OPERATING INSTRUCTIONS FOR
Model 3300ZB
Trace Oxygen Analyzer
P/N M3300ZB
2/11/2014
DANGER
for a time even after the power is turned off and disconnected. Only authorized personnel should conduct maintenance and/or servicing.
Before conducting any maintenance or servicing, consult with authorized supervisor/manager.
Teledyne Analytical Instruments
Model 3300ZB
Copyright © 2014 Teledyne Instruments/ Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne Instruments / Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91748.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free from defects of material and of construction, and that our liability shall be limited to replacing or repairing at our factory (without charge, except for transportation), or at customer plant at our option, any material or construction in which defects become apparent within one year from the date of shipment, except in cases where quotations or acknowledgements provide for a shorter period. Components manufactured by others bear the warranty of their manufacturer. This warranty does not cover defects caused by wear, accident, misuse, neglect or repairs other than those performed by TI/AI or an authorized service center. We assume no liability for direct or indirect damages of any kind and the purchaser by the acceptance of the equipment will assume all liability for any damage which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its use r, an d serves as a tool b y whic h valuable data can be gathered. The information provided by the instrument may assist the user in eliminating potential hazards caused by his process; however, it is essential that all personnel involved in the use of the instrument or its interface, with the process being measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process conditions. While this instrument may be able to provide early warning of imminent danger, it has no control over process conditions, and it can be misused. In particular, any alarm or control systems installed must be tested and understood, both as to how they operate and as to how they can be defeated. Any safeguards required such as locks, labels, or redundancy, must be provided by the user or specifically requested of TI/AI at the time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The purchaser is responsible for the training of personnel, for providing hazard warning methods and instrumentation per the appropriate standards, and for ensuring that hazard warning devices and instrumentation are maintained and operated properly.
Teledyne Instruments/ Analytical Instruments, the manufacturer of this instrument, cannot accept responsibility for conditions beyond its knowledge and control. No statement expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the user’s process conditions.
Teledyne Analytical Instruments ii
Trace Oxygen Analyzer
Safety Messages
Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully.
A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol. These symbols are found in the manual and inside the instrument. The definition of these symbols is described below:
GENERAL WARNING/CAUTION: Refer to the instructions for details on the specific danger. These cautions warn of specific procedures which if not followed could cause bodily Injury and/or damage the instrument.
No
Symbol
CAUTION: HOT SURFACE WARNING: This warning is specific to heated components within the instrument. Failure to heed the warning could result in serious burns to skin and underlying tissue.
WARNING: ELECTRICAL SHOCK HAZARD: Dangerous voltages appear within this instrument. This warning is specific to an electrical hazard existing at or nearby the component or procedure under discussion. Failure to heed this warning could result in injury and/or death from electrocution.
Technician Symbol: All operations marked with this symbol are to be performed by qualified maintenance personnel only.
NOTE: Additional information and comments regarding a specific component or procedure are highlighted in the form of a note.
Teledyne Analytical Instruments iii
Model 3300ZB
CAUTION: THE ANALYZER SHOULD ONLY BE USED FOR THE
PURPOSE AND IN THE MANNER DESCRIBED IN THIS MANUAL.
IF YOU USE THE ANALYZER IN A MANNER OTHER THAN THAT FOR WHICH IT WAS INTENDED, UNPREDICTABLE BEHAVIOR COULD RESULT POSSIBLY ACCOMPANIED WITH HAZARDOUS CONSEQUENCES.
This manual provides information designed to guide you through the installation, calibration operation and maintenance of your new analyzer. Please read this manual and keep it available.
Occasionally, some instruments are customized for a particular application or features and/or options added per customer requests. Please check the front of this manual for any additional information in the form of an Addendum which discusses specific information, procedures, cautions and warnings that may be peculiar to your instrument.
Manuals do get lost. Additional manuals can be obtained from TI/AI at the address given in the Appendix. Some of our manuals are available in electronic form via the internet. Please visit our website at:
www.teledyne-ai.com.
Teledyne Analytical Instruments iv
Trace Oxygen Analyzer
Table of Contents
Safety Messages .......................................................................... iii
Table of Contents .......................................................................... v
List of Figures .............................................................................. vii
Introduction ................................................................................... 1
1.1 Overview 1
1.2 Main Features of the Analyzer 1
1.3 Front Panel Description 2
1.4 Rear Panel Description 3
Operational Theory ....................................................................... 5
2.1 Introduction 5
2.2 Zirconium Oxide Sensors 5
2.2.1 Principles of Operation 5
2.2.2 The Effect of Pressure 6
2.3 Electronics 6
2.3.1 General 6
2.3.2 Signal Processing 7
Installation ..................................................................................... 9
3.1 Unpacking the Analyzer 9
3.2 Location and Mounting 10
3.2.1 Control Unit Installation 10
3.3 Electrical Connections 10
3.4 Gas Connections 13
3.5 Installation Checklist 14
Teledyne Analytical Instruments v
Model 3300ZB
Operation ..................................................................................... 15
4.1 Introduction 15
4.2 Using the Function and Data Entry Buttons 16
4.3 Setting the Analysis Ranges 16
4.3.1 HI Range 17
4.3.2 LO Range 17
4.4 Setting the Alarm Setpoints 17
4.4.1 HI Alarm 17
4.4.2 LO Alarm 17
4.4.3 Sensor Fail Alarm 18
4.5 Selecting a Fixed Range or Autoranging 18
4.6 Calibration 18
4.7 LED Display on StartUp 19
Maintenance ................................................................................ 21
5.1 Replacing the Fuse 21
5.2 Sensor Installation or Replacement 22
5.2.1 When to Replace a Sensor 22
Appendix ...................................................................................... 25
A.1 Specifications 25 A.2 Spare Parts List 26 A.3 Drawing List 27
Teledyne Analytical Instruments vi
Trace Oxygen Analyzer
List of Figures
Figure 1-1: Front Panel .................................................................. 2
Figure 1-2: Rear Panel .................................................................... 4
Figure 2-1: Block Diagram of the Signal Processing Electronics .... 7
Figure 3-1: Rear Panel Electrical Connectors ............................... 11
Figure 3-2: Contact ID for FAILSAFE Relay Operation ................. 12
Figure 3-3: Piping Diagram ............................................................ 14
Figure 4-1: Front Panel Controls and Indicators ............................ 15
Figure 5-1: AC Fuse Replacement ................................................ 21
Figure 5-2: ZrO2 Sensor and Cell Block ........................................ 23
Figure 5-3 (A&B): Cell Removal/Installation .................................. 23
Teledyne Analytical Instruments vii
Model 3300ZB
DANGER
COMBUSTIBLE GAS USAGE
This is a general purpose instrument designed for use in a non-hazardous area. It is the customer's responsibility to ensure safety especially when combustible gases are being analyzed since the potential of gas leaks always exist.
The customer should ensure that the principles of operating this equipment are well understood by the user. Misuse of this product in any manner, tampering with its components, or unauthorized substitution of any component may adversely affect the safety of this instrument.
WARNING
Since the use of this instrument is beyond the control of Teledyne Instruments/ Analytical Instruments, referred as TI/AI, no responsibility by TI/AI, its affiliates, and agents for damage or injury from misuse or neglect of this equipment is implied or assumed.
Teledyne Analytical Instruments viii
Trace Oxygen Analyzer
(This page intentionally left blank)
Teledyne Analytical Instruments ix
Trace Oxygen Analyzer Introduction
Introduction
1.1 Overview
The Teledyne Analytical Instruments (TAI) Model 3300ZB is a microprocessor-based trace oxygen analyzer for real-time measurement of trace amounts of oxygen in inert gases, or in a wide variety of gas mixtures. It features simple operation, fast response, and a compact, rugged construction.
1.2 Main Features of the Analyzer
The main features of the analyzer include:
High resolution, accurate readings of oxygen content from 0-10
parts per million (ppm) through 0-1,000 ppm. Large, bright, light-emitting-diode meter readout
Simple pushbutton controls Advanced zirconium oxide measuring technology sensors Fast response and recovery time Microprocessor based electronics: 8-bit CMOS with on-board
RAM and 16 KB ROM
Two user selectable ranges (depending on sensor configuration)
allow best match to user’s process and equipment
Operator can select auto-ranging, allowing the analyzer to
automatically select the proper preset range for a given measurement or lock the analyzer onto a single range
Two concentration alarms with adjustable set-points Three analog outputs: two for measurement (0–10 VDC and
negative ground 4–20 mADC) and one for range identification (0-10 VDC).
Teledyne Analytical Instruments 1
Introduction Model 3300ZB
Compact and rugged control unit, wall-mounting NEMA-4 rated
enclosure.
External probe
1.3 Front Panel Description
All controls and displays except the power switch are accessible from the front panel. See Figure 1-1. The front panel has seven pushbutton membrane switches, a digital meter, and an alarm indicator LED for operating the analyzer. These features are described briefly here and in greater detail in Chapter 4 Operation.
Figure 1-1: Front Panel
Function Keys: Seven pushbutton membrane switches are used to select the function performed by the analyzer:
Set HI Alarm Set the concentration ABOVE which an
alarm activates.
Teledyne Analytical Instruments 2