Tektronix VM Series User Manual

User Manual
x
VM Series Video Measurement System
071-2103-07
*P071210307*
071210307
User Manual
VM Series Video Measurement System
071-2103-07
This document applies to Option VGA software version 3.X and above.
www.tektronix.com
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its suppliers and are protected by United States copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserve d.
TEKTRONIX and T EK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
H In North America, call 1-800-833-9200. H Worldwide, visit www.tektronix.com to find contacts in your area.
Warranty 2
Tektronix warrants that this product will be free from defects in materials and workma nship for a period of one (1) year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance. All replaced parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to a ny other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or i nadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a ) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompa tible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Table of Contents
Getting Started
Preface xiii...................................................
About This Manual xiii...............................................
Related Manuals and Online Documents xiv..............................
General Safety Summary xv...................................
Environmental Considerations xvii...............................
Product Description 1--1........................................
Key Features 1--1....................................................
Compatibility of the VM Series Option HD with the VM5000HD 1--3..........
Firmware Upgrade 1--3...............................................
Recommended Oscilloscope Probes 1--3.................................
Installation 1--5...............................................
Unpacking 1--5......................................................
Checking the Environment Requirements 1--6.............................
Powering On the Instrument 1--7.......................................
Shutting Down the Instrument 1--7.....................................
Connecting to a Network 1--8..........................................
Adjusting Display Contrast 1--9........................................
Getting Acquainted with your Instrument 1--10.............................
Backing Up User Files 1--12............................................
Installing Software 1--13...............................................
Incoming Inspection 1--15.......................................
Self Tests 1--15.......................................................
Functional Test 1--16..................................................
Options and Accessories 1--23....................................
Options 1--23........................................................
Accessories 1-- 24.....................................................
Operating Basics
Operating Basics 2--1..........................................
Input Signal Connection Requirements 2--1...............................
Connecting Input Signals 2--3..........................................
Signal Path Compensation 2--17.........................................
Sync Pickoff Compensation -- VM6000 and DPO7000 Series,
Option SD/HD only 2--18...........................................
The Matrix Test Signal 2--21............................................
Starting the Application 2--25...........................................
The VM Series Application User Interface 2--27............................
Measurement Steps Overview 2--28......................................
Setting the Input Signal Format 2--29.....................................
Selecting the Measurements 2--33........................................
Setting the Measurement Run Operations 2--34.............................
Setting the Measurement Warnings 2--39..................................
Setting Reference and Limits Options 2--41................................
VM Series Video Measurement System User Manual
i
Table of Contents
Reference
Displaying the Input Signal as a Picture 2--43..............................
Changing Picture Display Settings 2--44...................................
Displaying the Input Signal in a Vectorscope Window for Option SD/HD 2--46....
Taking and Viewing the Measurement Results 2--48.........................
Taking Measurements Across Multiple Lines 2--51..........................
Changing Line Number or Averaging on the Results Screen 2--52..............
Saving Measurement Setups 2--54........................................
Recalling Measurement Settings 2--56....................................
Saving Measurement Results (Reports) 2--58...............................
Reference Overview 3--1........................................
Menu Descriptions 3--3.........................................
File Menu 3--3......................................................
Configuration Menu 3--5..............................................
Results Menu 3--19...................................................
Utilities Menu 3--32...................................................
Help Menu 3--33.....................................................
Option SD/HD Detailed Measurement Information 3--35.............
Matrix Signals and Line Numbers 3--35...................................
Averaging 3- -35......................................................
Channel Delay 3--36..................................................
Color Bars 3--37......................................................
Frequency Response 3--38..............................................
H Sync 3--41........................................................
H Sync Jitter 3--42....................................................
Levels 3--43.........................................................
Multiburst 3--44......................................................
Noise 3--45..........................................................
Non Linearity 3--46...................................................
Short Time Distortion 3--46.............................................
Spatial Distortion 3--48................................................
V Sync 3--54........................................................
Option VGA Detailed Measurement Information 3--57...............
Matrix Signals and Line Numbers 3--57...................................
Averaging 3- -57......................................................
Ch-Ch Mismatch Measurement 3--58.....................................
Ch-Ch Skew Measurement 3--59.........................................
Color Bars Measurement 3--60..........................................
H Sync Measurement 3--61.............................................
H Sync Jitter Measurement 3--64........................................
H Timing Measurement 3-- 65...........................................
Linearity Measurement 3--66............................................
Luma Levels Mea surement 3--68........................................
Noise Inj Ratio Measurement 3--69.......................................
V Sync Measurement 3--71.............................................
V Timing Measurement 3-- 74...........................................
Video Transient Measurement 3--75......................................
ii
VM Series Video Measurement System User Manual
Table of Contents
Setting Up Measurements 3--77...................................
Taking Measurements 3--77.............................................
Taking Measurements Relative to a Reference 3--78.........................
Creating a Custom Reference File 3--80...................................
Taking Measurements with Limit Testing 3--82.............................
Customizing a Limits File 3--84.........................................
Setting Up the Channel Delay Measurement 3--85...........................
Setting Up the Color Bars Measurement 3--86..............................
Setting Up the Frequency Response Measurement 3--88......................
Setting Up the H Sync Measurement 3--91.................................
Setting Up the H Sync Jitter Measurement 3--93............................
Setting Up the Levels Measurement 3--96.................................
Setting Up the Multiburst Measurement 3--98..............................
Setting Up the Noise Measurement 3--99..................................
Setting Up the Noise Spectrum Measurement 3--102..........................
Setting Up the Non Linearity Measurement 3--104...........................
Setting Up the Short Time Distortion Measurement 3--105.....................
Setting Up the Spatial Distortion Measurement 3--106........................
Setting Up the V Sync Measurement 3--108.................................
Configuring the Device Under Test for Option VGA 3--108....................
Setting Up the Ch-- Ch Mismatch Measurement 3--111.........................
Setting Up the Ch-- Ch Skew Measurement 3--112............................
Setting Up the Color Bars Measurement 3--114..............................
Setting Up the H Sync Measurement 3--115.................................
Setting Up the H Sync Jitter Measurement 3--116............................
Setting Up the H Timing Measurement 3--117...............................
Setting Up the Linearity Measurement 3--118...............................
Setting Up the Luma Levels Measurement 3--119............................
Setting Up the Noise Inj Ratio Measurement 3--120..........................
Setting Up the V Sync Measurement 3--121.................................
Setting Up the V Timing Measurement 3--122...............................
Setting Up the Video Transient Measurement 3--123..........................
Viewing Measurement Results 3--125...............................
Displaying the Results Summary 3--125....................................
Displaying Channel Delay Measurement Results 3--127.......................
Displaying Color Bars Measurement Results 3 --128..........................
Displaying Frequency Response Measurement Results 3--130...................
Displaying H Sync Measurement Results 3--131.............................
Displaying H Sync Jitter Measurement Results 3--133.........................
Displaying Levels Measurem ent Results 3--134..............................
Displaying Multiburst Measurement Results 3--135...........................
Displaying Noise Measurement Results 3--136...............................
Displaying Non Linearity Measurement Results 3--138........................
Displaying Short Time Distortion Measurement Results 3--140.................
Displaying Spatial Distortion Measurement Results 3--141.....................
Displaying V Sync Measurement Results 3--142.............................
Displaying Ch-Ch Mismatch Measurement Results 3--146.....................
Displaying Ch-Ch Skew Measurement Results 3--147.........................
Displaying Color Bars Measurement Results 3 --148..........................
Displaying H Sync Measurement Results 3--150.............................
Displaying H Sync Jitter Measurement Results 3--151.........................
Displaying H Timing Measurement Results 3--153...........................
VM Series Video Measurement System User Manual
iii
Table of Contents
Appendices
Displaying Linearity Measurement Results 3--154............................
Displaying Luma Levels Measurement Results 3--155.........................
Displaying Noise Inj Ratio Measurement Results 3--157.......................
Displaying V Sync Measurement Results 3--158.............................
Displaying V Timing Measurement Results 3--159...........................
Displaying Video Transient Measurement Results 3--161......................
Displaying Pass/Fail Results 3--163.......................................
Appendix A: Specifications A--1..................................
Option SD/HD A--2..................................................
Option VGA A--11....................................................
EMC Compliance Information A--17......................................
Appendix B: Performance Verification B--1........................
Option SD/HD Performance Verific ation B--1..............................
Prerequisites B--1....................................................
Required Equipment B--2..............................................
Option SD/HD Procedure B--2..........................................
RGBHV Measurement Interface Unit (012--1685--XX)
Performance Verification B--5.......................................
Prerequisites B--5....................................................
Required Equipment B--6..............................................
Test A: Initial Connection and Setup B--8.................................
Test B: RGB LF Mode DC Termination Accuracy B--9......................
Test C: RGB LF Mode and HV LF Mode DC Gain B--10.....................
Test D: HV Input LF Mode DC Termination Accuracy and
WB Mode DC Input Impedance B--14.................................
Test E: HV Input LF Mode Current Sources B--15...........................
Test F: RGB WB Mode Bandwidth B--16..................................
Test G: H/V Inputs WB Mode Bandwidth B--20.............................
Appendix C: Cleaning C --1......................................
Exterior Cleaning C--1................................................
Flat Panel Display Cleaning C--1........................................
Appendix D: Rebuilding the Hard Drive D--1......................
Appendix E: Signal Sources Package Contents E--1.................
Signal Source Discs for Option SD/HD E--1...............................
Baseband Test Signals CD E--2.........................................
ATSC Transport Stream CD E--4........................................
Advanced TV Elementary Streams CD
Standard Definition TV Elementary Streams CD E-- 5....................
480 and 576 Line DVDs E--7...........................................
H.264 SD & HD Streams CDs E--8......................................
Signal Source Disc for Option VGA E--10.................................
Index
iv
VM Series Video Measurement System User Manual
List of Figures
Table of Contents
Figure 1 --1: Instrument space requirements 1--6....................
Figure 1 --2: On/Standby switch location 1--7.......................
Figure 1--3: Powering off the instrument 1--8.......................
Figure 1--4: Connecting to a network 1--8..........................
Figure 1 --5: Front panel of the instrument 1 --10.....................
Figure 1 --6: Side and rear panels of the instrument 1--11..............
Figure 1--7: Displaying measurement values in
color bars measurement 1--18.................................
Figure 1--8: Format panel 1--20...................................
Figure 1 --9: Results Summary display 1--20.........................
Figure 1--10: Color Bars measurement results display 1--21...........
Figure 1--11: H Sync measurement results display 1--21...............
Figure 1--12: V Sync measurement results display 1--21...............
Figure 1 --13: Video Transient measurement results display 1--22.......
Figure 2--1: Connecting the T-connector to the
VM Series System 2--1......................................
Figure 2--2: RGBHV Measurement Interface Unit 2--2...............
Figure 2--3: VGA connector to 5X cable 2--2.......................
Figure 2--4: Connecting to a VM6000 / DPO7000
Series instrument 2--4.......................................
Figure 2--5: Connecting to a VM5000 / TDS5000
Series instrument 2--4.......................................
Figure 2--6: Connecting to a VM6000 / DPO7000
Series instrument 2--5.......................................
Figure 2--7: Connecting to a VM5000 / TDS5000
Series instrument 2--6.......................................
Figure 2--8: Connecting the 012--1680--00
Sync Pickoff to CH1 input 2--6...............................
Figure 2--9: Connecting to a VM6000 / DPO7000
Series instrument 2--7.......................................
Figure 2--10: Connecting to a VM5000 / TDS5000
Series instrument 2--7.......................................
Figure 2--11: Connecting to a VM6000 / DPO7000
Series instrument 2--8.......................................
Figure 2--12: Connecting to a VM5000 / TDS5000
Series instrument 2--8.......................................
VM Series Video Measurement System User Manual
v
Table of Contents
Figure 2--13: Connecting to a VM6000 / DPO7000
Series instrument 2--9.......................................
Figure 2--14: Connecting to a VM5000 / TDS5000
Series instrument 2--10.......................................
Figure 2--15: Connecting the DUT to the RGBHV
Measurement Interface Unit 2--11.............................
Figure 2--16: Connecting the RGBHV MIU to a
VM6000 / DPO7000 Series instrument 2--11.....................
Figure 2--17: Connecting the RGBHV MIU to a
VM5000 / TDS5000 Series instrument 2--12.....................
Figure 2--18: Connecting the RS-232C cable between the
MIU and the VM Series System 2--12...........................
Figure 2--19: Selecting the Use MIU option on the
Configuration > Operations tab 2--13..........................
Figure 2--20: Connecting the VGA to 5X cable to the
input channel 2--14..........................................
Figure 2--21: Building the 2.21 kiloohms and 301 ohms
terminations 2--15..........................................
Figure 2 --22: Connecting the PC video output signals to the
VM6000 / DPO7000 Series instrument 2--16.....................
Figure 2--23: Connecting the PC video ou tput signals
to the VM5000 / TDS5000 Series instrument 2--16................
Figure 2--24: Appearance of the pop-up warning
when making H Sync measurement 2--17.......................
Figure 2--25: Connecting the Sync Pickoff cable to
CH1 and CH4 2--18.........................................
Figure 2--26: Attaching the Sync Pickoff to a
BNC-to-minigrabbers adapter and connecting to
PROBE COMP 2--18........................................
Figure 2--27: Selecting AUTOSET 2--19............................
Figure 2--28: Properly compensated 2--19...........................
Figure 2--29: Under compensated 2--19.............................
Figure 2--30: Over compensated 2--20..............................
Figure 2--31: Adjusting the compensation 2--20......................
Figure 2--32: Option SD/HD Matrix signal 2--21.....................
Figure 2--33: Option VGA Matrix test signal 2--23...................
Figure 3--1: Signal processing overview 3--43........................
Figure 3--2: Block diagram for measuring video
spatial distortion both in horizontal and
vertical dimensions 3--49.....................................
vi
VM Series Video Measurement System User Manual
Table of Contents
Figure 3 --3: Block diagram for one dimensional refined
alignment measurement 3--50.................................
Figure 3--4: Signal definitions (H Sync) 3--61........................
Figure 3 --5: Jitter in H Sync period 3--64...........................
Figure 3 --6: Frequency response with 500 MHz filter for
Noise In Ratio measurement 3--69.............................
Figure 3--7: Voltage ripple and noise 3--70..........................
Figure 3--8: Signal definitions (V Sync) 3--71........................
Figure 3--9: Signal definitions (Video transient) 3--75.................
Figure 3--10: Results of the Color Bars measurement 3--78............
Figure 3--11: Enabling relative to reference results 3--79..............
Figure 3--12: Relative to reference results displayed
(Option SD/HD) 3--80........................................
Figure 3--13: Relative to reference results displayed
(Option VGA) 3--80.........................................
Figure 3--14: Editing a refer ence file 3--81..........................
Figure 3--15: Enabling limit testing 3--83...........................
Figure 3--16: Results when limit testing is available 3--83..............
Figure 3--17: Customizing a limits file 3--84.........................
Figure 3--18: Changing the filter type 3--100.........................
Figure 3--19: Changing the noise filter bandwidth frequency 3--101......
Figure 3--20: Changing the cursor frequency 3--103...................
Figure 3--21: Changing the channel 3-- 103...........................
Figure 3--22: Illustration of the BMP test signal in
full screen mode 3--110........................................
Figure 3--23: Illustration of the BMP image set as wallp aper 3--110......
Figure 3--24: Displaying the results summary
(Option SD/HD) 3--125........................................
Figure 3--25: Displaying the results summary (Option VGA) 3--125......
Figure 3--26: Displaying the Channel Delay measurement 3--127........
Figure 3--27: Displaying limit values for the
Channel Delay measurement 3--128.............................
Figure 3--28: Displaying the Color Bars measurement 3--128...........
Figure 3--29: Displaying limit values for the
Color Bars measurement 3--129................................
Figure 3 --30: Displaying the Frequency Response measurement 3--130...
Figure 3--31: Displaying limit values for the
Frequency Response measurement 3 --131........................
Figure 3--32: Displaying additional units for the
Frequency Response measurement 3 --131........................
VM Series Video Measurement System User Manual
vii
Table of Contents
Figure 3--33: Displaying the H S ync measurement 3--132...............
Figure 3--34: Displaying limit values for the
H Sync measurement 3--132...................................
Figure 3--35: Displaying the H S ync Jitter measurement 3--133.........
Figure 3--36: Displaying limit values for the
H Sync Jitter measurement 3--134..............................
Figure 3--37: Displaying the Levels measurement 3--134...............
Figure 3--38: Displaying limit values for the
Levels measurement 3--135....................................
Figure 3--39: Displaying the Multiburst measurement 3--135...........
Figure 3--40: Displaying limit values for the
Multiburst measurement 3--136................................
Figure 3--41: Displaying the Noise measurement 3--137................
Figure 3--42: Displaying limit values for the
Noise measurement 3--137.....................................
Figure 3--43: Displaying the Noise Sp ectrum measurement 3--138.......
Figure 3--44: Displaying the Noise Sp ectrum on the
VM Series video monitor 3--138................................
Figure 3--45: Displaying the Non Linearity measurement 3--139.........
Figure 3--46: Displaying limit values for the
Non Linearity measurement 3--139..............................
Figure 3--47: Displaying the Short Time Distortion
measurement 3--140.............................................
Figure 3--48: Displaying limit values for the
Short Time Distortion measurement 3--141.......................
Figure 3--49: Displaying the Spatial Distortion measurement 3--141......
Figure 3--50: Displaying limit values for the
Spatial Distortion measurement 3--142..........................
Figure 3--51: Displaying the V S ync measurement for
SD Interlaced format 3--142...................................
Figure 3--52: Displaying the V S ync measurement for
SD Progressive format 3--143..................................
Figure 3--53: Displaying the V S ync measurement for
SD Interlaced format 3--143...................................
Figure 3--54: Displaying the V S ync measurement for
HD Progressive format 3--143..................................
Figure 3--55: Displaying limit values for the
V Sync measurement for SD Interlaced format 3--144..............
Figure 3--56: Displaying limit values for the
V Sync measurement for SD Progressive format 3--144.............
viii
VM Series Video Measurement System User Manual
Table of Contents
Figure 3--57: Displaying limit values for the
V Sync measurement for HD Interlaced format 3--145.............
Figure 3--58: Displaying limit values for the
V Sync measurement for HD Progressive format 3--145............
Figure 3--59: Ch-Ch Mismatch measurement results 3--146.............
Figure 3--60: Limit values for the Ch-Ch Mismatch
measurement 3--147..........................................
Figure 3--61: Ch-Ch Skew measurement results 3--147.................
Figure 3--62: Limit values for the Ch-Ch Skew measurement 3--148.....
Figure 3--63: Color Bars measurement results 3--148..................
Figure 3--64: Limit values for the Color Bars measurement 3--149.......
Figure 3--65: H Sync measurement results 3--150.....................
Figure 3--66: Limit values for the H Sync measurement 3--151..........
Figure 3--67: H Sync Jitter measurement results 3--151................
Figure 3--68: Limit values for the H Sync Jitter measurement 3--152.....
Figure 3--69: H Timing measurement results 3--153...................
Figure 3--70: Limit values for the H Timing measurement 3--154........
Figure 3--71: Linearity measurement results 3--154....................
Figure 3--72: Limit values for the Linearity measurement 3--155........
Figure 3--73: Luma Levels measurement results 3--155................
Figure 3--74: Limit values for the Luma Levels measurement 3--156.....
Figure 3--75: Noise Inj Ratio measurement results with MIU 3--157......
Figure 3--76: Limit values for the Noise Inj Ratio measurement 3--158...
Figure 3--77: V Sync measurement results 3-- 158.....................
Figure 3--78: Limit values for the V Sync measurement 3--159..........
Figure 3--79: V Timing measurement results 3--159...................
Figure 3--80: Limit values for the V Timing measurement 3--160........
Figure 3 --81: Video Transient measurement results 3--161..............
Figure 3--82: Limit values for the Video Transient
measurement 3--162..........................................
Figure 3--83: Displaying Pass/Fail results in
results summary for Option SD/HD 3 --163.......................
Figure 3--84: Displaying Pass/Fail results in
results summary for Option VGA 3--163.........................
Figure 3--85: Displaying Pass/Fail results (Option SD/HD) 3--164........
Figure 3--86: Displaying Pass/Fail results (Option VGA) 3--164.........
VM Series Video Measurement System User Manual
ix
Table of Contents
Figure A--1: Graph of VM6000 Opt VGA Noise Floor A--18...........
Figure B--1: Noise floor accuracy test hookup B--3..................
Figure E--1: Using TG700 baseband test signals E--4................
Figure E--2: Using ATSC transport stream signals E--5..............
Figure E--3: Using elementary stream signals E--6...................
Figure E--4: Using DVD signals E--7..............................
x
VM Series Video Measurement System User Manual
List of Tables
Table of Contents
Table 1--1: Recommended Probes 1--3............................
Table 1--2: Instrument options 1--23...............................
T able 1--3: Standard Accessories 1--24.............................
T able 1--4: Recommended Accessories 1--26........................
T able 1--5: Optional Accessories 1--27.............................
Table 2--1: Required connections of the VGA signal,
VM Series System, and the terminator 2--17....................
Table 2--2: Option SD/HD Matrix signal elements 2--22..............
Table 2--3: Elements of Option VGA Matrix Signal 2--23.............
Table 2--4: Resolutions and line number ranges
supported at 60 Hz (DMT standard) for Option VGA 2--24........
T able 3--1: Overview of menu descriptions 3--1....................
Table 3--2: Scanning systems for SD in accordance
with the EIA--770.2-A standard 3--54...........................
Table 3--3: Scanning systems for HD in accordance
with the EIA-770.3-A standard High Definition
TV analog Component Video Interface 3--55....................
T able 3--4: Elements of the reference file 3--81......................
Table 3--5: Matching measurements to the appropriate
Full Field test pattern 3--109...................................
Table 3--6: Status Box Indicators in Results Summary pane 3 --126......
Table 3--7: Warning Box Indicators in Results Summary pane 3--126....
VM Series Video Measurement System User Manual
xi
Table of Contents
T able A--1: Video measurement specifications A--2..................
Table A--2: Sync Pick--Off Accessory (012--1680--01)
Specifications A--9.........................................
T able A--3: Option VGA Video Measurement Specifications A--1 1......
T able A--4: RGBHV Measurement Interface Unit
(MIU) Specifications A--16...................................
Table B--1: Required test equipment for Option SD/HD
performance verification B--2................................
Table B--2: The Noise (dB) Measured values when filter
bandwith is 30 MHz B--4....................................
Table B--3: The Noise (dB) Measured values when filter
bandwith is 60 MHz B--5....................................
Table B--4: Equipment required for performance verification B--6....
Table B--5: RGB WB mode bandwidth worksheet B--17..............
Table B--8: Sample Table 4 worksheet B--20........................
Table E--1: Signal Sources discs appropriate for
Option SD/HD E--1........................................
Table E--2: Test Signal Files -- File and Signal Formats E--2..........
Table E--3: TG700 DNL analog signal format files E--2..............
Table E--4: TG700 DNL digital signal format files E--3..............
Table E--5: Available MXT100 .trp signal setup files E--4............
T able E--6: Advanced TV Elementary Streams E--5.................
Table E--7: Advanced TV Bitmaps E--5...........................
Table E--8: Standard Definition TV Elementary Streams E--6........
Table E--9: Standard Definition TV Bitmaps E--6...................
Table E--10: Available formats of DVD files E--7....................
T able E--11: H.264 Elementary Streams E--8.......................
Table E--12: H.264 Transport Streams E--9........................
xii
VM Series Video Measurement System User Manual
Preface
About This Manual
This is the user manual for the VM Series Video Measurement System. This manual focuses on the automatic video measurement functionality; it does not cover the oscilloscope functionality. Refer to Related Manuals and Online Documents to locate information about using the oscilloscope functionality.
This manual is composed of the following chapters:
H Getting Started shows you how to configure and install your instrument and
provides a functional check procedure.
H Operating Basics, using a highly-graphical approach, walks you through the
basics of using the instrument user interface.
H Reference goes into more detail on the menu functions, understanding the
measurements, and interpreting your results.
H Appendices provides additional information including instrument specifica-
tions, performance verification, cleaning information, and signal source descriptions.
VM Series Video Measurement System User Manual
xiii
Preface
Related Manuals and Online Documents
This manual is part of a document set of manuals and online documentation. See the following list for other documents supporting operation and service:
Manual name Description
VM Series Quick Start User Manual (printed, and also available on the software CD as a PDF file)
The VM Series Quick Start User Manual provides information about installation and basic operating concepts to enable you to start using VM Series as quickly as possible.
VM Series Video Measurement System Online Help
VM Series Programmer Manual (only in PDF format)
DPO7000 Series Quick Start User Manual The Quick Start User Manual provides informati on about instal lation and basic
DPO7000 Series Digital Phosphor Oscilloscopes Service Manual
An integrated online help system, which can be accessed from the VM Series help menu (located in the middle of the screen when the VM Series application is open).
Documentation for GPIB commands, wi th syntax and description, and commands that are specific to the VM Series.
operating concepts to enable you t o easily use the basic oscilloscope functionality.
A description of how to service the instrument to the module level. This optional manual must be ordered separately.
xiv
VM Series Video Measurement System User Manual
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read the General Safety Summary in other system manuals for warnings and cautions related to operating the system.
ToAvoidFireor
Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and certified for the country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Do Not Operate Without Covers. Do not operate this product with metal covers or panels removed.
Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Provide Proper Ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
VM Series Video Measurement System User Manual
xv
General Safety Summary
Symbols and Terms
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
CAUTION
Refer to Manual
WARNING
High Voltage
Protective Ground
(Earth) Terminal
xvi
VM Series Video Measurement System User Manual
Environmental Considerations
This section provides information about the environmental impact of the product.
Product End-of-Life
Handling
Observe the following guidelines when recycling an instrument or component:
Equipment Recycling. Production of this equipment required the extraction and use of natural resources. The equipment may contain substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. In order to avoid release of such substances into the environment and to reduce the use of natural resources, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
This symbol indicates that this product complies with the applicable European Union requirements according to Directives 2002/96/EC and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries. For information about recycling options, check the Support/Service section of the Tektronix Web site (www.tektronix.com).
Mercury Notification. This product uses an LCD backlight lamp that contains mercury. Disposal may be regulated due to environmental considerations. Please contact your local authorities or, within the United States, refer to the E--cycling Central Web page (www.eiae.org) for disposal or recycling information.
Perchlorate Materials. This product contains one or more type CR lithium batteries. According to the state of California, CR lithium batteries are classified as perchlorate materials and require special handling. See www.dtsc.ca.gov/ha­zardouswaste/perchlorate for additional information.
Restriction of Hazardous
Substances
VM Series Video Measurement System User Manual
This product has been classified as Monitoring and Control equipment, and is outside the scope of the 2002/95/EC RoHS Directive.
xvii
Environmental Considerations
xviii
VM Series Video Measurement System User Manual
Getting Started
Product Description
This manual supports the VM Series Video Measurement System.
This chapter summarizes and highlights some of the key features of the instrument. Following this description are three sections:
H Installation shows you how to configure and install the instrument, as well
as how to reinstall the system software included with the product.
H Incoming Inspection procedure verifies basic operation and functionality.
H Accessories lists the standard and optional accessories for this product.
Key Features
The VM Series Video Measurement System is a high-performance, oscilloscope­based video measurement system. Key features include:
H Fast, accurate, and repeatable video measurements
H Comprehensive component analog signal analysis
H Support for SD, HDTV, and R GBHV component analog video formats using
options
H Automated testing of PC and DTV consumer video devices
H Automated VESA compliance testing for PC graphics devices
H Picture, Vector, and Waveform Displays
H Integrated Pass-Fail limit testing
H Companion Test Signal Package
H GPIB Remote Control and LAN Connectivity
H Complete 1 GHz bandwidth, 4 Channel DPO Functionality
H Time Saving Test Utilities and Automatic Report Generator
H Technology-specific Software Options for Jitter and Timing Measurements,
Power Measurements, Serial Data, Ethernet, and USB2.0 Compliance Testing
H Video Measurement Accessories and Pinpoint Triggering
VM Series Video Measurement System User Manual
1- 1
Product Description
H Large 12.1 inch XGA Touch Screen Display and CD-R/W drive (DVD Read
only)
Targeted to meet the needs of video professionals developing, testing, and manufacturing the next generation of Digital Television reception and play--out devices, PC and play-out devices, the instrument incorporates the power, features, and functionality needed in R&D, QA, and Production Test applica­tions. It enables manufacturers to ensure that DTV video signal (Option SD/HD), and PC video signal (Option VGA) quality is up to the challenge of high performance displays, and to clearly differentiate between input signal and display device impairments.
You can use the VM Series Video Measurement System for the following purposes:
H R&D
H Compliance and certification testing
H Quality/Audit testing
H Installation and Troubleshooting
H Automated production test
H Off-air video systems testing
The VM Series System automates a variety of component analog video measurements which verify the integrity and quality of DTV video signals (Option SD/HD) and PC video signals (Option VGA).
The VM Series System automatically assesses conformance of selected video signal parameters to applicable EIA--770--3 (for HD), SMPTE --274M, and 296M standards (Option SD/HD), or the VESA version 1 revision 2 standard (Option VGA).
The VM Series System also automates measurement of other industry-standard video parameters that quantify the performance of PCs or other consumer video reception and play-out devices with component analog interfaces.
By integrating automated measurement algorithms and a high-speed, wide bandwidth signal acquisition platform into one instrument, Tektronix offers a reliable means to make fast, objective, and accurate video measurements. Product verification activities that previously took hours or days to complete can now be completed in seconds or minutes. Offering near plug-and-play video measurement capability, even unskilled operators can reliably assess PC video output signal quality.
1- 2
As a fully integrated video analyzer, the VM Series System offers configuration menu settings that can readily be recalled or copied, eliminating complicated oscilloscope setups, tedious manual measurements, and time-consuming results
VM Series Video Measurement System User Manual
correlation exercises. Product quality is enhanced because test results can be reliably generated, easily replicated with other VM Series System instruments, and readily communicated across a global engineering, supply, or sales organization.
The DTV formats from 480i through 1080i and either RGB or YP bPr color space are supported in Options SD and HD. Option VGA supports common analog RGBHV signal resolutions from 640x480p though 2048x1536p, and refresh rates from 50 Hz through 100 Hz. The highest refresh rate for option VGA is 120 Hz and the highest resolution is 2048X2048.
The VM Series System runs the Windows operating system, which provides for easy operation with minimal training time. A 12.1 in (307 mm) color display provides a bright, clear, and crisp display of waveforms and measurement results.
Compatibility of the VM Series Option HD with the VM5000HD
Product Description
If you are upgrading an existing VM5000HD, some default settings have changed. The changes are as follows:
H The default trigger channel has changed from CH1 to CH4.
H The Operation menu ‘Run Mode’ default setting has changed from ‘Continu-
ously’ to ‘Once.’
Firmware Upgrade
Tektronix may offer firmware upgrade kits for the VM Series System instrument. Contact your Tektronix service representative for more information (refer to Contacting Tektronix at the beginning of the manual).
Recommended Oscilloscope Probes
Table 1--1 lists the recommended oscilloscope probes to use when using the instrument as an oscilloscope.
Table 1- 1: Recommended Probes
Probe type Tektronix probe
Passive, 10X, 500 MHz P5050
Passive, 100X, 250 MHz, high voltage P5100
Passive, 1000K, 100 MHz, high voltage P6015A
Passive, 15 MHz, 1X P6101B
VM Series Video Measurement System User Manual
1- 3
Product Description
Table 1- 1: Recommended Probes (Cont.)
Probe type Tektronix probe
Active, DC to 1.0 GHz, FET hi gh speed digital , for CMOS/TTL technology
Active, DC to 1.5 GHz, FET hi gh speed digital , for all technologies
Active, DC to 400 MHz, FET, high bandwidth differential P6246
Active, DC to 1.0 GHz, FET, high bandwidth differential P6247
Active, DC to 1.7 GHz, FET, high bandwidth differential P6248
Active, 100 MHz, high voltage differential P5205
Active, 50 MHz, high voltage differential P5210
AC Current P6021
AC Current P6022
DC/AC Current AM5030S
High Speed Current CT--1/CT--2
Current, 250 KHz to 2 GHz, 50 input, AC Only, for temporary in-circuit installat ion
Current, DC to 50 MHz, 15 A DC TCP202
Low Capacitance, 20X, 3 GHz P6158
Current Measurement TCPA3000
Current Measurement TCPA4000
Differential Preamp ADA400A
High Voltage Differential P5200
Optical/Electrical Converter P6701B
Optical/Electrical Converter P6703B
P6243
P6245
CT-6
1- 4
VM Series Video Measurement System User Manual
Installation
Unpacking
This section addresses instrument installation:
H Unpacking on page 1--5
H Checking the Environment Requirements on page 1--6
H Connecting the Mouse on page 1--10
H Powering On the Instrument on page 1--7
H Shutting Down the Instrument on page 1--7
H Backing Up User Files on page 1--12
H Installing Software on page 1--13
H ConnectingtoaNetworkon page 1--8
H Adjusting Display Contrast on page 1--9
Refer to the tables beginning on page 1--23 for a list of the options and accesso­ries shipped with this instrument. Remember to fill out and send in the customer registration card.
VM Series Video Measurement System User Manual
1- 5
Installation
Checking the Environment Requirements
Read this section before starting any installation procedures. This section describes site considerations, power requirements, and ground connections for your instrument.
Site Considerations
The instrument is designed to operate on a bench or on a cart in the normal position (on the bottom feet). For proper cooling, make sure that there is at least three inches (76.20 mm) of clearance on the left side of the instrument. See Figure 1--1. If you need to access the instrument CD-RW drive, you will also need at least 5.1 inches (129.54 mm) of clearance on the right side of the instrument.
Operating Requirements
1- 6
Figure 1- 1: Instrument space requirements
CAUTION. To ensure proper cooling, keep the bottom and left side of the instrument clear of obstructions.
The specifications in Appendix A list the power source, temperature, humidity, and altitude operating requirements for the instrument.
VM Series Video Measurement System User Manual
Powering On the Instrument
Follow these steps to power on the instrument for the first time:
CAUTION. To ensure proper operation, connect the keyboard, mouse, and other accessories before applying power to the product the first time.
1. Connect one end of the instrument power cord to the instrument and the other end to a power source.
2. Push the On/Standby switch to power on the instrument (see Figure 1--2 for the switch location).
Installation
Figure 1- 2: On/Standby switch location
Shutting Down the Instrument
To shut down the instrument, push the front-panel On/Standby switch. The instrument preserves settings and then starts a shutdown process (including a Windows shutdown) to put the instrument into standby mode. Avoid disconnect­ing the power cord or power source to shut down the instrument.
NOTE. If you remove power instead of using the On/Standby switch to shut down the instrument, the instrument will be in the factory Default Setup when powered on the next time.
If the shutdown process does not start immediately after pushing the ON/Standby switch, push and hold the On/Standby switch for up to 20 seconds to start the shutdown process.
To completely remove power to the instrument, push the On/Standby switch and then disconnect the power cord from the power source.
VM Series Video Measurement System User Manual
1- 7
Installation
Figure 1- 3: Powering off the instrument
Connecting to a Network
You can connect your instrument to a network for printing, file sharing, Internet access, and other functions. Consult with your network administrator and use the standard Windows utilities to configure the instrument for your network.
1- 8
Figure 1- 4: Connecting to a network
NOTE. If you want to disable network access for the instrument, perform the following procedure, except substitute Disabled for the command listed in step 8. The instrument will boot faster with network access disabled.
1. Begin with the instrument power off.
2. Attach a keyboard and mouse to the instrument.
3. Power on the instrument.
VM Series Video Measurement System User Manual
Installation
4. As the instrument begins to boot, press the keyboard F2 key repeatedly until the “Loading SETUP” message appears.
5. In the BIOS Setup Utility, use the keyboard right-arrow key to highlight the Advanced menu at the top of the screen.
6. Use the arrow down key to highlight Peripheral Configuration in the Advanced screen, and then press Enter.
7. Use the arrow down key to highlight LAN Device in the Peripheral Configuration screen, and then press Enter.
8. Use the arrow up or down key to highlight Enabled, and then press Enter.
9. Press the F10 key to save and quit. Confirm Save of the configuration
changes when you are prompted on screen.
10. Use the Windows network setup utility to define the instrument as a network client, and configure it for your network. You can find the network setup utility in the Windows Start menu if you select Settings > Control Panel and then double click Network. You should consult your network adminis­trator for specific instructions to make these settings.
Adjusting Display Contrast
If you find that the instrument screen is too bright or too dark, use these steps to adjust the LCD display contrast to your preference. You must have a mouse attached to perform this procedure.
1. Right click the Windows desktop.
2. Select Properties.
3. Select the Settings tab.
4. Click Advanced.
5. Select the Color tab.
6. Adjust the slider control for the desired gamma setting. The screen
brightness will change as you move the slider control.
7. Click Apply.
8. Click OK.
VM Series Video Measurement System User Manual
1- 9
Installation
Getting Acquainted with your Instrument
Front Panel
The front panel points of the instrument are shown in the following figure :
Figure 1- 5: Front panel of the instrument
1. DVD/CD-RW drive
2. Front panel controls
3. USB port
4. Ground terminal
5. Auxillary Trigger Input
6. Channel 1--4 inputs
7. Probe calibration output
8. Probe compensation output
9. Recovered clock output
10. Recovered data output
1- 10
VM Series Video Measurement System User Manual
Side and Rear Panels
Installation
Figure 1- 6: Side and rear panels of the instrument
VM Series Video Measurement System User Manual
1- 11
Installation
1. USB ports
2. Video port to connect a monitor for side-by-side display
3. Mic connector for microphone
4. Line Out connector for speaker
5. Scope Only XGA Out video port to connect a monitor
6. Printer connection
7. Line In connector
8. RJ--45 LAN connector to connect to a network
9. Centronics Parallel Port
10. COM1 serial port
11. PS--2 connector for mouse
Backing Up User Files
12. PS--2 connector for keyboard
13. Tek Link connector for future use
14. Removable hard disk drive
15. GPIB port to connect to controller
16. Auxiliary output
17. Channel 3 output
18. External reference input
Always back up your user files on a regular basis. Use the Windows Backup tool to back up files stored on the hard disk. The Backup tool is located in the System Tools folder in the Accessories folder.
Use these steps to back up your files:
1. Quit the instrument application by selecting the File menu Exit command.
2. Click the Windows Start button.
1- 12
3. Select Programs > Accessories > System Tools > Backup.
4. Use the Backup W izard to select the items you want to back up and the
destination. The destination can be the floppy drive or a third-party storage device over one of the side-panel ports.
VM Series Video Measurement System User Manual
Installing Software
Installation
The instrument system and application software is preinstalled at the factory. If you have to reinstall the software for any reason, refer to the instructions that accompany the CD-ROMs that are shipped with the instrument.
Software Release Notes
VM Series Remote
Command Information
Desktop Applications
Read the software release notes (README.TXT) on the product software CD-ROM before performing installation procedures. This file contains additional information that supercedes other product documentation.
To view README.TXT, start the Windows Notepad accessory and open the file from the VM Series Video Measurement System Product Software CD-ROM.
The GPIB programmer information is available in the VM Series Video Measurement System Programmer Manual PDF located on the VM Series product software CD--ROM.
You can install desktop application software on the instrument. The instrument has been tested with the following software products installed:
H Microsoft Office 98 (Word, Excel, PowerPoint)
H MathCad (Option SD/HD only)
H MATLAB (Option SD/HD only)
Other software products may be compatible but have not been tested by Tektronix. If the instrument malfunctions after you install software, uninstall the software and then reinstall the instrument software (if necessary) to restore proper operation.
Quitting the instrument Application. Before installing other desktop applications, quit the instrument application. To do this, select the File > Exit command.
VM Series Video Measurement System User Manual
1- 13
Installation
1- 14
VM Series Video Measurement System User Manual
Incoming Inspection
This section contains instructions for performing a self test and functional verification of the instrument. These procedures verify that the instrument is operating correctly after shipment, but they do not check product specifications.
If the instrument fails any test within this section, it may need service. To contact Tektronix for service, see Contacting Tektronix on the copyright page.
Make sure you have installed the instrument as described in Installation starting on page 1--5.
Self Tests
This procedure uses internal routines to verify that the instrument functions and was adjusted properly. No test equipment or hookups are required.
Equipment required None
Prerequisites Power on the instrument and allow a 20 minute warm-up.
Perform these substeps to verify internal diagnostics:
1. Display the system diagnostics menu as follows:
a. If the instrument is in the toolbar mode, click Menu (far right toolbar
button) to switch to menu bar mode.
b. From the Utilities menu, select Instrument Diagnostics to open the
Instrument Diagnostics control window.
c. In the Subsystem, Area,andTes t lists, ensure that you select ALL.
2. Run the system diagnostics as follows:
a. Disconnect all input signals from all four channels.
b. Click Run in the Instrument Diagnostics control window.
The internal diagnostics do an exhaustive verification of proper instrument function, taking five to fifteen minutes. When complete, the resulting status appears in the diagnostics control window.
3. Verify that no failures are found and reported: The instrument should pass all tests. If any failures occur, record the failure information and contact your local Tektronix service personnel for more information.
VM Series Video Measurement System User Manual
1- 15
Incoming Inspection
Functional Test
4. Run the signal path compensation routine as follows:
a. From the Utilities menu, select Instrument Calibration to open the
Instrument Calibration control window.
b. Click Calibrate to start the routine.
c. Confirm signal path compensation returns passed status: Signal path
compensation may take up to ten minutes to run. Verify that the word Pass appears in the Status text box.
5. Click Close to quit the Instrument Calibration control window.
Option SD/HD
This procedure uses the output from a video signal generator, TG700 or equivalent, to verify the VM Series System HD and SD video application functionality. Two functional tests are described, one for option HD and another for option SD. If you have both options, either test alone will verify functionality adequately.
Equipment required VM6000
75Ω precision BNC terminations (3)
BNC “T” connectors (3)
A TG700 video generator with either a AWVG7 (for HD) or AVG7 (for SD) (or equivalent generator)
NOTE. These procedures verify that the application features operate. They do not verify that they operate within limits. Therefore, when the instructions in these functional tests call for you to verify that a signal appears on screen “that is about five divisions in amplitude” or “has a period of about six horizontal divisions,” and so on, do not interpret the quantities given as limits.
Do not change the front-panel settings unless directed to by the procedure. If you change front-panel settings, you may obtain invalid results. In this case, redo the procedure from step 1.
1- 16
When you are instructed to press a front-panel button or screen button, the button may already be selected (its label is lighted). If this is the case, it is not necessary to press the button.
VM Series Video Measurement System User Manual
Incoming Inspection
1. Set the video generator to output 1080i/60 100% (for option HD) color bars, or 480i/60 100% color bars (for option SD).
2. Connect the signal generator to the VM Series System instrument as shown in the following diagram:
VM6000
BNC “T”
TG700 with
AWVG7 module
CH3 CH2 CH1
CH2CH1
CH3
connector
CH4
Y/G
Pr/RPb/B
Sync Pickoff
3. Power on the VM Series System instrument and wait for it to complete the power-on process.
4. Select File > Run Application > VM HD and SD Video to display the VM Series System application window.
5. In the application window, select File > Recall Defaulttoloadthe application default settings.
6. In the application window, select Configuration > Format to display the Format view .
7. Select the 1080i/60 (HD) or 480i/60 (SD) button.
8. In the application window, select Configuration > Operation to display the
Operation view.
VM Series Video Measurement System User Manual
1- 17
Incoming Inspection
9. In the application window, select the Run button to start the application measurement process.
The instrument begins taking measurements and displays a Color bars measurement view listing measurement values similar to those as shown (To update).
Option VGA
Figure 1- 7: Displaying measurement values in color bars measurement
10. Select the Picture button to display the VM Series Video Monitor Window. Verify that a color bars signal appears.
This completes the functional test. If your VM Series did not complete the functional test satisfactorily, please contact your local Tektronix Service representative.
This procedure uses the output of a PC Display Card to verify the functionality of the VM Series System Option VGA measurement set (comprising the VM Series VGA, RGBHV Measurement Interface Unit (MIU) and associated cables).
Equipment required VM6000 Option VGA
RGBHV Measurement Interface Unit
VGA to 5 BNC cable, part number 174--5126--XX
PC with VGA Display Card
1- 18
VM Series Video Measurement System User Manual
Incoming Inspection
NOTE. These procedures verify that the application features operate. They do not verify that they operate within limits. Therefore, when the instructions in these functional tests call for you to verify that a signal appears on screen “that is about five divisions in amplitude” or “has a period of about six horizontal divisions”, and so on, do not interpret the quantities given as limits.
Do not change the front-panel settings unless directed to by the procedure. If you change front-panel settings, you may obtain invalid results. In this case, redo the procedure from step 1.
When you are instructed to press a front-panel button or screen button, the button may already be selected (its label is lighted). If this is the case, it is not necessary to press the button.
1. Connect the RGBHV MIU to the VM S eries System as described in
Option VGA Setup 1: RGBHV Measurement Interface Unit on page 2--9.
2. Set the VGA display card under test to any built-in format (such as 800x600@60) supported by Option VGA. This can be done in the PC by using the Display Properties menu. S elect the corresponding Matrix BMP file (Resolution 800x600-Matrix Signal.bmp) from the PC Bitmap Graphics CD (part number 020--2634--XX) and display the BMP file in full screen mode.
3. Remove the PC monitor connection from the VGA display card and connect the VGA display card output to the input channels on the MIU using the VGAto5xBNCadaptercable.
4. Ensure that all four channels of the VM Series System are enabled. All four channel buttons must be glowing.
5. Press the Autoset button on the VM Series System. You should see the signals from all four channels on the display with some visible activity.
6. Launch the VM Series System application from the TekScope menu bar by selecting File > Run Application > VM VGA Video.
7. In the Format Panel (Configuration > Format), select the 800 x 600 resolution and 60 Hz refresh rate. Leave the Timing standard as the default (that is, DMT). If the VGA display card generates signals using a different Timing Standard, change the Timing selection accordingly. Refer to Figure 1--8.
VM Series Video Measurement System User Manual
1- 19
Incoming Inspection
Figure 1- 8: Format panel
8. In the Measurements panel (Configuration > Measurements), select the following measurements:
H Color Bars
H H Sync
H V Sync
H Video Transient
Leave the settings in other configuration panels at their default values.
9. Press the Run button.
10. If the VM Series System is functioning normally, the four selected
measurements will be executed without any warning message(s). The results should be displayed on the corresponding result panels (for example, Results > H Sync). Because multiple measurements are selected, the Results Summary panel will appear. Some representative screen shots are shown in the Figures 1--9 through 1--13.
1- 20
Figure 1- 9: Results Summary display
VM Series Video Measurement System User Manual
Incoming Inspection
11. To view results for individual measurements, go to their respective screens. For example, to see the results of the color bars measurement, select Results > Color Bars.
The results shown in Figure 1--10 through Figure 1--13 are typical results. Your results may vary.
Figure 1- 10: Color Bars measurement results display
Figure 1- 11: H Sync measurement results display
Figure 1- 12: V Sync measurement results display
VM Series Video Measurement System User Manual
1- 21
Incoming Inspection
Figure 1- 13: Video Transient measurement results display
Possible Errors and Troubleshooting Tips. If there is any problem associated with
the MIU or its connections, the following error message appears:
Error occurred in configuring the MIU
Perform the following steps to isolate the problem:
1. To further localize the problem, check the results display. For example, if there is no signal on the G channel, the corresponding results will not be displayed. So, check the G channel cables and their connection at the input side of the MIU. Between the MIU and VM Series System check the Ch 1 cable and its connections.
2. If expected results still do not appear, there might be some internal problem in the MIU.
3. Other warning messages can be due to an incorrect signal or no signal. If the warning is due to no signal, carry out the checks specified in step 1 above. If the warning is due to an incorrect signal, ensure that the VGA card is set to the same resolution, refresh rate, and timing standard as selected in the Format panel of the application and that the appropriate BMP file is used.
This completes the VM Series functional test for Option VGA. If your VM Series System did not complete the functional test satisfactorily, contact your local Tektronix Service representative.
1- 22
VM Series Video Measurement System User Manual
Options and Accessories
This section lists the product options, standard accessories, and optional accessories that are available for the instrument.
Options
You can order the options shown in Table 1--2 with the instrument. For the current list of accessories, upgrades, and options, including service options, available for your VM6000, visit the Tektronix Web site, www.tektronix.com.
Table 1- 2: Instrument options
Hardware 18
Mounting 016--1946--XX TDS5000B Series Rackmount Kit -- 1R
Software SM Series Communications mask testing
1
1R Rackmount kit
FHD Front panel removable hard drive
1K Scope Cart
655--4920--00 Thermal Printer -- 1P
2
CP2
3
HD
3
SD
3
VGA
DVD TDSDVD -- Optical storage analysis software
ET3 TDSET3 -- Ethernet compliance t est software
J2 TDSDDM2 -- Hard disk drive measurements
J3E TDSJIT3E -- Jitter and timing analysis software
JA3 TDSJIT3 V2.0 Advanced-- Jitter and timing analysis
JT3 TDSJIT3 V2.0 Essentials-- Jitter and timing analysis
PW3 TDSPWR3 -- Power measurements software
Touch-screen interface
TDSCPM2 -- ANSI/ITU telecom pulse compliance testing software (This requires option SM)
HD component analog video measurements & format support, includes test signal package
SD component analog video measurements & format support, excludes test signal package
RGBHV Video Measurements and VESA Compliance Tests
software
essentials
software
software
VM Series Video Measurement System User Manual
1- 23
Options and Accessories
Table 1- 2: Instrument options (Cont.)
USB TDSUSBS; USB 2.0 Compliance test software only
Service C3 Calibration Service 3 years
C5 Calibration Service 5 years
D1 Calibration Data Report
D3 Calibration Data Report 3 years (with Option C3)
D5 Calibration Data Report 5 years (with Option C5)
R3 Repair service 3 years (including warranty)
R5 Repair service 5 years (including warranty)
1
No upgrade path for this option, must be ordered at the time of initial purchase.
2
Requires option SM.
3
At least one of the Option HD, SD, or VGA is mandatory for each VM6000 Instrument.
Upgrades equivalent to original options can be ordered to extend instrument performance after initial purchase. Users can install upgrades without opening the instrument case or requiring on--site service. To upgrade a VM5000, order a VM5UP with one or more of the options listed above. Factory installation is available by specifying Option IF. For older VM5000HD instrument upgrades, order VM5HDUP with one or more of the following options: 1R, DVD, ET3, J2, J3E, JT 3, PW3, USB.
Accessories
This section lists the standard and optional accessories. The accessories listed in Table 1--3 are shipped standard with the instrument. Please specify the power plug and the manual language option when ordering.
Table 1- 3: Standard Accessories
Accessory Part number
VM Series Video Measurement System Quick Start User Manual in English (and a French, German, Spanish, Traditional Chinese, Simplified Chinese, or Japanese translation if a language option was ordered)
VM Series Video Measurement System User Manual 071--2103--XX
VM Series Video Measurement System Programmer Manual
in PDF format on VM Series Video Measurement System Product Software CD
DPO7000, DSA70000, and DPO70000 Digital Phosphor Oscilloscopes Quick Start User Manual
071--2091--XX
071--2104--XX
071--1733--XX
1- 24
VM Series Video Measurement System User Manual
Options and Accessories
Table 1- 3: Standard Accessories (cont.)
Accessory Part number
VM Series Video Measurement System Product Software CD-ROM (includes VM Series Video Measurement System product software with online help, VM Series Video Measurement System User Manual PDF, VM Series Video Measurement System Quick Start User Manual PDF, and release notes)
DPO7000, DSA70000, and DPO70000 Product Software CD-ROM 063--2693--XX
DPO7000, DSA70000, and DPO70000 Operating System Restore CD 063--2659--XX
Optional Applications Software CD-ROM and Documentation Kit 020--2700--XX
Tek Restore Software License for Recovery Medi a/OS Restore -- -- --
Getting Started with OpenChoiceTMSolutions Kit 020--2513--XX
020--2767--XX
DPO7000, DSA70000, and DPO70000 Online help (part of the Product Software)
DPO7000, DSA70000, and DPO70000 Performance Verification (a PDF file on the Product Software CD)
DPO7000, DSA70000, and DPO70000 Programmer Online Guide (files on the Product Software CD)
NIST, Z540--1, and ISO9001 Calibration Certificate -- -- --
Sync pick-off cable (Option SD/HD only) 012--1680--00
Option VGA
VGA to 5X BNC cable (3 foot long)
VGA to 5X BNC cable (6 inch l ong)
RGBHV Measurement Interface Unit RS-232 RGBHV MIU Control Cable
BNC-to-Binding Post Adapters (2)
Power Cord
North America (Option A0) 161--0104--00
Universal Euro (Option A1) 161--0104--06
United Kingdom (Option A2) 161--0104--07
Australia (Option A3) 161--0104--05
-- -- --
-- -- --
-- -- --
174--5126--XX
174--5147--XX
012--1685--XX 012--1692--00
011--0183--XX
240V North America (Option A4) 161--0104--08
Switzerland (Option A5) 161--0167--00
Japan (Option A6) 161--0005--00
China (Option A10) 161--0306--00
India (Option A11) 161--0324--00
No power cord or AC Adapter (Option A99) -- -- --
Mouse, optical 119--7054--XX
VM Series Video Measurement System User Manual
1- 25
Options and Accessories
Table 1- 3: Standard Accessories (cont.)
Accessory Part number
Mini-Keyboard, USB with 2-port hub 118--7083--XX
Front cover 200--4963--XX
Accessory pouch 016--1966--XX
Probe Calibration and Deskew Fi xture, with instructions 067--0405--XX
Nero OEM Software CD 063--3781--XX
TPA-BNC Service Replacement Adapter Option SD & HD -- 1 Option VGA -- 4
P5050 10X, 500 MHz, Passive Voltage Probe
Blank CDR / CDRW 020--2434--XX
Click ’N Burn Patch floppy disk
Precision 75 Ω terminations (4) 011--0102--XX
2.21 KΩ and 301 Ω terminator kit 020--2633--XX
BNC female-to-female connector 103--0028--XX
BNC male-to-male connector 103--0029--XX
BNC “T” connectors (4) 103--0030--XX
2.21 kresistor 322--3226--00
301 resistor 322--3143--00
VGA Sync Combiner cable 012--1664--XX
013--0355--02
P5050
063--XXXX--XX
Table 1- 4: Recommended Accessories
Accessories Part Number
VM Series Signal Sources, 480 line DVD and 580 line DVD 020--2770--XX
1- 26
VM Series Signal Sources, standard definition elementary streams
VM Series Signal Sources, advanced TV elementary streams 020--2772--XX
VM Series Signal Sources, ATSC transport streams 020--2773--XX
VM Series Signal Sources, baseband test signal s 020--2774--XX
VM Series Signal Sources, PC bitmap graphics test signal files 020--2775--XX
VM Series Signal Sources, H.264 SD & HD streams 020--2776--XX
DPO70000 Series Digital Phosphor Oscilloscopes Service Manual
Transit case 016--1937--XX
Probe calibration, compensat ion, and deskew adapter 067--0405--XX
020--2771--XX
071--1174--XX
VM Series Video Measurement System User Manual
Options and Accessories
Table 1- 4: (Cont.)Recommended Accessories
Accessories Part Number
Power deskew fixture 067--1478--00
Video display clamp 013--0278--XX
P5050 500 MHz 10x passive probe P5050
GPIB cable (3.3 ft. or 1m ) 012--0991--01
GPIB cable (6.6 ft. or 2m ) 012--0991--00
Centronics (printer) cable 012--1214--00
The accessories in Table 1--5 were orderable for use with the instrument at the time this manual was published. Consult a current Tektronix catalog or the Tektronix Web site for additions, changes, and details.
Table 1- 5: Optional Accessories
Accessories Part Number
Transit case 016--1937--00
Video display clamp 013--0278--XX
Thermal printer paper 016--1897--00
Probe calibration, compensat ion and deskew adapter requires a separate pulse generator
Power deskew fixture 067--1478--00
Extra front removable hard drive 065--0692--XX
P6101B 15 MHz 1X Passive Probe P6101B
Extra P5050 500 MHz 10X Passive Probe P5050
TCPA300, TCPA400 Series Current Measurement Systems (requires TCPA300 / TCPA400 style probe)
AMT75 1 GHz 75 adapter AMT75
AFTDS Differential signal adapter AFTDS
K420 Scope Cart with mouse accessory tray 436--0317--XX
TDS5000 Series, VM5000 Option HD rackmount kit 016--1887--XX
GPIB cable (3.3 ft / 1 m) 012--0991--01
GPIB cable (6.6 ft / 2m) 012--0991--00
067--0405--XX
TCPA300 / TCPA400
Centronics cable 012--1250--00
VM Series Video Measurement System User Manual
1- 27
Options and Accessories
1- 28
VM Series Video Measurement System User Manual
Operating Basics
Operating Basics
This section provides basic information on how to use the VM Series System. More detailed measurement information can be found in the Reference section.
Input Signal Connection Requirements
Option SD/HD
Use only 75 Ω coaxial cables to connect the video device under test to the instrument.
Install the included 75
Ω terminations between the end of the coaxial cables and
the instrument input channels, using the included BNC “T” connectors. See Figure 2--1. For accurate frequency response measurements, it is important to attach the “T” connector directly to the input channel BNCs.
Direct connection to VM6000
input channel BNC
75 Ω coaxial
cable
BNC “T” connector
(103-0030-00)
75 Ω precision
end-line termination
(011-0102-03)
Figure 2- 1: Connecting the T-connector to the VM Series System
Option VGA
The preferred method of connecting the video device under test to the instrument is to use the RGBHV Measurement Interface Unit. The Measurement Interface Unit contains an impedance-matching circuit to ensure the most accurate measurement results. For some measurements, a low-frequency path is used and for other measurements a wideband path is used.
VM Series Video Measurement System User Manual
2- 1
Operating Basics
The VM Series System selects which path is used through an RS -232 connection to the Measurement Interface Unit. See Figure 2--2.
Figure 2- 2: RGBHV Measurement Interface Unit
You can also connect the video device under test to the VM Series System by using the supplied VGA Connector to 5X cable. See Figure 2--3. If you use this method, you will need to connect the cable to the VM S eries System using the supplied precision terminations. Depending on which measurements you take, you may need to manually change terminations while taking measurements.
Figure 2- 3: VGA connector to 5X cable
2- 2
VM Series Video Measurement System User Manual
Connecting Input Signals
Use the following diagrams to determine the optimum input signal setup for your measurement needs.
Operating Basics
Setup 1 - Three-Wire
Analog Video with
Composite Sync (Option
SD/HD)
Setup 1A - Three-Wire
Video with Composite
Sync for Sync Pickoff
012- 1680- 01 (Option
SD/HD)
Setups 1A and 1B are the preferred setups for testing 3-wire component analog video signals (Y/G, Pb/B, Pr/R) with the composite sync signal on Y/G. These setups use the sync pickoff accessory to derive the trigger signal from the Y/G signal on CH1. The triggering source is set to Channel 4. This configuration provides more accurate low-level noise measurements (below --60 dB at 30 MHz bandwidth) than setup 2.
This type of connection allows you to make more accurate noise measurements on CH1 while triggering on CH4, without adding additional cabling between CH1 and CH4. The additional cabling could adversely impact frequency response and multiburst measurements.
To achieve the best frequency response and multiburst measurements, the75 termination should be connected as close to the input connector as possible, as shown in the setup illustration.
There are two ways to connect the Sync Pickoff to the VM Series System. One way connects the Sync Pickoff directly to the input and the other way connects through a T-connector attached to a 75 Ω termination. If you are using a 012--1680--01 Sync Pickoff (contains internal 75 Ω termination), connect the Sync Pickoff directly to the VM Series System CH1 input (using a TPA-BNC adapter) as shown in the following Figure 2--4. If you are using a 012--1680--00 Sync Pickoff (requires external 75 Ω termination), use the Figure 2--6 shown in Setup 1B.
Ω
VM Series Video Measurement System User Manual
2- 3
Operating Basics
Figure 2- 4: Connecting to a VM6000 / DPO7000 Series instrument
Figure 2- 5: Connecting to a VM5000 / TDS5000 Series instrument
2- 4
VM Series Video Measurement System User Manual
Operating Basics
Setup 1B - Three-Wire
Analog Video with
Composite Sync for Sync
Pickoff 012- 1680- 00
(Option SD/HD)
There are two ways to connect the Sync Pickoff to the VM Series System. One way connects the Sync Pickoff directly to the input and the other way connects through a T-connector attached to a 75 Ω termination. If you are using a 012--1680--00 Sync Pickoff (the part number is located on the label of the larger module), connect the Sync Pickoff to the VM Series System CH1 input as shown in the following Figure 2--6. If you are using a 012--1680--01 Sync Pickoff, use the Figure 2--4 shown in Setup 1A.
Figure 2- 6: Connecting to a VM6000 / DPO7000 Series instrument
VM Series Video Measurement System User Manual
2- 5
Operating Basics
Figure 2- 7: Connecting to a VM5000 / TDS5000 Series instrument
Connect to CH1 input
Sync Pickoff
(012-1680-00)
BNC Female to Female
(103-0028-00)
BNC Male to Male
(103-0029-00)
BNC “T” connector
(103-0030-00)
Connect to CH4 for trigger
Figure 2- 8: Connecting the 012- 1680- 00 Sync Pickoff to CH1 input
NOTE.
will be unable to take measurements.
Verify that the trigger channel is set to Channel 4 (the default setting). If the trigger setting
(Configuration > Operation) is not set to Channel 4, you will see signal warning messages and
75 Ω Precision
end-line termination
(011-0102-03)
2- 6
VM Series Video Measurement System User Manual
Operating Basics
Setup 2 - Three-Wire
Analog Video with
Composite Sync (Option
SD/HD)
Use this setup for testing 3-wire component analog video signals (Y/G, Pb/B, Pr/R) with the composite sync signal on Y/G. Set the triggering on the instrument to Channel 1 (the default trigger setting is CH4). This setup is simpler than the others mentioned in this section, but limits noise measurement on Y/G to --65 dB (30 MHz noise bandwidth).
Figure 2- 9: Connecting to a VM6000 / DPO7000 Series instrument
Figure 2- 10: Connecting to a VM5000 / TDS5000 Series instrument
VM Series Video Measurement System User Manual
2- 7
Operating Basics
Setup 3 - Four-Wire Video
with Separate Composite
Sync (Option SD/HD)
Use this setup for testing 4-wire component analog video signals (Y/G, Pb/B, Pr/R) with the composite sync signal on a separate output line. For CH4, termination is optional. For only CH4, you can connect the signal without using a termination or you can use a 75
Ω terminations that are provided with the instrument. See Figure 2--11.
75
Ω termination of lower accuracy than the
Figure 2- 11: Connecting to a VM6000 / DPO7000 Series instrument
Figure 2- 12: Connecting to a VM5000 / TDS5000 Series instrument
2- 8
VM Series Video Measurement System User Manual
Operating Basics
Option VGA Setup 1:
RGBHV Measurement
Interface Unit
NOTE.
will be unable to take measurements.
Use the following diagrams to determine the optimum input signal setup for your measurement needs.
This is the preferred setup for testing PC and play-out device video. Use the RGBHV Measurement Interface Unit to connect the PC video card to the VM Series System. Using the RGBHV MIU provides the most accurate measurement results.
Verify that the trigger channel is set to Channel 4 (the default setting). If the trigger setting
(Configuration > Operation) is not set to Channel 4, you will see signal warning messages and
Figure 2- 13: Connecting to a VM6000 / DPO7000 Series instrument
VM Series Video Measurement System User Manual
2- 9
Operating Basics
Figure 2- 14: Connecting to a VM5000 / TDS5000 Series instrument
2- 10
VM Series Video Measurement System User Manual
Operating Basics
To connect the RGBHV Measurement Interface Unit:
1. Connect the device-under-test (DUT) to the input side of the RGBHV
Measurement Interface Unit using one of the VGA-to-BNC adapter cables supplied with your VM Series System.
Figure 2- 15: Connecting the DUT to the RGBHV Measurement Interface Unit
2. Connect the outputs (Aux, CH1--CH4) of the RGBHV Measurement
Interface Unit to same inputs on the VM Series System (that is, connect the CH1 output to the CH1 input).
Figure 2- 16: Connecting the RGBHV MIU to a VM6000 / DPO7000 Series instrument
VM Series Video Measurement System User Manual
2- 11
Operating Basics
Figure 2- 17: Connecting the RGBHV MIU to a VM5000 / TDS5000 Series instrument
3. Connect the RS-232C cable between the RGBHV Measurement Interface Unit and the COM1 port on the VM Series System. See Figure 2--18.
Figure 2- 18: Connecting the RS-232C cable between the MIU and the VM Series System
4. Verify that the Use MIU check box on the Configuration > Operations tab is selected. See Figure 2--19.
2- 12
VM Series Video Measurement System User Manual
Operating Basics
Figure 2- 19: Selecting the Use MIU option on the Configuration > Operations tab
Option VGA Setup 2: VGA
Connector to 5X Cable
Use this setup for testing PC and play-out device video when it is not convenient to use the RGBHV Measurement Interface Unit. To ensure accurate measurements, connect the signals to the VM Series System using the T-connec­tors and precision terminations on CH1--CH4. You must use 75
Ω terminations
onCH1--CH3(usedforGBR)and2.21kΩ and 301 Ω terminations on CH4 (used for H Sync and V Sync measurements). This method requires you to change the signal and termination connections on CH4 for certain horizontal and vertical measurements. The VM Series displays a dialog box prompting you to make these changes when necessary (the RGBHV Measurement Interface Unit performs these connection and termination changes internally).
VM Series Video Measurement System User Manual
2- 13
Operating Basics
To connect the DUT to the VM Series System using a 5X cable:
1. Install the included 75 Ω terminations between the end of the BNC cables and the instrument input Channels 1 through 3, using the included BNC T-connectors. See Figure 2--20. To achieve accurate measurements, it is important to attach the T-connectors directly to the input channel BNCs.
Figure 2- 20: Connecting the VGA to 5X cable to the input channel
2. If you will be making H Sync, V Sync, or V Timing measurements, you will need to make 2.21 kΩ and 301 Ω terminations using the B NC--to--Binding Post adapters and resistors supplied as standard accessories.
2- 14
VM Series Video Measurement System User Manual
Attach each resistor to an adapter . See Figure 2--21.
Operating Basics
Figure 2- 21: Building the 2.21 kiloohms and 301 ohms terminations
3. Connect the PC video output signals to the VM Series System. See Figure 2--23. Connect the H S ync signal using the 2.21 kΩ adapter to begin with. The VM Series System will prompt you to change it to the 301 Ω adapter if necessary.
VM Series Video Measurement System User Manual
2- 15
Operating Basics
Figure 2- 22: Connecting the PC video output signals to the VM6000 / DPO7000 Series instrument
Figure 2- 23: Connecting the PC video output signals to the VM5000 / TDS5000 Series instrument
2- 16
A pop-up warning appears to prompt you to change signal termination on CH4 when making a H Sync measurement. See Figure 2--24.
VM Series Video Measurement System User Manual
Operating Basics
Figure 2- 24: Appearance of the pop-up warning when making H Sync measurement
Table 2--1 identifies which signal from the Device-Under-Test (DUT) should be connected to which input connector on the VM Series System and the terminator that should be used for each signal.
Table 2- 1:
VGA signal (Device Under Test)
Green CH1 75
Blue CH2 75
Red CH3 75
H Sync CH4
V Sync AUX None
1
Required connections of the VGA signal, VM Series System, and the terminator
VM Series System Input Connector Terminator
1
V Sync signal is connected to this channel only when making a V Timing or V Sync measurement. The VM Series System will prompt you for this connection change if necessary.
2.21 kand 301
Signal Path Compensation
To optimize measurement accuracy, see the oscilloscope online help to perform the instrument signal path compensation procedure. Use the procedure if the temperature has changed more than 5 _C since the last signal path compensation. Perform the signal path compensation once a week. Failure to do so may result in the instrument not meeting warranted performance levels at those volts/div settings.
VM Series Video Measurement System User Manual
2- 17
Operating Basics
Sync Pickoff Compensation - VM6000 and DPO7000 Series, Option SD/HD only
To properly compensate the sync pickoff:
1. Connect the sync pickoff cable to CH1 and CH4. the small housing, labeled “Connect to Trigger Channel”, should be connected to CH4.
Figure 2- 25: Connecting the Sync Pickoff cable to CH1 and CH4
2. Attach the sync pickoff BNC connector to a BNC-to-minigrabbers connector (Pomana Electronics model number 3789).
3. Connect the minigrabbers to the PROBE COMP connector.
Figure 2- 26: Attaching the Sync Pickoff to a BNC-to-minigrabbers adapter and connecting to PROBE COMP
2- 18
4. Select CH1 and CH4. Push FastAcq. Push AUTOSET.
VM Series Video Measurement System User Manual
Operating Basics
Both the Channel 1 and Channel 4 waveforms should be displayed. It is normal for the waveform to show tilt, caused by the 75 Sync Pickoff termination loading the PROBE COMP output signal.
5. Compare the Channel 1 and Channel 4 waveforms. If they do not have the same shape, adjust the Sync Pickoff compensation (the adjustment is accessed through the side of the small housing).
Figure 2- 27: Selecting AUTOSET
Figure 2- 28: Properly compensated
Figure 2- 29: Under compensated
VM Series Video Measurement System User Manual
2- 19
Operating Basics
Figure 2- 30: Over compensated
6. If necessary, adjust the sync pickoff. Repeat as needed
Figure 2- 31: Adjusting the compensation
2- 20
VM Series Video Measurement System User Manual
The Matrix Test Signal
1 2 3 4 5
6
7
8
9
10
11
12
13
14
15
16
17
18
Operating Basics
The Matrix test signal is available as optional accessories with this product. There is Matrix signal signal for Options SD/HD and a Matrix signal for Option VGA. The Matrix test signal incorporates all the analog video signal require­ments within one file (different files are used for Option SD/HD and Option VGA). Using this single test file eliminates the need to change test signals when running different measurements. You can select the setup file (.vmset) that matches the format of the Matrix test signal. The default line number setting is provided for each measurement type. You can change the line number as needed. See Appendix E, Signal Source Package Contents for a full description of the files in the Signal Source Disks.
Figure 2--32 shows the Option S D/HD Matrix test signal. Table 2--2 shows the signal elements.
Figure 2- 32: Option SD/HD Matrix signal
VM Series Video Measurement System User Manual
2- 21
Operating Basics
Table 2- 2: Option SD/HD Matrix signal elements
Signal Name
1. Color bars All 7. YPbPr Sweep
2. RGB Pulse
Bars
3. RGB Frequency Sweep
4. RGB Time Sweep
5. RGB Multiburst
6. RGB Pulse Bars
Format Signal Name Format Signal Name Format
Time
RGB 8. YPbPr
Multiburst
RGB 9. YPbPr
Multiburst
RGB 10. YPbPr
Multiburst
RGB 11. RGB Crosstalk RGB 17. Flat Field 50% All
RGB 12. YPbPr
Crosstalk
YPbPr 13. Valid Ramp All
YPbPr 14. YPbPr Shallow
Ramp
YPbPr 15. RGB Shallow
Ramp
YPbPr 16. Flat Field 7.5% All
YPbPr 18. Flat Field
100%
YPbPr
RGB
All
2- 22
VM Series Video Measurement System User Manual
Operating Basics
Figure 2--33 shows the Option VGA Matrix test signal. Table 2--3 shows the signal elements.
1
2
3
4
5
6
7
8
Figure 2- 33: Option VGA Matrix test signal
Table 2- 3: Elements of Option VGA Matrix Signal
Signal Name
1. Half White and Half Black 5. Black White Bl ack
2. Nine Colorbars 6. Full White
3. 32 Steps 7. Ramp
4. Alternate White and Black 8. Half White and Half Black
The addressable line number pattern of the matrix signal is based on the following equations:
Signal Name
Half White and Half Black Bars (1--9.5%) Nine Color Bars (10.5--29.5%)
VM Series Video Measurement System User Manual
2- 23
Operating Basics
32 Steps ( 30.5--39.5%) Alternate White and Black Bars( 40.5--49.5%) Black--White--Black B ars (50.5--59.5%) Full White (60.5--69.5%) Staircase (70.5--89.5%) Half White and Half Black Bars(90.5--100%)
The actual line number which user enters in the application is calculated based on the following equation:
Active Line Number = Vertical Sync Time + Back Porch + Addressable Line Number according percentage
For 1280x1024 resolution, 85 Hz refresh rate, and DMT timing standard the Vertical Sync Line is 3, V ertical Back Porch is 44.
For Half White and Half Black pattern, The Start Active Line Number = 3 + 44 + 1 = 48 The End Active Line Number = 3 + 44 + 97 = 144 The line range for Half White and Half Black for 1280x1024 at 85 Hz would be 48 -- 144.
Table 2--4 shows an example of the resolutions and the corresponding line number ranges the DMT timing standard supports at 60 Hz refresh rate for Option VGA matrix signal.
Table 2- 4: Resolutions and line number ranges supported at 60 Hz (DMT standard) for Option VGA
1280x
Pattern
1. Half White and Half Black
2. Nine Colorbar 77--169 90--204 116--262 149--343 175--403 168--396 210--484
3. 32 Steps 173--217 210--264 269--338 353--445 415--523 408--516 498--628
4. Alternate White and Black
5. BlackWhite­Black
6. Full White 317--361 390--444 500--569 661--753 775--883 768--876 930--1060
7. Ramp 365--457 450--564 576--722 763--957 895--1123 888--1116 1074--1348
8. Half White and Half Black
640x480 800x600 1024x768
28--73 28--84 36--108 42--138 50--163 43--156 60--196
221--265 270--324 346--415 456--548 535--643 528--636 642--772
269--313 330--384 423--492 558--650 655--763 648--756 786--916
461--507 570--627 730--803 968--1065 1135--1249 1128--1242 1362--1499
1024
1600x 1200
1920x 1200
1920x 1440
2- 24
VM Series Video Measurement System User Manual
Starting the Application
1. Power on the instrument. The instrument displays the default main screen.
Operating Basics
For other refresh rates, the actual starting line number can be noticeably different from the 60 Hz refresh rate. To view the active video line number location for a specific pattern at another refresh rate, locate the Microsoft Excel file, Matrix­Signals-LineNumbers.xls located on the VM Series software CD. This file lists the Actual Video line numbers for all supported refresh rates.
Ensure that you have set up the input signal sources before running the measure­ments (See page 2--28).
1
2. On the menu bar, click File > Run Application > VM HD and SD Video V3.2.
or Analyze > VM HD and SD Video V3.2 or
File > Run Application > VM VGA Video V3.2 or
Analyze > VM VGA Video V3.2.
Option SD/HD
2
Option VGA
2
VM Series Video Measurement System User Manual
2- 25
Operating Basics
3. The instrument reduces the waveform area to the upper part of the screen, and opens the VM Series application in the lower part of the screen.
Option SD/HD
3
Option VGA
2- 26
3
VM Series Video Measurement System User Manual
The VM Series Application User Interface
VGA
Operating Basics
1. Menu bar: The menu bar allows
you to configure the different parameters of the application. Use the menu bar to display the results information in the Results window.
2. Application view: The main
application window displays the configuration controls and the measurement results.
3. Run button: Click Run to run
the selected measurements and to display the results.
4. Stop button: Click Stop to stop
running the selected measurements.
1
2
2
1
22
Option SD/HD
3
5
8
Option SD/HD
Option VGA
Run button when multiple
3
lines is selected
4
6
7
Run button
3
when run mode is
set to continuous
7
Option
3
5
8
4
7
5. Picture button: Click the Picture button to display the picture of the signal.
6. Vectorscope button (Option SD/HD): Displays the signal in vectorscope format.
7. To access the Windows desktop, for example, to view reports, click Hide. In the main screen,
click File > Minimize, and then click Hide again.
VM Series Video Measurement System User Manual
2- 27
Operating Basics
8. Exit button: Click Exit to quit the application. From the Exit dialog box you can choose whether to retain the settings. The default setting for Restore Scope Setting? is No. Click the button to change the option to
Yes .
To quit the application, click OK. If you don’t want to quit the application, click Cancel.
9. To restore the application win­dow after hiding it, select the
Analyze > Restore Application.
Measurement Steps Overview
8
8
9
8
1. Use the Configuration > Format menu to set the input signal format standard (such as 1080i/60) and refresh rate (page 2--29).
2. Use the Configuration > Measurements menu to select one or more measurements to run (page 2--33).
3. Use the Configuration > Operations menu to set the measurement run operations, such as run mode and trigger channel (page 2--34).
4. Click Run to run the selected measurements and view the measurement results (page 2--48).
5. Use the Utilities > Generate Report menu to save the measurement results to a file (page 2--58).
2- 28
VM Series Video Measurement System User Manual
Setting the Input Signal Format
Operating Basics
1. On the menu bar, click
Configuration > Format.
2. The Format panel appears.
When Auto Detect is selected or a valid format is not available, then the format dependent measurement input parameters are disabled.
1
2
2
Option SD/HD
Option VGA
1
Option SD/HD
Option VGA
VM Series Video Measurement System User Manual
2- 29
Operating Basics
3. Select one of the standard video formats or define your own format using the User Defined Format option.
WhenAuto Detectisselected, the application automatically detects the input format once the run button is pressed without you having to specify a valid format, selects the associated measurement line number setup, and performs the selected measurements.
Option SD/HD
3
Option VGA
3
If relative and limit testing have been enabled, then the application will also load the appropriate reference and limit values from the files associated with the detected format.
The detected format name is displayed under the measurement name in the Results panel for each selected measurement. “------” will be displayed when a valid format has not been detected.
2- 30
VM Series Video Measurement System User Manual
Operating Basics
4. The application supports creat-
ing customized formats and provides the values for the list of parameters that are required to correctly define and identify a format. You can choose a user-defined custom format by selecting the “User Defined Format” in the Format Configu­ration pane.
To add a new user-defined for­mat, select User Defined For- mat and click Add.
To edit a user-defined format select User Defined Format and click Edit. The User De­fined Format window appears. Enter appropriate values for the horizontal and vertical parame­ters. To save your new format, enter a name in the Format Name field and click Save.
Option SD/HD
4
Option VGA
4
Option SD/HD
4
To delete a user defined format, select the User Defined format and click Delete.
4
VM Series Video Measurement System User Manual
Option VGA
2- 31
Operating Basics
5. Select the appropriate Color Space (Option SD/HD) or Refresh Rate (Option VGA).
6. Select the appropriate timing standard for the signal (Option VGA).
H DMT: The Display Monitor Timing standards and guidelines are used for
lower resolutions or refresh rates that are in common industry use in lower performance systems.
H CVT: The Coordinated Video Timing standard is a method for generating a
consistent and coordinated set of standard formats, display refresh rates, and timing specifications for computer display products, both those employing CRTs and those using other display technologies. CVT addresses problems found in GTF.
Option SD/HD
5
6
Option VGA
5
Option VGA
2- 32
H CVT--R: The Coordinated Video Timing Reduced Blanking standard is for
use with non--CRT displays that can accept reduced horizontal blanking times.
H GTF: The Generalized Timing Formula standard is a method for generating
general purpose display timings and new formats. This allows much greater flexibility in the choice of refresh rates and pixel formats than currently available with discrete monitor timings. GTF enables new formats and refresh rates to be developed, but is based on some set parameters to improve the likelihood of interchange.
VM Series Video Measurement System User Manual
If a timing standard or refresh rate is not available for the selected resolution, then the timing standard or refresh rate is disabled. Select the resolution, refresh rate, and the timing standard for the signal you are measuring. If a resolution is selected and the currently selected refresh rate is not available in the selected resolution, the refresh rate will automatically change to the default value of 60 Hz. A message informs you of the change. If a resolution is selected and the currently selected timing standard is not available for the selected resolution, the timing standard will automatically change to a supported timing standard in the order of CVT, DMT, and GTF. A message informs you of the change.
Selecting the Measurements
Operating Basics
1. On the menu bar, click
Configuration > Measurements.
2. Select from one to twelve
measurements in any combination.
Option SD/HD
1
1
Option VGA
Option SD/HD
2
Option VGA
2
VM Series Video Measurement System User Manual
2- 33
Operating Basics
Setting the Measurement Run Operations
1. Select Configuration > Operation.
2. The instrument displays the
Operations window. These settings control how the instrument sets up and acquires measurements when you click Run.
Option SD/HD
1
1
Option VGA
Option SD/HD
2
Option VGA
2
2- 34
3. Select how the instrument sets up and runs the measurements.
3
VM Series Video Measurement System User Manual
Operating Basics
Setup & Run: Configures the instrument setting and takes a measurement, for each selected
measurement, and for each measurement acquisition. Use this when taking multiple measurements that require different instrument setups.
Setup Only: Configures the instrument for a measurement but does not take the measurement. Use
this to set up customized settings. Try out your settings by pressing the R un button, making adjustments as needed (for example, change the horizontal scale), and then pressing the Run button again. Once you are satisfied with your customization, use the Run Only setting to collect results. Setup Only is not selectable for certain measurements like H Sync, V Sync, Noise Inj Ratio, V Timing, and Video Transient as they need multiple settings for the oscilloscope and MIU.
Run Only: Takes a measurement using the existing instrument configuration. Designed for taking
measurements with custom settings (see the Setup Only description above).
4. Select how the application
acquires measurements (Run Mode).
4
Once takes each selected measurement the number of times determined by the averaging setting,
then stops taking measurements.
Continuously continually takes measurements. You must select the Stop button to stop measure-
ments.
Once & Report takes each selected measurement the number of times determined by the averaging
setting, presents the report, and stops taking measurements.
VM Series Video Measurement System User Manual
2- 35
Operating Basics
5. Auto Scale fits the signal within the available vertical scale. Auto Scaling adjusts the vertical scale units so that the input signals fill the graticule as much as possible. Auto Scaling occurs once for each selected measurement. Select the Enabled check box to enable waveform Auto Scale. Clear the Enabled check box to disable waveform autoscaling.
In the initial values pane, there are three settings: Last Meas, Last Recalled, and Restore Default.
5
Last Meas.: is the default. When selected, the values from the last measurement will be used at the
start of the next Auto Scale step.
Last Recalled: when selected, the values recalled with the last recalled setup will be used at the
start of the next Auto Scale step. If a setup (File > Recall Setup) has not been recalled since the program was started, the application will load the Auto Scale settings from the last power up.
Restore Default: when selected, the Factory Default Auto Scale starting points will be used at the
start of the next Auto Scale step. See page 3--8 for details.
NOTE. If you take measurements on the wrong line or with the wrong scale setup, press the Restore Default button before taking measurements again.
2- 36
VM Series Video Measurement System User Manual
6. (Option SD/HD) Select the
instrument channel that has the sync signal for triggering.
If you use the wrong trigger channel (and Signal Warnings is on), you will get a signal warn­ing. See the setup diagrams and trigger requirements on page 2--3.
7. Choose the Line Select setting
as appropriate. In Single Line mode, each measurement is made on a single line (for ex­ample, line 153). In Multi Lines mode, measurements are made across a range of lines (for ex­ample, lines 153--185).
Operating Basics
6
7
Single Line takes each selected measurement on the line specified in the Line Number field of the
Results window.
Multi Lines takes each selected measurement across a range of lines. You acquire measurements
across multiple lines by specifying the starting and the ending line numbers. When Multi Lines is selected, the Generate Report button in the Utilities > Generate Report panel is not select­able. When Multi Lines is selected, the Run Mode operation settings Once and Once & Report are selectable and the report is generated based on the configuration set in the Utilities > Generate Report. The Results Summary panel is not accessible when Multi Lines is selected. For more information, see Taking Measurements Across Multi Lines, on page 2--51.
VM Series Video Measurement System User Manual
2- 37
Operating Basics
8. (Option VGA) Select Use MIU to automatically route H and V signals to the CH4 output depending on the selected measurements.
9. (Option VGA) The specified polarities for the H Sync and V Sync signals vary between the timing standards.
In the absence of the MIU and with Auto Detect selected, the VM Series will prompt you to connect the H Sync and V Sync signals to determine the polarities. If you use the MIU, the VM Series will automatically determine the H Sync and V Sync polarities.
Once the sync signal polarities are determined, the measurement cycle begins. This “polarity test” is performed at the beginning of every measurement cycle. To prevent this test from running at the beginning of every measurement cycle, take any measurement once with Auto Detect selected. Once a measurement cycle has been completed with Auto Detect selected, you can clear Auto Detect.
8
9
The VM Series remembers the polarity of the sync signals. Any time you change the format setup for the device-under-test, you should complete a measurement with Auto Detect selected.
10.In the RTF Report Mode pane, select Embed Screen Capture to
10
include a snapshot of the waveform display in the results report only in RTF format.
2- 38
VM Series Video Measurement System User Manual
Setting the Measurement Warnings
Operating Basics
1. On the menu bar, click
Configuration > Warnin gs.
2. The instrument displays the
Warnings view. These settings control how the instrument sets up the Warning Types Reported and the Control Options when you select the Run button.
3. Select the warning messages to
report when a warning occurs.
1
2
Option SD/HD
Option VGA
1
See page 3--10 for more information.
3
VM Series Video Measurement System User Manual
2- 39
Operating Basics
4. Select how to report warning messages by selecting the check boxes in the control options pane. Messages can be displayed in a window or logged to a file.
4
NOTE.IfLog Warnings To File is selected, the warnings are written to the Log.txt file (Option SD/HD at C:\VMApps\OptHDSD or Option VGA at C:\VMApps\OptVGA). Warnings are appended to the Log.txt file until it reaches 1 MB in size. At that point, it is renamed LogOld.txt and a new Log.txt file is created. If the new file reaches 1 MB in size, it will be renamed LogOld.txt and will be written over the existing LogOld.txt file. If you want to save the data in the LogOld.txt file, you must save the file under a different name before it is overwritten.
2- 40
VM Series Video Measurement System User Manual
Setting Reference and Limits Options
VGA
When you run a measurement relative to a reference, the VM Series runs the measurements and displays the difference between the measured value and the reference value. The reference values are stored in a reference file. Limit Testing provides a Pass/Fail indication by comparing the measured value to limit values. The limit values are the maximum and minimum values that are stored in limit files.
NOTE. The default reference file (or default limits file) is provided as is without warranty of any kind, and is intended to be used as a template only. WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE EXPLICITLY DISCLAIMED.
Operating Basics
1. Select Configuration >
Reference & Limits.
2. The instrument displays:
the Relative Result Display and the Limit Testing in the Configuration display view.
1
2
Option SD/HD
Option
1
VM Series Video Measurement System User Manual
2- 41
Operating Basics
3. When you select Enable Relative Result Display, the
VM Series enables the options in Reference Selection to select a reference file. For more details, see page 3--78.
Capture Result as Reference
saves measurements to a reference file using the actual measured values as the reference values. For more details, see page 3--82.
4. Select Enable Limit Testing to enable pass/fail testing. When limit testing is enabled, the choices under Limits Selection are used to specify the limits file used. By default, the VM Series application automatically selects a limits file from a set of supplied templates that matches the selected signal format. For more details, see page 3--82.
3
4
2- 42
When you select the
Stop on Limit Testing Failure
option, the measurement execu­tion stops when limit testing fails.
VM Series Video Measurement System User Manual
Displaying the Input Signal as a Picture
The VM Series will display a picture of the input signal for Options SD, HD, and VGA.
Operating Basics
1. Click the picture button.
2. This launches the video monitor
window. The instrument dis­plays the picture. Click the maximize button to expand the picture to full screen.
1
Option SD/HD
2
Option VGA
VM Series Video Measurement System User Manual
2
2- 43
Operating Basics
Changing Picture Display Settings
Option SD/HD
You can change the settings for the Picture and Vector displays as follows:
1. Select Configuration > Picture & Vector.
2. The instrument displays the
Picture & Vector panel. These settings control the display parameters of the picture and vectorscope displays.
3. Line Number sets the video line number that is displayed in the vectorscope display or that is highlighted in the picture display.
1
2
3
2- 44
Use the line number to identify features in the picture. To remove the line, set the line number to a line outside the active video. Note that the line number on the Picture & Vector tab is not used to specify where measurements are taken.
VM Series Video Measurement System User Manual
Loading...