Tektronix TPI4202 Combined User/Service

xx
TPI4200 Series Protocol Analyzers
ZZZ
Installation and Safety
Instructions
*P071299500*
071-2995-00
xx
TPI4200 Series Protocol Analyzers
ZZZ
Instructions
www.tektronix.com
071-2995-00
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 5 0 0 Beaverto USA
For product information, sales, service, and technical support:
n, OR 97077
In North America, call 1-800-833-9200. Worl dwid e, visi t www.tektronix.com to nd contacts in your area.
Table of Contents
Preface ............................................................................................................... 1
Documentation ................................................................................................. 1
General Safety Summary .......................................................................................... 2
Compliance
EMC Compliance............................. ................................ ................................ . 4
Safety Compliance............................................................................................. 5
Environmental Considerations ........................................ ................................ ....... 7
Operating Requirements ........................................................................................... 8
Electrical Ratings .............................................................................................. 8
Environ
Physical Specications........................................................................................ 9
Basic Installation Procedure.................................. ................................ .................... 10
Power-On and Power-Off Procedure ............................................................................ 11
Information........................................................................................... 4
mental Ratings........................................................................................ 9
TPI4200 Series Installation and Safety Instructions i
Table of Contents
ii TPI4200 Series Installation and Safety Instructions
Preface
Documentation
Table 1: Product documentation
This document contains the following information:
Important safety precautions to avoid injury and prevent damage to this product or any products connected to it
EMC (electromagnetic compliance), safety, and environmental standards with which the instrument complies
Voltage, power, and environmental requirements to use the product
Installation procedure
Power-on and power-off procedure
The following table lists the documentation that is available for the product and shows where you can nd it: in a printed manual, on the product documentation CD-ROM, or on the Tektronix Web site at www.tektronix.com.
Item Purpose Location
Installation and Safety Instructions
s manual)
(thi
Guides
User
ine Help
Onl
Provides safety and compliance information along
high-level installation instructions.
with
ide operation and application information.
Prov
trument operation and user interface help.
ins
Printed manual and also available in electronic format a t
ektronix.com/manuals
www.t
uct Documentation CD
Prod
he instrument
On t
TPI4200 Series Installation and Safety Instructions 1
General Safety Summary
General Safet
To Avoid Fire or Personal
Injury
ySummary
Review the fo this product or any products connected to it.
To avoid pot
Only qualied personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions r
Use Proper Power Cord. Use only the power cord specied for this product and certied for the country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test leads w
Ground the Product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
llowing safety precautions to avoid injury and prevent damage to
ential hazards, use this product only as specied.
elated to operating the system.
hile they are connected to a voltage source.
Observe All Terminal Ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
The inputs are not rated for connection to mains or Category II, III, or IV circuits.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating o f that terminal.
Power Disconnect. The power cord disconnects the product from the power source. Do not block the power cord; it must remain accessible to the user at all times.
Do Not Operate Without Covers. Do not operate this product with covers or panels removed.
Do Not Operate With Suspected Failures. If you suspect that there is damage to this product, have it inspected by qualied service personnel.
Avoid Exposed Circuitry. Do not touch exposed connections and components when powe r is p resent.
2 TPI4200 Series Installation and Safety Instructions
General Safety Summary
Terms in this Manual
Symbols and Terms on the
Product
Do Not Operate i
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Provide Prop
details on installing the product so it has proper ventilation.
These terms may appear in this manual:
WAR NI NG .
in injury or loss of life.
CAUTION
damage to this product or other property.
These t
. Caution statements identify conditions or practices that could result in
erms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the ma
n Wet/Damp Conditions.
er Ventilation. Refer to the manual's installation instructions for
Warning statements identify conditions or practices that could result
rking.
WARNING indicates an injury hazard not immediately accessible as you
the marking.
read
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
TPI4200 Series Installation and Safety Instructions 3
Compliance Information
Compliance In
EMC Compliance
EC Declaration of Conformity – EMC
formation
This section environmental standards with which the instrument complies.
Meets intent of Directive 2004/108/EC for Electromagnetic Compatibility. Compliance was demonstrated to the following specications as listed in the Ofcial Journal of the European Communities:
EN 61326-1 2006. EMC requirements for electrical equipment for measurement, control
CISPR 11:2003. Radiated and conducted emissions, Group 1, Class A
IEC 61000-4-2:2001. Electrostatic discharge immunity
IEC 61000-4-3:2002. RF electromagnetic eld immunity
IEC 61000-4-4:2004. Electrical fast transient / burst immunity
IEC 61000-4-5:2001. Power line surge immunity
lists the EMC (electromagnetic compliance), safety, and
, and laboratory use.
123
1000-4-6:2003. Conducted RF immunity
IEC 6
IEC 61000-4-11:2004. Voltage dips and interruptions immunity
4
EN 61000-3-2:2006. AC power line harmonic emissions
EN 61000-3-3:1995. Voltage changes, uctuations, and icker
European Contact.
ektronix UK, Ltd.
T Western Peninsula Western Road Bracknell, RG12 1RF United Kingdom
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test object.
3
To ensure compliance with the EMC standards listed here, high quality shielded interface cables should be used.
4
Performance Criterion C applied at the 70%/25 cycle Voltage-Dip and the 0%/250 cycle Voltage-Interruption test levels (IEC 61000-4-11).
4 TPI4200 Series Installation and Safety Instructions
Compliance Information
Australia / New Zealand
Declaration of
Conformity – EMC
Safety Compliance
EC Declaration of
Conformity – Low Voltage
U.S. Nationally Recognized
Testing Laboratory Listing
Canadian Certication
Complies with t following standard, in accordance with ACMA:
CISPR 11:2003 accordance with EN 61326-1:2006.
Compliance was demonstrated to the following specication as listed in the Ofcial Journal of the European Communities:
Low Voltage Directive 2006/95/EC.
EN 61010-1: 2001. Safety requirements for electrical equipment for measurement control and laboratory use.
UL 61010 equipment.
CAN/CSA-C22.2 No. 61010-1:2004. Safety requirements for electrical equipm
he EMC provision of the Radiocommunications Act per the
. Radiated and Conducted Emissions, Group 1, Class A, in
-1:2004, 2
ent for measurement, control, and laboratory use. Part 1.
nd
Edition. Standard for electrical measuring and test
Additional Compliances
pment Type
Equi
Safety Class
Pollution Degree
cription
Des
IEC 61010-1: 2001. Safety requirements for electrical equipment for measurement, control, and laboratory use.
Test and measuring equipment.
Class1–groundedproduct.
A measure of the contaminants that could occur in the environment around
within a product. Typically the internal e nvironment inside a product is
and considered to be the same as the external. Products should be used only in the environment for which they are rated.
Pollution Degree 1. No pollution or only dry, nonconductive pollution occurs. Products in this category are generally encapsulated, hermetically sealed, or located in clean rooms.
Pollution Degree 2. Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is caused by condensation must be expected. This location is a typical ofce/home environment. Temporary condensation occurs only when the product is out of service.
TPI4200 Series Installation and Safety Instructions 5
Compliance Information
Pollution Degree
Installation (Overvoltage)
Category Descriptions
Pollution Degr that becomes conductive due to condensation. These are sheltered locations where neither temperature nor humidity is controlled. The area is protected from direct sunshine, rain, or direct wind.
Pollution Degree 4. Pollution that generates persistent conductivity through conductive dust, rain, o r snow. Typical outdoor locations.
Pollution Degree 2 (as dened in IEC 61010-1). Note: Rated for indoor use only.
Terminals on this product may have different installation (overvoltage) category designations. The installation categories are:
Measurement Category IV. For meas urements performed at the source of low-voltage installation.
Measurement Category III. For measurements performed in the building installation.
Measurement Category II. For measurements performed on circuits directly connected to the low-voltage installation.
Measurement Category I. For measurements performed on circuits not directly connected to MAINS.
ee 3. Conductive pollution, or dry, nonconductive pollution
Overvoltage Category
MAINS overvoltage Category II
6 TPI4200 Series Installation and Safety Instructions
Compliance Information
Environmenta
Product End-of-Life
l Considerations
This section provides information about the environmental impact of the product.
Observe the following guidelines when recycling an instrument or component:
Handling
Equipment Recycling. Production of this equipment required the extraction and
use of natural resources. The equipment may contain substances that could be harmful to end of life. In order to avoid release of such substances into the environment and to reduce the use of natural resources, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused o r recycled appropriately.
orate Materials. This product contains one or more type CR lithium
Perchl
batteries. According to the state of California, CR lithium batteries are classied as perchlorate materials and require spe cial handling. See www.dtsc.ca.gov/hazardouswaste/perchlorate for additional information.
the environment or human health if improperly handled at the product’s
This sym Union requirements according to Directives 2002/96/EC and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries. For informa Tektronix Web site (www.tektronix.com).
bol indicates that this product complies with the applicable European
tion about recycling options, check the Support/Service section of the
Restriction of Hazardous
Substances
This product has been classied as Monitoring and Control equipment, and is outside the scope of the 2002/95/EC RoHS Directive.
TPI4200 Series Installation and Safety Instructions 7
Operating Requirements
Operating Req
Electrical Ratings
Power Requirements
uirements
This section product safely and correctly.
The product has a single-phase power source with one current-carrying conductor at or near earth-ground (the neutral conductor).
NOTE. Systems with both current-carrying conductors live with respect to ground
(such as phase-to-phase in multiphase systems) are not recommended as power sources.
The following table list the power requirements.
Table 2: Power requirements
Characteristic Description
Source voltage 100 to 240 VAC, ±10% (90 to 264 VAC
Frequency 50 or 60 Hz
r consumption
Powe
k inrush current
Pea
provides the specications that you need to know to operate your
RMS
202 600 W maximum, 100 to 240 V, 50 to 60 Hz, single
TPI4 phase
208 1200 W maximum, 100 to 240 V, 50 to 60 Hz, single
TPI4 phase
13 A peak at 240 VAC, 50 Hz
WARNING. To reduce risk of re and shock, ensure the mains supply voltage
ctuations do not exceed 10% of the operating voltage range.
u
Fuses
Batteries
8 TPI4200 Series Installation and Safety Instructions
Only the line conductor is fused for over-current protection. The fuse is internal
nd not user replaceable. Do not attempt to replace the fuse. If you suspect the
a fuse has blown, return the instrument to an authorized service center for repair.
The instrument does not contain any user-replaceable batteries .
Environmental Ratings
Operating Requirements
Table 3: TPI4200 environmental requirements
Feature Description
Temperature
Humidity
20% to 80%
Altitude
Cooling This product is provided with forced-air cooling; do not block
Operating
Nonoperating
Operating 5% to 95% relative humidity at up to +30 °C
Nonoperating
ng
Operati
Nonoperating
+5 °C to +45 °C with 15 °C/hour maximum gradient, non-condensing, derated 1.0 °C per 300 meters above 1500 m altitude.
–20 °C to +60 °C, with 15 °C/hour maximum gradient, with no media installed in disc drives.
5% to 45% re non-condensing, and as limited by a maximum wet bulb temperature of +29 °C (derates relative humidity to 45% relative h
5% to 95% r
5% to 45% relative humidity above +30 °C, non-condensing, and as limited by a maximum wet bulb temperature of +29 °C (derate
Up to 300 temperature by 1 °C per 300 m (984 ft) above 1500 m (4921 ft ) altitude
Up to 12,000 m (39,370 ft)
lation openings.
venti
lative humidity above +30 °C up to 40 °C,
umidity at +40 °C)
elative humidity at up to +30 °C
s relative humidity to 20% relative humidity at +60 °C)
0 m (9843 ft), derates maximum operating
Physical Specications
Table 4: Physical specications
Characteristic TPI4202 TPI4208
Dimensions
Weight
Height
Width
Depth
424 mm (16.7 in.) 179 mm (7.0 in.)
333 mm (13.1 in.) 483 mm (19.0 i n.)
173 mm (6.8 in.) 457 mm (18.0 i n.)
10,5 kg (23.0 lb.) 20.5 kg (45.0 lb.)
TPI4200 Series Installation and Safety Instructions 9
Basic Installation Procedure
Basic Installation Procedure
Verify that you have received all of the parts of your instrument and the following items:
Correct power cords for your geographical area
Standard accessories
All optional accessories that you ordered
Refer to your Quick Start user documentation that you received with your product for additional installation information.
10 TPI4200 Series Installation and Safety Instructions
Power-On and Power-Off Procedure
Power-On and P
Power-On
Power-Off
ower-Off Procedure
This instrum conductor at or near earth ground. The line conductor is fused for over-current protection. A protective ground c onnection through the grounding conductor in the power cord is essential for safe operation.
1. Connect the supplied power cord to the power connector.
2. Press the p
turn on.
NOTE. The Standby button on the front-panel does not disconnect mains power.
Only the power cord at the rear of the product can disconnect mains power.
1. Press the power button on the instrument front-panel to turn the instrument off.
2. If you want to remove power completely, disconnect the power cord from the
rear-panel of the instrument.
ent operates from a single-phase power source with the neutral
ower button on the instrument front-panel and the instrument will
TPI4200 Series Installation and Safety Instructions 11