testo 270 User guide

testo 270 · Deep-frying Oil Tester
Short manual
Pos: 1 /Kurzanleitungen/testo 270/Vor der Messung ( KurzBAL) @ 4\mod_124637611 8402_6.doc @ 45547 @ 111
Before the measurement
2. Wait until no more bubbles rise (approx. 5 minutes).
The temperature of the deep-frying oil to be measured must be at least 40 °C. The maximum operating temperature is 200 °C.
3. Switch on instrument [
4. Immerse sensor in hot oil.
> Observe Min./Max. marking.
> Sensor: Maintain a minimum distance of 1 cm from metal
parts.
/Hold].
During the measurement
1. Move the sensor in the oil (equalization period
approx. 20 s)
with Auto Hold:
- Auto Hold in the display shows the end of the
measurement. If set, an acoustic signal also sounds.
- Final value is held.
without Auto Hold:
- If the temperature display no longer changes, a stable
TPM final value has been reached.
> To hold readings: Briefly press [
2. To change back to Measuring Mode: Briefly press
[
/Hold] (< 1 s).
/Hold] (< 1 s).
After the measurement
Clean
WARNING
Risk of burns due to hot instrument parts (sensor and probe shaft)!
> Do not touch hot instrument parts with your hands.
> Allow instrument to cool sufficiently before cleaning.
> In the event of burns, immediately cool corresponding spot with cold water
and see a doctor if necessary.
CAUTION
Possible damage to the sensor!
> Do not remove cold oil residues from the sensor.
> Do not use any sharp-edged objects.
> Do not use aggressive cleaning agents and solvents.
TopSafe and wrist strap can be cleaned in a dishwasher.