GENERAL INSTRUCTIONS....................................................................................................................................4
GENERAL PREPARATION ..................................................................................................................................... 5
UNITS OF MEASUREMENT ..........................................................................................................................................7
DIRECT MEASUREMENT.............................................................................................................................................8
DIRECT MEASUREMENT OF CUPRIC ION (USING A STANDARD PH/MV METER)...........................................................8
DIRECT MEASUREMENT OF CUPRIC ION (USING AN ION METER) ..............................................................................10
LOW LEVEL CUPRIC DETERMINATION (USING A STANDARD PH/MV METER) ...........................................................11
LOW LEVEL CUPRIC DETERMINATION (USING AN ION METER).................................................................................12
TITRATION OF CUPRIC ION.......................................................................................................................................13
PRECIPITATION AND COMPLEXATION ......................................................................................................................15
TEMPERATURE INFLUENCES ....................................................................................................................................16
LIMITS OF DETECTION .............................................................................................................................................18
ELECTRODE LIFE .....................................................................................................................................................19
The Cupric Ion Electrodes are used to quickly, simply,
accurately, and economically measure cupric ion in aqueous
solutions.
Required Equipment
1. A pH/mV meter or an ion meter, either line operated or
portable.
2. Semi-logarithmic 4-cycle graph paper for calibration
curves when using the meter in the mV mode.
3. A magnetic stirrer.
4. The Cupric Ion Electrode (reference electrode
necessary), the Cupric Ion Combination Glass Electrode,
or the Cupric Ion Combination Epoxy Electrode.
5. The Double Junction Reference with Filling Solution in
the inner junction and with Filling Solution in the
outer junction.
6. Polishing Paper to polish dirty or etched electrode
membranes.
Required Solutions
1. Deionized or distilled water for solution and standard
preparation.
2. Ionic Strength Adjuster (ISA), 5M NaNO3. To prepare
this solution from your own laboratory stock, add 425
grams reagent-grade sodium nitrate, NaNO3, to a 1 liter
volumetric flask about half full of distilled water.
Swirl the flask to dissolve the solid. Dilute to the
mark with distilled water, cap the flask, and upend it
several times to thoroughly mix the solution. To each
100 ml of standard or sample, add 2 ml of ISA. The
background ionic strength of the resulting solution
will be 0.1M.
3. Cupric Nitrate Standard Solution, 0.1M. This solution
4
Instruction Manual Copper Electrode
may be prepared by adding 24.16 grams of reagent-grade
Cu(NO3)2 ∃ 3H2O to a 1 liter volumetric flask about
half full of distilled water. Swirl the flask to
dissolve the solid. Fill flask to the mark with
distilled water, cap the flask, and upend it several
times to mix the solution thoroughly.
4. Cupric Nitrate Standard Solution, 1000 ppm Cu+2. This
solution may be prepared by adding 3.80 grams of
reagent-grade Cu(NO3)2 ∃ 3H2O to a 1 liter volumetric
flask about half full of distilled water. Swirl the
flask to dissolve the solid. Fill the flask to the
mark with distilled water, cap the flask, and upend it
several times to mix the solution thoroughly.
5. EDTA Titrant (1M) for Cupric Titrations. To prepare
this solution from your own laboratory stock, add 37.2
grams of reagent-grade Na2EDTA ∃ 2H2O to a 100 ml
volumetric flask about three-fourths full of distilled
water. Swirl the flask to dissolve the solid. Add
distilled water to the mark, cap the flask, and upend
it several times to thoroughly mix the solution.
GENERAL PREPARATION
Electrode Preparation
Remove the rubber cap(s) covering the electrode tip(s) and the
rubber insert covering the filling hole of the cupric combination
ion electrode or the reference electrode. Fill the reference
electrode or the combination electrode with the filling solution
shipped with the electrode to a level just below the fill hole.
No preparation is required with a sealed reference electrode.
Connect the electrode(s) to the proper terminal(s) as recommended
by the meter manufacturer.
Electrode Slope Check (with standard pH/mV meter)
(check electrodes each day)
1. To a 150 ml beaker, add 100 ml of distilled water and 2
ml of ISA. Place the beaker on the magnetic stirrer
and begin stirring at a constant rate. After assuring
that the meter is in the mV mode, lower the electrode
tip(s) into the solution.
2. Using a pipette, add 1 ml of either the 0.1M or 1000
ppm cupric standard to the beaker. When the reading is
stable, record the millivolt reading.
5
Instruction Manual Copper Electrode
3. Using a pipette, add 10 ml of the standard used above
to the beaker. When the reading is stable, record the
millivolt reading.
4. Determine the difference between the two readings. A
difference of 27∀2 mV indicates correct electrode
operation, assuming the temperature is between 20o and
25oC. See the TROUBLESHOOTING GUIDE and
TROUBLESHOOTINGHINTS sections if the potential change
is not within this range.
Slope is defined as the change in potential observed when the
concentration changes by a factor of 10.
Electrode Slope Check (with ion meter)
(check electrodes each day)
1. Prepare standard cupric solutions whose concentrations
vary be tenfold. Use either the 0.1M or 1000 ppm stock
solutions and the serial dilution method for this
preparation.
2. To a 150 ml beaker, add 100 ml of the lower value
standard and 2 ml of ISA. Place the beaker on a
magnetic stirrer and begin stirring at a constant rate.
After assuring that the meter is in the concentration
mode, lower the electrode tip(s) into the solution.
3. After the reading has stabilized, adjust the meter to
the concentration of the standard and fix the value in
the memory according to the meter manufacturer's
instructions.
4. Rinse the electrode(s) with distilled water and blot
dry.
5. To a 150 ml beaker, add 100 ml of the higher value
standard and 2 ml of ISA. Place the beaker on the
magnetic stirrer and begin stirring at a constant rate.
Lower the electrode tip(s) into the solution.
6. After the reading has stabilized, adjust the meter to
the concentration of the standard and fix the value in
the memory.
7. Read the electrode slope according to the meter
manufacturer's instructions. Correct electrode
operation is indicated by a slope of 90-100%. If the
slope is not within this range, see the TROUBLESHOOTING GUIDE and TROUBLESHOOTING HINTS sections.
6
Instruction Manual Copper Electrode
MEASUREMENT
Measuring Hints
All samples and standards should be at the same temperature for
precise measurement.
Constant, but not violent, stirring is necessary for accurate
measurements. Magnetic stirrers can generate sufficient heat to
change the solution temperature. To counteract this effect,
place a piece of insulating material, such as styrofoam sheet,
between the stirrer and beaker.
Always rinse the electrode(s) with distilled water and blot dry
between measurements. Use a clean, dry tissue to prevent crosscontamination.
For samples with high ionic strength, prepare standards whose
composition is similar to the sample. Dilute concentrated
samples (>0.1M) before measurement.
Use fresh standards for calibration.
Use 2 ml of ISA for each 100 ml of sample or standard.
Always check to see that the membrane is free from air bubbles
after immersion into the standard or sample.
Sample Requirements
All samples must be aqueous and not contain organics which can
dissolve the epoxy electrode body and/or the cement bonding the
sensing crystal to the electrode body. Infrequent measurements
in solutions containing methanol, benzene, or acetonitrile are
permitted. Highly polar solvents slowly attack the electrode.
Please check with before using these electrodes in other organic
solvents.
The temperature of the standard and of the sample solution should
be the same and below 80 degrees C. About a 4% error in the
slope will occur for each 1oC difference in temperature.
Interferences should be absent. If they are present, use the
procedure found in the Interference and Electrode Response
sections to remove them.
Adjust sample pH with 1M HNO
to below pH 7 to avoid
3
precipitation of cupric hydroxide, Cu(OH)2.
Units of Measurement
Cupric ion concentrations are measured in units of parts per
million, moles per liter, equivalents per liter, or any other
7
Instruction Manual Copper Electrode
convenient concentration unit. Table 1 indicates some of these
concentration units.
TABLE 1: Concentration Unit Conversion Factors
ppm Cu+2 moles/liter
6354.0 1.0X10-1
635.0 1.0X10-2
63.5 1.0X10-3
6.4 1.0X10-4
MEASUREMENT PROCEDURES
Direct Measurement
A simple procedure for measuring a large number of samples. A
single meter reading is all that is required for each sample.
The ionic strength of samples and standards should be made the
same by adjustment with ISA. The temperature of both sample
solutions and standard solutions should be the same.
Direct Measurement of Cupric Ion (using a standard pH/mV meter)
1. By serial dilution of the 0.1M or 1000 ppm standards,
prepare 10-2, 10-3, and 10-4M or 100 and 10 ppm standards
for the cupric ion. To 100 ml of each standard, add 2
ml of ISA. Prepare standards with a composition
similar to the samples if the samples have an ionic
strength above 0.1M.
2. Place 100 ml of the 10-4M or 10 ppm standard in a 150 ml
beaker, place the beaker on the magnetic stirrer and
begin stirring at a constant rate.
3. Lower the electrode tip(s) into the solution. Make
sure that the meter is in the mV mode. When the
reading has stabilized, record the mV reading.
4. Place 100 ml of the 10
-3
M or 100 ppm standard in a 150
ml beaker, place the beaker on the magnetic stirrer and
begin stirring at a constant rate.
5. After rinsing the electrode(s) with distilled water and
blotting dry, immerse the electrode tip(s) in the
solution. When the reading has stabilized, record the
mV reading.
8
Instruction Manual Copper Electrode
6. Place 100 ml of the 10-2M or 1000 ppm standard in a 150
ml beaker, place the beaker on the magnetic stirrer and
begin stirring at a constant rate.
7. After rinsing the electrode(s) with distilled water and
blotting dry, immerse the electrode tip(s) in the
solution. When the reading has stabilized, record the
mV reading.
8. Using the semi-logarithmic graph paper, plot the mV
reading (linear axis) against concentration (log axis).
A typical calibration curve can be found in Figure 1.
A calibration curve is constructed on semilogarithmic paper when using a pH/mV meter
in the millivolt mode. The measured
electrode potential in mV (linear axis) is
plotted against the standard concentration
(log axis). In the linear region of the
curve, only three standards are necessary
to determine a calibration curve. In the
non-linear region, additional points must
be measured. The direct measurement
procedures given are for the linear portion
of the curve. The non-linear portion of
the curve requires the use of low level
procedures.
9. To a clean, dry 150 ml beaker, add 100 ml of the sample
and 2 ml of ISA. Place the beaker on the magnetic
stirrer and begin stirring at a constant rate.
10. After rinsing the electrode tip(s) in distilled water
9
Loading...
+ 19 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.