Transcend 365 miniCPAP Auto User Manual

Auto User Manual
Transcend 365 miniCPAP AutoTM User Manual Page ii
Notices
Revised Transcend 365 Auto User Manual 103765 Rev A
Published November 2018 and supersedes all previous versions.
Notice The information contained in this document is subject to change without notice.
Trademark Somnetics
International, Inc.; all rights reserved.
Copyright © Copyright 2018 Somnetics International, Inc.; all rights reserved.
Contact Somnetics International, Inc.
®
and Transcend® are registered trademarks of Somnetics
Corporate headquarters Somnetics International, Inc.
Email info@somnetics.com
Web http://www.mytranscend.com/
Telephone 651.621.1800
Toll-free telephone 877.621.9626
Fax 651.204.0064
Somnetics International, Inc. 33 5th Avenue NW, Suite 500 New Brighton, Minnesota 55112 USA
Authorized Representative (MDD 93/42/EEC) Atlantico Systems Ltd.
Rx Only
34 Oldfield
Kingston
Galway, Ireland
TEL. +35391443609
Transcend 365 miniCPAP AutoTM User Manual Page iii
About this user manual
Note Training is required for the use of this device. All of the information required to
operate and maintain this device is found within this manual. It is expected that users will use this manual for their training.
Note Some images displayed may differ from the current product presentation
Transcend 365 miniCPAP AutoTM User Manual Page iv
Contents—
Introduction ............................................................................................... 6
Indications for use ............................................................................................ 6
Contraindications .............................................................................................. 6
Precautions for use ........................................................................................... 7
Warnings .................................................................................................... 7
Cautions ..................................................................................................... 9
Symbols ......................................................................................................... 11
Components of the Transcend® 365 miniCPAP Auto™ ..................................... 13
Accessories (sold separately)............................................................................. 13
The 365 miniCPAP Auto ............................................................................... 14
Assembling the 365 miniCPAP Auto™ ................................................................. 14
Getting Started With the 365 miniCPAP Auto™ Device ........................................... 15
Home Screen and Control panel .................................................................... 16
Power connection jack and USB port .............................................................. 18
Using the 365 miniCPAP Auto™ .................................................................... 19
Standard user modes ....................................................................................... 19
Starting therapy .............................................................................................. 19
Ending therapy and entering Dry Mode ............................................................... 20
The 365 miniCPAP Auto Humidifier ..................................................................... 21
Filling the Reservoir ......................................................................................... 21
Humidifier Parts ................................................................................................. 23
Replacing the Heater Wick ................................................................................ 24
Heater Cartridge Replacement ........................................................................... 24
Using the LCD Screen ........................................................................................ 25
Therapy Settings ............................................................................................. 25
Therapy Mode ............................................................................................. 25
Pressure Settings ........................................................................................ 25
Ramp Settings ............................................................................................ 26
EZEX pressure relief settings ........................................................................ 27
Language Settings ........................................................................................... 28
Last Session Report ......................................................................................... 29
Humidifier ...................................................................................................... 30
Compliance Report ........................................................................................... 31
Humidifier LCD indicators in Standby Mode ......................................................... 33
Water Status Indicators .................................................................................... 33
Heater Status Indicators ................................................................................... 35
Combination Indicators ..................................................................................... 37
Humidifier LCD indicators in Therapy Mode
......................................................... 38
Transcend 365 miniCPAP AutoTM User Manual Page v
Powering the 365 miniCPAP Auto ........................................................................ 40
Using the AC Power Supply ............................................................................... 41
Using the P
Charging the P
Battery ........................................................................................ 42
10
10 Battery
................................................................................... 43
Battery life expectancy................................................................................. 44
Using the Battery in-line with AC power .......................................................... 44
Caring for your Transcend 365 miniCPAP Auto™ ............................................ 45
Cleaning the unit ............................................................................................. 45
Caring for the humidifier reservoir
…………………………………………………………………………45
Cleaning Your Mask and Air Hose……………………………………………………………………………………46 Replacing the filter media
……………………………………………………………………….. ............ 46
Environmental Information ................................................................................ 46
Software User Guide ................................................................................... 47
Fault and alert codes ................................................................................. 56
Troubleshooting ........................................................................................ 58
Appendix: Part numbers ............................................................................. 59
Disposable parts .............................................................................................. 59
Accessories ..................................................................................................... 59
Replacement parts ........................................................................................... 59
Appendix: Specifications ............................................................................. 60
365 miniCPAP Auto™ ....................................................................................... 60
AC power supply PSA3 .................................................................................... 60
Battery ...................................................................................................... 60
P
10
365 miniCPAP Auto™performance ...................................................................... 61
Manufacturer’s declaration ................................................................................ 62
Electromagnetic emissions ............................................................................ 62
Electromagnetic immunity ............................................................................ 62
IEC 60601-1 compliance .............................................................................. 64
Performance ............................................................................................... 64
Appendix: Limited warranty ......................................................................... 66
Contact…………………..………………………………………………………………………………..…67
Transcend 365 miniCPAP AutoTM User Manual Page 6
Introduction
The 365 miniCPAP™ Auto by Transcend® is a medical device that provides positive airway pressure to users in the range of 4 to 20 cm
The 365 miniCPAP Auto may be used every day on a bedside table and can be
packed with you when you travel.
An interactive LCD interface makes the 365 miniCPAP easy to operate. The 365 miniCPAP Auto™ is the only continuous positive airway pressure (CPAP)
device licensed to use integrated Capillary Force Vaporization™ (CFV) humidification technology.* CFV instantaneously combines humidified vapor with the airstream when you inhale, providing a warm, moist, comfortable airflow. This humidification method requires less water than other humidification systems, which reduces the size of the humidifier.
A DC power jack and USB port are also incorporated into the 365 miniCPAP Auto™.
*Capillary Force Vaporization™ (CFV) humidification technology may be covered by one or more U.S. or foreign patents (patent 6,634,864, patent 7,431,570, patent 7,920,777, patent 8,201,752, patent 7,942,644, patent 5,008,207) and other U.S. or foreign patents pending.
H2O as prescribed by your clinician.
Indications for use
The ‘Transcend 365 miniCPAP Auto System’ provides positive airway pressure to support treatment of adults over 66 pounds (30 kg) with Obstructive Sleep Apnea (OSA). The device is intended for home and hospital/institutional use. The integrated humidifier provides humidification of air delivered from the Transcend 365 devices. The use of the
humidifier is optional.
Contraindications
The 365 miniCPAP Auto™ is contraindicated in patients with the following conditions:
Bullous lung disease
Pathologically low blood pressure
Pneumothorax or pneumomediastinum.
Pneumocephalus has been reported in some users using nasal PAP.
Caution should be used when prescribing PAP for susceptible users such as those with any of these conditions:
Cerebral spinal fluid (CSF) leaks
Abnormalities of the cribriform plate
A prior history of head trauma
Pneumocephalus
Transcend 365 miniCPAP AutoTM User Manual Page 7
Adverse effects
You should report unusual chest pain, severe headache or increased breathlessness to your prescribing physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment.
The following side effects have been reported by users of airway delivery devices during CPAP therapy.
congestion or mucus in the throat
sneezing or cough
bloating
nocturnal wakening
feelings of claustrophobia
burn
Irritation/dryness of the mouth, nose or throat
Precautions for use
This section describes the warnings and cautions associated with use of the 365 miniCPAP Auto™. The following guidelines apply to this document:
Warning Indicates the possibility of serious injury or death to yourself or others.
Caution Indicates the possibility of minor injury or damage to the equipment.
Note Indicates a tip, explanation or feature to aid in understanding, or efficient operation
of the device.
Warnings
Do not allow water to enter this device other than as specified in the humidifier
reservoir. The 365 miniCPAP Auto™ should not be exposed to environmental conditions where the system may get wet. If liquids are spilled onto the machine, unplug the machine and let the parts dry before plugging it back in.
This device is not intended for life support.
The 365 miniCPAP Auto™ must be set up and adjusted by a trained provider before
being used.
Do not block or otherwise obstruct the exhalation ports of the mask. Follow the
manufacturer’s instructions included with your mask.
Transcend 365 miniCPAP AutoTM User Manual Page 8
This equipment is not suitable for use with oxygen or in the presence of a
flammable anesthetic mixture with air or oxygen, or with nitrous oxide. Sources of oxygen must be located more than 1 m from the equipment to avoid the risk of fire and burns.
The 365 miniCPAP Auto™ is only to be used with the supplied or recommended
accessories. Use of accessories not recommended may result in increased electromagnetic emissions or decreased electromagnetic immunity of the PAP system and may be potentially unsafe.
The 365 miniCPAP Auto™ is not defibrillation proof.
Do not attempt to sterilize 365 miniCPAP Auto™.
The device is only compatible with appropriately marketed FDA cleared masks and
hoses. A mask should not be used unless the device is turned on and is properly delivering Ramp or therapy pressure. The proper placement and positioning of the mask on the face is critical to the consistent operation of this equipment. The exhalation port(s) associated with the mask should never be blocked. Explanation of the Warning: The device is intended to be used with masks or connectors specifically designed to have exhalation ports to allow continuous flow of air out of the mask. When the device is turned on and functioning properly air flow from the device flushes the exhaled air out through the mask exhalation port. When the device is not operating, however, fresh air will not be provided through the mask and exhaled air may be rebreathed.
Failure to use a mask or accessory that minimizes rebreathing of carbon dioxide or
permits spontaneous breathing can cause asphyxiation.
Strangulation hazard from power cord and air tube. These can become wrapped
around a neck and strangle.
Keep power cord and air tube more than 3 feet from a baby’s crib and out of a
baby’s reach. Keep cord and tube out of children’s reach.
Small parts are unlikely to be expelled from the 365 miniCPAP Auto™ enclosure,
but in case of severe damage internal components may fragment and create a swallowing or choking hazard if they get out of the enclosure.
Do not position the equipment in bed. Covering breathing tubes with a blanket or
heating them can affect the quality of therapy or injure the user.
To prevent disconnection of the tubing during use, only tubes in compliance with
ISO 5367 or ISO 80601-2-74 should be used.
Do not use the Transcend 365 miniCPAP Auto at an altitude above 8000’ (2438 m)
or outside a temperature of 41-95F (5-35 C). Using the humidifier outside of this temperature range or above this altitude can affect the quality of the therapy or injure the user.
Do not add any attachments or accessories to the humidifier or CPAP that are not
approved for use by this instruction. Unapproved accessories may lead to improper functioning of the device and affect the quality of the therapy or injure the user.
Transcend 365 miniCPAP AutoTM User Manual Page 9
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify they are operating normally.
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Transcend 365 miniCPAP Auto including cables specified by the manufacturer. Otherwise degradation of the performance of this equipment could result.
Cautions
Federal law (United States) restricts this device to sale by, or on the order of, a
physician.
Power the 365 miniCPAP Auto™ only with the Somnetics-supplied power supplies,
or batteries. See Appendix: Part Numbers.
Discontinue use of the 365 miniCPAP Auto™ and contact your physician if
respiratory or skin irritations occur.
Do not introduce objects into the 365 miniCPAP Auto™ air inlet or air outlet.
Inspect the power supply for signs of wear or damage before each use. Replace the
power cord if necessary.
Operating the humidifier with any substance other than distilled water will damage
the heating element and it will require replacement.
Do not use a nebulizer with the device.
To protect the environment, some parts and accessories of the 365 miniCPAP
Auto™, including optional batteries, must be disposed of in accordance with local regulations.
The equipment must not be positioned in an environment that can unacceptably
affect the basic safety and performance of the equipment. Examples of this would include:
o The equipment must not be positioned in a bed o The equipment should not be placed anywhere other than on a firm, flat
surface
o Do not position in a location where pets or children can access equipment o Do not position in an open window or other location where dust, sunlight, or
pests (insects) can affect equipment safety and/or performance
Transcend 365 miniCPAP AutoTM User Manual Page 10
o Magnets used to secure the humidifier reservoir to the CPAP may interfere
with other medical devices. Do not place the 365 miniCPAP within 6 inches (15 cm) of medical devices such as cardiac pacemakers, ICDs, and implantable neurostimulators.
The equipment must not be covered or positioned in such a way that the operation
or performance of the equipment is adversely affected. Examples of this would include:
o Do not position next to a curtain that blocks the flow of cooling air, thereby
causing the equipment to overheat
o Do not block the air intake port, thereby interfering with therapy
DBP
Device contains DBP (a phthalate). Significant exposure to DBP may interfere with the normal development of the male reproductive tract. Testing has demonstrated that DBP exposure levels are below established limits. Women who are pregnant or nursing may wish to discuss the benefits and risks of this device with a physician.
Transcend 365 miniCPAP AutoTM User Manual Page 11
Symbols
IP22
(the following symbols may appear on the product or packaging)
Caution
Protected against finger-sized objects and against dripping water when tilted 15 degrees from specified orientation
Type BF Applied Part
UL Seal of Approval demonstrating quality, safety and professional manufacturing of medical product
Temperature limit
Separate collection for electrical and electronic equipment per EC Directive 2002/96/EC. – Waste Electrical and Electronic Equipment (WEEE)
Non-Condensing
Rx Only
Consult instructions for use
Humidity limitation
Prescription only. U.S. federal law restricts this device to sale by or on the order of a physician or properly licensed practitioner.
Batch code
Date of Manufacture
Precedes reference or item number
Authorized Representative in the European Community
Transcend 365 miniCPAP AutoTM User Manual Page 12
Manufacturer
Fragile, handle with care
Keep dry
Serial Number
Use distilled water only. Use of any liquid other than distilled water will result in immediate damage to the humidifier and will void heater warranty.
Transcend 365 miniCPAP AutoTM User Manual Page 13
Components of the Transcend® 365
miniCPAP Auto™
Begin by unpacking all items from the 365 miniCPAP Auto™ travel bag and inspecting them to ensure they were not damaged during shipment. Report any missing or damaged items to the home healthcare provider that sold the 365 miniCPAP Auto™ to you.
Included with 365 miniCPAP Auto
365 miniCPAP Auto™ device and integrated humidifier
Universal power supply (PSA3) and power cords
Travel bag
Transcend® 365 miniCPAP Auto™ Quick Guide
USB cable
Accessories (sold separately)
Air Supply Tube (Compatible with standard 22 mm connector)
P
Battery™
10
Patient Mask
Transcend 365 miniCPAP AutoTM User Manual Page 14
The 365 miniCPAP Auto™ device
Your 365 miniCPAP Auto™ comes ready to generate and regulate continuous, positive airway pressure therapy when powered with the included AC power supply or optional battery.
No allergic reactions to materials used in the equipment have been identified with the 365 miniCPAP Auto™.
Assembling the 365 miniCPAP Auto Device
Your mask connects directly to the 365 miniCPAP Auto™ via the air supply tube.
Follow these steps to assemble the Transcend 365 miniCPAP Auto™:
1. Connect your mask (sold separately) to an air supply tube with a standard 22mm
connector (sold separately).
2. Connect the air supply tube directly to the 365 miniCPAP Auto™ device, making
sure it is fully seated.
3. Plug the power supply into the 365 miniCPAP Auto™. Place device on a flat, stable
surface.
4. Connect the appropriate power cord to the power supply and connect the power
cord to a wall outlet.
Transcend 365 miniCPAP AutoTM User Manual Page 15
Getting Started with the 365 miniCPAP Auto
Allow the 365 miniCPAP Auto to adjust to ambient temperature after removing from storage. To begin therapy, plug in the power supply to your 365 miniCPAP Auto and press the Center button to begin therapy. Press the down button to activate Ramp, if desired.
Note: To begin therapy from the home screen, first ensure that no icons are
highlighted, then press the Center button. You may also start therapy from any screen by holding the Center power button for three seconds.
Press Center button to begin therapy.
Then press down button to begin ramp.
Transcend 365 miniCPAP AutoTM User Manual Page 16
Note: If the miniCPAP™ device loses power while delivering therapy, the CPAP will power
on as soon as power is restored. The user must hit the Center button to resume therapy.
Home screen and Control Panel
Home Screen
Once power is supplied to the device, the LCD display lights up with the Home screen. The top right white numbers on the Home screen show the set pressure (large number) and the minimum and maximum pressures (small, lighter numbers - 4 and 20 in this image). The Home screen display quick menu options: Compliance Report (left), Settings (top), Last Session Report (right) and Humidification (bottom).
These functions are described in more detail in upcoming sections of the manual.
Control Panel
Use the buttons on the Control Panel below it to navigate through your options on the LCD screen. Pressing one of the four directional buttons (Up, Down, Left or Right) on the Control Panel will immediately bring up the section you highlight on the Home screen. When in a section, these buttons help you navigate to highlight areas and make choices.
The Center button on the Control Panel starts the PAP blower and takes you to the
Therapy Mode screen from the Home Screen. The Center button is used to ‘Select’,
‘Save Settings’ and begin CPAP therapy. Note: When changing settings, be sure to
press the Center button to save your choice.
The Down button takes you to the Humidifier section from the Home Screen. The
Down button is used to toggle to various selections and also activates Ramp once the CPAP is delivering therapy.
The Right button takes you to the Last Session Report section. It is also used to toggle
to various selections.
The Left button navigates you to your Compliance Report. It also acts as the ‘Back’
button while navigating through screens and will take you to the Home Screen.
The Up button brings you to the Settings screen from the Home Screen. It is also used
to toggle to various selections.
Transcend 365 miniCPAP AutoTM User Manual Page 17
(Settings & Toggle)
Up
Center
(Select, SAVE & Begin therapy)
Left
(Back & Home
screen)
Right
(Last Session &
Toggle)
Down
(Ramp & Toggle)
Transcend 365 miniCPAP AutoTM User Manual Page 18
Power connection jack and USB port
The power jack accepts the barrel plug of the output cable from the AC power source to operate the 365 miniCPAP Auto™.
An optional P
battery pack is also available
10
to power the 365 miniCPAP Auto™.
USB port
A mini-AB USB port is provided for direct data exchange between the 365 miniCPAP Auto™ and a computer. The 365 miniCPAP Auto™ LCD screen interface allows for basic settings adjustment and data viewing necessary for the patient. A key sequence allows clinicians to change therapy settings via the LCD interface.
The USB connection allows the clinician to configure the 365 miniCPAP Auto™ for prescription pressure, Ramp settings and Auto settings and also access therapy compliance information with the Transcend desktop software (password protected).
USB port
Power port
Transcend 365 miniCPAP AutoTM User Manual Page 19
Using the 365 miniCPAP Auto™
Standard user modes
Normal operation consists of four modes:
Off When the device is not connected to a power source, the device is off.
Standby When power is applied to the device the LCD screen lights up and the CPAP
enters Standby Mode. Standby Mode is also initiated by pressing the power button when the device is in On Mode or if the mask is removed while in On Mode. As long as power is supplied to the device it will remain in Standby Mode until On or Drying Mode is initiated.
On When in On Mode, the blower is working and regulated device therapy
pressure is being generated. The LCD lights remain off. On Mode is initiated by pressing the center button when the device is in Standby Mode and the mask is worn by the patient.
Drying Mode
This mode is accessed from Standby mode. Press and hold the Up and Down buttons at the same time for three seconds to start Drying Mode. This mode will last 30 minutes unless you interrupt it by pressing any button. Drying mode flushes your CPAP device, tubing and mask of residual moisture left from a therapy session.
Starting therapy
Press the Center button on the Control Panel to start the CPAP blower and go to the Therapy Mode screen. Press the down button to activate Ramp, if desired.
The average pressure readout will be visible. On the bottom of the screen you will see the set pressure (bottom left) and the set humidification level (bottom right).
Press the Left/Back button to return to the Home screen.
Transcend 365 miniCPAP AutoTM User Manual Page 20
Ending therapy and entering Drying Mode
When you are finished with your therapy session, press the Center button to return your CPAP to Standby mode.
Drying Mode
Press and hold the Up and Down buttons for three seconds to start Drying Mode. This mode will last 30 minutes unless you interrupt it by pressing any button. After the 30­minute drying cycle, the blower will turn off and the device will automatically enter Standby Mode. Dry mode flushes your CPAP device, tubing and mask of any residual moisture left from a therapy session.
Push and hold the Up and Down buttons to enter Drying Mode.
Transcend 365 miniCPAP AutoTM User Manual Page 21
The 365 miniCPAP Humidifier_________
Filling the Reservoir
1. To remove the reservoir from the CPAP, first disconnect the power source to your
CPAP, then set the device on a flat, stable surface and secure the reservoir with one
hand while you lift and remove the CPAP with your other hand. The pieces are
simply joined by magnets- there is not a button or latch that releases the humidifier from the CPAP device.
2. Hold the reservoir at a slight angle and fill it with distilled water only. This step is
best performed over a sink in case you spill water while filling the reservoir. (See image below.)
Caution! Distilled water is required for use with this device because of the
humidifier’s unique heating element. Failure to use distilled water will void component warranty. Failure to use distilled water will also damage the humidifier and its heating element, requiring these pieces to be replaced.
Caution! Do not perform this operation near electronic equipment. Always unplug the power source to your CPAP before attempting to remove the reservoir.