VX1410A & VX1420A IntelliFrame
VXIbus Mainframes
070-9645-01
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to the Safety Summary prior to performing service.
Copyright Tektronix, Inc. All rights reserved.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
Printed in the U.S.A.
Tektronix, Inc., P.O. Box 1000, Wilsonville, OR 97070±1000
TEKTRONIX, TEK, and IntelliFrame are registered trademarks of Tektronix, Inc.
Super Cooler is a trademark of Tektronix, Inc.
WARRANTY
Tektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for a period of three (3) years from the date of shipment. If a product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
TEKTRONIX SERVICE OPTIONS
Tektronix offers the following service options. These options are modular, flexible, and easy to order with your instrument. Designed to support tracking of calibration to requirements of ISO9000, and to provide for extended repair coverage, these options help fix your long-term maintenance costs and eliminate unplanned expenditures.
Three years of calibration services |
Option C3 |
Provides initial Certification on delivery, plus |
|
|
two more annual calibrations from your ser- |
|
|
vice center |
Five years of calibration services |
Option C5 |
Provides initial Certification on delivery, plus |
|
|
four more annual calibrations from your ser- |
|
|
vice center |
Five years repair coverage |
Option R5 |
Provides Five years of repair coverage for the |
|
|
instrument, including displays and accessories |
|
|
sold with the instrument |
Tektronix Service Options are available at the time you order your instrument.
Contact your local Tektronix Sales Office for more information.
General Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
ix |
Service Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
xi |
Contacting Tektronix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
xiv |
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
xv |
Service Offerings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
xv |
Getting Started |
|
Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
1±1 |
Standard Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
1±4 |
Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
1±5 |
Selecting the Correct Power Cord and Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
1±6 |
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
1±6 |
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
1±20 |
Functional Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
1±25 |
Operating Basics |
|
Enhanced Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
2±2 |
Reference |
|
Passive Monitor Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±1 |
Enhanced Monitor RS-232 Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±4 |
Backplane Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±5 |
Enhanced Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±9 |
Command Groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±13 |
Display Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±13 |
Status Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±13 |
System Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±16 |
Test Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±19 |
VXI Monitor Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±19 |
IEEE 488.2 Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±20 |
Syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±21 |
SCPI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±21 |
Command Syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±21 |
Command Headers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±22 |
Parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±22 |
Message Terminator and Message Separator . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±24 |
Entering Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±24 |
Command Characters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±24 |
Abbreviating Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±24 |
Combining Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±25 |
DISPlay Subsystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±27 |
DISPlay:WINDow:TEXT:DATA? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±28 |
Syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±28 |
Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±28 |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
i |
Table of Contents
Related Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±28 |
DISPlay:TEXT:MARQuee? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±29 |
Syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±29 |
Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±29 |
Related Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±29 |
STATus Subsystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±31 |
SYSTem Subsystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±111 |
TEST Subsystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±149 |
VXIMonitor Subsystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±155 |
IEEE 488.2 Common Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±161 |
Status and Events . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±173 |
Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±183 |
Example 1: Monitor Power Supply Voltages . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±183 |
Example 2: Monitor the Delta Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±183 |
Example 3: Summarize Status Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±184 |
Example 4: Monitor the Power Supply Currents . . . . . . . . . . . . . . . . . . . . . . . . . |
3±184 |
Example 5: VXIMonitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±184 |
Specifications |
|
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
4±1 |
Performance Verification |
|
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
5±1 |
DC Voltage Regulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
5±1 |
DC Load Ripple and Noise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
5±2 |
Theory of Operation |
|
IntelliFrame Mainframe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
6±1 |
Mainframe with Option 1M . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
6±3 |
Adjustment Procedures |
|
Adjustment Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
7±±1 |
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±1 |
Service Strategy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±1 |
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±1 |
Inspection and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±2 |
Interior Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±2 |
Exterior Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±2 |
Cleaning the Fan or Blower Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±3 |
Removal and Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±4 |
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±4 |
Procedure 1: Removing the Mainframe Cover . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±5 |
Procedure 2: Removing the Fan or Blower Assembly . . . . . . . . . . . . . . . . . . . . . |
8±6 |
Procedure 3: Removing the Enhanced Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±10 |
Procedure 4: Removing the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±12 |
Procedure 5: Removing the Card Guides . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±12 |
Procedure 6: Replacing the Top and Bottom Nut Rails . . . . . . . . . . . . . . . . . . . . |
8±15 |
ii |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
Table of Contents
Procedure 7: Removing the Temperature Sense Board . . . . . . . . . . . . . . . . . . . . |
8±16 |
Procedure 8: Removing the Front Panel Display Assembly . . . . . . . . . . . . . . . . |
8±17 |
Procedure 9: Removing the EMI DIN shields . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±18 |
Procedure 10: Removing the Backplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±19 |
Troubleshooting Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±21 |
Diagnostic Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±21 |
Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±21 |
Fault Isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±21 |
Replaceable Parts |
|
Parts Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
9±1 |
IntelliFrameR mainframe Common Replaceable Parts . . . . . . . . . . . . . . . . . . . . |
9±2 |
Using the Replaceable Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
9±3 |
Appendices |
|
Appendix A: Power Budget Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . |
A±1 |
Appendix B: Special Configuration Options . . . . . . . . . . . . . . . . . . . . |
B±1 |
Remote Power Switch Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
B±1 |
Front Panel SYSReset Cable Connector Configuration . . . . . . . . . . . . . . . . . . . |
B±3 |
Enhanced Monitor Board Jumpers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
B±4 |
Appendix C: SCPI Conformance Information . . . . . . . . . . . . . . . . . . . |
C±1 |
Glossary
Index
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
iii |
Table of Contents
Figure 1±1: Power cord identification chart . . . . . . . . . . . . . . . . . . . . . |
1±6 |
Figure 1±2: Install Option 1U cable tray . . . . . . . . . . . . . . . . . . . . . . . . |
1±8 |
Figure 1±3: Install Option 1U cable tray . . . . . . . . . . . . . . . . . . . . . . . . |
1±9 |
Figure 1±4: Install Option 2U cable tray . . . . . . . . . . . . . . . . . . . . . . . . |
1±10 |
Figure 1±5: Install Option 2U cable tray . . . . . . . . . . . . . . . . . . . . . . . . |
1±11 |
Figure 1±6: Installing the rackmount kit hardware . . . . . . . . . . . . . . . |
1±14 |
Figure 1±7: Assembling the slide-out track assemblies . . . . . . . . . . . . |
1±15 |
Figure 1±8: Vertical clearance for rack installation |
|
(standard mainframe) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
1±16 |
Figure 1±9: Installing the slide-out track assemblies |
|
in the rack (Top view) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
1±17 |
Figure 1±10: R1 or R2 rackmounting screw |
|
positions (front view). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
1±18 |
Figure 1±11: Installing the rackmount door (Option 2R) . . . . . . . . . . |
1±19 |
Figure 1±12: Installing the empty slot panel fillers . . . . . . . . . . . . . . . |
1±21 |
Figure 1±13: Locations of safety and chassis ground connections . . . |
1±22 |
Figure 1±14: Installing the EMI intermodule shield and |
|
front panel EMI fillers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
1±23 |
Figure 1±15: Logical address switch Settings . . . . . . . . . . . . . . . . . . . . |
1±24 |
Figure 2±1: Front view of the VX1410A |
|
IntelliFrame Mainframe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
2±1 |
Figure 2±2: Partial front view of a VX1410A with Option 1M . . . . . . |
2±2 |
Figure 2±3: Rear view of a IntelliFrame Mainframe . . . . . . . . . . . . . . |
2±3 |
Figure 2±4: Rear view of the Enhanced Monitor . . . . . . . . . . . . . . . . . |
2±4 |
Figure 3±1: Passive monitor connector . . . . . . . . . . . . . . . . . . . . . . . . . |
3±1 |
Figure 3±2: RS-232 connector pinout . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±4 |
Figure 3±3: Tree hierarchy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±22 |
Figure 3±4: Command header . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±22 |
Figure 3±5: Command header with parameter . . . . . . . . . . . . . . . . . . |
3±22 |
Figure 3±6: DISPlay subsystem hierarchy . . . . . . . . . . . . . . . . . . . . . . |
3±27 |
Figure 3±7: STATus subsystem hierarchy . . . . . . . . . . . . . . . . . . . . . . . |
3±31 |
Figure 3±8: SYSTem subsystem hierarchy . . . . . . . . . . . . . . . . . . . . . . |
3±111 |
Figure 3±9: TEST subsystem hierarchy . . . . . . . . . . . . . . . . . . . . . . . . |
3±149 |
Figure 3±10: VXIMonitor subsystem hierarchy . . . . . . . . . . . . . . . . . . |
3±155 |
iv |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
Table of Contents
Figure 3±11: SCPI and IEEE status and event registers . . . . . . . . . . . |
3±174 |
Figure 3±12: VXI Interrupt Monitor status register . . . . . . . . . . . . . . |
3±175 |
Figure 3±13: VXI Monitor status register . . . . . . . . . . . . . . . . . . . . . . . |
3±175 |
Figure 3±14: VXI TTL Trigger Monitor status register . . . . . . . . . . . |
3±176 |
Figure 3±15: Power status register . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±177 |
Figure 3±16: Current Status register . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±177 |
Figure 3±17: Voltage status register . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±178 |
Figure 3±18: Temperature status register . . . . . . . . . . . . . . . . . . . . . . . |
3±178 |
Figure 3±19: Blower status register . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±179 |
Figure 4±1: Mainframe cooling curve . . . . . . . . . . . . . . . . . . . . . . . . . . |
4±6 |
Figure 4±2: Front view and side view of VX1410A |
|
IntelliFrame Mainframe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
4±11 |
Figure 4±3: Front view and side view of VX1420A |
|
IntelliFrameR Super Cooler Mainframe . . . . . . . . . . . . . . . . . . . . |
4±12 |
Figure 4±4: Front view and side view of IntelliFrame |
|
Mainframe with Option 1U and rackmount option . . . . . . . . . . . |
4±12 |
Figure 4±5: Front view and side view of IntelliFrame |
|
Mainframe with Option 2U and Rackmount Option . . . . . . . . . . |
4±13 |
Figure 6±1: Mainframe block diagram . . . . . . . . . . . . . . . . . . . . . . . . . |
6±2 |
Figure 8±1: Removing the fan/blower filters . . . . . . . . . . . . . . . . . . . . |
8±3 |
Figure 8±2: Removing the mainframe cover . . . . . . . . . . . . . . . . . . . . |
8±5 |
Figure 8±3: Location of fan assembly screws on the |
|
rear of the mainframe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±7 |
Figure 8±4: VX1410A±Removing the fan assembly . . . . . . . . . . . . . . . |
8±8 |
Figure 8±5: VX1410A±Removing the fans . . . . . . . . . . . . . . . . . . . . . . |
8±8 |
Figure 8±6: VX1420A±Removing the blower assembly . . . . . . . . . . . . |
8±9 |
Figure 8±7: VX1420A±Removing the blower . . . . . . . . . . . . . . . . . . . . |
8±9 |
Figure 8±8: Removing the Enhanced Monitor . . . . . . . . . . . . . . . . . . . |
8±10 |
Figure 8±9: Removing the power supply . . . . . . . . . . . . . . . . . . . . . . . . |
8±12 |
Figure 8±10: Removing the top and bottom card guides . . . . . . . . . . . |
8±14 |
Figure 8±11: Replacing the top and bottom nut rails . . . . . . . . . . . . . . |
8±15 |
Figure 8±12: Removing the temperature sense board . . . . . . . . . . . . . |
8±16 |
Figure 8±13: Removing the front panel monitor display . . . . . . . . . . . |
8±17 |
Figure 8±14: Removing the backplane EMI DIN shields . . . . . . . . . . |
8±18 |
Figure 8±15: Removing the backplane . . . . . . . . . . . . . . . . . . . . . . . . . |
8±20 |
Figure 9±1: Cabinet and chassis assembly . . . . . . . . . . . . . . . . . . . . . . |
9±7 |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
v |
Table of Contents
Figure 9±2: Circuit boards and chassis parts . . . . . . . . . . . . . . . . . . . . |
9±9 |
Figure 9±3: Power supply, monitor, and VX1410A |
|
fan assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
9±11 |
Figure 9±4: VX1420A or Option VXF20A blower assembly . . . . . . . . |
9±13 |
Figure 9±5: 1R Rackmount assembly . . . . . . . . . . . . . . . . . . . . . . . . . . |
9±15 |
Figure 9±6: 2R Rackmount and door assembly . . . . . . . . . . . . . . . . . . |
9±17 |
Figure 9±7: Option 1U cable tray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
9±19 |
Figure 9±8: Option 2U cable tray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
9±21 |
Figure 9±9: Display panel, VX1420A or Option 1M . . . . . . . . . . . . . . |
9±23 |
Figure B±1: Power switch connectors . . . . . . . . . . . . . . . . . . . . . . . . . . |
B±2 |
Figure B±2: SYSReset cable connector . . . . . . . . . . . . . . . . . . . . . . . . . |
B±3 |
Figure B±3: Enhanced monitor board jumpers . . . . . . . . . . . . . . . . . |
B±4 |
vi |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
Table of Contents
Table 1±1: IntelliFrame Mainframe options and upgrade kits . . . . . |
1±3 |
Table 1±2: Power cord identification . . . . . . . . . . . . . . . . . . . . . . . . . . |
1±4 |
Table 1±3: Power supply voltages at the 25-pin connector . . . . . . . . . |
1±25 |
Table 3±1: Passive monitor connector pinouts . . . . . . . . . . . . . . . . . . |
3±2 |
Table 3±2: RS-232 pin connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±5 |
Table 3±3: P1 connector pinouts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±6 |
Table 3±4: P2 connector pinouts for slots 1 ± 12 . . . . . . . . . . . . . . . . . |
3±7 |
Table 3±5: P2 connector pinouts for slot 0 . . . . . . . . . . . . . . . . . . . . . . |
3±8 |
Table 3±6: Front panel display messages . . . . . . . . . . . . . . . . . . . . . . . |
3±9 |
Table 3±7: Display commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±13 |
Table 3±8: Status commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±13 |
Table 3±9: System commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±16 |
Table 3±10: Test commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±19 |
Table 3±11: VXI Monitor commands . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±19 |
Table 3±12: IEEE 488.2 common commands . . . . . . . . . . . . . . . . . . . . |
3±20 |
Table 3±13: Parameter types for syntax descriptions . . . . . . . . . . . . . |
3±23 |
Table 3±14: Default factory RS-232 communication settings . . . . . . . |
3±111 |
Table 3±15: Effects of :PRESet on serial port parameters . . . . . . . . . |
3±120 |
Table 3±16: Super Cooler factory default settings . . . . . . . . . . . . . . . |
3±131 |
Table 3±17: NVRecall device settings . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±136 |
Table 3±18: NVSave device settings . . . . . . . . . . . . . . . . . . . . . . . . . . . |
3±137 |
Table 3±19: IntelliFrame Mainframeself test . . . . . . . . . . . . . . . . . . . . |
3±149 |
Table 3±20: IntelliFrame Mainframe error and event codes . . . . . . . |
3±181 |
Table 4±1: AC Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
4±1 |
Table 4±2: Secondary Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
4±2 |
Table 4±3: Cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
4±5 |
Table 4±4: Certifications and compliances . . . . . . . . . . . . . . . . . . . . . . |
4±7 |
Table 4±5: Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
4±9 |
Table 4±6: Backplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
4±10 |
Table 4±7: Display System VX1410A Option 1M |
|
or VX1420A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
4±10 |
Table 4±8: Mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
4±10 |
Table 4±9: VXI and Plug & Play . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
4±13 |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
vii |
Table of Contents
Table 5±1: DC load ripple and noise check . . . . . . . . . . . . . . . . . . . . . |
5±2 |
Table 7±1: PC RS-232 settings for the calibration Procedure . . . . . . |
7±2 |
Table 8±1: IntelliFrame Mainframe Enhanced Monitor |
|
jumper configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±11 |
Table 8±2: Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
8±21 |
Table 9±1: IntelliFrame Mainframe common |
|
replaceable parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
9±2 |
Table C±1: SCPI conformance information . . . . . . . . . . . . . . . . . . . . . |
C±1 |
viii |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
To Avoid Fire or Personal Injury
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read the General Safety Summary in other system manuals for warnings and cautions related to operating the system.
Use Proper Power Cord. Use only the power cord specified for this product and certified for the country of use.
Ground the Product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
The common terminal is at ground potential. Do not connect the common terminal to elevated voltages.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do Not Operate Without Covers. Do not operate this product with covers or panels removed.
Use Proper Fuse. Use only the fuse type and rating specified for this product.
Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Provide Proper Ventilation. Refer to the manual's installation instructions for details on installing the product so it has proper ventilation.
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
ix |
General Safety Summary
Symbols and Terms Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
WARNING |
Protective Ground |
CAUTION |
Double |
High Voltage |
(Earth) Terminal |
Refer to Manual |
Insulated |
x |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present.
Disconnect Power. To avoid electric shock, disconnect the main power by means of the power cord or, if provided, the power switch.
Use Care When Servicing With Power On. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
To avoid electric shock, do not touch exposed connections.
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
xi |
Service Safety Summary
xii |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
The VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual provides information on using the mainframe. The manual includes the following chapters:
HGetting Started provides a brief description of the IntelliFrame Mainframe, a list of accessories that accompany the mainframe, installation and configuration procedures, and a functional check procedure.
HOperating Basics provides a functional overview of the mainframe and includes illustrations of the main components.
HSpecifications lists all of the warranted characteristics, nominal traits, and typical characteristics of the mainframe.
HReference provides reference information on the various connectors on the mainframe. It also provides details on the command set and programming information for the Enhanced Monitor option.
HPerformance Verification includes the performance verification procedures used to verify the advertised specifications of the mainframe.
HTheory of Operation includes a brief overview of the theory of operation for the IntelliFrame Mainframe.
HAdjustment chapter describes adjustment procedures for the IntelliFrame Mainframe.
HMaintenance is intended for qualified service personnel who need to service the mainframe. Module-level removal and replacement procedures and troubleshooting procedures are included in this chapter.
HReplaceable Parts includes a list of replaceable mechanical and electrical parts for the mainframe to be used with the maintenance chapter. It also includes the exploded view diagrams of the mainframe and options.
HAppendices includes a power budget worksheet, a description of special configuration options and SCPI conformance information.
HGlossary.
HIndex.
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
xiii |
Preface
Product |
For application-oriented questions about a Tektronix measure- |
Support |
ment product, call toll free in North America: |
|
1-800-TEK-WIDE (1-800-835-9433 ext. 2400) |
|
6:00 a.m. ± 5:00 p.m. Pacific time |
|
Or contact us by e-mail: |
|
tm_app_supp@tek.com |
|
For product support outside of North America, contact your |
|
local Tektronix distributor or sales office. |
Service |
Contact your local Tektronix distributor or sales office. Or visit |
Support |
our web site for a listing of worldwide service locations. |
|
http://www.tek.com |
For other |
In North America: |
information |
1-800-TEK-WIDE (1-800-835-9433) |
|
An operator will direct your call. |
To write us |
Tektronix, Inc. |
|
P.O. Box 1000 |
|
Wilsonville, OR 97070-1000 |
xiv |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
This manual contains information needed to properly service the IntelliFrame Mainframe This manual does not support component-level fault isolation and replacement.
Tektronix provides service to cover repair under warranty as well as other services that are designed to meet your specific service needs.
Whether providing warranty repair service or any of the other services listed below, Tektronix service technicians are well equipped to service the IntelliFrame Mainframe. Tektronix technicians train on Tektronix products; they have access to the latest information on improvements to the IntelliFrame Mainframe as well as the latest new product upgrades. Services are provided at Tektronix
Services Centers and on-site at your facility, depending on your location.
Warranty Repair Service Tektronix warrants this product for three years from date of purchase. (The warranty appears behind the title page in this manual.) Tektronix technicians provide warranty service at most Tektronix service locations worldwide. The Tektronix product catalog lists all service locations worldwide or you can visit us on our Customer Services World Center web site at http://www.tek.com/Measurement/Service. See our latest service offerings and contact us by email.
Calibration and Repair
Service
In addition to warranty repair, Tektronix Service offers calibration and other services which provide cost-effective solutions to your service needs and qualitystandards compliance requirements. Our instruments are supported worldwide by the leading-edge design, manufacturing, and service resources of Tektronix to provide the best possible service.
The following services can be tailored to fit your requirements for calibration and/or repair of the IntelliFrame Mainframe.
Service Options. Tektronix service options can be selected at the time you purchase your instrument. You select these options to provide the services that best meet your service needs. These service options are listed on the Tektronix Service Options page following the Warranty page of this manual.
Service Agreements. If service options are not added to the instrument purchase, then service agreements are available on an annual basis to provide calibration services or post-warranty repair coverage for the IntelliFrame Mainframe.
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
xv |
Introduction
Service agreements may be customized to meet special turn-around time and/or on-site requirements.
Service on Demand. Tektronix also offers calibration and repair services on a per incident basis that is available with standard prices for many products.
Self Service. Tektronix supports repair to the replaceable-part level by providing for circuit board exchange.
Use this service to reduce down-time for repair by exchanging circuit boards for remanufactured ones. Tektronix ships updated and tested exchange boards. Each board comes with a 90-day service warranty.
For More Information. Contact your local Tektronix service center or sales engineer for more information on any of the Calibration and Repair Services just described.
xvi |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
Getting Started
This chapter provides an overview of the Tektronix IntelliFrame Mainframe family. It includes installation and configuration information on user-installed options.
The IntelliFrame Mainframes are intelligent, 13-slot, C-size, VXIbus Mainframes. They fully comply with the VXIbus Specification 1.4 requirements and are VXIplug&play compatible. The mainframes can be configured as standard benchtop or optional rackmount instruments. The mainframes have the following features:
HHigh output power supply
HIntelligent cooling for installed modules
HSuper CoolerT (VX1420A)
HHigh quality auto-configuration backplane
HEnhanced Monitor
HFront panel display (Optional for VX1410A)
HEMI DIN shields for superior EMI protection
HOptional adjustable rackmounts and cable trays
These features are explained in more detail in the following paragraphs.
Power Supply The IntelliFrame Mainframe provides all VXIbus voltages at levels necessary for the most demanding ATE applications. The power supply has a volt-ampere capacity of 1783 VA with 875 watts usable by VXI instrument modules at any one time (refer to Appendix A: Power Budget Worksheet). The mainframe accommodates exclusively digital applications, exclusively analog applications, or combinations of both.
The power supply plugs directly into the rear of the mainframe and has no cables to disconnect. The power supply can be replaced in less than three minutes, minimizing downtime. The power supply automatically sets the ranges for the appropriate input voltage and frequency.
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
1±1 |
Getting Started
Intelligent Cooling The IntelliFrame Mainframe provides optimal cooling for all installed modules. The mainframe uses an intelligent, adaptive cooling scheme to efficiently remove heat while maintaining quiet operation. The mainframe monitors the temperature rise of the exhaust air above each module slot and automatically adjusts the fan speed to keep the temperature rise to user specified temperature rise or less for the each module. You can also manually set the fan to full speed to provide maximum cooling.
Air is exhausted at the sides of the mainframe, allowing you to stack the mainframe with other rackmount equipment. You don't have to worry about restricting airflow out of the top of the mainframe or creating a chimney effect.
Cooling is not wasted on empty slots. IntelliGuides automatically direct air to only those slots containing modules; empty slots are shut off. The mainframe directs airflow across each installed module from the bottom to the top of each slot. Baffles balance the airflow from the front to back and across occupied slots in the mainframe.
You can remove the air filter from the rear of the mainframe. The fans can be replaced without disassembling the entire mainframe.
High Cooling The VX1420A Super Cooler has all the features of the VX1410A IntelliFrame Mainframe, but with higher cooling capacity and very quiet operation. The higher cooling fan operates at a low fan speed with equivalent airflow.
Backplane The IntelliFrame Mainframe has an auto configuration, solid-state backplane that uses electronic jumpering of the IACK and BUS GRANT signal lines. There are no jumpers to move or install on the backplane. This results in a high quality backplane that is ideal for ATE systems that must be reconfigured frequently. The EMI DIN shields are standard to provide superior EMI protection.
Full differential distribution of the CLK10 signals provides a clean timing source for your VXIbus instruments.
Optional Rackmounts and You can order the IntelliFrame Mainframe with user-installed rackmounts and Cable Trays cable trays. The cable trays provide a means to conveniently route cables from the front of the modules to the rear of the mainframe. Options 1U and 2U cable
tray depths range from 1.75 to 6.0 inches. Install Option 1U to provide a 1.75-inch, 2.5-inch, or 4.25-inch deep cable tray. You can install Option 2U to provide a 3.5-inch, 4.25-inch, or 6.0-inch deep cable tray.
The rackmount adapters let you recess the mainframe from the front panel mounting surface at specified depths from 0.735 inches to 5.735 inches in 0.5-inch increments. Similarly, you can configure the mainframe to extend out of the front of the rack at specified depths (0.57 inches to 5.57 inches in 0.5-inch increments).
1±2 |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
Getting Started
You can install the rackmount kits along with the cable trays to provide vertical cable access to easily route cables to instruments above or below the mainframe while keeping lead lengths as short as possible.
You can also install an optional Lexan door with the rackmount kit.
NOTE. If you choose to extend the mainframe out of the front of the rack, the mainframe cannot accept a door.
Instructions for installing the cable trays begin on page 1±7; instructions for installing the rackmounts kits begin on page 1±12.
Enhanced Monitor The IntelliFrame Mainframe enhanced monitor lets you monitor the power supply voltages, power supply currents, slot temperature rises, fan speeds, and other VXIbus system activity via the VXIbus or a rear RS-232 interface.
Optional Front Panel The VX1410A Option 1M includes a 20-character alphanumeric display on the Display front panel with scroll keys for manual viewing for mainframe status. The
VX1420A comes standard with this option.
Options and Upgrade Kits Table 1±1 lists the IntelliFrame Mainframe options and upgrade kits that you can purchase and install separately. Contact your local Tektronix representative for information on the upgrade kits.
Table 1±1: IntelliFrame Mainframe options and upgrade kits
|
|
Upgrade or |
Description |
Option |
Part number |
|
|
|
Rackmount Kit with Rack Slides |
1R |
020-2221-XX |
|
|
|
Rackmount Kit with Rack Slides and Lexan Door |
2R |
020-2222-XX |
|
|
|
Cable Tray Kit (1U/2U) |
1U |
020-2223-XX |
|
|
|
Cable Tray Kit (1U/2U) |
2U |
020-2224-XX |
|
|
|
VX1410A Front Panel Display Upgrade (Option 1M) |
1M |
VXF1M |
|
|
|
VX1410A Option 1M to VX1420A Upgrade |
± |
VXF20A |
|
|
|
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
1±3 |
Getting Started
The IntelliFrame Mainframe comes with the following standard accessories:
HIntelliFrame Mainframe Instruction Manual
HOne single-slot panel filler
HFive double-slot panel fillers
HTwo power cords (voltage range and plug type varies according to country, see Table 1±2)
HFuse holder
HLine fuse, 15 A, 250 V
HLine fuse, 20 A, 250 V
HLine fuse, 6.3 A, 250 V
HVXIplug&play Floppy disk (See the Readme file on disk for installation information)
Table 1±2: Power cord identification
Plug configuration |
Normal usage |
Option number |
|
|
|
|
North America |
Standard |
|
125 V/15 A Plug |
|
|
NEMA 5-15P |
|
|
|
|
|
North America |
Standard |
|
125 V/20 A Plug |
|
|
NEMA 5-20P |
|
|
|
|
|
Europe |
A1 |
|
230 V |
|
|
|
|
|
United Kingdom |
A2 |
|
230 V |
|
|
|
|
1±4 |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
Getting Started
Table 1±2: Power cord identification (Cont.)
Plug configuration |
Normal usage |
Option number |
|
|
|
|
Australia |
A3 |
|
230 V |
|
|
|
|
|
North America |
A4 |
|
230 V |
|
|
|
|
|
Switzerland |
A5 |
|
230 V |
|
|
|
|
The IntelliFrame Mainframe has the following optional accessories:
HRS-232C Cable, DB9 female X DB9 female, serial null modem (Tektronix part number 012-1379-XX)
HAC to DC low voltage converter supply to externally supply power to the Enhanced Monitor (Tektronix part number 119-5061-XX)
HUS Power cord for the external supply (Tektronix part number 161-0213-00)
HOption A1 Power cord for the external supply (Tektronix part number 161-0209-00)
HOption A2 Power cord for the external supply (Tektronix part number 161-0210-00)
HOption A3 Power cord for the external supply (Tektronix part number 161-0211-00)
HOption A4 Power cord for the external supply (Tektronix part number 161-0208-00)
HOption A5 Power cord for the external supply (Tektronix part number 161-0212-00)
HSingle-wide EMI panel fillers
HDouble-wide EMI panel fillers
HEMI intermodule shields
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
1±5 |
Getting Started
The standard IntelliFrame Mainframe comes with two power cords and three fuses. Before installing the mainframe, you must determine the correct fuse and power cord for your application. Use the following information to determine the appropriate power cord and line fuse for either the VX1410A or the VX1420A mainframes. This information is important to avoid overloading the power distribution system and to comply with the National Electrical Code.
For mainframe loads in the nonshaded region of Figure 1±1, use the power cord with the 15 A plug (Tektronix part number 161-0213-XX) or the power cord with the 20 A plug (Tektronix part number 161-0218-XX). For high mainframe loads combined with low input line voltages (shaded region), use only the power cord with the 20 A plug.
Select the proper fuse based on the ranges shown in Figure 1±1.
Mainframe Power Consumption (Watts)
875 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Power Cord with 20A plug (NEMA 5±20P) ONLY. |
|
|
|
|
|||||||||
800 |
|
|
|
|
|
|
|
|
|
|||||||||
|
|
|
|
|
|
|
|
|
||||||||||
|
|
|
|
|
|
|
|
|
||||||||||
700 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
600 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
500 |
|
|
|
|
|
|
Power Cord with 15A plug (NEMA 5±15P) |
|
|
|
|
|||||||
|
|
|
|
|
|
|
|
|
|
|||||||||
400 |
|
|
|
|
|
|
|
|
|
|
||||||||
|
|
|
|
|
|
or Power Cord with 20A plug (NEMA 5±20P) |
|
|
|
|
||||||||
|
|
|
|
|
|
|
|
|
|
|||||||||
300 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
200 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
100 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
110 |
130 |
150 |
170 |
190 |
210 |
230 |
250 |
|||||||||
90 |
115
20AT Fuse
15AF Fuse
6.3AF Fuse
Figure 1±1: Power cord identification chart
This section describes the procedures to install your mainframe. It includes procedures for installing the mainframe with the rackmount or cable tray options.
This section does not include specific procedures for installing modules in the mainframe. Refer to your module documentation for information regarding the module installation instructions.
1±6 |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
Getting Started
Site Considerations The IntelliFrame Mainframe is designed to operate on a bench or in a rackmount environment. For proper cooling, allow at least 2 in (5.1 cm) of clearance on the rear and sides of the mainframe. Determine how you will use your mainframe with cable trays, and then follow the appropriate installation instructions. If you ordered your mainframe with a cable tray option and rackmount option, install the cable tray before installing the rackmount.
WARNING. To avoid personal injury, use care when lifting or moving the mainframe to avoid personal injury while performing the installation procedures.
Cable Tray Installation The IntelliFrame Mainframe comes a with a 0.75-inch cable access area on the bottom of the mainframe. Use the following procedure to install your cable tray option:
You will need a T-20 Torx and Phillips driver to install Options 1U or 2U.
Options 1U and 2U can be installed three different ways; without rails, with rails, or with rails installed at 90_.
Figure 1±2 on page 1±8 and Figure 1±3 on page 1±9 show how to install Option 1U (without rails, with rails, rails at 90_) to provide a cable tray depth of 1.75, 2.50, or 4.25 inches.
Figure 1±4 on page 1±10 and Figure 1±5 on page 1±11 show how to install Option 2U (without rails, with rails, rails at 90_) to provide a cable tray depth of 3.50, 4.25, or 6.00 inches.
Both options come with butch plates to mount cable connectors at the rear of the cable trays. The butch plate will not fit on Option 1U when the foot rails are not used. You can modify the butch plate to accept cable connectors at the rear of the mainframe. Specific instructions for modifying the butch plate are not given since the locations, types, and size of connectors vary with each application.
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |
1±7 |
Getting Started
Remove feet (4)
Opt 1U without rails 1.75±inch depth
Install feet. (optional)
Install 4 screws, and leave 1/8 inch clearance from the chassis. Slide the cable tray over the screws; then tighten screws to secure cable tray.
Figure 1±2: Install Option 1U cable tray
1±8 |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |