Testo 310 Operating Manual

testo 310 flue gas analyser
Brief instructions
These brief instructions provide an overview over the most important functions of the product. Before using the product, you must read and follow the instruction manual!
Commissioning/instrument configuration menu
Area version, measurement units, time and date can be set.
When the instrument is switched on initially, the configuration menu is opened automatically.
1. Switch the instrument on: hold down [] until all segments are shown on the display.
2. Open instrument configuration menu: Press right function key [set] during the initialisation phase.
3. Make settings:
[] switches to the next parameter, while [esc] takes you to the
previous parameter.
Parameter Explanation
ArEA (area
version)
Pressure, temperature
Time, date
Selecting the area version activates different calculation formulas and associated measurement parameters.
> Select area version code: [] and [].
1: USA, HU, IN, KR / 2: GB, RU, IT, CZ, DK, AU, JP, CN / 3: NL, SE, TR, RO / 4: DE, AT, CH, FR, ES, BE, PL, PT,
AR, BR
> Confirm the entry: [OK].
> Select the unit: [] and [], confirm the entry: [OK].
> Select the parameter: [], set the values: [] and [],
confirm the entry: [OK].
Measurements configuration menu
Fuels and measurement units can be set.
1. Switch the instrument on: hold down [] until all segments are shown on
the display.
2. Open Measurements configuration menu: After the instrument initialisation
phase, press the right function key ([set]).
3. Make settings:
[] switches to the next parameter, while [esc] takes you to the
previous parameter.
Parameter Explanation
Fuel
> Select the fuel: [] and [], switch to the next parameter:
[].
Pressure
> Select the unit: [] and [], switch to the next parameter:
[].
Temperature
> Select the unit: [] and [], confirm the entries: [OK].
Preparing measurements
After switching on the instrument or selecting a flue gas ( ) or ambient CO
probe ( ) measurement, the gas sensors are zeroed and the combustion air temperature is measured: the flue gas probe must be close to the burner's fresh air inlet duct!
1. Switch the instrument on: hold down [] until all segments are shown on
the display.
2. Select the fuel: [] and [], confirm the entry: [OK].
Performing the measurement
Measure-
Explanation
ment type
Flue gas
> Select : [] [OK].
1. Start measurement: [Start]. > Edit reading display line 1/line 2: [] / [].
2. End measurement: [Stop]. > Remove flue gas probe from the flue gas duct and purge with
fresh air.