Troxler 3430 User Manual

Page 1
Manual of Operation and Instruction
Model 3430
(and Model 3430-M)
Surface
Moisture-Density
Troxler Electronic Laboratories, Inc.
3008 Cornwallis Rd. P.O. Box 12057
Research Triangle Park, NC 27709
Phone: 1.877.TROXLER
Outside the USA: +1.919.549.8661
Fax: +1.919.549.0761
www.troxlerlabs.com
Page 2
The Troxler Model 3430 Surface Moisture-Density Gauge
is protected by U.S. and Foreign Patents
Copyright © 1990–2006
Troxler Electronic Laboratories, Inc.
All Rights Reserved
No part of this manual may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopying, recording, or information storage and retrieval systems, for any purpose without the express written permission of Troxler Electronic Laboratories, Inc.
Fantastic is a trademark of Dow Consumer Products, Inc. Federal Express is a registered trademark of the Federal Express
Corporation.
409 is a trademark of the Clorox Company. Magnalube-G is a registered trademark of Carleton-Stuart
Corporation.
WD-40 is a registered trademark of the WD-40 Company.
PN 105305 December 2006 Edition 8.1
ii
Page 3

TROXLER SERVICE CENTERS

Troxler Corporate Headquarters
3008 Cornwallis Road
P.O. Box 12057
Research Triangle Park, NC 27709
Phone: 1.877.TROXLER (1.877.876.9537)
Outside the U.S.A.: +1.919.549.8661
Fax: +1.919.549.0761
Web: www.troxlerlabs.com
Technical Support
Phone: 1.877.TROXLER (1.877.876.9537)
E-mail: TroxTechSupport@troxlerlabs.com
Midwestern Branch Office
1430 Brook Drive Downers Grove, IL 60515 Fax: 630.261.9341
Western Regional Branch Office
11300 Sanders Drive, Suite 7 Rancho Cordova, CA 95742 Fax: 916.631.0541
Southwestern Branch Office
2016 East Randol Mill Road Suite 406 Arlington, TX 76011 Fax: 817.275.8562
To locate an independent, Troxler-authorized service center near you, call 1.877.TROXLER (1.877.876.9537).
Florida Service Center
2376 Forsyth Road Orlando, FL 32807 Fax: 407.681.3188
Troxler European Subsidiary
Troxler Electronics GmbH Gilchinger Strasse 33 D.82239 Alling nr. Munich, Germany Phone: ++49.8141.71063 Fax: ++49.8141.80731 E-mail: troxler@t-online.de
NOTE
Model 3430 iii
Page 4
CAUTIONS AND WARNINGS
Units intended for use in countries that are members of the European Community are shipped with a CE-approved AC adapter, Troxler bin number 108354.
Gauge cover is to be removed by trained service personnel only. There are no user-serviceable components inside. Note that components behind the cover can have voltage potentials in excess of 50 volts during normal operation of the gauge.
Appendix A, Radiation Theory and Safety should be read carefully and understood before using the gauge.
See page C-8, Alkaline Battery Use.
The source rod should automatically retract to the SAFE position when the gauge is lifted by the handle.
See page C-12, Tungsten Sliding Block.
iv
Page 5

EU DECLARATION OF CONFORMITY

Application of Council EMC Directive 89/336/EEC and Low Voltage Directive 73/23/EEC
Standards to which Conformity is Declared:
EN 61010-1 EN 55011 Group 1, Class A EN 50082-2
An EMC Technical Report/Certificate has been issued in accordance with Part IV (Reg 50) of the UK Regulations (SI 1992 No. 2372) by a UK appointed Competent Body, namely,
Interference Technology International Limited
41-42 Shrivenham, Hundred Business Park
Shrivenham, Swindon, Wiltshire SN6 8TZ
Certificate Number C283TRO.1ABS Dated 16th January 1997 Troxler Document Number 108205
Manufacturer: Troxler Electronic Laboratories, Inc. PO Box 12057 3008 Cornwallis Road Research Triangle Park, North Carolina 27709 USA
Apparatus: Model 3430 Surface Moisture-Density Gauge
Year of Declaration: 1997 (Original)
2006 (Corrected)
Model 3430 v
Page 6
HOW TO USE THIS MANUAL
Congratulations on the purchase of the Troxler Model 3430 Surface Moisture-Density Gauge.
The Model 3430 Manual of Operation and Instruction contains information on how the Model 3430 operates, and provides directions on the use of this gauge. Site selection, basic parameter setup, moisture and density determination, storage, and advanced operations are included, along with radiological information and system troubleshooting.
GUIDE TO SYMBOLS AND FORMATTING
Throughout this manual, symbols and special formatting are used to reveal the purpose of the text as follows:
CAUTION
NOTE Indicates important information that must be read to
Diamonds indicate a list of things needed (such as
Check marks indicate the performance of an action.
Triangles indicate that more than one option is
KEY Angle brackets and a different typestyle indicate a
DISPLAY A different typestyle is also used to indicate text
Indicates conditions or procedures that, if not followed correctly, may cause personal injury or equipment damage.
ensure proper operation.
equipment) or things to know.
With lists of check marks, follow the instructions in the order of the check marks.
available. Carefully select the option that applies.
key or character (number or letter) to press on the control unit keypad. For example, “Press STD” means to press the key labeled STD.
displayed on the control unit.
vi
Page 7

TABLE OF CONTENTS

CHAPTER 1. INTRODUCTION TO THE MODEL 3430
Introduction....................................................................................1-2
Gauge Parts and Accessories.........................................................1-4
Unpacking and Inspection .............................................................1-6
Storage Site Selection....................................................................1-7
CHAPTER 2. THEORY OF OPERATION
Density...........................................................................................2-2
Moisture.........................................................................................2-5
CHAPTER 3. OPERATING THE GAUGE
The Keypad....................................................................................3-2
Source Rod Positions.....................................................................3-4
Daily Inspection.............................................................................3-5
Turning the Gauge On ................................................................... 3-6
Gauge Parameter Setup..................................................................3-7
Taking the Standard Count..........................................................3-10
Site Preparation/Gauge Positioning.............................................3-12
Taking a Measurement – Soil Mode............................................ 3-15
Taking a Measurement – Asphalt Mode......................................3-17
CHAPTER 4. ADVANCED GAUGE OPERATION
Offsets............................................................................................4-2
Special Calibrations.......................................................................4-7
Thin Layer Measurements...........................................................4-11
CHAPTER 5. SPECIAL FUNCTIONS
Recall.............................................................................................5-2
Offset .............................................................................................5-2
Stat Test.........................................................................................5-2
Drift Test........................................................................................5-5
Special Calibration.........................................................................5-7
Specific Gravity.............................................................................5-7
Voidless Density............................................................................5-8
Set Units.........................................................................................5-8
Calibration Constants.....................................................................5-9
Memory Reset..............................................................................5-10
Test Reading................................................................................5-10
Language......................................................................................5-11
Model 3430 vii
Page 8
TABLE OF CONTENTS (Continued)
APPENDIX A. RADIOLOGICAL INFORMATION
Radiation Theory...........................................................................A-2
Radiation Safety............................................................................A-5
APPENDIX B. SPECIFICATIONS
Measurement.................................................................................B-2
Radiological ..................................................................................B-4
Electrical........................................................................................B-5
Mechanical ....................................................................................B-6
APPENDIX C. TROUBLESHOOTING AND SERVICE
Troubleshooting.............................................................................C-2
Battery Charging ...........................................................................C-8
Mechanical Maintenance.............................................................C-10
Leak Testing................................................................................C-15
Replacement Parts.......................................................................C-16
Returning the Gauge for Service.................................................C-26
Troxler Service Centers............................................................... C-28
APPENDIX D. TRANSPORTATION AND SHIPPING
U.S. Requirements.........................................................................D-2
Canadian Shipping Requirements .................................................D-4
APPENDIX E. STANDARD COUNT LOG
APPENDIX F. UNIT CONVERSION
Measurement Units........................................................................F-2
Radiological Units......................................................................... F-2
INDEX
WARRANTY
viii
Page 9

LIST OF FIGURES

Figure
1-1 Model 3430 Gauge and Accessories..........................1-5
2-1 Direct Transmission Geometry ..................................2-3
2-2 Backscatter Geometry................................................2-3
2-3 Backscatter Surface Density Effects (Top Layer
Effect Curves) ............................................................2-4
2-4 Effect of Moisture on Depth of Measurement............2-6
3-1 Model 3430 Keypad...................................................3-2
3-2 Source Rod Positions .................................................3-4
3-3 Standard Count Position...........................................3-11
3-4 Drill Rod Positioning ...............................................3-12
3-5 Marking the Test Area .............................................3-13
3-6 Voids Illustration......................................................3-16
A-1 Diagram of an Atom..................................................A-2
A-2 Variation of Radioactive Emission ...........................A-4
A-3 Effect of Distance on Exposure................................. A-6
A-4 Model 3430 Gauge and Transport Case.................... A-8
C-1 Removing the Tungsten Sliding Block ................... C-13
C-2 3430 Gauge Assembly ............................................ C-17
C-3 3430 Gauge Base M echanical Assembly................ C-19
C-4 3430 Source Rod Handle Assembly........................ C-19
C-5 3430 Gauge Preamplifier Assembly ....................... C-21
C-6 3430 Gauge Scaler Assembly ................................. C-23
Title Page
Model 3430 ix
Page 10

LIST OF TABLES

Title Page
Table
3-1 Model 3430 Keypad Functions.................................. 3-3
4-1 K Values for Thin Lift Overlays.............................. 4-13
A-1 Radiation Profile for Model 3430 Gauge...................A-9
A-2 Radiation Profile for Model 3430-M Gauge............A-10
x
Page 11
ATTENTION GAUGE OWNER
This gauge contains functions that require an ACCESS CODE. This code must be entered before these functions may be used. For more information on using the access code refer to the specific function in Chapter 5.
The ACCESS CODE for this gauge is:
4678
This page should be removed if the access code is not to be distributed to other parties or users of this gauge.
Model 3430 xi
Page 12
NOTES
xii
Page 13
CHAPTER 1
INTRODUCTION TO THE MODEL 3430
This chapter provides a general introduction to the Model 3430 Surface Moisture-Density Gauge and its applications. Also included are a list of the gauge parts and accessories, instructions for unpacking and inspecting the system, and guidelines for site selections.
CONTENTS
Introduction....................................................................................1-2
Gauge Parts and Accessories.........................................................1-4
Unpacking and Inspection .............................................................1-6
Storage Site Selection....................................................................1-7
INTRODUCTION
Model 3430 1-1
Page 14

INTRODUCTION

The Model 3430 Surface Moisture-Density Gauge can quickly and precisely determine the moisture and density of soils, soil bases, aggregate, concrete and asphaltic concrete without the use of core samples or other destructive methods.
Using direct transmission or backscattered gamma radiation, the 3430 gauge determines the density of materials by counting the number of photons emitted by a cesium-137 source. Geiger-Mueller (G-M) detectors located in the gauge base detect the gamma radiation and a microprocessor converts the counts into a density reading.
Using the principle of neutron thermalization, the Model 3430 determines the moisture content of soils and soil-like materials. Hydrogen (water) in the material slows neutrons emitted from an americium-241:beryllium source (or californium-252 in the Model 3430-M). Helium-3 detectors located in the gauge base detect the slowed neutrons.
The nuclear method of testing density and moisture has been approved by the American Society of Testing and Materials (ASTM). The Model 3430 meets or exceeds all the requirements of ASTM Standards C1040, D2922, D2950, and D3017.
Some information contained in this manual is used in training courses offered by Troxler Electronic Laboratories, Inc. and to assist purchasers in obtaining a Radioactive Materials License from the U.S. Nuclear Regulatory Commission or an Agreement State. Owners of this gauge must maintain a current radioactive materials license as long as they own the gauge, even if it is in storage and not actively being used.
1-2
Page 15
Any licensing issues discussed in this manual are for the United States. To purchase a Model 3430 in Canada, owners must obtain a radioisotope license from the Canadian Nuclear Safety Commission (CNSC). The owner should obtain copies of the CNSC Regulations and the Transportation of Dangerous Goods Act and Regulations. This manual provides a guide to Canadian shipping requirements in Appendix D.
Owners are encouraged to require study of this manual by users before allowing any use of the instrument. To monitor exposure to radiation, personnel should wear a dosimeter while operating or cleaning the gauge. The sections of the manual covering radiation safety should be required reading for all operators and potential operators. If these sections are not completely understood, users
should seek assistance from Troxler, an appointed Troxler representative or others designated within the user organization. Additional radiation safety information is available
by attending a Troxler Nuclear Gauge Training Course.
As changes are made to local, state, and federal regulations on a continuing basis, the owner/user must maintain a knowledge of these regulations. The responsibility for compliance ultimately falls upon the owner. The owner may also wish to purchase and subscribe to Titles 10 and 49 of the Code of Federal Regulations in addition to applicable local/state regulations.
NOTE
This manual also contains radiological information for the Model 3430-M. The 3430 and 3430-M have the same functions and operational aspects but different neutron sources.
INTRODUCTION
Model 3430 1-3
Page 16

GAUGE PARTS AND ACCESSORIES

Figure 1-1 shows the Model 3430 gauge and its accessories. Use this figure and the list below to identify the gauge and parts as they are unpacked.
1. The Gauge is a portable instrument containing all electronic
modules, battery packs, detectors, and radioactive sources.
2. The Reference Standard Block provides a uniform reference
material for gauge adjustment to compensate for source decay.
3. The Scraper Plate/Drill Rod Guide is used to prepare the test
site and aid in guiding the drill rod into the soil.
4. The Drill Rod is used to prepare a hole for a direct transmission reading. Do Not Use the Source Rod For This Purpose!
5. Two Chargers/Adapters are supplied: one for DC (12 VDC) and one for AC (115/230 VAC 50/60 Hz.)
6. The Transport Case is a container designed for transportation of the gauge and associated parts.
7. The Extraction Tool provides a means of removing the drill rod from the test material after use.
1-4
Page 17
INTRODUCTION
Figure 1-1. Model 3430 Gauge and Accessories
Model 3430 1-5
Page 18

UNPACKING AND INSPECTION

Upon receipt of the gauge from the factory, a complete inspection and inventory should be performed. If the shipping case, any other part of the container, or the gauge appears to be damaged, notify the carrier and your Troxler Representative immediately.
For shipping to another location or back to the factory, save the box and any packing material. For shipping instructions and regulations, please see Appendix D.
Check to see if the following literature and components have been included:
Manual of Operation and Instruction Gauge Warranty Source Certificate 3430 Gauge Scraper Plate Drill Rod AC Battery Charger DC Charger/Adapter One handle lock with keys
Lift the gauge from the case. Inspect the gauge for damage. Check the lock on the handle. Ensure the keys fit the lock.
1-6
Page 19

STORAGE SITE SELECTION

When deciding where to store the gauge, take into consideration the rules governing the storage of low-level radioactive devices that are set forth by your regulatory agency and the conditions of your gauge license.
Ë The handle should be locked and the gauge stored in its
transport case.
Ë It is recommended that the gauge and transport case be stored at
least 15 ft (5 m) from work areas, preferably in a locked closet/storage area in a dry location (indoors).
Ë The storage area should be marked with a radiation sign that
reads “CAUTION RADIOACTIVE MATERIALS (can be obtained from Troxler).
Ë The storage of a nuclear gauge in a motor vehicle is not
recommended.
INTRODUCTION
Model 3430 1-7
Page 20
NOTES
1-8
Page 21
CHAPTER 2
THEORY OF OPERATION
This chapter contains a brief description of the theory of operation of the Model 3430 Surface Moisture-Density Gauge. The direct transmission and backscatter modes of operation are illustrated along with a brief explanation of the cesium-137 source, americium­241:beryllium or californium-252 source, and detector geometry.
CONTENTS
Density...........................................................................................2-2
Moisture.........................................................................................2-5
OPERATION THEORY
Model 3430 2-1
Page 22

DENSITY

The Troxler Model 3430 gauge utilizes two modes of operation: direct transmission mode (source rod extended into the material) and backscatter mode. Figures 2-1 and 2-2 illustrate the two modes of operation.
In direct transmission mode (Figure 2-1), the rod containing the cesium-137 (8 mCi/0.3 GBq) source is lowered to the desired depth. The detectors (G-M tubes) in the gauge base measure the radiation emitted by the source rod. Gamma photons reaching the detectors must first pass through the material, colliding with electrons present in the material. In general, the lower the number of photons that reach the detectors, the higher the material density.
In backscatter mode (Figure 2-2), the gamma photons that enter the material must be scattered (or reflected) at least once to reach the detectors in the gauge. With the rod locked in the first notch below the SAFE position, the source and detectors are in the same plane, referred to as the backscatter position. Photons emitted from the source penetrate the material, and the detectors measure the scattered photons. Shielding between the source and detectors greatly reduces the number of photons reaching the detectors in a direct path with the source.
While the direct transmission geometry measures the average density of the material from the source to the surface, the backscatter geometry yields an average heavily weighted by the density close to the surface. Figure 2-3 shows two normalized top layer effect curves, illustrating the percentages of photons at the detectors for various depths. The two curves can be used to compute the gauge response to layered material of different densities. For example, the density of the top inch of a surface layer accounts for approximately 52% of the backscatter density measurement.
2-2
Page 23
OPERATION THEORY
Figure 2-1. Direct Transmission Geometry
Figure 2-2. Backscatter Geometry
Model 3430 2-3
Page 24
Figure 2-3. Backscatter Surface Density Effects
(Top Layer Effect Curves)
2-4 Model 3430 2-5
Page 25

MOISTURE

The Model 3430 gauge uses a 40 mCi (1.48 GBq) americium­241:beryllium neutron source to measure the hydrogen content (consequently the water content) of the material. The 3430-M gauge employs a 60 µCi (2.22 MBq) californium-252 source.
Neutrons emitted by the Am-241:Be (or Cf-252) source penetrate the material and are thermalized (or slowed). Thermalization is the process where neutrons are slowed to the point where further collisions with hydrogen or other materials will not continue to slow the neutron.
The 3430 gauge contains a helium-3 neutron detector that is sensitive to thermalized neutrons. This detector is insensitive to non-thermalized, or “fast” neutrons and, as a result, the counts obtained are directly proportional to the amount of hydrogen/moisture present in the material.
The depth of measurement, or depth at which 98% of the counted neutrons pass before reaching the detector, is a function of moisture content:
Depth (inches) = 11 – (0.17
Depth (mm) = 280 - (0.27
Therefore, the higher the moisture content in the material being measured, the smaller the depth of measurement. The normalized curve set shown in Figure 2-4 illustrates the effect of moisture content on the depth of measurement.
×
M), where: M = moisture in pcf
or
×
M), where: M = moisture in kg/m3
OPERATION THEORY
Page 26
Figure 2-4. Effect of Moisture on Depth of Measurement
2-6 Model 3430 3-1
Page 27
CHAPTER 3
OPERATING THE GAUGE
This chapter explains the basic operation of the Model 3430 Surface Moisture-Density Gauge. Instructions for conducting a daily gauge inspection, setting gauge parameters, taking the daily standard count, preparing the site, positioning the gauge, and taking moisture and density measurements are included.
CONTENTS
The Keypad....................................................................................3-2
Source Rod Positions.....................................................................3-4
Daily Inspection.............................................................................3-5
Turning the Gauge On ................................................................... 3-6
Gauge Parameter Setup..................................................................3-7
Setting Measurement Units.....................................................3-7
Setting the Count Time ...........................................................3-8
Setting the Depth.....................................................................3-8
Selecting the Mode (Marshall/Proctor)...................................3-8
Taking the Standard Count..........................................................3-10
Site Preparation/Gauge Positioning.............................................3-12
Soil and Base Course Preparation.........................................3-12
Asphalt Surface Preparation..................................................3-14
Taking a Measurement – Soil Mode............................................ 3-15
Taking a Measurement – Asphalt Mode......................................3-17
GAUGE OPERATION
Page 28

THE KEYPAD

Figure 3-1. Model 3430 Keypad
The keypad (Figure 3-1) of the Model 3430 Surface Moisture­Density Gauge consists of ten keys — an eight-function keypad and the ON/YES and OFF/NO keys. The gauge is equipped with a beeper to verify keystrokes. If a beep is not heard when a key is pressed, the keystroke was not recognized and should be repeated. The ON/YES and OFF/NO keys are used for responses to specific questions displayed on the screen and to turn the gauge on and off.
The up and down arrows allow the operator to scroll through various function lists displayed by the gauge.
Table 3-1 provides a more detailed description of the individual keys and the location in the manual where the functions are described.
3-2 Model 3430 3-3
Page 29
Table 3-1. Model 3430 Keypad Functions
KEYS
ON
YES
OFF
NO
MA PR
TIME
STD
DEPTH
SPECIAL
DESCRIPTION PAGE
Turns on the gauge and answers Yes to prompts.
Turns gauge off and answers No to prompts
Scrolls the display up. Scrolls the display down. Allows entering or enabling of a
Proctor or Marshall value. Allows the operator to change the
count time. Use to access the Standard Count
mode. Allows entry of the source rod depth. 3-8
Provides access to the Special functions.
3-6
3-9
3-8
3-10
3-7, 4-3, 4-4, 4-7,
5-1
GAUGE OPERATION
START ENTER
Starts a measurement or completes answer entry.
Page 30

SOURCE ROD POSITIONS

Figure 3-2 shows the available positions of the source rod.
NOTE
The source rod should always be in the SAFE position when the gauge is not in use.
3-4
Figure 3-2. Source Rod Positions
Page 31

DAILY INSPECTION DAILY INSPECTION

The gauge should be inspected daily before use to ensure proper operation of all safety features as follows:
Push the source rod down into the backscatter position, and then
raise it back to the SAFE (shielded) position. The source rod opening in the bottom of the gauge is equipped with a spring­loaded tungsten sliding block that shuts when the source rod is in the SAFE position. Turn the gauge over and verify that the sliding block is completely shut. If any portion of the opening is uncovered, the sliding block should be cleaned before using, transporting, or storing the gauge. Refer to the Mechanical Maintenance section of Appendix C for instructions on cleaning the tungsten sliding block.
CAUTION
Do not store or transport the gauge unless the sliding block is completely closed. Increased radiation levels may violate transportation regulations and cause excessive personnel radiation exposure.
GAUGE OPERATIONGE OPERATION
If a radiation survey instrument is available, verify that the
radioactive gamma source is in place by measuring the exposure rate at the surface of the gauge. The exposure rate should be approximately 10 – 20 mrem per hour. A reading of about 1 mrem or less indicates either that the survey instrument is not working properly or that the cesium-137 source may be missing. Refer to the Troubleshooting section of Appendix C for further instructions.
Model 3430 3-5
Page 32

TURNING THE GAUGE ON

The gauge uses rechargeable NiCad batteries (included) as a power source. When first turned on, the control panel displays test characters before proceeding to the self-test.
NOTE
If the gauge turns off immediately after it is turned on, the battery charge may be low or the gauge may be wet inside. Refer to the information in Appendix C on battery charging and gasket replacement.
To turn the gauge on, press ON/YES. The gauge performs a test of its liquid crystal display (LCD):
Testing LCD
0123456789ABCDEFG
After the 300-second self-test, the gauge will enter the Ready mode. In this state any of the gauge functions may be accessed.
The Ready mode display is:
READY xx min
Depth: xx inches
The first line of the display indicates the current count time. The second line of the display indicates the source rod depth that has been selected.
NOTE
The gauge will automatically turn off after five hours if no keys are pressed.
3-6
Page 33

GAUGE PARAMETER SETUP

After unpacking the gauge and turning it on, there are several parameters that can be initialized. These parameters do not usually require changing and include the measurement units and count time.

SETTING MEASUREMENT UNITS

The 3430 gauge allows measurement results to be displayed in either metric or US units. To set the measurement units, first access the Special function menu by pressing SPECIAL.
– RECALL –
( or ENTER)
Press the down arrow seven times to display:
– SET UNITS – ( or ENTER)
To select Set Units, press START/ENTER.
Units: PCF (↑ ↓ or ENTER)
Use the up and down arrows to scroll through the available units. When the desired units are displayed, press START/ENTER.
Model 3430 3-7
GAUGE OPERATION
Page 34

SETTING THE COUNT TIME

The count time defines how long the gauge reads. Longer count times produce better measurement precision. Troxler recommends a count time of one minute for most sample measurements.
To change the count time, press TIMEto display:
Time: xx min. (↑ ↓ or ENTER)
Use the up and down arrows to scroll through the available count times. When the desired count time is displayed, press
START/ENTER.

SETTING THE DEPTH

To change the depth of measurement, press 〈DEPTH.
Depth: xx in. (↑ ↓ or ENTER)
Use the up and down arrows to scroll through the available measurement depths. When the desired depth is displayed, press
START/ENTER.

SELECTING THE MODE (MARSHALL/PROCTOR)

The 3430 gauge may be used on construction materials (soils, asphalt, concrete, and so on). To select the Soil mode, enter or activate a Proctor value. To select the Asphalt mode, enter or activate a Marshall value. Only one Marshall and one Proctor can be stored in the gauge at one time.
3-8
Page 35
NOTE
To measure concrete, use either the Asphalt or Soil mode. For moisture results select the Soil mode. For density measurement only, use the Asphalt mode.
To enter or activate a Marshall or Proctor value, press MA/PR. The display will be one of the following:
MA:    (↑ ↓)
Change MA value?
PR:   (↑ ↓)
Change PR value?
To switch from a Marshall value to a Proctor value, or vice versa, use the arrow keys.
To activate the displayed value, press OFF/NO.
To change the displayed value, press ON/YES. The first digit of the value will flash. Use the arrow keys to scroll through the possible entries (09 and .). When the correct value for the current digit is displayed, press START/ENTER. The gauge will proceed to the next digit to the right.
When the value entry is complete, the gauge activates the value and returns to the Ready mode.
GAUGE OPERATION
Model 3430 3-9
Page 36

TAKING THE STANDARD COUNT

The 3430 gauge uses a cesium-137 and an americium-241:beryllium source for taking measurements. These radioactive sources undergo a natural decay process, resulting in a gradual loss in the intensity of their radiation. The time required for the source strength to diminish by 50% is referred to as the half-life.
To compensate for the source decay and to check proper operation of the gauge, a daily reference standard count should be performed. To ensure the highest accuracy possible with the gauge, it is important to take a daily standard count.
The gauge is equipped with a reference standard block for taking the standard count. Place the reference standard block on a dry, flat surface at least three meters (10 ft) from any large vertical surface and at least ten meters (33 ft) from any other radioactive source. The surface should be asphalt, concrete or soil at least ten centimeters (4 in) thick and with a density of at least 100 pcf. The right side of the gauge, farthest from the handle, should be against the metal butt plate (Figure 3-3).
To begin the standard count procedure, press STD.
DS=xxxx MS=xxxx New Std Cnt?
To take a new standard count, press ON/YES.
Press START for Standard Count
Ensure that the gauge is positioned as shown in Figure 3-3. To initiate the standard count, press START/ENTER. After the count is complete, the display will be:
Standard Count: DS=xxxx MS=xxxx
3-10
Page 37
Figure 3-3. Standard Count Position
Troxler recommends that the operator keep a daily log of the moisture and density standard counts (see Appendix E). To verify gauge stability, compare the daily standard count to a reliable reference as follows:
During the first four days of operation of a new or recalibrated
gauge, compare the daily standard count to the factory calibrated values.
After the first four days of operation (or after taking four
standard counts), compare the daily standard count to the average of the last four counts. Acceptable standard count limits are:
±1% each day for DS (density standard) and ±2% each day for MS (moisture standard).
After recording the standard counts, return to the Ready mode by pressing ON/YES.
Model 3430 3-11
GAUGE OPERATION
Page 38

SITE PREPARATION/GAUGE POSITIONING

Preparation of the test site surface is critical to gauge performance. This section provides site preparation procedures for both soils and base courses and asphalt surfaces. To ensure the most accurate gauge readings, the appropriate preparation procedure should be followed.

SOIL AND BASE COURSE PREPARATION

Since soil surface conditions are critical to accurate
measurements, locate a level site free from any large holes, cracks, or debris.
Smooth the surface by moving the scraper plate in a back and
forth motion. Filler such as fine sand may be used to decrease the surface voids.
NOTE
Use only enough filler to fill the voids. Too much filler will cause an error in the measurement.
For direct transmission measurements, put the drill rod through
the extraction tool and then through one of the guides on the plate (see Figure 3-4).
3-12
Figure 3-4. Drill Rod Positioning
Page 39
Wearing a radiation badge and safety glasses (or other
Wearing a radiation badge and safety glasses (or other
locally approved safety devices), step on the plate and hammer
locally approved safety devices), step on the plate and hammer
the drill rod at least 50 millimeters (2 in) deeper than the desired
the drill rod at least 50 millimeters (2 in) deeper than the desired test depth. The drill rod increments include the additional depth.
test depth. The drill rod increments include the additional depth.
Remove the drill rod by pulling straight up and twisting the
Remove the drill rod by pulling straight up and twisting the
extraction tool. Do not loosen the drill rod by tapping from
extraction tool. Do not loosen the drill rod by tapping from
side to side with a hammer. This will distort the hole or cause
side to side with a hammer. This will distort the hole or cause
loose material to fall into the hole.
loose material to fall into the hole.
To ensure accurate placement of the gauge, before removing the
To ensure accurate placement of the gauge, before removing the
scraper plate mark the test area using the drill rod as shown in
scraper plate mark the test area using the drill rod as shown in Figure 3-5.
Figure 3-5.
Carefully pick up the scraper plate and place the gauge on the
Carefully pick up the scraper plate and place the gauge on the
surface prepared by the plate. Insert the source rod into the hole
surface prepared by the plate. Insert the source rod into the hole made by the drill rod. Use care when inserting the source rod,
made by the drill rod. Use care when inserting the source rod,
trying not to disturb the soil around the hole.
trying not to disturb the soil around the hole.
Lower the source rod into the hole. Release the trigger and lock
Lower the source rod into the hole. Release the trigger and lock
the source rod into the correct position. A click should be heard
the source rod into the correct position. A click should be heard when the source rod is locked into position.
when the source rod is locked into position.
Gently slide the gauge toward the keypad so the source rod
Gently slide the gauge toward the keypad so the source rod
makes contact with the wall of the hole.
makes contact with the wall of the hole.
MARK FOR SCRAPER PLATE CENTER
EDGE MARKS
GAUGE OPERATIONGE OPERATION
SCRAPER
PLATE
METHOD 1
MARK FOR DRILL ROD CENTER
SCRAPER
PLATE
METHOD 2
Figure 3-5. Marking the Test Area Figure 3-5. Marking the Test Area
Model 3430 3-13
Page 40

ASPHALT SURFACE PREPARATION

It is possible, but usually not necessary, to take direct transmission readings on asphalt. Drilling a hole in asphalt can be difficult, and may require the use of a drill (rather than the drill rod) if the asphalt has cooled and hardened.
Under normal conditions, a backscatter reading provides an accurate measurement of asphalt density.
Find a smooth, level location on the asphalt. The operator may
want to fill the voids on open mixes with sand or cement. Take care to leave the asphalt exposed. The gauge base must rest on the asphalt, not the fill material!
Ensure that the gauge does not “rock.” It must remain steady. If
rocking occurs, find a more suitable test site. If taking a measurement around a core, the gauge may be moved a few inches away from the hole to level the gauge.
3-14
Page 41
TAKING A MEASUREMENT – SOIL MODE TAKING A MEASUREMENT – SOIL MODE
The Soil mode is automatically selected when a Proctor value is enabled (see page 3-8).
NOTE
When not taking measurements, always keep the source rod in the SAFE position. For added operator safety, the source rod on the 3430 gauge automatically retracts to the SAFE position when the gauge is lifted by the handle.
If you do not hear a click when the source rod is raised to the SAFE position, look at the bottom of the gauge to verify that the tungsten sliding block is completely closed. If the gauge base opening is not completely closed by the sliding block, the sliding block may require cleaning. Refer to Appendix C for cleaning instructions.
CAUTION
Do not store or transport the gauge unless the tungsten sliding block is completely closed. Increased radiation levels may violate transportation regulations and cause excessive personnel exposure.
GAUGE OPERATIONGE OPERATION
Place the gauge over the test site. Release the gauge handle and push it down until it is in the correct position. Ensure that the pin engages the notch in the index rod.
Press START/ENTER.
Depth: xx in. Time: xxx sec.
After the count time has elapsed, the gauge displays the measurement results in a series of six screens, as follows. Use the up and down arrows to scroll through the various screens.
Model 3430 3-15
Page 42
WD: xxxxx PCF (Use & keys)
DD: xxxxx PCF %PR: xx.x %
Moist: xxxxx PCF % Moist: xx.x %
Air Void: xx.x% Void Ratio
MOIST CR: xx.x DENS CR: xx.x
3-16
M Count: xxxxx D Count: xxxxx
where:
WD = Wet density in kg/m DD = Dry density in kg/m %PR = Percent Proctor (This value is valid only if an
appropriate target has been entered for the material being tested.)
MOIST = Moisture value in kg/m % MOIST = Percent moisture Air Void = See description below Void Ratio = See description below MOIST CR = Moisture count ratio DENS. CR = Density count ratio M Count = Moisture counts as read by the gauge D Count = Density counts as read by the gauge
3
or pcf
3
or pcf
3
or pcf
Page 43
Figure 3-6 illustrates the terms void ratio and % air voids. The void ratio is the ratio of the volume occupied by air and water in the soil to the volume occupied by solid particles. The term % air voids refers to the volume of air voids only as a percentage of the total volume.
The following formulas are used to calculate the % air voids and void ratio values.
% AIR VOIDS = 100 (1 – (Vs/Vt) – (Vw/Vt))
where:
Vs = Volume of Soil Vt = Total Volume Vw = Volume of Water
or,
% AIR VOIDS = 100 (1 – (DD / SG(Dw)) – (M / (Dw)))
where:
Dw = Density of Water SG = Specific Gravity of Soil Particles DD = Dry Density M = Moisture
VOID RATIO = Volume of Voids / Volume of Soil
= ( SG(Dw) – DD ) / DD
GAUGE OPERATION
Figure 3-6. Voids Illustration
Model 3430 3-17
Page 44
TAKING A MEASUREMENT – ASPHALT MODE
The Asphalt mode is automatically selected when a Marshall value is enabled (see page 3-8).
NOTE
When not taking measurements, always keep the source rod in the SAFE position. For added operator safety, the source rod on the 3430 gauge automatically retracts to the SAFE position when the gauge is picked up by the handle.
If you do not hear a click when the source rod is raised to the SAFE position, look at the bottom of the gauge to verify that the tungsten sliding block is completely closed. If the gauge base opening is not completely closed by the sliding block, the sliding block may require cleaning. Refer to Appendix C for cleaning instructions.
CAUTION
Do not store or transport the gauge unless the tungsten sliding block is completely closed. Increased radiation levels may violate transportation regulations, and may cause personnel exposure.
Place the gauge over the test site. Release the gauge handle and push it into the backscatter position. Set the depth to Backscatter. Ensure that the pin engages the notch in the index rod. Gently tap the handle down to ensure proper source rod seating.
Press START/ENTER.
Depth: BACKSCAT. Time: xxx sec.
After the count time has elapsed, the gauge displays the measurement results in a series of six screens, as follows. Use the up and down arrows to scroll through the various screens.
3-18
Page 45
WD: xxxxx PCF % MA xx.x %
DD: xxxxx PCF (Use & keys)
Moist: xxxxx PCF % Moist: xx.x %
% VOIDS xx.x % 100 – % MA xx.x %
MOIST CR: xx.x DENS CR: xx.x
M Count: xxxxx
GAUGE OPERATION
D Count: xxxxx
where:
WD = Wet density in kg/m
3
or pcf
% MA = Percent Marshall (This value is valid only if an
appropriate target has been entered for the material being tested.)
DD = Dry density in kg/m MOIST = Moisture value in kg/m
3
or pcf
3
or pcf
% MOIST = Percent moisture % VOIDS = 100 (1 – WD/VOIDLESS) 100 – % MA = Value given by subtracting the percent
Marshall value from 100
MOIST CR = Moisture count ratio DENS. CR = Density count ratio M Count = Moisture counts as read by the gauge D Count = Density counts as read by the gauge
Model 3430 3-19
Page 46
NOTES
3-20
Page 47
CHAPTER 4
ADVANCED GAUGE OPERATION
This chapter provides instructions for using the Model 3430 Surface Moisture-Density Gauge in special circumstances where the gauge may require an offset or special calibration. This chapter also explains the procedure for using the gauge to measure thin layers of asphalt.
CONTENTS
Offsets............................................................................................4-2
Density Offset .........................................................................4-3
Moisture Offset .......................................................................4-4
Trench Offsets.........................................................................4-6
Special Calibrations.......................................................................4-7
Entering a New B Value..........................................................4-8
Gauge-calculated Calibration..................................................4-9
Thin Layer Measurements...........................................................4-11
Model 3430 4-1
ADVANCED OPERATION
Page 48

OFFSETS

The Model 3430 is factory-calibrated for soils, asphalt, and concrete with an approximate density range of 1100 to 2700 kg/m pcf). With an offset, the operator can adjust the gauge readings to correlate to traditional laboratory methods, such as core samples. The 3430 gauge provides three offsets: density, moisture, and trench.
NOTE
When an offset has been enabled, all future readings will automatically be adjusted with the offset factor regardless of the test site. It is very important that the operator disable the offset function prior to taking readings on materials that do not require an offset
Offsets are disabled if the gauge is turned off for more than 10 seconds.
Density offsets are common when the material being measured is outside the range of 70 to 170 pcf (1121 to 2723 kg/m material composition varies from average soil/asphalt on which the factory calibration is based.
Moisture offsets are required for accurate measurements if the material to be measured contains elements that can cause the gauge to yield erroneous results. A negative offset is required if the material to be measured is high in hydrogenous components such as cement, gypsum, coal, or lime. A positive offset is required if the material is high in neutron-absorbing material such as boron or cadmium.
The 3430 gauge requires an offset if measurements are to be taken inside a trench or close to vertical structures. Vertical structures can scatter neutrons and gamma photons back to the gauge, increasing the possibility of moisture or density errors due to high counts.
3
(70 to 170
.
3
) or if the
4-2
Page 49

DENSITY OFFSET

To access the Special functions, press SPECIAL.
Press the down arrow key once to access the Offset function. Press START/ENTER to display:
Offset: Density (↑ ↓ or ENTER)
Press START/ENTER.
Dens. Offset OFF Want to Enable?
To enable the Density Offset function, press ON/YES.
D off= 0.0 PCF (↑ ↓ or ENTER)
Input the difference between the gauge and actual density readings. To input a minus sign (for a negative offset), press the down arrow first. To scroll through the numerals, press up and down arrows.
To select the next digit and/or exit, press START/ENTER.
The display will be:
Dens. Offset ON
Model 3430 4-3
ADVANCED OPERATION
Page 50

MOISTURE OFFSET

Some soils contain hydrogen sources other than water and/or neutron absorbers. Since the 3430 gauge measures moisture by determining the hydrogen content of the material and relating this to the water content, both types of material could cause gauge readings that differ from the true moisture. If measuring such materials, use a moisture offset to adjust the readings.
The offset factor (k) is determined by comparing the moisture content of a laboratory sample with the moisture content determined by a gauge reading. To determine the offset factor, use the following procedure:
Take a gauge reading at the site. Record the reading (%M
Remove a sample from the measurement site, then use
laboratory methods methods (for example, oven dry, and so on) to determine the moisture content of the sample (%M
LAB
Multiple samples and measurements may be taken. Calculate the average moisture of the samples. This average value should be used for the offset factor calculation.
Calculate the offset factor (k).
k =
%M 100 + %M
LAB
– %M
GAUGE
GAUGE
×
1000
NOTE
If the k value is negative, a minus sign (–) must be entered by pressing the down arrow before entering the first digit.
To access the Special functions, press SPECIAL.
GAUGE
).
).
4-4
Page 51
Press the down arrow key once to access the Offset function. Press START/ENTER to display:
Offset: Density (↑ ↓ or ENTER)
To enter a moisture offset, press the down arrow once and press START/ENTER.
Moist Offset OFF Want to enable?
To enable the Moisture Offset function, press ON/YES.
K= 0.0 (↑ ↓ or ENTER)
The first digit will flash. To input a minus (–) sign (for a negative offset), press the down arrow first values for each digit, press the arrow key. To select the next digit, press START/ENTER. When all digits are entered, the gauge will enable the offset. The display will be:
! To scroll through the possible
Moist Offset ON
Model 3430 4-5
ADVANCED OPERATION
Page 52

TRENCH OFFSETS

If the Model 3430 gauge is to be used for moisture or density measurements in a trench or within two feet (0.6 m) of a large vertical structure, a trench offset may be required. If used, the trench offset adjusts all moisture measurements but only the density measurements from backscatter through four inches (10 cm).
To perform a trench offset:
Take the daily standard count (outside the trench) and record the
density count (DS) and moisture count (MS)values.
Place the gauge on the reference standard block in the trench the
same distance from the wall as the anticipated readings. Do not take another standard count.
Set the count time to four minutes.
With the source rod in the SAFE (standard count) position, take
a four-minute count. To start the count, press the START/ENTER key.
Record the trench density count (DC
(MC
Trench
.).
Subtract the daily standard count values from the trench count
values:
Dens Const = (DC
Trench
) – DS
) and moisture count
Trench
Moist Const = (MC
Trench
) – MS
To enable a trench offset, choose Trench from the Offset options under the Special functions. The gauge requests the Dens Const and Moist Const values determined above. The procedure for entering the values is the same as for moisture and density offsets, ignoring
the ± sign on the display.
4-6
Page 53

SPECIAL CALIBRATIONS

Troxler gauges are calibrated to “average soil.” Average soil is defined as material consisting of 50% limestone (calcareous) and 50% granite (siliceous). This factory calibration provides accurate results for the majority of materials encountered in the field. However, there are situations when varying material compositions could affect the gauge accuracy. In these special cases the gauge B value can be recalculated either by the gauge or by considering the mass attenuation (µ/ρ) of the material.
If the chemical composition of the soil is known, Troxler can provide a procedure for recalculating the B value for manual entry (see page 4-8) to accurately measure the soil density. This calculation requires in-depth knowledge of the gauge geometry and the detected energy spectrum of the Cs-137 source.
The Special Calibration function allows the Model 3430 gauge to be recalibrated for material densities and compositions other than those covered by the factory calibration.
The true density of a sample of the material must be obtained prior to calculating a special calibration. This density may be obtained from a laboratory sample.
To access the Special functions, press SPECIAL.
To access the Special Calibration function, press the down arrow four times. Press START/ENTER to display:
SPECIAL CALIB. Want to Recalib?
ADVANCED OPERATION
Model 3430 4-7
Page 54
To recalibrate the gauge for the densities outside the factory calibration range, press ON/YES. To disable the Special
Calibration feature, press OFF/NO〉 at the above display and ON/YES at the disable inquiry. After disabling this feature, the
gauge will return to the Ready mode.
SPECIAL CALIB. Enter B Value?
To enter a known B value obtained with the procedure available from Troxler, press ON/YES. To have the gauge calculate the recalibration, press OFF/NO. If entering a new known B value, see the following explanation. For gauge-calculated special calibration, see page 4-9.

ENTERING A NEW B VALUE

Depth = xx in (↑ ↓ or ENTER)
To change the value of the flashing digit for the measurement depth, use the up and down arrows. To accept the flashing value and select the next digit, press START/ENTER.
B VAL = x.xxxx (↑ ↓ or ENTER)
The gauge displays the current B value. To change the value of the flashing digit, use the up and down arrows. To accept the flashing digit and select the next digit, press START/ENTER.
4-8
Page 55
Upon entry completion, the gauge will indicate that the special calibration is enabled and return to the Ready mode. Note that when the gauge is turned off the Special Calibration is disabled.
SPECIAL CALIB. ENABLED!

GAUGE-CALCULATED CALIBRATION

The true density of the sample and a gauge reading must be performed on the material for the special calibration routine to adjust the gauge calibration.
NOTE
When using destructive methods such as drilling cores or sample removal for true density measurement, take gauge readings before removing samples.
To select the depth and/or exit, press START/ENTER.
Depth: xx in. (↑ ↓ or ENTER)
To scroll through the numerals for the depth of the measurement, press the up and down keys. To select the next field and/or exit, press START/ENTER.
If calibration counts have not been taken, the gauge will take four one-minute counts.
The gauge provides the operator with a partial calibration feature. Since a partial calibration allows the operator to enter the density after taking counts, it is helpful to those performing destructive material testing. If the operator has previously taken calibration counts, the gauge asks if these counts should be used in calibrating the gauge. To use the previous counts, press ON/YES. The gauge will then request the density. To take new counts, press OFF/NO.
Model 3430 4-9
ADVANCED OPERATION
Page 56
Press START for Reading # x
Place the gauge on the test material. To begin taking the four one­minute counts, press START/ENTER. After each count is complete, the operator must initiate the next count by pressing START/ENTER.
Enter known density now?
To create a partial calibration and return to the Ready mode, press OFF/NO. To complete the special calibration by entering the density, press ON/YES.
Density: xx (↑ ↓ or ENTER)
To change the value of a digit, press the up and down arrows. To select the next digit, press START/ENTER. After the density value is entered, the special calibration routine readjusts the gauge for the new material and indicates that the special calibration is enabled. The special calibration is only valid for the depth selected during the special calibration.
SPECIAL CALIB. ENABLED!
4-10
Page 57

THIN LAYER MEASUREMENTS

With the increase in thin lift overlay applications and the limitations of conventional backscatter gauges to measure these overlays the following method (formula) has been developed:
DT =
where:
DT = Overlay density WD = Density read by gauge DB = Bottom layer density K = Effect of top layer thickness on the gauge
To use the above method of overlay measurement, follow the procedure below:
Determine the density of the bottom layer (underlying material)
(DB).
Apply the thin lift overlay.
Determine the thickness of the overlay and select the
corresponding (k) value from Table 4-1.
Measure the thin lift overlay density with the gauge in
backscatter position (WD).
Enter all values into the above equation and calculate the
overlay density (DT).
WD – DB
1 – K
×
K
ADVANCED OPERATION
Model 3430 4-11
Page 58
EXAMPLE
Given the following values:
Bottom Density (DB) = 135 pcf (2162 kg/m
Overlay Thickness = 1.2 inches (30 mm)
K (from Table 4-1) = 0.38235
Density read by gauge (WD) = 142.0 pcf (2275 kg/m
×
DT =
142.0 – (135 1 – 0.38235
0.38235)
DT = 146.3 pcf
or,
DT =
2275 – (2162 1 – 0.38235
DT = 2345 kg/m
×
0.38235)
3
NOTE
The majority of the backscattered gamma rays reaching the detectors are the result of interactions in the top 3.3 inches (84 mm) of the overlay. In applications where the overlay thickness is greater than 3.3 inches (84 mm), use (0) for the k value or use the actual gauge readings (WD).
3
)
3
)
4-12
Page 59
Table 4-1. K Values for Thin Lift Overlays
Thickness
(inches)
1.0 25 0.46159 55 0.13459 26 0.44787 2.2 56 0.12880 27 0.43414 57 0.12078
1.1 28 0.42042 2.3 58 0.11275 29 0.40138 59 0.10781
1.2 30 0.38235 60 0.10285 31 0.36475 2.4 61 0.09790 32 0.35889 62 0.09104
1.3 33 0.34716 2.5 63 0.08418 34 0.33631 64 0.07995 35 0.32547 65 0.07572
1.4 36 0.31462 2.6 66 0.07149 37 0.29958 67 0.06562
1.5 38 0.28454 2.7 68 0.05976 39 0.27527 69 0.05615 40 0.26600 70 0.05253
1.6 41 0.25673 2.8 71 0.04892 42 0.24387 72 0.04390
1.7 43 0.23102 2.9 73 0.03889 44 0.22310 74 0.03580 45 0.21517 75 0.03271
1.8 46 0.20725 3.0 76 0.02962 47 0.19626 77 0.02676
1.9 48 0.18527 78 0.02391 49 0.17850 3.1 79 0.02105 50 0.17172 80 0.01709
2.0 51 0.16495 3.2 81 0.01313 52 0.15556 82 0.01069
2.1 53 0.14617 83 0.00825 54 0.14038 3.3 84 0.00581
Thickness
(mm)
K
Thickness
(inches)
Thickness
(mm)
K
ADVANCED OPERATION
Model 3430 4-13
Page 60
NOTES
4-14
Page 61
CHAPTER 5
SPECIAL FUNCTIONS
This chapter gives brief explanations of the Special functions available on the Model 3430 Surface Moisture-Density Gauge, or directs the reader to the appropriate section dealing with a particular function.
CONTENTS
Recall.............................................................................................5-2
Offset .............................................................................................5-2
Stat Test.........................................................................................5-2
Drift Test........................................................................................5-5
SPECIAL FUNCTIONS
Special Calibration.........................................................................5-7
Specific Gravity.............................................................................5-7
Voidless Density............................................................................5-8
Set Units.........................................................................................5-8
Calibration Constants.....................................................................5-9
Memory Reset..............................................................................5-10
Test Reading................................................................................5-10
Language......................................................................................5-11
15-second Inhibit.........................................................................5-12
Model 3430 5-1
Page 62

RECALL

The Recall function allows the operator to view the data from the last reading. Even though the 3430 gauge does not store multiple readings, this function displays the latest data.
To access the Special functions, press SPECIAL.
To access the Recall feature, press START/ENTER.
The gauge displays the data from the last measurement. Scroll through the screens using the up and down arrow keys.

OFFSET

For information on offsetting gauge readings, refer to Chapter 4.

STAT TEST

The statistical stability test, or stat test, may be performed to validate the normal operation of the gauge. Erratic readings or readings that seem to fluctuate may indicate a problem with the gauge. In the event the readings are suspect, a stat test may be executed.
A stat test consists of twenty 1-minute counts. After the twenty counts, the gauge calculates the standard deviation. This standard deviation is compared to a theoretical standard deviation value. Ideally this ratio should be one. However, the 3430 gauge pre-scales (or divides) the counts by 16, resulting in an ideal ratio of 0.25. The acceptable limits for the ratio are from 0.17 to 0.33. The gauge is considered to be unstable if the ratio is outside these limits.
5-2
Page 63
To perform a stat test on the 3430 gauge, place the gauge on the reference standard block in the standard count position (see Chapter 3).
To access the Special functions, press SPECIAL.
To access the Stat Test feature, press the down arrow twice and press START/ENTER.
press START for
20 min Stat Test
To begin the twenty counts, press START/ENTER.
The gauge will display the stat test count progress as shown below.
– STAT TEST –
Rdg.#:xx xx sec
Upon completion of the stat test, the gauge displays the pass/fail status. If the stat test fails, repeat the test twice more. If two out of three stat tests fail, contact Troxler. If the stat test passes, the display is:
SPECIAL FUNCTIONS
D: PASS M: PASS to view data
To view the stat test data, use the up and down arrow keys.
Dens. R = xxxx ↑ ↓ to view data
Dens. Avg. xxxx ↑ ↓ to view data
Model 3430 5-3
Page 64
Moist R = xxxx ↑ ↓ to view data
Moist Avg. xxxx ↑ ↓ to view data
#20 D xxxx M xxxx (Use & keys)
5-4
Page 65

DRIFT TEST

If the stat test has already been performed (and passed), but gauge readings seem to drift between tests, the drift test can check the long-term drift of the 3430 gauge.
A drift test consists of five 4-minute counts taken approximately 3 to 8 hours after completion of a stat test with no movement of the gauge between tests. Pass/fail limits are set using the percent difference between the average of the stat and drift test results. If the percent difference exceeds 0.5% for density or 1% for moisture, the drift test fails.
The gauge should not be turned off between the stat test and drift test. The stat test must be current.
In addition, the gauge must not be moved between the stat and drift tests to eliminate possible failure due to positioning changes.
With the gauge still in the standard count position (on the reference standard block), press SPECIAL.
From the Special functions, select the Drift Test feature by pressing the down arrow three times and START/ENTER.
NOTE
SPECIAL FUNCTIONS
press START for 20 min Drift Test
To begin the five counts, press START/ENTER.
– DRIFT TEST –
Rdg.#:xx xxx sec
As with the stat test, the gauge indicates the count progress during the drift test.
Model 3430 5-5
Page 66
After the five counts have been completed, the display is:
D: PASS M: PASS to view data
To view the stat test data, use the up and down arrow keys.
D% Drift xxxx ↑ ↓ to view data
Dens. Avg. xxxx ↑ ↓ to view data
M% Drift xxxx ↑ ↓ to view data
5-6
#5 D xxxx M xxxx (Use and keys)
Page 67

SPECIAL CALIBRATION

For information on performing a special calibration, see Chapter 4.

SPECIFIC GRAVITY

The specific gravity of a solid is defined as the density of the material divided by the density of water. The Specific Gravity function allows the operator to input the specific gravity of a material into the gauge. This value (SG) is used in the calculation of % Air Voids and Void Ratio (see Chapter 3).
To access the Special functions, press SPECIAL.
To access the Specific Gravity feature, press the down arrow five times and press START/ENTER.
SG = 2.70 (↑ ↓ or ENTER)
To change the value of the flashing digit, use the up and down arrows. To accept the flashing value and select the next digit, press START/ENTER.
If a value is not entered, the default value is the specific gravity for soil grains (2.70).
SPECIAL FUNCTIONS
Model 3430 5-7
Page 68

VOIDLESS DENSITY

The Voidless Density function allows the input of the theoretical voidless density value of the material being measured. This value is used in the % Voids calculation.
To access the Special functions, press SPECIAL.
To access the Voidless Density feature, press the down arrow six times and press START/ENTER.
VD = xxx.x (↑ ↓ or ENTER)
To change the value of the flashing digit, use the up and down arrows. To accept the flashing value and select the next digit, press
START/ENTER.

SET UNITS

For information on the Set Units feature, see Chapter 3.
5-8
Page 69

CALIBRATION CONSTANTS

The Calibration Constants function allows the operator to change the mathematical constants used for calculating a test result. If the gauge has been repaired or the memory has been lost, the constants must be verified or re-entered.
NOTE
Each 3430 gauge contains a unique set of constants. The constants used in one gauge will not work in another gauge! The constants for your 3430 gauge are developed at the factory and are reflected on the factory calibration sheet.
To access the Special functions, press SPECIAL.
To access the Calibration Constants feature, press the down arrow key eight times and press START/ENTER.
Enter code –
0
This feature requires the input of the access code found in the front of this manual. Using the up and down arrow keys to select the correct number for the flashing digit, enter the access code. To accept the flashing value and select the next digit, press START/ENTER.
The gauge will prompt for the input of the E value.
NOTE
If the value is negative, enter a minus sign (–) by pressing the down arrow key prior to entering the first digit. Leading zeros must be entered (for example:
0.012345).
To change the value of the flashing digit for the E value, use the up and down arrows. To accept the flashing value and select the next digit, press START/ENTER.
SPECIAL FUNCTIONS
Model 3430 5-9
Page 70
Enter the remaining constants for each depth.
NOTE
If the calibration sheet lists B and F values, instead of B*1000 and F*1000 values, then the calibration is in English, rather than metric, units. Therefore, the B and F values must be converted to metric values and multiplied by 1000 before they are entered into the gauge. A more direct method is to multiply the B and F values by 62.4298, then enter the resulting products into the gauge.

MEMORY RESET

This function is for authorized service personnel only!
NOTE
CAUTION
Memory Reset erases all data stored in the gauge and sets all constants, except calibration constants, to the default values.

TEST READING

This function is for authorized service personnel only!
5-10
NOTE
Page 71

LANGUAGE

The 3430 gauge supports three language displays (English, French, and Spanish). Troxler also offers keypad inserts in each of these three languages (see Appendix C).
To change the display language, first access the Special functions by pressing SPECIAL. Press the up arrow once.
– LANGUAGE – (↑ ↓ Or ENTER)
To access the Language feature, press START/ENTER.
Enter code –
0
To prevent unauthorized language changes, this feature requires the input of the access code found in the front of this manual. Using the up and down arrow keys to select the correct number for the flashing digit, enter the access code. To accept the flashing value and select the next digit, press START/ENTER.
SPECIAL FUNCTIONS
English (↑ ↓ Or ENTER)
To display all gauge operations in English, press START/ENTER. The gauge returns to the Ready mode.
To access either of the other two languages, use the arrow keys. When the desired language is displayed, press START/ENTER. The gauge returns to the Ready mode.
Model 3430 5-11
Page 72

15-SECOND INHIBIT

The 15-second Inhibit function enables the gauge owner or operator to disable the 15-second count option. When this function is enabled, the 3430 gauge can only conduct one- or four-minute counts.
To disable the 15-second count option, first access the Special functions by pressing SPECIAL. Use the up or down arrows to display:
– 15 SECONDS – (↑ ↓ Or ENTER)
To access the 15-second Inhibit function, press START/ENTER.
Enter Code –
0
This feature requires the input of the access code found in the front of this manual. Using the up and down arrow keys to select the correct number for the flashing digit, enter the access code. To accept the flashing value and select the next digit, press START/ENTER.
If the 15-second count option is currently enabled, the gauge
displays:
– 15 SECONDS – Want to DISABLE?
Press ON/YES to disable the 15-second count option. The gauge returns to the Ready mode.
If the 15-second count option is currently disabled, the gauge
displays:
– 15 SECONDS – Want to ENABLE?
Press ON/YES to enable the 15-second count option. The gauge returns to the Ready mode.
5-12
Page 73
APPENDIX A
RADIATION THEORY AND SAFETY
This appendix is required reading for anyone who will operate the Model 3430 Surface Moisture-Density Gauge. This information covers radiation theory, along with a brief explanation of radiation statistics and radiation terminology.
CONTENTS
Radiation Theory .......................................................................... A-2
Atomic Structure.................................................................... A-2
Radiation Terminology ..........................................................A-3
Radiation Theory.................................................................... A-3
Radiation Statistics.................................................................A-4
Radiation Safety............................................................................A-5
Types of Radiation................................................................. A-5
Limiting Exposure.................................................................. A-5
Monitoring Radiation.............................................................A-7
3430 Radiation Profile ...........................................................A-8
Source Encapsulation...........................................................A-11
Emergency Procedures......................................................... A-12
RADIATION THEORY
Model 3430 A-1
Page 74
N
RADIATION THEORY
A more detailed discussion of radiological theory can be found in the Troxler Nuclear Gauge Safety Training Program manual, provided at the Troxler safety class.

ATOMIC STRUCTURE

All materials consist of chemical elements that can not decompose by ordinary chemical methods. Some examples are:
(H) Hydrogen (C) Carbon (O) Oxygen (U) Uranium (Cf) Californium (Co) Cobalt
Each element contains an atom with a unique structure. The atom consists of protons, neutrons, and electrons (Figure A-1). The protons and neutrons are grouped together in the nucleus. The electrons orbit the nucleus. An atom is normally electrically neutral because the positive charge of the protons cancels out the negative charge of the electrons.
ELECTRONS
NEUTRON
PROTO
Figure A-1. Diagram of an Atom
Protons carry a positive charge and are described as having a mass of one. Neutrons have no charge and also have a mass of one. Electrons carry a negative charge and essentially have no mass.
MASS (ATOMIC WEIGHT SCALE)
Protons 1.0073 +1 Neutrons 1.0087 0 Electrons 0.0006 –1
CHARGE
A-2 Model 3430 A-3
Page 75
Since protons and neutrons are clustered together in the nucleus, the mass of an atom is concentrated in the nucleus. The atom shown in Figure A-1 has two protons and two neutrons; therefore, it is a helium atom. The atomic weight of an atom is the sum of the number of protons and the number of neutrons.

RADIATION TERMINOLOGY

The curie, defined as the quantity of radioactive material giving 3.7
10
× 10
disintegrations per second, is equal to the number of disintegrations/second of one gram of radium-226. Note that the source used in the 3430 is small, with quantities expressed in millicurie (mCi). The SI unit of radiation is the becquerel (Bq). A becquerel equals one disintegration per second. Therefore, one curie equals 3.7 × 10
10
Bq.
The rad, or radiation absorbed dose, is the unit of absorbed dose equal to 0.01 joules/kg in any medium. To account for the effect of various types of radiation on biological tissue, the roentgen equivalent man (rem), or, more appropriate for Troxler operators, the millirem is used when measuring radiation dose. The unit rem is derived by multiplying the radiation absorbed dose (rad) by a quality factor (QF). One rad is equal to the exposure of one rem of photon radiation. For example, the average neutron energy of an americium-241:beryllium source is approximately 4.5 MeV. The quality factor (QF) for this radiation is about 10. The absorbed dose of 1 rad of neutron radiation gives a dose equivalent (absorbed dose × QF) of 10 rem (0.1 Sv).

RADIATION THEORY

Radioactivity is the spontaneous breakdown of unstable nuclei (radioisotopes) with the resulting emission of radiation. The strength of radioactive material is measured by its activity, or rate of decay. This activity decreases with time. The length of time it takes a given amount of radioactive material to decay to half of its original strength is referred to as the half-life. The half-life of cesium-137 is 30 years, while that of americium-241 is 432 years. Californium­252 has a half-life of 2.6 years.
RADIATION THEORY
Page 76
σ-1σ
σ2σ

RADIATION STATISTICS

Radioactive emission is a random process. The number of emissions in a given time period is not constant but varies statistically about an average value. The variation about the true mean value is a Poisson distribution. In this distribution, the standard deviation ( mean (n) is defined as:
σ
= √n
When the mean is greater than 100, the Poisson distribution can be closely approximated by the normal distribution (Figure A-2). The normal distribution predicts the probability that any given count rate will fall within a selected region about the mean.
Normal Distribution
σ
) about the
σ
-3
-2
MEAN
68.3%
95.4%
99.7%
σ
1
3
Figure A-2. Variation of Radioactive Emission
Using the average of a large number of counts to approximate the true mean, the distribution shows that 68.3% of the time the count rate obtained will be within ±1 standard deviation of the mean. Figure A-2 shows the probability of counts falling within three standard deviations of the mean. The operator may perform a statistical stability test (stat test) to compare the experimental standard deviation to the theoretical standard deviation (see Chapter 5).
A-4 Model 3430 A-5
Page 77

RADIATION SAFETY

This section provides a brief discussion of general radiation safety. The exposure profile for the Model 3430 gauge is also included, along with a discussion of the source encapsulation.

TYPES OF RADIATION

The radioactive sources in the Model 3430 produce four types of radiation:
Alpha Particles Beta Particles Photons (Gamma Rays) Neutrons
The alpha and beta particles are stopped by the source capsule. Only the photons and neutrons contribute to any occupational radiation exposure.
Photon (gamma) radiation is electromagnetic radiation, as are x­rays, radio waves, and visible light. Photons have no mass, no electrical charge, and travel at the speed of light. Photons are energetic and penetrating. Dense materials (such as lead, tungsten, and so on) provide the best shielding against photon radiation.
Neutron radiation allows measurement of the hydrogen (water) content in a material because the neutrons are slowed by collisions with materials containing hydrogen atoms (i.e. water, polyethylene, etc). Neutrons have no charge and are very penetrating.

LIMITING EXPOSURE

Government agencies set occupational exposure limits. The current limit in the United States and many other countries is 5,000 mrem per year. Under average conditions, a full-time employee working with the 3430 gauge will receive less than 200 mrem per year. By comparison, people in the US receive an average of 360 mrem per year from natural background radiation and medical radiation.
RADIATION THEORY
Page 78
Taking advantage of all available means to limit radiation exposure is always recommended. The three methods of limiting exposure are:
Time Distance Shielding
These methods are a part of the ALARA (As Low As Reasonably Achievable) philosophy of radiation protection.
Time
The simplest way to reduce exposure is to keep the time spent around a radioactive source to a minimum. If time is cut in half, then exposure is cut in half, provided all other factors remain constant.
Distance
Distance is another effective means to reduce radiation exposure. A formula known as the inverse square law relates the radiation exposure rate to distance (Figure A-3). Doubling the distance from a radiation source reduces the exposure to one-fourth its original value. If the distance is tripled, the exposure is reduced by a factor of nine, and so on.
Figure A-3. Effect of Distance on Exposure
A-6 Model 3430 A-7
Page 79
Shielding
Shielding is any material used to reduce the radiation reaching the operator from a radioactive source. While some types of radiation (such as alpha particles) may be stopped by a single sheet of paper, other radiation (such as photons and neutrons) require much more shielding. Dense materials, such as lead, shield photons. Materials containing large amounts of hydrogen, such as polyethylene, shield neutrons. The Model 3430 gauge has shielding built into the system to reduce the exposure.

MONITORING RADIATION

In the United States, anyone working with or near radioactive materials is subject to the limits on occupational exposure mentioned earlier and must complete a radiation safety training course to be designated an authorized user. To verify that occupational exposures do not exceed the regulatory limits, authorized users may be monitored using personnel dosimeters. The most common methods of personnel monitoring are thermoluminescent dosimeter (TLD) badges and film badges. Troxler recommends using TLD badges, since they can measure both gamma and neutron radiation. Film badges are not suitable for measuring neutrons.
In Canada, nuclear gauge users are not normally classified as Atomic Radiation Workers. In such cases, the general public dose limit of 0.5 rem per year would apply to nuclear gauge users. Users may not be required to wear a dosimeter. To establish the personnel monitoring requirements for your application, consult the conditions of your radioactive materials license and the Canadian Nuclear Safety Commission (CNSC) regulatory document R91, Monitoring and Dose Recording for the Individual.
RADIATION THEORY
Page 80

RADIATION PROFILE

Table A-1 shows the radiation profile for the Model 3430 gauge; Table A-2 shows the profile for the Model 3430-M gauge. Each table lists the radiation dose equivalent rates (in mrem/hour) for each side of the gauge and transport case shown in Figure A-4.
TRANSPORT CASE GAUGE
Figure A-4. Model 3430 Gauge and Transport Case
A-8 Model 3430 A-9
Page 81
Table A-1. Radiation Profile for Model 3430 Gauge
SURFACE 10 CENTIMETERS 30 CENTIMETERS 1 METER
Gamma
Neutron
Total
Gamma
Neutron
Total
Gamma
Neutron
Total
Gamma
Neutron
Total
GAUGE Front 13.0 1.7 14.7 5.0 1.7 6.7 1.1 0.3 1.4 0.3
Back 26.0 1.4 27.4 8.0 1.4 9.4 2.5 0.5 3.0 0.4 Left Side 13.0 0.5 14.0 4.0 0.5 4.5 0.7 0.25 0.95 0.1 Right Side 12.0 0.7 13.0 8.0 0.7 8.7 2.5 0.25 2.75 0.4 Top 19.0 1.7 20.7 8.0 1.7 9.7 0.6 0.7 1.3 0.15 0.1 0.25 Bottom 18.0 6.0 24.0 2.5 6.0 8.5 0.6 0.9 1.5 0.1 0.1 0.2
TRANSPORT CASE WITH GAUGE Front 10.0 0.7 10.7 5.0 0.7 5.7 1.2 0.45 1.6 0.25  0.25
Back 7.0 0.8 7.8 3.0 0.8 3.8 0.8 0.25 1.1 0.1  0.1 Left Side 0.3 0.1 1.4 0.25 0.1 0.35 0.1 0.1 0.2 Right Side 5.0 3.0 8.0 2.0 3.0 5.0 0.6 0.75 1.3 0.2 0.1 0.3 Top 10.0 0.4 10.4 2.5 0.4 2.95 0.6 0.3 0.9 0.1 Bottom 10.0 0.7 10.7 5.0 0.7 5.7 2.0 0.2 2.2 0.3
0.3
0.4
0.1
0.45
0.1
0.3
RADIATION THEORY
NOTES:
1. All readings in mrem/hr.
2. indicates a reading less than or equal to 0.1 mrem/hr.
3. Gamma measurements made with Ludlum Model 14C Survey Meter, calibrated
March 9, 1990.
4.
Neutron measurements made with Nuclear Research Corp., Model NP-2 Survey Meter, calibrated March 22, 1990.
5. Dose rates measured by the State of North Carolina Department of
Environment, Health, and Natural Resources, Division of Radiation Protection.
Page 82
Table A-2. Radiation Profile for Model 3430-M Gauge
SURFACE 30 CENTIMETERS 1 METER
Gamma
Neutron
Total
Gamma
Neutron
Total
Gamma
GAUGE Front 12.0 4.5 16.5 1.5 1.4 2.9 0.3 0.2 0.5
Back 20.0 4.5 24.5 2.0 2.0 4.0 0.4 0.2 0.6 Left Side 13.5 2.0 15.5 3.0 0.8 3.8 0.7 0.2 0.9 Right Side 19.0 1.5 20.5 0.9 0.8 1.7 0.2 Top 18.0 5.0 23.0 0.8 1.8 2.6 0.3 0.2 0.5 Bottom 18.0 16.0 34.0 0.7 3.0 3.7 0.2 0.4 0.6
TRANSPORT CASE WITH GAUGE Front 10.0 1.9 11.9 1.4 0.6 2.0 0.3 0.2 0.5
Back 6.0 1.5 7.5 1.2 0.3 1.5 0.2 Left Side 8.0 5.0 13.0 0.8 1.6 2.4 0.2 0.3 0.5 Right Side 0.4 0.2 0.6 Top 10.0 1.2 11.2 0.9 0.4 1.3 0.2 Bottom 7.0 1.3 8.3 2.5 0.5 3.0 0.6
0.2 0.2
Neutron
 
Total
0.2
0.2
0.2
0.6
NOTES:
1. All readings in mrem/hr.
2. indicates a reading less than or equal to 0.1 mrem/hr.
3. Gamma measurements made with Ludlum Model 14C Survey Meter, calibrated January 7, 1991.
4. Neutron measurements made with Nuclear Research Corp., Model NP-2 Survey Meter, calibrated April 18, 1991.
A-10
Page 83

SOURCE ENCAPSULATION

The source in the Model 3430 gauge meets regulatory requirements of U.S. and international authorities as “Special Form” sealed source material. The sources are encapsulated to prevent leakage of radioactive material and meet radiation safety requirements.
The neutron source (americium-241:beryllium in the Model 3430 or californium-252 in the Model 3430-M) is compressed and then welded inside stainless steel capsules.
The photon (gamma) source (cesium-137) is sealed in a welded capsule.
Proper use of this instrument (following the instructions in this manual) and the shielding design of the instrument will keep the exposure levels at a minimum under normal conditions. The operator may, however, be required to wear personnel dosimetry when using the Model 3430 gauge.
RADIATION THEORY
Model 3430 A-11
Page 84

EMERGENCY PROCEDURES

If the nuclear gauge is lost or stolen, then immediately notify the gauge owner’s Radiation Safety Officer (RSO).
The gauge owner should complete the emergency contact information on the lines furnished below. (Note that company refers to the gauge owner’s company, not Troxler Electronic Laboratories.) This information should be readily available to the gauge operator at all times.
The company RSO is ____________________________________ Call the RSO at _________________________________________
The regulatory agency is _________________________________ Call the agency at _______________________________________
If a gauge is damaged, then follow the steps below:
Locate the gauge and/or source.
Do not touch or move the gauge.
Immediately cordon off an area around the nuclear gauge and/or
source. A radius of fifteen feet (5 m) will be sufficient. Do not leave the area unattended.
Keep all unauthorized personnel from the nuclear gauge.
If a vehicle is involved, it must be stopped until the extent of
contamination, if any, can be established.
The gauge operator should perform a visual inspection of the
nuclear gauge to determine if the source housing and/or shielding has been damaged.
Use a survey meter to measure the dose rate at a distance of
three feet (1 m) from the gauge.
A-12
Page 85
Contact the company RSO (name and number given at the
beginning of this section). Provide the RSO with the following:
The date, time, and location of the accident
The gauge model and serial number
The nature of the accident
The location and condition of the gauge and/or source
The dose rate at three feet (1 m) from the gauge.
If you are unable to reach the RSO, then call your regulatory
agency (name and number given at the beginning of this section).
Follow the instructions of the RSO. The RSO should report the
incident to the regulatory agency. The RSO may also be required to notify the U.S. DOT of accidents during transport.
Before shipping a damaged gauge to Troxler, obtain an RGA
(Returned Goods Authorization) number from the Troxler RSO as described in the Returning the Gauge for Service section of Appendix C.
RADIATION THEORY
Model 3430 A-13
Page 86
NOTES
A-14
Page 87
APPENDIX B
3430 SPECIFICATIONS
This appendix contains gauge and measurement specifications for the Model 3430 and Model 3430-M Surface Moisture-Density Gauges.
CONTENTS
Measurement Specifications......................................................... B-2
Density at 2000 kg/m Moisture at 240 kg/m
Density at 125 pcf .................................................................. B-3
Moisture at 15 pcf .................................................................. B-3
Radiological Specifications .......................................................... B-4
Electrical Specifications ............................................................... B-5
Mechanical Specifications............................................................ B-6
3
............................................................ B-2
3
............................................................ B-2
SPECIFICATIONS
Model 3430 B-1
Page 88

MEASUREMENT SPECIFICATIONS

DENSITY AT 2000 KG/M3 Direct Transmission (150 mm)
.25 min
1 min 4 min
Precision (kg/m Composition error (kg/m3) ±20.0 ±20.0 ±20.0 Surface error (kg/m
3
) ±6.8 ±3.40 ±1.70
3
) –17.0 –17.0 –17.0
(100% Void)
Backscatter (98%, 100 mm)
.25 min
1 min 4 min
Precision (kg/m Composition error (kg/m3) ±40.0 ±40.0 ±40.0 Surface error (kg/m
3
) ±16.0 ±8.00 ±4.00
3
) –75.0 –75.0 –75.0
(100% Void)

MOISTURE AT 240 KG/M3

.25 min
Precision (kg/m
3
) ±10.3 ±5.1 ±2.6
Precision is defined as ±one (1) standard deviation in density readings. This number is calculated by the ratio of the standard deviation in the counting rate and the slope of the calibration curve at a given density.
1 min 4 min
B-2
Page 89
DENSITY AT 125 PCF Direct Transmission (6 inches)
.25 min
Precision (pcf) ±0.42 ±0.21 ±0.11 Composition error (pcf) ±1.25 ±1.25 ±1.25 Surface error (pcf) –1.06 –1.06 –1.06 (100% Void)
1 min 4 min
Backscatter (98%, 4 inches)
.25 min
Precision (pcf) ±1.00 ±0.50 ±0.25 Composition error (pcf) ±2.50 ±2.50 ±2.50 Surface error (pcf) –4.68 –4.68 –4.68 (100% Void)
1 min 4 min

MOISTURE AT 15 PCF

.25 min
Precision (pcf) ±0.64 ±0.32 ±0.16
1 min 4 min
SPECIFICATIONS
Model 3430 B-3
Page 90

RADIOLOGICAL SPECIFICATIONS

Gamma Source 0.3 ±10% GBq (8 ±10% mCi) cesium-
137
Neutron Source Model 3430 1.48 ±10% GBq (40 ±10% mCi)
americium-241:beryllium
Model 3430-M 2.22 ±10% MBq (60 ±10% µCi)
californium-252
Source Type Sealed Source - Special Form
Source Housing Stainless Steel
Shielding Tungsten, Lead and Cadmium
Surface Dose Rate See Radiation Profiles in Appendix A.
Source Rod Material Stainless Steel
Shipping Case DOT 7A, Type A, Yellow II TI. 0.3 (0.6 for Model 3430-M)
B-4
Page 91

ELECTRICAL SPECIFICATIONS

Stored Power 15 watt-hours
Battery Recharge Time 14 to 16 hours
Gauge Charging 12 V dc, 500 mA minimum Requirements
Liquid Crystal Display 2 line x 16 character alphanumeric
Keypad 10-key sealed membrane
Power Consumption < 0.10 watts average
SPECIFICATIONS
Model 3430 B-5
Page 92

MECHANICAL SPECIFICATIONS

Base Cast Aluminum
Gauge Size (w/o handle) 14.45 × 8.85 × 6.45 in 367 × 225 × 164 mm
Gauge Height (w/ handle) 23.25 in (591 mm) for 12" rod
19.25 in (489 mm) for 8" rod
Transportation Case 29.5 × 14.0 × 17.0 in
74.9 × 35.6 × 43.2 cm
Weight 29 lb (13 kg)
Shipping Weight 86 lb (39 kg)
Operating Temperature Ambient: 14 to 158 °F –10 to 70 °C
Surface: 350 °F 175 °C
Storage Temperature –70 to 185 °F –55 to 85 °C
Vibration Test 0.1 in (2.54 mm) at 12.5 Hz
Drop Test 300 mm onto 25 mm steel ball
B-6
Page 93
APPENDIX C
PERIODIC MAINTENANCE AND SERVICE
This appendix contains information for maintaining and servicing the Model 3430 Surface Moisture-Density Gauge. The following procedures should be performed to keep the 3430 gauge in good working order. If a serious problem with the gauge arises, contact the nearest Troxler Service Center or representative for instructions.
CONTENTS
Troubleshooting............................................................................ C-2
Possible Malfunction Indicators............................................. C-6
Error Messages....................................................................... C-7
Battery Charging........................................................................... C-8
Alkaline Battery Use.............................................................. C-8
Mechanical Maintenance............................................................ C-10
Inspecting the Source Rod.................................................... C-10
Cleaning ............................................................................... C-10
Source Rod Bearing ............................................................. C-11
Tungsten Sliding Block........................................................ C-12
Gasket Replacement............................................................. C-14
Leak Testing ............................................................................... C-15
Replacement Parts ...................................................................... C-16
Gauge Assembly .................................................................. C-16
Base Mechanical Assembly ................................................. C-18
Source Rod Handle Assembly.............................................. C-18
Preamplifier Assembly......................................................... C-20
Scaler Assembly................................................................... C-22
Other Standard Parts............................................................. C-24
Accessories........................................................................... C-25
Maintenance Supplies .......................................................... C-25
Returning the Gauge for Service ................................................ C-26
Troxler Service Centers.............................................................. C-28
MAINTENANCE
Model 3430 C-1
Page 94

TROUBLESHOOTING

GAUGE FAILS STANDARD COUNTS
Ensure that the source rod opening on the gauge bottom is
completely closed or covered by the tungsten sliding block. If any opening is visible, the sliding block should be cleaned as described later in this chapter. If the sliding block still does not close completely, contact the nearest Troxler Service Center.
NO DENSITY READINGS
The most likely reason for no density readings is an electronic
problem, such as a failure of the detector preamplifier. However, as a precaution, ensure that the tip of the source rod is intact and undamaged (that is, ensure that the source is not missing). Use a radiation survey meter to check the radiation levels on contact with the surface of the gauge base (without extending the source rod). A maximum reading of 10-20 mrem/hr is normal, and indicates the source is present. However, if the maximum reading is less than 1 mrem/hr or if a survey meter is not available, perform a visual inspection of the source rod tip as follows to confirm its integrity:
1. Extend the source rod just far enough to see the source rod tip. The tip should appear flat to slightly rounded and smooth.
2. Stay at least three feet away from the tip of the unshielded source rod and complete the inspection as quickly as possible to minimize exposure (the dose rate at three feet from the unshielded source is about 2.7 mrem/hr).
If the visual inspection indicates that the source rod tip is broken off (source is missing):
1. Immediately contact your Radiation Safety Officer (RSO).
2. Initiate a search for the source starting at the location where the gauge was last used.
C-2
Page 95
3. Report lost or missing radioactive sources to your state or federal radiation control agency in accordance with applicable regulatory requirements.
4. Contact the Troxler Radiation Safety Department for further advice.
GAUGE READINGS APPEAR ERRATIC
Ensure that the source rod is properly locked in the desired
backscatter or direct transmission position, and is not resting on the test material.
Check the inside of the gauge for moisture. To dry the gauge
interior, remove the keypad. If necessary, use a hair dryer (on low heat) to circulate warm air for one to three hours.
Remove any foreign objects from inside the gauge.
Ensure the hardware mounting screws are tight and in place.
Check count time – a four-minute count will give the highest
precision with a repeatability of ±1 pcf.
Erratic density readings may be caused by a dirty sliding block.
Clean the sliding block as described on page C-12.
Perform a statistical stability (stat) test.
If test passes, proceed with job.
If test fails, repeat two more times. If test fails two out of
three times, contact the nearest Troxler Service Center.
NOTE
To aid in verifying gauge readings, after a gauge has been calibrated, mark a test area on a concrete floor, sidewalk, or equivalent and measure the density (WD). This measurement can then be used as a reference to verify later gauge readings.
MAINTENANCE
Model 3430 C-3
Page 96
GARBAGE OR XXXXXX IS DISPLAYED
Check the standard counts in memory. If the standard counts are
suspect, perform new standard counts. If counts equal zero for both systems, replace high voltage board (contact the nearest Troxler Service Center).
Check gauge for water damage. If the gauge is wet, dry the
gauge interior with hairdryer (on low heat) for 3 hours.
Check the calibration constants. They should match the
constants on your calibration data sheet if your calibration sheet is in metric units. (Refer to the note on page 5-10 to determine if your calibration sheet is in metric units and, if not, for instructions on converting B and F values from English to metric units.)
If necessary, perform a statistical stability (stat) test, record the
results and contact the nearest Troxler Service Center.
GAUGE TURNS OFF AFTER IT IS TURNED ON
The gauge automatically turns off after five hours if no keys are
pressed. Try to turn the gauge on again.
The gauge may be wet. Do not turn the gauge on until moisture
is removed from gauge interior! Component damage may result.
If the battery is below 3.0 volts, recharge or replace the
batteries.
The scaler may be defective. To test, replace the suspect scaler
with a good scaler.
SHORT BATTERY LIFE AFTER RECHARGING
NiCad batteries may be charged up to 100 full charge-discharge
cycles. The batteries may be reaching end of life cycle - replace. Note that all information stored in the gauge except the calibration constants and the chosen language is lost when the batteries are disconnected.
Charger/adapter may not be supplying full charge – check the
AC outlet and the DC output (12 VDC).
C-4
Page 97
Check the output voltage of your charger. The correct output
voltage is indicated on the charger unit.
Check that you are using the correct charger.
Remove any loose screws or foreign objects from the gauge
interior that may cause an electrical short to ground.
The AC charger may be defective. Check voltage output of
charge with a voltmeter, or use the DC charger to charge the batteries.
SATISFACTORY COUNTS, BUT RESULTS ARE IN ERROR
Ensure the measurement depth corresponds to the actual source
rod depth.
Check calibration constants.
Check to see if an offset (density, moisture, trench or special) is
enabled.
Ensure that the standard counts are correct.
Ensure that the index rod is seated in bottom of notch.
Model 3430 C-5
MAINTENANCE
Page 98
POSSIBLE MALFUNCTION INDICATORS CPU Board
Display Malfunctions No Keypad Response RAM Test Fails Batteries Do Not Recharge Battery Low Indicator Does
Not Function Correctly
Display Test Fails Gauge Doesn't Turn "Off" Beeper Stops (or is erratic) Gauge Does Not Turn On
When Charger Is Connected
Preamp Board
No Moisture or Density Counts Batteries Do Not Recharge
Gauge Fails Tube Test Fails Stability or Drift Tests
HV Board
No Moisture or Density Counts Moisture or Density Counts are
Unstable
Batteries Discharge
Prematurely
Gauge Fails Stability or Drift
Tests
C-6
Page 99

ERROR MESSAGES

The following error messages are not user-serviceable. Contact Troxler Customer Service for more information.
KEY PAD TEST ERROR! GM TUBE TEST ERROR! HELIUM TUBE TEST ERROR! DISPLAY TEST ERROR!
Model 3430 C-7
MAINTENANCE
Page 100

BATTERY CHARGING

With fully charged batteries, the 3430 gauge will remain operational for approximately eight weeks under normal (8-hour day) conditions.
If the batteries become discharged, the following message will be displayed on the gauge:
*** WARNING ***
Battery Low!
When this display appears, there are a few hours remaining before the battery must be recharged. In an emergency, a 30-minute recharge with the DC or AC charger gives several hours of use.
Although batteries cannot be “overcharged,” rechargeable batteries have a “memory” and repeated unnecessary recharging will shorten the battery life. If possible, run the batteries down before recharging.
NOTE Batteries should not be recharged unless the Battery Low! indication is displayed!

ALKALINE BATTERY USE

CAUTION
Do not mix alkaline and rechargeable batteries in the gauge. Charging may cause alkaline batteries to explode!
If recharging the NiCad batteries is not an option, alkaline batteries may be used. A separate battery case (PN 104148) can be purchased as an option. Note that all information stored in the gauge except the calibration constants and the chosen language is lost when the batteries are disconnected.
C-8
Loading...