OVERVIEW
The model GFC7000EU is a close derivative of the Model GFC7000E CO Analyzer.
The optical bench is longer (14 m), making the instrument more sensitive at low CO
2
levels. The instrument has a user selectable full scale range of 0-100 PPB to 0-100 PPM
of CO2.
The instrument does not come equipped with a sample flow rate sensor. Hence, disregard
any reference to the flow sensor in the users manual.
Because the GFC7000EU is a close derivative of GFC7000E, the Model GFC7000E
Instruction Manual continues to be valid as the reference manual covering the details of
the instrument’s components. Only the sections of the manual that are different between
the GFC7000E and a GFC7000EU are listed in this addendum.
2. SPECIFICATIONS, AGENCY
APPROVALS, AND WARRANTY
2.1 Specifications
Table: Model GFC7000EU Specifications
User selectable to any full scale range from 0-100 ppb to 0-100 ppm
ppb, ppm, µg/m3, mg/m3 (user selectable)
Zero Noise
< 0.5% of reading RMS over 5 ppm
(1) (3)
< 0.5% of reading
(2) (4)
Better than 1% of Range
(5)
800 scc/min. ± 10%
Temperature Range
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE
(Addendum to GFC7000E Manual PN 04584)
05134 Rev -
0-95% RH, Non-Condensing
Temp Coefficient
< 0.05 % of reading per °C (5 ppb/°C minimum)
(6)
< 0.05 % of reading per V
Dimensions
(HxWxD)
7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)
100V 50/60 Hz (3.25A), 115 V 60 Hz (3.0A),
220 – 240 V 50/60 Hz (2.5A)
Environmental
Conditions
Installation Category (Over voltage Category) II
Pollution Degree 2
Analog Outputs
Three (3) Outputs
Analog Output
Ranges
100 mV, 1 V, 5 V, 10 V, 2-20 or 4-20 mA isolated
current loop.
All Ranges with 5% Under/Over Range
Analog Output
Resolution
1 part in 4096 of selected full-scale voltage
8 Status outputs from opto-isolators
Control Inputs
6 Control Inputs, 2 defined, 4 spare
I/O
One (1) RS-232; One (1) RS-485/RS-232/Ethernet
Baud Rate : 300 - 115200
Certifications
CE: EN61010-1:90 + A1:92 + A2:95, EN61326 - Class
A
(1)
As defined by the USEPA
(2)
At constant temperature and voltage
(3)
Or 0.2 ppm, whichever is greater
(4)
Or 0.1 ppm, whichever is greater
(5)
Above 10 ppm range, otherwise 0.2 ppm for lower ranges
3.1.2. Pneumatic Connections
This section is similar to the one in the manual, with the following exception. The
extra port labeled “From Purge” is the exhaust port for the for the purge gas. The
gas needs to exhausted with a gas line not more than 10 meters long.
In addition, the GFC7000EU uses a different sample pressure sensor than the
GFC7000E. The GFc7000EU pressure sensor is rated for oxygen service. The part
number is listed in the appendix – spare parts.
6.13.1.3 Relay alarm outputs