Teledyne SofTA 300S ELSD Quick Start Manual

SofTA 300S ELSD Quick Start Guide
Instruction Sheet 8-243109
May 2016
Overview
This quick start guide will take you through the steps to successfully set up and use your SofTA 300S ELSD. Please see the SofTA 300S ELSD User Manual for more information regarding this product.
Connections
steel tube. Use the supplied exhaust adaptor and additional tubing to connect to a fume hood dur­ing operation.
Communication: An output jack can be found on
the back of the ELSD for connection to a chart recorder, computing integrator, or computer­ized data system. The maximum signal output is 1 V.
Gas Connections: Connect a supply of clean, dry,
inert gas regulated to 65 psi to the on the back of the unit. The internal gas regula­tor maintains the gas pressure at the factory set value displayed on the front panel.
Power Connection: The ELSD operates at either
120 V or 240 V, 50/60 Hz.
Fluid Connections: Connect the outlet from your
HPLC injector to the inlet line on the front of the unit. The length and volume between the injec­tor outlet and the detector inlet should be kept as short as possible to avoid unnecessary band broadening.
Thermo-Split Drain: Place the ¼” stainless steel
drain adapter, provided in the accessory kit, on the drain port located on the front of the instru­ment. Direct the outlet of the tube to a collec­tion vessel below the drain. Place the open end of the flexible ¼” elastometric tubing in a suit­able waste container, and trim the length as nec­essary. Watch the liquid level in the container during operation and empty it when it is full (especially when running the spray chamber at sub-ambient temperature).
Location
All components of the system should be placed on a sturdy table or bench in a room with a room tempera­ture of 20 ± 5 °C. The detector is designed to fit alongside an HPLC system. The detector must be placed with at least 4” of clearance on the right and back of the detector to allow for proper air flow through the detector. Access to the back of the unit is required to make the appropriate connections and provide for safe removal of the power cord.
GAS INLET port
Operation
Instrument Controls
The ELSD is controlled via six touch-sensitive keys on the front panel labeled PWR (power), A/Z (A/Z), ESC (escape), behind the front panel. The touch-sensitive keys are activated by lightly placing a finger over the light repre­senting the key. An audible beep is heard when the key is activated. Touching the parameters section to review operation conditions. If there is a * at the upper right corner, touching ESC key will bring up a screen to explain what’s wrong.
Menu
Touching the ENT key will bring up the MENU screen.
Start-up
(up), (down) and ENT (enter) found
or keys will circulate the
SC = Spray Chamber (setpoint: 10 °C to 60 °C; readout range: -10.0 °C to 120.0 °C) DT = Drift Tube (setpoint: 22°C to 120°C; readout range: -10.0 °C to 120.0 °C) OC = Optical Cell (setpoint: 22 °C to 70 °C; readout range: -10.0 °C to 120.0 °C) RST= Reset baseline to about 20 mV. BLN = Baseline reading (5000.000 – 0.000 mV) GAS = Gas (0.0 to 130.0 psi) FLT or BFT = Baseline Filter (OFF, weight 1 to 10) FS = Full Scale (1 V or 10 mV) GAIN = Gain (Normal or Low) CAL = Calibration (50% to 200%)
1. Make all connections (gas, liquid, power, com­munications, exhaust, and drain) as described above. Do not turn on the gas supply yet.
2. Press the M nostic tests. When the tests are complete, press­ing the M mode.
3. Confirm that the correct conditions are set by touching the
parameters. If not to display the M
keys to select set points. Touch M key to accept the changes. The A functions as an escape key, and abandons any changes made.
4. Start your HPLC pump to deliver 1 mL/min mobile phase to the detector. When you see liq­uid coming out of the drain port, the internal “p” trap is filled with the mobile phase. (This may
ENU | Power key to begin self-diag-
ENU | Power key again will exit standby
or keys to cycle through the
, touch the MENU | Power key
ENU screen. Touch the or
ENU | Power
UTOZERO key
take up to 10 minutes.) Stop the pump at this time.
5. Before conditions are met, a * blinks at the upper right corner of the display and the green light above Ready is not lit. Press the
ENU | Power key to display the error condition.
M Allow up to 30 minutes for temperatures to reach their set points and the electronics to stabilize.
6. Start the gas flow.
7. When the gas flow is stable, start the mobile phase flow.
8. When the R
UTOZERO key to reset the baseline. Cycle
A
EADY light illuminates, press the
through the settings to BLN. This is the voltage offset by the A
UTOZERO button. Record this value.
It is important to record this value for each set of operating conditions. If the value is high (greater than 500 mV), the mobile phase is not being evaporated and the detector temperatures should be modified. Monitor your data system and ELSD signal for stability. The signal should be stable within 2 mV. When stable, begin your analysis.
Shut Down Procedure
Stop the flow of mobile phase to the system. Wait for the pump pressure to return to zero. Select the E
TANDBY button from the HOME page or activate the
S power key and follow the on screen directions. Allow the standby timer to count down. In standby, the gas flow is shut off and the laser and heaters are turned off.
NTER
performance. The following table can provide a starting point for some common mobile phases. Optimization for the best signal-to-noise ratio will provide the best sensitivity.
Solvent @ 1.0mL/min DT, °C SC, °C
90/10 Water/METHANOL 45 15 90/10 Methanol/WATER 65 50 50/50 Methanol/Water 60 30 Acetonitrile 60 50 Methanol 60 50 Water 45 10
Choosing Operating Conditions
The drift tube temperature and the Thermo-Split spray chamber temperature are selected to provide the maximum detector response with minimum baseline noise. The temperatures are selected based on the sol­vent volatility and mobile phase flowrate. Some experimentation will be required to optimize the ELSD.
When setting the ELSD temperatures for a new method, select 25 °C for spray chamber temperature and 55 °C for drift tube temperature. These tempera­tures should then be adjusted for the best signal-to-noise ratio during method optimization. A mobile phase that is highly organic and volatile will require a moderately high drift tube temperature and an ambient or elevated spray chamber temperature. Highly aqueous or high boiling point organic mobile phases will perform best at moderate drift tube temperatures and sub-ambient spray chamber temperatures.
Each mobile phase can require a different set of oper­ating conditions to achieve the best detector
Teledyne Isco
P.O. Box 82531, Lincoln, Nebraska, 68501 USA Toll-free: (800) 775-2965 • Phone: (402) 464-0231 • Fax: (402) 465-3001 E-mail: IscoService@teledyne.com
Teledyne Isco is continually improving its products and reserves the right to change product specifications, replacement parts, schematics, and instructions without notice.
Loading...