Teledyne GC-Pro-TCD User Manual

INSTRUCTION, OPERATING AND
MAINTENANCE MANUAL FOR
GC-PRO TCD
P/N M89443
DATE 4/05/13
a time even after the power is turned off and disconnected. Only authorized personnel should conduct maintenance and/or servicing. Before
conducting any maintenance or servicing, consult with authorized supervisor/manager.
Teledyne Analytical Instruments
DANGER
GC-Pro TCD
Copyright © 2013 Teledyne Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91748.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free from defects of material and of construction, and that our liability shall be limited to replacing or repairing at our factory (without charge, except for transportation), or at customer plant at our option, any material or construction in which defects become apparent within one year from the date of shipment, except in cases where quotations or acknowledgements provide for a shorter period. Components manufactured by others bear the warranty of their manufacturer. This warranty does not cover defects caused by wear, accident, misuse, neglect or repairs other than those performed by Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and the purchaser by the acceptance of the equipment will assume all liability for any damage which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its use r, an d serves as a tool b y whic h valuable data can be gathered. The information provided by the instrument may assist the user in eliminating potential hazards caused by his process; however, it is essential that all personnel involved in the use of the instrument or its interface be properly trained in the process being measured, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process conditions. While this instrument may be able to provide early warning of imminent danger, it has no control over process conditions, and it can be misused. In particular, any alarm or control systems installed must be tested and understood, both as to how they operate and as to how they can be defeated. Any safeguards required such as locks, labels, or redundancy, must be provided by the user or specifically requested of Teledyne at the time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The purchaser is responsible for the training of personnel, for providing hazard warning methods and instrumentation per the appropriate standards, and for ensuring that hazard warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot accept responsibility for conditions beyond its knowledge and control. No statement expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the user’s process conditions.
Teledyne Analytical Instruments ii
GC-Pro TCD
Safety Messages
Your safety and the safety of others are very important. We have provided many important safety messages in this manual. Please read these messages carefully.
A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol. These symbols are found in the manual and inside the instrument. The definition of these symbols is described below:
GENERAL WARNING/CAUTION: Refer to the instructions for details on the specific danger. These cautions warn of specific procedures which if not followed could cause bodily Injury and/or damage the instrument.
No
Symbol
CAUTION: HOT SURFACE WARNING: This warning is specific to heated components within the instrument. Failure to heed the warning could result in serious burns to skin and underlying tissue.
WARNING: ELECTRICAL SHOCK HAZARD: Dangerous voltages appear within this instrument. This warning is specific to an electrical hazard existing at or nearby the component or procedure under discussion. Failure to heed this warning could result in injury and/or death from electrocution.
Technician Symbol: All operations marked with this symbol are to be performed by qualified maintenance personnel only.
NOTE: Additional information and comments regarding a specific component or procedure are highlighted in the form of a note.
STAND-BY: This symbol indicates that the instrument is on Stand-by but circuits are active.
Teledyne Analytical Instruments iii
GC-Pro TCD
CAUTION: THE ANALYZER SHOULD ONLY BE USED FOR THE
PURPOSE AND IN THE MANNER DESCRIBED IN THIS MANUAL.
IF YOU USE THE ANALYZER IN A MANNER OTHER THAN THAT FOR WHICH IT WAS INTENDED, UNPREDICTABLE BEHAVIOR COULD RESULT POSSIBLY ACCOMPANIED WITH HAZARDOUS CONSEQUENCES.
This manual provides information designed to guide you through the installation, calibration and operation of your new analyzer. Please read this manual and keep it available.
Occasionally, some instruments are customized for a particular application or features and/or options added per customer requests. Please check the front of this manual for any additional information in the form of an Addendum which discusses specific information, procedures, cautions and warnings that may be specific to your instrument.
Manuals do get misplaced. Additional manuals can be obtained from Teledyne at the address given in the Appendix. Some of our manuals are available in electronic form via the internet. Please visit our website at: www.teledyne-ai.com.
Teledyne Analytical Instruments iv
GC-Pro TCD
Additional Safety Information
DANGER
COMBUSTIBLE GAS USAGE
This is a general purpose instrument designed for use in a non-hazardous area. It is the customer's responsibility to ensure safety especially when combustible gases are being analyzed since the potential of gas leaks always exist.
WARNING
The customer should ensure that the principles of operating of this equipment are well understood by the user. Misuse of this product in any manner, tampering with its components, or unauthorized substitution of any component may adversely affect the safety of this instrument.
Since the use of this instrument is beyond the control of Teledyne, no responsibility by Teledyne, its affiliates, and agents for damage or injury from misuse or neglect of this equipment is implied or assumed.
WARNING: If toxic or flammable gases are used, always purge
the entire system before performing any maintenance and always leak check the system after removing any tubing or fittings on the sample system. See the procedures for purging and leak checking this instrument on the following pages.
If toxic gases or other hazardous materials are introduced into the sample system, the same precautions regarding leak checking and purging apply to the sample lines and sample supply or delivery lines.
Teledyne Analytical Instruments v
GC-Pro TCD
WARNING: ELECTRICAL SHOCK HAZARD.
ONLY AUTHORIZED AND SUITABLY TRAINED PERSONNEL SHOULD PERFORM WORK INSIDE OF THE INSTRUMENT. COMPONENTS WITHIN THE COVER, INSIDE THE ISOTHERMAL CHAMBER (SAMPLE SYSTEM), AND ON PC BOARDS MAY CONTAIN DANGEROUSLY HIGH VOLTAGE SUFFICIENT TO CAUSE SERIOUS INJURY OR DEATH.
There are the following three types of inaccessible
shock hazards within the Analyzer:
1. Line voltages and line related voltages such as 115 VAC which exists within the 230 VAC versions as well. These voltages stop when the Analyzer is turned off and the mains (line) cord is removed from the instrument.
2. External hazardous voltages which may be connected to the Analyzer alarm relay connections.
WARNING: THIS INSTRUMENT IS DESIGNED TO BE OPERATED
IN A NONHAZARDOUS AREA. DEPENDING ON THE APPLICATION, TOXIC GASES OR FLAMMABLE/EXPLOSIVE GASES MAY BE PRESENT. IT IS THEREFORE, THE CUSTOMER'S RESPONSIBILITY TO ENSURE THAT PROPER TRAINING AND UNDERSTANDING OF THE PRINCIPLES OF OPERATION OF THIS EQUIPMENT ARE UNDERSTOOD BY THE USER. SINCE THE USE OF THIS INSTRUMENT IS BEYOND THE CONTROL OF TELEDYNE, NO RESPONSIBILITY BY TELEDYNE, ITS AFFILIATES AND AGENTS FOR DAMAGE OR INJURY RESULTING FROM MISUSE OR NEGLECT OF THIS INSTRUMENT IS IMPLIED OR ASSUMED. MISUSE OF THIS PRODUCT IN ANY MANNER, TAMPERING WITH ITS COMPONENTS OR UNAUTHORIZED SUBSTITUTION OF ANY COMPONENT MAY ADVERSELY AFFECT THE SAFETY OF THIS INSTRUMENT.
CAUTION: WHEN OPERATING THIS INSTRUMENT, ALL
COVERS SECURELY FASTENED. THE GAUGES
Teledyne Analytical Instruments vi
GC-Pro TCD
MUST BE IN PROPER WORKING ORDER. DO NOT OVERPRESSURIZE THE SYSTEM.
READ THIS MANUAL BEFORE OPERATING THE
INSTRUMENT AND ADHERE TO ALL WARNINGS INCLUDED IN THIS MANUAL.
Teledyne Analytical Instruments vii
GC-Pro TCD
Table of Contents
Safety Messages .......................................................................... iii
Additional Safety Information ...................................................... v
List of Figures ............................................................................. xii
List of Tables .............................................................................. xiii
Introduction ................................................................................... 1
1.1 Main Features of the Analyzer 1
1.2 Principle of Operation 3
1.3 Analyzer Description 3
1.4 Applications 7
Operational Theory ....................................................................... 9
2.1 Introduction 9
2.2 Modes of Operation 10
2.3 Analyzer Subsystems 11
2.3.1 Sample System 11
2.3.2 Gas Flow Control System 12
2.3.3 Gas Separation System 14
2.4 Detector Cell 14
Installation ................................................................................... 17
3.1 Unpacking the Analyzer 17
3.2 Mounting the Analyzer 17
3.3 User Connections 18
3.3.1 Electrical Power Connections 18
3.3.2 Electronic Connections 18
3.3.2.1 Primary Input Power 19
3.3.2.2 Fuse Installation 20
Teledyne Analytical Instruments viii
GC-Pro TCD
3.3.2.3 50-Pin Equipment Interface Connector 20
3.3.2.4 Analog Output 20
3.3.2.5 Alarm Relays 22
3.3.2.6 Digital Remote Cal Inputs 24
3.3.2.7 ‘Measure Once’ (or one shot measurement) contacts 25
3.3.2.9 Pin Out Table 25
3.3.2.10 RS-232 Port 28
3.3.3 Gas Connections 29
3.3.3.1 Effluent 31
3.3.3.2 Sensor Vent 31
3.3.3.3 Carrier Gas Connection 31
3.3.4 Sample and Span Gas Connections 31
3.4 Placing the System in Operation 32
Operation ..................................................................................... 33
4.1 Equipment 34
4.2 Preliminary Power-Off Check List 34
4.3 Powering Up the Unit 35
4.4 Activating the Support Gases 38
4.4.1 Carrier Gas 38
4.4.2 Span Gas 38
4.5 Analyzer Operation 38
4.5.1 Default Parameters 39
4.5.2 The HOME Screen 40
4.5.3 The MENU Screen 41
4.5.4 Standby 42
4.5.5 Overlay Chromatogram 44
4.5.6 Timing 46
4.5.6.1 TCD Amplifier Gain 50
4.5.7 Temperature 50
Teledyne Analytical Instruments ix
GC-Pro TCD
4.5.7.1 Thermocouple 51
4.5.7.2 Settings 52
4.5.7.3 PID Settings 53
4.5.7.4 Status 54
4.5.8 Group Setup 55
4.5.9 Analog Adjust 58
4.5.10 Calibration 59
4.5.10.1 Span 60
4.5.10.2 Alt Span 62
4.5.10.3 Auto cal 65
4.5.11 Self Test 66
4.5.12 Settings 68
4.5.12.1 Alarms 69
4.5.12.2 Range 72
4.5.12.3 Change Stream 74
4.5.12.4 Time 75
4.5.12.5 Password 76
4.5.12.6 Communication 76
4.5.12.7 Detector 78
4.5.12.8 Set TCD 79
4.5.12.9 Cycle Mode 80
4.5.13 Analysis Mode 80
4.5.13.1 T-Time 81
4.5.13.2 T_Cycle 82
4.5.13.3 Chromatogram 83
4.5.13.4 VS Count 85
4.5.13.5 Cold Boot 85
Maintenance & Troubleshooting ................................................ 89
Appendix ...................................................................................... 91
A.1 Specifications and Initial Settings: 91
Teledyne Analytical Instruments x
GC-Pro TCD
A.2 Recommended Spare Parts List 92 A.3 Drawing List 93
Appendix B .................................................................................. 95
B1. VNC for GC-Pro TCD 95
Appendix C ................................................................................ 101
C1 Addendum and Testing Results 101
Teledyne Analytical Instruments xi
GC-Pro TCD
List of Figures
Figure 1-1: GC-Pro TCD Front Panel .............................................. 4
Figure 1-2: GC-Pro TCD Rear Panel .............................................. 5
Figure 1-3: Internal PCB Arrangement ............................................ 6
Figure 1-4: GC-Pro TCD Internal View ............................................ 6
Figure 2-1: Internal Temperature Controllers ................................ 12
Figure 2-2: Typical Piping Diagram Used in the GC-Pro TCD ....... 13
Figure 2-3: Thermal Conductivity Cell Operating Principle ............ 14
Figure 3-1: GC-Pro TCD Rear Panel with Optional Gas Manifold . 19
Figure 3-2: Equipment Interface Connector Pin Arrangement....... 20
Teledyne Analytical Instruments xii
GC-Pro TCD
List of Tables
Table 3-1: Analog Output Connections — Middle Connector ........ 21
Table 3-2: Analog Concentration Output—Example ...................... 22
Table 3-3: Alarm Relay Contact Pins — Top Connector ............... 23
Table 3-4: Remote Calibration Connections — Bottom Connector 24
Table 3-5: ‘Measure Once’ Relay Connections ............................. 25
Table 3-6: Pin out of Alarm Relay O/P (Top) 50 pin D-Sub
Connector .................................................................... 26
Table 3-7: Pin out of Analog Signal (Middle) 50 pin D-Sub
Connector .................................................................... 26
Table 3-8: Pin out of Standard (Bottom) 50 pin D-Sub Connectors27
Table 3-9: Commands via RS-232 Input ....................................... 28
Table 3-10: Required RS-232 Options .......................................... 28
Teledyne Analytical Instruments xiii
GC-Pro TCD
Teledyne Analytical Instruments xiv
GC-Pro TCD Introduction
Introduction
The GC-Pro TCD Analyzer is a microprocessor controlled digital instrument incorporating a Thermal Conductivity Detector (TCD) coupled with a gas separation column and switching valve designed to measure the quantity of aromatic hydrocarbons present in a positive pressure sample gas. Up to 8 channels of output are present for measuring specific analytes in a sample stream. In using a thermal conductivity detector after gas separation through a column, the range of analysis is extended to include additional phase separable inert, or other difficult to analyze gas phase components like N2, H2. etc. that are present in a positive pressure sample gas.
The GC-Pro TCD Analyzer has multiple channels for measurement with enhanced versatility and capabilities. Due to its flexibility, it is vitally important that you refer to Appendix C for the specific factory setup for your application.
The GC-Pro TCD features a modern user interface with a touch screen front panel facilitating operation, calibration, data collection and display. With the Ethernet connectivity feature, this new interface allows the instrument to be operated remotely with a VNC software application from another digital device whereby the analyzer can be accessed, controlled, data displayed etc. as if the analyzer were physically present.
Although the analyzer arrives already setup for your specific application, a gas chromatograph mode allows the user to program the instrument for measuring different species present in the sample.
1.1 Main Features of the Analyzer
The GC-Pro TCD Analyzer is sophisticated yet simple to use. A touch screen display on the front panel provides access all phases of setup, calibration, operation, and troubleshooting on the GC-Pro TCD.
The main features of the analyzer include:
Stable, accurate thermal conductivity detector for analyzing
inert gases and other species not amenable to flame ionization detection.
Teledyne Analytical Instruments 1
Introduction GC-Pro TCD
Data acquisition and control functions supporting:
Pressure measurements, 5-Channels (Future option) Flow (MFC) measurements, 3-Channels (Future option) Oxygen measurements, 6-Channels (Future option) Temperature measurements and control, 4-Channels 6-Channel Dew point Hygrometer measurements, 6-
Channel (Future option)
Relay, (10) and solenoid outputs, (4) Alarm, (3) and over range indication outputs, (4) 4-20mA current loop outputs, 10 channels HMI through Maple touch sensitive display
The system operates at single phase AC, 110 or 220V 50/60
Hz (select at time of order).
High resolution, accurate readings of concentration. Versatile analysis with three user-definable analysis ranges. Microprocessor based electronics: 8-bit CMOS
microprocessor with 32 kB RAM and 128 kB ROM.
Auto ranging allows analyzer to automatically select the
proper preset range for a given measurement. Manual override allows the user to lock onto a specific range of interest.
Two adjustable concentration alarms and a system failure
alarm.
Extensive self-diagnostic testing at startup and on demand
with continuous power supply monitoring.
RS-232 serial digital port for use with a computer or other
digital communication device.
8-Analog outputs for concentration and range identification
(0-1 VDC standard and isolated 4-20 mA dc).
Superior Accuracy
Teledyne Analytical Instruments 2
GC-Pro TCD Introduction
1.2 Principle of Operation
The analyzer uses a sample valve which is a specialized electronic­controlled valve responsible for extracting a precise volume of sample and delivering it to the separation column. This valve uses four micro cavities and rotates between two positions in a timed fashion controlled by the timing function in the control section.
The gas separation column separates the component of interest in the sample gas based on its retention time in the packed column. Using a carrier gas and a microprocessor actuated switching valve, the eluted gas is analyzed using a Thermal Conductivity Detector (TCD). After each injection cycle the column is back flushed to accept the fresh sample in the next cycle. The result is a series of peaks over time corresponding to the detector output for each sample cycle. The integrated area under the peak is directly related to concentration and after signal processing, this result is displayed in the appropriate units on the screen. The screen is updated after each cycle.
1.3 Analyzer Description
The standard analyzer is a rack mountable instrument designed to fit into a standard 19” instrument rack. The front interface screen is mounted on the left side of the panel. The right side of the panel includes the gas controls, pressure gauges and flowmeter. The front panel is shown in Figure 1-1.
Teledyne Analytical Instruments 3
Introduction GC-Pro TCD
Figure 1-1: GC-Pro TCD Front Panel
Gas pressure gauges and controls are mounted on the front panel adjacent to the touch screen (human/machine interface abbreviated as HMI) display as shown in Figure 1-1.
At the rear of the instrument are ports for the introduction of air, zero, span, and sample gas as well as vent port connections. Three 50­pin user-interface cable connectors route input/output and alarm signals to external devices. RS232 and a network port are also available at the rear panel for connection to a remote computer or other digital communication device. Internally, there are two USB user connections which are used for updating firmware to the PCBs. Figure 1-2 shows the rear panel including the user connections.
The analyzer is set up for either 120 VAC 60 Hz or 230 50/60 Hz operation depending on the customer’s requirements. The appropriate power cord for your unit is included with the analyzer.
Teledyne Analytical Instruments 4
GC-Pro TCD Introduction
Figure 1-2: GC-Pro TCD Rear Panel
Internally, there are three isothermal chambers each separately controlled for precise control and separation of analyzed species. In the electronics section, internal PC boards are stacked in an assembly in an order based on inter connectivity needs between the PCBs as well as their association with other sub assemblies distributed within the system. See Figure 1-3.
Following is the card stack-up from top to bottom planned inside the system
Heater card 9261 controller card Analog data acquisition card Analog signal conditioning card 89C5131 controller card
Teledyne Analytical Instruments 5
Introduction GC-Pro TCD
Figure 1-3: Internal PCB Arrangement
Figure 1-4: GC-Pro TCD Internal View
Teledyne Analytical Instruments 6
GC-Pro TCD Introduction
1.4 Applications
Monitoring the benzene concentration of carbon dioxide gas
for use in the beverage industry.
Monitoring aromatic contamination in air liquefaction and
other gas production processes.
Gas purity certification. Detecting trace benzene in ambient air. Detecting atmospheric pollutants.
Teledyne Analytical Instruments 7
GC-Pro TCD Operational Theory
Operational Theory
2.1 Introduction
The GC-Pro TCD Analyzer uses a Thermal Conductivity (TC) Detector and a Gas Chromatograph (GC) Column to separate and analyze different factions of a volatile hydrocarbon sample. Using a carrier gas and a microprocessor actuated switching valve, a fixed volume of sample is pushed into the column maintained at a constant temperature. The eluted gas is analyzed for the specific components configured for your application using the TC detector and the column is back flushed to accept the fresh sample for the next cycle. Actual separating and detecting sequence may vary depending on compounds of interest in the application for which the analyzer is configured.
The GC-Pro TCD Analyzer uses the thermal conductivity detector to sense a variety of components that are separated through a column and eluted to the sensor in a time sequence determined by their respective retention time. It compares the difference in conductivity between a known volume of injected sample driven by carrier gas at a fixed temperature to a constant flowing reference gas. Thermal conductivity is a fast, accurate and reproducible tool for detection and measurement under certain conditions. In general, thermal conductivity measurements are binary and non-specific in nature, that is, they respond to the total difference in thermal conductivity between two samples. If the sample contains more than two components, the thermal conductivity measurement is ill defined and cannot be used to determine the concentration of the mixture without extensive calibration. Similarly, if the thermal conductivity difference between components of a binary mixture is low, sensitivity is negatively impacted.
The GC-Pro TC Analyzer avoids these issues by coupling the TC detector with a gas separation capability in the GC column. Therefore, even if the sample stream contains more than one compound, at any time during analysis only a binary mixture which includes the component of interest based on its separation properties, plus carrier gas is delivered to the detector. A suitable carrier gas can be selected that among other properties, maximizes the thermal conductivity difference for enhanced
Teledyne Analytical Instruments 9
Operational Theory GC-Pro TCD
sensitivity. And since the sampling valve is designed to inject a precise and known volume of sample at a fixed temperature, the actual concentration of the compound of interest can be determined.
A stainless steel packed column containing Chromosorb Diatomite or other material depending on the application, is held at a constant temperature. The temperature as well as the carrier gas and its pressure are critical and depend on the application. The actual temperature setpoint and compound separation for your instrument have been determined at the factory. They are listed in the Addendum and Testing Results section of Appendix C.
Using a specific carrier gas to inject the sample gas into a known in volume, the specific component of interest elutes from the column within a few minutes. The exact timing is characteristic of the material and has been determined at the factory during testing. Additionally, a clear separation is observed from other components in the sample gas however each has a characteristic timing so the instrument can select only the peak associated with the compound of interest for processing. All necessary information regarding carrier gas type, sample loop volume, temperature and pressure settings, and internal timing settings are given in the Addendum and Testing Results section of Appendix C.
2.2 Modes of Operation
The analyzer has 2 modes of operation depending on the position of the GC Sampling Valve (See Piping Diagram in Figure 2-2). They are: Sample Mode (position A) and Analysis Mode (position B).
1. Valve Position A—Sampling Mode
In this mode the analyzer configures the operational valve to back flush the column and charge the sample loop. The Sample Mode is programmed to continue for a 7 minute period by factory default; however, it is usually modified to match the needs of a particular process. The specific timing interval for your system is listed in the Addendum and Testing Results section of Appendix C. It is possible to set the Sample Mode duration up to 25 minutes.
2. Valve Position B—Analysis Mode
In this mode the analyzer configures the sampling valve to feed the gas in the sample loop through the column and to the detector. The eluted sample from the column is fed to the TC
Teledyne Analytical Instruments 10
GC-Pro TCD Operational Theory
detector for the analysis of the compound of interest in the stream. If configured to do so, the analyzer may read the detector for a programmed 5 second period at some point within this mode cycle to obtain a baseline reading for further calculation. This mode is usually programmed to continue until the compound of interest is fully eluted from the column. It is possible to set it up to 25 minutes.
During the Analysis Mode, a ‘Peak Detect’ period is programmed at which time the analyzer reads the detector output. The analyzer integrates the peak area during this time in conjunction with the baseline settings to calculate the concentration of the compound of interest. The calculation is performed at the end of the ‘Peak Detector’ period and the result is displayed at the end of Analysis Mode.
2.3 Analyzer Subsystems
The Model Analyzer is composed of four subsystems:
1. Sample System
2. GC Column
3. Detector Cell
4. Electronic Signal Processing, Display and Control
2.3.1 Sample System
All components used to control the sample and supporting gases are located on the front panel or inside the analyzer chassis. For the internal components, they are accessible after removing the top cover of the analyzer. Adjustments are made to the carrier gas pressure and sample flow using the controls on the front panel. See Figure 1-1. Other adjustments are made using the touch screen and are described in Section 4.
The analyzer contains three separate isothermal chambers ‘SAMPLE’, ‘Sensor’, and ‘COLUMN’ that are controlled individually. Temperature and control information are available onscreen. See Section
4.5.7. The chambers and the switching valve are identified in Figure 2-
1.
Teledyne Analytical Instruments 11
Operational Theory GC-Pro TCD
Figure 2-1: Internal Temperature Controllers
The sample chamber contains the 10-port GC switching valve and 2 sample loops. The ‘TC Sensor’ chamber contains the thermal conductivity sensor. The ‘COLUMN’ is housed in a separate ‘COLUMN’ enclosure and maintained at a temperature of 70° C. Chamber temperatures have been set at the factory however they can be changed by the user from the Settings/Thermocouple/Settings screen as described in Section 4.5.7.2. The actual temperature setpoint for your instrument may be different depending on the application. It is listed in the Addendum and Testing Results section of Appendix C.
2.3.2 Gas Flow Control System
The analyzer is equipped with ports for the introduction of carrier gas, span, and sample gas. Depending on the application and the options chosen, some instruments may have a separate port for admitting a purge gas.
It is imperative that the sample, carrier and span gases be supplied at constant pressure using two stage stainless steel diaphragm gas regulators. The recommended pressure range is 30 to 80 psig for sample and carrier gas; the span gas should be supplied at a pressure of 20 psig
Teledyne Analytical Instruments 12
GC-Pro TCD Operational Theory
to the restrictor fitting either on the optional auto calibration module or the span inlet on the rear panel.
The Piping Diagram for the standard instrument is shown in Figure 2-2. A 10-port 2-position GC sampling valve is used to control and direct gas flows including sampling, back flush, and carrier gas. The fixed volume sample loop ensures the same volume of sample injection in the column every cycle.
Figure 2-2: Typical Piping Diagram Used in the GC-Pro TCD
If your instrument is fitted with the optional auto calibration module, a separate compartment containing of a pair of solenoid valves is installed for controlling the introduction of sample or span gas to the detector. Calibration can be performed automatically on a programmed schedule or manually using the front panel interface.
Other detector options are available for the GC-Pro series instrument which may alter the appearance of the front and rear panels and include additional gas ports such as fuel and air for the FID option. These options, if included, will be described in an accompanying addendum to this manual.
Teledyne Analytical Instruments 13
Operational Theory GC-Pro TCD
2.3.3 Gas Separation System
The carrier gas pressure and column temperature are the critical parameters in the separation process. A heater is used in column housing and the temperature controlled by the Temperature PCB. A constant carrier gas flow is provided by a pressure regulator on the front panel set to a specific pressure with an inlet carrier gas supply of 80 psig at the rear panel. Thus, a stable flow is achieved by maintaining a constant pressure across restrictors upstream from the cell. Actual carrier gas flow rate through the column, which is crucial for separation retention time, will depend on the carrier gas pressure and the type of column used for each application. It is important that all temperature and pressure settings are set to the values listed in the Addendum and Testing Results section of Appendix C.
2.4 Detector Cell
The thermal conductivity sensor contains two chambers, one for the reference gas of known conductivity (which is of the same as the carrier gas) and one for the sample gas which is driven by the carrier gas. Each chamber contains a pair of heated filaments. Depending on its thermal conductivity, each of the gases conducts a quantity of heat away from the filaments in its chamber. See Figure 2-3(a).
The resistance of the filaments depends on their temperature. These filaments are parts of the two legs of a bridge circuit that unbalances if the resistances of its two legs do not match. See Figure 2-1(b).
Figure 2-3: Thermal Conductivity Cell Operating Principle
Teledyne Analytical Instruments 14
GC-Pro TCD Operational Theory
Teledyne Analytical Instruments 15
Loading...
+ 85 hidden pages