Topcon SL-D301 User Manual

USER MANUAL SLIT LAMP
SL-D301
INTRODUCTION
Thank you for purchasing the SLIT LAMP SL-D301.
INTENDED USE / INDICATIONS FOR USE
FEATURES
This instrument has the following features:
• Various accessories to extend the system
• Ergonomic cable layout
• Clear fluorescent cornea observation and photography of cornea
• Clear eyeground observation and photography by color conversion filter
PURPOSE OF THIS MANUAL
To get the best usage from the instrument, please read DISPLAYS FOR SAFE USE and GENERAL SAFETY INFORMATION. Keep this manual with the instrument for future reference.
1
2
1. No part of this manual may be copied or reprinted, in whole or in part, without prior written permission.
2. The contents of this manual are subject to change without prior notice and without legal obligation.
3. The contents of this manual are correct to the best of our knowledge. Please inform us of any ambiguous or erroneous descriptions, missing information, etc.
4. Original Instructions This manual was originally written in English.
©2016 TOPCON CORPORATION
ALL RIGHTS RESERVED
CONTENTS
INTRODUCTION ......................................................................................................................... 1
DISPLAYS AND SYMBOLS FOR SAFE USE .................................................................... 5
GENERAL SAFETY INFORMATION .................................................................................... 6
HOW TO USE THIS MANUAL ................................................................................................ 8
GENERAL MAINTENANCE INFORMATION ...................................................................... 8
DISCLAIMERS ............................................................................................................................ 8
POSITIONS OF WARNING AND CAUTION INDICATIONS .......................................... 9
SYSTEM DIAGRAM .......................................................................................................................... 10
COMPONENT NAMES .................................................................................................................. 10
COMPOSITION OF PARTS THAT COME IN CONTACT WITH THE PATIENT ........................... 10
STANDARD ACCESSORIES ......................................................................................................... 11
COMPONENTS .................................................................................................................................. 12
COMPONENTS .............................................................................................................................. 12
ASSEMBLY PROCEDURE ............................................................................................................. 13
SECURING THE POWER UNIT .................................................................................................... 13
SECURING THE TABLETOP SO-TABLE06 (OPTIONAL ACCESSORY) ................................... 13
SECURING THE TABLETOP SO-TABLE07 (OPTIONAL ACCESSORY) .................................... 13
SECURING THE CHINREST ......................................................................................................... 14
SECURING THE BASE UNIT AND RAIL COVER ......................................................................... 15
SECURING THE BINOCULAR TUBES ......................................................................................... 15
SECURING THE ILLUMINATION UNIT ......................................................................................... 15
REMOVING THE ILLUMINATION UNIT PAD ................................................................................ 16
CONNECTING THE CABLES ........................................................................................................ 17
FITTING THE CHINREST TISSUE ................................................................................................ 17
FITTING THE CAP ......................................................................................................................... 17
FITTING THE DIFFUSION LENS SO-DFL01 (OPTIONAL ACCESSORY) ................................... 18
SECURING THE TONOMETER MOUNT SO-TM1 (OPTIONAL ACCESSORY) .......................... 19
PREPARATIONS ............................................................................................................................... 20
POWERING ON ............................................................................................................................. 20
ADJUSTING THE DIOPTER AND PUPILLARY DISTANCE (PD) ................................................. 20
OPERATION PROCEDURE ........................................................................................................... 22
FIXING THE PATIENT'S FACE AND FIXATION ........................................................................... 22
OPERATING THE MICROSCOPE UNIT ....................................................................................... 23
OPERATING THE BASE AND FOCUSING ................................................................................... 24
OPERATING THE ILLUMINATION UNIT ...................................................................................... 25
ENDING PROCEDURE ................................................................................................................. 29
MAINTENANCE AND CHECKUPS .............................................................................................. 30
PERIODIC MAINTENANCE ...........................................................................................................30
DAILY CARE .................................................................................................................................. 30
PLACING AN ORDER FOR CONSUMABLES .............................................................................. 30
USER MAINTENANCE ITEMS ...................................................................................................... 30
DEALER MAINTENANCE ITEMS .................................................................................................. 30
REPLACING ILLUMINATION LAMPS ........................................................................................... 31
REPLACING SOCKETS ................................................................................................................ 32
RESTOCKING CHINREST TISSUE .............................................................................................. 32
DAILY CARE .................................................................................................................................. 33
3
CLEANING APPLIED PARTS ........................................................................................................ 33
CLEANING LENSES AND MIRRORS ........................................................................................... 33
TROUBLESHOOTING ................................................................................................................... 34
TROUBLESHOOTING GUIDE ....................................................................................................... 34
SPECIFICATIONS AND PERFORMANCE .................................................................................... 35
GENERAL INFORMATION ON USAGE AND MAINTENANCE ........................................... 36
INTENDED PATIENT POPULATION ............................................................................................. 36
INTENDED USER PROFILE ..........................................................................................................36
ENVIRONMENTAL CONDITIONS FOR USE ................................................................................ 36
STORAGE, USAGE PERIOD ........................................................................................................ 36
ENVIRONMENTAL CONDITIONS FOR PACKAGING IN STORAGE ........................................... 37
ENVIRONMENTAL CONDITIONS FOR PACKAGING IN TRANSPORTAION ............................. 37
ELECTRIC RATING ....................................................................................................................... 37
DIMENSIONS AND WEIGHT ......................................................................................................... 37
SYSTEM CLASSIFICATION ..........................................................................................................38
OPERATION PRINCIPLES ............................................................................................................ 38
CHECKPOINTS FOR MAINTENANCE .......................................................................................... 38
DISPOSAL ..................................................................................................................................... 39
PATIENT’S ENVIRONMENT ......................................................................................................... 40
REQUIREMENTS FOR THE EXTERNAL DEVICE ....................................................................... 40
ELECTROMAGNETIC COMPATIBILITY ....................................................................................... 41
OPTICAL RADIATION HAZARD .................................................................................................... 45
RELATIVE SPECTRAL DISTRIBUTION OF ILLUMINATION LIGHT ............................................ 45
TYPE OF PLUG ............................................................................................................................. 46
OPTIONAL ACCESSORIES ........................................................................................................... 47
SYSTEM CONFIGURATION ......................................................................................................... 47
DIGITAL CAMERA UNIT DC-4 ...................................................................................................... 48
BACKGROUND ILLUMINATION BG-2GN ..................................................................................... 48
YELLOW FILTER UNIT .................................................................................................................. 48
IRIS DIAPHRAGM SO-DF01 .........................................................................................................48
12.5X EYEPIECE ........................................................................................................................... 48
20X EYEPIECE .............................................................................................................................. 48
TONOMETER MOUNT SO-TM1 .................................................................................................... 49
HRUBY LENS ................................................................................................................................ 49
PARALLEL BINOCULAR TUBE PB-2 ............................................................................................ 49
ADAPT COVER SO-AC7 ............................................................................................................... 49
DIFFUSION LENS SO-DFL01 ....................................................................................................... 49
TABLETOP SO-TABLE06, 07 ........................................................................................................50
4
DISPLAYS AND SYMBOLS FOR SAFE USE
To encourage safe and proper use and to prevent injury to the operator and others or potential damage to property, important messages are put on the instrument body and inserted in this manual. We suggest that everyone understand the meaning of the following displays, icons and text before read­ing the GENERAL SAFETY INFORMATION and observe all listed instructions.
DISPLAY
Display Meaning
Situations in which the device should not be used because the
CONTRAINDICATION
WARNING
CAUTION
NOTE
risk of use clearly outweighs any possible benefit.
Incorrect handling by ignoring this display may lead to a risk of death or serious injury.
Incorrect handling by ignoring this display may lead to personal injury or physical damage.
Useful functions to know. Paying attention to these will prevent the noted problems.
SYMBOL
Symbol IEC/ISO Publication Description Description (French)
IEC 60417-5032 Alternating Current Courant alternatif
IEC 60878-02-02 Type B applied part Partie appliquée du Type B
ISO 7010-W001 General warning sign
ISO 7010-M002
ISO 7000-2497 Date of manufacture Date de fabrication
ISO 7000-2498 Serial number Numéro de série
ISO 7000-3082
ISO 15223-1
Refer to instruction manual/ booklet
Manufacturer Fabricant
Authorised Representative in the European Community
Symbole d'avertissement général
Voir le manuel/la brochure
Représentant autorité pour l’Union européenne
IEC 60417-5264
IEC 60417-5265
“On” (only for a part of EQUIP­MENT)
“Off” (only for a part of EQUIP­MENT)
Mise en service d'une partie d'appareil
Mise hors service d'une partie d'appareil
5
GENERAL SAFETY INFORMATION
CONTRAINDICATION
This instrument must not be used for the following patients:
• Patients who are hypersensitive to light.
• Patients who recently underwent photodynamic therapy (PDT).
• Patients taking medication that causes photosensitivity.
WARNING
Ensuring the Safety of Patients and Operators
Use this instrument carefully on the following patients.
• Patients who have epidemic corneitis, conjunctivitis or any other infectious disease.
• Patients who are taking medications that cause light hypersensitivity.
To avoid injury to the patient’s eye and nose, pay particular attention while operating the instrument body. (The patient may be injured.)
The Topcon SL-D301 is a medical device. The software and hardware has been designed in accor­dance with U.S., European and other international medical device design and manufacturing stan­dards. Unauthorized modification of the Topcon SL-D301 software or hardware, or any addition or deletion of any application in any way can jeopardize the safety of operators and patients, the perfor­mance of the instrument, and the integrity of patient data.
Because prolonged intense light exposure can damage the retina, the use of the device for ocular examination should not be unnecessarily prolonged, and the brightness setting should not exceed what is needed to provide clear visualization of the target structures. The retinal exposure dose for a photochemical hazard is a product of the radiance and the exposure
time. If the value of radiance were reduced in half, twice the time would be needed to reach the maxi­mum exposure limit.
While no acute optical radiation hazards have been identified for direct or indirect ophthalmoscopes, it is recommended that the intensity of light directed into the patient's eye be limited to the minimum level which is necessary for diagnosis. Infants, aphakes and persons with diseased eyes will be at greater risk. The risk may also be increased if the person being examined has had any exposure to the same instrument or any other ophthalmic instrument using a visible light source during the previous 24 hours. This will apply particularly if the eye has been exposed to retinal photography.
Equipment is not suitable for use in the presence of a Flammable Anesthetic Mixture with Air, Oxygen, or Nitrous Oxide.
The Topcon SL-D301 has no special protection against harmful ingress of water or other liquids (clas­sified IPX0). To avoid damage to the instrument and cause a safety hazard, the cleaning solutions, including water, should not be directly applied to the device. Using a dampened cloth (without drip­ping), is a good method to clean the exterior surface of the enclosure.
Handling the cord on this product or cords associated with accessories sold with this product, will expose you to lead, a chemical known to the State of California to cause birth detects or other reproductive harm.
Wash hands after handling.
6
Preventing Electric Shock and Fire.
To avoid fire in the event of an instrument malfunction, immediately turn off the power switch and remove the power plug from the outlet if you see smoke coming from the instrument or if you detect other problems. Don't install the instrument where it is difficult to disconnect the power plug from the outlet. Ask your dealer for repairs.
CAUTION
Ensuring the Safety of Patients and Operators
Be careful not to let the patient touch this instrument. The patient's hand may be pinched by a movable part.
To avoid burns caused by heat, do not replace the lamp with a new one immediately after it goes off.
To avoid injury to the patient's head, incline the illumination unit slowly while holding the base unit.
When operating the base unit, please note the following:
•Beware of catching fingers in the moving parts.
•Avoid hitting the patient's eyes or nose.
Preventing Electric Shock and Burn
To avoid injury or fire caused by electric shock, turn off the power switch and unplug the power cord when not in use.
To avoid injury caused by electric shock, turn off the power switch when replacing the lamp.
Electromagnetic Compatibility (EMC)
This instrument has been tested (with 100/120/230V) and found to 2007as class B(classified according to CISPR11). This instrument radiates radio frequency energy within standard and may affect other devices in the vicinity. If you have discovered that turning on/off the instrument affects other devices, we recommend you change its position, keep a proper distance from other devices, or plug it into a different outlet. Please consult your authorized dealer if you have any additional questions.
comply with IEC60601-1-2 Ed.3.0:
7
HOW TO USE THIS MANUAL
• Read the instructions on pages 1 to 9 before using the machine.
• If you would like an overview of the system, begin by reading OPERATION PROCEDURE (page 22).
GENERAL MAINTENANCE INFORMATION
USER MAINTENANCE
To maintain the safety and performance of the instrument, unless done by an authorized service engi­neer, never attempt to do maintenance of items other than those specified here in. For details about maintenance, read the description of this manual.
Replacing the Illumination Lamp
The illumination lamp can be replaced if necessary. For specific instructions, see page 30.
DISCLAIMERS
TOPCON is not responsible for damage due to fire, earthquakes, actions or inactions of third persons or other accidents, or damage due to negligence and misuse by the user and any use under unusual conditions.
TOPCON is not responsible for damage derived from inability to properly use this equipment, such as loss of business profits and suspension of business.
TOPCON is not responsible for damage caused by operations other than those described in this user manual.
The device does not provide a diagnose of any condition or lack thereof or any recommendation for appropriate treatment. The relevant healthcare provider is fully responsible for all diagnose and treatment decisions and recommendations.
8
POSITIONS OF WARNING AND CAUTION INDICATIONS
2
1
3
4
5
To ensure safety, warning labels are provided on the instrument body. Use the instrument following these warning instructions. If any of the following labels are missing, con­tact your dealer or TOPCON (see the back cover).Fabricant
No. Label Meaning Signification
CAUTION
To prevent electric shocks, switch off
the power supply and remove the
1
power cord before replacing the lamp.
Do not replace the lamp immediately
after switching it off: the high temper­atures could cause burns.
• Afin d’éviter les chocs électriques, coupez l’alimentation électrique et débranchez le câble d’alimentation avant de remplacer la lampe.
• Ne pas remplacer la lampe immédiatement après l’avoir éteinte: la température élevée peut provoquer des brûlures.
CAUTION
2
To avoid injury to the patient’s head, incline the illumination unit slowly while holding the base unit.
Afin d’éviter de blesser le patient à la tête, inclinez lentement l’élément lumineux tout en maintenant la base de l’appareil.
CAUTION
When operating the base unit, please
3
4
5
note the following:
Beware of catching fingers in the
 moving parts.
Avoid hitting the patient’s eyes or
nose.
Degree of protection against electric shock : TYPE B APPLIED PART
AC adapter Model :BPM050S18F03 Manufacturer :Bridge Power Corporation
Lorsque vous maniez la base de l’appareil, veuillez noter les points suivants:
• Faites attention à ne pas vous coincer les doigts dans les parties en mouvement.
• Évitez de heurter les yeux ou le nez du patient.
Degré de protection contre les chocs électriques : TYPE B PARTIE D'APPLICATION
Adaptateur secteur Modèle :BPM050S18F03
Fabricant :Bridge Power Corporation
PRÉCAUTION
PRÉCAUTION
PRÉCAUTION
9

SYSTEM DIAGRAM

Plug Lamp house cover
Aperture/slit length display window
Filter selector lever Aperture/slit-length
selector knob
Canthus marker
Mirror
Centering knob Inclination lever Slit width controls
Photography button Control lever
Base locking knob Base
Forehead rest
Chinrest
Objective lens
Magnification selector handle
Fixing knob
Diopter adjusting ring
12.5× eyepiece
Magnification index mark
Chinrest adjuster
Microscope arm
Illumination arm locking knob
Power switch
Microscope arm locking knob
Illumination Unit
Tabletop
Power Supply
Microscope Unit
Base Unit
Chinrest Unit
Bightness adjustment knob
Fixation target
Tabletop*
Illumination arm

COMPONENT NAMES

COMPOSITION OF PARTS THAT COME IN CONTACT WITH THE PATIENT

* Tabetop is opptional accessory.
Forehead rest: Polyamide resin Chinrest : Polyamide resin
10
SYSTEM DIAGRAM

STANDARD ACCESSORIES

D301
Make sure that all the following standard accessories are included. Figures in parentheses are the quantities.
Chinrest tissue (1) Dust cover (1) Test rod (1)
Crosshead screwdriver (1) Screwdriver (1) AC adapter (1)
(model:BPM050S18F03)
Chinrest tissue pin (2) Square mirror (1)
Luminous fixation target (1) Cap (1) Cleaning brush (1)
Illumination lamp (1) Socket (1) User manual (1),
Accessory case (1)
Instruction manual (1)
Power cord (1) *
* More than one power cord can be included on certain occasions.
For optional accessories, see “Optional Accessories” on page 47.
11
SYSTEM DIAGRAM

COMPONENTS

COMPONENTS

(1) Illumination unit (2) Binocular tubes (3) Base unit
(4) Power Supply (5) Rail cover (6) Chinrest unit
(7) Rail cover fixing screw
Article name Qty Article name Qty
(1) Illumination unit 1 (5)Rail cover 2
(2) Binocular tubes 1 (6)Chinrest unit 1
(3) Base unit 1 (7) Rail cover fixing screw 4
(4) Power Supply 1
12
COMPONENTS

ASSEMBLY PROCEDURE

* The illustration depicts the SO-TABLE06.

SECURING THE POWER UNIT

• The tabletop SO-TABLE06 and SO-TABLE07 is optional accessory.
1 Affix the power supply to the back of the tabletop with 4 screws.

SECURING THE TABLETOP SO-TABLE06 (OPTIONAL ACCESSORY)

CAUTION
FITTING TO AUTOMATIC INSTRUMENT TABLE AIT-16
To prevent falling during use and movement, secure each unit.
1 Place the tabletop on the instrument table, and secure it with the 4 bolts attached to the instru-
ment table.

SECURING THE TABLETOP SO-TABLE07 (OPTIONAL ACCESSORY)

1 Peel off the tape which secures the plastic washer to the mounting bracket’s shaft.
ASSEMBLY PROCEDURE
13
2 Insert the plastic washer, together with the shaft, into the cavity for the ophthalmic unit arm.
Washer
Shaft

SECURING THE CHINREST

1 Take out the chinrest from the upper section of the container box.
2 Attach the chinrest to the bottom face of the table, and affix it to the table using 4 screws.
3 Chinrest fitted to the tabletop.
14
ASSEMBLY PROCEDURE

SECURING THE BASE UNIT AND RAIL COVER

Rail cover
Fixing knob
Microscope unit
30° - 60°
Indices
1 Align the wheel of the base unit with the rail of the tabletop. 2 Secure the rail covers with 4 screws (7): (2 screws each on the right and left sides).

SECURING THE BINOCULAR TUBES

1 Align the pin of the microscope unit with the groove on the binocular tubes, and fit the binocular
tubes with the fixing knob.
NOTE
Make sure you do not touch the lens surfaces.

SECURING THE ILLUMINATION UNIT

Loosen the microscope arm locking-knob of the base unit, manually turn the shaft and tilt the guide rod­shaft index 30-60°, then refasten the microscope-arm locking-knob.
15
ASSEMBLY PROCEDURE
2 Loosen the fixing screw on the outside of the fitting cavity of the illumination unit with a screw-
Indices
Protection pad (for transportation)
driver. Align indices and slowly lower the illumination unit onto the shaft of the base unit.
NOTE
While assembling the illumination unit, take care not to get your fingers caught.
3 Firmly tighten the fixing screw with a screwdriver.

REMOVING THE ILLUMINATION UNIT PAD

1 Remove the rubber band and slowly withdraw the protection pad from the slit operation mecha-
nism of the illumination unit.
16
ASSEMBLY PROCEDURE

CONNECTING THE CABLES

Plug
Illumination cable
power cord
Fixation target cable
• Use the power cord applicable to the voltage of the commercial power supply being used. Using a power cord improper to the commercial
CAUTION
power supplymay cause the instrument to malfunction.
• The power cable in standard accessories for this instrument cannot use besides this instrument.
1 Remove the tape from the lamp house cover of the illumination unit. Plug the cable from the
upper part of the chinrest into the illumination unit.
2 Connect the cable from the lower part of the chinrest unit and the power cord to the power supply.
3 Pull the base assembly toward the operator's side fully. Attach the cables to the rear side of the
table with the cable clip to adjust the loosening of the cord.
4 Move the base assembly and illumination assembly and make sure there is enough cable to
allow free movement of the base assembly in all directions.

FITTING THE CHINREST TISSUE

1 Remove the chinrest tissue pins. 2 Take approximately one-fifth of the pad of chinrest tissues and secure this at each end with the pins.

FITTING THE CAP

1 Fit the cap to the shaft aligning the guide rod with the groove in the cap.
17
ASSEMBLY PROCEDURE

FITTING THE DIFFUSION LENS SO-DFL01 (OPTIONAL ACCESSORY)

Mirror stopper
Frame
Hole
Pivot
DIFFUSION LENS SO-DFL01
Flat side
Concave side
Concave side
Hole
Pivot
1 Insert one pivot of frame of SO-DFL01 into the hole of mirror stopper.
Be careful not to make mistake of the flat side / concave side of SO-DFL01.
NOTE
Set the concave side as shown below.
2 Insert another pivot into the opposite hole of mirror stopper.
• Be careful not to extend the frame excessively. The frame may be deformed and, if so, it cannot be attached.
• Make sure you do not touch the mirror surface,If you touch the mirror surface, please clean this according to the process on page 33 Cleaning Lenses and Mirrors.
NOTE
18
ASSEMBLY PROCEDURE

SECURING THE TONOMETER MOUNT SO-TM1 (OPTIONAL ACCESSORY)

Screw
Locating Pin
The holes
Screw
Fixing knob
1 Align the locating pin of SO-TM1 into the holes of the microscope, and fasten the screw.
2 Remove the fixing knob of the microscope, and secure the eyepieceunit, etc with the packaged
screw.
3 Applanation tonometer R900 type, Photokeratoscope attachment, etc could be mounted on SO-
TM1.
19
ASSEMBLY PROCEDURE

PREPARATIONS

Tes t ro d
Magnification selector handle
Slit adjustment knob
Aperture/slit-length selector knob
Eyepiece

POWERING ON

To avoid fire and electric shock in case of leakage, be sure to use a
WARNING
CAUTION
1 Connect the power cord. 2 Turn ON the POWER switch.

ADJUSTING THE DIOPTER AND PUPILLARY DISTANCE (PD)

grounded outlet. Do not connect to outlets that are not grounded.
• Use the power cord applicable to the voltage of the commercial power supply being used. Using a power cord improper to the commercial power supplymay cause the instrument to malfunction.
• The power cable in standard accessories for this instrument cannot use besides this instrument.
NOTE
In case that no test rod is provided, set the diopter scale to your diopter by turning the diopter adjust­ment ring.
To ensure sharp observation of slit images, always carry out the diopter and PD adjustments.
1 Insert the test rod into the rotation shaft cavity, and set the black face square with the micro-
scope.
2 Place the brightness adjustment knob (see page 25) in at the 2/3. 3 Adjust the illumination to φ9mm by adjusting the slit width control knob and aperture/slit length
control knob.
4 Turn the diopter adjusting ring of the eyepiece at one side fully counter-clockwise. 5 Turn the diopter adjusting ring clockwise and stop when the test rod can be clearly seen. 6 Adjust the diopter of another eyepiece in the same way as above.
20
PREPARATIONS
7 After adjusting the diopter, turn the slit width control knob until the slit width is about 1mm, then
Diopter adjusting ring
Prism box
check if the slit image projected on the test rod is properly in focus.
8 Holding the prism box, look through the eyepiece with both eyes, and adjust the pupillary dis-
tance so that the image projected on the test rod can be seen without diplopia (double vision), and appear to be three dimensional.
21
PREPARATIONS

OPERATION PROCEDURE

Forehead rest
Canthus marker
Chinrest
Chinrest adjuster
(A) Fixation target with
diopter adjustment
(B) Luminous fixation target
Diopter adjusting ring
Fixation target unit
Fixation target lever

FIXING THE PATIENT'S FACE AND FIXATION

1 Place the patient's chin on the chinrest with his forehead against the forehead rest. 2 By rotating the chinrest adjuster, align the patient's eye with the canthus marker on the chinrest
frame.
3 Ask the patient to look at the fixation target with the eye that is not being examined.To change the
patient's fixation point, hold the fixation target at the end opposite to the target and adjust accord­ingly.
• When using the fixation target with diopter adjustment (A), slide the diopter adjustment knob so that the patient can see the following target ( ).
NOTE
• The ring target can be adjusted within a range of -15D to +10D.
• The luminous fixation target is used for myopia of -15D or more.
• When replacing targets, remove the target by pulling gently whilst supporting the opposite end.
22
OPERATION PROCEDURE

OPERATING THE MICROSCOPE UNIT

Magnification selector handle
Magnification index mark
Turn the magnification selector to set a magnification value against the magnification index mark.
For the overall magnification in conjunction with magnification marks of the
NOTE
magnification selector handle, see page 35.
23
OPERATION PROCEDURE

OPERATING THE BASE AND FOCUSING

Base locking knob
Base
Control lever
• To avoid injury to the eye and nose whilst moving the base unit, make sure that you have a clear view of the slit lamp and the patient's face.
CAUTION
• To prevent dropping the base locking knob from the base, do not loosen the
NOTE
• Rough focusing is carried out with major movements, following step 1-3.
• Fine focusing is done with the microscope, following steps 2 and 3.
• For the safety of the operator and the patient, do not place fingers between moving parts.
knob too much.
1 For major horizontal movements, hold the control lever in the upright position and move the
entire base.
2 For fine adjustments, move the control lever in the required direction. 3 The base can be raised by turning the control lever clockwise, and lowered by turning the control
lever counter-clockwise.
4 To fix the base, fasten the base locking knob.
24
OPERATION PROCEDURE

OPERATING THE ILLUMINATION UNIT

Brightness adjustment knob
Slit-width control knob
• To avoid injury to the patient’s head, incline the illumination unit hold­ing the base unit.
CAUTION
• Adjust the slit width according to the results of the investigation.
NOTE
ADJUSTING THE BRIGHTNESS
Turn the brightness adjustment knob. The brightness of the illumination light can be adjusted to the preferred illumination setting.
• The slit-width scale should be used as a guideline.
• When using the square mirror, incline the illumination unit at least 10º.
• To avoid causing discomfort to the patient or any damage to the patient's eye, keep the illumination at its minimum during adjustment.
The surrounding number of the brightness adjustment
NOTE
knob indicate the ratio of brightness, where 1 is the maximum value.
ADJUSTING THE SLIT WIDTH
Turn the slit-width control knob. The slit width can be changed gradually between 0 and 9mm (9mm=circle).
25
OPERATION PROCEDURE
CHANGING THE APERTURE/SLIT LENGTH
Aperture/slit-length display window
Aperture/slit-length control knob
Aperture/slit-length control knob
Slit-angle scale
Centering knob
Turn the aperture/slit-length control knob. When the slit is fully opened, 7 types of spot illumination (φ9, φ8, φ5, φ3, φ2, φ1, φ0.2mm) are availa- ble. The slit width can be changed gradually from 1mm to 8mm.
NOTE
The spot illumination size and slit length are displayed on the aperture/slit­length display window.
TURNING THE SLIT
Horizontally rotate the aperture/slit-length control knob. This directly changes the slit image from vertical to horizontal. In this mode, the slit angle can be read off the angle scale.
SWINGING THE SLIT SIDEWAYS
Loosen the centering knob and swing the illumination unit right and left. This provides indirect illumination displacing the slit light from the microscope center. By fastening the centering knob, the slit light returns to the center of the vision field.
NOTE
26
OPERATION PROCEDURE
This function is used for scanning observation and observation with indirect illu­mination.
INCLINED ILLUMINATION
Inclination lever
Square mirrorBattledore mirror
Press to unlock the inclination lever and pull. The illumination unit is inclined for inclined illumination up to 20° in 5° steps.
NOTE
This function is used for observing a horizontal cross section, and for corner angle and fundus observation.
REFLECTION MIRROR
For this instrument, a battledore mirror and a square mirror are available. For normal observation, the battledore mirror is used. However, if the arm angle scale, which represents the angle formed by the illumination arm and microscope arm, reads approx. 3° to 10° and the observation light flux is disturbed by the battledore mirror, then the square mirror should be used. The square mirror is to be used when the arm angle is opened to more than 10°
.
NOTE
The square mirror is standard accessory.
27
OPERATION PROCEDURE
REPLACING REFLECTION MIRRORS
Square mirrorBattledore mirror
Green-colored
No filter
ND filter
(13transmission)
Blue filter
Red-free filter
Blue-colored
Filter selector lever
Replace mirrors as follows, taking care not to touch the mirror and lens surfaces:
1 Open the microscope arm and illumination arm 30° or more. 2 Incline the illumination unit 10° or more. 3 Pull out the battledore mirror, holding the slender part on both sides. To reinsert the battledore
mirror, hold the slender part on both sides and insert.
4 Insert the square mirror from the side recessed on the back. 5 To pull out the square mirror, which has no handle, push the square mirror up as illustrated
below.
NOTE
CHANGING FILTERS
Move the filter selector lever right and left to select the required filter from the 4 types.
If you touch the mirror or lens surface, please clean this according to the proc­ess on page 33 "Cleaning Lenses and Mirrors".
28
OPERATION PROCEDURE
DIFFUSION LENS SO-DF 01 (OPTIONAL ACCESSORY)
Diffusion lens
When in use, the diffusion lens is set vertically in front of the reflection mirror. When not in use, remove the diffusion lens from the light path.
When using the diffusion lens, open the microscope arm and illumination arm
NOTE
30°to avoid friction of the diffusion lens or illumination support. Also, fully open the slit; otherwise this prevents sufficient light from entering.

ENDING PROCEDURE

Turn OFF the Power switch.
29
OPERATION PROCEDURE

MAINTENANCE AND CHECKUPS

PERIODIC MAINTENANCE

Before using, confirm the following:
•Adjust the diopter and eye width following ADJUSTING THE DIOPTER AND PUPILLARY DIS­TANCE (PD) on page 20 and turn the Slit width controls and make the slit width about 1mm: The slit image projected on the test rod is seen clearly.
•Move the base forward-backward and right-left: The base moves smoothly.
•Component parts, including the eyepiece unit, are fitted in place.
•The chinrest base is firmly fitted to the table.
•Cables and plugs are firmly connected.
•The slit width doesn't narrow by disengaging a hand from the Slit width control knob.
•The illumination unit smoothly swings.

Daily Care

This instrument may be affected adversely by dust. Apply the dust cover when not using.

PLACING AN ORDER FOR CONSUMABLES

When ordering consumable items, contact your dealer or TOPCON (see the back cover). Specify the article name, product code and quantity.
Article name Product code
Light source
Chinrest tissue 403104082
Halogen lamp 447032530 Socket 405261249

USER MAINTENANCE ITEMS

Item Inspection time Contents
Inspection Before using • Adjusting the diopter and pupillary distance
• Focus of slit image
• The base unit must move smoothly.
• The components must be fitted in place correctly.
• The chinrest unit must be fitted to the table unit correctly.
• The cables and plugs must be connected correctly.
• The objective lens, eyepiece and mirror must not be stained or damaged.
Cleaning When the part is
stained
Replacement As required • Halogen lamp
Supply As required • Chinrest tissue
• Objective lens
• Eyepiece
•Mirror
• Sliding plate, rail and wheel shaft unit
• Forehead rest and chinrest unit
• Socket

DEALER MAINTENANCE ITEMS

Item Inspection time Contents
Adjustment
Within 12 months from the
last maintenance
• Slit width control knob torque
• Inclination torque of illumination unit
30
MAINTENANCE AND CHECKUPS

REPLACING ILLUMINATION LAMPS

Lamp house cover
Plug
Socket fixing
Lamp unit
lever
L
Pin
• When replacing the lamp, switch off the power supply and remove the
CAUTION
• To ensure perfect illumination, make sure that the socket flange and notch are
NOTE
• Use a soft cloth and do not touch the illumination lamp with bare fingers: fin-
firmly fitted to the lamp house.
gerprints and stains may affect illumination and cause premature failure of the lamp.
power cord to avoid electric shock.
• Beware of high temperatures when replacing the lamp immediately after switching it off: these could cause burns.
1 Turn OFF ( ) the POWER switch and remove the cable
plug. Pull out the plug from the lamp house cover. Turn the lamp house cover counterclockwise and remove upward.
2 Lightly pull the socket fixing lever and turn in the direction indicated
by the arrow.
3 Pull out the lamp unit. 4 Remove the illumination lamp from the socket and fit the new lamp.
When fitting the new lamp, make sure the direction of the illumination lamp and socket is correct. For the article name and product code of the illumination lamp,
refer to PLACING AN ORDER FOR CONSUMABLES on P.30.
Check the marked "L" on the illumination lamp. If "L" is not marked, it can not use.
NOTE
5 Fit the pin into the groove on the lamp house cover as shown below. Turn the lamp house cover clockwise
to fix it.
6 Connect the plug.
NOTE
Unless the pin is fitted into the groove correctly, the plug is not connected prop­erly and the illumination is not turned on.
31
MAINTENANCE AND CHECKUPS

REPLACING SOCKETS

Cable-fixing terminal
NOTE
The socket may deteriorate due to the constant heat: therefore, it should be replaced after the lamps have been changed two or three times.
1 Remove the lamp following steps 1 - 4 of REPLACING ILLUMINATION LAMPS. 2 Loosen the cable-fixing terminal by turning it counterclockwise, remove the cable and replace the
socket with a new one. *For the article name and product code of the socket, refer to PLACING AN ORDER FOR CON­SUMABLES on P.
30.
3 Turn the cable-fixing terminal clockwise to fix the cable securely.
NOTE
Unless the cable-fixing terminal is fixed securely, the illumination may not be turned on.
4 Install the lamp house cover and plug following steps 5 - 6 of REPLACING ILLUMINATION
LAMPS.

RESTOCKING CHINREST TISSUE

When the chinrest tissue supply is depleted, pull out the chinrest tissue pins and replace tissue.
32
MAINTENANCE AND CHECKUPS

DAILY CARE

• Before carrying out daily care, remove the power cord (to avoid elec-
CAUTION
• To prevent the chinrest, forehead rest and other plastic parts from discolora-
NOTE
• Wipe parts with a cloth moistened with a tepid solution of neutral kitchen detergent.
tion and deterioration, do not use volatile solvents for cleaning, including ben­zine, thinner, ether, gasoline, etc.
tric shocks) and wait until the lamp house has cooled (to avoid burns).
• Do not touch parts inside the lamp house cover during operation and immediately after switching off the power supply: this could cause burns.

CLEANING APPLIED PARTS

Wipe the forehead rest, the chinrest with a cloth moistened with a tepid solution of neutral detergent for kitchenware.

CLEANING LENSES AND MIRRORS

REMOVING STAINS
NOTE
To prevent damaging lens surfaces, do not hold gauze with tweezers.
1 Prepare a solution of ethyl alcohol 20% and ether 80%. 2 Remove dust from lens and mirror surfaces with the cleaning brush, or a blower. 3 Using clean gauze or lint-free tissue, lightly clean with a rotating movement from the center of the
lens/mirror outwards.
4 If the stain remains, repeat this 2 to 3 times. 5 If stains are persistent, call your dealer or TOPCON (see the back cover).
33
MAINTENANCE AND CHECKUPS

TROUBLESHOOTING

TROUBLESHOOTING GUIDE

To avoid electric shocks, do not attempt overhauling, rebuilding or
CAUTION
If you suspect a problem, check the possible cause by means of the check list below. If the check list below does not solve the problem, or if the problem is not included in the list, contact your dealer or TOPCON (see the back cover).
Problem Possible cause Check Page
Plug of lamp house cover is switched off Insert plug. 17 Cable connection is disconnected Check cable connection. 20 POWER switch is OFF Turn ON POWER switch. 20
Illumination lamp does not work
Illumination field is not uniform/is shady/ is dark
Brightness adjustment knob is the mini­mum
Illumination lamp is broken Replace it with a new illumination lamp. 30 Socket has deteriorated Replace it with a new socket. 32
Slit width is the minimum Rotate the Slit width controls
Filter selector lever is out of position Click filter selector lever. 28
repairs. Ask your dealer for repair.
Check List
Turn up brightness adjustment knob. 25
25
34
MAINTENANCE AND CHECKUPS

SPECIFICATIONS AND PERFORMANCE

Specifications and Performance
Microscope unit
Type Galileo type Magnification Drum, 3-step magnification Magnification steps 10/16/25
10.00 (φ22.5mm)
Overall magnification (actual vision field)
Eyepiece lens
PD adjustment range
Illumination unit
Illumination field
Aperture diameter φ9, 8, 5, 3, 2, 1, 0.2mm Slit direction Vertical to horizontal, can be altered gradually
Inclination
Filter
Illumination lamp Halogen type: 6V 20W
Base unit
Forward-backward movement Right-left movement length 100mm Vertical movement length 30mm Fine adjustment in all directions length 12mm
length 90mm
Chinrest unit
Adjustable range of the vertical direction 80mm
Fixation target
15.98 (φ14.1mm)
25.53 (φ8.8mm) Magnification: 12.5x
Diopter adjustment range: -5D to +5D 55 to 78mm
Slit width: 0 to 9mm, can be altered gradually (9mm=circle) Slit length: 1 to 8mm, can be altered gradually
5°, 10°, 15,° 20° from below
Available for use side swing
Blue filter, red-free filter, ND filter (13% transmission), UV cut filter (normal use), IR cut filter (normal use)
Fixation target with diopter adjustment Light source for fixation target: LED (red)
• The specification and design of the product can be altered for improvements without prior notice.
The following statement is the Essential performance provided for by IEC60601-1.
NOTE
The halogen lamp is not to be turned off. There are no component failures.
MAINTENANCE AND CHECKUPS
35

GENERAL INFORMATION ON USAGE AND MAINTENANCE

INTENDED PATIENT POPULATION

The patient who undergoes an examination by this instrument must maintain concentration for a few minutes and keep to the following instructions:
To fix the face to the chinrest, forehead rest. To keep the eye open. To understand and follow instructions when undergoing an examination.

INTENDED USER PROFILE

The SL-D301 SLIT LAMP is an electric instrument for medical use. Use this instrument under a doctor's guidance.

ENVIRONMENTAL CONDITIONS FOR USE

Temperature : 10°C to 40°C Humidity : 30% to 90% (non-condensing) Air pressure : 700hPa to 1060hPa

STORAGE, USAGE PERIOD

1. Environmental conditions (without package) * Temperature : 10°C to 40°C Humidity : 10% to 95% (without dew condensation) Air pressure : 700hPa to 1060hPa
*THIS INSTRUMENT DOES NOT MEET THE TEMPERATURE REQUIREMENTS OF ISO 15004-1 FOR STORAGE. DO NOT STORE THIS INSTRUMENT IN CONDITIONS WHERE THE TEMPER­ATURE MAY RISE ABOVE 40°C OR FALL BELOW 10°C.
2. When storing the instrument, ensure that the following conditions are met: (1) The instrument must not be splashed with water. (2) Do not store the instrument in an environment where air pressure, temperature, humidity, venti-
lation, sunlight, dust, salty/sulfurous air, etc. could cause damage.
(3) Do not store or transport the instrument on a slanted or uneven surface or in an area where it is
subject to vibrations or instability.
(4) Do not store the instrument where chemicals are stored or gas is generated.
3. Normal life span of the instrument: 8 years from delivery providing regular maintenance is performed (according to the self-certification [TOPCON data])
36
GENERAL INFORMATION ON USAGE AND MAINTENANCE

ENVIRONMENTAL CONDITIONS FOR PACKAGING IN STORAGE

Temperature : -20°C to 50°C Humidity : 10% to 95% Air pressure : 700hPa to 1060hPa

ENVIRONMENTAL CONDITIONS FOR PACKAGING IN TRANSPORTAION

Temperature : -40°C to 70°C Humidity : 10% to 95% Air pressure : 700hPa to 1060hPa

ELECTRIC RATING

Source voltage : AC100 - 240V Frequency : 50 - 60Hz Power input : 80VA

DIMENSIONS AND WEIGHT

Dimensions, Weight
Dimensions: w/o Table
w/ Unit Table 440mm(W) x 410mm(D) x 760 to 790mm(H) w/ AIT Table 550mm(W) x 430mm(D) x 760 to 790mm(H) Weight: w/o Table 12kg w/ Unit Table 17kg
w/ AIT Table Height from the top to patient’s eye 375mm
312mm(W) x 296mm(D) x 646 to 676mm(H)
17.5kg
37
GENERAL INFORMATION ON USAGE AND MAINTENANCE

SYSTEM CLASSIFICATION

• Type of protection against electric shocks: Class Ι equipment Class Ι equipment does not depend on basic insulation only for protection against electric shocks. It can also be earthed; therefore, the metal parts with which one comes into contact do not become conduc­tive if the basic insulation fails.
• Degree of protection against electric shocks: Type B applied part Type B applied part is the applied part complying with the specified requirements of the Standard IEC 60601-1 to provide protection against electric shock, particularly regarding allowable LEAKAGE CUR­RENT.
• Degree of protection against harmful ingress of water: IPx0 SL-D301 has no protection against ingress of water. (The degree of protection against harmful ingress of water defined in IEC 60529 is IPx0)
• Classification according to the methods of sterilization or disinfection recommended by the manufac­ture: not applicable. SL-D301 has no part to be sterilized or be disinfected.
• Not AP or APG equipment
• Classification according to the degree of safety of application in the presence of a flammable anaes­thetic mixture with air or with oxygen or nitrous oxide: Equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide. SL-D301 should be used in environments where no flammable anesthetics and/or flammable gases are presents.
• Classification according to the mode of operation: Continuous operation. Continuos operation is the operation under normal load for an unlimited period, without the specified limits of temperature being exceeded.

OPERATION PRINCIPLES

Illuminates the observed part by the illumination light emitted from the illumination optical system and allows enlargement observation by binocular stereoscopic microscope.

CHECKPOINTS FOR MAINTENANCE

1. Regularly maintain and check the instrument and its parts.
2. When using the instrument after a prolonged period of inactivity, confirm normal and safe operation beforehand.
3. To take a good picture, be careful not to stain the objective lens with fingerprints or dust.
4. When this instrument is not in use, cap the objective lens and cover the instrument with the dust cover.
5. When the objective lens is stained, clean it according to Cleaning Lenses and Mirrors in this man­ual.
38
GENERAL INFORMATION ON USAGE AND MAINTENANCE

DISPOSAL

This symbol is applicable for EU member countries only. To avoid potential damage to the environment and possibly human health, this instrument should be disposed of (i) for EU member coun­tries - in accordance with WEEE (Directive on Waste Electrical and Electronic Equipment), or (ii) for all other countries, in accordance with local disposal and recycling laws.
The base contains strong springs. Do not attempt to disassemble or
CAUTION
• When disposing of SL-D301 parts, follow the local regulations for disposal and recycling.
burn the base, as the springs could cause injury by shooting out of it.
NOTE
GENERAL INFORMATION ON USAGE AND MAINTENANCE
39

PATIENT’S ENVIRONMENT

Do not use the power strip in the patient's environment. Connect the power supply of the device to the commercial power supply.
Radius 1.5m
1.5m
2.5m
1.5m
Devices applicable to the
Digital camera
patient’s environment
When the patient or inspector may touch the devices (including the connecting devices) or when the patient or inspector may touch the person that comes into contact with the devices (including the connecting devices), the patient's environment is shown below. In the patient's environment, use the device conforming to IEC60601-1. If you are compelled to use any device not conforming to IEC60601-1, use an insulation transformer.

Requirements for the EXTERNAL DEVICE

The external device connected to the analog and digital interfaces must comply with the respective IEC or ISO standards (e.g. IEC 60950-1 for data processing equipment and IEC 60601-1 for medical equipment). Anybody connecting additional equipment to medical electrical equipment configures a medical system and is therefore responsible that the system complies with the requirements for medial electrical systems. Attention is drawn to the fact that local laws take priority over the above mentioned requirements. If in doubt, contact your dealer or TOPCON (see the back cover).
40
GENERAL INFORMATION ON USAGE AND MAINTENANCE

ELECTROMAGNETIC COMPATIBILITY

This product conforms to the EMC standard (IEC 60601-1-2 Ed.3: 2007).
a) MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in the ACCOMPANYING
DOCUMENTS. b) Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT. c) The use of ACCESSORIES, transducers and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of the EQUIPMENT or SYSTEM as replacement
parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the
EQUIPMENT or SYSTEM. d) The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equipment. If adja-
cent or stacked use is necessary, the EQUIPMENT or SYSTEM should be observed to verify normal
operation in the configuration in which it will be used. e) The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other than those
specified may result in increased EMISSION or decreased IMMUNITY of the EQUIPMENT or SYS-
TEM.
Item
ACCESSORIES DIGITAL CAMERA UNIT DC-4 - - ­BACKGROUND ILLUMINATION BG-2GN - - ­CABLES AC Power cord (for AC adapter) Not Used Not Used 3.0 DC-4 TRIGGER cable Used Used 1.0 LAN cable Used Used 3.0 ILLUMINATION cable Used Not Used 1.6 AC adapter cable Not Used Used 0.15
Cable shield
Ferrite
core
Length(m)
Guidance and manufacturer's declaration - electromagnetic emissions
The SL-D301 is intended for use in the electromagnetic environment specified below. The customer or the user of the SL-D301 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The SL-D301 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Harmonic emissions IEC61000-3-2
Voltage fluctuations/ flicker emissions IEC61000-3-3
Class B
Class A
Complies
The SL-D301 is suitable for use in all establishments includ­ing domestic and those directly connected to the public low­voltage power supply network that supplies buildings used for domestic purposes.
41
GENERAL INFORMATION ON USAGE AND MAINTENANCE
Guidance and manufacturer's declaration - electromagnetic immunity
The SL-D301 is intended for use in the electromagnetic environment specified below. The customer or the user of the SL-D301 should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transient/burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions and Voltage variations on power supply input lines IEC 61000-4-11
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5% U
t
(>95% dip in Ut) for 0.5 cycle 40% U
t
(60% dip in Ut) for 5 cycles 70% U
t
(30% dip in Ut) for 25 cycles <5% U
t
(>95% dip in Ut) for 5 sec
Compliance
level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5% U
t
(>95% dip in Ut) for 0.5 cycle 40% U
t
(60% dip in Ut) for 5 cycles 70% U
t
(30% dip in Ut) for 25 cycles <5% U
t
(>95% dip in Ut) for 5 sec
Electromagnetic environment -
guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Main power quality should be that of a typical commercial or hospital environ­ment.
Main power quality should be that of a typical commercial or hospital environ­ment.
Main power quality should be that of a typical commercial or hospital environ­ment. If the user or the SL-D301 requires continued operation during main power interruptions, it is recom­mended that the SL-D301 be powered from an uninterruptible power supply or battery.
Power frequency (50/60 Hz) magnetic field
3 A/m 3 A/m
IEC 61000-4-8 NOTE U
is the a.c. main voltage prior to application of the test level.
t
42
GENERAL INFORMATION ON USAGE AND MAINTENANCE
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Guidance and manufacturer's declaration - electromagnetic immunity
PPP
The SL-D301 is intended for use in the electromagnetic environment specified below. The customer or the user of the SL-D301 should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 Vrms 150kHz to 80MHz
3 V/m 80MHz to 2.5GHz
Compliance
level
3 V
3 V/m
Electromagnetic environment -
guidance
Portable and mobile RF communications equipment should be used no closer to any part of the SL-D301, including cables, than the recommended separation distance calcu­lated from the equation applicable to the fre­quency of the transmitter.
Recommended separation distance
d = 1.2
d = 1.2 80MHz to 800MHz d = 2.3 800MHz to 2.5GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the rec- ommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site sur-
a
vey,
should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 NOTE 2
abField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be pre­dicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF trans­mitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the SL-D301 is used exceeds the applicable RF compliance level above, the SL­D301 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the SL-D301.
At 80 MHz and 800 MHz, the higher frequency range applies. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
43
GENERAL INFORMATION ON USAGE AND MAINTENANCE
Recommended separation distance between
PPP
portable and mobile RF communications equipment and the SL-D301
The SL-D301 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the SL-D301 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the SL-D301 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73 1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation dis­tance d in meters (m) can be estimated using the equation applicable to the frequency of the transmit­ter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 NOTE 2
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter
m
150kHz to 80MHz
d = 1.2
80MHz to 800MHz
d = 1.2
800MHz to 2.5GHz
d = 2.3
44
GENERAL INFORMATION ON USAGE AND MAINTENANCE

OPTICAL RADIATION HAZARD

• The light radiated from the instrument is potentially hazardous. The longer the exposure time is, the higher the risk of causing disorder to
CAUTION

RELATIVE SPECTRAL DISTRIBUTION OF ILLUMINATION LIGHT

the eye is.
• When operating with the maximum intensity, the light radiation reaches values exceeding the safety guideline in about 309 sec.
Because prolonged intense light exposure can damage the retina, the use of the device for ocular examination should not be unnecessarily prolonged, and the brightness setting should not exceed what is needed to provide clear visualization of the target structures.
The retinal exposure dose for a photochemical hazard is a product of the radiance and the exposure time. If the value of radiance were reduced in half, twice the time would be needed to reach the maximum exposure limit.
While no acute optical radiation hazards have been identified for slit lamps, it is recommended that the intensity of light directed into the patient's eye be limited to the minimum level which is necessary for diagnosis. Infants, aphakes and persons with diseased eyes will be at greater risk. The risk may also be increased if the person being examined has had any exposure with the same instrument or any other ophthalmic instrument using a vis­ible light source during the previous 24 hours. This will apply particularly if the eye has been exposed to retinal photography.
45
GENERAL INFORMATION ON USAGE AND MAINTENANCE

TYPE OF PLUG

Country Voltage/frequency Type of plug
Mexico 110V/50Hz Type C&E Argentina 220V/60Hz Type A Peru 220V/60Hz Type A Venezuela 110V/50Hz Type C&E Bolivia & Paraguay 220V/60Hz Type A (Most common)
Type H (Infrequently) Chile 220V/60Hz Type A Colombia 110V/50Hz Type C Brazil 220V/60Hz
127V/60Hz Ecuador 110V/50Hz Type C&E USA 120V/60Hz Type A (Hospital Grade) Canada 120V/60Hz Type A (Hospital Grade)
Type A Type C
46
GENERAL INFORMATION ON USAGE AND MAINTENANCE

OPTIONAL ACCESSORIES

TOPCON SLIT LAMP SL-D301 provides the following optional accessories for imaging. For inquiries, please call your dealer or TOPCON (see the back cover)
To prevent falling during use and movement, secure optional accesso-
CAUTION
•For details, please refer to the instruction manuals of each product.

SYSTEM CONFIGURATION

CAUTION
System Chart
TOPCON product
Products of other
ries.
To avoid electric shock, do not touch the external connection terminal and the patient at the same time.
Background Illumination BG-2GN
Diffusion Lens SO-DFL01
Hruby Lens
Applanation Tonometer Model R900
Tonometer Mount
SO-TM1
Iris Diaphragm
Yellow Filter
Parallel binoculer Tubes PB-2
SO-DF01
Adapt Cover SO-AC7
Digital Camera Unit DC-4
Trigger cable SO-TRIG01
12.5X Eyepiece
The one there is no click but with measurement scale.
12.5X Eyepiece
The one there is click, and without scale.
12.5X Eyepiece
The one there is click, and with scale.
12.5X Eyepiece
The one there is click, and with measurement scale.
20X Eyepiece
LAN cable SO-LAN01
insulating transformer
Tabletop SO-TABLE06
For AIT-16
Tabletop SO-TABLE07
For Unit table
47
OPTIONAL ACCESSORIES

DIGITAL CAMERA UNIT DC-4

FEATURES
• Simple operation and high-quality digital image.
• Optimized functions for slit lamp photography.
• Compact body and internal cabling for a neat look.
• All DC-4 functions are software driven.

BACKGROUND ILLUMINATION BG-2GN

FEATURES
• Used for background illumination.

YELLOW FILTER UNIT

FEATURES
• Combines with the blue filter prepared in the main body for a high-contrast fluorescence observation.
• Easy switching between filter insertion & removal.

IRIS DIAPHRAGM SO-DF01

FEATURES
When taking picture, you can adjust the light intensity and the depth according to the need.

12.5x Eyepiece

FEATURES
There are four types as below:
• One with internal measurement scale and no click stops.
• One without internal scale and click stops on each adjustment step.
• One with internal scale and click stops on each adjustment step.
• One with internal measurement scale and click stops on each adjustment step.

20X EYEPIECE

FEATURES
• Replaces the normal eyepiece for high magnification observation.
48
OPTIONAL ACCESSORIES

TONOMETER MOUNT SO-TM1

FEATURES
• For the measuring of the intraocular pressure, models R900 type and T900 type, Haag-Streit, are available.
* If the R900 type is being used for the SL-D301, the tonometer mount SO-
TM1 is required.
(Depending on specification, SO-TM1 may be included in standard accesso-
ries.)
* If the T900 type is in use, the tonometer guide plate (for T-900 type) is
required.

HRUBY LENS

FEATURES
Normally, observation is possible only up to the anterior vitreous body due to the refractive power of the cornea and lens. With the Hruby lens, the posterior vitreous body and the retina can be observed.

PARALLEL BINOCULAR TUBE PB-2

FEATURES
• Can observe a parallel view of the object.

ADAPT COVER SO-AC7

FEATURES
Used to cover the gaps with the microscope arm and hide cables when attaching the digital camera unit DC-4.
SO-AC7: For SL-D301 + DC-4 +SO-DF01

DIFFUSION LENS SO-DFL01

FEATURES
Large area can be illuminated with diffuse illumination.
49
OPTIONAL ACCESSORIES

TABLETOP SO-TABLE06, 07

FEATURES
• The tabletop to support the base unit, the chinrest unit and the power supply.
SO-TABLE06 For AIT-16
SO-TABLE07 For Unit table
50
OPTIONAL ACCESSORIES
52
OPTIONAL ACCESSORIES
Please provide the following information when contacting us regarding questions about this instrument:
Model name: SL-D301
Serial No.: This is printed on the rating nameplate on the right side of the
power supply unit.
Period of use: Please inform us of the date of purchase.
Defective condition: Please provide us with as much detail as possible on the
problem.
SLIT LAMP SL-D301
USER MANUAL
Rev.0 June 1, 2016
Published by TOPCON CORPORATION
75-1 Hasunuma-cho, Itabashi-ku, Tokyo, 174-8580 Japan.
SLIT LAMP
0
SL-D301
Printed in Japan 1606100TH
44703 95850
Loading...