Tektronix TPR1000,TPR4000 Compliance,Safety User manual

xx
TPR1000 and TPR4000 Active Power Rail Probes
Complian
ce and Safety Instructions
Termsinthismanual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or
practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or
practices that could result in damage to this product or other property.
Symbols and terms on the product
When this symbol is marked on the product, be sure to consult the manual to nd out the nature of the potenti to be taken to avoid them. (This symbol may also be used to refer the user to ratings in the manual.)
al hazards and any actions which have
The following symbol(s) may appear on the product:
Complianc
e information
This product is intended for use by professionals and trained personnel
only; it is not designed for use in households or by
children.
Questions about the following compliance information may be directed to the following address:
Tektronix, Inc. PO Box 500, MS 19-045 Beaverton, OR 97077, USA www.tek.com
Environmental considerations
This section provides information about the environmental impact of the product.
ion of hazardous substances
Restrict
Complies with RoHS2 Directive 2011/65/EU.
Key performance specications
Specication Description
Oscilloscope compatibility
Bandwidth in DC coupling mode
Bandwidth in DC reject mode
Linear dynamic range
Offset range
Attenuation 1.25x
Measurement accuracy
Noise
Input impedance
Offset ±60 V offset range
1
2
3
4
23
DPO70000 oscilloscopes require the optional TCA-VPI50 adapter. Frequency response optimized for <1 source impedance. Through SMA-to-SMA cable or Solder Micro-Coax tip. Max AC RMS of 1 V.
6 series MSO, 5 series MSO, MDO3000, MDO4000C, MSO/DPO5000B, DPO7000C, and DPO70000C/DX/SX
TPR1000: DC to 1 GHz
23
TPR4000: DC to 4 GHz
TPR1000: 10 kHz to 1 GHz
TPR4000: 10 kHz to 4 GHz
Up to 60 V DC, ±1 V
±60 V
2
DC linearity: <0.1%
Step response long-term aberrations: ±1%
<300 μV BW Limit)
<1 mV Bandwidth)
50 kDC to 10 kHz
50 AC > 1
Offset setting error: ±2 mV max, ±0.4 μV typical
noiseon6SeriesMSO(20MHz
p-p
noise on 6 Series MSO (Full
p-p
p-p
00 kHz
1
oscilloscopes
to bandwidth
4
Standard accessories
Each probe is shipped with one TPR4KIT accessory kit containing the following items:
Item Description
1.3 m cable, SMA male-to-MMCX male, 50
1
*P0713
63701*
071-3637-01
Important Safety information
This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe condition.
To safely perform service on this product, additional information is provided at the end of this section in the Service safety summary.
Use the product only as specied. Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. Carefully read all instructions. Retain these instructions for future reference.
Probes and test leads
e connecting probes or test leads, connect the power cord
Befor from the power connector to a properly grounded power outlet.
Inspect the probe and accessories. Before each use, inspect
and accessories for damage (cuts, tears, or defects in
probe the probe body, accessories, or cable jacket). Do not use if
ged.
dama
High temperature probe tips
WARNING. To prevent a burn injury, when using a solder
micro-coax or ex tip in a high temperature application, be sure to allow the tip to cool down before handling the tip.
Service safety summary
The Service safety summary section contains additional information required to safely perform service on the product. Only qualied personnel should perform service procedures. Read this Service safety summary and the General safety summary before performing any service procedures.
Do not service alone. Do not perform internal service or adjustments of this product unless another person capable of rendering rst aid and resuscitation is present.
Product end-of-life handling
Observe the following guidelines when recycling an instrument or component:
Equipment recycling. Production of this equipment required the extraction and use of natural resources. The equipment may contain substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. To avoid release of such substances into the environment and to reduce the use of natural resources, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
This symbol indicates that this product complies with the applicable European Union requirements according to Directives 2012/19/EU and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries. For information
cycling options, check the Tektronix Web
about re site (www.tek.com/productrecycling).
Operating overview
The TPR1000 and TPR4000 probes provide a low noise, large offset range solution for measurement of ripple on DC power rails ranging from –60 to +60 VD C and measurement of AC ripple between 200 μV
and1V
p-p
Refer to one of the following TPR1000 and TPR4000 Instruction Manuals for complete operating information and product specications. The manuals are available for download at www.tek.com\downloads.
Language Tektronix part number
English 077-1542-xx
Japanese 077-1543-xx
Simplied Chinese
077-1544-xx
Environmental specications
Specication Description
Maximum input voltage
Temperature
Humidity
Altitude 0 to 3000 m
DC: ±60 V
AC: ±30 V RMS
Pk-Pk: ±42 V peak
Probe body: 0 to +55 °C
Standard accessories: –40 to +125 °C
High-temp accessories: –55 to +155 °C
5 to 95% up to +40 °C, derate to 85% above +40 °C
atupto4GHz.
p-p
1.3 m cable, SMA male-to-SMA male, 50
Y-lead adapter, MMCX female-to-0.8 mm sockets
er cable, MMCX female-to-U.FL
Adapt female, 50
Adapter, MMCX female-to-square pin (0.062 centers)
nterface solder pins, set of 20
DUT i
Soldering aide tool, 0.062 solder pins over SMT
Solder-in cable adapter, MMCX female-to-solder micro-coax tip, 50 Ω, setof3
Solder-in cable adapter, MMCX female-to-solder ex-paddle tip, 50 Ω, setof3
Wire card, solderable enameled self-uxing copper wire (for use with the solder-in tips)
Probe tip tripod support (with living hinge)
Marker bands, set of 5 (for probe identication)
x
Optional high-temperature accessory kit
If you ordered the optional TPR4KITHT high-temperature accessory kit, you received the following items:
Item Description
2 m high-temperature cable, SMA male-to-MMCX male, 50
Solder-in cable adapter, MMCX female-to-solder micro-coax tip, 50 Ω, setof3
Solder-in cable adapter, MMCX female-to-solder ex-paddle tip, 50 Ω, setof3
Optional 1 GHz browser accessory kit
If you ordered the optional TPRBRWSR1G browser accessory kit, you received the following items:
Item Description
Browser
Ground leads (blade, 0.5 mm spring, 15 cm alligator)
Y-lead adapter, browser tip-to-0.8 mm sockets
1. To install the probe, slide the probe body into the FlexChannel or VPI receptacle. The probe clicks into place when fully engaged.
2. Attach one of the following probe cables to the SMA connector on the probe body:
SMA-to-SMA standard cable (standard accessory)
SMA-to-MMCX standard cable (standard accessory)
SMA-to-MMCX high-temperature cable (optional
accessory)
1 GHz browser probe cable (optional accessory)
CAUTION. To prevent damage to the probe, do not exceed
8 in-lbs torque on the SMA nut.
3. To disconnect the probe, press the latch release button and pull away from the instrument.
Attaching MMCX accessories
Gently insert the MMCX end of the cable into one of the following accessories: micro-coax tip, solder ex tip, u. adapter or MMCX to square pin Y-lead adapter, until you feel the connector engage. To remove an accessory, gently pull from the MMCX connection point.
Attaching solder-in accessories
Micro-coax tip. For convenient rst-time use, the solder
micro-coax tips are shipped pre-trimmed and ready t o be soldered to the test point. You can reuse a tip by removing the tip from the solder joint and then trimming the wire insulation back to expose the center pin and ground shield on the tip cable. To achieve the best measurements results with a trimmed tip, refer to the TPR1000 and TPR4000 User Manual for guidelines and a description of best practices.
Flex tip. To attach the solder ex tip, rst solder the enameled self-uxing copper wire (standard accessory) to the test point. Feed the wire through the vias on the end of the ex tip, and then apply a small amount of solder to the vias to attach the wire to the tip.
Using the optional 1 GHz browser
The optional 1 GHz browser kit contains the following parts: 1 GHz browser probe, square pin Y-lead adapter, micro-SMD clip, three ground leads (alligator, blade, spring), and four replacement probe-tip pins (two rigid, two spring loaded).
WARNING. To prevent injury to the operator or damage
to the probe, oscilloscope and device under test, do not touch the probe ground to any point that is not at the same potential as the chassis ground of the oscilloscope. The probe ground must be connected to the same potential as the chassis ground of the oscilloscope.
Installing ground leads. To obtain accurate measurements, always attach a ground lead to the probe tip before making measurements. It is recommended that you use the shortest ground lead that will function in your electrical application. The following illustration shows the browser probe tip, the tip cover and the three types of grounds leads supplied with the browser.
To install the ground leads:
Spring: Slide the ground lead over the probe tip until it seats around the metal portion of the probe-tip housing.
Alligator: Slide the ground lead prongs over the exposed metal between the plastic probe-tip sections.
Blade: Loc
ate the slot in the probe-tip housing as shown below. Slide the ground lead over the probe tip until the blade slid
es into the slot.
Micro-SMD clip
Replacement 0.5 mm browser tips (2 solid tips, 2 spring tips)
Optional solder-in, micro -coax tip accessory kit
If you ordered the optional TPR4SIACOAX accessory kit, you received the following items:
Item Description
Solder-in cable adapter, MMCX female-to-solder micro-coax tip, 50 Ω, setof3
Optional solder-in, ex-paddle tip accessory kit
If you ordered the optional TPR4SIAFLEX accessory kit, you received the following items:
Item Description
Solder-in cable adapter, MMCX female-to-solder ex-paddle tip, 50 Ω, setof3
Installation
NOTE. Your FlexChannel or VPI instrument may require a
software upgrade to support full functionality of the TPR1000 and TPR4000 probes. Before you connect the probe to an oscilloscope, refer to the required software versions table at the end of this document to check the version requirements.
Using the solder-pin installation tool
The supplied set of solder pins are intended to be installed on DUT circuit boards and used with the supplied MMCX-to-square-pin adapter. To install the solder pins, use the supplied soldering-aide tool as described below .
NOTE. The solder pins are extremely small and can be
challenging to handle. It is recommended to use tweezers and a magnifying tool when installing pins on a circuit board.
1. Carefully insert the solder pins into the soldering aide tool as shown below.
2. Use the soldering a ide tool to hold the solder pins in place while soldering the pins to the circuit board.
3. If necessary, apply a small amount of adhesive to further strengthen the connection to the circuit board. However, keep the height of the adhesive to a minimum to provide good electrical contact for the adapter.
Using the tripod
The tripod probe support adds stability to square-pin mounted test points. For more stability, use glue to attach the tripod legs to the DUT circuit board.
Connecting the Y-lead adapter and micro-SMD clip. The browser
udes a Y-lead adapter and a micro-SMD clip that connect
kit incl as shown below. The Y-lead adapter can also connect to square pins.
ng browser-tip pins. To remove the browser-tip pin, use
Replaci
pliers to grasp the pin and gently pull it out of the tip housing.
ll a new browser-tip pin, select between a solid (silver
To insta colored) or spring-loaded (gold colored) pin, and then use pliers
y insert t he pin into the browser-tip housing until you
to gentl feel the pin press against the bottom of the housing.
Required oscilloscope software versions
Oscilloscope Required software version
5and6SeriesMSO
MDO3000
MDO4000C
MSO/DPO5000B
DPO7000C
DPO70000C/DX/SX
1
The probe may operate with older versions of oscilloscope software. However, older software versions than those listed are not guaranteed to provide ful probe functionality.
1.12.5
1.274
1.09354
10.8.3.3
10.8.3.3
.1
10.9
62
1
l