Tektronix TPP0850 Instructions Manual

xx
TPP0850 800 MHz 50X High Voltage Probe
Instructions
Operating Information
The TPP0850 Probe is a high impedance probe with 50X attenuation that is designed for use with Tektronix DPO/MSO4000B & DPO/MSO5000 series ground-referenced oscilloscopes.
Connect the probe as shown in the illustrations below.
Compensating the Probe (Required for Initial Use)
must
1.
2.
If the probe compensation fails, verify that the signal and ground connections are secure at the Probe Comp connections on the oscilloscope, and secure at the probe head and tip.
compensate the probe after you attach it to an
Connect the probe to an oscilloscope channel.
On MSO/DPO5000 instruments:
a.
Connect the probe to the probe compensation output
terminals on the oscilloscope front panel.
b.
Select Vertical > Probe Cal....
c.
Select the channel number tab of the channel that you
connected the probe to.
d.
In the Calibration section, click Calibrate Probe.
On MSO/DPO4000B instruments:
a.
Press the Channel Menu front panel button for the
channel that you connected the probe to.
b.
Push the More button until Probe Setup is selected.
c.
Push the Calibrate Probe button and follow the
on-screen instructions.
Standard Accessories
WARNING.
To avoid electric shock when using the probe or accessories, keep your ngers behind the nger guard of the probe body and away from the shaded area shown in the accessory illustrations below.
Item Description
Small hook tip (TASH)
Use this tip to access test points in tight spaces.
Screw the hook tip onto the probe tip and then clamp the hook onto the circuit.
Rating: 2300 V CAT I 1000 V CAT II
Reorder Tektronix part number 013-0386-xx
Large hook tip (TALH)
Use this tip to connect to large components and terminals.
Screw the hook tip onto the probe tip and then clamp the hook onto the circuit.
Rating: 2300 V CAT I 1000 V CAT II
Reorder Tektronix part number 013-0384-xx
Ground lead (TAGL)
Slidethegroundleadoverthe probe head and snap it into place.
Connect the banana plug end directly to your circuit ground, or use the crocodile clip below.
Reorder Tektronix part numbers:
6 inch: 196-3524-xx
18 inch: 196-3525-xx
Item Description
Crocodile clip
Attach the clip to the banana end of the ground lead and then to your circuit ground.
Reorder Tektronix part number 344-0461-xx
Color bands
Use these bands to identify the oscilloscope channel at the probe head.
Reorder Tektronix part number 016-1886-xx (5 pairs)
Ground spring
To limit aberrations on high frequency signals caused by ground path inductance, bend the spring to reach nearby ground connections (~1 in).
Reorder Tektronix part number 214-5298-xx
Optional Accessories
The accessories shown below are available for the probes and are rated 30 V unless indicated otherwise.
Accessory
BNC to Tip Adapter, Unterminated
Rubber Spring Tip
Part number
013-0291-xx
206-0060-xx
*P071288400*
x
071-2884-00
Specications
Table 1: Electrical and mechanical specications
Bandwidth (–3 dB)
System rise time (typical)
System input capacitanc e (typical)
System attenuation accuracy 50:1 ±2.2%
System input resistance @DC 40 M±1%
Propagation delay ~6.1 ns
Cable length
Maximum input voltage
Performance Graphs
(see
800 MHz
<525 ps
1.8 pF
1.3 m
1000 V
)
CAT II (2500 Vpk)
RMS
Performance Graphs
Table 2: Environmental specications
Characteristics Description
Temperature
Operating Nonoperating
Humidity
Operating 5% to 95% relative humidity (RH) up to
Nonoperating
Altitude
Operating Nonoperating
0°Cto+50°C(+32°Fto+122°F) –40 °C to +71 °C (–40 °F to +160 °F)
+30 °C 5% to 75% RH above +30 °C to +65 °C. Noncondensing
5% to 45% RH +65 °C to +85 °C. Noncondensing
3.0 km (9,842 ft) maximum
15.4 km (50,000 ft) maximum
Table 3: Cer tications and compliances
Characteristics Description
EC Declaration of Conformity
Measurement Category Product Examples
Compliance was demonstrated to the following specication as listed in the Ofcial Journal of the European Communities:
Low Voltage Directive 2006/95/EC: EN61010-031/A1: 2008
CAT III: Distribution-level mains, fixed installation
CAT II: Local-level mains, appliances, portable equipment
CAT I: Circuits not directly connected to mains.
Characteristics Description
Pollution Degree 2
Additional Safety Standards
Do not operate in environments where cond­uctive pollutants may be present (as dened in IEC 61010-1). Rated for indoor use only.
IEC61010-031; IEC 61010-031/A1:2008
Equipment Recycling.
with the European Union’s requirements according to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE). For more information about recycling options, check the Support/Service section of the Tektronix Web site (www.tektronix.com).
This product complies
Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specied. Using the probe or accessories in a manner not specied could result in a shock or re hazard.
To Avoid Fire or Personal Injury
Ground-Referenced Oscilloscope Use.
Do not oat the reference lead of this probe when using with ground referenced oscilloscopes (for example, DPO, MSO, and TDS series oscilloscopes). The reference lead must be connected to earth potential (0 V).
Connect and Disconnect Properly.
Connect the probe output to the measurement instrument before connecting the probe to the circuit under test. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement instrument.
Avoid Electric Shock.
To avoid injury or loss of life, do not connect or disconnect probes or test leads while they are connected to a voltage source.
Observe All Terminal Ratings.
To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Avoid Electric Shock.
When using probe accessories, never exceed the lowest rating of the probe or its accessory, whichever is less, including the measurement category and voltage rating.
Avoid Electric Overload.
To avoid injury or re hazard, do not apply potential to any input, including the reference inputs, that varies from ground by more than the maximum rating for that input.
Avoid Exposed Circuitry and Do not Operate Without Covers.
Do not touch exposed connections and components
when power is present.
Inspect The Probe And Accessories.
Before each use, inspect probe and accessories for damage (cuts, tears, defects in the probe body, accessories, cable jacket, etc.). Do not use if damaged.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Safety Terms and Symbols Terms in This Manual.
These terms may appear in this manual:
WARNING.
Warning statements identify conditions or
practices that could result in injury or loss of life.
CAUTION.
Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbols on the Product.
These symbols may appear on
the product:
Contacting Tektronix
Web site: www.tektronix.com
Phone: 1-800-833-9200
Address: Tektronix, Inc.
Email:
Department or name (if known) 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
techsupport@tektronix.com
Warranty Information
For warranty information, go to www.tektronix.com/warranty.
Copyright © Tektronix, Inc. All rights reserved. www.tektronix.com
Loading...