Unauthorized duplication of Teledyne LeCroy documentation materials other than for internal sales and
distribution purposes is strictly prohibited. However, clients are encouraged to distribute and duplicate
Teledyne LeCroy documentation for their own internal educational purposes.
HDO6000 and Teledyne LeCroy are trademarks of Teledyne LeCroy, Inc. Other product or brand names are
trademarks or requested trademarks of their respective holders. Information in this publication supersedes
all earlier versions. Specifications are subject to change without notice.
Setting Up the Oscilloscope3
Powering On/Off5
Software Activation5
Inputs/Outputs6
Front Input/Output Panel6
Side Input/Output Panel6
Back Input/Output Panel7
Analog Inputs7
Probes7
Digital Inputs8
Touch Screen9
Menu Bar9
Signal Display Grid10
Descriptor Boxes11
Dialogs13
Shortcut Toolbar13
Control Application Window13
Enter/Select Data14
Annotate Traces16
Print Screen17
Screen Saver17
Front Panel18
Front Panel Trigger Controls19
Front Panel Horizontal Controls19
Front Panel Vertical Controls20
Front Panel Math, Zoom, and Mem(ory) Buttons20
Front Panel Cursor Controls20
Front Panel Adjust and Intensity Controls20
Miscellaneous Front Panel Controls21
Turn On/Off Traces22
Analog Traces22
Digital Traces22
Other Traces22
Activate Trace22
Zooming Waveforms23
Create Zoom23
922499 Rev B
i
HDO6000 High Definition Oscilloscope
Turn off Zoom24
Quick Zoom24
Multi-Zoom24
Zoom Controls24
Rescale Memory or Math Function Traces26
Multi-Zoom26
Display Setup49
Moving Traces from Grid to Grid49
XY Displays50
Persistence Overview50
Cursors53
Cursor Types53
Cursor Settings54
Measure55
Measure Gate55
Level and Slope55
Set Up Measurement Parameter56
List of Standard Parameters57
Quick Measurements60
View Statistics61
View Histicon61
ii
922499 Rev B
Operator's Manual
Help Markers62
Qualified Parameters63
Math on Parameters64
Calculating Measurements66
Math69
Single vs. Dual Operation Functions69
Graphing69
Set Up Math Function70
Enable/Disable Math Function71
List of Math Functions71
Interpolation73
Sparser Function74
Rescaling and Assigning Units74
Enhanced Resolution76
Averaging Waveforms78
FFT80
Copy Function82
Analysis83
View Histogram83
View Persistence Histogram84
Track and Trend85
WaveScan87
History Mode92
Pass/Fail Testing93
View Configurations96
Utilities97
Utilities Settings97
System Status98
Remote Control Settings98
Print (Hardcopy) Settings100
Auxiliary Output Settings102
Date/Time Settings103
Disk Utilities104
Preferences Settings105
Calibration Settings106
Acquisition Settings107
Color Settings108
E-Mail108
Miscellaneous Settings109
Save/Recall Overview110
Save/Recall Setups110
Save/Recall Waveforms111
Save Table Data114
Cleaning122
Calibration122
Touch Screen Calibration122
Language Selection122
Add Software Option123
X-Stream Firmware Update124
HDO System Recovery125
Technical Support127
Returning a Product for Service128
Contact Teledyne LeCroy129
Certifications130
EMC Compliance130
Safety Compliance131
Environmental Compliance132
ISO Certification132
Warranty132
Windows License Agreement133
iv
922499 Rev B
Operator's Manual
Welcome
Thank you for purchasing a Teledyne LeCroy High Definition Oscilloscope. We're certain you'll be pleased
with the detailed features so unique to our instruments.
The manual is arranged in the following manner:
Safety contains important precautions and information relating to power and cooling.
The sections from Start Up through Maintenance cover everything you need to know about the operation
and care of the oscilloscope.
Documentation for compatible software options is available from the Teledyne LeCroy website at
teledynelecroy.com. Our website maintains the most current product specifications and should be
checked for frequent updates.
Remember...
When your product is delivered, verify that all items on the packing list or invoice copy have been shipped
to you. Contact your nearest Teledyne LeCroy customer service center or national distributor if anything is
missing or damaged. We can only be responsible for replacement if you contact us immediately.
Thank You
We truly hope you enjoy using Teledyne LeCroy's fine products.
Sincerely,
David C. Graef
Teledyne LeCroy
Vice President and Chief Technology Officer
922499 Rev B
v
HDO6000 High Definition Oscilloscope
vi
922499 Rev B
Operator's Manual
Safety Instructions
Observe these instructions to keep the instrument operating in a correct and safe condition. You are
required to follow generally accepted safety procedures in addition to the precautions specified in this
section. The overall safety of any system incorporating this instrument is the responsibility of the
assembler of the system.
Symbols
These symbols may appear on the instrument's front or rear panels and in its documentation to alert you
to important safety considerations:
CAUTION of potential damage to instrument, or WARNING of potential bodily injury. Do not
proceed until the information is fully understood and conditions are met.
High voltage. Risk of electric shock or burn.
Measurement ground connection.
Safety (protective) ground connection.
Alternating Current.
Standby Power (front of instrument).
Precautions
Use proper power cord. Use only the power cord shipped with this instrument and certified for the country
of use.
Maintain ground. This product is grounded through the power cord grounding conductor. To avoid electric
shock, connect only to a grounded mating outlet.
Connect and disconnect properly. Do not connect/disconnect probes or test leads while they are
connected to a voltage source.
Observe all terminal ratings. Do not apply a voltage to any input (C1, C2, C3, C4 or EXT) that exceeds the
maximum rating of that input. Refer to the front of the oscilloscope for maximum input ratings.
Use only within operational environment listed. Do not use in wet or explosive atmospheres.
Use indoors only.
Keep product surfaces clean and dry. See Cleaning in the Maintenance section.
Do not block the cooling vents. Leave a minimum six-inch (15 cm) gap between the instrument and the
nearest object. Keep the underside clear of papers and other objects.
Do not remove the covers or inside parts. Refer all maintenance to qualified service personnel.
Do not operate with suspected failures. Do not use the product if any part is damaged. Obviously incorrect
measurement behaviors (such as failure to calibrate) might indicate impairment due to hazardous live
electrical quantities. Cease operation immediately and sequester the instrument from inadvertent use.
922499 Rev B
1
HDO6000 High Definition Oscilloscope
Operating Environment
Temperature: 5 to 40° C.
Humidity: Maximum relative humidity 90 % for temperatures up to 31° C, decreasing linearly to 50%
relative humidity at 40° C.
Altitude: Up to 3,048 m (10,000 ft) at or below 30° C.
Cooling
The instrument relies on forced air cooling with internal fans and vents. Take care to avoid restricting the
airflow to any part. Around the sides and rear, leave a minimum of 15 cm (6 inches) between the
instrument and the nearest object.The feet (up or down) provide adequate bottom clearance.
CAUTION. Do not block cooling vents. Always keep the area beneath the instrument clear of
paper and other items.
The instrument also has internal fan control circuitry that regulates the fan speed based on the ambient
temperature. This is performed automatically after start-up.
Power
AC Power
The instrument operates from a single-phase, 100 to 240 Vrms (± 10%) AC power source at 50/60/400 Hz
(± 10%). Manual voltage selection is not required because the instrument automatically adapts to the line
voltage.
Power Consumption
Maximum power consumption with all accessories installed (e.g., active probes, USB peripherals, digital
leadset) is 320 W (320 VA). Power consumption in standby mode is 4 W.
Ground
The AC inlet ground is connected directly to the frame of the instrument. For adequate protection again
electric shock, connect to a mating outlet with a safety ground contact.
WARNING. Only use the power cord provided with your instrument. Interrupting the protective
conductor inside or outside the oscilloscope, or disconnecting the safety ground terminal,
creates a hazardous situation. Intentional interruption is prohibited.
2
922499 Rev B
Start Up
Setting Up the Oscilloscope
Checking Shipment
Verify that all items on the materials list below have been shipped to you:
l 1 oscilloscope
l 4 passive probes (one for each channel)
l 1 AC line (power) cord rated for country of use
l 1 protective front cover
l 1 Getting Started Guide
l 1 Oscilloscope Security Certificate
l 1 Oscilloscope Registration Card
l 1 Calibration Document
Operator's Manual
Mixed-signal (-MS) model oscilloscopes also ship with:
l 1 digital leadset
l 5 flying ground leads
l 20 ground extenders
l 22 XL microgrippers
Contact your nearest Teledyne LeCroy customer service center or national distributor if anything is
missing or damaged. We can only be responsible for replacement if you contact us immediately.
Carrying and Placing the Oscilloscope
The oscilloscope’s case contains a built-in carrying handle. Lift the handle away from the oscilloscope
body, grasp firmly and lift the instrument. Always unplug the instrument from the power source before
lifting and carrying it.
Place the instrument where it will have a minimum 15 cm (6 inch) clearance from the nearest object. Be
sure there are no papers or other debris beneath the oscilloscope or blocking the cooling vents.
CAUTION. Do not place the instrument so that it is difficult to reach the power cord in case you
need to quickly disconnect from power.
922499 Rev B
3
HDO6000 High Definition Oscilloscope
Positioning the Feet
The HDO is equipped with rotating, tilting feet to allow four different viewing positions.
To tilt the body back slightly for bench top viewing, pull the small flaps on
the bottom of the feet away from the body of the oscilloscope.
To tilt the body forward, rotate both feet to the back. This position is
useful when placing the oscilloscope on a high shelf. Pulling out the flaps
in this position increases the angle of the tilt.
Connecting to Other Devices/Systems
Make the desired cable connections. All except for the power connection are optional.
After start up, configure the connection on the oscilloscope using the menu options listed below. More
detailed instructions are provided later in this manual.
POWER
Connect the line cord rated for your country to the AC power inlet on the back of the instrument, then plug
it into a grounded AC power outlet. (See Power and Ground Connections in General Safety Information.)
LAN
Connect a cable from either Ethernet port on the side panel to a network access device. On the
oscilloscope, use the standard Windows Network dialog to configure the network connection. Go to
Utilities > Preference Setup > Email to configure email settings.
USB PERIPHERALS
Connect the device to a USB port on the front or side of the instrument. Go to Utilities > Utilities Setup >
Hardcopy to configure printer settings.
EXTERNAL MONITOR
Connect the monitor cable to a video output on the side of the instrument (VGA, DVI, and HDMI are all
supported). Go to Display > Display Setup > Open Monitor Control Panel to configure the display settings.
4
922499 Rev B
Operator's Manual
EXTERNAL CONTROLLER
Connect a USB-A/B cable from the USBTMC port on the back of the instrument to the controller. Go to
Utilities > Preference Setup > Remote to configure remote control.
OTHER INSTRUMENT (FOR REFERENCE CLOCK)
Connect a BNC cable from Ref In/Out on the back of the oscilloscope to the other instrument. Go to
Timebase > Horizontal Setup > Reference Clock to configure the clock.
OTHER AUXILIARY DEVICE
Connect a BNC cable from Aux Out on the back of the instrument to the other device. Go to Utilities >
Utilities Setup > Aux Output to configure the output.
Powering On/Off
The Standby Power button at the lower, left front of the oscilloscope controls the operational state
of the instrument.
Press the button to switch the instrument into Standby mode (reduced power); press it again to return to
full operation.
CAUTION. Do not change the instrument’s Windows Power Options setting from the default
Never to System Standby or System Hibernate. Doing so can cause the system to fail.
CAUTION. Do not power on or calibrate the oscilloscope with a signal attached.
Always use the File > Shutdown menu option to execute a proper shut down process and preserve
settings before powering down. Pressing and holding the Standby button will execute a “hard” shutdown,
the same as on a computer, but we do not recommend doing this because it does not allow the Windows
operating system to shut down properly. Do not power off by pulling the power cord from the socket or
shutting off a connected power strip without first shutting down properly.
The Standby button does not disconnect the oscilloscope from the AC power supply. The only way to fully
power down the instrument is to unplug the AC power cord from the outlet.
We recommend unplugging the instrument if it will be unused for a long period of time.
Software Activation
The oscilloscope operating software (firmware and standard applications) is active upon delivery. Upon
power-up, the oscilloscopes loads the software automatically.
Firmware
Free firmware updates are available periodically from the Teledyne LeCroy website at
teledynelecroy.com/support/softwaredownload. Registered users can receive an email notification when
a new update is released. Follow the instructions on the website to download and install the software.
Purchased Options
If you decide to purchase an option, you will receive a license key via email that activates the optional
features on the oscilloscope. See Add a New Software Option.
922499 Rev B
5
HDO6000 High Definition Oscilloscope
Inputs/Outputs
Front Input/Output Panel
A. The Power button turns on/off the oscilloscope.
B. BNC connectors for analog input on Channels 1–4 (or 1–2 depending on model), and EXT for con-
necting an external trigger device.
C. Mixed signal interface for digital inputs.
D. Ground and calibration output terminals are used to compensate passive probes.
E. Two (2) front-mounted host USB ports can be used for transferring data or connecting peripherals
such as a mouse or keyboard.
Side Input/Output Panel
A. Video Output VGA, DVI, and HDMI ports connect the oscilloscope to external
monitors.
B. USB Ports (4) allow you to connect external USB devices, such as storage
drives.
C. Ethernet Ports (2) connect the oscilloscope to networks.
D. Audio Input/Output Mic, Speaker, and Line-In jacks connect the oscilloscope to
external audio devices.
6
922499 Rev B
Operator's Manual
Back Input/Output Panel
A. Aux Out connector sends device trigger enabled, trigger out, or pass/fail output to another device.
B. Ref In/Out connector allows you to input an external Reference Clock, or to output a Reference Clock
to another instrument.
C. USBTMC Port enables remote control of the oscilloscope.
D. AC Power Inlet connects the AC line cord.
See the general set up instructions for more information about configuring connections to other devices.
Analog Inputs
A series of BNC connectors arranged on the front and back of the oscilloscope are used to input analog
signal on Channels 1-4 and AUX In, or an external trigger pulse on EXT.
HDO connectors use the ProBus interface and are compatible with any Teledyne LeCroy ProBus type
probes rated for the oscilloscope's bandwidth.
The ProBus interface contains a 6-pin power and communication connection and a BNC signal connection
to the probe. It offers both 50 Ω/1 MΩ input impedance and provides probe power and control for a wide
range of probes such as high impedance passive probes, high impedance active probes, current probes,
high voltage probes, and differential probes. ProBus also includes sense rings for detecting passive
probes. The ProBus interface may also have a BNC-terminated cable connected directly to it.
The interfaces power probes and completely integrate the probe with the oscilloscope channel. Upon
connection, the probe type is recognized and some setup information, such as input coupling and
attenuation, is performed automatically. This information is displayed on the Channel Probe Dialog.
System (probe plus oscilloscope) gain settings are automatically calculated and displayed based on the
probe attenuation.
Probes
HDO6000 oscilloscopes are compatible with the included passive probes and all Teledyne LeCroy ProBus
active probes that are rated for the oscilloscope’s bandwidth. Probe specifications and documentation
are available at teledynelecroy.com/hdo6000.
The passive probes supplied with your oscilloscope are matched to the input impedance of the
instrument, but may need further compensation; refer to the probe manual for the procedure. If using
other passive probes with your oscilloscope, be sure to perform a low frequency calibration using the Cal
signal available from the HDO's front panel before using them to measure signal.Follow the directions in
the probe instruction manual to compensate the low and/or high frequency response of the probes.
922499 Rev B
7
HDO6000 High Definition Oscilloscope
Digital Inputs
Standard with all HDO6000-MS model oscilloscopes, the
digital leadset enables input of up-to-16 lines of digital
data. Lines can be organized into four logical groups and
can be named appropriately.
The digital leadset features two digital banks with separate
threshold and hysteresis controls, making it possible to
simultaneously view data from different logic families.
Connecting/Disconnecting the Leadset
To connect the leadset to the oscilloscope, push the connector into the mixed signal interface below the
front panel until you hear a click.
To remove the leadset, press in and hold the buttons on each side of the connector, then pull out to
release it.
Grounding Leads
Each flying lead has a signal and a ground connection. A variety of ground extenders and flying ground
leads are available for different probing needs.
In order to achieve optimal signal integrity, you should connect the ground at the tip of the flying lead for
each input used in your measurements. Use either the provided ground extenders or ground flying leads to
make the ground connection.
8
922499 Rev B
Operator's Manual
Touch Screen
The touch screen is the principal viewing and control center of the oscilloscope. The entire display area is
active: use your finger or the stylus to touch, double-touch, touch-and-drag, touch-and-hold (right click)
or draw a selection box. Many controls that display information also work as “buttons” to access other
functions.
If you have a mouse installed, you can click anywhere you can touch to activate a control; in fact, you can
alternate between clicking and touching, whichever is convenient for you.
The touch screen is divided into the following major control groups:
Menu Bar
The top of the screen contains a complete menu of oscilloscope functions. Making a selection here
changes the dialogs displayed at the bottom of the screen.
Many common oscilloscope operations can also be performed from the Front Panel or launched via the
Descriptor Boxes. However, the menu bar is the best way to access dialogs for Save/Recall (File)
functions, Display functions, Status, LabNotebook, Pass/Fail setup, and Utilities/Preferences setup.
If an action can be “undone” (such as a zoom/rescale of a trace), a smallUndo button appears at
the far right of the menu bar. Click this to return to the previous oscilloscope display.
922499 Rev B
9
HDO6000 High Definition Oscilloscope
Signal Display Grid
The grid area displays the waveform traces. It is sectioned into 10 Horizontal (Time) divisions and 8
Vertical (Voltage) divisions.
Multiple Grid Display
You can divide the display to simultaneously view multiple traces in different grids. By default, the
oscilloscope has Auto Grid enabled. This divides the display into additional grids each time a new trace is
opened, up to 16 grids for simultaneous viewing.
There are Display menu options to show all traces on a Single Grid, or to manually divide the display into
different grid sizes and formats. When you manually divide the display, zooms and measurement markers
appear on the same grid as the source channel, while math and memory traces appear in new grids until
none are available.
Manually move traces from grid to grid by activating the trace and touching the Next Grid shortcut button.
Of special note, you can also move a trace to another grid by dragging its descriptor box to the desired
grid.
Different types of traces opening in separate grids.
Adjusting Grid Brightness
You can adjust the brightness of the grid lines to make either the grid or traces more visible. Go to
Display > Display Setup and enter a new Grid Intensity percentage. The higher the number, the brighter
and bolder the grid lines.
Grid Indicators
These indicators appear over the grid to mark important points on the display. They are matched to the
color of the trace to which they apply.
Trigger Position - A small triangle along the bottom (horizontal) edge of the grid shows the time
the oscilloscope is set to trigger an acquisition. Unless Delay is set, this indicator is at the zero
(center) point of the grid. Trigger Delay is shown at the top right of the Timebase descriptor box.
10
922499 Rev B
Operator's Manual
Pre/Post-trigger Delay - A small arrow to the bottom left or right of the grid indicates that a preor post-trigger Delay has shifted the Trigger Position indicator to a point in time not displayed
on the grid. All trigger Delay values are shown on the Timebase Descriptor Box.
Trigger Level - This small triangle at the right edge of the grid tracks the trigger voltage level. If
you change the trigger level when in Stop trigger mode, or in Normal or Single mode without a
valid trigger, a hollow triangle of the same color appears at the new trigger level. The trigger
level indicator is not shown if the triggering channel is not displayed.
Zero Volts Level - This indicator is located at the left edge of the grid. One appears for each
open trace on the grid, sharing the number and color of the trace.
Various Cursor lines appear over the grid to indicate specific voltage and time values on the
waveform. Touch-and-drag cursor indicators to quickly reposition them.
Signal Display Grid Pop-Up Menu
Touching/clicking a trace opens a pop-up menu with shortcuts to the appropriate trace setup dialog, or
the Math and Measure setup dialogs. You can also use it to turn off the trace or place an annotation label
on it.
Descriptor Boxes
Shown just beneath the grid display, these boxes provide a summary of your channel, timebase and
trigger settings. They also act as convenient navigation tools.
Descriptor boxes appear when a trace is turned on. Touch the descriptor box once to activate the
corresponding trace. Touch the descriptor box a second time to open its corresponding setup dialog.
When a trace is active, its corresponding descriptor Box is shown highlighted, and Front Panel controls
will work for that trace.
Highlighted descriptor box (left) is active. Controls will work for this trace.
Channel Descriptor Box
Channel trace descriptor boxes correspond to analog signal inputs. They show
Vertical settings and any cursor selection: (clockwise from top left) Trace Number
(Cx), Pre-Processing List (summarizes changes from default state), Coupling, Gain
Setting, Offset Setting, and Averaging Sweeps Count.
Codes are used to indicate pre-processing that has been applied to the input. The codes have a long and
short form. When several processes are in effect, the short form is used.
922499 Rev B
11
HDO6000 High Definition Oscilloscope
Preprocessing Symbols on Descriptor Boxes
Pre-Processing TypeLong FormShort Form
Sin X InterpolationSINXS
AveragingAVGA
InversionINVI
DeskewDSQDQ
CouplingDC50, DC1M or AC1MD50, D1M, or A1
GroundGNDG
Bandwidth LimitingBWLB
Similar descriptor boxes appear for zoom (Zx), math (Fx), and memory (Mx) traces. These descriptor
boxes show any Horizontal scaling that differs from the signal Timebase.
Digital Descriptor Box
Digital descriptor boxes appear whenever a digital line group is enabled on a mixedsignal model oscilloscope. Like Channel descriptors, they are numbered 1-4
corresponding to one of the four line groups.They show the number of digital lines in
the group, the digital sample rate, and the digital memory.
Timebase Descriptor Box
The TimeBase descriptor box shows: (clockwise from top right) Trigger Delay
(position), Time/div, Sample Rate, Number of Samples, and Sampling Mode (blank
when in real-time mode).
Trigger Descriptor Box
Trigger descriptor box shows: (clockwise from top right) Trigger Source and Coupling,
Trigger Level (V), Slope, Trigger Type, Trigger Mode.
Setup information for Horizontal cursors, including the time between cursors and the
frequency, is shown beneath the TimeBase and
Trigger descriptor boxes. See the Cursors section for more information.
12
922499 Rev B
Operator's Manual
Dialogs
Dialogs appear at the bottom of the display for entering setup data. The top dialog will be the main entry
point for the selected setup option. For convenience, related dialogs appear as a series of tabs behind the
main dialog. Touch the tab to open the dialog.
Dialogs may also display right-hand dialogs (sub-tabs) or pop-up dialogs. These often change depending
on the other selections made on the left-hand dialog.
Many dialog settings can be made using either the touch screen or the Front Panel buttons.
Shortcut Toolbar
Several setup dialogs contain a row of buttons at the bottom of the dialog. These provide a shortcut to
common functions without having to leave the underlying set up dialog.
Measure - Opens the Measure pop-up to set measurement parameters on the active trace.
Zoom - Creates a zoom trace of the active trace.
Math - Opens the Math pop-up to apply math functions to the active trace and create a new math trace.
Decode - Opens the main Serial Decode dialog where serial data decoders can be configured and applied.
Store - Loads the active trace into the corresponding memory location (C1, F1 and Z1 to M1; C2, F2 and
Z2 to M2, etc.).
Find Scale - Automatically performs a vertical scaling that fits the waveform into the grid.
Next Grid - Automatically moves the active trace to the next grid. If you have only one grid displayed, a
new grid will be created automatically, and the trace moved.
Label - Opens the Label pop-up to annotate the active trace.
The following buttons appearing at the bottom of the Measure (Px) dialogs. They allow you to create a
Math function to draw the corresponding type of plot (Histogram, Trend, or Track) while remaining on the
Measure setup dialog.
Control Application Window
The oscilloscope applications runs on a Windows Embedded Standard 7P Operating System and functions
exactly as do other Windows applications. The application software loads automatically when you turn on
the oscilloscope using the Power button.
To minimize the application window and show the Windows desktop, touch the minimize button or choose
File > Minimize. To restore the window after minimizing, touch the oscilloscope display icon in the lower
right corner of the desktop.
To exit the application window, choose File > Exit. When you exit the application, the oscilloscope
operating system continues to run. To reload the application after exiting, touch the Start DSO desktop
shortcut.
922499 Rev B
13
HDO6000 High Definition Oscilloscope
To restart Windows (reboot the oscilloscope), choose File > Shutdown. Wait 10 seconds then press the
Power button on the front of the oscilloscope.
Enter/Select Data
Touch & Type
Touching once activates a control. In some cases, you’ll immediately see a pop-up menu of options.
Touch one to select it.
In other cases, data entry fields appear highlighted on the display. When a data entry
field is highlighted, it is active and can be modified by using the Front Panel Adjust knob.
If you have a keyboard installed, you can type your entry in the active field. Or, you can
touch again, then select your entry from the pop-up menu or keypad.
You’ll see a pop-up keypad when you double-touch a numerical data entry field. Touch the soft keys to
use it exactly as you would a calculator. When you touch OK, the calculated value is entered in the field.
Touch & Drag
Touch-and-drag waveforms, cursors, and trigger indicators
to reposition them on the grid; this is the same as setting
the values on the dialog.
Quickly zoom areas of the grid by touching and dragging to
draw a selection box around a portion of the trace.
14
922499 Rev B
Touch & Swipe
Stylus
Operator's Manual
Touch and swipe the screen in an up or down direction to
scroll long lists of values. You can also use scroll bars or
Up/Down arrow keys to navigate to the desired value.
Use the stylus when you want a more precise selection tool
than your finger. It is especially helpful for selecting exact
areas of the grid or values that lie close together on pop-up
menus.
922499 Rev B
15
HDO6000 High Definition Oscilloscope
Annotate Traces
The Label function gives you the ability to add custom annotations to traces that are
shown on the display. Labels are numbered sequentially in the order they were
created. Once placed, labels can be moved to new positions and turned on/off.
Create Label
1. Touch the trace, then choose Set label... from the pop-up menu.
OR
Touch the trace descriptor box twice, then touch the Label shortcut button on the setup dialog.
2. On the Trace Annotation pop-up, touch Add Label.
3. Enter the Label Text.
4. Optionally, enter the Horizontal Pos. and Vertical Pos. at which to place the label. The default position
is 0 ns horizontal. You can optionally check Use Trace Vertical Position instead of entering a Vertical
Pos.
5. Close the dialog.
Edit/Remove Label
1. Touch the label and choose Set Label from the pop-up menu.
2. Select the Label number. You can use the Up/Down arrow keys to scroll the list.
3. Change the Label Text and/or Horizontal Pos., or touch Remove Label to delete it.
4. Close the dialog.
Turn On/Off Labels
After labels have been placed on a grid, you can turn on/off all labels at once by opening the Trace
Annotation dialog and selecting/deselecting the View labels checkbox.
16
922499 Rev B
Operator's Manual
Print Screen
Print captures an image of the display and outputs it according to your Hardcopy settings, which may be
to send it to a network printer, e-mail it, save it to a file, or copy it to the clip-board to paste into another
application.
There are three ways to print:
l Touch the Front Panel Print button.
l Choose File > Print.
l Choose Utilities > Utilities Setup > Hardcopy tab and touch the Print button to the far right of the
dialog.
NOTE: The Front Panel Print button can be configured to capture the screen as a LabNotebook entry. In
this case, only the File and Utilities menu print options will function according to your Hardcopy setup.
Screen Saver
The screen saver is activated the same as on any Windows PC. Minimize the instrument display by
choosing File → Minimize from the menu bar. Then, open the Windows ControlPanel and change
Appearance and Personalization settings.
Touch the oscilloscope icon at the bottom right of the desktop to restore the instrument display.
922499 Rev B
17
HDO6000 High Definition Oscilloscope
Front Panel
Most Front Panel controls duplicate functionality available
through the touch screen display and are described on the
following pages.
Shortcut buttons arranged across the top of the Front Panel
give quick access to commonly used functions. Other
shortcut buttons arranged across the bottom open special
applications.
All the knobs on the Front Panel function one way if turned
and another if pushed like a button. The top label describes
the knob’s principal “turn” action, and the bottom label
describes its “push” action.
Front panel buttons light up to indicate which traces and
functions are active. Actions performed from the Front
Panel always apply to the active trace.
18
922499 Rev B
Operator's Manual
Front Panel Trigger Controls
Level knob- Changes the trigger threshold level (V). The
number is shown on the Trigger descriptor box. Pushing the
knob sets the trigger level to the 50% point of the input
signal.
READY and TRIG'D Indicators - The READY indicator is lit
when the trigger is armed. TRIG'D is lit momentarily when a
trigger occurs. A fast trigger rate causes the light to stay lit
continuously.
Setup - CCorresponds to the menu selection Trigger → Trigger Setup. Press it once to open the Trigger
Setup dialog and again to close the dialog.
Auto - Sets Auto trigger mode, which triggers the oscilloscope after a time-out, even if the trigger
conditions are not met.
Normal - Sets Normal trigger mode, which triggers the oscilloscope each time a signal is present that
meets the conditions set for the type of trigger selected.
Single - Sets Single trigger mode, which arms the oscilloscope to trigger once (single-shot acquisition)
when the input signal meets the trigger conditions set for the type of trigger selected. If the scope is
already armed, it will force a trigger.
Stop - Prevents the scope from triggering on a signal. If you boot up the instrument with the trigger in
Stop mode, a "No trace available" message is shown. Press the Auto button to display a trace.
Front Panel Horizontal Controls
The Horizontal Front Panel group corresponds to the Timebase dialog.
Delay knob - Turn to change the Trigger Delay value (S). Push the knob to reset Delay to
zero.
Horizontal Adjust knob - If the trace source is an input channel, turn this knob to set the
Time/division (S) of the oscilloscope acquisition system. The value is shown on the
Timebase descriptor box. When using this control, the oscilloscope allocates memory as
needed to maintain the highest sample rate possible for the timebase setting. If the
trace source is a zoom, memory or math function, turn the knob to change the horizontal
scale of the trace, effectively "zooming" in or out. The value is shown on the
corresponding descriptor box. Push the knob to change the setting in fine increments;
push it again to return to 1, 2, 5, 10 step increments.
922499 Rev B
19
HDO6000 High Definition Oscilloscope
Front Panel Vertical Controls
Channel buttons- Turn on a channel that is off, or activate a channel that is
already on. When the channel is active, pushing its channel button turns it off. A
lit button shows the active channel (here, C2 is lit).
Offset knob - Adjusts the zero level of the trace (this makes it appear to move up
or down relative to the center axis of the grid). The value appears on the trace
descriptor box. Push it to reset Offset to zero.
Gain knob- Sets Vertical Gain (V/div). The value appears on the trace descriptor
box. Push it once to adjust V/div in fine increments; push it again to adjust in 1, 2,
5, 10-step increments.
Dig button - Enables digital input on -MS models.
Front Panel Math, Zoom, and Mem(ory) Buttons
The Zoom button creates a quick zoom for each open channel trace. The resulting zoom trace
(s) will be 1/10 of the channel timebase and centered on the display. Touch the zoom trace
descriptor box to display the zoom controls.
The Math and Mem(ory) buttons open the corresponding setup dialogs.
If a Zoom, Math or Memory trace is active, the button illuminates to indicate that the Vertical
and Horizontal knobs will now control that trace.
Front Panel Cursor Controls
Cursors identify specific voltage and time values on the waveform. The white
cursor lines help make these points more visible, as well as provide a simple way
to reposition them. A readout of the values appears on the trace descriptor box.
There are five preset cursor types, each with a unique appearance on the display:
Horizontal (Time), Horizontal + Vertical, Vertical (Amplitude), Horizontal
(Frequency), and Horizontal (Event). These are described in more detail in the Cursors section.
Type - Press to apply or remove cursors. Continue pressing to cycle through all cursor types until the
desired type is found ("no cursors" will appear in the cycle).
Cursor knob - Turn to reposition the selected cursor line. Push to select a different cursor line to adjust.
Front Panel Adjust and Intensity Controls
The Adjust knob changes the value in any highlighted data entry field when
turned. Pushing the Adjust knob toggles between coarse (large increment) or fine
(small increment) adjustments when the knob is turned.
The Intensity button sets the Adjust knob to control the trace intensity. When
more data is available than can actually be displayed, the Intensity button helps
to visualize significant events by applying an algorithm that dims less frequently occurring samples. This
feature can also be accessed from the Display > Display Setup dialog.
20
922499 Rev B
Operator's Manual
Intensity 40% (left) dims samples that occur ≤ 40% of the time to highlight the more frequent samples, vs.
intensity 100% (right) which shows all samples at the same intensity.
Miscellaneous Front Panel Controls
Top Row
Auto Setup - Performs an Auto Setup. After the first press, you will be prompted for a confirmation. Press
the button again or use the touch screen to confirm.
Default Setup - Resets the oscilloscope to the factory default configuration.
Print - Captures the entire screen and outputs it according to your Hardcopy settings. It can also be
configured to output a LabNotebook entry.
Touch Screen - Enables or disables touch screen functionality.
Clear Sweeps - Resets the acquisition counter and any cumulative measurements.
Bottom Row
Decode - Opens the Serial Decode dialog if you have serial data decoder options installed.
WaveScan - Opens the WaveScan dialog.
Spectrum - Opens the Spectrum Analyzer dialog if you have that option installed.
History - Opens the History Mode dialog.
922499 Rev B
21
HDO6000 High Definition Oscilloscope
Turn On/Off Traces
Analog Traces
From the display, choose Vertical > Channel <#> Setup to turn on the trace. To turn it off, clear the Trace
On checkbox on the corresponding Channel dialog, or right-click on the descriptor box and choose Off.
From the Front Panel, press the Channel button (1-4) to turn on the trace; press again to turn it off.
NOTE: The default is to display each trace in its own grid . Use the Display menu to change how traces
are arranged.
Digital Traces
From the display, choose Vertical > Digital <#> Setup.
From the Front Panel, press the Dig button, then check Group on the Digital<#> trace dialog. Clear Group
to turn off the trace.
Other Traces
You can quickly create zoom or math traces without leaving the setup dialogs by touching the Zoom or
Math shortcut button at the bottom of the dialog.
You can also use the Front Panel Zoom, Math, or Mem(ory) buttons to quickly create traces. The Zoom
control automatically creates zoom trace(s) that are 1/10 of the original waveform(s). The middle of the
grid is used as the center of the zoom trace.
Activate Trace
A trace descriptor box appears on the display for each enabled channel, digital, zoom, math, or memory
trace. Touch this box at any time to activate the trace and open its setup dialog. A highlighted descriptor
box indicates the active trace to which all actions apply.
Active trace descriptor (left), inactive trace descriptor (right).
Although several traces may be open and appear on the grid, only one at a time is active. Whenever you
activate a trace , the dialog at the bottom of the screen automatically switches to the appropriate setup
dialog for that trace. The tab at the top of the dialog shows to which trace it applies.
The Zoom function magnifies a selected region of a trace. On HDO6000 model oscilloscopes, you can
display up to eight zoom traces (Z1 - Z8) taken from any channel, math, or memory trace.
You can also use the Multi-Zoom Math function to create time-locked zoom traces for selected
waveforms. For more information, refer to Multi-Zoom).
Create Zoom
To create a zoom, touch -and-drag to draw a selection box around any part of the source waveform.
Selected portion of trace.
The zoom will resize the selected portion to fit the full width of the grid. The degree of vertical and
horizontal magnification, therefore, depends on the size of the rectangle that you draw.
The zoom opens in a new grid, or the next empty grid, with the zoomed portion of the source trace
highlighted. If there are no more available grids, zooms will open in the same grid as the source trace.
922499 Rev B
Zoomed area of original trace highlighted.
23
HDO6000 High Definition Oscilloscope
New zooms are turned on and visible by default. However, you can turn off a particular zoom if the display
becomes too crowded, and the zoom settings are saved in its location, ready to be turned on again when
desired.
The zoom's Vertical and Horizontal units will differ from the source trace, as seen from a comparison of
the trace descriptor boxes, because the zoom is showing a scale, not a measured level.
Channel descriptor box and its Zoom descriptor box.
You can further adjust these settings using the Front Panel knobs, or by changing the settings on the
Zoom dialog. Touch the zoom descriptor box to activate it, then touch it again to display the Zx tab.
Because it is a calculated and not a sampled trace, you can adjust the zoom's Horizontal Scale without
changing the oscilloscope's Timebase (a characteristic shared with math and memory traces).
Turn off Zoom
Turn off a zoom trace the same as you would any other trace:
l Deselect the Trace On checkbox on the Zoom dialog.
l Touch-and-hold (right-click) the descriptor box until the pop-up menu opens, then choose Off.
Quick Zoom
Use the Front Panel Zoom button to quickly create one zoom trace for each displayed channel trace.
NOTE: Quick zooms are created at the same vertical scale as the source trace and 10x horizontal
magnification.
To turn off the quick zooms, press the Zoom button again.
Multi-Zoom
The Multi-Zoom feature creates time-locked zoom traces for only the waveforms that you choose to
include. The zooms are of the same X-axis section of each waveform. As you scroll through a waveform,
all included zooms scroll in unison.
Zoom Controls
Once the zoom trace has been created, adjust its Vertical and Horizontal Scale to further "zoom" in or out.
You can do this by activating the zoom trace and using the Front Panel Vertical and Horizontal knobs, or
by modifying settings on the Zoom dialog.
24
922499 Rev B
Operator's Manual
To access it the Zoom dialog, double-touch any zoom trace descriptor box, or choose Math > Zoom Setup
from the menu bar.
The main Zoom dialog contains selection boxes for turning on/off zoom traces. There are also options to:
l Reset All - returns all zooms to x1 magnification.
l Quick Zoom - creates a corresponding zoom trace for each open channel trace, same as the Front
Panel Zoom button.
l MultiZoom
Behind the main Zoom dialog is a separate tab for each potential zoom trace (Z1-Z8). Each dialog reflects
the current scale settings for that zoom.
Trace Controls
Trace On - displays the zoom trace. Select/deselect this box to show/hide the zoom.
Source - lets you change the source for this zoom to any channel, math, or memory trace while
maintaining all other settings.
Segment Controls
These controls are used only in Sequence Sampling Mode.
Rescale Controls
These controls on the Zx dialogs are the same used to rescale any trace, and you will see them
throughout the oscilloscope software. They work the same wherever they appear.
Out and In buttons - increase or decrease the magnification of the zoom, and consequently change the
Horizontal andVertical Scale settings. Continue to touch either button until you've achieved the desired
level of zoom.
Var.checkbox - enables variable zooming in increments finer than the default 1, 2, 5, 10 step increments.
When checked, each touch of the zoom control buttons changes the degree of magnification by a single
increment.
922499 Rev B
25
HDO6000 High Definition Oscilloscope
Horizontal Scale/div- sets the amount of time represented by each horizontal division of the grid. It is the
equivalent of Time/div, only unlike the Timebase setting, it may be set differently for each zoom, math
function, or memory trace.
Vertical Scale/div - sets the voltage level represented by each vertical division of the grid; it's the
equivalent of V/div used for channel settings.
Horizontal/Vertical Center - sets the voltage or time that is to be at the center of the screen on the zoom
trace. The horizontal center is the same for all zoom traces.
Reset Zoom - returns the zoom to x1 magnification.
Rescale Memory or Math Function Traces
Unlike channel traces, memory (M1 - M4) or math function (F1 - F8) traces can be rescaled directly
without having to create a separate zoom trace. The same set of controls used to rescale zoom traces
appear on the Zoom right-hand dialog, or on one of the trace setup dialogs. This applies to any trace that
is created as a math function (Fx) trace, including those that are generated through analysis options.
You can, however, create a separate zoom trace from a memory or function trace by drawing a selection
box around a portion of the waveform. In this case, you choose one of the zoom locations in which to draw
the trace, Z1-Z8, but the source trace remains at the original scale.
Multi-Zoom
Multi-Zoom creates time-locked zoom traces for only the waveforms that you choose to include. The
zooms are of the same X-axis section of each waveform. As you scroll through a waveform, all included
zooms scroll in unison.
Set Up Multi-Zoom
1. Choose Math → Zoom Seutp... to open the Zoom dialog, then touch the Multi-Zoom tab or Multi-Zoom
Setup... button.
2. On the Multi-Zoom dialog, turn Multi-Zoom On and select all the traces that are In the Multi-Zoom
group.
26
922499 Rev B
Operator's Manual
Scroll Waveforms
The Auto-Scroll controls appear at the right of the Multi-Zoom dialog. They work similarly to A/V controls
to allow you to continuously scroll all the selected zoom traces together in time-locked steps from the
beginning to the end of the acquisition.
They are (from left to right, top to bottom row):
Scroll Left Fast - back in time
Scroll Left Slow - back in time
Pause - stop scrolling
Scroll Right Slow - forward in time
Scroll Right Fast - forward in time
Jump to Start - go to beginning of acquisition
Jump to End - go to end of acquisition
In/Out - increase or decrease magnification level of zooms
Var - zoom In/Out in finer increments than the default 1, 2, 5, 10 steps
Reset Zoom- return all zooms to same scale as the source trace.
Turn Off Multi-Zoom
1. From the menu bar, touch Math → Zoom Setup....
2. On the main Zoom dialog, deselect the MultiZoom checkbox.
922499 Rev B
27
HDO6000 High Definition Oscilloscope
Vertical
Vertical, also called Channel, settings usually relate to voltage level and control the trace along the Y axis.
NOTE:While Digital settings can be accessed through the Vertical menu on -MS model oscilloscopes,
they are handled quite differently. See Digital Overview.
The amount of voltage displayed by one vertical division of the grid, or Vertical Scale (V/div), is most
quickly adjusted by using the Front Panel Vertical knob. The Channel descriptor box (Cx) always shows
the current Vertical Scale setting.
More extensive Vertical settings are made on the Channel dialog, which will be labeled Cx after the
corresponding channel. To access the Channel dialog, choose Vertical > Channel <#> Setup from the
menu bar, or touch the Channel descriptor box.
The Channel dialog contains:
l Vertical Controls for vertical scale, offset, coupling, bandwidth, and probe settings.
l Pre-Processing Controls to set up pre-acquisition processes that will affect the waveform, such as
noise filtering and interpolation.
If a probe is connected to the channel, the Channel dialog also contains a tab for the Probe dialog.
Vertical Settings
Vertical Scale - Set the vertical scale or sensitivity, and choose fixed or variable gain adjustment.
Vertical Offset - Select between zero vertical offset or to set the offset to a specific value.
Coupling - Select from DC 50 Ω, DC1M, AC1M and GROUND.
CAUTION. The maximum input voltage depends on the input used. Limits are displayed on the
front of the oscilloscope. Whenever the voltage exceeds this limit, the coupling mode
automatically switches to GROUND. You then have to manually reset the coupling to its
previous state. While the unit does provide this protection, damage can still occur if extreme
voltages are applied.
Bandwidth - Bandwidth filters are available at a variety of fixed bandwidth settings. The exact settings
vary by model.
Probe Attenuation -Enables you to set probe attenuation manually if using a third-party probe. The
oscilloscope's inputs automatically sense Teledyne LeCroy probes and sets probe attenuation and
Vertical units for you.
Vertical Unit Override - Allows the units of the selected channel to be changed from Volts (V) to Amperes
(A). This is useful when using a third party current probe that is not auto-detected or when probing across
a current sense resistor.
28
922499 Rev B
Operator's Manual
Pre-Processing Settings
Averaging - performs continuous averaging or the repeated addition, with unequal weight, of successive
source waveforms. It is particularly useful for reducing noise on signals drifting very slowly in time or
amplitude. The most recently acquired waveform has more weight than all the previously acquired ones:
the continuous average is dominated by the statistical fluctuations of the most recently acquired
waveform. The weight of old waveforms in the continuous average gradually tends to zero (following an
exponential rule) at a rate that decreases as the weight increases.
Deskew - adjusts the horizontal time offset by the amount entered in order to compensate for
propagation delays caused by different probes or cable lengths. The valid range is dependent on the
current timebase setting. The Pre-processing deskew and the Math deskew functions perform the same
activity. See Deskew Channels.
Invert - Inverts the waveform for the selected channel.
Interpolation - Linear interpolation, which inserts a straight line between sample points, is best used to
reconstruct straight-edged signals such as square waves. (Sinx)/x interpolation, on the other hand, is
suitable for reconstructing curved or irregular wave shapes, especially when the sample rate is 3 to 5
times the system bandwidth.
Noise Filter (ERes) - Enhanced Resolution (ERes) filtering increases vertical resolution, allowing you to
distinguish closely spaced voltage levels. The tradeoff is reduced bandwidth. The functioning of the
instrument's ERes is similar to smoothing the signal with a simple, moving-average filter. Use ERes on
single-shot waveforms, or where the data record is slowly repetitive (when you cannot use averaging).
Use it to reduce noise when your signal is noticeably noisy, but you do not need to perform noise
measurements. It also may be used when performing high-precision voltage measurements: zooming
with high vertical gain, for example. For more information, see Enhanced Resolution.
Probe Dialog
The Probe Dialog displays probe attributes and (depending on the probe type) allows you to AutoZero or
DeGauss Teledyne LeCroy probes from the oscilloscope touch screen. When a probe is not connected, the
Channel dialog shows only the C1 tab for vertical setup.
Channel dialog with tab for connected probe.
Probe Information on Channel Dialog
After a Teledyne LeCroy probe is connected, it is recognized by the oscilloscope.
l For passive probes, attenuation is automatically set, and these fields are disabled on the Channel
setup dialog.
l For active voltage and current probes, an additional tab with the probe model name is displayed to
the right of the C1 tab. Click on the tab to display the probe dialog.
922499 Rev B
29
HDO6000 High Definition Oscilloscope
When using third-party probes, the Probe Attenuation and Deskew values may be entered manually on the
Channel dialog.
Probe Information on Probe Dialog
This additional tab contains specific information on the connected probe. Default values for the probes
coupling and attenuation are automatically downloaded from the probe, and these settings along with
other attributes are shown on the dialog. Other controls may be available depending on the probe model or
input device type.
Probe dialog showing the connected probe's control attributes.
Deskew Channels
The signal input channels are deskewed at the factory prior to shipment and should not require any
further action.
Follow this generic deskew procedure to compensate for propagation delays due to different lengths of
cables, probes, or anything else that might cause timing mismatches between signals.
1. Connect all probes to the desired channels, then probe a common signal with each probe.
2. Turn on two channels, one of which will be the reference channel for the entire deskew procedure.
3. Set an Edge trigger on the reference channel.
4. Switch to Display > Single Grid so both traces appear on the same grid.
5. Turn the Front Panel Horizontal knob to adjust Time/div so that you can clearly see the edges of each
trace.
6. Touch the channel descriptor box for the second channel twice to open the setup dialog.
7. Touch the Deskew field to activate it.
8. Turn the Front Panel Adjust knob until the trace aligns with the reference waveform.
9. Repeat Steps 2 through 8 for each input, using the same channel as the reference and same trigger
each time.
30
922499 Rev B
Operator's Manual
Digital (Mixed Signal)
The digital leadset (standard with -MS model oscilloscopes) inputs up-to-16 lines of digital data. Leads
are organized into two banks of eight leads each, and you assign each bank a standard Logic Family or a
custom Threshold and Hysteresis to capture the digital signals.
The Digital set up dialog has four tabs each corresponding to one of four possible digital groups, labeled
Digital1 to Digital4. You choose which lines from among the 16 make up each digital group, what they are
named, and how the group appears on the display. Initially, logical lines are numbered the same as the
physical lead they represent, although any line number can be re-assigned to any lead.
Digital Traces
When a digital group is enabled, digital Line traces show which lines are high, low, or transitioning relative
to the threshold. You can also view a digital Bus trace that collapses all the lines in a group into their Hex
values.
Three Digital Line traces and a Bus trace displayed with a Vertical Position of positive 4.0 divisions (top of
grid) and a Group Height 4.0 divisions (half the grid).
Activity Indicators
Activity indicators at the bottom of the Digital<#> dialogs show which lines are High (up arrow), Low
(down arrow), or Transitioning (up an down arrows) relative to the Logic Threshold value. They provide a
quick view of which lines are active and of interest to display on screen.
922499 Rev B
31
HDO6000 High Definition Oscilloscope
Digital Group Set Up
1. From the menu bar, choose Vertical > Digital <#> Setup, or press the Front Panel Dig button and select
the desired Digital<#> tab.
2. On the Digital<#> set up dialog, check the boxes for lines D0 through D15 that comprise the group.
Touch the Display D0-D7 and Display D8-D15 buttons to quickly turn on the entire digital bank, or
touch the Right and Left Arrow buttons to switch between each digital bank as you make line
selections.
NOTE: Each group can consist of anywhere from 1 to 16 of the leads that are (or will be) connected to
signal, from either digital bank regardless of the Logic set on the bank. It does not matter if the some
or all of the lines have been included in other groups.
3. When all group members are selected, optionally rename them.
4. Go on to set up the digital display for the group. Check Group to enable the display.
5. When you're finished on the Digital<#> dialog, touch the Logic Setup tab and choose the Logic Family
that applies to each digital bank, or set custom Threshhold and Hysteresis values.
Digital Display Set Up
You can choose the type and position of the digital traces that appear on screen for each digital group.
1. Set up the digital group.
2. Touch Display Mode and choose from:
l Lines - the default display, which shows a time-correlated trace indicating high, low, and tran-
sitioning points (relative to the Threshold) for every digital line in the group. The size and placement of the lines depend on the number of lines, the Vertical Position and Group Height
settings.
32
l Bus - which collapses the lines in a group into their Hex values. It appears immediately below
all the Line traces when both are selected.
l Lines & Bus - which displays both line and bus traces at once.
922499 Rev B
Operator's Manual
3. In Vertical Position, enter the number of divisions (positive or negative) relative to the zero line of the
grid where the display begins.The top of the first trace appears at this position.
In the example above, the first Line trace (D0) starts at positive 4.0 divisions off the zero line, or at the
very top of the eight vertical division grid.
4. In Group Height, enter the total number of grid divisions the entire display should occupy. All the
selected traces (Line and Bus) will appear in this much space.
Individual traces are resized to fit the total number of divisions available. The example above shows a
group of three Line traces plus the Bus trace occupying a Group Height of 4.0 divisions. Each trace
takes up one division.
5. Check the Group box to enable the display.
TIP: Because a new grid opens to accommodate each enabled group, you may wish to enable groups
one or two at a time when they have many lines to maximize the total amount of screen space
available for the each grid. Closing the set up dialogs will also increase available screen space.
To close traces, uncheck the Group box , or touch-and-hold on the Digital<#> descriptor box and
choose Off from the pop-up menu.
Renaming Digital Lines
The labels used to name each line can be changed to make the user interface more intuitive. Also, labels
can be "swapped" between lines.
Changing Labels
1. Set up the digital group.
2. Touch Label and select from:
l Data - the default, which appends "D." to the front of each line number.
l Address - appends "A." to the front of each line number.
l Custom - lets you create your own labels line by line.
3. If using Custom labels:
l Touch the Line number button below the corresponding checkbox. If necessary, use the
Left/Right Arrow buttons to switch between banks.
l Use the virtual keyboard to enter the name, then press OK.
The button and any active line traces are renamed accordingly.
Swapping Lines
This procedure helps in cases where the physical lead number is different from the logical line number
you would like to assign to that input (e.g., a group is set up for lines 0-4, but lead 5 was accidentally
attached to the probing point). It can save time having to re-attach leads or re-configure groups.
1. Select a Label of Data or Address.
2. Touch the Line number button below the corresponding checkbox. If necessary, use the Left/RightArrow buttons to switch between banks.
3. From the popup, choose the line with which you want to swap labels.
The button and any active line traces are renumbered accordingly.
922499 Rev B
33
HDO6000 High Definition Oscilloscope
Timebase
Timebase, also known as Horizontal, settings control the trace along the X axis. These settings are
shared by all channel traces.
The time represented by each horizontal division of the grid, or Time/Division, is most easily adjusted
using the Front Panel Horizontal knob. Full Timebase set up, including sampling mode and clock source
selection, is done on the Timebase dialog, which can be accessed by either choosing Timebase >Horizontal Setup from the menu bar, or touching the Timebase descriptor box.
The main Timebase dialog contains settings for Sampling Mode, Timebase Mode, and Real Time
Memory. Related tabs open dialogs to set up Sequence Mode and Clock Source.
Timebase Settings
Sampling Mode
Real Time, Sequence, Roll, or RIS mode.
Timebase Mode
These controls set the timebase shared by all channels.
Time/Division - the time represented by one horizontal division of the grid. Touch the Up/Down Arrow
buttons on the Timebase dialog or turn the Front Panel Horizontal knob to adjust this value.
Delay - The time relative to the trigger event to display on the grid. In Real Time sampling mode, the
trigger event is placed at time zero on the grid. Delay may be time pre-trigger, entered as a negative
value, or post-trigger, entered as a positive value. Raising/lowering the Delay value has the effect of
shifting the trace to the right/left, enabling you to focus on the relevant portion of longer acquisitions.
Set to Zero - returns Delay to zero.
Real Time Memory
These controls set how the oscilloscope samples when in Real Time mode.
Sampling Rate - the number of samples taken per time division when using a Fixed Sampling Rate. It
changes to Max. Sampling Points, the number of samples taken per acquisition, if you choose to Set
Maximum Memory.
Set Maximum Memory - automatically adjusts the sampling rate to take the maximum number of
samples possible given the amount of pre- or post-trigger delay and the Time/div, up to the
oscilloscope's maximum record length. This is a quick way to optimize the sample rate for fast
timebases when in Real Time mode.
Fixed Sampling Rate - activates the Sampling Rate field for you to set your own rate. Lowering the rate
can extend the acquisition to accommodate slower timebases or longer delays.
34
922499 Rev B
Operator's Manual
Sampling Modes
Real Time Sampling Mode
Real Time sampling mode is a series of digitized voltage values sampled on the input signal at a uniform
rate. These samples are displayed as a series of measured data values associated with a single trigger
event. By default, the waveform is horizontally positioned so that the trigger event is time zero on the grid.
The relationship between sample rate, memory, and time can be expressed as:
Capture Interval = 1/Sample Rate X Memory
Capture Interval/10 = Time Per Division
In Real Time sampling mode, the acquisition can be displayed for a specific period of time (or number of
samples) either before or after the trigger event occurs, known as trigger delay. This allows you to isolate
and display a time/event of interest that occurs before or after the trigger event.
l Pre-trigger delay displays the time prior to the trigger event. This can be set from a time well before
the trigger event to the moment the event occurs, up to the oscilloscope's maximum sample record
length. How much actual time this represents depends on your timebase setting. When set to the
maximum allowed pre-trigger delay, the trigger position (and zero point) is off the grid (indicated by
the trigger delay arrow at the lower right corner), and everything you see represents pre-trigger
time.
l Post-trigger delay displays time following the trigger event. Post-trigger delay can cover a much
greater lapse of time than pre-trigger delay, up to the equivalent of 10,000 time divisions after the
trigger event occurred. When set to the maximum allowed post-trigger delay, the trigger point may
actually be off the grid far to the left of the time displayed.
Usually, on fast timebase settings, the maximum sample rate is used when in Real Time mode. For
slower timebase settings, the sample rate is decreased so that the maximum number of data samples is
maintained over time.
Roll Mode
Roll mode displays, in real time, incoming points in single-shot acquisitions that have a sufficiently low
data rate. This mode can be invoked for slow acquisitions where the time per division is 100 ms/div or
slower. Roll mode samples at ≤ 2.5 MS/s.
The oscilloscope appears to "roll" the incoming data continuously across the screen until a trigger event
is detected and the acquisition is complete. The parameters or math functions connected to each
channel are updated every time the roll mode buffer is updated, as if new data is available. This resets
statistics on every step of Roll mode that is valid because of new data.
NOTE: If the processing time is greater than the acquire time, the data in memory is overwritten. In this
case, the instrument issues the warning, "Channel data is not continuous in ROLL mode!!!" and rolling
starts again.
RIS Sampling Mode
RIS (Random Interleaved Sampling) is an acquisition technique that allows effective sampling rates
higher than the maximum single-shot sampling rate. It is used on repetitive waveforms with a stable
trigger. The maximum effective RIS sampling rate is achieved by making multiple single-shot
acquisitions at maximum real-time sample rate. The bins thus acquired are positioned approximately 8
922499 Rev B
35
HDO6000 High Definition Oscilloscope
ps (125 GS/s) apart. The process of acquiring these bins and satisfying the time constraint is a random
one. The relative time between ADC sampling instants and the event trigger provides the necessary
variation.
The instrument requires multiple triggers to complete an acquisition. The number depends on the sample
rate: the higher the sample rate, the more triggers are required. It then interleaves these segments (as
shown in the following illustration) to provide a waveform covering a time interval that is a multiple of the
maximum single-shot sampling rate. However, the real-time interval over which the instrument collects
the waveform data is much longer, and depends on the trigger rate and the amount of interleaving
required.
Interleaving of sample in RIS sampling mode.
Sequence Sampling Mode
In Sequence Mode, the complete waveform consists of a number of fixed-size segments (see the
instrument specifications at teledynelecroy.com for the limits). The oscilloscope uses the sequence
timebase setting to determine the capture duration of each segment as 10 x time/div. With this setting,
the oscilloscope uses the desired number of segments, maximum segment length, and total available
memory to determine the actual number of samples or segments, and time or points.
Sequence Mode is ideal when capturing many fast pulses in quick succession or when capturing few
events separated by long time periods. The instrument can capture complicated sequences of events
over large time intervals in fine detail, while ignoring the uninteresting periods between the events. You
can also make time measurements between events on selected segments using the full precision of the
acquisition timebase.
36
Capturing segments in Sequence sampling mode.
922499 Rev B
SEQUENCE DISPLAY MODES
The instrument gives you a choice of five ways to display your segments:
l Adjacent
l Waterfall (cascaded)
l Mosaic (tiled)
Operator's Manual
l Overlay
l Perspective
NOTE: some display modes have limitations on the number of segments that can be shown at one time.
922499 Rev B
37
HDO6000 High Definition Oscilloscope
SET UP SEQUENCE MODE
When setting up Sequence Mode, you define the number of fixed-size segments acquired in single-shot
mode (see the instrument specifications for the limits). The oscilloscope uses the sequence timebase
setting to determine the capture duration of each segment. Along with this setting, the oscilloscope uses
the number of segments, maximum segment length, and total available memory to determine the actual
number of samples or segments, and time or points.
1. From the menu bar, choose Timebase → Horizontal Setup....
2. Choose Sequence Sampling Mode.
3. On the Sequence tab under Acquisition Settings, touch Number of Segments and enter a value.
NOTE: The number of segments displayed can be less than the total number of segments acquired.
4. To stop acquisition in case no valid trigger event occurs within a certain timeframe, check the Enable
Timeout box, then touch Timeout and provide a timeout value.
NOTE: While optional, Timeout ensures that the acquisition will complete in a reasonable amount of
time and control of the oscilloscope will return to the operator/controller without having to manually
stop the acquisition.
5. Touch Display mode and select a sequence display mode from the pop-up menu.
6. Touch the one of the Front Panel Trigger buttons to begin acquisition.
NOTE: Once acquisition has started, you can interrupt it at any time by pressing the Stop Front Panel
button. In this case, the segments already acquired will be retained in memory.
VIEW SEGMENTS IN SEQUENCE MODE
When in Sequence Mode, you can view individual segments easily using the Zoom dialog. The Zoom trace
defaults to Segment 1. You can move to later segments by changing the values in First segment to
display and Num(ber) of segments to display at once.
Tip: By changing the Num field value to 1, you can use the Front Panel Adjust knob to scroll through each
segment in order.
Channel descriptor boxes indicate the total number of segments acquired. Zoom descriptor boxes show
the number of first segment displayed and total number of segments displayed ([#] #). As with all other
Zoom traces, the zoomed segments are highlighted on the source trace.
Example: You may have acquired 1000 segments. You chose to display segments number 4-6. The
Channel descriptor box will read 1000. The Zoom descriptor box will read [4]3.
Use the Zoom controls to change the scale factors of the trace.
38
922499 Rev B
Operator's Manual
VIEW SEGMENT AS MATH FUNCTION
Besides using the Zoom feature, you can also create a Math (Fx) trace to display individual segments.
1. From the menu bar, choose Math → Math Setup...
2. Touch a Function (Fx) tab to display its corresponding dialog.
3. On the dialog, touch Operator1 and select the Segment button from the pop-up menu.
4. Touch the Select right-hand dialog tab.
5. Touch First Selected and choose the first segment to display.
6. Touch Number of Selected and enter the number of segments to display at once.
VIEW SEGMENT TIME STAMPS
To view time stamps for each segment:
1. From the menu bar, choose Timebase → Acquisition Status or Vertical → Channel Status.
2. Touch the Trigger Time tab.
3. Under Show Status For, choose Time .
4. In Select Segment, enter the segment number of interest.
You can also touch the Up/Down Arrow buttons to scroll through segment times.
Clock Source Settings
Sample Clock
These settings determine the clock that controls when the oscilloscope's digitizers sample the input
waveforms. The default setting is to use the oscilloscope's Internal clock. To use an external sample
clock:
1. Connect the clock source to the Ext input on the front of the oscilloscope using a BNC cable.
2. Go to TImebase > Horizontal Setup and choose Real-time Sampling Mode.
3. On the Clock Source tab under Sample Clock choose from 0V, ECL, TTL.
4. Choose an External Coupling of 50 Ohms, Ground, or 1 M Ohm.
Reference Clock
These settings control the Timebase reference used to synchronize acquisition across all channels. The
default setting is to use the oscilloscope's Internal 10 MHz clock. To use an external reference clock:
1. Connect the clock source to the Ref In/Out 10 MHz input on the back of the oscilloscope using a BNC
cable.
2. Go to Timebase > Horizontal Setup and choose Real-Time Sampling Mode.
3. On the Clock Source tab under Reference Clock choose External.
922499 Rev B
39
HDO6000 High Definition Oscilloscope
External Reference Clock vs. External Sample Clock
An external reference clock is used to synchronize the oscilloscope's internal timebase to an external
frequency source. This allows multiple instruments to lock their timebases to a common source.
An external sampling clock, applied via the Ext input, replaces the oscilloscope's internal timebase as the
sampling clock. This means that the external sampling clock controls when the oscilloscope's digitizers
sample the input waveforms.
Since the external sampling clock uses the Ext input, an external trigger cannot be used when the external
sampling clock is in use.
Auto Setup
Auto Setup quickly configures the essential oscilloscope settings based on the first input signal it finds,
starting with Channel 1. If nothing is connected to Channel 1, it searches Channel 2 and so forth until it
finds a signal.
Vertical Scale (V/div), Offset, Timebase (Time/div), and Trigger are set so that there is an Edge trigger on
the first, non-zero-level amplitude, and the entire waveform is visible for at least 10 cycles over the 10
horizontal divisions.
To run Auto Setup, you can either:
l Press the Auto Setup Front Panel button.
l Choose Auto Setup from the Vertical, Timebase, or Trigger menus. All these options perform the
same Auto Setup function.
To confirm Auto Setup, press the Auto Setup button again or use the touch screen display.
Restore Default Setup
Restore the oscilloscope to its factory default state by pressing the Front Panel Default Setup button. You
can also restore default settings by choosing File > Recall Setup > Recall Default.
Default settings for your oscilloscope include the following:
Channel/VerticalC1-C4 on at 50 mV/div Scale, 0 V Offset, Linear Interpolation
TimebaseReal Time Sampling at 50 ns/div, 0 Delay, 1.25 kS at 2.5 GS/s, 1.0 MS Memory
TriggerC1 with an Auto Positive Edge, DC Coupling, 0 V Level
DisplayAuto Grid
CursorsOff
MeasurementsCleared
MathCleared
40
922499 Rev B
Operator's Manual
Trigger
While the oscilloscope is continuously sampling signal when it is turned on, it can only display up to its
maximum memory in data samples. Triggers select an exact event/time in the waveform to display on
the oscilloscope screen so that memory is not wasted on insignificant periods of the signal.
All trigger types allow for pre-trigger or post-trigger delay, the display time relative to the trigger event
(although the trigger itself may not be visible), or let you set the time between sweeps, how often the
display is refreshed. Unless modified by a pre- or post-trigger delay, the trigger event occurs at point zero
at the center of the grid, and an equal period of time before and after this point is shown to the left and
right of it.
Trigger capabilities include:
l Simple Triggers - activated by basic waveform features such as an edge with a positive or negative
slope or width.
l Pattern Triggers - trigger the oscilloscope when a pattern condition, from false to true, occurs on
selected input channel and external input.
l SMART Triggers - sophisticated triggers that enable you to use basic or complex conditions for trig-
gering. Use SMART Triggers for signals with rare features, like glitches.
l Measurement Trigger - triggers that allow you to leverage parameter measurements as waveform
trigger conditions. A measurement trigger is either the only trigger or the final trigger in a chain of
trigger events including hardware triggers.
l MultiStage Triggers - varied forms of triggers including Cascaded, QualFirst, and Qualified allowing
varied combinations of triggers and trigger stages.
l Serial Triggers - provide triggers specific to a wide variety of serial data protocols.
l TV Triggers - provide the ability to trigger on multiple types of video signal.
In addition to the trigger type, the trigger mode determines how the oscilloscope behaves in the presence
or absence of a trigger event.
Trigger Modes
The trigger mode determines how the oscilloscope sweeps, or refreshes, the display. This can be set
from the Trigger menu, or from the Front Panel Trigger control group.
Auto mode causes the oscilloscope to sweep without a set trigger. An internal timer triggers the sweep
after a preset timeout period so that the display refreshes continuously. Otherwise, Auto functions the
same as Normal when a trigger condition is found.
In Normal mode, the oscilloscope sweeps only if the input signal reaches the set trigger point. Otherwise
it continues to display the last acquired waveform.
In Single mode, one sweep occurs each time you choose Trigger >Single or press the Front Panel Single
button.
Stop pauses sweeps until you select one of the other three modes.
922499 Rev B
41
HDO6000 High Definition Oscilloscope
Trigger Types
These are the trigger types available for selection. If the trigger is part of a subgroup (e.g., Smart), first
choose the subgroup from among the basic types to display all the trigger options.
Basic Triggers
Edge triggers upon a achieving a certain voltage level in the positive or negative slope of the wave.
Width triggers upon finding a positive- or negative-going pulse width when measured at the specified
voltage level.
Pattern triggers on a logical combination of analog or digital inputs: CH1, CH2, CH3, CH4, EXT, and D0D15. You have a choice of four Boolean operators (AND, NAND, OR, NOR) and can stipulate the high or low
voltage logic level for each input independently.
NOTE: Only the AND Boolean operator is available when combining analog and digital inputs.
Measurement triggers when a certain parameter measurement is found. A measurement trigger is either
the only trigger or the final trigger in a chain of trigger events including hardware triggers.
TV triggers on standard (PAL, SECAM, NTSC, HDTV) or custom composite video signals.
Smart Triggers
Window triggers when a signal enters or exits a window defined by voltage thresholds.
Interval triggers upon finding a specific interval, the time (period) between two consecutive edges of the
same polarity: positive to positive or negative to negative. Use the interval trigger to capture intervals that
fall short of, or exceed, a specified range.
Glitch triggers upon finding a fixed pulse-width time or time range.
Dropout triggers when a signal loss is detected. The trigger is generated at the end of the timeout period
following the last trigger source transition. It is used primarily in single-shot applications with a pretrigger delay.
Runt triggers when a pulse crosses a first threshold, but fails to cross a second threshold before recrossing the first. Other defining conditions for this trigger are the edge (triggers on the slope opposite to
that selected) and runt width.
SlewRate triggers when the rising or falling edge of a pulse crosses an upper and a lower level. The pulse
edge must cross the thresholds faster or slower than a selected period of time.
42
922499 Rev B
Operator's Manual
MultiStage Triggers
A type of MultiStage trigger, Cascaded triggers when a succession of criteria in Stages A-D are met. Each
stage can result in different trigger actions, such as arm only, trigger only, or trigger and rearm.
QualFirst arms the oscilloscope on the A event, then triggers on all subsequent B events.
NOTE: This button is enabled when using the sequence sampling mode. It is commonly used in sequence
mode for disk drive applications with the index pulse defined as the A qualifier signal and the servo gate
signal as the B triggering events.
Qualified arms the oscilloscope on the A event, then triggers on the B event. In Normal trigger mode, it
automatically resets after the B event. A (arm) can be Edge, Pattern, State, or PatState events; B (trigger)
can be Edge or Pattern events.
Only available as a sub-type of Qualified triggers, PatState triggers when the qualifying signal goes above
or below a specified voltage level. You can specify the number of these events that must occur to trigger.
Serial Triggers
Protocol-enabled serial triggers are available as options on some oscilloscope models. This trigger type
will be available only if you have such an option installed.
Trigger Settings
To access the Trigger setup dialogs, choose Trigger > Trigger Setup from the menu bar, or press the Front
Panel Trigger Setup button.
The dialogs you see and the options on them will vary depending on your trigger type selection. The main
Trigger dialog contains settings that are required for most trigger types.
The trigger setup is summarized in a preview window at the far right of the Trigger dialog.
922499 Rev B
43
HDO6000 High Definition Oscilloscope
Type - see Trigger Types for details. This selection drives the remainder of the trigger setup. The default
selection is Edge.
Source - the channel signal upon which to base the trigger. If a trigger is designed to work with multiple
inputs, like a Pattern trigger, you do not have to choose a single source, but will be given controls for
setting the conditions on each source.
Coupling - the type of signal coupling at the input. Choices are:
l DC - All the signal’s frequency components are coupled to the trigger circuit for high frequency
bursts or where the use of AC coupling would shift the effective trigger level.
l AC - The signal is capacitively coupled. DC levels are rejected, and frequencies below 50 Hz are
attenuated.
l LFREJ - The signal is coupled through a capacitive high-pass filter network, DC is rejected and sig-
nal frequencies below 50 kHz are attenuated. For stable triggering on medium to high frequency signals.
l HFREJ - Signals are DC coupled to the trigger circuit, and a low-pass filter network attenuates
frequencies above 50 kHz (used for triggering on low frequencies).
Level - the source Voltage level or levels that mark the threshold for the trigger to fire. Trigger levels
specified in Volts normally remain unchanged when the vertical gain or offset is modified.
Find Level - where available, this button sets the Level to the signal mean.
Trigger Holdoff
Holdoff is an additional condition that may be set for Edge and Pattern triggers. It can be expressed either
as a period of time or an event count. Holdoff disables the trigger temporarily, even if the trigger
conditions are met, until the holdoff conditions are also met. The trigger fires when the holdoff has
elapsed.
Use holdoff to obtain a stable trigger for repetitive, composite waveforms. For example, if the number or
duration of sub-signals is known, you can disable them by choosing an appropriate holdoff value.
Qualified triggers operate using conditions similar to holdoff.
Hold Off by Time
This is a period of time to wait to fire the trigger, either since the beginning of the acquisition or since the
trigger conditions were met.
Sometimes you can achieve a stable display of complex, repetitive waveforms by placing a holdoff
condition on the time between each successive Edge trigger event. This time would otherwise be limited
only by the input signal, the coupling, and the instrument's bandwidth. Select a positive or negative slope,
and a minimum time between triggers.
In the figure below, the bold edges on the trigger source indicate that a positive slope has been selected.
The broken upward-pointing arrows indicate potential triggers, which would occur if other conditions are
met. The bold arrows indicate where the triggers actually occur when the holdoff time has been
exceeded.
44
922499 Rev B
Operator's Manual
Edge trigger with holdoff by time.
Hold Off by Events
For purposes of Hold Off, Events refers to the number of times the trigger conditions have been met,
counted either from the beginning of the acquisition or since the last trigger. For example, if the hold-off
number of Events is 2 counted from the beginning of the acquisition, the trigger fires on the third event.
In the figure below, the bold edges on the trigger source indicate that a positive slope has been selected.
The broken, upward-pointing arrows indicate potential triggers, while the bold ones show where triggers
actually occur after the holdoff expires.
Edge trigger with holdoff by events.
922499 Rev B
45
HDO6000 High Definition Oscilloscope
Holdoff Settings
To access the Trigger Holdoff dialog, choose Triggers > Trigger Setup from the menu bar or press the
Front Panel Trigger Setup button, then touch the Holdoff tab.
Holdoff by - type of holdoff to use with trigger: None, Time (clock), or Event.
Time - if using Holdoff by Time, the time in S to wait before triggering.
Events - if using Holdoff by Events, the number of events to count before triggering.
Starts Holdoff Counter On - whether to count holdoff time/events from Acquisition Start or Last Trigger
Time before triggering again.
Software Assisted Trigger
Software Assisted Trigger is used to find the trigger-level crossing point closest to the hardware trigger
point. It then adjusts the time offset of the waveform so that it is aligned with the specified trigger level
and slope. Software Assisted Trigger provides a quick way to create eye diagrams.
NOTE: This feature can only be used with an Edge trigger type in Normal trigger mode.
1. From the menu, choose Triggers > Trigger Setup, then touch the Software Assisted Trigger tab.
2. Touch Enable.
3. Create a trigger window by entering a Hysteresis value. This value sets a boundary above and below
the main trigger level to exclude noise.
4. Choose Auto orNormal; this determine the trigger behavior when trigger crossings are not found in the
trigger source waveform.
l Auto mode allows all waveforms through the channel.
l Normal mode allows waveforms only with a trigger crossing within the horizontal gate region
through the channel.
5. Set Start and Stop time values on the Horizontal Gate part of the Software Assisted Trigger tab. These
values control where in the waveform the software-assisted trigger processing searches for trigger
crossings.
46
922499 Rev B
Operator's Manual
TriggerScan
TriggerScan is a debugging tool (available for any trigger type) that helps you quickly find rare waveform
glitches and anomalies. With TriggerScan, you can build a list of trigger setups to look for rare events and
automatically sequence through each one. TriggerScan can use any type of trigger setup available
including edge, width, and qualify as well as Smart Triggers (such as, glitch and runt triggers).
TriggerScan automates two key processes in triggering rare events:
l Trains the system by looking at normal acquired waveforms. During the training, the oscilloscope
analyzes the waveforms to determine what waveforms normally look like. Using this information, it
generates a list of smart trigger setups to trigger on abnormal situations.
l Loads the smart trigger setups from the Trainer and cycles through these. As triggers occur, they
are overlaid on the screen. All acquisition settings are preserved and you can use all the functions
of the oscilloscope to find the root cause of these anomalies including, WaveScan, Histograms, and
advanced analysis.
Training TriggerScan
The TriggerScan Trainer inspects the current acquisition and automatically builds a list of trigger setups
that could potentially be used to find events of interest.
NOTE: Run the Trainer if you want to change the trigger types or if you change the channel or signal. You
must acquire and display at least 3 cycles of a signal before running the Trainer.
1. Touch Trigger → Trigger Setup... from the menu bar, then touch the TriggerScan tab.
2. Touch the Trainer button.
3. Choose the Source channel on which to train, and select all the trigger types you want to set up.
4. Touch the Start Training button. The training begins. When it is complete, a list of smart trigger setups
is displayed in the Trigger List.
Modify Trigger List
The Trigger List displays a list of of the triggers created by the Trainer. Follow these steps to add or
remove triggers, or update their individual setups. Once you have made any changes to the Trigger List,
you are ready to start scanning.
1. If not already there, choose Trigger → Trigger Setup... from the menu bar, then touch the TriggerScan
tab.
2. Make any of the following modifications to the Trigger List:
l To add a new trigger setup to the list, touch the Trigger tab and set up the new trigger as desired
on the Trigger dialog. Then, back on the TriggerScan dialog, touch the Add New button to append
the new trigger to the Trigger List.
922499 Rev B
47
HDO6000 High Definition Oscilloscope
l To replace a Trigger List setup with the setup on the Trigger dialog, highlight the setup in the
Trigger List and touch the Update Selected button.
l To use a trigger from the Trigger List, highlight its corresponding row on the list, and then touch
the Load Selected button.
l To delete a trigger setup, highlight the setup in the Trigger List and touch the Delete Selected
button.
All trigger setups can be deleted regardless of selections on the Trigger List with one step by touching
the Delete All button.
3. Once you have made the desired changes to the Trigger List, touch the Trainer button and restart the
scan by touching the Start Training button on the Trigger Scan Trainer pop-up. The oscilloscope
automatically cycles through all the trigger setups.
NOTE:
l Use Dwell Time to tune the time that the oscilloscope waits before loading the next trigger.
l If you want TriggerScan to stop when the oscilloscope next triggers, check the Stop On Trigger
checkbox. You can use this to isolate trigger setups.
l If you have Persistence enabled, all trigger events are recorded on the display.
Saving TriggerScan Setups
Save TriggerScan setups whenever you have modified the Trigger List. The current Trigger List is not
preserved after exiting the application unless you manually save it.
1. On the TriggerScan dialog, touch Setup File Name and enter a file name, or touch the Browse button
and select a location and file name.
2. Touch the Save Setup... button.
NOTE: You can load previously saved TriggerScan setups by touching the Browse button, locating the file,
then touching Load Setup....
48
922499 Rev B
Operator's Manual
Display
Display settings affect the number and style of grids that appear on screen and some of the visual
characteristics of traces, such as persistence.
See Utilities for settings related to screen resolution, color, and screen savers.
Display Setup
Follow this procedure to adjust how traces appear on the touch screen display.
1. From the menu bar, choose Display → Display Setup...
2. Touch the Grid button, then select one of the grid types (the image on the icon shows the resulting
grid arrangement).
Auto, the default, automatically adds or deletes grids as you open or close traces, up to the maximum
number supported.
3. To dim or brighten the background grid lines, touch Grid Intensity and provide a value from 0 to 100.
4. Optionally, check Grid on top to superimpose the grid over the waveform.
NOTE: Depending on the grid intensity, some of your waveforms may be hidden from view when the
grid is placed on top. To view them, simply uncheck Grid on top.
5. Optionally, check Axis labels to display the values of the top and bottom grid lines (calculated from
volts/div) and the extreme left and right grid lines (calculated from the timebase).
6. Choose a line style for your traces: solid Line or a disconnected series of sample Points.
7. To highlight more frequent samples, touch Intensity and enter a value from 0 to 100. For more information, see Adjust and Intensity.
8. If you selected to display an XY grid, select the source channels to Input X and Input Y.
9. If you have an external monitor installed, touch Open Monitor Control Panel and set up the external dis-
play.
Moving Traces from Grid to Grid
You can move traces from grid to grid in several ways.
Next Grid Shortcut Button
Open the Channel setup dialog for the trace you want to move, then touch the Next Grid shortcut button at
the bottom of the dialog.
NOTE: If you have only one grid open, a second grid opens automatically when you select Next Grid.
922499 Rev B
49
HDO6000 High Definition Oscilloscope
Drag-and-Drop Descriptor Box
You can also move a trace from one grid to another by dragging its descriptor box to the desired grid. This
is a convenient way to quickly re-arrange traces on the display.
XY Displays
XY displays plot the phase shift between otherwise identical signals. They can be used to display either
voltage or frequency on both axes, each axis now corresponding to a different signal input, rather than a
different parameter. The shape of the resulting pattern reveals information about phase difference and
frequency ratio.
The sources must have the same X-axis scale.
1. Set up the desired source traces.
2. Go to Display > Display Setup and choose:
l XY to display only the XY plot.
l XYSingle to display the XY plot next to a single grid containing both source traces.
l XYDual to display the XY plot next to two grids, each containing one of the source traces.
3. Touch Input X and Input Y and select your sources from the pop-up menu.
NOTE: The inputs can be any combination of channels, math functions, or memories.
Persistence Overview
The Persistence feature displays waveforms in a manner that helps reveal idiosyncrasies or anomalies in
a repetitive signal. Use Persistence to accumulate on-screen points from many acquisitions to see your
signal change over time. The instrument persistence modes show the most frequent signal path in threedimensional intensities of the same color, or graded in a spectrum of colors.
You can show persistence for up to eight inputs for any channel, math function, or memory location (M1
to M4).
Persistence Mode
The Persistence display is generated by repeated sampling of the amplitudes of events over time, and the
accumulation of the sampled data into display maps. These maps create an analog-style display.
Statistical integrity is preserved because the duration (decay) is proportional to the persistence
population for each amplitude or time combination in the data.
ANALOG MODE
When you select Analog mode, each trace is assigned a different color.
As a persistence data map develops, different intensities of that color are assigned to the range between
a minimum and a maximum population. The maximum population automatically gets the highest
intensity, the minimum population gets the lowest intensity, and intermediate populations get intensities
in between these extremes.
The information in the lower populations (for example, down at the noise level) could be of greater
interest to you than the rest. The Analog persistence view highlights the distribution of data so that you
can examine it in detail.
50
922499 Rev B
Operator's Manual
COLOR MODE
Color mode persistence works on the same principle as Analog persistence, but instead uses the entire
color spectrum to map signal intensity: violet for minimum population, red for maximum population. In
this mode, all traces use all colors, which is helpful for comparing amplitudes by seeking like colors
among the traces.
3D MODE
3d persistence creates a topographical view of your waveform from a selection of shadings, textures, and
hues. The advantage of the topographical view is that areas of highest and lowest intensity are shown as
peaks and valleys, in addition to color or brightness. The shape of the peaks (pointed or flat) can reveal
further information about the frequency of occurrences in your waveform.
In this mode, you can also turn the X and Y axes of the waveform through 180° of rotation from -90° to
+90°.
In the solid view of color-graded persistence, saturation is set at 50%, with red
areas indicating highest intensity. The X-axis has been rotated 60%; the Y-axis
has been rotated 15%.
In the monochrome (analog) view, the lightest areas indicate highest intensity,
corresponding to the red areas in the solid view.
The shaded (projected light) view emphasizes the shape of the pulses.
In the wire frame view, lines of equal intensity are used to construct the
persistence map.
Saturation Level
Besides the different modes, you can select a saturation level as a percentage of the maximum
population. All populations above the saturation population are then assigned the highest color intensity:
that is, they are saturated. At the same time, all populations below the saturation level are assigned the
remaining intensities. Data populations are dynamically updated as data from new acquisitions is
accumulated.
A saturation level of 100% spreads the intensity variation across the entire distribution; at lower
saturation levels the intensity will saturate (become brighter) at the percentage value specified. Lowering
this percentage causes the pixels to be saturated at a lower population and makes visible those events
rarely seen at higher saturation levels.
Persistence Time
Persistence time is, quite simply, the duration of time (in seconds) after which persistence data is erased
from the display.
922499 Rev B
51
HDO6000 High Definition Oscilloscope
Persistence Setup
This procedure explains how to set up the persistence display on traces. Persistence can be quickly
added to all traces or applied individually.
TURN ON PERSISTENCE
1. Access the Persistence dialog by choosing Display > Persistence Setupand touching the Persistence
tab.
2. Check Persistence On.
3. To set up all traces together, touch All Locked.
To set up traces individually, touch Per Trace.
4. Select the persistence mode: None, Color, Analog, or 3d
If you're doing individual setup, repeat the mode selection for each channel, then go to that channel's
tab and make the remaining settings.
5. If using Analog or Color mode, optionally check Show last trace to superimpose the channel trace over
the persistence display.
If using 3d mode, complete setup for 3-D persistence.
6. Optionally, also change Saturation level and Persistence Time, and enable/disable Dot Joined.
SET UP 3-D PERSISTENCE
1. Touch the 3d button and, if necessary, open the channel tab.
2. Under 3D settings, touch Quality and choose wire frame, solid, or shaded.
3. Check MonoChrome if you prefer a single-color representation. In this case, intensity will be used
instead of color to indicate more frequently occurring events.
4. Optionally, change the angle of display by entering new Axis X Rotation and Axis Y Rotation values
from -90° to +90°.
TIP: A quick way to rotate the display is to grab a corner and drag it in the desired direction.
TURN OFF PERSISTENCE
To turn off the persistence display, access the Persistence dialog and choose Reset All, or select an
individual channel's None(left-most) persistence mode button.
52
922499 Rev B
Operator's Manual
Cursors
Cursors are markers (lines, cross-hairs, or arrows) that identify specific voltage and time values on the
waveform. Use cursors to make fast, accurate measurements of specific points in the waveform. There
are five, standard cursor types available.
The cursor measurement values can be read on the descriptor box for the trace. The Show buttons let you
change which set of values are shown on the descriptor box. The available selections depend on the type
of cursor.
The easiest way to position cursors is to touch and drag them to the desired locations. Use the Front
Panel Cursors knob or the Position data entry controls at the right side of the Standard Cursors dialog to
place the cursors precisely.
Cursor Types
Standard Cursors
These five cursors can be placed on most any Channel, Memory, Math or Zoom trace.
Horizontal (Time) cursors place vertical lines through a desired point along the horizontal axis to read the
signal's amplitude at the selected time. There are two main types:
l Horizontal Abs - displays a single, dashed, vertical line. The readout shows the absolute value at
the cursor location.
l Horizontal Rel - displays two, dashed, vertical lines. The readout depends on the Show option
selected.
Vertical (Amplitude) cursors place horizontal lines through a point on the vertical axis to read the
amplitude of the signal at that point. The two types are:
l Vertical Abs - displays a single dashed, horizontal line. The readout shows the absolute value at the
cursor location.
l Vertical Rel - displays two dashed, horizontal lines. The readout depends on the Show option
selected.
An option exists to place Both Horizontal (Time) and Vertical (Amplitude) types at once.
Special Cursors
Some cursors are offered only in special circumstances:
l Horizontal (Frequency) cursors look the same as Horizontal (Time) cursors except that they are
placed on waveforms that have frequency on the x-axis, such as FFTs.
l Horizontal (Event) cursors are placed only on Trend waveforms.
In addition, some optional software packages provide cursors and help markers that are specific to the
application.
Cursors on Math Functions
Cursors can be placed on math functions whose X-axis has a dimension other than time, such as an FFT.
When there is at least one math trace open, the Standard Cursors dialog contains an X-Axis control where
you can choose the units measured by the horizontal cursors. The options will be appropriate to the types
922499 Rev B
53
HDO6000 High Definition Oscilloscope
of function traces open; for example, if there is an FFT trace, there is an option for Hz. The cursor lines
are placed on the traces that normally display X-axis values in the selected units.
Cursor Settings
Display Cursors
Use either of the following methods to quickly turn on/off cursors:
l From the menu bar, choose Cursors then select the desired from the drop-down list.
l On the Front Panel, press the Cursor Type button repeatedly to scroll through all the cursor types.
Stop when the desired type is displayed
Position Cursors
With the cursor on, turn the Front Panel Cursors knob. If there is more than one cursor line, push the
Cursor knob until the correct line is selected, then turn the knob to move it.
OR
Touch and drag the cursor line to a new position.
Standard Cursors Dialog
These controls can be used in lieu of the Front Panel controls to set cursors. Access the dialog by
choosing Cursors > Cursors Setup from the menu bar.
Cursors On displays or hide cursor lines. When first checked, the last selected cursor type is displayed.
Cursor Type buttons select the type of cursor displayed on the grid.
The Show controls determine which values appear on the trace descriptor box readout, particularly when
using relative cursors:
l Absolute - shows specific voltages for the two cursor locations.
l Delta - shows the difference between the specific voltages at the cursor locations.
l Abs+Delta - shows both the specific voltages and the difference between the specific voltages at
the cursor locations.
l Slope - shows the slope of the waveform between the cursor locations.
The Position controls at the right-side of the Standard Cursors dialog display the current cursor location
and can be used to set a new location. The options available depend on the Cursor Type and Show
settings.
l X 1/2 - positive or negative time from the zero point.
l Y 1/2 - number of positive or negative divisions from the zero level. May be a fraction of a division.
l Track - locks both cursor lines so that they move together, maintaining their same relative distance
from each other.
54
922499 Rev B
Operator's Manual
Measure
Measurement parameters are tools that give you access to a wide range of waveform properties. Use
them to analyze many attributes of your waveform like rise-time, rms voltage, and peak-to-peak voltage,
for example.
The oscilloscope offers a selection of:
l Standard parameters for measuring amplitude and time
l Custom parameters you configure
l Specialized parameters for applications such as pass/fail testing or serial data decode (if you have
those options installed)
You can configure and display up to eight measurement parameters at once.
Measure Gate
By using gates, you can narrow the span of the waveform on which to perform parameter measurements,
allowing you to focus on the area of greatest interest. For example, if you "gate" five rising edges of the
waveform, the parameter calculations for rise time are performed only on the five pulses bounded by the
gate posts.
The default starting positions of the gate posts are 0 div and 10 div, which coincide with the left and right
ends of the grid. The gate, therefore, initially encloses the entire waveform.
The quickest way to set a gate is by dragging the gate posts located at the far left and right of the grid to
the desired positions.
You can refine this setting by specifying a position down to hundredths of a division in the Gate Start and
Stop fields on the Gate right-hand dialog.
For Standard Horizontal or Standard Vertical parameters, all parameters share the same gate.
Touch the Default button to return gates to the width of the trace.
Level and Slope
For several time-based measurements, you can choose to begin the measurement on positive, negative,
or both slopes. For two-input parameters, such as Dtime@level, you can specify the slope for each input,
as well as the level and type (percent or absolute).
Make Level selection on the right-hand Level dialog when it appears.
922499 Rev B
55
HDO6000 High Definition Oscilloscope
Set Up Measurement Parameter
To configure custom measurements to add to the table of parameter readouts:
1. From the menu bar, choose Measure > Measure Setup.
2. Choose Measure Mode My Measure.
3. Touch the Pxtab or button of an unused location (or one that you want to change).
4. Select a Type:
l Measure On Waveforms- measures directly on the waveform selected as Source1.
l Math On Parameters - performs math (addition, subtraction, multiplication, division) on the
parameters selected as Source1 and Source2. These must be two other custom parameters you
have or will configure and saved to those slots.
l Advanced Web Edit - uses Teledyne LeCroy's Processing Web for measurement setup. This fea-
ture, available with the XWEB option, allows you to chain practically unlimited math functions for
operation on your waveform measurements.
5. Touch Source1 and select the channel, math trace, memory trace, or other waveform to be measured.
If using Math on Parameters, choose the parameters, rather than the source trace, in Source1 and
Source2.
6. If you selected Measure On Waveforms, touch the Measure field and select the parameter from the
pop-up menu.
7. Make any further selections on the right-hand dialogs that appear after your Measure selection.
These are explained on the dialog.
8. Optionally, set a measurement gate by dragging the gate posts to reposition them or by entering a
Start and Stop division on the Gate right-hand dialog..
9. Check On to enable the parameter and add it to the measurement readout table.
10. Check Show Table to display the readout on screen.
56
922499 Rev B
Operator's Manual
List of Standard Parameters
Standard measurement parameters are listed below alphabetically.
NOTE: There may be additional parameters available depending on the software options installed on the
oscilloscope.
ParameterDescription
Measures the difference between upper and lower levels in two-level signals. Differs from pkpk in that
Amplitude
(ampl)
noise, overshoot, undershoot, and ringing do not affect the measurement. Amplitude is calculated by using
the formula Top – Base. On signals not having two major levels (such as triangle or saw-tooth waves), the
amplitude parameter returns the same value as peak-to-peak.
Area
Base
Cycles
(cycles)
Delay
Delta Delay
(ddelay)
Dperiod@level
(dper@lv)
Dtime@level
(dt@lv)
Dtrig Time
(dtrig)
Duration
(dur)
Integral of data: Computes area of the waveform relative to zero level. Values greater than zero contribute
positively to the area; values less than zero, negatively.
Lower of two most probable states (higher is top). Measures lower level in two-level signals. Differs from
min in that noise, overshoot, undershoot, and ringing do not affect measurement. On signals not having two
major levels (such as triangle or saw-tooth waves), the amplitude parameter returns the same value as
minimum.
Determines number of cycles of a periodic waveform lying between cursors. First cycle begins at first
transition after the left cursor. Transition may be positive- or negative-going.
Time from trigger to transition: Measures time between trigger and first 50% crossing of specifies signal.
Delay can be used to measure the propagation delay between two signals by triggering on one and
determining delay of other.
Computes time between 50% level of two sources.
Adjacent cycle deviation (cycle-to-cycle jitter) of the period measurement for each cycle in a waveform. The
reference level for this measurement can be specified.
Computes the time between transitions of the selected sources at the specified levels. Only positive going
transitions are counted.
Time from last trigger to this trigger
For single sweep waveforms, dur is 0; for sequence waveforms: time from first to last segment's trigger; for
single segments of sequence waveforms: time from previous segment's to current segment's trigger; for
waveforms produced by a history function: time from first to last accumulated waveform's trigger.
Duty CyclePercent of period for which data are above or below the 50% level of the signal.
Duty@level
(duty@lv)
Edge@level
(edge@lv)
Fall 80-20%
(fall8020)
Fall time
(fall)
Fall@level
(fall@lv)
922499 Rev B
Percent of period for which data are above or below a specified level.
Number of positive edges in waveform that cross the specified threshold level.
Duration of pulse waveform's falling transition from 80% to 20% of the amplitude averaged for all falling
transitions between the measurement gates. On signals not having two major levels (triangle or saw-tooth
waves, for example), top and base can default to maximum and minimum, giving less predictable results.
Duration of pulse waveform's falling transition from 90% to 10% of the amplitude averaged for all falling
transitions between the measurement gates. On signals not having two major levels (triangle or saw-tooth
waves, for example), top and base can default to maximum and minimum, giving less predictable results.
Fall at level: Duration of pulse waveform's falling edges between user-specified transition levels.
57
HDO6000 High Definition Oscilloscope
ParameterDescription
Threshold arguments specify two vertical values on each edge used to compute fall time. Formulas for
upper and lower values:
lower = lower thresh. x amp/100 + base
upper = upper thresh. x amp/100 + base
FirstIndicates value of horizontal axis at left cursor.
Frequency
(freq)
Freq@level
(freq@lv)
FWxxMeasures the width of the largest area histogram peak at xx% of the population of the highest peak.
Half Period
(hper)
Hist Ampl
(hampl)
LastTime from trigger to last (rightmost) cursor.
Level@X
(lvl@x)
MATLABProduces a parameter using a user-specified MATLAB function.
Maximum
(max)
Mean
Period of cyclic signal measured as time between every other pair of 50% crossings. Starting with first
transition after left measurement gate. The period is measured for each transition pair. The reciprocal of
each period measurement is calculated as the frequency.
Period of cyclic signal measured as time between every other pair at the specified level. Starting with first
transition after left measurement gate. The period is measured for each transition pair. The reciprocal of
each period measurement is calculated as the frequency.
Half period of a waveform.
Difference in value between the two most populated peaks in a histogram.
Gives the vertical value at the specified x position. If the x position is between two points, it gives the
interpolated value. When the Nearest point checkbox is checked, it gives the vertical value of the nearest
data point.
Measures highest point in waveform. Unlike top, does not assume waveform has two levels.
Average of data for time domain waveform. Computed as centroid of distribution for a histogram of the data
values.
MedianThe average of base and top values.
Minimum
(min)
N-cycle JitterPeak-to-peak jitter between edges spaced n UI apart.
NoneDisables parameter calculation
Num Points
(npoints)
Overshoot-
Overshoot+
PeaksNumber of peaks in a histogram.
Peak to PeakDifference between highest and lowest points in waveform. Unlike ampl, does not assume the waveform
58
Measures the lowest point in a waveform. Unlike base, does not assume waveform has two levels.
Number of points in the waveform between the measurement gates.
Amount of overshoot following a falling edge. This is represented as percentage of amplitude. Overshootis calculated using the formula (base - min.)/ampl x 100. On signals not having two major levels (triangle or
saw-tooth waves, for example), may not give predictable results.
Amount of overshoot following a rising edge specified This is represented as a percentage of amplitude.
Overshoot+ is calculated using the formula (max. - top)/ampl x 100. On signals not having two major levels
(triangle or saw-tooth waves, for example), may not give predictable results.
922499 Rev B
ParameterDescription
(pkpk)has two levels. Peak to peak is calculated using the formula maximum – minimum.
Operator's Manual
Percentile
(pctl)
Period
Period@level
(per@lv)
Phase
Rise
Rise 20-80%
(rise2080)
Rise@level
(rise@lv)
Horizontal data value that divides a histogram so the population to the left is xx% of the total.
The time between every other pair of 50% crossings. Starting with first transition after left measurement
gate, period is measured for each transition pair, with values averaged to give final result.
The time between every other pair of at the level specified. Starting with first transition after left
measurement gate, period is measured for each transition pair, with values averaged to give final result.
Phase difference between signal analyzed and signal used as reference. Both signals are measured from
the 50% point of their rising edges.
Duration of pulse waveform's rising transition from 10% to 90% of the amplitude averaged for all rising
transitions between the measurement gates. On signals not having two major levels (triangle or saw-tooth
waves, for example), top and base can default to maximum and minimum, giving less predictable results.
Duration of pulse waveform's rising transition from 20% to 80% of the amplitude averaged for all rising
transitions between the measurement gates. On signals not having two major levels (triangle or saw-tooth
waves, for example), top and base can default to maximum and minimum, giving less predictable results.
Rise at level: Duration of pulse waveform's rising edges between user-defined transition levels.
Threshold arguments specify two vertical values on each edge used to compute rise time.
Formulas for upper and lower values:
lower = lower thresh. x amp/100 + base
upper = upper thresh. x amp/100 + base
Root Mean Square of data between the measure gates calculated using the formula:
RMS
Where: vi denotes measured sample values, and N = number of data points within the periods found up to
maximum of 100 periods.
SetupTime from the data edge to the clock edge.
Skew
Slew Rate
(slew)
Std Dev
(sdev)
Time of clock1 edge minus time of nearest clock2 edge. Both signals are measured from the 50% point of
their rising edges.
Slew rate or local dV/dt in a transition zone
Standard deviation of the data between the measure gates using the formula:
Where: vi denotes measured sample values, and N = number of data points within the periods found up to
maximum of 100 periods. This is equivalent to the rms for a zero-mean waveform. Also referred to as AC
RMS
922499 Rev B
59
HDO6000 High Definition Oscilloscope
ParameterDescription
Difference between the measured times of crossing a given slope and level and the ideal expected time.
TIE@level
(tie@lv)
For Slope you can choose positive, negative, or both. For output units you can choose time or unit interval
(UI). A unit interval equals one clock period. The Virtual Clock setup gives you a choice of Standard (1.544
MHz) or Custom reference clocks. You can also use a mathematically derived Golden PLL to filter low
frequency jitter. The cutoff frequency is user selectable.
Time@level
(time@lv)l
Top
Total Pop
(totp)
Width
Width@level
(wid@lv)
WidthN
(widn)
X@maxDetermines the horizontal axis location of the maximum value between the measure gate.
X@minDetermines the horizontal axis location of the minimum value between the measure gate.
Time from trigger (t=0) to crossing at a specified level.
Higher of two most probable states (base is lower). Measures higher level in two-level signals. Differs from
max in that noise, overshoot, undershoot, and ringing do not affect measurement. On signals not having
two major levels (such as triangle or saw-tooth waves), the amplitude parameter returns the same value as
minimum.
Total population of a histogram.
Width of cyclic signal determined by examining 50% crossings in data input. If first transition after left cursor
is a rising edge, waveform is considered to consist of positive pulses and width the time between adjacent
rising and falling edges. Conversely, if falling edge, pulses are considered negative and width the time
between adjacent falling and rising edges. For both cases, widths of all waveform pulses are averaged for
the final result.
Width measured at a user-specified level.
Time of cyclic signal determined by examining 50% crossings in data input. The widthN is measured from
falling edge to rising edge.
Quick Measurements
Once you have set custom parameters in an available location, you can quickly hide or display them by
choosing Measure > My Measure and checking On next to each parameter you want to display. You do not
have to go through the entire setup process.
There are also standard parameter sets available for quick display. From the menu bar, choose:
l Measure > Standard Horizontal for a full set of common time parameters: freq, period, width, rise,
fall, delay, duty, num points.
l Measure > Standard Horizontal for a full set of common voltage parameters: mean, sdev, max., min.,
ampl, pkpk, top, base.
Mark the Show Table checkbox to display the parameter readout table below the grid.
60
922499 Rev B
Operator's Manual
Table of measurements open beneath grid. Far left cell opens the Measure dialog.
To quickly access the Measure Setup dialog if it is closed, touch the far left cell of the readout table
labeled Measure.
View Statistics
You can add the statistical measures value(last), mean, min., max., sdev, and num(ber of measurements
computed) to the measurement parameter readout table.
To turn on statistics, access the Measure dialog and check Statistics On.Clear the checkbox to remove
statistics from the readout. You can also choose Measure > Statistics from the menu bar.
The num statistic is the number of measurements computed. For any parameter that computes on an
entire waveform (like amplitude, mean, minimum, maximum, etc.) the value displayed represents the
number of sweeps.
For any parameter that computes on every event, the value displayed is equal to the number of events per
acquired waveform. If x waveforms were acquired, the value represents x times the number of cycles per
waveform. The value(last) statistic is equal to the measurement of the last cycle on the last acquisition.
To reset the statistics counter, touch Clear Sweeps on the display or Front Panel.
View Histicon
Histicons are miniature histograms of measurement parameters that appear on the measurement table.
These thumbnail histograms let you see at a glance the statistical distribution of each parameter.
1. Choose Measure > Measure Setup from the menu bar to access the Measure dialog.
2. Select Show Table.
3. Check On to enable the parameters you wish to display.
4. Select Statistics Histicons.
NOTE: You can quickly display a full histogram by touching the histicon you want to enlarge. The enlarged
histogram appears superimposed over its source trace.
922499 Rev B
61
HDO6000 High Definition Oscilloscope
Help Markers
Help Markers clarify measurements by displaying cursor lines and labels marking the points being
measured. For at-level parameters, markers make it easier to see where your waveform intersects the
chosen level. This feature also displays any hysteresis band that you have set about that level.
You can choose to use Simple markers, which are only the lines, or Detailed markers, which include the
measurement point labels.
NOTE: Unlike regular cursors, which are white and can be moved, help markers are blue and only augment
the display; they cannot be moved, and they do not reset the measurement points. Some optional analysis
software packages include markers designed specially for that domain of reference, which are
documented in the option manual.
You also have the option, by means of the Always On checkbox, to leave the Help Markers displayed over
traces after you have closed the Measure dialogs or readout table. If you change the set of parameters
displayed, the markers will change, as well.
62
Standard Horizontal Parameter Help Markers
Standard Vertical Parameter Help Markers
922499 Rev B
Operator's Manual
Turn On Markers
1. From the menu bar, choose Measure > Measure Setup.
2. Select a Measure Mode: Std Vertical, Std Horizontal, or My Measure.
3. Touch the Show All button to display Help Markers for each enabled parameter.
The type of markers last selected appear on the display.
NOTE:If you choose My Measure but have not yet set up or enabled any parameters, you will not see
any markers, either.
4. To change the marker type, open any parameter (Px) dialog and in Markers select either:
l Simple - produces cursors and gate posts. The gate posts are independently placeable for each
parameter.
l Detailed - produces cursors, gate posts, a label identifying the parameter being measured, and a
level indicator and hysteresis band for "at level" parameters.
NOTE: The Markers setting is applied to all parameters at the same time. If you choose Simple
markers on any parameter dialog, all parameters are then displayed in this mode.
5. Select the Always On checkbox if you wish to continue to display Help Markers on open traces.
Turn Off Markers
From the Measure setup dialog, choose Help Markers Clear All.
From any Px dialog, choose Markers Off.
Qualified Parameters
Some Teledyne LeCroy software packages give you the ability to constrain parameter measurements to a
vertically or horizontally limited range, or to occurrences gated by a second waveform. Furthermore, both
constraints can operate together. This capability enables you to exclude unwanted characteristics from
your measurements. It is much more restrictive than See "Measure Gate" on page 55 which is used only
to narrow the span of analysis along the horizontal axis.
NOTE: Since this feature operates on only a subset of the data, possible alerts or status indicators
concerning the measurement (such as Data range too low) are not displayed.
Range Limited Parameters
1. From the menu bar, choose Measure → Measure Setup....
2. Touch a Px tab to open its corresponding dialog.
3. Now, on the dialog, touch inside the Source control and select a source from the pop-up.
4. Touch inside the Measure control and select a parameter from the pop-up menu.
5. Touch the Accept tab of the dialog on the right, then touch the Values In Range checkbox.
NOTE: Depending on whether you select a vertical or horizontal parameter, the correct units will be
automatically displayed (V, s, Hz, dB) in the Between and And controls. Or, if you select a simple ratio
parameter (such as power factor) that yields a dimensionless number, no units will be displayed.
6. Touch the Find Range button to quickly display the most recent value of the parameter measurement.
922499 Rev B
63
HDO6000 High Definition Oscilloscope
Waveform Gated Parameters
1. From the menu bar, choose Measure → Measure Setup.
2. Touch any Px tab to open the setup dialog.
3. Touch Source and select a source from the pop-up menu.
4. Touch Measure and select a parameter from the pop-up menu.
5. Touch the Acceptrigh-hand dialog tab, then check the Values Based on Waveform State box.
6. Touch When Wform and select the gating source.
7. Touch State Is and select High or Low from the pop-up menu. Parameter measurements on the
subject waveform will only be taken when the gating waveform is in the selected state.
8. TouchLevel Type and select Absolute or Percent from the pop-up menu.
9. TouchLevel and enter the crossing level value at which you want measurements to begin..
You can instead touch the Find Level button to automatically select the 50% level of your gating
waveform.
Math on Parameters
Besides reading parameter measurements, you can set up a custom parameter that performs arithmetic
operations (addition, subtraction, multiplication, division) on two other parameter measurements.
Alternatively, you can apply mathematical functions (for example, invert) to a single parameter
measurement.
The setup for Math on Parameters is much like other custom parameter setup. The only significant
difference is the choice of Math on Parameters instead of Measure on Waveforms and the selection of
source parameters instead of source traces. There is added functionality for using custom scripts to
calculate the results.
Math on Parameters differs fromMath in that the input and the output are still numerical values, as are
all parameter measurements. Math functions, on the other hand, input and output waveform traces. Math
on Parameters results display in the parameter readout table.
Exclusions
LOGARITHMIC PARAMETERS
The parameter math feature prevents multiplication and division of parameters that return logarithmic
values. These parameters include:
l auto-correlation signal-to-noise ratio (ACSN)
l narrow-band power (NBPW)
l media signal-to-noise ratio (MSNR)
l residual signal-to-noise ratio (RSNR)
l top-to-base ratio when the units are in dB (TBR)
OTHER EXCLUDED PARAMETERS
Parameters that are already the result of parameter math operations are excluded. If they are included in
a remote control setup command, an error message is generated and the setup canceled.
64
922499 Rev B
l delta clock-to-data near (DC2D)
l delta clock-to-data next (DC2DPOS)
l delta clock-to-data previous (DC2DNEG)
l delta delay (DDLY)
l delta time at level (DTLEV)
l phase (PHASE)
l resolution (RES)
l mTnTmT shift (BEES)
l mTnTmT shift sigma (BEESS)
l mTnTmT shift sigma – list (BEESS)
Set Up Math on Parameters
1. Touch Measure → Measure Setup... on the menu bar.
2. Choose Measure Mode My Measure.
3. Touch output Px tab or button to display the parameter setup dialog.
Operator's Manual
4. Touch the Math on Parameters button.
5. Touch Math Operator and select a math operation from the Select Measurement menu.
If you select an operation that requires two input parameters, the Source1 field will expand to two
fields.
6. Touch Source1 and Source2 and select two input parameters (P1 to P8) other than the parameter you
are now setting up.
To apply math to a single parameter (for example, Invert), just select it in Source1.
7. Check On to enable the new output parameter and add it to the display.
Using Scripts
In addition to the arithmetic operations, you can write your own VBScript or JavaScript to apply to one or
two measurement parameters. When setting up the output parameter, choose the Math Operator P Script.
Scripting can be done in the Script Editor window directly on the instrument, or you can import an existing
script.
PARAM SCRIPT VS. P SCRIPT
Param Script is a VBScript or JavaScript that operates on one or two waveforms and outputs a parameter
measurement, as shown in the figure below. P Script, on the other hand, is another VBScript or JavaScript
that takes as input one or two parameters and performs a math operation on them to produce another
parameter output.
The inputs to Param Script can also be math (Fx) or memory (Mx) traces. The inputs to P Script can be
the results of any parameter measurement, not necessarily Param Script.
922499 Rev B
65
HDO6000 High Definition Oscilloscope
SET UP MATH ON PARAMETERS USING SCRIPT
1. Touch Measure → My Measure... on the menu bar.
2. Touch the output Px tab or button to display the parameter setup dialog.
3. Touch the Math on Parameters button.
4. Touch Source1 and Source2 and select the input parameters (P1 to P8).
If you are applying math to a single parameter (for example, invert), just select it in Source1.
5. Touch Math Operator and choose P Script from the Select Measurement menu.
6. In the right-hand Script Math dialog, touch Script Language and choose either VBScript or JScript.
7. Touch the Edit Code button.
The Script Editor window opens.
8. Enter code in the script editor, or call up an existing script from a file storage location.
If you create your script in this window, you can export it to a new file.
Calculating Measurements
Determining Top and Base Lines
Proper determination of the top and base reference lines is fundamental for ensuring correct parameter
calculations. The analysis begins by computing a histogram of the waveform data over the time interval
spanned by the left and right measurement gates. For example, the histogram of a waveform transitioning
in two states will contain two peaks (see figure). The analysis will attempt to identify the two clusters
that contain the largest data density. Then the most probable state (centroids) associated with these two
clusters will be computed to determine the top and base reference levels: the top line corresponds to the
top and the base line to the bottom centroid.
66
922499 Rev B
Operator's Manual
Determining Rise and Fall Times
Once top and base are estimated, calculation of the rise and fall times is easily done (see figure). The
appropriate threshold levels are automatically determined by the instrument, using the amplitude (ampl)
parameter.
Threshold levels for rise or fall time can also be selected using absolute or relative settings (r@level,
f@level) if these parameters are included in your oscilloscope. If absolute settings are chosen, the rise or
fall time is measured as the time interval separating the two crossing points on a rising or falling edge.
But when relative settings are chosen, the vertical interval spanned between the base and top lines is
subdivided into a percentile scale (base = 0 %, top = 100 %) to determine the vertical position of the
crossing points.
The time interval separating the points on the rising or falling edges is then estimated to yield the rise or
fall time. These results are averaged over the number of transition edges that occur within the
observation window.
Rising Edge Duration
Falling Edge Duration
Where Mr is the number of leading edges found, Mf the number of trailing edges found,the time when
rising edge i crosses the x% level,and the time when falling edge i crosses the x% level.
Determining Time Parameters
Time parameter measurements such as width, period and delay are carried out with respect to the mesial
reference level, located halfway (50%) between the top and base reference lines or with respect to the
specified level for @level parameters.
Time-parameter estimation depends on the number of cycles included within the observation window. If
the number of cycles is not an integer, parameter measurements such as rms or mean will be biased.
However, only the last value is actually displayed, the mean being available when statistics are enabled.
To avoid these bias effects, cyclic parameters can be chosen, including crms and cmean, that restrict the
calculation to an integer number of cycles.
922499 Rev B
67
HDO6000 High Definition Oscilloscope
Determining Differential Time Measurements
The instrument enables accurate differential time measurements between two traces: for example,
propagation, setup and hold delays (see figure).
If included in your oscilloscope, parameters such as Delta c2d± require the transition polarity of the clock
and data signals to be specified.
Moreover, a hysteresis range may be specified to ignore any spurious transition that does not exceed the
boundaries of the hysteresis interval. In the figure, Delta c2d- (1, 2) measures the time interval separating
the rising edge of the clock (trigger) from the first negative transition of the data signal. Similarly, Delta
c2d+ (1, 2) measures the time interval between the trigger and the next transition of the data signal.
68
922499 Rev B
Operator's Manual
Math
Teledyne LeCroy offers a deep and always growing toolset of math functions.
Math functions can be applied to any channel (Cx), zoom (Zx), memory (Mx), or even other math traces
(Fx), allowing you to chain operations. For example, trace F2 can show the average of C1, while trace F3
provides the integral of F2.
In addition to the extensive math capabilities that are standard with every oscilloscope, enhanced math
analysis tools customized for various industries and applications are offered through optional software
packages. To learn about math tools available in each optional package, see the datasheets on the
Teledyne LeCroy website at teledynelecroy.com. If you have installed software options, these capabilities
are accessed through the oscilloscope Analysis menu, rather than the Math menu, although special
measure parameters and math functions will be available when using Measure and Math dialogs.
Single vs. Dual Operation Functions
Single functions perform one operation on one or two input sources.
Dual functions chain two operations to arrive at a single result. This saves you the effort of having to
chain two separate math functions together.
As with single functions, the number of sources required will vary based on the operation. You may need
only one source for Operator1, but two for Operator2 (the result of the first operation counts as one
source):
Graphing
The Graph button on the Math Function (Fx) dialogs allows you to create math functions that plot the
results of an applied measurement parameter: histogram, track, or trend. Choose the source, the
measurement parameter, and the type of plot to draw. The plots are the same as those you would create
using the shortcut buttons on the Measure Parameter (Px) dialog. See About Histograms and Track vs.
Trend.
As with other math functions, any configurable settings will appear on right-hand dialogs, after the plot
type is selected.
922499 Rev B
69
HDO6000 High Definition Oscilloscope
Set Up Math Function
This procedure explains how to set up single or dual operator math function (Fx) traces. Function traces
take as input one or more channel, zoom, memory or math traces and output a new math trace.
For more information about creating math traces that plot the results of applied measurements, see View
Trend, View Track, and View Histogram.
1. From the menu bar, choose Math > Math Setup.
TIP: If you know which function location you'll be using, you can select Fx Setup right from the Math
menu.
2. Choose a location by touching one of the Fx tabs (F1-F8).
3. On the Fx dialog, choose a single f(x) or dual g(f(x) operator function.
4. Choose math Operator1 to perform.
5. The choice of operator drives the number of Source fields you will see displayed. Make a selection in
each field.
A Summary of the function you are building appears on the dialog. Refer to this to be sure your
sources are in the proper order to yield the function you want (e.g., C1-C2 vs. C2-C1).
6. If the operator you've selected has any other configurable settings, you'll see a right-hand dialog of the
same name as the operator. Touch the tab to open the dialog and make any further settings. These
are explained on the dialog.
There will also be a Zoom dialog where you can optionally rescale the math trace. This does not
affect the scale of any other traces.
7. If you're creating a dual function, repeat Steps 4 through 6 for the second operation.
70
922499 Rev B
Operator's Manual
8. Check Trace On to display the new math trace.
Enable/Disable Math Function
Once a math function has been created and saved in one of the Fx locations, just use the main Math
dialog to quickly enable/disable it.
Touch the Front Panel Math button, or from the menu bar, choose Math > Math Setup, then check the On
box next to each function you wish to display.
Clear the On box to disable the function and close the trace.
List of Math Functions
Standard math functions are listed below alphabetically.
NOTE: There may be additional math functions available depending on the software options installed on
the oscilloscope.
FunctionDefinition
AbsoluteFor every point in the waveform the distance away from zero is calculated. For values greater
than zero this is the same as the value. For values less than zero, the magnitude of this value
without regard to its sign is used.
AverageCalculates either a summed or continuous average of a selected number of sweeps. See
Averaging Waveforms. The maximum number of sweeps is determined by the oscilloscope
model and memory. See the specifications at teledynelecroy.com.
CopyCopies waveform in its unprocessed state to the first available memory location.
CorrelationCalculates a measure of similarity of two waveforms, or a waveform against itself, as a function
of a time-lag applied to one of them.
DerivativeCalculates the derivative of adjacent samples using the formula:
(next sample value – current sample value) / (horizontal sample interval)
DeskewShifts trace in time the amount of the deskew factor.
DIfferenceFor every point in the waveform, the value of Source2 is subtracted from the value of Source1.
Source1 and Source2 must have the same horizontal units and scale and the same vertical
units.
922499 Rev B
71
HDO6000 High Definition Oscilloscope
FunctionDefinition
DigitalANDAND function between two digital waveforms (-MS models only).
DigitalFlipFlopInput1 is clocked in a hold when a rising edge of input2 occurs (-MS models only).
DigitalNANDNAND function between two digital waveforms (-MS models only).
DigitalNORNOR function between two digital waveforms (-MS models only).
DigitalNOTNOT function (inverter) of a digital waveform (-MS models only).
DigitalOROR function between two digital waveforms (-MS models only).
DigitalXORXOR function between two digital waveforms (-MS models only).
EnvelopeCalculates highest and lowest vertical values of a waveform at each horizontal value for a
specified number of sweeps.
EResApplies a noise reduction and smoothing filter by adding a specified number of bits. See
Enhanced Resolution.
ExpCalculates the antilog to the base e of the source; that is, e raised to the power equal to the
source.
Exp10Same as Exp, using base 10.
FFTComputes a frequency spectrum with optional Rectangular, Von Hann, Flat Topp, Hamming,
Blackman-Harris, and Hanning windows. Calculates up to 1 Mpts. Also allows FFT Averaging
through use of a second math operator. See FFT.
FloorCalculates the lowest vertical values of a waveform at each horizontal value for a specified
number of sweeps.
HistogramPlots the number of data points that fall into statistically significant intervals or bins. Bar height
relates to the frequency at which data points fall into each interval/bin.
IntegralCalculates the linearly rescaled integral (with multiplier and adder) of a waveform input starting
from the left edge of the screen using the formula:
(current sample value + next sample value) * (horizontal sample interval)
Each calculated area is summed with the previous sum of areas. The multiplier and adder are
applied before the integration function.
InterpolateInserts points between sampled points (upsamples) according to one of three algorithms: Linear
(straight line), Sinx/x (curved), and Cubic (spine). Interpolation factor of 2 to 50 determines
number of points in the upsample.
InvertFor every point in the waveform, the inverse of that point is calculated.
LnPeforms a natural log of a waveform. Values less than or equal to zero are set to underflow.
Log10Performs a log base 10 of a waveform. Values less than or equal to zero are set to underflow.
MatLab mathApplies a pre-programmed MatLab math function to the source waveform. Requires XDEV
option to edit functions through the oscilloscope GUI using MatLab Script.
phistogramCreates a histogram based on the displayed pixels of a waveform falling within a user defined
vertical or horizontal box (slice).
ProductFor every point in the waveform, the value of Source1 is multiplied by the value of Source 2.
Source1 and Source2 must have the same horizontal units and scale.
ptrace meanPlots the mean value of each sample point in a persistence map.
ptrace rangeGenerates a waveform with a width derived from the population range of a persistence map.
ptrace sigmaGenerates a waveform with a width derived from the sigma (sum) of a persistence map.
72
922499 Rev B
Operator's Manual
FunctionDefinition
RatioFor every point in the waveform, the value of Source1 is divided by the value of Source2.
Source1 and Source2 must have the same horizontal units and scale.
ReciprocalFor every point in the waveform the inverse is calculated using the formula:
1 / (sample value)
RescaleFor every point in the waveform the sample value is multiplied by the specified multiplier and
then add to with the specified adder. See Rescaling and Assigning Units.
RoofCalculates the highest vertical values of a waveform at each horizontal value for a specified
number of sweeps.
SegmentSelects one segment from a source waveform to place in a sequence waveform. Used in
Sequence sampling mode.
Sinx/xPerforms10 -to-1 interpolation using a Sin(x)/x filter.
Sparse“Thins,” or decimates, an incoming acquisition by dropping sample points at regular intervals.
Sparsing factor specifies the number of points to drop between retained samples (e.g., factor of
4 retains 1 then drops 4). Sparsing offset specifies the point at which to begin applying the
sparsing factor (e.g., offset of 3 begins count on the third sample (3), then drops the number of
samples specified by the sparsing factor (4).
SquareFor every point in the waveform, the square of the sample value is calculated.
Square RootFor every point in the waveform, the square root of the sample value is calculated.
SumFor every point in the waveform, the value of Source1 is added to the value of Source 2.Source1
and Source2 must have the same horizontal units and scale and the same vertical units.
TrackGenerates a waveform composed of parameter measurements that is time synchronous with the
source waveform. The vertical units are those of the source parameter value and the horizontal
units are seconds. Parameter values are posted at the sampling rate.
TrkSame as Track, with alternate transition types.
TrendProduces a waveform composed of a series of parameter measurements in the order the
measurements were taken. The vertical units are those of the source parameter, the horizontal
unit is measurement number. The trend contains a single value for each measurement. See
View Trend.
ZoomProduces a magnified trace of a selected portion of the input waveform. See Zooming Traces.
Interpolation
Linear interpolation, which inserts a straight line between sample points, is best used to reconstruct
straight-edged signals such as square waves. (Sinx)/x interpolation, on the other hand, is suitable for
reconstructing curved or irregular waveshapes, especially when the sampling rate is 3 to 5 times the
system bandwidth. The instrument also gives you a choice of Cubic interpolation. For each method, you
can select a factor from 2 to 50 points by which to interpolate (upsample).
1. Follow the usual steps to set up a math function, selecting Interpolate from the Filter submenu.
2. Touch the Interpolate tab in the mini setup dialog to the right of the main dialog.
3. Touch inside the Algorithm control and select an interpolation type.
4. Touch inside the Upsample by control (Upsampling is the factor by which sampling is increased) and
enter a value.
922499 Rev B
73
HDO6000 High Definition Oscilloscope
Sparser Function
The Sparse math function allows you to thin out an incoming waveform by skipping points at regular
intervals, and by starting acquisition at a particular offset (point). The Sparsing factor specifies the
number of sample points to reduce the input waveform by. A sparsing factor of 4, for example, tells the
oscilloscope to retain only one out of every 4 samples. A Sparsing offset of 3, on the other hand, tells the
oscilloscope to begin on the third sample, then skip the number of samples specified by the sparsing
factor (4). In this way, the sample rate is effectively reduced.
For the sparsing factor (interval), you can set a value from 1 to 1,000,000 points. For the sparsing offset
you can set a value from 0 to 999,999.
NOTE: The maximum sparsing offset that can be entered for any sparsing factor equals Sparsing Factor 1.
1. Follow the usual steps to set up a math function, selecting Sparse from the Misc submenu.
2. Touch the Sparsing factor control and provide a Bandwidth Limit value.
3. Touch the Sparsing offset control and provide a value.
Rescaling and Assigning Units
This feature allows you to apply a multiplication factor (a) and additive constant (b) to your waveform: aX
+ b. You can do it in the unit of your choice, depending on the type of application.
Set Up Rescaling
1. Follow the usual steps to set up a math function, selecting Rescale from the Functions submenu.
2. Touch the Rescale right-hand dialog tab.
3. To apply a multiplication factor:
l Check the First multiply by: box and enter a value for a, the multiplication factor.
l Touch then add: and enter a value for b, the additive constant.
4. To change the output unit of measure from that of the source waveform:
l Check Override units.
l In Output enter the abbreviation for the unit the measure you wish to use.
You can also enter combinations of the unit abbreviations following these rules:
l For the quotient of two units, use the character ":/"
l For the product of two units, use the character "."
l For exponents, append the digit to the unit without a space: S2 = seconds squared.
NOTE: Some units may be converted to simple units (e.g., V.A will display as W).
74
922499 Rev B
Abbreviated Units of Measure
AbbreviationMeasureAbbreviationMeasure
(blank)No unitsNNewton
AAmpereOHMOhm
CCoulombPALPascal
CYCLECyclesPCTPercent
DBDecibelPOISEPoise
DBCDecibel referred to carrierPPMParts per million
DBMDecibel MilliwattRADRadian
DBVDecibel VoltsDEGDegree (of arc)
DBUZDecibel MicroampMNTMinute (of arc)
DECDecadeSAMPLESample
DIVDivisionsSWEEPSweeps
EventEventsSECSecond (of arc)
Operator's Manual
FFaradSSecond
GGramSIESiemens
HHenryTTesla
HZHertzUIUnit interval
JJouleVVolt
KDegree KelvinVAVolt amps
CELDegree CelsiusWWatt
FARDegree FahrenheitWBWeber
LLiterMINMin
MMeterHOURHour
FTFootDAYDay
INInchWEEKWeek
YARDYard
MILEMile
922499 Rev B
75
HDO6000 High Definition Oscilloscope
Enhanced Resolution
ERes (Enhanced Resolution) filtering increases vertical resolution, allowing you to distinguish closely
spaced voltage levels. The instrument's ERes function is similar to smoothing the signal with a simple,
moving-average filter. However, it is more efficient concerning bandwidth and pass-band filtering.
Use ERes:
l On single-shot acquisitions, or where the data record is slowly repetitive (cases where you cannot
use averaging).
l To reduce noise on noticeably noisy signals when you do not need to perform noise measurements.
l When performing high-precision voltage measurements (e.g., zooming with high vertical gain).
ERes can be applied as a form of Pre-Processing, or as a Math function.
Set Up Enhanced Resolution (ERes)
To quickly set up ERes, open the Channel setup dialog and in the Pre-Processing section select a Noise
Filter (ERes) bit size .
To apply ERes as a Math function:
1. Follow the usual steps to set up a math function, selecting ERes from the Filter submenu.
2. Touch the Trace On checkbox.
3. Touch the ERes right-hand dialog tab , then touch bits and make a selection from the pop-up menu.
How the Instrument Enhances Resolution
The instrument's enhanced resolution feature improves vertical resolution by a fixed amount for each
filter. This real increase in resolution occurs whether or not the signal is noisy, or your signal is singleshot or repetitive. The signal-to-noise ratio (SNR) improvement you gain is dependent on the form of the
noise in the original signal. The enhanced resolution filtering decreases the bandwidth of the signal,
filtering out some of the noise.
The instrument's constant phase finite impulse response (FIR) filters provide fast computation, excellent
step response in 0.5 bit steps, and minimum bandwidth reduction for resolution improvements of between
0.5 and 3 bits. Each step corresponds to a bandwidth reduction factor of two, allowing easy control of the
bandwidth resolution trade-off. The parameters of the six filters are given in the following table.
Resolution increased by
0.50.52
1.00.2415
1.50.12110
-3 dB Bandwidth
(x Nyquist)
Filter Length
(Samples)
2.00.05824
2.50.02951
3.00.016117
With low-pass filters, the actual SNR increase obtained in any particular situation depends on the power
spectral density of the noise on the signal.
76
922499 Rev B
Operator's Manual
The improvement in SNR corresponds to the improvement in resolution if the noise in the signal is white
(evenly distributed across the frequency spectrum).
If the noise power is biased towards high frequencies, the SNR improvement will be better than the
resolution improvement.
The opposite may be true if the noise is mostly at lower frequencies. SNR improvement due to the
removal of coherent noise signals - feed-through of clock signals, for example - is determined by the fall
of the dominant frequency components of the signal in the passband. This is easily ascertained using
spectral analysis. The filters have a precisely constant zero-phase response. This has two benefits. First,
the filters do not distort the relative position of different events in the waveform, even if the events'
frequency content is different. Second, because the waveforms are stored, the delay normally associated
with filtering (between the input and output waveforms) can be exactly compensated during the
computation of the filtered waveform.
The filters have been given exact unity gain at low frequency. Enhanced resolution should therefore not
cause overflow if the source data is not overflowed. If part of the source trace were to overflow, filtering
would be allowed, but the results in the vicinity of the overflowed data -- the filter impulse response
length - would be incorrect. This is because in some circumstances an overflow may be a spike of only
one or two samples, and the energy in this spike may not be enough to significantly affect the results. It
would then be undesirable to disallow the whole trace.
Example ERes Applications
The following examples illustrate how you might use the instrument's enhanced resolution function.
GraphFunction
In low-pass filtering: The spectrum of a square signal before (left top) and after (left
bottom) enhanced resolution processing. The result clearly illustrates how the filter
rejects high-frequency components from the signal. The higher the bit enhancement,
the lower the resulting bandwidth.
To increase vertical resolution: In the example at left, the lower (inner) trace has
been significantly enhanced by a three-bit enhanced resolution function.
922499 Rev B
77
HDO6000 High Definition Oscilloscope
GraphFunction
To reduce noise: The example at left shows enhanced resolution of a noisy signal.
The original trace (left top) has been processed by a 2-bit enhanced resolution filter.
The result (left bottom) shows a smooth trace, where most of the noise has been
eliminated.
NOTE: While enhanced resolution can only improve the resolution of a trace, it cannot improve the
accuracy or linearity of the original quantization. The pass-band causes signal attenuation for signals
near the cut-off frequency. The highest frequencies passed may be slightly attenuated. Perform the
filtering on finite record lengths. Data is lost at the start and end of the waveform and the trace ends up
slightly shorter after filtering. The number of samples lost is exactly equal to the length of the impulse
response of the filter used: between 2 and 117 samples. Normally this loss (just 0.2 % of a 50,000 point
trace) is not noticed. However, you might filter a record so short that no data is output. In that case,
however, the instrument would not allow you to use the ERes feature.
Averaging Waveforms
Summed Averaging
Summed Averaging is the repeated addition, with equal weight, of successive source waveform records. If
a stable trigger is available, the resulting average has a random noise component lower than that of a
single-shot record. Whenever the maximum number of sweeps is reached, the averaging process stops.
In Summed averaging, you specify the number of acquisitions to be averaged. The averaged data is
updated at regular intervals and presented on the screen.
An even larger number of records can be accumulated simply by changing the number in the dialog.
However, the other parameters must be left unchanged or a new averaging calculation will be started. You
can pause the averaging by changing the trigger mode from NORM/AUTO to STOP. The instrument
resumes averaging when you change the trigger mode back to NORM/AUTO.
You can reset the accumulated average by pushing the CLEAR SWEEPS button or by changing an
acquisition parameter such as input gain, offset, coupling, trigger condition, timebase, or bandwidth limit.
The number of current averaged waveforms of the function, or its zoom, is shown in the acquisition status
dialog. When summed averaging is performed, the display is updated at a reduced rate to increase the
averaging speed (points and events per second).
Continuous Averaging
NOTE: Continuous Averaging may be set up from either the Channel dialog under Pre-Processing, or as a
Math function.
Continuous Averaging, the default setting, is the repeated addition, with unequal weight, of successive
source waveforms. It is particularly useful for reducing noise on signals that drift very slowly in time or
78
922499 Rev B
Operator's Manual
amplitude. The most recently acquired waveform has more weight than all the previously acquired ones:
the continuous average is dominated by the statistical fluctuations of the most recently acquired
waveform. The weight of ‘old' waveforms in the continuous average gradually tends to zero (following an
exponential rule) at a rate that decreases as the weight increases.
You determine the importance of new data vs. old data by assigning a weighting factor. Continuous
averaging allows you to make adjustments to a system under test and to see the results immediately.
The formula for continuous averaging is:
new average = (new data + weight * old average)/(weight + 1)
This is also the formula used to compute summed averaging. But by setting a "sweeps" value, you
establish a fixed weight that is assigned to the old average once the number of "sweeps" is reached. For
example, for a sweeps (weight) value of 4:
1stsweep (no old average yet): new average = (new data +0 * old average)/(0 + 1) = new data only
2ndsweep: new average = (new data + 1*old average)/(1 + 1) = 1/2 new data +1/2 old average
3rdsweep: new average = (new data + 2 * old average)/(2 + 1) = 1/3 new data + 2/3 old average
4thsweep: new average = (new data + 3 * old average)/(3 + 1) = 1/4 new data + 3/4 old average
5thsweep: new average = (new data + 4 * old average)/(4 + 1) = 1/5 new data + 4/5 old average
6thsweep: new average = (new data + 4 * old average)/(4 + 1) = 1/5 new data + 4/5 old average
7thsweep: new average = (new data + 4 * old average)/(4 + 1) = 1/5 new data + 4/5 old average
In this way, for sweeps > 4 the importance of the old average begins to decrease exponentially.
NOTE: The number of sweeps used to compute the average will be displayed in the bottom line of the
trace descriptor label:
Set Up Averaging
To quickly set up Continuous Averaging (only), access the Channel setup dialog and enter the number of
sweeps to average in Averaging. The valid range is 1 to 1,000,000 sweeps.
To apply Continuous or Summed Averaging as a Math function:
1. Follow the usual steps to set up a math fuction, selecting Average from the Basic Math submenu.
2. On the Average right-hand dialog, choose Summed or Continuous.
3. Touch Sweeps and provide a value. The valid range is 1 to 1,000,000 sweeps.
922499 Rev B
79
HDO6000 High Definition Oscilloscope
FFT
For a large class of signals, you can gain greater insight by looking at spectral representation rather than
time description. Signals encountered in the frequency response of amplifiers, oscillator phase noise and
those in mechanical vibration analysis, for example, are easier to observe in the frequency domain.
If sampling is done at a rate fast enough to faithfully approximate the original waveform (usually five
times the highest frequency component in the signal), the resulting discrete data series will uniquely
describe the analog signal. This is of particular value when dealing with transient signals because, unlike
FFT, conventional swept spectrum analyzers cannot handle them.
Because of its versatility, FFT analysis has become a popular analysis tool. However, some care must be
taken with it. In most instances, incorrect positioning of the signal within the display grid will significantly
alter the spectrum, producing effects such as leakage and aliasing that distort the spectrum.
An effective way to reduce these effects is to maximize the acquisition record length. Record length
directly conditions the effective sampling rate of the oscilloscope and therefore determines the frequency
resolution and span at which spectral analysis can be carried out.
Set Up FFT
1. Follow the usual steps to set up a math function, selecting FFT from the Frequency Analysis submenu.
2. Open the FFT right-hand dialog.
3. Choose to either:
l trunc(ate) - When the FFT transform size does not match the record length, truncate the record
and perform an FFT on the shorter record. This option increases the resolution bandwidth.
l zero-fill - When the source data for the FFT comes from a math operation that shortens the rec-
ord (as is commonly encountered in filtering operations like ERes), replace the missing data
points wityh data values whose amplitudes are interpolated to fit between the last data point
and the first data point in the record. This guarantees that there is not a first-order discontinuity
in the filled data. Since the data at the end of the record is filled data, it is advisable to select a
weighting window other than rectangular to minimize the effect of the fill on the resulting spectrum.
4. Check the Suppress DC box to make the DC bin go to zero. Otherwise, leave it unchecked.
5. Choose an Output type.
6. Optionally, choose a weighting Window. See the section below for more information about FFT
weighting windows.
7. Touch Algorithm and choose either:
80
922499 Rev B
Operator's Manual
l Least Prime (default) - a least primes algorithm that computes FFTs on transform sizes having
lengths that can be expressed as factors of 2N*5K. This is very compatible with the record
lengths encountered in the oscilloscope, which are often multiples of 1, 2, 4, 5, or 10.
l Power of 2 - a power of 2 algorithm where the record lengths are in the form of 2
N
. The power of
2 algorithm generally runs faster than the least primes algorithm. The price that is paid is a record length that is not the same as the acquired signal. The power of 2 FFT truncates to the nearest power of 2 less than record length (if truncate is chosen) or fill data to nearest power of 2
greater than the record length (if zero fill is selected).
8. Depending on your Output Type selection, you may also make selections for :
l Group Delay Shift
l Line Impedence - by default, the FFT function assumes that the oscilloscope is terminated in 50
Ohms. If an external terminator is being used, this setting can be changed to properly calculate
the FFT based on the new termination value.
Choosing a Window
The choice of a spectral window is dictated by the signal's characteristics. Weighting functions control
the filter response shape, and affect noise bandwidth as well as side lobe levels. Ideally, the main lobe
should be as narrow and flat as possible to effectively discriminate all spectral components, while all
side lobes should be infinitely attenuated. The window type defines the bandwidth and shape of the
equivalent filter to be used in the FFT processing.
Rectangular windows provide the highest frequency resolution and are thus useful for estimating the type
of harmonics present in the signal. Because the rectangular window decays as a (sinx)/x function in the
spectral domain, slight attenuation will be induced. Alternative functions with less attenuation (Flat Top
and Blackman-Harris) provide maximum amplitude at the expense of frequency resolution. Whereas,
Hamming and Von Hann are good for general purpose use with continuous waveforms.
Window TypeApplications and Limitations
RectangularThese are normally used when the signal is transient (completely contained in the time-
domain window) or known to have a fundamental frequency component that is an integer
multiple of the fundamental frequency of the window. Signals other than these types will
show varying amounts of spectral leakage and scallop loss, which can be corrected by
selecting another type of window.
Hanning (Von Hann)These reduce leakage and improve amplitude accuracy. However, frequency resolution is
also reduced.
HammingThese reduce leakage and improve amplitude accuracy. However, frequency resolution is
also reduced.
Flat TopThis window provides excellent amplitude accuracy with moderate reduction of leakage,
but with reduced frequency resolution.
Blackman-HarrisIt reduces the leakage to a minimum, but with reduced frequency resolution.
922499 Rev B
81
HDO6000 High Definition Oscilloscope
FFT Window Filter Parameters
Highest
Window Type
Rectangular-133.921.00.0
Von Hann-321.421.5-6.02
Hamming-431.781.37-5.35
Flat Top-440.013.43-11.05
Blackman-Harris-671.131.71-7.53
Side
Lobe
(dB)
Scallop Loss (dB)ENBW (bins)Coherent Gain (dB)
Copy Function
The Copy math function saves a copy of your present waveform in its unprocessed state to the first
available memory location. While processing may continue on the original waveform, the copy enables
faster throughput in some cases by preserving the original data. That is, no calculations need to be
undone on the copy before additional math can be calculated.
This benefit of faster throughput, however, comes at the expense of memory usage.
Follow the ususal steps to set up a math function, selecting Copy from the Misc submenu.
On the Wform Copy right-hand dialog, you can optionally Reset Count or Change BatchSize.
82
922499 Rev B
Operator's Manual
Analysis
Most Teledyne LeCroy oscilloscopes calculate measurements for all instances in the acquisition,
enabling you to rapidly and thoroughly analyze a long memory acquisition of thousands or millions of
parameter values to find anomalous measurements, or to apply a variety of mathematical functions to the
waveform trace.
These measurements and manipulations of the original input signal can be viewed in several graphical
formats to facilitate your analysis.
l Histograms display the distribution of measured values for a given parameter as a bar chart. See
About Histograms.
l Tracks provide a time-correlated view of a measurement parameter compared to other acquired
channels or calculated math traces. A common usage for track is to observe the modulation of a
signal, such as amplitude, frequency, or pulse width modulation. See View Track.
l Trends provide a view of a measurement parameter over an extended period of time and over mul-
tiple acquisitions. See View Trend and Track vs. Trend to better understand what a Trend provides
compared to a Track.
There are also conditional tests that can be applied to the data to find particular events:
l Pass/Fail Testing, including Pass/Fail Testing, finds normal/abnormal measurements as indicated
by whether or not they meet a set of defined criteria.
l WaveScan searches a single acquisition for events that meet specific criteria, enabling you to zoom
in on anomalies in the waveform, or scans multiple acquisitions with allowable trigger actions when
conditions are met. It can also be used to filter measurements. A variety of views help you understand the behavior of waveforms.
Finally, History Mode facilitates analysis by enabling you to quickly return the waveform display to any
point in an acquisition history.
Optional software packages may be purchases that simplify specialized analysis, such as various Serial
Data Decode options. These all add new methods to those available on the oscilloscope Analysis menu.
View Histogram
1. If you are not already on the Measure or Math dialog, choose Measure → Measure Setup... or Math →
Math Setup...from the menu bar.
2. Touch the tab for the measurement parameter or math function you wish to histogramand check
Trace On.
3. If you're already on the Fx dialog, touch the graph button and skip to Step 5.
OR
Touch the Histogram button at the bottom of the Px dialog and choose the math trace (F1-F12) in
which to display the histogram.
The histogram opens in a new grid along with its function descriptor box.
4. Touch the new Fx descriptor box to display the Fx dialog.
5. Touch the Histogram tab at the right to display the Histogram right-hand dialog.
922499 Rev B
83
HDO6000 High Definition Oscilloscope
6. Enter the maximum #Valuesin one bin of the histogram. This determines the number of samples that
are represented by the bar at full height.
7. Touch #Bins and enter the number of bins that comprise the histogram. This determines how many
bars appear in the histogram.
8. To let the oscilloscope determine the range of values represented by each bin/bar, check Enable AutoFind, then touch the Find Center and Width button.
OR
To set your own range, enter Center and Width values.
View Persistence Histogram
You can create a histogram of a persistence display, which graphs a horizontal or vertical “slice” of a
waveform.
NOTE: This math operation is different than the Histogram math operation and is not affected by Center
and Width settings made on any existing Histograms.
1. Choose Math → Math Setup... from the menu bar to access the Math dialog.
2. Touch an open Fx button and select Phistogram from the pop-up menu.
3. Touch the Fx tab to open the Function dialog, then touch Source1 and select a source trace from the
pop-up.
4. Touch the Phistogram tab at the right to open the Phistogram dialog.
5. Touch Slice Direction and select Horizontal or Vertical slice from the pop-up menu.
6. Touch Slice Center and use the pop-up keypad to enter a value.
7. Touch Slice Width and use the pop-up keypad to enter a value.
84
922499 Rev B
Operator's Manual
Track and Trend
Both Track and Trend are tools that can be used to plot measurement data and observe variations with
respect to time. Differences between Track and Trend are summarized in the following table:
CharacteristicTrackTrend
RepresentationParameter value vs. timeParameter value vs. event
BehaviorNon-cumulative (resets after every
acquisition). Unlimited number of
events
Time Correlation to Other DataYesNo
Monitors an Evolution in the
Frequency Domain
Monitors the Evolution of a
Measurement Parameter over Several
Acquisitions
Ensures No Lost Measurement DataYes. Maximum time period that can be
YesNo. Trend points are not evenly
No. Track resets after every
acquisition.
captured is limited by acquisition
memory and sampling rate.
Cumulative over several acquisitions
up to 1 million events
spaced in time and therefore cannot be
used for an FFT.
Yes
No. Since data can be accumulated
over many acquisitions, and since the
oscilloscope takes time to calculate
measurement values and to display
data before the trigger is re-armed,
data can be missed.
In general, Track is the tool to use if you want to capture a continuous stream of data spaced closely
together. To understand the change in a parameter with time, Trend can be used if your data is spaced
widely apart and longer than the dead-time of the oscilloscope between acquisitions. Think of Trend as a
strip chart recorder for your oscilloscope.
View Track
This procedure explains how to view the Track of a measurement parameter applied to a waveform. A
track is a waveform composed of parameter measurements that is time synchronous with the source
waveform. The vertical units are those of the source parameter and the horizontal units are seconds. In
order to maintain time synchronism, the parameter values are posted at the sampling rate. Track values
are redundant in that the same value is repeated every sample period until the measurement changes.
Although a Track plots measurement parameter values, it is created as a function and controlled on the
Math dialog.
1. If not already on the Measurement dialog, choose Measure → Measure Setup....
2. Touch the Px tab for the parameter you wish to plot.
3. Touch the Track button at the bottom of the Px dialog and select a math function (Fx) in which to draw
the Track.
The Track is displayed on a new grid, along with its function descriptor box.
4. To rescale the Track plot:
l Touch the Track function descriptor box to open the Fx dialog, then touch the Track tab.
922499 Rev B
85
HDO6000 High Definition Oscilloscope
l On theTrack right-hand dialog, uncheck Auto Find Scale and enter a new Center and Height/div.
View Trend
This procedure explains how to view the trend of a measurement parameter. A trend is a waveform
composed of a series of parameter measurements in the order the measurements were taken. The
vertical units are those of the source parameter, the horizontal unit is measurement number. The trend
contains a single value for each measurement.
Although the trend plots measurement values, the plot is drawn as a math function and controlled through
the Math dialog.
1. If you're not already on the Measure dialog, choose Measure → Measure Setup....
2. Touch the Px tab for the parameter you wish to plot.
3. Touch the Trend buttonat the bottom of the dialog and choose a math function Fx in which to draw the
Trend.
The Trend is displayed in a new grid, along with its function descriptor box.
4. To rescale the Trend plot:
l Touch the Trend function descriptor box to open the Math dialog, then touch the Trend tab at the
far right of the dialog.
l On the Trend right-hand dialog, uncheck Auto Find Scale and enter the new Center and Height
values.
86
922499 Rev B
Operator's Manual
WaveScan
The WaveScan®Search and Find tool enables you to search for unusual events in a single capture, or to
scan for a particular event in many acquisitions over a long period of time. Each Scan Mode is optimized
to find a different type of event. The results are time stamped, tabulated, and can be selected for
individual viewing.
There are two principal approaches to using WaveScan.
Capture & Search -- Make a single acquisition, then use Measurement Mode to search for parameter
measurements that fit your filter criteria.
Scan -- Set up the scan mode, then scan for matching events across multiple acquisitions.
Customize the presentation by choosing different WaveScan display features, or Scan Views. Optionally,
set Actions to occur automatically when unusual events are found, such as stopping the acquisition or
sounding an alarm.
NOTE: Whenever WaveScan is enabled, the instrument reverts to Real-time sampling mode.
922499 Rev B
WaveScan window with all scan "views" turned on.
87
HDO6000 High Definition Oscilloscope
Scan Modes
The scan mode determines the type of search to be performed. Select the Mode along with the Source
trace to be searched on the main WaveScan dialog.
For each mode, different controls appear on the WaveScan dialog, providing additional inputs to the
search criteria. Make the appropriate entries in these fields before starting the search.
EDGE MODE
Edge Mode is used for detecting the occurrence of edges. Events that meet the threshold level are
captured and tabulated. When the acquisition is stopped, scan filters can be applied to the edges to find
specific characteristics. Additional settings for Edge Mode are:
l Slope -- choose Pos, Neg, or Both.
l Level is -- choose Percent or Absolute.
l Percent/Absolute Level -- Enter a threshold value as a percentage of Top to Base or voltage level. A
marker displayed over the source trace indicates the level.
NON-MONOTONIC MODE
Non-monotonic Mode looks for edges that cross a threshold more than once between high and low levels.
All events that meet the criteria of slope, hysteresis, and level are presented in a table and highlighted in
the source trace. The value displayed in the table is the difference of the max. and min. of the nonmonotonicity. This can be confirmed with cursors. The hysteresis value is used to eliminate noise. A nonmonotonicity is detected only when its amplitude is greater than the hysteresis. Therefore, when setting a
hysteresis level, set a value that is greater than the amplitude of the noise. Additional settings for Nonmonotonic Mode are:
l Slope -- choose Pos, Neg, or Both.
l Hysteresis is -- choose Division, Percent, Absolute.
l Division/Percent/Absolute -- enter the hysterisis level in the units you selected.
l Levels are -- choose Percent, Absolute, or Pk-Pk%.
l High Level and Low Level -- Enter the top and bottom thresholds in the units you selected.
RUNT MODE
Runt Mode looks for pulses that fail to cross a specified threshhold. You can search for positive-going or
negative-going runts, or both. An adjustable hysteresis band is provided to eliminate noise.
In the case of negative-going runt pulses, the value displayed in the table is the difference (delta) of the
high level of the signal and the runt amplitude (i.e., where the runt bottoms out). This can be confirmed by
placing cursors on the runt pulse and reading the delta Y value in the trace labels. In the case of positivegoing runt pulses, the value displayed in the table is the absolute value of the amplitude of the runt pulse.
Additional settings for Runt Mode are:
l Runt Type -- choose Both, Pos, or Neg.
l Hysteresis -- enter the hysteresis level as a percentage or voltage.
l Low Threshold and High Threshold -- enter the levels as a percentage or voltage.
l Absolute Levels -- check this box if you want to enter levels as absolute voltage instead of per-
centage.
88
922499 Rev B
Operator's Manual
MEASUREMENT MODE
Measurement Mode is used for applying filters to measurements to find those that meet your defined
criteria, helping to isolate particular events within many samples. Markers appear over the source trace
to indicate the location of measurement, while the table displays values for the selected parameter that
meet the criteria. Additional Settings for Measurement Mode are:
l Measurement -- choose the measurement parameter you wish to search.
l Filter Method -- choose the operator that indicates the desired relationship to the Filter Limit. Only
measurements that meet this criteria are returned.
l Filter Limit -- enter the value that completes the filter criteria.
Alternatively, you can use the Filter Wizard to create the filter criteria.
SERIAL PATTERN MODE
Serial Pattern Mode is used for finding 2- to 64-bit patterns in digital sequences; ideal for bursted
patterns where a PLL cannot lock.Additional settings for Serial Pattern Mode are:
l Viewing -- choose to enter the pattern as Binary or Hex.
l Binary/Hex -- enter the pattern.
l Num. Patterns to detect -- enter a whole number.
BUS PATTERN MODE
Bus Pattern Mode (-MS models only) is used for finding 2- to 16-bit patterns across the digital lines.
Additional settings for Bus Pattern Mode are:
l Viewing -- choose to enter the pattern as Binary or Hex.
l Binary/Hex -- enter the pattern.
l Num. Patterns to detect -- enter a whole number.
Scan Views
Scan Views are different ways to view your WaveScan results. You can choose to display views
simultaneously or visit them sequentially. Just check the boxes at the bottom of the WaveScan dialog for
those views you wish to display. Uncheck the box to turn off the view.
NOTE: The number of grids displayed varies from one to three grids depending on which views are
enabled. WaveScan handles this function automatically, and there is no option to move traces from one
grid to another, as would be the case under normal operation.
You'll find additional controls for manipulating views like Scan Overlay and Zoom on their respective
dialogs. If you turn on these traces from their dialogs, you must turn them off from there, too.
SOURCE TRACE
By default, the source trace is displayed in the top grid, with markers indicating points in the trace that
meet the search criteria.
TABLE AND TIMES
Table view displays a table of measurements relevant to your chosen Search Mode next to the source
trace. Times adds columns to the table showing Start and Stop Times for each event.
922499 Rev B
89
HDO6000 High Definition Oscilloscope
SCAN OVERLAY
Scan Overlay view plots the location of captured events in a new trace.
To apply monochromatic persistence to the scan overlay:
1. Check Persistence On.
2. Enter aSaturation value. This controls...
3. Choose a Persistence Time. The higher the time, the more static the persistence display.
To rescale the scan overlay to effectively "zoom" in or out: touch the In/Out buttons, or touch Scale andDelay and enter new values. Check Var. to adjust values in finer steps than the default 1, 2, 5, 10.
Scan Histogram provides a statistical view of edges that meet your search criteria.
ZOOM
Zoom view works exactly as it does elsewhere in the oscilloscope software, opening a close-up of the
source trace in a new grid that you can rescale vertically and horizontally. A Zx tab appears by default
when you launch WaveScan; see Zoom Controls for an explanation of the remainder of the controls found
on this dialog.
One unique feature of the WaveScan Zoom is that you can automatically zoom the events captured from
the source trace by touching the Prev/Next buttons on the Zx dialog. You can also select the event from
the Table display, and you are automatically relocated to that event on the zoom trace.
90
922499 Rev B
Operator's Manual
Set Up WaveScan
This procedure explains how to set up WaveScan to search an acquisition for events of interest. Set up
your source channel and triggers before setting up the scan.
1. Press the Front Panel Stop button to stop acquisition.
2. Choose Analysis > WaveScan.
3. Check Enable.
4. Choose the Source waveform.
5. Choose the Scan Mode and enter values for any additional settings that appear at the right of the
dialog based on your selection.
6. If you're using Measurement Mode, set up the filter in one of the following ways:
l Touch Filter and choose an operator, then enter the Filter Limit.
l Touch Filter Wizard and choose one of the pre-set filters. The Filter and Filter Limit are auto-
matically set based on your selection.
7. Select each Scan View in which you wish to display results by checking the box at the bottom of the
dialog. Each view selected is displayed simultaneously.
8. If you're using Scan Overlay view, on the Scan Overlay dialog Clear Sweeps. If desired, set up the Per-
sistence display.
9. Optionally, choose an Action to trigger when an event that meets your scan criteria is found.
10. Restart acquisition.
922499 Rev B
91
HDO6000 High Definition Oscilloscope
History Mode
History Mode allows you to review any acquisition saved in the oscilloscope's history buffer, which
automatically stores all acquisition records until full. Not only can individual acquisitions be restored to
the grid, you can "scroll" backward and forward through the history at varying speeds to capture individual
details or changes in the waveforms over time.
Each record is indexed and time-stamped, and you can choose to view the absolute time of acquisition or
the time relative to when you entered History Mode. In the latter case, the last acquisition is time zero,
and all others are stamped with a negative time. The maximum number of records stored depends on
your acquisition settings and the size of the oscilloscope memory.
To view history:
1. Press the Front Panel History Mode button, or choose Timebase > History Mode.
2. Select View History to enable the history display, and View Table to display the index of records.
Optionally, select to view Relative Times on the table.
3. Choose a single acquisition to view by entering its Index number on the dialog or selecting it from the
table of acquisitions. You can also use the Navigation buttons or the slider bar at the bottom of the
dialog to "scroll" the history of acquisitions.
l The top row of buttons scrolls continuously and are (left to right): Fast Backward, Slow Back-
ward, Pause, Slow Forward, Fast Forward.
l The bottom row of buttons steps one record at a time and are (left to right): Back to Start, Back
One, Go to Index (#), Forward One, Forward to End.
Entering History Mode automatically stops new acquisitions. To leave History Mode, press the Front
PanelHistory Mode button again or clear the View History checkbox on the History dialog. Restart
acquisition by pressing one of the Front Panel Trigger Mode buttons.
92
922499 Rev B
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.