Instructions
TPS2000 BNC
Replacement Kit
063-3829-00
*P063382900*
063382900

TPS2000 BNC
Removal and Replacement Procedure
NOTE. This procedure is to be performed by a
qualified service person.
BNC Removal
1. Use a flat-blade screwdriver to pop off the plastic
BNC cover from the front of the TPS2000
oscilloscope.
2. Use a 0.625” (16 mm) Hex socket, with a thin wall to
loosen and remove the BNC nut (Tektronix part
number 220-0265-xx).
3. Pinch and pull off the plastic BNC shell from the
channel connector. If needed, use a flat-blade
screwdriver to gently pry off the shell.
Plastic BNC shell
0.625” BNC Nut
BNC Cover