TPP0051
50 MHz 10X Passive Probe
Instructions (Revision A)
Operating Information
The TPP0051 10X Passive Probe is a high impedance,
passive probe with 10X attenuation that is designed for use
with the following Tektronix oscilloscopes:
TBS1000B and TBS1000B-EDU oscilloscopes, which
have 20 pF of input capacitance
The probe has no user- or Tektronix-serviceable parts.
WARNING.
except TPS2000 Series Oscilloscopes.
Do not float the reference lead of this probe to >30 V
when used with a TPS2000 Series Oscilloscope. Use
either the P5120 or P5150 probes (floatable to 600 V
CAT II), a similarly-rated passive high voltage probe, or
an appropriately-rated high voltage differential probe
when floating the reference lead above 30 V
to the ratings of the high voltage probe used.
Connecting the Probe to the Oscilloscope
Connect the probe as shown in the illustrations below.
Do not float this probe on any oscilloscope
, subject
RMS
RMS
RMS
3.
Push AUTOSET or otherwise adjust your oscilloscope
to display a stable waveform.
4.
Adjust the trimmer in the probe until you see a perfectly
flat-top square wave on the display. (See illustration.)
WARNIN G.
To avoid electric shock, only use the
insulated adjustment tool when making compensation
adjustments.
Connecting the Probe to the Circuit
Use the standard accessories included with the probe to
connect to your circuit.
WARNIN G.
To avoid electric shock when using the probe
or accessories, keep fingers behind the finger guard of
the probe body and accessories.
To reduce risk of shock, ensure the ground lead and
ground spring are fully mated before connecting the
probe to the circuit under test.
Standard Accessories
The accessories included with the probe are shown below.
Item Description
Ground lead, 6-inch
Secure thelead to the probe
head ground and then to
your circuit ground.
Reorder Tektronix part
number 196-3521-xx
Ground spring
The ground spring
minimizes aberrations
on high-frequency signals
caused by the inductance of
the ground path, giving you
measurements with good
signal fidelity.
Attach the spring to the
ground band on the probe
tip. You can bend the spring
out to ~0.75 in. away from
the signal test point.
Reorder Tektronix part
number 016-2028-xx (2 ea.)
Adjustment tool
Reorder Tektronix part
number 003-1433-xx
*P071324800*
x
071-3248-00
Compensating the Probe
Due to variations in oscilloscope input characteristics,
the low-frequency compensation of the probe may need
adjustment after moving the probe from one oscilloscope
channel to another.
If a 1 kHz calibrated square wave displayed at 1 ms/division
shows significant differences between the leading and
trailing edges, perform the following steps to optimize
low-frequency compensation:
1.
Connect the probe to the oscilloscope channel that you
plan to use for your measurements.
2.
Connect the probe to the probe compensation output
terminals on the oscilloscope front panel.
WARNING.
To avoid electric shock, only connect to the
Probe Comp signal on the oscilloscope when making
this adjustment.
Item Description
Color bands
Use these bands to identify
the oscilloscope channel at
the probe head.
Reorder Tektronix part
number 016-0633-xx
(5 pairs)
Hook tip
Press the hook tip onto
probe tip and then clamp
the hook onto the circuit.
Reorder Tektronix part
number 013-0362-xx
Optional Accessories
You can order the following accessories for your probe.
Accessory Part number
Alligator Ground Lead, 12 in
6” Clip-on Ground Lead
Ground Spring, Short, 2 ea.
MicroCKT Test Tip
Micro Hook Tip 013-0363-xx
Universal IC Cap
Circuit Board Test Point/PCB Adapter
Wire, spool, 32 AWG
196-3512-xx
196-3198-xx
016-2034-xx
206-0569-xx
013-0366-xx
016-2016-xx
020-3045-xx

Specifications
Table 1: Electrical and mechanical specifications
Characteristic
Bandwidth (–3 dB) DC to 50 MHz
System atten–
uation accuracy
Compensation range
System input resistance
@DC
System input capacitance
System rise time (typical)
Propagation delay ~6.1 ns
Maximum input voltage
Cable length
10:1 ±3.2%
15 pF – 25 pF
10 MΩ ±1.5%
<12 pF
<7.0 ns
300 V
CAT II
RMS
1.3 m
Table 2: Environmental specifications
Characteristics Description
Temperature
Operating
Nonoperating
Humidity
Operating and
Non-Operating
Altitude
Operating
Nonoperating
–10°Cto+55°C(14°Fto+131°F)
–51 °C to +71 °C (–60 °F to +160 °F)
5% to 95% relative humidity (%RH) up
to +30 °C (86 °F), 5% to 65% RH above
+30° C up to +55 °C (131 °F)
3.0 km (10,000 ft) maximum
12.2 km (40,000 ft) maximum
Performance Graphs
Table 3: Cert ifications and compliances
Characteristics Description
EC
Declaration
of
Conformity
Safety
Standards
Measurement
Category
Descriptions
Pollution
Degree 2
Compliance was demonstrated to the following
specification as listed in the Official Journal of
the European Communities:
Low Voltage Directive 2006/95/EC:
EN61010-031: 2002
UL61010-031;2007 (AM1)
CAN/CSA C22.2 No. 61010-031-07 (AM1)
IEC61010-031; IEC 61010-031/A1:2008
Category Examples of Products in this
CAT III Distribution-level mains, fixed
CAT II
CAT I Circuits not directly connected to
Do not operate in environments where cond–
uctive pollutants may be present (as defined in
IEC 61010-1). Rated for indoor use only.
Category
installation
Local-level mains, appliances,
portable equipment
mains.
Equipment Recycling.
with the European Union’s requirements
according to Directive 2002/96/EC on waste
electrical and electronic equipment (WEEE). For
more information about recycling options, check
the Support/Service section of the Tektronix
Web site (www.tektronix.com).
This product complies
Safety Summary
Review the following safety precautions to avoid injury and
prevent damage to this product or any products connected
to it. To avoid potential hazards, use this product only as
specified. Using the probe or accessories in a manner not
specified could result in a shock or fire hazard.
To Avoid Fire or Personal Injury
Ground-Referenced Oscilloscope Use.
Do not float the
reference lead of this probe when using with ground
referenced oscilloscopes (for example, TBS, DPO, MSO,
and TDS series oscilloscopes). The reference lead must be
connected to earth potential (0 V).
TPS2000 Series Oscilloscope Use.
lead of this probe above the rated float voltage (30 V
Connect and Disconnect Properly.
Do not float the reference
).
RMS
Connect the probe output
to the measurement instrument before connecting the probe
to the circuit under test. Disconnect the probe input and
the probe reference lead from the circuit under test before
disconnecting the probe from the measurement instrument.
Avoid Electric Shock.
Do not connect or disconnect probes
or test leads while they are connected to a voltage source.
Observe All Terminal Ratings.
To avoid fire or shock hazard,
observe all ratings and markings on the product. Consult
the product manual for further ratings information before
making connections to the product.
Avoid Electric Shock.
When using probe accessories, never
exceed the lowest rating of the probe or its accessory,
whichever is less, including the measurement category and
voltage rating.
Inspect the Probe and Accessories.
Before each use, inspect
the probe and accessories for damage (cuts, tears, defects
in the probe body, accessories, cable jacket, etc.). Do not
use if damaged.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Safety Terms and Symbols Terms in This Manual.
These terms may appear in this manual:
WARNING.
Warning statements identify conditions or
practices that could result in injury or loss of life.
CAUTION.
Caution statements identify conditions or
practices that could result in damage to this product or
other property.
Symbols on the Product.
These symbols may appear on
the product:
Contacting Tektronix
Web site: www.tektronix.com
Phone: 1-800-833-9200