
PGA309
Quick Start System Reference
Guide
by Art Kay
High-Precision Linear Products
SBOA103C
Jan 2006

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
3
Required Equipment……………………………………………………..………4
Definition of sensor specifications………………………………………………5-12
PGA309 Absolute Calibration Example..……….………………………………13
Step 1: Will the PGA309 work for your application?..…..………………14-17
Step 2: Set up hardware……………………………………………..…..…18-21
Step 3: Configure PGA309 for initial scaling.……………………………..22-31
Step 4: Configure Sensor-Emulator-EVM to emulate
sensor……………………………………………………...………….…..…32-38
Step 5: Use the Calibration Spreadsheet to perform the calibration …39-59
PGA309 Ratiometric Calibration Example…………………………………..…60-86
PGA309 With Output Scaling Example………………………………………...87-95
PGA309 In Three-Wire Mode …………………………………………………..96-100
PGA309 Quick Start Contents

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
4
Required items for Quick Start
Hardware
• PGA309EVM –This is an evaluation kit that allows you to communicate with and interface to the PGA309. It
contains a PC Interface Board and a Sensor Interface Board combined with a PGA309 and EEPROM.
• Sensor-Emulator-EVM –This is an evaluation kit that uses rotary switches and trim potentiometers to
generate voltage excited bridge sensor output signals and temperature sensor output signals.
• +/-12V supply – Any low noise dc supply for the sensor emulator.
• Precision DVM – Any five or six digit meter that can read into microvolts(e.g., HP3458, HP34401).
• Slotted Jeweler’s Screwdriver –The best tool to quickly adjust the potentiometer.
Software
• PGA309DK Board Interface –This software is used to communicate with the PGA309EVM. See
http://focus.ti.com/docs/toolsw/folders/print/pga309evm-eu.html under support software for free download.
• PGA309 Calculator –This software is used to do initial gain scaling and verify that the design does not violate any
PGA309 specifications. Software is bundled with PGA309DK Board Interface software.
• PGA309 Calibration Spreadsheet – This spreadsheet uses PGA309 / Sensor readings over temperature and at
different applied stimulus levels to generate the calibration table used to correct for the sensor errors. Software is
bundled with PGA309DK Board Interface software.
• Generate_Emulator_Values.xls –This spreadsheet translates sensor specifications into voltage settings for the
Sensor-Emulator-EVM. See http://focus.ti.com/docs/toolsw/folders/print/sensor-emulator-evm.html under support
software for free download.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
5
Specifications
There are several key specifications that are used throughout our literature.
The mathematical definitions are listed below.
• Offset – the normalized output of a sensor (in V/V) with no applied stimulus.
• OffsetTC1 –The linear drift of the sensors’offset given in % of span/oC.
• NonlinOffsetDrift –The second order (quadratic) drift of the offset. This
coefficient is in % of span at room temperature.
• OffsetTC2 –The second order (quadratic) drift of the offset. This
coefficient is in % of span/oC2at room temperature.
• Span – the amount of change in normalized output voltage (in V/V) of the
sensor over the entire range of applied stimulus.
• SpanTC1 –The linear drift of the sensors’span given in % of span/oC.
• NonlinSpanDrift –The second order (quadratic) drift of the offset. This
coefficient is in % of span at room temperature.
• SpanTC2 –The second order (quadratic) drift of the span. This coefficient
is in % of span/oC2at room temperature.
• PressureNonlinearity–The second order (quadratic) nonlinearity versus
applied signal given in % of span.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
6
Span –the amount of change in normalized output voltage (in V/V) of the
sensor over the entire range of applied stimulus.
Offset –the normalized output of a sensor (in V/V) with no applied
stimulus.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
7
OffsetTC1
Offset3Offset
1
−
Span2T3T1−
( )
⋅
OffsetTC1
2.963 10
6−
× 1.624− 10
4−
×
( )
−
3.673 10
3−
× 85 40−( )−[ ]⋅
3.602 10
4−
×
% of span/oC
Bridge Sensitivity vs Temp
-5.0E-04
0.0E+00
5.0E-04
1.0E-03
1.5E-03
2.0E-03
2.5E-03
3.0E-03
3.5E-03
4.0E-03
-50 0 50 100 150
Temp, degC
Kbridge, V/V
offset
span
(T2, Span2)
(22.5C, 3.67E-3)
(T3, Offset3)
(85C, 2.96E-6)
(T2, Offset2)
(22.5C, 1.02E-3)
(T1, Offset1)
(-40C, -1.62E-3)
Linear end point fit is used to
determine the linear drift.
The linear drift of the
sensor’s offset given in
% of span/oC.
OffsetTC1

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
8
NonLinOffsetDrift
Offset
2
Offset1Offset
3
+
Span
2
NonLinOffsetDrift
1.023104−⋅
1.624− 104−⋅
( )
2.963106−⋅
( )
+
2
−
3.673 103−×
4.956 102−×
% of span
OffsetTC2
NonLinOffsetDrift
T3T1−
( )
2
2
4.956 102−×
85 40−( )−[ ]
2
2
1.269 105−×
% of span/oC
2
Bridge Sensitivity vs Temp
-5.0E-04
0.0E+00
5.0E-04
1.0E-03
1.5E-03
2.0E-03
2.5E-03
3.0E-03
3.5E-03
4.0E-03
-50 0 50 100 150
Temp, degC
Kbridge, V/V
offset
span
OffsetTC2:
The second order (quadratic) drift of
the offset. This coefficient is in %of
span/oC2at room temperature.
NonlinOffsetDrift:
The second order (quadratic) drift
of the offset. This coefficient is in
% of span at room temperature.
NonlinOffsetDrift
OffsetTC2

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
9
Bridge Sensitivity vs Temp
-5.0E-04
0.0E+00
5.0E-04
1.0E-03
1.5E-03
2.0E-03
2.5E-03
3.0E-03
3.5E-03
4.0E-03
-50 0 50 100 150
Temp, degC
Kbridge, V/V
offset
span
SpanTC1
Span3Span
1
−
Span2T3T1−
( )
⋅
SpanTC1
3.7284103−⋅ 3.4412103−⋅−
( )
3.6734 10
3−
× 85 40−( )−[ ]⋅
6.255 10
4−
×
% of span/oC
The linear drift of the
sensors’span given
in % of span/oC.
SpanTC1

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
10
NonLinSpanDrift
Span
2
Span1Span
3
+
( )
2
−
Span
2
NonLinSpanDrift
3.6734 103−×
3.4412103−⋅
( )
3.7284103−⋅
( )
+
2
−
3.6734 103−×
2.412 102−×
% of span
SpanTC2
NonLinSpanDrift
T3T1−
( )
2
2
2.412 102−×
85 40−( )−[ ]
2
2
6.175 106−×
% of span/oC
2
Bridge Sensitivity vs Temp
-5.0E-04
0.0E+00
5.0E-04
1.0E-03
1.5E-03
2.0E-03
2.5E-03
3.0E-03
3.5E-03
4.0E-03
-50 0 50 100 150
Temp, degC
Kbridge, V/V
offset
span
(T3, Span3)
(85C, 3.73E-3)
(T1, Span1)
(-40C, 3.44E-3)
(T2, Span2)
(22.5C, 3.67E-3)
The nonlinear coefficient assumes T2 is
equal distant t o T1 and T3. Thus the
vertex of the parabola will be at T2.
NonlinSpanDrift:
The second order (quadratic) drift of
the offset. This coefficient is in % of
span at room temperature.
NonlinSpanDrift
SpanTC2
SpanTC2:
The second order (quadratic) drift of
the span. This coefficient is in % of
span/oC2at room temperature.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
11
Sensor Output vs Applied Stimulus
0.00E+00
5.00E-04
1.00E-03
1.50E-03
2.00E-03
2.50E-03
3.00E-03
3.50E-03
4.00E-03
0 10 20 30 40 50 60 70 80 90 100 110
Applied Stimulus (%)
Sensor output (V/V)
ideal span
span
slope
real_sensor
100
real_sensor
0
−
stim
100
stim0−
( )
3.67103−⋅ 0−
100 0−( )
3.67 105−×
ideal_sensor stim( ) slopestim⋅
ideal_sensor 50( ) 3.67 105−×
( )
50⋅ 1.835 103−×
PresureNonlinearity
real_sensor50ideal_sensor
50
−
( )
real_sensor
100
100⋅
1.66 103−× 1.835 103−×−
( )
3.67 103−×
( )
100⋅ 4.768− %
The second order (quadratic)
nonlinearity versus applied signal
given in % of span.
Note: These readings were all
taken at room temperature. So,
real_sensor100 is the span of the
sensor at room temperature.
Pressure
Nonlinearity

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
12
P
nonlin
P() P 4Nonlinearity_pct⋅ 100⋅
P
100P100
⋅+
Span_TCT() SpanTC1T T
room
−
( )
⋅ SpanTC2T T
room
−
( )
2
⋅−
Offset_TCT() OffsetTC1T T
room
−
( )
⋅ OffsetTC2T T
room
−
( )
2
⋅−
SensorOutputPT,( ) Offset
room
Span
room
Offset_TCT()⋅+ Span
Nonlinearity_pct
100
1 Span_TCT()+( )⋅
The equations use the constants defined on the previous slides. These
equations are used in the generate_emu_settings.xls spreadsheet* to
compute the voltage settings for the Sensor-Emulator-EVM.
Sensor Output Equations
* Available for download at www.ti.com as SBOC065

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
13
For this quick start example the specifications below and the example hardware configuration will be used.
The Sensor-Emulator-EVM will create an equivalent for the illustrated Real World Inputs.
“generate_sim_values.xls
“Offset and Span”Tab
PGA309 Absolute
Calibration Example

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
14
Step 1: Will the PGA309
work for my sensor?
•Use your sensor’s specifications with the PGA309
Calculator software tool (SLVC073) to see if the PGA309
has the gain and offset adjustment range required to
accommodate your sensor.
•Use the PGA309 Calculator software tool to verify that
your design does not violate any of the most critical
PGA309 specifications (internal or external nodes).

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
15
Enter
information
here. For
example,
enter the
values shown.
Press Compute
Constants and
the resulting gain
settings will be
displayed here.
Enter your sensor parameters and your PGA309
configuration parameters to get the gain scaling.
If your design
generates values
for gain and offset
that are out of the
PGA309’s range,
the software will
flag the problem.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
16
The program selects values to allow
the Gain DAC and Zero DAC to have
the maximum adjustable range. The
Set Additional Constraints button is a
way to force the front end gain or
coarse offset to a constant. For this
example, set the coarse offset zero to
minimize noise. Click Apply
Constraints and then click Compute
Constants. In this case the range of
adjustment for the Zero DAC is
reduced but is still adequate to
correct for the sensor drift.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
17
After the gains and offsets of the
PGA309 have been calculated, press
Simulate Device to see if any internal
nodes are out of range.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
18
Step 2: Connect the hardware
Example of a Typical Engineering Bench Setup Using the Sensor Emulator
This diagram illustrates an example of how the Sensor-Emulator-EVM would be used in an engineering bench
setup. The PGA309 is a programmable sensor signal conditioning chip. The Sensor-Emulator-EVM can be used
in conjunction with the PGA309EVM (both versions) to facilitate the development of the PGA309 application.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
19
Jumper setup of PGA309EVM-xx and connections
to PC, power, and the Sensor-Emulator-EVM

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
20
Required Electrical Connections to
Sensor-Emulator-EVM

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
21
These jumpers
must be set to the
position shown to
allow the on-board
voltage reference
to generate the
emulated diode
voltages.
Sensor-Emulator-EVM Jumper Setup
These three
channels are used
to set the
temperature output
signal in the diode
mode. The Rt
channels are not
used in this mode.
Set the jumper
JUMP1 to the
position shown to
connect the
Diode
temperature
emulation.
Set the jumper
JUMP5 to the
position shown
to connect GND
to the bottom of
the bridge
emulator.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
22
Step 3: Do initial setup of the
PGA309 using the PGA309 DK
Program
•Copy the PGA309 Calculator results into the
PGA309DK software.
•Configure the PGA309 Temp ADC
•Calibrate the ADS1100 (ADC on PGA309EVM-xx
PC Interface Board Used to read the PGA309
output; read via software).

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
23
Start the PGA309 Designer’s Kit
Control Program. When it starts,
a message box will ask if you want
to load from the EEPROM (Press
No). Another box will indicate that
“the PGA309 EVM was detected
using the One-Wire interface.”
If the PGA309EVM does not work
properly, refer to the PGA309EVM
Users Guide.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
24
For this example, we will measure the PGA309 output voltage using an delta-sigma A/D converter on
the PGA309 PC Interface Board (the ADS1100). For optimal accuracy the ADS1100 should be
calibrated. To calibrate the ADS1100, measure the supply voltage Vs on the PGA309 PC Interface
Board (this should be close to 5V).

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
25
Press the Board
Settings button to
enter the calibration
factors. Enter the
measured value for
Vs then click
Read ADS1100.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
26
For the next part of the ADS1100
calibration, the input of the ADS1100
is shorted. The two boards must be
separated so that the PGA309 output
is not shorted by the calibration.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
27
Press calibrate
ADS1100. This will short
the input to the ADS1100
and measure the offset.
The calibration will take a
few seconds. When it is
complete close the
window.
When this step is done,
plug the two PGA309EVM
boards back together. At
this point the calibration
is complete.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
28
Step B:
Make sure PGA309 Test Pin HIGH
is checked. During calibration, the
PGA309 test pin must be set high.
This pin prevents the PGA309 from
reading the EEPROM during
calibration.
Step A:
It is a good practice to
press reset at the
beginning of a calibration
to insure all the registers
are in a known state.
Initial Configuration for the PGA309

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
29
Set the reference and
bridge excitation
voltage to the proper
values used in the
PGA309 Calculator.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
30
Copy the gain and
offset settings from the
PGA309 Calculator to
the PGA309
Designer’s Kit Control
Program.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
31
Step A:
Configure the
temperature ADC by
pressing the ADC
Config button. The
example settings
shown are good for a
diode measurement.
Step B:
Enter the example
settings shown for a
diode temperature
measurement. Press
OK when done.
Step C:
Press Write PGA309 to copy
all the information entered in
the program into the registers
of the PGA309.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
32
Step 4: Configure
Sensor-Emulator-EVM to
Emulate the Bridge Sensor
1. In order to use the Sensor-Emulator-EVM, you have to adjust a
number of trim potentiometers to configure the Sensor-EmulatorEVM so that it acts like your sensor. If the sensor’s raw output
characteristics are known, this step is simple: you adjust the SensorEmulator-EVM output to mimic your sensor.
2. In the case where you want to use a sensor data sheet to configure
the Sensor-Emulator-EVM, you can use the
generate_emu_settings.xls to translate your specifications to
Sensor-Emulator-EVM settings. Unfortunately, sensor manufacturers
may have specifications that do not conform to a standard, and
sometimes the specifications are difficult to understand. For our
tools we will mathematically define the specifications. You mayhave
to translate your particular specifications to our format.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
33
1. Offset and Span: Generates the bridge output voltages.
2. Diode Vo: Generates the temperature sensor output voltages for the diode method.
3. Rt-: Generates the temperature sensor voltages for the Rt-method.
4. Rt+: Generates the temperature sensor voltages for the Rt-method.
5. PGA309 Error: Allows you to read the PGA309 via the ADS1100 (The ADS1100 is the
delta-sigma A/D converter that is a part of the PGA309EVM-xx).
6. PGA309 RatioMetricError: Allows you to read and compute error for a ratiometric PGA309
setup.
If the raw output of the sensor is not known, the “Generate_Sim_Values.xls”spreadsheet can be used to
translate the specifications of your bridge sensor and temperature sensor to system voltage levels. The
spreadsheet contains five sections (Offset and Span, Diode Vo, Rt-, Rt+, PGA309 Error, Ratiometric Error):
The temperature measurement methods, Diode, Rt-, and Rt+ are described in detail in the SensorEmulator-EVM System Reference Guide (SBOA102) and the PGA309 Users Guide (SBOU024).
Configuring the Emulator to
Emulate a Real World Sensor

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
34
All the areas shown in light blue are either sensor specifications or system requirements. Enter these values and the
spreadsheet will generate output voltage settings for each channel on the sensor emulator. The next several pages will
show how the voltages listed in the spreadsheet are used to program the Sensor-Emulator-EVM.
Enter these
for our
example
Set Sensor-Emulator-EVM
potentiometers to generate these
voltages as detailed in pages 35-36
Generates the bridge output voltages from sensor specifications (“Generate_Sim_Values.xls”)
Offset and Span:

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
35
Bridge Sensitivity vs Temp
-1.0E-03
-5.0E-04
0.0E+00
5.0E-04
1.0E-03
1.5E-03
2.0E-03
2.5E-03
3.0E-03
3.5E-03
4.0E-03
4.5E-03
-50 0 50 100 150
Temp, degC
Kbr idge, V/V
offset
span
Each channel on the top section of the sensor emulator
represents a applied stimulus and temperature combination for
the sensor. Adjust the potentiometers coarse first, then fine, to
match the values computed by the Generate_Sim_Values.xls
spreadsheet for cold (0%. 50%, 100%), room (0%, 25%, 50%,
75%, 100%), and hot (0%, 50%, 100%). For example, the
sensor output at cold temperature and 0% of applied stimulus is
emulated by this channel. The rotary switch S1 is used to select
this channel. When the channel is selected, LED D101 will light
to indicate that the correct channel is selected.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
36
Bridge Sensitivity vs Temp
-1.0E-03
-5.0E-04
0.0E+00
5.0E-04
1.0E-03
1.5E-03
2.0E-03
2.5E-03
3.0E-03
3.5E-03
4.0E-03
4.5E-03
-50 0 50 100 150
Temp, degC
Kbr idge, V/V
offset
span
This is another example illustrating how a particular
channel on the sensor emulator represents an
applied stimulus and temperature combination for the
sensor. In this example, the sensor output at cold
temperature and 100% of applied stimulus is
emulated by this channel. The rotary switch S1 is
used to select this channel. When the channel is
selected, LED D103 will light to indicate that the
correct channel is selected.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
37
The second tab in the Generate_Sim_Values.xls spreadsheet allows the user to enter the temperature
range and room temperature diode voltage (light blue areas). Thespreadsheet calculates the diode
voltages and displays the results in the yellow areas. Note thatthe Temp ADC areas are specific to the
PGA309 sensor signal conditioning chip. The Temp ADC values willbe used in the computation of the
Counts for the temp ADC. The next several pages will show how the diode voltages are used to program
the sensor Sensor-Emulator-EVM.
Adjust the Diode section
potentiometer on the
Sensor-Emulator-EVM to
generate the counts as
detailed on page 38.
PGA309 Temp ADC
generate_sim_values.xls
Diode Vo Tab
Diode Vo: Generate Diode Voltages
based on Operating Temperature Range

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
38
Note when emulating Diode
temperature control, the Rt
temperature section is not used.
Each channel on the bottom section of the SensorEmulator-EVM represents the output of the
emulated temperature sensor. Using the Temp
DVM, adjust the respective potentiometers, coarse
first, the fine, to match the values computed by the
Generate_Sim_Values.xls spreadsheet for
Diode/Cold, Diode/Room, and Diode/Hot. For this
example, the temperature output signal at cold
temperature (-45oC) is emulated by this channel.
The rotary switch S2 is used to select this channel.
When the channel is selected, LED D201 will light
to indicate that the correct channel is selected.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
39
Step 5: Use the PGA309
Calibration Spreadsheet
•Select the calibration algorithm
•Copy the PGA309 registers into the spreadsheet
•Use the Sensor-Emulator-EVM to generate the
sensor outputs over temperature.
•Store calibration results into a file. Load this into
the PGA309 external EEPROM.
•Measure the post-calibration error. Perform a
second calibration to improve accuracy.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
40
For this example, use the PGA309 Calibration Spreadsheet. This tool uses
measured data (pressure and temperature) to create a lookup table that the
PGA309 will use to compensate for offset and gain drift. The spreadsheet will
also generate a coefficient that the PGA309 will use to correct for nonlinearity
verses applied pressure. Note: you will need to enable macros and load the
analysis toolpackto get this Excel sheet to work properly. Information
regarding configuration of Excel is detailed in the PGA309EVM Users’Guide.
When you bring up the spreadsheet, it will
ask you if you want to start the program.
Press No, because the program should
already be up from Step 2.
PGA309 Calibration
Spreadsheet, Main Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
41
Press Load
registers from
PGA309 to copy
the registers
from the
evaluation
fixture into the
spreadsheet.
PGA309 Calibration
Spreadsheet, Main Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
42
Press Prepare Calibration Sheet to select
the algorithm. In this example, we will do a
3 temperature3 pressure calibration.
Press OK after you have
selected 3 Temperature 3
Pressure calibration.
PGA309 Calibration
Spreadsheet, Main Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
43
Next the program will ask what
type of Temperature
Measurement Method you want
to use. For this example, we use
the diode method.
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
44
When the template for your
calibration algorithm is loaded
this box will pop up.
Press OK.
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
45
When the Load registers from
PGA309 button on the Main
sheet was pressed, the
PGA309 registers were copied
into this section of the Sensor
Curvefit sheet.
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
46
The appropriate values need to be
entered manually for the measurement
temperatures. The measurement
temperatures are the temperatures that
the calibration measurements are made
at.
The appropriate values need to be
entered manually for the temperature
range. This is the range that the curve fit
is done over. Enter the values shown for
our example.
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
47
The easiest way of doing
this is to select a cell and
press the Insert
TempADCreading in
active cell button. This
will insert a PGA309
Temp ADC in counts into
that cell.
The measured PGA309 Temp
ADC readings need to be
recorded at the respective
applied temperatures. Use the
temperature selector switch on
the Sensor-Emulator-EVM to
generate room,hot,and cold
readings for this example.
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
48
The easiest way of
doing this is to select a
cell and press the Insert
Voutreading in active
cell button. This will
insert a PGA309 output
voltage reading from the
ADS1100 delta sigma
ADC into that cell.
The PGA309 output voltage needs to
be recorded at the appropriate
applied pressure and temperature.
Use the bridge selector switch on the
Sensor-Emulator-EVM to generate
the respective room (0%, 50%,
100%), hot (0%, 25%, 50%, 75%,
100%), and cold (0%, 50%, 100%).
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
49
After the calibration
measurements are
complete look at the
graphs located on
the sensor Curvefit
sheet. These
graphs are an easy
way to check for
gross problems.
The graphs shown
are indicative of
typical results for
example.
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
50
Enter the output
voltage scale, the
number of points
in the table, and
the look up table
temperature
range. For our
example, enter
the values shown.
PGA309 Calibration
Spreadsheet, Calibration
Results Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
51
Range3: This is the range of
measurements made during
calibration. This range must be a
subset of Range1. It is OK for range 1
and range 3 to be equivalent.
Range1: This is the range of
the mathematical model of the
sensor that is developed by
the spreadsheet.
Range2: This is the range that the
look up table is developed over. This
range must be a subset of Range1. It
is ok for Range2 and Range3 to be
equivalent. This range over which the
calibrated sensor will correct for
temperature drift.
Note about the temperature ranges
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab
PGA309 Calibration
Spreadsheet, Calibration
Results Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
52
The Result
Sanity Check
will flag any
problems with
gain and offset
ranges in the
calibration
table.
The Voutmax and min
calibrated result graph gives
an idea of what errors you will
see based on the resolution
of internal components. Note
the output should
approximately match the
values entered in the Enter
Output Scale section.
PGA309 Calibration
Spreadsheet, Calibration
Results Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
53
The calibration table
that will be loaded into
the EEPROM is
displayed on the
Calibration Results tab
on the spreadsheet. At
this point the initial
calibration is complete
and the table can be
uploaded into the
PGA309 EEPROM.
PGA309 Calibration
Spreadsheet, Calibration
Results Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
54
Press the Save Registers + Lookup Table button. This will
store the lookup table into a file that can be loaded into the
PGA309 EEPROM.
PGA309 Calibration
Spreadsheet, Main Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
55
Step B
Make sure the PGA309 Test Pin
High box is not checked. When in
this mode the PGA309 will read the
EEPROM and adjust offset and
gain for each temperature
conversion
Step D
Press the Read PGA309 to see
the updated register values.
After Step D, the initial
calibration is complete.
Step C
Press Write EEPROM to
store the lookup table in
the external EEPROM .
Step A
Press the Open File
button to get the file
containing the
calibration results.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
56
The initial post
calibration results will
typically have errors
ranging from 0.1% to
0.3%.
The PGA309 Error tab on the
generate_sim_values.xls is a
convenient way to do a post calibration
error analysis. To use it select the
blue cell corresponding to the current
setup, and press the Insert Vout
reading in active cell button. This will
insert the PGA309 output reading from
the ADS1100.
generate_sim_values.xls,
PGA309 Error Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
57
The post first
calibration results are
made at room
temperature and
entered here. For this
example, use the
Sensor-Emulator-EVM
to generate 0% and
100% pressure at
room temperature.
Note the correction
factors are
developed based
on these readings.
These are used to
calibrate the
Lin_Dacerrors not
previously
accounted for.
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
58
The final calibration is
complete at this point
After making the secondary
calibration measurements,
store the calibration results
into a file and load them into
the PGA309 as with the first
calibration. The file for this
example calibration is saved
on the PGA309 Quick Start
Disk and is called
quick_start_second.txt.
Your results should be
similar to this file.
PGA309 Calibration
Spreadsheet, Main Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
59
The secondary
calibration can be
done to significantly
reduce the error. Postsecondary calibration
errors are typically on
the order of 0.05%.
The secondary
calibration involves
making two
measurements at
room temperature.
generate_sim_values.xls,
PGA309 Error Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
60
PGA309 Ratiometric
Calibration Example
This example walks through the PGA309 ratiometric calibration technique.
The PGA309 Absolute Calibration example is a more detailed description of a
calibration, and so, it is recommended that you review this example first. This
document describes the key elements that are required in a ratiometric
calibration, but does not fully explain how to use the PGA309 Gain Calculator,
Sensor-Emulator-EVM, or the Designers Kit Control Program.
For information on these development tools, please see the PGA309 Product
Folder on the TI website, at www.ti.com.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
61
This is the hardware
configuration that this
ratiometric calibration example
details. In this example, the
Sensor-Emulator-EVM is used
to emulate the bridge sensor
and the Diode. Note that the
device power supply is used to
provide excitation for the
sensor. So for this
configuration, the Vexcpin on
the PGA309 is not used and
consequently, the PGA309
cannot correct for nonlinearity
verses applied stimulus.
Temperature nonlinearities of
span and offset will still be
corrected.
Vexc
Vin+
Vin-
Temp
PGA309
Vsa
5
V
GND
Ref_In/
Ref_Out
Real World Inputs
PGA309 Ratiometric
Example

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
62
This diagram illustrates PGA309EVM jumper settings for a ratiometric system.
Sensor-Emulator-EVM connections and power connections are also shown.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
63
Required Electrical Connections to
Sensor-Emulator-EVM

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
64
These jumpers must
be set to the position
shown to allow the
on-board voltage
reference to generate
the emulated diode
voltages.
These three channels
are used to set the
temperature output
signal in the diode
mode. The Rt
channels are not used
in this mode.
Set the jumper
JUMP1 to the
position shown to
connect the Diode
temperature
emulation.
Set the jumper
JUMP5 to the
position shown to
connect GND to
the bottom of the
bridge emulator.
Sensor-Emulator-EVM Jumper Setup

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
65
The PGA309 Calculator can be
used to compute the gain and
offset settings for the PGA309.
These are the values used for
this ratiometric example
configuration.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
66
In the ratiometric configuration, the power supply (Vs) is beingused as the
reference. Thus, it is very important that the supply is measured during calibration.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
67
Configure the initial settings of the PGA309
Step B
The gain and offset values
computed by the calculator need
to be written into the PGA309
using the PGA309 Designer’s Kit
Control Program.
Step A
During calibration, the PGA309
Test Pin High must be checked to
prevent the PGA309 from reading
the EEPROM during calibration.
Step C
The value measured for Vs must
be typed in here. After all the
values are entered, press Write
PGA309.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
68
The configuration shown was selected
for this example (diode measurement
using the built-in 2.048V reference). It
is important to use the built in ADC
reference because the diode
measurement is absolute and all the
other references are relative to the
power supply for this configuration.
Configure the Temp ADC as
shown and click OK. From the
main window, press Write
PGA309.
Set up the PGA309
Temperature ADC

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
69
The sensor’s raw output is computed
and displayed here. These values are
used to setup the sensor emulator. The
sensor emulator EVM will need to be
adjusted to these levels.
The sensor specifications are
entered here. The definitions of
the different parameters is
described earlier in this document.
Note that Vexc= Vs for
ratiometric.
Enter these for our
example.
Note that Pressure Nonlin is zero. The sensor
must be linear for this configuration because the
sensor excitation is the power supply and so the
nonlinearity correction circuit cannot be used.
generate_sim_values.xls,
Offset and Span Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
70
When the sensor’s specifications
have been entered, the
spreadsheet will display the bridge
output versus temperature.
The bridge output versus applied
stimulus is also displayed. This
must be a linear function for a
ratiometric setup that does not use
Vexcfor bridge excitation.
generate_sim_values.xls,
Offset and Span Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
71
Bridge Sensitivity vs Temp
-1.0E-03
-5.0E-04
0.0E+00
5.0E-04
1.0E-03
1.5E-03
2.0E-03
2.5E-03
3.0E-03
3.5E-03
4.0E-03
4.5E-03
-50 0 50 100 150
Temp, degC
Kbr idge, V/V
offset
span
Each channel on the top section of the sensor emulator
represents a applied stimulus and temperature combination
for the sensor. Adjust the potentiometers coarse first, then
fine, to match the values computed by the
Generate_Sim_Values.xls spreadsheet for cold (0%. 50%,
100%), room (0%, 25%, 50%, 75%, 100%), and hot (0%,
50%, 100%). For example, the sensor output at cold
temperature and 0% of applied stimulus is emulated by this
channel. The rotary switch S1 is used to select this channel.
When the channel is selected, LED D101 will light to indicate
that the correct channel is selected.
For this ratiometric example adjust
the potentiometer on the SensorEmulator-EVM to the bridge
section to produce the respective
voltages shown.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
72
Note when emulating Diode
temperature control, the Rt
temperature section is not used.
Each channel on the bottom section of the SensorEmulator-EVM represents the output of the
emulated temperature sensor. Using the Temp
DVM adjust the respective potentiometers, coarse
first, the fine, to match the values computed by the
Generate_Sim_Values.xls spreadsheet for
Diode/Cold, Diode/Room, and Diode/Hot. For this
example, the temperature output signal at cold
temperature (-45oC) is emulated by this channel.
The rotary switch S2 is used to select this channel.
When the channel is selected, LED D201 will light
to indicate that the correct channel is selected.
This is the Diode Vo tab on the
Generate_Sim_Values.xls
spreadsheet.
It is used to compute diode voltages that are
used to set up the Sensor-Emulator-EVM.
For this ratiometric example, adjust the
potentiometer on the Sensor-EmulatorEVM to the diode temperature section to
product the respective counts shown for
temperature.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
73
Press Load
registers from
PGA309 to copy
the registers
from the
evaluation
fixture into the
spreadsheet.
For this ratiometric example, use the PGA309 Calibration Spreadsheet. This
tool uses measured data (pressure and temperature) to create a lookup table that
the PGA309 will use to compensate for offset and gain drift. The spreadsheet will
also generate a coefficient that the PGA309 will use to correct for nonlinearity
verses applied pressure. Note: you will need to enable macros and load the
analysis toolpackto get this Excel sheet to work properly. Information
regarding configuration of Excel is detailed in the PGA309EVM Users Guide.
PGA309 Calibration
Spreadsheet, Main Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
74
Press Prepare Calibration Sheet to select
the algorithm. In this example, we will do a
3 pressure 3 temperature calibration.
Press OK after you have
selected 3 Temperature 3
Pressure calibration.
PGA309 Calibration
Spreadsheet, Main Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
75
Next, the program will ask
what type of Temperature
Measurement Method you
want to use. For this example,
we use the diode method.
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
76
This are contains the PGA309 settings. These settings
are loaded into these cells when the Load Registers From
PGA309 button was pressed from the main tab.
The temperature
ranges and pressure
ranges need to be
entered by hand.
The TempADC
readings and
VoutMeasvalues
need to be
measured. This can
be done using the
Insert TempADC
reading in active cell
and Insert Vout
reading in active cell
buttons.
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
77
Note that the 3 pressure 3 temperature calibration algorithm
will compute values for nonlinearity error. This value needs
to be very small for this configuration because nonlinearity
correction is not used. This value will not be used to
generate the calibration tables in the EEPROM.
The value of Klinstored in the
EEPROM will be zero for this mode
because Vexcis disabled.
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
78
For the ratiometric calibration method,
the secondary calibration is not
necessary. The secondary calibration is
used to correct for errors introduced by
the LinDac. So, for this example, this
section is left blank.
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
79
Select the desired
post-calibration
output range.
Select the
temperature range of
the look-up-table.
Make sure the Result
Sanity Check passes.
PGA309 Calibration
Spreadsheet, Calibration
Results Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
80
Press Save
Registers+LookupTable.
The spreadsheet will let you know that
Excitation is Disabled. This is normal
for the ratiometric method.
PGA309 Calibration
Spreadsheet, Main Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
81
Step B
Make sure the PGA309 Test Pin High
box is not checked. When in this
mode, the PGA309 will read the
EEPROM and adjust offset and gain for
each temperature conversion
Step D
Press the Read PGA309 to see the
updated register values.
Step C
Press Write EEPROM to
store the table on the
EEPROM .
Step A
Press the Open File
button to get the file
containing the
calibration results.
After Step D, the calibration is
complete.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
82
The post-calibration
results will typically have
errors less than 0.1%.
The PGA309 RatioMetricError tab on
the generate_sim_values.xls is a
convenient way to do a post-calibration
error analysis. To use it, select the
blue cell corresponding to the current
setup, and press the Insert Vout
reading in active cell button. This will
insert the PGA309 output reading from
the ADS1100. This spreadsheet page
provides for error calculations at two
different power supply voltages. The
initial supply is Vs= 4.963V (you need
to enter your measured Vs).
generate_sim_values.xls,
PGA309 Ratiometric
Error Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
83
For the ratiometric calibration
method it is useful to adjust the
power supply to see how PSR
affects the PGA309 calibrated
accuracy. A 10% power supply
deviation is used in this example
because it is a typical worst case
deviation for ratiometric systems.
Connecting a 53kΩ resistor
between the 3V pin and the
center pin on JA will cause the
power supply to shift from 5V to
4.5V. You can adjust the value of
the shunt resistance to get more
or less power supply deviation.
A short will cause the power
supply to deviate from 5V to 3V.
Vout
ADJ
GND
Vin
Vin
NC
EN
Vout
53k
3V
JA
5V
JA
JB
JF
JC
JD
JE
VOUT
5V
PROG
3V
PC
1PU
INPU
ON
SDN
RTS
DTR
DIS
PGA309 PC Interface Board
To PC
Serial
Port
6V dc Power
from wall
adaptor
53k
Schematic
Mechanical Diagram

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
84
Make sure that you measure the
supply voltage (Vs) and enter it into
the PGA309 Designer’s Kit Control
Program.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
85
Make sure that you
measure the supply voltage
(Vs) and enter it into the
PGA309 Designer’s Kit
Control Program.
Measure the PGA309 post calibration error
at a different supply voltage to see the
affect of PSR on error. For this example,
the supply was changed from Vs= 4.963V
to Vs=4.457V and the average error
changed from -0.06% to -0.03%.
generate_sim_values.xls,
PGA309 Ratiometric
Error Tab

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
86
PGA309 With Output Scaling
(0 to 10V)
V
o
R
f
R
1
R
f
R
2
+ 1+
V
ref
⋅−
VoExtraGain( )Vin⋅ ExtraOffset( )V
ref
⋅+
In many applications an external gain stage is used to get an output swing beyond the range of the
PGA309. The circuit shown below is a typical example of gain scaling with an offset shift. The PGA309
calibration spreadsheet can accommodate external gain and offsetscaling. Doing the calibration by
measuring the output of the external gain stage will calibrate out errors caused by resistor tolerance in the
external stage.
+
-
R1
R2
Rf
Vref
Vo
PGA309
Vin
Vout

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
87
PGA309 With Output Scaling
(0 to 10V)
This calculation shows how
the example configuration
shown on the previous page
can be used to take the 0.5V
to 4.5V output of the PGA309
and re-scale it to 0V to 10V.
This equation must be broken
down into an “ExtraGain”and
“ExtraOffset”factor for the
spreadsheet.
ExtraOffset 0.5−=
ExtraOffset
Rf−
10103⋅
:=
ExtraGain 2.5=
ExtraGain
R
f
R
1
R
f
R
2
+ 1+
:=
For the spreadsheet you need to break the function into "extra gain" and "extra
offset" as shown below.
Rf5000.
Substitute the value of R2 into
equation 2 to solve for Rf
R25000.
Substitute equation 2 into
equation 3 and solve for R2
Equation 3: Vin=4.5V, Vo = 10V
10
R
f
10103⋅
R
f
R
2
+ 1+
2.5⋅−
Equation 2: solve Equation 1 for Rf
Rf2500.
R
2
R22500.−
⋅
Equation 1: Vin = 0.5V, Vo = 0V
0
R
f
10103⋅
R
f
R
2
+ 1+

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
88
PGA309 With Output Scaling
(0 to 10V)
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab
The “ExtraGain”and
“ExtraOffset”factor are
entered here on the
spreadsheet. Normally these
are set to ExtraGain= 1.0 and
ExtraOffset= 0.0.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
89
PGA309 With Output Scaling
(0 to 10V)
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab
The data that is measured at
the output of the external
amplifier is entered directly
into the spreadsheet.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
90
PGA309 With Output Scaling
(0 to 10V)
The output range must be
include the scaling stage.
PGA309 Calibration
Spreadsheet, Calibration
Results Tab
Other then these few minor
changes, the calibration
method is the same as the
other examples.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
91
I
outIinIref
+
100⋅
I
out
V
in
R
1
V
ref
R
2
+
V
ref
⋅+
I
out
ExtraGain( )Vin⋅ ExtraOffset( )2.5⋅+
The spreadsheet can also be used to calibrate a system using a PGA309 with a 4mA to 20mA output
scaling.
PGA309 With Output Scaling
(4mA to 20mA)

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
92
4103−⋅
0.5
R
1
2.5
R
2
+
100⋅
Equation 1: Vin = 0.5V, Iout = 4mA
R112500.
R
2
R262500.−
⋅
Equation 2: solve Equation 1 for R1
20103−⋅
4.5
R
1
2.5
R
2
+
100⋅
Equation 3: Vin=4.5V, Iout = 20mA
Substitute equation 2 into
equation 3 and solve for R2
R2125103⋅
Substitute the value of R2 into
equation 2 to solve for R1
R125103⋅
For the spreadsheet you need to break the function into "extra gain" and "extra
offset" as shown below.
I
out
100
R
1
V
ref
⋅+
I
out
100
25103⋅
2.5⋅+
I
out
ExtraGain( )Vin⋅ ExtraOffset( )2.5⋅+
I
out
4 103−×
( )
Vin⋅ 800 106−×
( )
2.5⋅+
This calculation shows how
the example configuration
shown on the previous page
can be used to take the 0.5V
to 4.5V output of the PGA309
and re-scale it to 4mA to
20mA.
This equation must be broken
down into an “ExtraGain”and
“ExtraOffset”factor for the
spreadsheet.
PGA309 With Output Scaling
(4mA to 20mA)

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
93
The “ExtraGain”and
“ExtraOffset”factor are
entered here on the
spreadsheet. Normally these
are set to ExtraGain= 1.0 and
ExtraOffset= 0.0.
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab
PGA309 With Output Scaling
(4mA to 20mA)

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
94
The data that is measured at
the output of the voltage to
current converter is entered
directly into the spreadsheet
(in Amps).
PGA309 Calibration
Spreadsheet, Sensor
CurvefitTab
PGA309 With Output Scaling
(4mA to 20mA)

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
95
The output range must be
include the scaling stage (in
Amps).
Other then these few minor
changes, the calibration
method is the same as the
other examples.
PGA309 Calibration
Spreadsheet, Calibration
Results Tab
PGA309 With Output Scaling
(4mA to 20mA)

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
96
PGA309 In Three Wire Mode
In many cases the PGA309 is connected in a configuration referred to as a three wire connection. In this
configuration the only wires that need to connect to the sensor module are power, ground, and Vout. In
this configuration the One-Wire digital communication line is connected to the Voutpin. When the
PGA309 is initially powered up, the Voutpin is placed in a high impedance mode for 15mS. If
communication is established using the One-Wire during this time, the PGA309 will keep Voutin high
impedance until the communications is complete. After the communication is complete the PGA309 Vout
pin will become active and remain active until power is cycled again. While using the EVM to
communicate in Three Wire Mode, the EVM will cycle power before each One-Wire communication.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
97
PGA309 In Three Wire Mode
If the “Set PreCalEE”feature is used the test pin is normally grounded (leave “PGA309 Test Pin
HIGH”box unchecked).

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
98
PGA309 In Three Wire Mode
A key technique used in calibration is to use the test pin on the PGA309. The test pin is typically used
during calibration to place the PGA309 into test mode. The mainbenefit of test mode is that the Gain
DAC and Offset DAC are forced to remain at the last values written to their respective registers.
In the case of three wire mode the test pin is grounded and cannot be used. In this case, an EEPROM
table can be built that will force that Gain DAC and Offset DAC to be constant. The PGA309 Designers
Kit Control Program “Set PrecalEE”feature simplifies the creation of this table.

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
99
PGA309 In Three Wire Mode
When using this feature, first set all the registers to values your application requires. Then press the “Set
PreCalEE”button.
Step 1
Step 2

SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
100
PGA309 In Three Wire Mode
After Pressing the “Set PreCalEE”a dialogue box will pop up that verifies the value of the Zero DAC and
Gain Dacyou want in your EEPROM configuration. After creating the EEPROM table, the PGA309
Designer’s Kit Control Program is ready for to be used with the calibration spreadsheet.
After pressing
Generate and
Write EEPROM
Table, the lookup
table will be
updated to force a
constant Gain
Dacand Zero Dac
for PreCal
settings.