Teledyne GFC7000EU User Manual

MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE
(Addendum to GFC7000E Manual PN 04584)
05134 Rev -
1
MODEL GFC7000EU
CO2 ANALYZER
(Addendum to GFC7000E Manual PN 04584)
Teledyne Analytical Instruments
16830 Chestnut St.
City of Industry, Ca. 91748
TEL: 626-961-9221 FAX: 626-961-2538 TEL: 626-934-1500 E-MAIL: ai-customerservice@teledyne.com
WEB SITE: http://www.teledyne-ai.com
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE
(Addendum to GFC7000E Manual PN 04584)
05134 Rev -
2
OVERVIEW
The model GFC7000EU is a close derivative of the Model GFC7000E CO Analyzer. The optical bench is longer (14 m), making the instrument more sensitive at low CO
2
levels. The instrument has a user selectable full scale range of 0-100 PPB to 0-100 PPM of CO2. The instrument does not come equipped with a sample flow rate sensor. Hence, disregard any reference to the flow sensor in the users manual.
Because the GFC7000EU is a close derivative of GFC7000E, the Model GFC7000E Instruction Manual continues to be valid as the reference manual covering the details of the instrument’s components. Only the sections of the manual that are different between the GFC7000E and a GFC7000EU are listed in this addendum.
2. SPECIFICATIONS, AGENCY APPROVALS, AND WARRANTY
2.1 Specifications
Table: Model GFC7000EU Specifications
Ranges
User selectable to any full scale range from 0-100 ppb to 0-100 ppm
Measurement Units
ppb, ppm, µg/m3, mg/m3 (user selectable) Zero Noise
0.01 ppm RMS
(1)
Span Noise
< 0.5% of reading RMS over 5 ppm
(1) (3)
Lower Detectable Limit
0.02 ppm
(1)
Zero Drift (24 hours)
< 10 ppb
(2)
Zero Drift (7 days)
< 20 ppb
(2)
Span Drift (24 hours)
< 0.5% of reading
(2) (4)
Span Drift (7 days)
< 1% of reading
(2) (4)
Linearity
Better than 1% of Range
(5)
Precision
0.5% reading
(1) (5)
Lag Time
<10 sec
(1)
Rise/Fall Time
<60 sec to 95%
(1)
Sample Flow Rate
800 scc/min. ± 10% Temperature Range
5 - 40°C operating
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE
(Addendum to GFC7000E Manual PN 04584)
05134 Rev -
3
Humidity Range
0-95% RH, Non-Condensing Temp Coefficient
< 0.05 % of reading per °C (5 ppb/°C minimum)
(6)
Voltage Coefficient
< 0.05 % of reading per V Dimensions (HxWxD)
7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)
Weight
50 lb (22.7 kg) AC Power
100V 50/60 Hz (3.25A), 115 V 60 Hz (3.0A), 220 – 240 V 50/60 Hz (2.5A) Environmental Conditions
Installation Category (Over voltage Category) II Pollution Degree 2 Analog Outputs
Three (3) Outputs Analog Output Ranges
100 mV, 1 V, 5 V, 10 V, 2-20 or 4-20 mA isolated current loop. All Ranges with 5% Under/Over Range Analog Output Resolution
1 part in 4096 of selected full-scale voltage
Status Outputs
8 Status outputs from opto-isolators Control Inputs
6 Control Inputs, 2 defined, 4 spare I/O
One (1) RS-232; One (1) RS-485/RS-232/Ethernet Baud Rate : 300 - 115200 Certifications
CE: EN61010-1:90 + A1:92 + A2:95, EN61326 - Class A
(1)
As defined by the USEPA
(2)
At constant temperature and voltage
(3)
Or 0.2 ppm, whichever is greater
(4)
Or 0.1 ppm, whichever is greater
(5)
Above 10 ppm range, otherwise 0.2 ppm for lower ranges
3.1.2. Pneumatic Connections
This section is similar to the one in the manual, with the following exception. The extra port labeled “From Purge” is the exhaust port for the for the purge gas. The gas needs to exhausted with a gas line not more than 10 meters long.
In addition, the GFC7000EU uses a different sample pressure sensor than the GFC7000E. The GFc7000EU pressure sensor is rated for oxygen service. The part number is listed in the appendix – spare parts.
6.13.1.3 Relay alarm outputs