Tektronix TPP0051 Series User manual

xx
TPP0051 50 MHz 10X Passive Probe
Instructions (Revision A)
Operating Information
The TPP0051 10X Passive Probe is a high impedance, passive probe with 10X attenuation that is designed for use with the following Tektronix oscilloscopes:
TBS1000B and TBS1000B-EDU oscilloscopes, which have 20 pF of input capacitance
The probe has no user- or Tektronix-serviceable parts.
except TPS2000 Series Oscilloscopes.
Do not oat the reference lead of this probe to >30 V when used with a TPS2000 Series Oscilloscope. Use either the P5120 or P5150 probes (oatable to 600 V CAT II), a similarly-rated passive high voltage probe, or an appropriately-rated high voltage differential probe when oating the reference lead above 30 V to the ratings of the high voltage probe used.
Connecting the Probe to the Oscilloscope
Connect the probe as shown in the illustrations below.
Do not oat this probe on any oscilloscope
, subject
RMS
RMS
RMS
3.
Push AUTOSET or otherwise adjust your oscilloscope to display a stable waveform.
4.
Adjust the trimmer in the probe until you see a perfectly at-top square wave on the display. (See illustration.)
WARNIN G.
To avoid electric shock, only use the insulated adjustment tool when making compensation adjustments.
Connecting the Probe to the Circuit
Use the standard accessories included with the probe to connect to your circuit.
WARNIN G.
To avoid electric shock when using the probe or accessories, keep ngers behind the nger guard of the probe body and accessories.
To reduce risk of shock, ensure the ground lead and ground spring are fully mated before connecting the probe to the circuit under test.
Standard Accessories
The accessories included with the probe are shown below.
Item Description
Ground lead, 6-inch
Secure thelead to the probe head ground and then to your circuit ground.
Reorder Tektronix part number 196-3521-xx
Ground spring
The ground spring minimizes aberrations on high-frequency signals caused by the inductance of the ground path, giving you measurements with good signal delity.
Attach the spring to the ground band on the probe tip. You can bend the spring out to ~0.75 in. away from the signal test point.
Reorder Tektronix part number 016-2028-xx (2 ea.)
Adjustment tool
Reorder Tektronix part number 003-1433-xx
*P071324800*
x
071-3248-00
Compensating the Probe
Due to variations in oscilloscope input characteristics, the low-frequency compensation of the probe may need adjustment after moving the probe from one oscilloscope channel to another.
If a 1 kHz calibrated square wave displayed at 1 ms/division shows signicant differences between the leading and trailing edges, perform the following steps to optimize low-frequency compensation:
1.
Connect the probe to the oscilloscope channel that you plan to use for your measurements.
2.
Connect the probe to the probe compensation output terminals on the oscilloscope front panel.
To avoid electric shock, only connect to the Probe Comp signal on the oscilloscope when making this adjustment.
Item Description
Color bands
Use these bands to identify the oscilloscope channel at the probe head.
Reorder Tektronix part number 016-0633-xx (5 pairs)
Hook tip
Press the hook tip onto probe tip and then clamp the hook onto the circuit.
Reorder Tektronix part number 013-0362-xx
Optional Accessories
You can order the following accessories for your probe.
Accessory Part number
Alligator Ground Lead, 12 in
6” Clip-on Ground Lead
Ground Spring, Short, 2 ea.
MicroCKT Test Tip
Micro Hook Tip 013-0363-xx
Universal IC Cap
Circuit Board Test Point/PCB Adapter
Wire, spool, 32 AWG
196-3512-xx
196-3198-xx
016-2034-xx
206-0569-xx
013-0366-xx
016-2016-xx
020-3045-xx
Specications
Table 1: Electrical and mechanical specications
Characteristic
Bandwidth (–3 dB) DC to 50 MHz
System atten– uation accuracy
Compensation range
System input resistance @DC
System input capacitance
System rise time (typical)
Propagation delay ~6.1 ns
Maximum input voltage
Cable length
10:1 ±3.2%
15 pF – 25 pF
10 MΩ ±1.5%
<12 pF
<7.0 ns
300 V
CAT II
RMS
1.3 m
Table 2: Environmental specications
Characteristics Description
Temperature
Operating Nonoperating
Humidity Operating and Non-Operating
Altitude
Operating Nonoperating
–10°Cto+55°C(14°Fto+131°F) –51 °C to +71 °C (–60 °F to +160 °F)
5% to 95% relative humidity (%RH) up to +30 °C (86 °F), 5% to 65% RH above +30° C up to +55 °C (131 °F)
3.0 km (10,000 ft) maximum
12.2 km (40,000 ft) maximum
Performance Graphs
Table 3: Cert ications and compliances
Characteristics Description
EC Declaration of Conformity
Safety Standards
Measurement Category Descriptions
Pollution Degree 2
Compliance was demonstrated to the following specication as listed in the Ofcial Journal of the European Communities:
Low Voltage Directive 2006/95/EC:
EN61010-031: 2002
UL61010-031;2007 (AM1) CAN/CSA C22.2 No. 61010-031-07 (AM1) IEC61010-031; IEC 61010-031/A1:2008
Category Examples of Products in this
CAT III Distribution-level mains, xed
CAT II
CAT I Circuits not directly connected to
Do not operate in environments where cond– uctive pollutants may be present (as defined in IEC 61010-1). Rated for indoor use only.
Category
installation
Local-level mains, appliances, portable equipment
mains.
Equipment Recycling.
with the European Union’s requirements according to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE). For more information about recycling options, check the Support/Service section of the Tektronix Web site (www.tektronix.com).
This product complies
Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specied. Using the probe or accessories in a manner not specied could result in a shock or re hazard.
To Avoid Fire or Personal Injury
Ground-Referenced Oscilloscope Use.
Do not oat the reference lead of this probe when using with ground referenced oscilloscopes (for example, TBS, DPO, MSO, and TDS series oscilloscopes). The reference lead must be connected to earth potential (0 V).
TPS2000 Series Oscilloscope Use.
lead of this probe above the rated oat voltage (30 V
Connect and Disconnect Properly.
Do not oat the reference
).
RMS
Connect the probe output to the measurement instrument before connecting the probe to the circuit under test. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement instrument.
Avoid Electric Shock.
Do not connect or disconnect probes
or test leads while they are connected to a voltage source.
Observe All Terminal Ratings.
To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Avoid Electric Shock.
When using probe accessories, never exceed the lowest rating of the probe or its accessory, whichever is less, including the measurement category and voltage rating.
Inspect the Probe and Accessories.
Before each use, inspect the probe and accessories for damage (cuts, tears, defects in the probe body, accessories, cable jacket, etc.). Do not use if damaged.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Safety Terms and Symbols Terms in This Manual.
These terms may appear in this manual:
WARNING.
Warning statements identify conditions or
practices that could result in injury or loss of life.
CAUTION.
Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbols on the Product.
These symbols may appear on
the product:
Contacting Tektronix
Web site: www.tektronix.com
Phone: 1-800-833-9200
Address: Tektronix, Inc.
Email:
Department or name (if known) 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
techsupport@tektronix.com
Warranty Information
For warranty information, go to www.tektronix.com/warranty
Copyright © Tektronix, Inc. All rights reserved. www.tektronix.com