Tektronix TPS2000/B BNC Field Install Instructions

TPS2000/B BNC Shell
ZZZ
Replacement Kit
TPS2000/B BNC Sh
ell Removal and
Replacement Procedure
BNC Shell Removal
1. Use a at-blade screwdriver to pop off the plastic BNC cover from the front of the TPS2000 or TPS2000B oscilloscope.
2. Use a 0.625” (16 mm) Hex socket, with a thin wall, to lo nut (Tektronix part number 220-0265-xx).
3. Pinch and pull off the plastic BNC shell from the channel connector. If needed, use a at-blade screwdriver to gently pry off the shell.
osen and remove the BNC
BNC Shell Replac
1. Remove from BNC kit number 050-3598-xx
2. Install the BNC
desired.
3. Hand start the nut onto the BNC shell desired and make sure that it seats properly. Be careful not to cross-thread the nut.
4. Torque down the nut using a 12 inch-lb. (1.35 N-M) driver and a 0.625” (16 mm) Hex socket, with a thin wall.
CAUTION. Do not overtighten the nut. Doing so can damage the nut or BNC shell.
5. Push the BNC cover back on.
Copyright © Tektronix, Inc. All rights reserved.
www.tektronix.com
ement
a nut, Tektronix part number
220-0265-xx
a BNC shell, Tektronix part number
131-7528-xx
shell onto the channel
*P071289500*
071-2895-00
Loading...