The print history shown below lists the printing dates of all revisions and addenda created for this
manual. The revision level letter increases alphabetically as the manual undergoes subsequent
updates. Addenda, which are released between revisions, contain important change information that
the user should incorporate immediately into the manual. Addenda are numbered sequentially.
When a new revision is created, all addenda associated with the previous revision of the manual are
incorporated into the new revision of the manual. Each new revision includes a revised copy of this
print history page.
Revision A (Document Number 154-032000) ............................................................. March 2000
page 2
Hastings Instruments reserves the right to change or modify the design of its equipment without
any obligation to provide notification of change or intent to change.
The Hastings Power Supply (HPS-10) supplies stable ±15 VDC specifically for Hastings mass
flowmeters (HFM-200 and HFM-300 series). The flowmeter returns a 0-5VDC output that is
proportional to flow that is digitally displayed.
Specifications
Output voltage ................................................................................................. ±15 VDC ± 2.5 %
Power input ............................................................................................. 115 VAC ±10%, 50/60 Hz
Signal input ..................................................................................................................... 0 - 5 VDC
Carefully unpack the HPS-10. Inspect all items received at this time for any obvious damage that
may have occurred during shipment. Immediately advise the shipping agent if there is any damage
to items in your shipment. Identify each item and the quantity received with those listed on the
packing list.
Mounting
Cut out panel as shown in the diagram below.
SECTION 2SECTION 2
SECTION 2
SECTION 2SECTION 2
Installation
page 7
Page 8
To panel mount the HPS-10:
1) Slide power supply into panel from the front.
2) Attach brackets to the side of the power supply.
3) Tighten screws against panel.
See Figure below:
page 8
Page 9
Hastings HPS-10 Power Supply Cable Connections
The HPS-10 is supplied with a power cable. The 3-wire cable plugs into the back of the instrument
in the socket marked “AC POWER”.
A 5-wire shielded cable connects the instrument to the flowmeter. The diagram below shows the
proper connections.
BLACK
ORANGE
YELLOW
115 +5% V AC
ANALOG OUT
GREEN + SHIELD
BLUE
page 9
Page 10
page 10
Page 11
Normal Start-up
1) Connect the interface cable from the HPS-10 to the flowmeter. (See figure on page 2-4 of
this manual)
2) Verify that the AC power to be applied to the HPS-10 is correct (115 VAC).
3) Apply AC power to the HPS-10.
4) The HFM instrument has a local zero adjustment. Follow the zero adjustment procedure
described in the HFM-200 or HFM-300 series manual.
Before applying power to the instrument, ensure that the line voltage being
applied is correct. Failure to do so may result in damage to the instrument
not covered by the warranty.
Caution
SECTION 3SECTION 3
SECTION 3
SECTION 3SECTION 3
Operating Information
Analog Output
The HPS-10 generates an analog signal which is directly proportional to the displayed pressure.
The analog output signal is given by:
The output signal is a DC voltage with a full scale reading for a given flowmeter corresponding to
+5 VDC. For example, if one has a 100 sccm flowmeter with a displayed flow of 35 sccm, then the
output voltage will be:
Conversely, if the analog output voltage is measured, then the gas flow reading is given by:
V5
OUT
V5
OUT
GAS FLOW
DISPLAY GAS FLOW
*
FLOWMETER RANGE
35 sccm
*
100 sccm
V (V)*FLOWMETER RANGE
OUT
1.75 Volts
5
page 11
Page 12
Internal Adjustments
The internal adjustments are factory set and should not be changed. However, if the need arises
and suitable standards are available the following adjustments can be made. (see figure on page 3-4)
Full ScaleFull Scale
Full Scale
Full ScaleFull Scale
The full scale potentiometer changes the display by changing the sensitivity of the instrument.
Apply a calibrated +5 VDC signal to the
supply. The display of the instrument should read “100.0”, “10.00” or “1.000”depending upon the
configuration. Adjust the full scale potentiometer (R11) if needed.
Analog OutputAnalog Output
Analog Output
Analog OutputAnalog Output
The analog output potentiometer changes the analog output voltage by changing the gain on the
output buffer amplifier. Apply a calibrated pressure standard to the transducer and refer to
equation (1). Adjust the potentiometer (R18) as needed.
Notes on internal adjustments:
*Both the full scale reading and the analog output voltage can change by only ± 10%.
** Zero adjustment can only be performed on the flowmeter itself. Consult your flowmeters
manual for making these adjustments.
Sig HiSig Hi
Sig Hi and
Sig HiSig Hi
Sig LoSig Lo
Sig Lo pins on the back of the power
Sig LoSig Lo
Display ConfigurationDisplay Configuration
Display Configuration
Display ConfigurationDisplay Configuration
The jumper position on the power supply circuit board determines the location of the decimal on
the display. The jumper is factory set to match the configuration specified by the customer. Should
the need arise to change the display configuration, this jumper can be easily moved (see figure on
page 13).
page 12
Page 13
configuration jumpers
Jumper LocationJumper Location
Jumper Location
Jumper LocationJumper Location
11
1
11
44
4
44
33
3
33
22
2
22
Display ConfigurationDisplay Configuration
Display Configuration
Display ConfigurationDisplay Configuration
100.0100.0
100.0
100.0100.0
not usednot used
not used
not usednot used
1.0001.000
1.000
1.0001.000
10.0010.00
10.00
10.0010.00
page 13
Page 14
page 14
Page 15
Warranty Policy
Hastings Instruments warrants this product for a period of one year from the date of shipment to
be free from defects in material and workmanship. This warranty does not apply to defects or
failures resulting from unauthorized modification, misuse or mishandling of the product. This
warranty does not apply to batteries or other expendable parts, nor to damage caused by leaking
batteries or any similar occurrence. This warranty does not apply to any instrument which has had
a tamper seal removed or broken.
This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty
as to fitness for a particular use. Hastings Instruments shall not be liable for any indirect or consequential damages.
Hastings Instruments will, at its option, repair, replace or refund the selling price of the product if
Hastings Instruments determines, in good faith, that it is defective in materials or workmanship
during the warranty period. Defective instruments should be returned to Hastings Instruments
together with a written statement of the problem and a Return Material Authorization (RMA)
number. Please consult the factory for your RMA number before returning any product for repair.
SECTION 4SECTION 4
SECTION 4
SECTION 4SECTION 4
Warranty and Repair
Non-Warranty Repair Policy
Any product returned for a non-warranty repair must be accompanied by a purchase order, RMA
form and a written description of the problem with the instrument. If the repair cost is higher, you
will be contacted for authorization before we proceed with any repairs. If you then choose not to
have the product repaired, a minimum will be charged to cover the processing and inspection.
Please consult the factory for your RMA number before returning any product for repair.
TELEDYNE HASTINGS
804 NEWCOMBE AVENUE
HAMPTON, VIRGINIA 23669 U.S.A.
ATTENTION: REPAIR DEPARTMENT
TELEPHONE(757) 723-6531
FAX(757) 723-3925
1-800-950-2468
page 15
Page 16
page 16
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.