The 365 miniCPAP™ Auto by Transcend® is a medical device that provides positive
airway pressure to users in the range of 4 to 20 cm
The 365 miniCPAP Auto may be used every day on a bedside table and can be
packed with you when you travel.
An interactive LCD interface makes the 365 miniCPAP easy to operate.
The 365 miniCPAP Auto™ is the only continuous positive airway pressure (CPAP)
device licensed to use integrated Capillary Force Vaporization™ (CFV)
humidification technology.* CFV instantaneously combines humidified vapor with
the airstream when you inhale, providing a warm, moist, comfortable airflow. This
humidification method requires less water than other humidification systems, which
reduces the size of the humidifier.
A DC power jack and USB port are also incorporated into the 365 miniCPAP Auto™.
*Capillary Force Vaporization™ (CFV) humidification technology may be covered by one or more U.S. or foreign
patents (patent 6,634,864, patent 7,431,570, patent 7,920,777, patent 8,201,752, patent 7,942,644, patent
5,008,207) and other U.S. or foreign patents pending.
H2O as prescribed by your clinician.
Indications for use
The ‘Transcend 365 miniCPAP Auto System’ provides positive airway pressure to support
treatment of adults over 66 pounds (30 kg) with Obstructive Sleep Apnea (OSA). The
device is intended for home and hospital/institutional use. The integrated humidifier
provides humidification of air delivered from the Transcend 365 devices. The use of the
humidifier is optional.
Contraindications
The 365 miniCPAP Auto™ is contraindicated in patients with the following conditions:
Bullous lung disease
Pathologically low blood pressure
Pneumothorax or pneumomediastinum.
Pneumocephalus has been reported in some users using nasal PAP.
Caution should be used when prescribing PAP for susceptible users such as those with any
of these conditions:
Cerebral spinal fluid (CSF) leaks
Abnormalities of the cribriform plate
A prior history of head trauma
Pneumocephalus
Transcend 365 miniCPAP AutoTM User Manual Page 7
Adverse effects
You should report unusual chest pain, severe headache or increased breathlessness to
your prescribing physician. An acute upper respiratory tract infection may require
temporary discontinuation of treatment.
The following side effects have been reported by users of airway delivery devices during
CPAP therapy.
congestion or mucus in the throat
sneezing or cough
bloating
nocturnal wakening
feelings of claustrophobia
burn
Irritation/dryness of the mouth, nose or throat
Precautions for use
This section describes the warnings and cautions associated with use of the 365 miniCPAP
Auto™. The following guidelines apply to this document:
WarningIndicates the possibility of serious injury or death to yourself or others.
CautionIndicates the possibility of minor injury or damage to the equipment.
Note Indicates a tip, explanation or feature to aid in understanding, or efficient operation
of the device.
Warnings
Do not allow water to enter this device other than as specified in the humidifier
reservoir. The 365 miniCPAP Auto™ should not be exposed to environmental
conditions where the system may get wet. If liquids are spilled onto the machine,
unplug the machine and let the parts dry before plugging it back in.
This device is not intended for life support.
The 365 miniCPAP Auto™ must be set up and adjusted by a trained provider before
being used.
Do not block or otherwise obstruct the exhalation ports of the mask. Follow the
manufacturer’s instructions included with your mask.
Transcend 365 miniCPAP AutoTM User Manual Page 8
This equipment is not suitable for use with oxygen or in the presence of a
flammable anesthetic mixture with air or oxygen, or with nitrous oxide. Sources of
oxygen must be located more than 1 m from the equipment to avoid the risk of fire
and burns.
The 365 miniCPAP Auto™ is only to be used with the supplied or recommended
accessories. Use of accessories not recommended may result in increased
electromagnetic emissions or decreased electromagnetic immunity of the PAP
system and may be potentially unsafe.
The 365 miniCPAP Auto™ is not defibrillation proof.
Do not attempt to sterilize 365 miniCPAP Auto™.
The device is only compatible with appropriately marketed FDA cleared masks and
hoses. A mask should not be used unless the device is turned on and is properly
delivering Ramp or therapy pressure. The proper placement and positioning of the
mask on the face is critical to the consistent operation of this equipment. The
exhalation port(s) associated with the mask should never be blocked. Explanation
of the Warning: The device is intended to be used with masks or connectors
specifically designed to have exhalation ports to allow continuous flow of air out of
the mask. When the device is turned on and functioning properly air flow from the
device flushes the exhaled air out through the mask exhalation port. When the
device is not operating, however, fresh air will not be provided through the mask
and exhaled air may be rebreathed.
Failure to use a mask or accessory that minimizes rebreathing of carbon dioxide or
permits spontaneous breathing can cause asphyxiation.
Strangulation hazard from power cord and air tube. These can become wrapped
around a neck and strangle.
Keep power cord and air tube more than 3 feet from a baby’s crib and out of a
baby’s reach. Keep cord and tube out of children’s reach.
Small parts are unlikely to be expelled from the 365 miniCPAP Auto™ enclosure,
but in case of severe damage internal components may fragment and create a
swallowing or choking hazard if they get out of the enclosure.
Do not position the equipment in bed. Covering breathing tubes with a blanket or
heating them can affect the quality of therapy or injure the user.
To prevent disconnection of the tubing during use, only tubes in compliance with
ISO 5367 or ISO 80601-2-74 should be used.
Do not use the Transcend 365 miniCPAP Auto at an altitude above 8000’ (2438 m)
or outside a temperature of 41-95F (5-35 C). Using the humidifier outside of this
temperature range or above this altitude can affect the quality of the therapy or
injure the user.
Do not add any attachments or accessories to the humidifier or CPAP that are not
approved for use by this instruction. Unapproved accessories may lead to improper
functioning of the device and affect the quality of the therapy or injure the user.
Transcend 365 miniCPAP AutoTM User Manual Page 9
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify they are
operating normally.
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the Transcend 365 miniCPAP Auto including cables specified by the
manufacturer. Otherwise degradation of the performance of this equipment could
result.
Cautions
Federal law (United States) restricts this device to sale by, or on the order of, a
physician.
Power the 365 miniCPAP Auto™ only with the Somnetics-supplied power supplies,
or batteries. See Appendix: Part Numbers.
Discontinue use of the 365 miniCPAP Auto™ and contact your physician if
respiratory or skin irritations occur.
Do not introduce objects into the 365 miniCPAP Auto™ air inlet or air outlet.
Inspect the power supply for signs of wear or damage before each use. Replace the
power cord if necessary.
Operating the humidifier with any substance other than distilled water will damage
the heating element and it will require replacement.
Do not use a nebulizer with the device.
To protect the environment, some parts and accessories of the 365 miniCPAP
Auto™, including optional batteries, must be disposed of in accordance with local
regulations.
The equipment must not be positioned in an environment that can unacceptably
affect the basic safety and performance of the equipment. Examples of this would
include:
o The equipment must not be positioned in a bed
o The equipment should not be placed anywhere other than on a firm, flat
surface
o Do not position in a location where pets or children can access equipment
o Do not position in an open window or other location where dust, sunlight, or
pests (insects) can affect equipment safety and/or performance
Transcend 365 miniCPAP AutoTM User Manual Page 10
o Magnets used to secure the humidifier reservoir to the CPAP may interfere
with other medical devices. Do not place the 365 miniCPAP within 6 inches
(15 cm) of medical devices such as cardiac pacemakers, ICDs, and
implantable neurostimulators.
The equipment must not be covered or positioned in such a way that the operation
or performance of the equipment is adversely affected. Examples of this would
include:
o Do not position next to a curtain that blocks the flow of cooling air, thereby
causing the equipment to overheat
o Do not block the air intake port, thereby interfering with therapy
DBP
Device contains DBP (a phthalate). Significant exposure to DBP
may interfere with the normal development of the male
reproductive tract. Testing has demonstrated that DBP exposure
levels are below established limits. Women who are pregnant or
nursing may wish to discuss the benefits and risks of this device
with a physician.
Transcend 365 miniCPAP AutoTM User Manual Page 11
Symbols
IP22
(the following symbols may appear on the product or packaging)
Caution
Protected against finger-sized objects and against dripping water when tilted 15
degrees from specified orientation
Type BF Applied Part
UL Seal of Approval demonstrating quality, safety and professional manufacturing
of medical product
Temperature limit
Separate collection for electrical and electronic equipment per EC Directive
2002/96/EC. – Waste Electrical and Electronic Equipment (WEEE)
Non-Condensing
Rx Only
Consult instructions for use
Humidity limitation
Prescription only. U.S. federal law restricts this device to sale by or on the order
of a physician or properly licensed practitioner.
Batch code
Date of Manufacture
Precedes reference or item number
Authorized Representative in the European Community
Transcend 365 miniCPAP AutoTM User Manual Page 12
Manufacturer
Fragile, handle with care
Keep dry
Serial Number
Use distilled water only. Use of any liquid other than distilled water will result in
immediate damage to the humidifier and will void heater warranty.
Transcend 365 miniCPAP AutoTM User Manual Page 13
Components of the Transcend® 365
miniCPAP Auto™
Begin by unpacking all items from the 365 miniCPAP Auto™ travel bag and inspecting
them to ensure they were not damaged during shipment. Report any missing or damaged
items to the home healthcare provider that sold the 365 miniCPAP Auto™ to you.
Included with 365 miniCPAP Auto
365 miniCPAP Auto™ device and integrated humidifier
Universal power supply (PSA3) and power cords
Travel bag
Transcend® 365 miniCPAP Auto™ Quick Guide
USB cable
Accessories (sold separately)
Air Supply Tube (Compatible with standard 22 mm connector)
P
Battery™
10
Patient Mask
Transcend 365 miniCPAP AutoTM User Manual Page 14
The 365 miniCPAP Auto™ device
Your 365 miniCPAP Auto™ comes ready to generate and regulate continuous, positive
airway pressure therapy when powered with the included AC power supply or optional
battery.
No allergic reactions to materials used in the equipment have been identified with the
365 miniCPAP Auto™.
Assembling the 365 miniCPAP Auto Device
Your mask connects directly to the 365 miniCPAP Auto™ via the air supply tube.
Follow these steps to assemble the Transcend 365 miniCPAP Auto™:
1. Connect your mask (sold separately) to an air supply tube with a standard 22mm
connector (sold separately).
2. Connect the air supply tube directly to the 365 miniCPAP Auto™ device, making
sure it is fully seated.
3. Plug the power supply into the 365 miniCPAP Auto™. Place device on a flat, stable
surface.
4. Connect the appropriate power cord to the power supply and connect the power
cord to a wall outlet.
Transcend 365 miniCPAP AutoTM User Manual Page 15
Getting Started with the 365 miniCPAP Auto
Allow the 365 miniCPAP Auto to adjust to ambient temperature after removing from
storage. To begin therapy, plug in the power supply to your 365 miniCPAP Auto and
press the Center button to begin therapy. Press the down button to activate Ramp, if
desired.
Note: To begin therapy from the home screen, first ensure that no icons are
highlighted, then press the Center button. You may also start therapy from any screen
by holding the Center power button for three seconds.
Press Center button
to begin therapy.
Then press down
button to begin ramp.
Transcend 365 miniCPAP AutoTM User Manual Page 16
Note:If the miniCPAP™ device loses power while delivering therapy, the CPAP will power
on as soon as power is restored. The user must hit the Center button to resume therapy.
Home screen and Control Panel
Home Screen
Once power is supplied to the device, the LCD display lights up with the Home screen.
The top right white numbers on the Home screen show the set pressure (large number)
and the minimum and maximum pressures (small, lighter numbers - 4 and 20 in this
image). The Home screen display quick menu options: Compliance Report (left), Settings
(top), Last Session Report (right) and Humidification (bottom).
These functions are described in more detail in upcoming sections of the manual.
Control Panel
Use the buttons on the Control Panel below it to navigate through your options on the LCD
screen. Pressing one of the four directional buttons (Up, Down, Left or Right) on the
Control Panel will immediately bring up the section you highlight on the Home screen.
When in a section, these buttons help you navigate to highlight areas and make choices.
The Center button on the Control Panel starts the PAP blower and takes you to the
Therapy Mode screen from the Home Screen. The Center button is used to ‘Select’,
‘Save Settings’ and begin CPAP therapy. Note: When changing settings, be sure to
press the Center button to save your choice.
The Down button takes you to the Humidifier section from the Home Screen. The
Down button is used to toggle to various selections and also activates Ramp once the
CPAP is delivering therapy.
The Right button takes you to the Last Session Report section. It is also used to toggle
to various selections.
The Left button navigates you to your Compliance Report. It also acts as the ‘Back’
button while navigating through screens and will take you to the Home Screen.
The Up button brings you to the Settings screen from the Home Screen. It is also used
to toggle to various selections.
Transcend 365 miniCPAP AutoTM User Manual Page 17
(Settings & Toggle)
Up
Center
(Select, SAVE &
Begin therapy)
Left
(Back & Home
screen)
Right
(Last Session &
Toggle)
Down
(Ramp & Toggle)
Transcend 365 miniCPAP AutoTM User Manual Page 18
Power connection jack and USB port
The power jack accepts the barrel plug of
the output cable from the AC power source
to operate the 365 miniCPAP Auto™.
An optional P
battery pack is also available
10
to power the 365 miniCPAP Auto™.
USB port
A mini-AB USB port is provided for direct
data exchange between the 365 miniCPAP
Auto™ and a computer. The 365 miniCPAP
Auto™ LCD screen interface allows for basic
settings adjustment and data viewing
necessary for the patient. A key sequence
allows clinicians to change therapy settings
via the LCD interface.
The USB connection allows the clinician to
configure the 365 miniCPAP Auto™ for
prescription pressure, Ramp settings and
Auto settings and also access therapy
compliance information with the Transcend
desktop software (password protected).
USB port
Power port
Transcend 365 miniCPAP AutoTM User Manual Page 19
Using the 365 miniCPAP Auto™
Standard user modes
Normal operation consists of four modes:
Off When the device is not connected to a power source, the device is off.
Standby When power is applied to the device the LCD screen lights up and the CPAP
enters Standby Mode. Standby Mode is also initiated by pressing the power
button when the device is in On Mode or if the mask is removed while in On
Mode. As long as power is supplied to the device it will remain in Standby
Mode until On or Drying Mode is initiated.
On When in On Mode, the blower is working and regulated device therapy
pressure is being generated. The LCD lights remain off. On Mode is initiated by
pressing the center button when the device is in Standby Mode and the mask
is worn by the patient.
Drying
Mode
This mode is accessed from Standby mode. Press and hold the Up and Down
buttons at the same time for three seconds to start Drying Mode. This mode
will last 30 minutes unless you interrupt it by pressing any button. Drying
mode flushes your CPAP device, tubing and mask of residual moisture left from
a therapy session.
Starting therapy
Press the Center button on the Control
Panel to start the CPAP blower and go to
the Therapy Mode screen. Press the down
button to activate Ramp, if desired.
The average pressure readout will be
visible. On the bottom of the screen you
will see the set pressure (bottom left) and
the set humidification level (bottom right).
Press the Left/Back button to return to the
Home screen.
Transcend 365 miniCPAP AutoTM User Manual Page 20
Ending therapy and entering Drying Mode
When you are finished with your therapy session, press the Center button to return your
CPAP to Standby mode.
Drying Mode
Press and hold the Up and Down buttons for three seconds to start Drying Mode. This
mode will last 30 minutes unless you interrupt it by pressing any button. After the 30minute drying cycle, the blower will turn off and the device will automatically enter
Standby Mode. Dry mode flushes your CPAP device, tubing and mask of any residual
moisture left from a therapy session.
Push and hold the Up and
Down buttons to enter
Drying Mode.
Transcend 365 miniCPAP AutoTM User Manual Page 21
The 365 miniCPAP Humidifier_________
Filling the Reservoir
1. To remove the reservoir from the CPAP, first disconnect the power source to your
CPAP, then set the device on a flat, stable surface and secure the reservoir with one
hand while you lift and remove the CPAP with your other hand. The pieces are
simply joined by magnets- there is not a button or latch that releases the
humidifier from the CPAP device.
2. Hold the reservoir at a slight angle and fill it with distilled water only. This step is
best performed over a sink in case you spill water while filling the reservoir. (See
image below.)
Caution!Distilled water is required for use with this device because of the
humidifier’s unique heating element. Failure to use distilled water will void component
warranty. Failure to use distilled water will also damage the humidifier and its heating
element, requiring these pieces to be replaced.
Caution! Do not perform this operation near electronic equipment. Always unplug the
power source to your CPAP before attempting to remove the reservoir.