage and accident.
l The Meter is suitable for indoor use. l Turn the Meter power off when it is not in use
and take out the battery when not using for a long time. l Constantly check the battery as it may leak when it has been using for some time, replace the battery as soon as leaking appears. A leaking battery will damage the Meter.
Model UT50A/B/C: OPERATING MANUAL
7
Page 8
International Electrical Symbols
AC (Alternating Current). DC (Direct Current). AC or DC. Grounding. Double Insulated. Deficiency of Built-In Battery. Continuity Test. Diode. Capacitance Test. Fuse. Warning. Refer to the Operating Manual. Conforms to Standards of European Union.
Model UT50A/B/C: OPERATING MANUAL
8
Page 9
The multimeter Structure (see figure 1)
9
( figure 1)
1 LCD Display 2 Data Hold Button. 3 Rotary Switch 4 Other Input Terminals 5 COM Input Terminal 6 20A Input Terminal
7 The Model UT50A:µAmA Input terminal The Model UT50B/UT50C:mA Input terminal
8 Power
Model UT50A/B/C: OPERATING MANUAL
Page 10
10
Functional Buttons
Button Operation Performed
Turn the Meter on and off. l Press down the POWER to turn on the Meter. l Press up the POWER to turn off the Meter.
HOLD
(Blue Button)
l Press HOLD once to enter hold mode. l Press HOLD again to exit hold mode. l In Hold mode, is displayed and the present value is shown.
POWER
(Yellow Button)
Below table indicated for information about the functional button operations
Model UT50A/B/C: OPERATING MANUAL
Page 11
11
Display Symbols (see figure 2)
( figure 2)
No. Symbol Meaning
1
Dangerous Voltages.
2
The battery is low.
Warning: To avoid false readings, which could lead to possible electric shock or personal injury , replace the battery as soon as the battery
indicator appears.
3
Indicator for AC voltage or current.The displayed value is the mean value.
4
Indicates negative reading.
5
Test of diode.
6
Data hold is active.
7
The continuity buzzer is on.
8
Centigrade temperature
o
C
o
F
9
Fahrenheit temperature
µA, mA, A
A: Amperes (amps). The unit of current. mA: Milliamp. 1 x 10-3 or 0.001 amperes.
µA: Microamp. 1x 10
-6
or 0.000001
amperes.
10
V: Volts. The unit of voltage. mV: Millivolt. 1 x 10-3 or 0.001 volts.
mV, V
11
Model UT50A/B/C: OPERATING MANUAL
10-14
Page 12
12
Display Symbols (see figure 2)
No. Symbol Meaning
nF, µF
12
F: Farad. The unit of capacitance. µF: Microfarad.1 x 10
-6
or 0.000001 farads. nF: Nanofarad. 1 x 10-9 or 0.000000001 farads.
kHz
13
The unit of frequency in cycles/ second. Kilohertz. 1 x 10-3 or
1,000 hertz.
: Ohm. The unit of resistance. k: kilohm. 1 x 103 or 1000
ohms. M: Megaohm. 1 x 106 or 1,000,000
ohms.
,k,M
14
Model UT50A/B/C: OPERATING MANUAL
Page 13
Measurement Operation(1)
A. DC Voltage Measurement (see figure 3)
Warning To avoid harms to you or damages to the Meter from electric shock, please do not attempt to measure voltages higher than 1000V or 750V rms although readings may be obtained.
The DC Voltage ranges are: 200mV, 2V, 20V, 200V and 1000V. To measure DC voltage, connect the Meter as follows:
( figure 3)
13
1. Insert the red test lead into the V input terminal and the black test lead into the COM input terminal.
2. Set the rotary switch to an appropriate measurement position in V range.
3. Connect the test leads across with the object being measured. The measured value shows on the display.
l Make sure the Sleep Mode is not on if you found there is no display on the LCD after turning on the Meter. l Make sure the Low Battery Display is not on, otherwise false readings may be provided. l Pay extra attention to the
symbol which is located besides the input terminals of the Meter before carrying out measeurement.
red black
Model UT50A/B/C: OPERATING MANUAL
Page 14
14
Measurement Operation(2)
Note
l If the value of voltage to be measured is unknown, use the maximum measurement position (1000V) and reduce the range step by step until a satisfactory reading is obtained. l The LCD displays “1” indicating the existing selected range is overloaded, it is required to select a higher range in order to obtain a correct reading. l In each range, the Meter has an input impedance of approx. 10MThis loading effect can cause measurement errors in high impedance circuits. If the circuit impedance is less than or equal to 10k, the error is negligible (0.1% or less). l When DC voltage measurement has been completed, disconnect the connection between the testing leads and the circuit under test.
B. AC Voltage Measurement (see figure 4)
Warning To avoid harms to you or damages to the Meter from electric shock, please do not attempt to measure voltages higher than 1000V or 750V rms although readings may be obtained.
( figure 4)
red black
Model UT50A/B/C: OPERATING MANUAL
Page 15
15
The Model UT50A: The AC voltage measurement has
5 measurement positions on the rotary switch: 200mV, 2V, 20V, 200V and 750V The Model UT50B/UT50C: The AC voltage measurement has 4 measurement positions on the rotary switch: 2V, 20V, 200V and 750V
To measure AC Voltage, connect the Meter as follows:
1. Insert the red test lead into the V terminal and the black test lead into the COM terminal.
2. Set the rotary switch to an appropriate measurement position in V range.
3. Connect the test leads across with the object being measured. The measured value shows on the display, which is effective value of sine wave (mean value response).
Note
l If the value of voltage to be measured is unknown, use the maximum measurement position (750V) and reduce the range step by step until a satisfactory reading is obtained.
l The LCD displays “1” indicating the existing selected range is overloaded, it is required to select a higher range in order to obtain a correct reading. l In each range, the Meter has an input impedance of approx. 10M This loading effect can cause measurement errors in high impedance circuits. If the circuit impedance is less than or equal to 10k, the error is negligible (0.1% or less). l When AC voltage measurement has been completed, disconnect the connection between the testing leads and the circuit under test.
Measurement Operation(3)
Model UT50A/B/C: OPERATING MANUAL
Page 16
16
Measurement Operation(4)
C. DC Current Measurement (see figure 5)
Warning Never attempt an in-circuit current measurement where the open circuit voltage between terminals and ground is greater than 60V DC or 30V rms. If the fuse burns out during measurement, the Meter may be damaged or the operator himself may be hurt. Use proper terminals, function, and range for the measurement. When the testing leads are connected to the current terminals, do not parallel them across any circuit.
The Model UT50A: The DC current measurement has
4 measurement positions on the rotary switch: 20µA, 2mA, 200mA and 20A. The Model UT50B: The DC current measurement has 4 measurement positions on the rotary switch: 2mA, 20mA, 200mA and 20A. The Model UT50C: The DC current measurement has 3 measurement positions on the rotary switch: 20mA, 200mA and 20A.
To measure current, do the following:
1. Turn off power to the circuit. Discharge all high­ voltage capacitors.
( figure 5)
red
black
Model UT50A/B/C: OPERATING MANUAL
Page 17
2. Insert the red test lead into the µAmA (The Model
UT50B/UT50C) or 20A terminal and the black test
lead into the COM terminal.
3. Set the rotary switch to an appropriate measurement position in A range.
4. Break the current path to be tested. Connect the red test lead to the more positive side of the break and the black test lead to the more negative side of the break.
5. Turn on power to the circuit. The measured value shows on the display.
Measurement Operation(5)
Note
l If the value of current to be measured is unknown, use the maximum measurement position (20A) and 20A terminal, and reduce the range step by step until a satisfactory reading is obtained. l When current measurement has been completed, disconnect the connection between the testing leads and the circuit under test.
D. AC Current Measurement (see figure 6)
( figure 6)
17
red
black
Model UT50A/B/C: OPERATING MANUAL
Page 18
18
Measurement Operation(6)
Warning Never attempt an in-circuit current measurement where the voltage between terminals and ground is greater than 60V. If the fuse burns out during measurement, the Meter may be damaged or the operator himself may be hurt. Use proper terminals, function, and range for the measurement. When the testing leads are connected to the current terminals, do not parallel them across any circuit.
The Model UT50A: The AC current measurement has
5 measurement positions on the rotary switch: 200µA, 2mA, 20mA, 200mA and 20A. The Model UT50B/UT50C: The AC current measurement has 3 measurement positions on the rotary switch: 20mA, 200mA and 20A.
Note
l If the value of current to be measured is unknown, use the maximum measurement position (20A) and 20A terminal and reduce the range step by step until a satisfactory reading is obtained. l When current measurement has been completed, disconnect the connection between the testing leads and the circuit under test.
To measure current, do the following:
1. Turn of f power to the circuit.Discharge all high-voltage capacitors.
2. Insert the red test lead into the µAmA (The Model
UT50A) or mA (The Model UT50B/UT50C) or 20A terminal and the black test lead into the COM terminal.
3. Set the rotary switch to an appropriate measurement position in A range.
4. Break the current path to be tested. Connect the red test lead to the more positive side of the break and the black test lead to the more negative side of the break.
5. Turn on power to the circuit. The measured value shows on the display.
Model UT50A/B/C: OPERATING MANUAL
Page 19
19
E. Measuring Resistance (see figure 7)
Warning T o avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring resistance.
The resistance ranges are: The Model UT50A/UT50B: 200, 2k, 20k, 200k,
2M, 20Mand 200M. The Model UT50C: 200, 2k, 20k, 200k, 2M and 200M.
To measure resistance, connect the Meter as follows:
1. Insert the red test lead into the V terminal and the black test lead into the COM terminal.
2. Set the rotary switch to an appropriate measurement position in range.
3. Connect the test leads across with the object being measured.The measured value shows on the display .
Note
l The test leads can add 0.1to 0.3 of error to the resistance measurement.To obtain precision readings in low-resistance, that is the range of 200, short­ circuit the input terminals beforehand and record the reading obtained (called this reading as X). (X) is the additional resistance from the test lead. Then use the equation:
Measurement Operation(7)
( figure 7)
red black
Model UT50A/B/C: OPERATING MANUAL
Page 20
20
measured resistance value (Y) – (X) = precision readings of resistance. l For high resistance (>1M), it is normal taking several seconds to obtain a stable reading. l When there is no input, for example in open circuit condition, the Meter displays “1”. l When resistance measurement has been completed, disconnect the connection between the testing leads and the circuit under test.
Measurement Operation(8)
F. The Model UT50C: Frequency Measurement (see figure 8)
Warning To avoid harm to you or damages to the Meter, do not attempt to measure voltages higher than 60V in DC or 30V rms in AC although readings may be obtained. When the frequency signal to be tested is higher than 30V rms, the Meter cannot guarantee accuracy of the measurement.
( figure 8)
red black
The frequency measurement ranges are 2kHz and 20kHz. To measure frequency, connect the Meter as follows:
1. Insert the red test lead into the V terminal and the black test lead into the COM terminal.
2. Set the rotary switch to an appropriate measurement position in kHz range.
3. Connect the test leads across with the object being measured. The measured value shows on the display.
Model UT50A/B/C: OPERATING MANUAL
Page 21
21
Note
l When Hz measurement has been completed, disconnect the connection between the testing leads and the circuit under test.
G. The Model UT50B/UT50C: Temperature Measurement (see figure 9)
Warning To avoid harm to you or damages to the Meter, do not attempt to measure voltages higher than 60V in DC or 30V rms in AC although readings may be obtained.
The temperature measurement range is from -40 oC~ 1000 oC or -40 oF~1832 oF. To measure temperature, connect the Meter as follows:
Note
l The Meter automatically displays the temperature value inside the Meter when there is no temperature probe connection.
Measurement Operation(9)
1. Insert the red temperature into the V terminal and the black temperature probe into the COM terminal.
2. Set the rotary switch to oC or oF.
3. Place the temperature probe to the object being measured. The measured value shows on the display.
( figure 9)
Model UT50A/B/C: OPERATING MANUAL
Page 22
22
l The included temperature probe can only be measured up to 250oC For any mesaurement higher than that, the rod type temperature probe must be used instead. l When temperature measeuremnet has been completed, disconnect the connection between the testing leads and te circuit under test.
Measurement Operation(10)
H.Capacitance Measurement (see figure 10)
( figure 10)
The Model UT50A:Capacitance measurement has 4 measurement positions on the rotary switch : 2nF, 20nF, 2µF and 100µF. The Model UT50B/UT50C: Capacitance measurement has 4 measurement positions on the rotary switch: 20nF, 200nF, 2µF and 100µF.
To measure capacitance, connect the Meter as follows:
1. Insert the red test clip or red test lead into the V terminal and the black test clip or black test lead into the µmA (The Model: UT50A) or mA (The Model: UT50B/UT50C) terminal.
2. Set the rotary switch to an appropriate measurement position in Fcx range.
3. Connect the test leads across with the object being measured. The measured value shows on the display.
Model UT50A/B/C: OPERATING MANUAL
Page 23
23
Measurement Operation(11)
Note
l For testing the capacitor with polarity, connect the red test lead to anode & black test lead to cathode l When the tested capacitor is shorted or the capacitor value is overloaded, the LCD display “1”. l To minimize the measurement error caused by the distributed capacitor, the testing lead or testing clip should be as short as possible.
When the tested capacitor is greater than 30 µF, reading is for reference only.
l It is normal to take a while for zeroing when changing over the measurement range. This process will not affect the accuracy of the final readings obtained.
I. Measuring Diodes & Continuity
Warning To avoid damage to the Meter or to the equipment under test, disconnect circuit power and discharge all high-voltage capacitors before measuring diodes and continuity .
Testing Diodes Use the diode test to check diodes, transistors, and other
semiconductor devices. The diode test sends a current through the semiconductor junction, and then measures the voltage drop across the junction. A good silicon junction drops between 0.5V and 0.8V.
To test out a diode out of a circuit, connect the Meter as follows:
1. Insert the red test lead into the V terminal and the black test lead into the COM terminal.
2. Set the rotary switch to
.
3. For forward voltage drop readings on any semiconductor component, place the red test lead on the component’s anode and place the black test lead on the component’s cathode. The measured value shows on the display.
Model UT50A/B/C: OPERATING MANUAL
Page 24
24
Measurement Operation(12)
Note
l In a circuit, a good diode should still produce a forward voltage drop reading of 0.5V to 0.8V; however; the reverse voltage drop reading can vary depending on the resistance of other pathways between the probe tips. l Connect the test leads to the proper terminals as said above to avoid error display. The LCD will display “1” indicating open-circuit for wrong connection. The unit of diode is Volt (V), displaying the positive­ connection voltage-drop value. l When diode testing has been completed, disconnect the connection between the testing leads and the circuit under test.
Testing for Continuity To test for continuity, connect the Meter as below:
1. Insert the red test lead into V terminal and the black test lead into the COM terminal.
2. Set the rotary switch to
.
3. Connect the test leads across with the object being measured. The buzzer sounds if the resistance of a circuit under test is less than 70Ω . The LCD displays the resistance value of a circuit under test.
Note
l The LCD display “1” indicating the circuit being tested is open. l When continuity testing has been completed, disconnect the connection between the testing leads and the circuit under test.
Model UT50A/B/C: OPERATING MANUAL
Page 25
Sleep Mode
25
Turning on the Auto Display Backlight
The Meter has a built-in sensor. Therefore the Display Backlight turns on and off automatically depending on the brightness of the environment. In a dim condition, the Display Backlight turns on automatically while in a bright condition, the Display Backlight turns off automatically .
Model UT50A/B/C: OPERATING MANUAL
Page 26
26
General Specifications
l Maximum voltage between
any Terminals and
Grounding : 1000V rms.
l
Fused Protection for mA Input Terminal : 0.5A, 250V fast type, 5x20mm.
l
Fused Protection for 20A Input Terminal : Un-fused.
l Range : Manual ranging. l Maximum Display : Display :1999. l Mesaurement Speed : Updates 2~3 times/second. l Temperature : Operating: 0oC~40oC
(32oF~104oF); Storage : -10oC~50oC( 14oF~122oF). l Relative Humidity :
75% @ 0oC~30oC;
50% @ 31oC~40oC. l Altitude: Operating : 2000m; Storage : 10000m. l Battery Type : 9V NEDA1604 or 6F22 or 006P.
l Battery Deficiency : Display “
”.
l Data Holding : Display “
l Negative reading : Display “ “ l Overloading : Display “1”. l Dimensions (HxWxL) : 165x80x38.3mm. l Weight : Approx.275g (battery
included). l Safety/Compliances : IEC61010 CAT II 1000V overvoltage and double insulation standard. l Certificate :
Model UT50A/B/C: OPERATING MANUAL
Page 27
27
Accuracy Specifications(1)
Accuracy: (a% reading + b digits),guarantee for 1 year. Operating temperature:23oC
5oC.
Relative humidity:<75%. Temperature coefficient: 0.1 x (specified accuracy) / 1oC.
A. DC Voltage
Remark: Input impedance: 10M B. AC Voltage
200mV
2V
20V 200V 750V
Remarks:
l Input impedance: 10M. l Frequency response: 40Hz~400Hz. l Display effective value of sine wave (mean value
response).
200mV
2V
20V
200V
1000V
100µV
1mV
10mV
100mV
1V
Range Resolution
(0.5%+1)
Overload Protection
UT50A UT50B UT50C
Accuracy
250VDCor ACrms.
(0.8%+2)
1000V DC or 750V AC rms.
100µV
1mV
10mV
100mV
1V
Range
Resolution
(1.2%+3)
Overload Protection
UT50A UT50B UT50C
Accuracy
250V DC or AC rms
(0.8%+3) (1.2%+3)
1000V DC or 750V AC rms.
Model UT50A/B/C: OPERATING MANUAL
Page 28
28
Accuracy Specifications(2)
C. DC Current
Range
Resolution
(0.8%+1)
Overload Protection
UT50A UT50B UT50C
Accuracy
(2%+5)
20µA
2mA
20mA
200mA
20A
0.01µA 1µA
10µA
100µA
10mA
(0.8%+1)
(0.8%+1)
0.5A. 250V fast type
fuse, 5x20 mm
Un-Fused
(1.5%+1)
Remarks:
l At 20A Range: For continuous measurement
10
seconds and interval not less than 15 minutes.
l Measurement voltage drop: Full range at 200mV. D. AC Current
Range
Resolution
(1%+3)
Overload Protection
UT50A UT50B UT50C
Accuracy
(3%+5)
(1.8%+3)
200µA
2mA
20mA
200mA
20A
0.1µA 1µA
10µA
100µA
10mA
(1%+3)
0.5A. 250V fast type
fuse, 5x20 mm
Un-Fused
Remarks:
l At 20A Range: For continuous measurement
10
seconds and interval not less than 15 minutes.
l Measurement voltage drop: Full range at 200mV. l Frequency reaponse: 40Hz~400Hz l Display effective value of sine wave (mean value
response).
Model UT50A/B/C: OPERATING MANUAL
Page 29
29
E. Resistance Test
Accuracy Specifications(3)
200
2k
20k
200k
2M
20M
200M
Range
Resolution
(0.8%+3)
Overload Protection
UT50A UT50B UT50C
Accuracy
(1%+2)
0.1 1
10
100
1k
10k
100k
[5%(reading-10)+10]
250V DC
or AC rms
(0.8%+1)
Remarks:
l Open circuit voltage:
700mV (At 200Mrange, it is approx. 2.8V) l At 200Mrange , test lead is in short circuit , and it is normal to display 10 digits. During measurement, subtract the 10 digits from the reading.
F. The Model UT50C: Frequency
Remarks: Input Sensitivity: 200mV.
G. The Model UT50B/UT50C: Temperature
Range
Resolution
(3%+3)
Overload Protection
UT50B UT50C
Accuracy
250V AC
-40oC~0oC
0oC~400oC
400oC~1000oC
-40oF~32oF
32oF~752oF
752oF~1832oF
(1%+3)
2.5% (3%+4) (1%+4)
(1.5%+15)
1oC
1oF
1Hz
10Hz
Range Resolution Accuracy Overload Protection
2kHz
20kHz
(2%+5)
(1.5%+5)
250V AC
Model UT50A/B/C: OPERATING MANUAL
Page 30
30
Accuracy Specifications(4)
Range
Resolution
(4%+3)
Overload Protection
UT50A UT50B UT50C
Accuracy
(4%+3)
H. Capacitance
2nF
20nF
200nF
2µF
100µF
1pF
10pF
0.1nF 1nF
0.1µF
(5%+4)
(4%+3)
250V AC
Remarks:Testing signal:approx. 400Hz 40mV rms. When the tested capacitor is greater than 30 µF, reading is for reference only.
I. Diodes and Continuity Test
1mV
1
Range Resolution Input Protection Remark
Open circuit voltage approx.2.8V
250V DC or AC
<70buzzer beeps continuously
Model UT50A/B/C: OPERATING MANUAL
Page 31
31
Maintenance(1)
This section provides basic maintenance information including battery and fuse replacement instruction.
Warning Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service information. To avoid electrical shock or damage to the Meter, do not get water inside the case.
A. General Service
l Periodically wipe the case with damp cloth and mild detergent. Do not use chemical solvent. l To clean the terminals with cotton bar with detergent, as dirt or moisture in the terminals can affect readings. l Turn the Meter OFF when it is not in use and take out the battery when not using for a long time. l Do not store the Meter in place of humidity, high temperature, explosive, inflammable and strong magnetic field
B. Replacing the Battery (see figure 11)
(figure 11)
Model UT50A/B/C: OPERATING MANUAL
Page 32
32
Maintenance(2)
Warning
T o avoid false readings, which could lead to possible electric shock or personal injury ,replace the battery
as soon as the battery indicator appears.
To replace battery:
1. Disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.
2. Turn the Meter OFF
3. Remove the rubber feet and screws from the battery compartment, and separate the battery compartment from the case bottom
4. Remove the battery from the battery compartment.
5. Replace the battery with a new 9V battery (NEDA 1604 or 6F22 or 006P).
6. Rejoin the battery compartment and the case bottom, and install the screw and rubber feet
C. Replacing the Fuses
Warning
To avoid electrical shock or arc blast, or personal injury or damage to the Meter, use specified fuses ONLY in accordance with the following procedure.
To replace the Meter’s fuse:
1. Disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.
2. Turn the Meter OFF.
3. Remove the rubber feet and screws from the case bottom, and separate the case bottom from the case top.
4. Remove the fuse by gently prying one end loose, and then take out the fuse from its bracket.
Model UT50A/B/C: OPERATING MANUAL
Page 33
33
Model UT50A/B/C: OPERATING MANUAL
Maintenance(3)
1. Install ONLY replacement fuses with the identical type and specification as follows and make sure the fuse is fixed firmly in the bracket. 0.5A. 250V fast type fuse, φ 5x20mm.
2. Rejoin the case bottom and the case top, and install the screw and rubber feet Replacement of fuses is seldom required. Burning of a fuse always results from the improper operation.
~ END ~
This operating manual is subject to change without notice.
Page 34
34
Manufacturer: UNI-TREND TECHNOLOGY(DONG GUAN)LIMITED Address: Dong Fang Da Dao, Bei Shan Dong Fang Industrial Development District, Hu Men Town, Dong Guan City, Guang Dong Province, China Headquarters: Uni-Trend International Limited Address: Rm901, 9/F, Nanyang Plaza 57 Hung To Road Kwun Tong Kowloon, Hong Kong Tel: (852) 2950 9168 Fax: (852) 2950 9303 Email: info@uni-trend.com http://www.uni-trend.com
Copyright 2001 Uni-Trend International Limited.
All rights reserved.
Model UT50A/B/C: OPERATING MANUAL
Loading...