Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read
the General Safety Summary in other system manuals for warnings and cautions
related to operating the system.
Injury Precautions
Avoid Electric Overload
Ground the Product
Do Not Operate Without
Covers
Use Proper Fuse
Do Not Operate in
Wet/Damp Conditions
Do Not Operate in
Explosive Atmosphere
Avoid Exposed Circuitry
To avoid electric shock or fire hazard, do not apply a voltage to a terminal that is
outside the range specified for that terminal.
This product is indirectly grounded through the grounding conductor of the
power cord. To avoid electric shock, the grounding conductor must be connected
to earth ground. Before making connections to the input or output terminals of
the product, ensure that the product is properly grounded.
To avoid electric shock or fire hazard, do not operate this product with covers or
panels removed.
To avoid fire hazard, use only the fuse type and rating specified for this product.
To avoid electric shock, do not operate this product in wet or damp conditions.
To avoid injury or fire hazard, do not operate this product in an explosive
atmosphere.
To avoid injury, remove jewelry such as rings, watches, and other metallic
objects. Do not touch exposed connections and components when power is
present.
Product Damage Precautions
Provide Proper Ventilation
VX4234
To prevent product overheating, provide proper ventilation.
iii
General Safety Summary
Do Not Operate With
Suspected Failures
If you suspect there is damage to this product, have it inspected by qualified
service personnel.
Safety Terms and Symbols
Terms in This Manual
Terms on the Product
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you read the
marking.
CAUTION indicates a hazard to property including the product.
iv
VX4234
General Safety Summary
Symbols on the Product
The following symbols may appear on the product:
DANGER
High Voltage
Protective Ground
(Earth) T erminal
ATTENTION
Refer to
Manual
Double
Insulated
VX4234
v
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone
Disconnect Power
Use Care When Servicing
With Power On
Do not perform internal service or adjustments of this product unless another
person capable of rendering first aid and resuscitation is present.
To avoid electric shock, disconnect the main power by means of the power cord
or, if provided, the power switch.
Dangerous voltages or currents may exist in this product. Disconnect power,
remove battery (if applicable), and disconnect test leads before removing
protective panels, soldering, or replacing components.
To avoid electric shock, do not touch exposed connections.
This Performance Verification contains test sequences suitable for determining if
the VX4234 is functioning properly and meets the characteristics as warranted.
The following skills are required to perform this procedure:
Thorough knowledge of test instrument operation and proper measurement
techniques
Knowledge of VXIbus system components and command language
programming
Ability and facility to construct interconnections and fixtures as needed to
perform the procedure
General Information and Conventions
The following conventions apply throughout this procedure:
Each test sequence begins with a table, similar to the one below, which
provides information and requirements specific to that section. The item
number appearing after each piece of equipment refers to an entry in Table 1,
Required Test Equipment. Following the table, you will be given instructions
for interconnecting the VX4234 under-test and for checking the performance
parameters. Results may be recorded on a photocopy of the Test Record on
page A–24.
VX4234
Equipment
Requirements
PrerequisitesAll prerequisites listed on page A–22
This procedure assumes that you will be using a VX4521 Slot 0 Resource
Manager and a National Instruments PC-GPIB controller configuration as
described in Table 3. If you are using a different controller and software,
simply substitute the commands for that system to achieve the equivalent
results. In the test sequences, you will be instructed to issue Interface Bus
Interactive Control (ibic) commands to set up the VX4234 under-test and
other associated VXIbus test instruments. Please refer to the NI-488.2M
User Manual for additional information.
Commands to the VX4234 may be entered in upper or lower case. To avoid
confusion between alphanumeric characters, for example between a one (1)
The verification sequences in this procedure are valid under the following
conditions:
The VX4234 module covers are in place and the module is installed in an
The VX4234 has passed the power-on self test
The VX4234 has been operating for a warm-up period of 10 minutes in an
and an “L” or a zero (0) and and the letter “o”, all commands in this
procedure are illustrated in the case which provides the greatest visual
distinction. However, when entering these commands you may use any
combination of upper and/or lower case.
approved VXIbus mainframe as described in Section 2 of the User Manual
ambient environment as specified in Section 1 of the User Manual
Equipment Required
This procedure uses traceable test equipment as specified in Table 1 to directly
check warranted characteristics. You may use instrumentation other than the
recommended example if it meets the minimum requirements.
T able 1: Required Test Equipment
Item Number and DescriptionMinimum RequirementsExamplePurpose
1. Calibrator-GeneratorVariable DC Voltage to ± 700 V DC,
accuracy to 0.005%
Variable DC Current to 2.0 Amps,
accuracy to 0.005%
Variable True RMS to 700 V AC, 50 Hz
to 7500 Hz; accuracy to 0.005%
Variable Resistance to 10 M ,
accuracy to 0.005%
2. Coaxial BNC Cable
(two required)
3. Adapter, BNC female to dual
banana (four required)
4. 1 k , Resistor1 k , Watt, 5%Tektronix part number
50 , 36 inch, male to male BNC
connectors
50 , BNC female to dual bananaTektronix part number
Fluke 5700A CalibratorChecking AC & DC Voltage,
DC Amps, and Resistance
accuracy .
Tektronix part number
012-0482-00
103-0090-00
315-0102-00
Interconnecting electrical
signals
Interconnecting electrical
signals
Checking CMRR
A–22
VX4234
T able 1: Required Test Equipment (Cont.)
Item Number and DescriptionPurposeExampleMinimum Requirements
Appendix F: Performance Verification
5. Adapter, BNC male to Dual
Binding post
6. Adapter, BNC female to
Clip Leads
50 , BNC male to Dual Binding postTektronix part number
103-0035-00
50 , BNC female to dual Clip LeadsTektronix part number
013-0076-00
Interconnecting electrical
signals
Interconnecting electrical
signals
VX4234 Under-Test Configuration
The VX4234 under-test must be installed in an approved VXIbus system. At a
minimum, the system must contain the elements listed in Table 2.
T able 2: Elements of a Minimum VX4234 Under-Test System
Item Number and DescriptionMinimum RequirementsExamplePurpose
1. VXIbus MainframeOne available slot (in addition to
the Slot 0 Resource Manager) for
the VX4234 under-test
Table 3 describes the VXIbus system configuration assumed in this procedure. If
your configuration is different, please note that you will observe your device
names and addresses in test sequences. No secondary addressing is assumed.
The VX4234 includes a built-in self test function (BITE) which runs automatically each time the power is turned on and when the internal self-tests (V0 and
V1) are executed.
BITE uses internal routines and reference circuitry to confirm the instrument’s
capability to measure voltage, current, and resistance, and convert the measurement to a digital result. You may additionally execute the extended AC and DC
mid–scale tests with the V1 command.
NOTE. Because these tests verify operation up to the front panel, you must
disconnect all input signals before executing the extended verification.
In addition to BITE, the front panel indicator lights display the current status of
module power, module self test results (including the central processor), and the
assertion of SYSFAIL*. If the module loses any of its power voltages, the Failed
light will be on, the Power light will be off, and SYSFAIL* will be asserted.
Following a successful VXIbus system startup sequence, the green PWR light on
the VX4234 front panel indicates that the self test has passed and that the +5 V
power supply is operational.
NOTE. If you experience an error indication from the VX4234-under-test, or any
other VXIbus module, investigate and correct the problem before proceeding.
Common items to check are logical address conflicts (primary and secondary;
see Table 3), breaks in the VXIbus daisy chain signals, improper seating of a
module, loose GPIB cable, or loose or blown fuses.
Performance Verification Tests
This Performance Verification procedure may be executed in any order. You may
use any VXIbus system which meets the requirements described in Table 2,
however, the test sequences are structured for a system configuration as
described in Table 3.
NOTE. If at any time you do not observe the expected result, check the front panel
ERROR light. If it is on, perform error Queries (ibwrt ”e”<cr> ibrd 100<cr>)
until you receive a response of “99 . .”. Following a response of “99 . .”, the
front panel ERROR light should be off.
VX4234
A–27
Appendix F: Performance Verification
VXIbus Interface
This sequence verifies that the VX4234 configures correctly and
communicates
properly with your system controller. It assumes a VXIbus system configuration
as specified in Table 3, and in particular, utilizes the VX4521 Resource Manager
“table” command for verification of the system configuration. If your VXIbus
system uses a different Resource Manager, you must substitute commands
specific to that Resource Manager-Controller to verify your system configuration.
Equipment
Requirements
PrerequisitesAll prerequisites listed on Page A–22
No test equipment required.
1. If you are using the VX4521 Slot 0 Resource Manager, send the TABLE
command to verify the system configuration. (If you are using a different
Controller, perform the equivalent function to confirm the responses shown
in Table 5.) Your configuration may not be identical, but the information
should be the same.
2. To verify overall module functionality, send the extended self-test (verifica-
tion) command and an Error query. Then confirm that the front panel
ERROR light is off and that there are no errors (code 99 indicates no errors).
ibfind VX4234
ibwrt "V1E" (Observe that the ERROR light is off)
and the tables and figures are thru numbered Arabic. ibrd 100
(Observe a response of: 99 . .)
A–28
VX4234
Appendix F: Performance Verification
DC Voltage Accuracy
This sequence verifies the accuracy of the six DC voltage ranges.
Equipment
Requirements
PrerequisitesAll prerequisites listed on page A–22
Calibrator (item 1)
Coaxial Cable, one (item 2)
Dual Banana Connectors, two (item 3)
1. Set the Calibrator to +0.0000 mV DC .
2. Connect the Calibrator to the VX4234 Voltage Input using two Dual-Banana
connectors and a Coaxial cable.
3. Set the Calibrator to +18.0 mV DC.
4. With the following commands, set the VX4234 to DC volts and for the
20 mV range (Function 1, Range 1). Then perform an acquisition and verify
the response relative to the limits shown in the Test Record.
set VX4234
ibwrt "F1R1"
ibrd 100 (Observe: +.01800 ±0.0000212 VDC, see Table 6)
5. Set the Calibrator to –18.0 mV DC.
6. Re-acquire and verify the response to the limits shown in the Test Record:
ibrd 100 (Observe: –.01800 ±0.0000212 VDC, see Table 6)
7. Check the additional voltages listed in Table 6 . (Note that the first entry in
Table 6 is a repeat of steps 4 through 6 above .)
Table 6: DC Voltage Accuracy
Set Calibrator
+18.0000 mV DC
–18.0000 mV DC
(Steps 3 to 6 repeated for table
continuity)
+180.000 mV DC
–180.000 mV DC
+1.80000 VDC
–1.80000 VDC
Command To VX4234
ibwrt "F1R1"
ibrd 100
ibrd 100
ibwrt "R2"
ibrd 100
ibrd 100
ibwrt "R3"
ibrd 100
ibrd 100
Response to Verify
+.018021 . . to +.017979 . .
–.017979 . . to –.018021 . .
+.18011 . . to +.17989 . .
–.17989 . . to –.18011 . .
+1.8008 . . to +1.7992 . .
–1.7992 . . to –1.8008 . .
VX4234
A–29
Appendix F: Performance Verification
Table 6: DC Voltage Accuracy (Cont.)
DC Common Mode
Rejection
Set CalibratorResponse to Verify
+18.0000 VDC
Command To
ibwrt "R4"
ibrd 100
–18.0000 VDC
+180.000 VDC
ibrd 100
ibwrt "R5"
ibrd 100
–180.000 VDC
+630.00 VDC
ibrd 100
ibwrt "R6"
ibrd 100
–630.00 VDC
ibrd 100
VX4234
+18.013 . . to + 17.987 . .
–17.987 . . to –18.013 . .
+180.10 . . to +179.90 . .
–179.90 . . to –180.10 . .
+630.3 . . to +629.7 . .
–629.7 . . to –630.3 . .
8. Set the Calibrator to +0.0000 mV.
This sequence verifies the capability of the VX4234 to reject Common Mode
voltage when operating in the DC Voltage measurement mode.
Equipment
Requirements
Calibrator (item 1)
Coaxial Cable, one (item 2)
Dual Banana Connectors, two (item 3)
Binding Post Connector (item 5)
Clip Lead Connector (item 6)
1 k Resistor ( ¼ Watt, 5%)
A–30
PrerequisitesAll prerequisites listed on page A–22
1. Set the Calibrator to +0.0000 mV DC.
2. Connect the Calibrator to the VX4234 as follows:
a. Connect a BNC to Dual Banana connector to the Voltage input of the
VX4234.
b. Connect a BNC to Dual Binding post connector to the BNC to Dual
Banana connector on the VX4234.
c. Connect a 1 k resistor between the binding posts.
d. Connect a BNC to Dual Banana adapter to the calibrator then connect
one end of a coaxial cable to the Dual Banana adapter and the opposite
end of the coaxial cable to a BNC to Clip Lead adapter.
VX4234
Appendix F: Performance Verification
e. Connect the red lead of the Clip Lead adapter to the red Dual Binding
post adapter on the VX4234 (one side of the resistor).
f.Connect the black lead of the Clip Lead adapter to the Chassis Ground
binding post of the VX4234.
3. Set the Calibrator to +100.0 VDC, and enable the output.
4. Using the following commands, set the VX4234 to read DC volts on the
20 mV range (Function 1, Range 1). Perform two acquisitions and verify the
second response to be ≤ 177.8 mV (Note: The first reading returned is stale
data).
set VX4234
ibwrt "F1R1"
ibrd 100 (Disregard this reading.)
ibrd 100 (Observe: ≤ [+/-].000018 . .)
DC Normal Mode
Rejection
5. Set the Calibrator to +100.0 V, 50 or 60 Hz. Perform two acquisitions and
verify the second response to be ≤ 56.8 mV
This sequence verifies the ability of the VX4234 to reject residual AC voltage
present on the input when operating in a DC Voltage mode.
Equipment
Requirements
PrerequisitesAll prerequisites listed on page A–22
Calibrator (item 1)
Coaxial Cable, one (item 2)
Dual Banana Connectors, two (item 3)
1. Set the Calibrator to +0.0000 mV DC.
2. Verify the VX4234-under-test 50/60 Hz reject switch (side panel) is in the
60 Hz position.
VX4234
3. Connect the Calibrator to the VX4234 Voltage Input using two Dual-Banana
connectors and a Coaxial cable.
4. Set the Calibrator to +10.0 V, 60 Hz and enable the output.
A–31
Appendix F: Performance Verification
5. Using the following commands, set the VX4234 to read DC volts on the
6. Power down the VXI system, remove the VX4234-under-test and set the
7. Set the Calibrator to +10.0 V, 50 Hz.
8. Perform two acquisitions and verify the second response to be ≤ 2.24 mV.
20 mV range (Function 1, Range 1). Perform two acquisitions and verify the
second response to be ≤ 2.24 mV (Note: The first reading returned is stale
data).
7. Acquire a reading from the VX4234 and verify the response to the limits
shown in the Test Record:
ibrd 100 (Observe: –.00018000 ±0.00001407 A {table 7})
8. Check the additional currents listed in Table 7 and 8. (Note that the first
entry in Table 7 is a repeat of steps 3 through 7 above.)
T able 7: DC Current Accuracy
Set Calibrator
+180.000 mA
–180.000 mA
Command to VX4234
ibwrt "F6R1"
ibrd 100
ibrd 100
Response to Verify
+.00019407 to +.00016593
–.00016593 to –.00019407
(Steps 3 to 7 repeated for
continuity)
+1.80000 mA
–1.80000 mA
+18.0000 mA
–18.0000 mA
+180.000 mA
–180.000 mA
ibwrt "R2"
ibrd 100
ibrd 100
ibwrt "R3"
ibrd 100
ibrd 100
ibwrt "R4"
ibrd 100
ibrd 100
+.0018407 to +.0017593
–.0017593 to –.0018407
+.018047 . . to +.017953 . .
–.017953 . . to –.018047 . .
+.18011 . . to + .17989 . .
–.17989 . . to –.18011 . .
9. Remove the Calibrator signal from the VX4234 “200 mA max” input and
connect to the “10 A max” input. Continue the DC Current checks as
directed in Table 8. (Be careful to reconnect the Calibrator cable in the
proper polarity. This will require rotating the dual banana connector 180
degrees.)
VX4234
A–33
Appendix F: Performance Verification
T able 8: High Range DC Current Accuracy
AC Voltage Accuracy
Set Calibrator
+1.80000 A
–1.80000 A
+1.90000 A
–1.90000 A
Command to VX4234
ibwrt "R5"
ibrd 100
ibrd 100
ibwrt "R6"
ibrd 100
ibrd 100
Response to Verify
+1.8008 . . to +1.7992 . .
–1.7992 . . to –1.8008 . .
+1.901 . . to +1.899 . .
–1.899 . . to –1.901 . .
10. Set the Calibrator to +0.0000 mV.
This sequence verifies the accuracy of the five AC RMS voltage ranges.
Equipment
Requirements
PrerequisitesAll prerequisites listed on page A–22
Calibrator (item 1)
Dual Banana Connectors, two (item 3)
Coaxial cable, one (item 2)
1. Set the Calibrator to 180.000 mV and 6.0 kHz.
2. Connect the Calibrator to the VX4234 Voltage input using two Dual-Banana
connectors and a Coaxial cable.
3. Set the VX4234 to AC volts and the 200 mV range (Function 5, Range 2).
(Note: There is no Range 1 for AC). Acquire a reading from the VX4234 and
verify the response relative to the limits shown in the Test Record (limits are
also shown in Table 9):
set VX4234
ibwrt "F5R2" (Note: there is no R1 range for AC Volts)
ibrd 100 (Observe: +.18000±.00208, see Table 9)
4. Check the additional voltages as directed in Table 9 . (Note that the first
This sequence verifies the accuracy of the six (6) Resistance ranges.
Equipment
Requirements
Calibrator (item 1)
Coaxial Cable, two (item 2)
Dual Banana Connectors, four (item 3)
PrerequisitesAll prerequisites listed on page A–22
1. Set the Calibrator to 100.
2. Connect the Calibrator Hi/Lo Output to the top input (R+/R-) of the
VX4234 using two Dual-Banana connectors and a Coaxial cable.
3. Connect the Calibrator Hi/Lo Sense to the bottom input (R+/R-) of the
VX4234 using a second Coaxial cable and two Dual-Banana connectors.
4. Set the VX4234 to 4-Wire and the 200 range (Function 4, Range 1).
Acquire a reading from the VX4234 and verify the response relative to the
limits shown in the Test Record (the limits are also shown in Table 9):
set VX4234
ibwrt "F4R1"
ibrd 100 (Observe: +100.00 +0.170W, see Table 10)
VX4234
A–35
Appendix F: Performance Verification
5. Check the additional resistance ranges listed in Table 10. (Note that the first
T able 10: Resistance Accuracy
Set CalibratorCommand to VX4234Response to Verify
+100.000
(Step 4 repeated for continuity)
+1000.00
+10.0000 k
+100.000 k
+1.00000 M
+10.0000 M
entry in Table 10 is a repeat of steps 4 above.)
ibwrt "F4R1"
ibrd 100
ibwrt "R2"
ibrd 100
ibwrt "R3"
ibrd 100
ibwrt "R4"
ibrd 100
ibwrt "R5"
ibrd 100
ibwrt "R6"
ibrd 100
+100.17 . . to +99.83 . .
+1001.1 . . to +998.9 . .
+10011.0 . . to +9989.0 . .
+100110.0 . . to + 99890.0 . .
+1001700.0 to +998300.0 . .
+10020000.0 to +9980000.0
This completes the VX4234 verification procedure.
A–36
VX4234
WARNING
The following servicing instructions are for use only by qualified personnel. To
avoid injury, do not perform any servicing other than that stated in the operating
instructions unless you are qualified to do so. Refer to all Safety Summaries before
performing any service.
37
Appendix G
38
Appendix G
39
Appendix G
40
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.