Tektronix TMSSM3 User manual

Instruction Manual
TMSSM3 700-Pin Socket Hardware Support
071-1387-00
www.tektronix.com
Copyright © Tektronix, Inc. All rights reserved.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this public ation supercedes that in all previously published material. Specifications and price c hange privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.

WARRANTY 2

Tektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If a product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incom patible equipm ent; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER W ARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADV ANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

WARRANTY 9b

Tektronix warrants that the media on which this software product is furnished and the encoding of the programs on the media will be free from defects in materials and workmanship for a period of three (3) months from the date of shipment. If a medium or encoding proves defect ive during the warranty period, Tektronix will provide a replacem ent in exchange for the defective medium. Except as to the me dia on which this software product is furnished, this software product is provided “as is” without warranty of any kind, either express or implied. Tektronix does not warrant that the functions contained in this software product will meet Customer’s requirements or that the operation of the programs will be uninterrupted or error-free.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period. If Tektronix is unable to provide a replacement that is free from defects in materials and workmanship within a reasonable time thereafter, Customer may terminate the license for this software product and return this software product and any associated materia ls for credit or refund.
THIS WARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER W ARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPLACE DEFECTIVE MEDIA OR REFUND CUSTOMER’S PAYMENT IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS W ARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Table of Contents

Getting Started
General Safety Summary v...................................
Service Safety Summary vii....................................
Preface ix...................................................
Manual Conventions ix..............................................
Contacting Tektronix x..............................................
Product Description 1--1..............................................
Product Software Compatibility 1--1.................................
Logic Analyzer Software Compatibility 1--1...............................
Logic Analyzer Configuration 1--1......................................
P6860 Probes 1--2................................................
Labeling P6860 Probes 1--2........................................
LAI Cables 1--2..................................................
Labeling LAI Cables 1--2..........................................
Standard and Optional Accessories 1--2...................................
Probe Adapter Review 1--2............................................
System Clock Rate 1--3............................................
Acquisition before Reset 1--3.......................................
Data Bus 1--3....................................................
Address Bus 1--3.................................................
Disabling the Cache 1--3...........................................
Configuring the Probe Adapter 1--4......................................
BINIT OBS Jumper 1--4...........................................
Connect the Logic Analyzer to the Target System 1--5.......................
Removing the Probe Head from the Target System 1--8...................
Applying LAI Labels 1--9.............................................
Connect the Probes and the LAI Cables 1--10...............................
PUB64G4, IA64G4, or IA64G4_C5 Setups 1--12........................
Applying and Removing Power 1--13.....................................
Installing the Software 1--14............................................
Care and Maintenance 1--15.............................................
Short-Term Storage 1--16...........................................
Long-Term Storage 1--17...........................................
Shipping the Probe Adapter 1--19........................................
Specifications
Circuit Description 2--1...............................................
Probe Adapter Loading Diagrams 2--3....................................
Specification Tables 2--5..............................................
TMSSM3 700-Pin Socket Support
Latched Operation 2--1............................................
Signal Probing 2--2...............................................
Bus Tracking Logic 2--2...........................................
Common Clock 2--2..............................................
i
Table of Contents
Reference
Replaceable Parts List
Index
Reference Tables 3--1..........................................
Viewing Reference Tables 3--1..........................................
To Extract Groups 3--2............................................
Parts Ordering Information 4--1.........................................
Using the Replaceable Parts List 4--2.....................................
ii
TMSSM3 700-Pin Socket Support

List of Figures

Table of Contents
Figure 1--1: Jumper location 1--4.................................
Figure 1--2: Attach the brackets and probe head 1--6................
Figure 1--3: Connect the microprocessor and attach your heatsink 1--7.
Figure 1--4: Apply LAI labels 1--9................................
Figure 1--5: LAI Cable, P6860 probe, and preprocessor unit 1--10......
Figure 1--6: Module, cable, and probe configuration 1--12.............
Figure 1--7: Place the preprocessor unit and probe head in
static-shielding bags. 1--17....................................
Figure 1--8: Place foam in the cardboard carton 1--17................
Figure 1--9: Place the preprocessor unit in the cardboard carton 1--18..
Figure 1--10: Place the cables carefully over the preprocessor unit 1--18.
Figure 1--11: Place the foam and probe head in place 1--18............
Figure 1--12: Place accessories in the accessory tray 1--19.............

List of Tables

Figure 2--1: Pin header electrical load model for typical signals 2--3...
Figure 2--2: Pin header electrical load model for typical signals 2--4...
Figure 2--3: ZIF electrical load model for typical signals 2--4.........
Figure 2--4: Mated Mictor model 2--4.............................
Figure 2--5: BCLK model 2--4...................................
Figure 2--6: Dimensions of the probe head 2--7.....................
Figure 2--7: Dimensions of the preprocessor unit 2--8................
Figure 4--1: Probe adapter exploded view 4--5......................
Figure 4--2: Preprocessor unit exploded view 4--6...................
Table 1--1: Jumper information 1--4..............................
Table 1--2: P6860 probe and LAI Cable configuration 1--11...........
Table 2--1: Electrical specifications for the target system 2--5........
Table 2--2: Electrical specifications for the AC input 2--5............
Table 2--3: BCLK timing and electrical specifications 2--5.........
Table 2--4: Environmental specifications 2--6......................
Table 2--5: Certifications and compliances 2--6....................
TMSSM3 700-Pin Socket Support
iii
Table of Contents
iv
TMSSM3 700-Pin Socket Support

General Safety Summary

Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read the General Safety Summary in other system manuals for warnings and cautions related to operating the system.
ToAvoidFireor
Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and certified for the country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Connect the ground lead of the probe to earth ground only.
Do Not Operate Without Covers. Do not operate this product with covers or panels removed.
Use Proper Fuse. Use only the fuse type and rating specified for this product.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Provide Proper Ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
TMSSM3 700-Pin Socket Support
v
General Safety Summary
Symbols and Terms
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
CAUTION
Refer to Manual
WARNING
High Voltage
Protective Ground
(Earth) Terminal
vi
TMSSM3 700-Pin Socket Support

Service Safety Summary

Only qualified personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present.
Disconnect Power. To avoid electric shock, switch off the instrument power, then disconnect the power cord from the mains power.
Use Care When Servicing With Power On. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
To avoid electric shock, do not touch exposed connections.
TMSSM3 700-Pin Socket Support
vii
Service Safety Summary
viii
TMSSM3 700-Pin Socket Support

Preface

Manual Conventions

This instruction manual contains specific information about the TMSSM3 microprocessor support product and is part of a set of information on how to operate this product on compatible Tektronix logic analyzers.
If you are familiar with operating microprocessor support products on the logic analyzer, all you need is this instruction manual to set up and run the support product.
If you are not familiar with operating microprocessor support products, you need to supplement this instruction manual with information on basic operations from the Tektronix logic analyzer online help to set up and run the support product.
This manual uses the following conventions:
H The phrase “information on basic operations” refers to basic information in
the Tektronix logic analyzer online help.
H The terms “Master” and “Slave” refer to modules that are located in
numbered slots (see Figure 1--6 on page 1--12).
H The term “LAI” is an acronym for Logic Analyzer Interface.
TMSSM3 700-Pin Socket Support
ix
Preface

Contacting Tektronix

Phone 1-800-833-9200*
Address Tektronix, Inc.
Department or name (if known) 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
Web site www.tektronix.com
Sales support 1-800-833-9200, select option 1*
Service support 1-800-833-9200, select option 2*
Technical support Email: techsupport@tektronix.com
1-800-833-9200, select option 3*
6:00 a.m. -- 5:00 p.m. Pacific time
* This phone number is toll free in North America. After office hours, please leave a
voice mail message. Outside North America, contact a Tektronix sales office or distributor; see the Tektronix web site for a list of offices.
x
TMSSM3 700-Pin Socket Support
Getting Started

Getting Started

Product Description

This section contains information about configuring the probe adapter and applying power to the TMSSM3 product.
The TMSSM3 product is an interposer design that allows the logic analyzer to acquire data from a microprocessor in an operating environment with little effect on the target system.
The probe adapter is connected to the target system and the microprocessor is connected to the probe adapter. Signals from the microprocessor-based system flow through the probe adapter to the logic analyzer.
Product Software
Compatibility
The TMSSM3 product is shipped with the PUB64G4 software. The PUB64G4 software acquires signals for viewing state and timing signals.
NOTE. The TMS123 (IA64G4, IA64G4_T, IA64G4_C5, IA64G4_T_C5, and IA64G4_SST) software is compatible with the TMSSM3 product. The TMS123 software provides disassembly and timing analysis capabilities, synchronous transactions, and instruction decoding. The TMS123 software product is available only to customers with a valid, restricted, secret nondisclo­sure agreement (RS-NDA) with Intel.

Logic Analyzer Software Compatibility

Software version 4.3 SP1 of the logic analyzer software is compatible with the TMSSM3 PUB64G4 software and the TMS123 software products.

Logic Analyzer Configuration

To use the probe adapter to acquire either disassembly or timing signals, you need a TLA7AAX or TLA7ABX logic analyzer equipped with a minimum of four, 136-channel, 450 MHz modules. The modules must be configured and merged. The memory depth is chosen automatically based on the shallowest memory depth of the modules.
TMSSM3 700-Pin Socket Support
1--1
Getting Started
Two additional merged modules can be added to acquire both disassembly and timing signals. The TLA7XM mainframe is needed when using six modules.
P6860 Probes
Labeling P6860 Probes
LAI Cables
You can use the logic analyzer modules and P6860 probes to connect to the TMSSM3 probe adapter.
For more information on connecting the P6860 Probes to the preprocessor unit, seepage1--10.
The probe adapter relies on the default channel mapping and labeling scheme for the probes. Apply labels using the instructions described in the P6860 High Density Logic Analyzer Probe Label Instructions, Tektronix part number 071-1123-XX. This manual can be accessed from the Tektronix Web site (see the contact page at the front of this manual. This information is also located in the logic analyzer online help.
Refer to the P6810, P6860, and P6880 Logic Analyzer Probes Instruction manual, Tektronix part number 071-1059-XX, for more information. You can find information about the P6860 probes in the logic analyzer online help.
The LAI cables are designed to connect the logic analyzer modules to the TMSSM3 probe adapter.
For more information on connecting the LAI cables to the preprocessor unit, see page 1--10.
Labeling LAI Cables
For information on how to label the LAI cables, see page 1--9.

Standard and Optional Accessories

A complete list of standard and optional accessories is provided in the Replace ­able Parts List on page 4--4.
1--2
TMSSM3 700-Pin Socket Support

Probe Adapter Review

Getting Started
Along with this TMSSM3 instruction manual, each logic analyzer includes additional information about performing tasks that are common to microproces­sor support packages on that platform.
This additional information is located in the logic analyzer online help, an installation manual, or a user manual. Review the additional information if you are unfamiliar with using support products on a logic analyzer product.
Review electrical, environmental, and mechanical specifications in the Specifica- tions section on page 2--1 as they pertain to the target system, as well as the following information.
System Clock Rate
Acquisition before Reset
Data Bus
Address Bus
Disabling the Cache
The TMSSM3 hardware support can acquire data from the microprocessor operating at speeds of up to 333 MHz and 400 MHz. The tested clock rate is 200 MHz.
Contact the Tektronix sales representative for current information on the fastest devices supported.
If data is acquired just before a power on Reset signal is observed by the target system, the data acquired by the logic analyzer will be inaccurate.
The TMSSM3 probe adapter supports only a double data bus.
The TMSSM3 probe adapter supports only a common clock address bus.
The cache bus is not monitored; therefore, the cache must be disabled. Disabling the cache makes all instruction prefetches visible on the bus so that they are acquired and displayed and correctly disassembled.
TMSSM3 700-Pin Socket Support
1--3
Getting Started

Configuring the Probe Adapter

Position the jumper on the probe adapter as described in Table 1--1. Figure 1--1 shows the location of the jumper on the preprocessor unit.
Table 1--1: Jumper information
Jumper designation
BINIT_OBS 2--3 Observe BINIT
Pin numbers Jumper name
1--2 Ignore BINIT assertion
BINIT OBS Jumper
Place the jumper in the pin 2 to pin 3 position to select BINIT observed, allowing the BINIT signal to act as another reset input. Place the jumper in the pin 1 to pin 2 position to select BINIT unobserved.
Preprocessor
unit
BINIT_OBS
Figure 1--1: Jumper location
1--4
TMSSM3 700-Pin Socket Support

Connect the Logic Analyzer to the Target System

This probe-adaptor product comes with different mechanical kits that allow you to attach the probe adapter to specific platforms, processors, and power-pod adapters. If you do not know which of these mechanical kits you need to use, contact your Tektronix application engineer.
WARNING. To prevent harm to yourself or damage to the preprocessor unit, do not open the preprocessor unit. There are no operator serviceable parts inside the preprocessor unit. Refer servicing of internal parts in the preprocessor unit to Tektronix authorized personnel only. External parts may be replaced by qualified service personnel.
To prevent static damage to the microprocessor, the probe adapter, the probes, and the module, you must handle components only in a static-free environment. Always wear a grounding wrist strap, heel strap, or similar device while handling the microprocessor and probe adapter.
Getting Started
Tools
You must allow airflow clearance for the preprocessor unit. Refer to the dimensions described in Table 2--4 on page 2--6 for the required clearance.
Read the following instructions before installing parts.
The following tools are required:
H Phillips (P1) to attach the brackets to the target system.
3
H
/32-inch (2.5 mm) hexagonal wrench for the ZIF sockets
H Dremel (hand grinder) or wire cutters to trim off the heatsink alignment pins.
H (optional) A torque wrench helps to ensure reliable connections by meeting
the nominal torque values. When attaching screws to the probe head, use
4 in-lbs (0.451 Newton meters) of torque.
To connect the logic analyzer to the target system, follow these steps:
1. Power off the target system. It is not necessary to power off the logic analyzer.
2. Power off any probe adapters that may be attached to your target system.
3. T o discharge static electricity, touch the ground connector located on the logic
analyzer.
4. Remove the microprocessor from your target system.
5. Follow the steps in Figures 1--2 and 1--3 to install the probe head.
TMSSM3 700-Pin Socket Support
1--5
Getting Started
Lower bracket
Unlock the target
system ZIF socket
Target system
Nonspecific bracket — Use the bracket instruction sheet that came with your black case to attach the lower bracket to the target system.
1
2
Remove the pin protector,
and then connect the
probe head to the
target-system socket
3
Figure 1--2: Attach the brackets and probe head
Probe head
Target system
1--6
TMSSM3 700-Pin Socket Support
Getting Started
Lock the target
system ZIF socket
Using the four
captured screws,
attach the heatsink
to the top bracket.
Upper bracket
8
Nonspecific bracket — Use the bracket instruction sheet that came with your black case to attach the top bracket to the target
5
system.
4
Target system
Heatsink
Connect the
microprocessor assembly
to the probe-head socket
Lock the probe
head ZIF socket
7
Power pod
6
Target system
Figure 1--3: Connect the microprocessor and attach your heatsink
TMSSM3 700-Pin Socket Support
1--7
Getting Started
Removing the Probe Head
from the Target System
Follow these steps to remove the probe head from the target system:
1. Power off the target system, and unplug the AC power cord on the preproces-
sor unit. The power switch for the probe adapter is located on the back of the preprocessor unit. It is not necessary to power off the logic analyzer.
2. Reverse the steps in Figures 1--2 and 1--3 to remove the probe head.
3. Store the probe head in the original packing material.
1--8
TMSSM3 700-Pin Socket Support

Applying LAI Labels

Getting Started
The LAI cables are shipped without labels; you must attach labels to the module end and the preprocessor end of the LAI cables. Read the following note before you begin attaching labels.
NOTE. Always use flat-nosed tweezers to remove the labels from the sheet of labels. Never peel labels with your fingers. The labels are made of soft vinyl and can stretch and distort easily. To avoid stretching a label, always grasp it from the top right corner while removing it from the sheet of labels.
The adhesive on the vinyl labels is extremely strong. Carefully align the label to the indented outline on the module end and preprocessor unit end. Once labels are placed on the LAI cables, they are difficult to remove.
To attach labels, perform the following steps:
1. Determine which channel groups you are planning to use and identify the
matching labels.
2. Follow the steps in Figure 1--4 while attaching the labels.
Module end
Blank
4
Align and place the labels
in the label indents.
Repeat for all steps.
LAI cable
3
Blank
1
Match the
color and the
channel name.
Probe-head
end
2
Figure 1--4: Apply LAI labels
TMSSM3 700-Pin Socket Support
1--9
Getting Started

Connect the Probes and the LAI Cables

To configure the P6860 probes and LAI cables for the disassembly and timing support functions, you must determine if you are using the four or six module configuration (see Table 1--2 on page 1--11). Then follow the steps on page 1--12 to connect the logic analyzer modules to the preprocessor unit.
To apply labels to the LAI cables, see page 1--9. Figure 1--5 shows a P6860 probe, an LAI cable, and the preprocessor unit.
P6860 probe
LAI cable
Preprocessor unit
1--10
Figure 1--5: LAI cable, P6860 probe, and preprocessor unit
TMSSM3 700-Pin Socket Support
Table 1--2 lists the 4- or 6-module hardware and software configurations.
Table 1--2: P6860 probe and LAI cable configuration
Getting Started
Software loaded
PUB64G4 State 4 7 (M,S1,S2,S3**) -- -- -- Acquires state only
IA64G4 or IA64G4_C5
IA64G4_T or IA32G4_T_C5
IA64G4 and IA64G4_SST
* You need a TLA7XM mainframe for six modules.
Function Modules LAI cables
Disassembly 4 7 (M,S1,S2,S3**) -- -- -- Acquires disassembly only:
Timing 4 3 (M,S3**) 6(S1,S2) Acquires timing only:
Disassembly, plus Timing4,+2*
7 (M,S1,S2,S3**) 6 Acquires complete disassembly and timing:
P6860 probes
Description
S Can on trigger common clock signals S Can view common clock signals in timing S Can trigger on data content S Cannot view source synchronous data in
timing
S Can trigger on common clock signals S Can view common clock signals in timing S Cannot trigger on data content S Can view source synchronous data in timing
S Can trigger on common clock signals S Can view common clock signals in timing S Can trigger on data content S Can view source synchronous data in timing
on a separately unmerged modules (IA64G4_SST software on two auxiliary modules)
** Do not connect to RESERVED on the Slave3 module.
TMSSM3 700-Pin Socket Support
1--11
Getting Started
Figure 1--6 shows the configuration for a 4-wide and 6-wide module merge. An expanded mainframe is needed for 6 modules and the order of the modules does not matter.
PUB64G4 , IA64G4, or IA64G4_C5
7 LAI cables
IA64G4_T
M A S T E R
or IA32G4_T_C5
IA64G4
7 LAI cables
S L A V E 1
S L A V E 3
and
S L A V E 2
4 LAI cables and 6 P6860 probes
Figure 1--6: Module, cable, and probe configuration
Expanded mainframe
M A S T E
R
IA64G4_SST
6 P6860 probes
TLA7XM
S L A V E
1
PUB64G4, IA64G4, or
IA64G4_C5 Setups
IA64G4_T or IA32G4_T_C5
Setups
Choose your software support, and then follow the steps to connect your logic analyzer module to the preprocessor unit:
1. Match the A, D, C, and E LAI cables from the Master module with the
corresponding M_ D3/D2 and A3/A2, M_D1/D0 and A1/A0, M_C1/C0 and C3/C2, and M_E3/E2 and E1/E0 connector labels on the preprocessor unit. The LAI cable connector is keyed for correct alignment to the preprocessor unit. Connect the LAI cable to the preprocessor.
2. Match and connect the rest of the LAI cables between the Slave1 (S1),
Slave2 (S2), and Slave3 (S3) modules to the top of the preprocessor unit. Do not connect to RESERVED on the Slave3 module.
1. Match the A, D, C, and E LAI cables from the Master module with the
corresponding M_ D3/D2 and A3/A2, M_D1/D0 and A1/A0, M_C1/C0 and C3/C2, and M_E3/E2 and E1/E0 connector labels on the preprocessor unit. The LAI cable connector is keyed for correct alignment to the preprocessor unit. Connect the LAI cables to the top of the preprocessor.
2. Match and connect an LAI cable to the Slave3(S3) module and preprocessor.
Do not connect to RESERVED on the Slave3 module.
1--12
TMSSM3 700-Pin Socket Support
Getting Started
3. Match and connect the P6860 probes between the Slave1 and Slave2
modules and the top of the preprocessor unit.
CAUTION. To prevent damage to the P6860 probe and the preprocessor unit, you must always position the probes perpendicular to the footprint on the PCB. Incorrect handling of the P6860 probe while connecting to or disconnecting from the preprocessor unit can damage the probe.
NOTE. For more detailed information about how to attach a P6860 probe, refer to the P6860 probe manual references listed on page 1--2.
IA64G4 and IA64G4_SST
Setups
1. Match the A, D, C, and E LAI cables from the Master module with the
corresponding M_ D3/D2 and A3/A2, M_D1/D0 and A1/A0, M_C1/C0 and
C3/C2, and M_E3/E2 and E1/E0 connector labels on top of the preprocessor
unit. Connect the LAI cable.
2. Match and connect the LAI cables between the Slave1 (S1), Slave2 (S2), and
Slave3 (S3) modules on top of the preprocessor unit. Do not connect to
RESERVED on the Slave3 module.
3. Match and connect the P6860 probes between the Master and Slave1
modules in the expanded mainframe and the top of the preprocessor unit.

Applying and Removing Power

To apply power to the probe adapter and target system, follow these steps:
1. Check that the power switch on the preprocessor unit is in the off position. If
powered off, the zero (0) is visible on the power switch.
2. Plug the AC power cord into the IEC connector on the back of the preproces-
sor unit.
3. Plug the AC power cord into an electrical outlet that you know is working
properly.
4. Power on the probe adapter at the back of the preprocessor unit. A green,
5. Power on the target system.
To remove power from the target system and the probe adapter, reverse the preceding steps. You can skip step 1.
TMSSM3 700-Pin Socket Support
power-on LED lights on the front of the preprocessor unit, indicating that the
probe adapter is active.
1--13
Getting Started

Installing the Software

If you have not used a support product, you may want to refer to microprocessor support information in the Tektronix logic analyzer online help or the user manual.
NOTE. Before you install any software, verify that the microprocessor support software is compatible with the logic analyzer software (see page 1--1).
To install the software on your Tektronix logic analyzer, follow these steps:
1. Insert the disk in the disk drive.
2. Click the Windows Start button, point to Settings, and click Control Panel.
3. In the Control Panel window, double-click Add/Remove Programs.
4. Follow the instructions on the screen for installing the software from the
disk.
To remove or uninstall software, follow the above instructions except select Uninstall. You must close all windows before you uninstall any software.
1--14
TMSSM3 700-Pin Socket Support

Care and Maintenance

Getting Started
Before cleaning this product, read the following information:
CAUTION. To prevent static damage, you must handle components only in a static-free environment. Static discharge can damage the microprocessor, the probe adapter, the probes, and the module.
The probe adapter, consisting of the probe head and preprocessor unit, does not require scheduled or periodic maintenance. However, to keep good electrical contact and efficient heat dissipation, keep the probe adapter free of dirt, dust, and contaminants. When not in use, store the probe adapter in the original shipping bags and cardboard carton.
External Cleaning Only
Fuses
Clean dirt and dust with a soft bristle brush. For more extensive cleaning, use only a damp cloth moistened with deionized water; do not use any chemical cleaning agents.
WARNING. To prevent harm to yourself or damage to the preprocessor unit, do not open the preprocessor unit for cleaning and do not allow any moisture to get inside the preprocessor unit. There are no operator serviceable parts inside the preprocessor unit. Refer servicing of internal parts in the preprocessor unit to Tektronix authorized personnel only. External parts may be replaced by qualified service personnel.
All fuses in the probe adapter are not replaceable by the customer (operator). If the probe adapter is not functioning correctly, contact your Tektronix sales representative.
TMSSM3 700-Pin Socket Support
1--15
Getting Started
Short-Term Storage
Follow steps 1 through 4 for short-term storage of the probe head:
CAUTION. To prevent static damage to the microprocessor, the probe adapter, the probes, and the module, handle components only in a static-free environment.
Always wear a grounding wrist strap, heel strap, or similar device while handling the microprocessor and probe adapter.
1. Power off the target system, and unplug the AC power cord on the preproces-
sor unit. The power switch for the probe adapter is located on the back of the preprocessor unit. It is not necessary to power off the logic analyzer.
2. To remove the probe head, reverse the instructions in Figures 1--2 and 1--3 on
pages 1--6 and 1--7.
CAUTION. To prevent damage to the sensitive probe head cables, you must position the cables so that they are not pinched or contacting any sharp objects. When you fold the cables, use a minimum radius of 0.25 (0.64 cm) at the fold.
3. Using antistatic nongenerating tape, tape the pin-protector board onto the pin
header on the bottom of the probe head.
4. Store the probe head in an antistatic bag.
1--16
TMSSM3 700-Pin Socket Support
Getting Started
Long-Term Storage
For long-term storage, use the existing cardboard carton and packaging, and follow these steps:
1. Disconnect the preprocessor unit from the logic analyzer by removing the
probes and LAI cables from the top of the preprocessor unit.
2. Follow Figures 1--7 through 1 --12 to repackage the probe adapter.
Figure 1--7: Place the preprocessor unit and probe head in static-shielding bags.
Figure 1--8: Place foam in the cardboard carton
TMSSM3 700-Pin Socket Support
1--17
Getting Started
Place the foam end caps on both sides of the preprocessor unit.
Figure 1--9: Place the preprocessor unit in the cardboard carton
Figure 1--10: Place the cables carefully over the preprocessor unit
Figure 1--11: Place the foam and probe head in place
1--18
TMSSM3 700-Pin Socket Support
Figure 1--12: Place accessories in the accessory tray
3. Close and tape the cardboard carton.
To ship the probe adapter, refer to Shipping the Probe Adapter.
Getting Started

Shipping the Probe Adapter

To commercially transport the TMSSM3 probe adapter, package as follows:
1. Use the existing cardboard shipping carton and cushioning material to ship
the probe adapter.
If the existing shipping carton is not available, use a double-walled,
corrugated cardboard shipping carton that allows a 3 inch (7.62 cm)
minimum on all sides of the product.
2. Follow Figures 1--7 through 1 --12 on pages 1--17 through 1--19 to package
the probe head and preprocessor unit.
3. If you are shipping a probe adapter to a Tektronix service center for Warranty
service, attach a tag to the probe adapter showing the following:
H Owner’s name and address
H Name of a person who can be contacted
H Probe adapter type and serial number
H Description of the problem
TMSSM3 700-Pin Socket Support
1--19
Getting Started
1--20
TMSSM3 700-Pin Socket Support
Specifications

Specifications

Circuit Description

This section contains information regarding the s pecifications of the TMSSM3 hardware support.
The TMSSM3 probe adapter hardware uses a custom ASIC to preprocess all signals on the analyzer. The custom ASIC performs the following functions:
H Latches signals within a narrow valid window
H Demultiplexes double-pumped, source-synchronous signals
H Deterministically synchronizes source-synchronous signals to BCLK
700-pin microprocessor before being captured by the logic
Latched Operation
1x Common-Clock Signals
These 2x double-pumped signals include D[127:00]# and DEP[15:0]#. The ASIC latches these signals using their dedicated strobes: STBN[...]#, STBP[...]#. The ASIC then performs two-way demultiplexing on these signals and synchronizes the STB[7:0]# results to BCLK. The resulting signals are sent out by the ASIC four clock cycles after they were received on the system bus.
These signals include all of the remaining front-side bus signals. The ASIC latches these signals using the rising edge of BCLK. The ASIC then delays the signals so that the final results are sent out four clock cycles after they were received onto the system bus.
TMSSM3 700-Pin Socket Support
2--1
Specifications
Derived Signals
The TMSSM3 probe adapter hardware derives several custom signals from the front-side bus signals captured by the ASIC. These signals are used by the logic analyzer support software to provide clocking, transaction phase linking, and disassembly. Following is a description of custom signals:
PHASE_D. This signal can be used by the logic analyzer to store only bus cycles that contain active information. The PHASE_D signal is asserted when any of the following signals are asserted: ADS#, DRDY#, INIT#, RESET#, RS[2:0]#, and SNOOP_D.
SNOOP_D. This signal is asserted at the end of each successful snoop phase, taking into account any stall conditions that might have occurred. SNOOP_D is asserted only once per each request and is not asserted during a stall condition.
RCNT_D[2:0]. The request counter tracks the number of outstanding requests on the front-side bus. The request counter is incremented by each new request phase and decrememted by each response phase. The maximum allowable count is 8.
SCNT_D[2:0]. The snoop counter tracks the number of outstanding snoops on the front-side bus. The snoop counter is incremented by each new request phase and decrememted by each successful snoop phase. The maximum allowable count is
8.
Signal Probing
Bus Tracking Logic
Common Clock
TRACK_ERR_D. This signal is asserted whenever the request or snoop counters exceed their maximum value of 8, or a minimum value of 0. This signal is also asserted when ADS# has been observed active for two clock cycles in a row.
The
700-Pin Socket probe adapter uses passive series isolation to acquire data.
The
700-Pin Socket probe adapter uses a bus tracking PAL to aid the disassembly
software in linking various bus phases.
The TMSSM3 software allows disassembly from a data bus operating at up to 200 MHz. The setup and hold sample points are set to default timing numbers based on FSB specifications.
2--2
TMSSM3 700-Pin Socket Support

Probe Adapter Loading Diagrams

Specifications
700-pin socket microprossor
header1
sig1
pin_header
sig2
vss
out1
out2
C26
0.5 pF
+
C25
0.5 pF
sig1
Socket1
probe
mck_load
in
out1
probe
mck_load
in
zif_socket
sig2
+
+
R56 180
vss
TLSE9
a=2 L=57.5 m td=0.384 n z0=78
out2
+
C27
0.5 pF
TLSE10 a=1.5 L=1.219 td=5.97 n z0=75
TLSE16
a=2 L=50.8 m td=0.324 n z0=75
Tip resistor varies +--10%
TLSE18
a=2 L=50.8 td=0.324 n z0=75
R60 180
TLSE12
a=2 L=57.5 m td=0.384 n
+
z0=78
C30
+
TLSE17 a=1.5 L=1.219 td=5.97 n z0=75
0.5 pF
Logic Board
mictor mtd16
mictor_mtd
+
C28
0.5 pF
TLSE11
a=2 L=50.8 m td=0.324 n z0=75
mictor mtd17
mictor_mtd
+
C31
0.5 pF
TLSE15
a=2 L=50.8 m td=0.324 n z0=75
Figure 2--1: Pin header electrical load model for typical signals
+
V28
0.75
+
V31
0.75
+
+
R57 75
R61
75
C29
0.5 pF
ecl_buffer6
buffer
C32
0.5 pF ecl_buffer7
buffer
TMSSM3 700-Pin Socket Support
2--3
Specifications
TF2_1
K12= 0.082 L1=1.685 n L2=1.685 n
Signal In 1
Signal In 2
Vss
+
C1
0.44 pF
+
C2
0.44 pF
+
Signal out 1
Signal out 2
C3
0.44 pF
Vss
Figure 2--2: Pin header electrical load model for typical signals
TF2_1
K12= 0.318 L1=5.20 n L2=5.20 n
Signal In 1
Signal In 2
Vss
+
C1
0.66 pF
+
C2
1.3 pF
+
Signal out 1
Signal out 2
C3
1.3 pF
Vss
Figure 2--3: ZIF electrical load model for typical signals
Signal In 1
R1
10 m
C4
0.5 pF
3.2 n
C3
0.7 pF
L1
Signal Out 2
C2
0.5 pF
Figure 2--4: Mated Mictor model
BCLKP
BCLKN
To TLSE11 (see Figure 2--1)
To TLSE15 (see Figure 2--1)
75
0.8 V
75
0.8 V
0.5 p
0.5 pF
Figure 2--5: BCLK model
buffer
2--4
TMSSM3 700-Pin Socket Support

Specification Tables

Specifications
These specifications are for a probe adapter connected between a compatible Tektronix logic analyzer and a target system. Signal voltage swing in your target system must be at least 600 mV
around the GTL+ reference voltage.
p-p
Table 2--1 lists the electrical requirements of the target system. Table 2--2 lists the electrical requirements for the power supply that provides power to the
700-Pin Socket probe adapter. Table 2--3 lists the BCLK timing and electrical
specifications. Table 2--4 lists the environmental specifications.
Table 2--1: Electrical specifications for the target system
Characteristics Requirements
Target system DC power requirements
Voltage, V
Current, VREF6 I maximum <1 mA, I typical <1 mA
Target system clock rate Maximum 333 MHz and 400 MHz
Target system tested clock rate Maximum 200 MHz
Common clock capture
T
su
T
hd
Source Synchronous capture
T
su
T
hd
TT
1.2 V 5%
0.4 ns
0.1 ns
0.3 ns
0.3 ns
Table 2--2: Electrical specifications for the AC input
Characteristic Description
Input Voltage rating 100 -- 240 VAC CAT II
Input Frequency Rating 50 -- 60 Hz
Input Current Rating 6.0 A maximum
TMSSM3 700-Pin Socket Support
Table 2--3: BCLK timing and electrical specifications
Characteristics Minimum Maximum Units Notes
Vin(lo) 0.4 V
Vin(hi) 1.0 V
2--5
Specifications
Table 2--3: BCLK timing and electrical specifications (cont.)
Characteristics NotesUnitsMaximumMinimum
Duty Cycle 25 75 %
t
lh
t
hl
1.25 ns Monotonically increasing
1.25 ns Monotonically decreasing
Table 2--4: Environmental specifications
Characteristic
1
Description
Temperature
Maximum operating +50 °C (+122 °F)
2
Minimum operating 0 °C (+32 °F)
Nonoperating -- 5 5 °Cto+75°C(--67° to +167 °F)
Humidity 10 to 95% relative humidity, noncondensing
Altitude
Operating 3 km (10,000 ft) maximum
Nonoperating 15 km (50,000 ft) maximum
Electrostatic immunity The probe adapter is static sensitive
Required airflow clearances for the preprocessor unit
Front, top, left side 5.08 cm (2 in)
Back 7.62 cm (3 in)
Bottom, right side 0.635 cm (0.250 in)
1
Designed to meet Tektronix standard 062-2847-00 class 5.
2
Not to exceed microprocessor thermal considerations. Customer supplied cooling might be required across the CPU.
Table 2--5: Certifications and compliances
EC Declaration of Conformity -­Low Voltage
Compliance was demonstrated to the following specification as listed in the Official Journal of the European Communities:
Low Voltage Directive 73/23/EEC, Amended by 93/68/EEC
EN 61010-1:1993/A2:1995 Safety requirements for electrical equipment for measurement,
2--6
control, and laboratory use
TMSSM3 700-Pin Socket Support
Specifications
Table 2--5: Certifications and compliances (cont.)
Approvals UL3111-1 -- Standard for electrical measuring and test equipment
CAN/CSA C22.2 No. 1010.1--92 -- Safety requirements for measurement, control, and laboratory use.
Installation Category Descriptions Terminals on this product may have different installation category designations. The installation
categories are:
CAT III Distribution-level mains (usually permanently connected). Equipment at this level is
typically in a fixed industrial location
CAT II Local-level mains (wall sockets). Equipment at this level includes appliances, portable
tools, and similar products. Equipment is usually cord-connected
CAT I Secondary (signal level) or battery operated circuits of electronic equipment
71.12 mm (2.800 in)
12.70 mm (0.500 in)
Dimensions
Figure 2--6 shows the dimensions of the TMSSM3 probe head.
190.50 mm (7.500 in)
Figure 2--6: Dimensions of the probe head
TMSSM3 700-Pin Socket Support
2--7
Specifications
Figure 2--7 shows the dimensions of the preprocessor unit. For required clearances around the preprocessor unit, refer to Table 2--4 on page 2--6).
474.32 mm (18.674 in)
425.45 mm (16.750 in)
420.37 mm (16.550 in)
139.70 mm (5.500 in)
Figure 2--7: Dimensions of the preprocessor unit
160.27 mm (6.310 in)
2--8
TMSSM3 700-Pin Socket Support
Reference

Reference Tables

The TMSSM3 software disk contains the following reference tables. Viewing a table is not a requirement when preparing the module for use. You can view the reference tables without connecting the TLA to your target system. These tables are provided to assist you when you are debugging a target system.
H Channel and Group Definition Tables — The TMSSM3 software
automatically defines channel groups. These groups define the channel assignments and the groups for a software support setup. If you want to know what the group definition is for a channel group or which signal is in which channel group, refer to these tables.
H Symbol Tables — Symbol files are files of alphanumeric symbol names and
associated data values. Symbol files are used to map a channel group value to a text string and/or color. For example, a symbol file could map an address group value to a function name. This is faster and easier to interpret than a numeric value.
NOTE. The Color symbol table is not associated with any channel groups.

Viewing Reference Tables

You must load your support and merge the modules before completing the following steps.
To view a symbol table, follow these steps:
1. In the System window under System, select Symbol to display the Symbol
2. Select Load to display the Load Symbol File window.
3. Select and right click one of the .tsf files to display the Select menu.
4. Select Open and select the appropriate editor, (Word Pad for example) to
window.
view the symbol table.
TMSSM3 700-Pin Socket Support
3- 1
Reference Tables
To Extract Groups
To extract a channel group or a group definition table, follow these steps:
1. Open a System window.
2. Select the Setup icon.
3. In the Setup window, open the Window menu and select Setup: <support file
name>.
4. In the Setup window under Group Name, s elect a group.
5. Open File, select Export Channel Setup to display a Channel Setup Export
window. Notice the Export Path name to locate your exported file later.
In the Channel Setup Export window under Export S elections, check that the Export Channel Information and the Export Group Information boxes are checked.
The Field delimiter box controls how the tables are displayed. See the Tektronix logic analyzer online help for more information.
6. Select Export.
3- 2
TMSSM3 700-Pin Socket Support
Replaceable Parts List

Replaceable Parts List

This section contains a list of the replaceable components or modules for the TMSSM3 Hardware Support package.

Parts Ordering Information

Replacement parts are available through your local Tektronix field office or representative.
Changes to Tektronix products are sometimes made to accommodate improved components as they become available and to give you the benefit of the latest improvements. Therefore, when ordering parts, it is important to include the following information in your order:
H Part number
H Instrument type or model number
H Instrument serial number
H Instrument modification number, if applicable
Module Servicing
If you order a part that has been replaced with a different or improved part, your local Tektronix field office or representative will contact you concerning any change in part number.
Modules can be serviced by selecting one of the following three options. Contact your local Tektronix service center or representative for repair assistance.
Module Exchange. In some cases you may exchange your module for a reman­ufactured module. These modules cost significantly less than new modules and meet the same factory specifications. For more information about the module exchange program, call 1-800-833-9200. Outside North America, contact a Tektronix sales office or distributor; see the Tektronix Web site for a list of offices: www.tektronix.com.
Module Repair and Return. You may ship your module to us for repair, after which we will return it to you.
New Modules. You may purchase replacement modules in the same way as other replacement parts.
TMSSM3 700-Pin Socket Support
4--1
Replaceable Parts List

Using the Replaceable Parts List

This section contains a list of the mechanical and/or electrical components that are replaceable for the TMSSM3 Hardware Support package. Use this list to identify and order replacement parts. The following table describes each column in the parts list.
Parts list column descriptions
Column Column name Description
1 Figure & index number Items in this section are referenced by figure and index numbers to the exploded view illustrations
that follow.
2 Tektronix part number Use this part number when ordering replacement parts from Tektronix.
3 and 4 Serial number Column three indicates the serial number at which the part was first effective. Column four
indicates the serial number at which the part was discontinued. No entry indicates the part is good for all serial numbers.
5 Qty This indicates the quantity of parts used.
6 Name & description An item name is separated from the description by a colon (:). Because of space limitations, an
item name may sometimes appear as incomplete. Use the U.S. Federal Catalog handbook H6-1 for further item name identification.
7 Mfr. code This indicates the code of the actual manufacturer of the part.
8 Mfr. part number This indicates the actual manufacturer’s or vendor’s part number.
Abbreviations
Mfr. Code to Manufacturer
Cross Index
Abbreviations conform to American National Standard ANSI Y1.1--1972.
The table titled Manufacturers Cross Index shows codes, names, and addresses of manufacturers or vendors of components listed in the parts list.
4--2
TMSSM3 700-Pin Socket Support
Replaceable Parts List
Manufacturers cross index
Mfr. code
060D9 TENSOLITE COMPANY PRECISION HARNESS AND ASSEMBLY,3000
0B445 ELECTRI--CORD MFG CO INC 312 EAST MAIN STREET WESTFIELD, PA 16950
0KB01 STAUFFER SUPPLY CO 810 SE SHERMAN PORTLAND, OR 97214--4657
S3109 FELLER U.S. CORPORATION 10B VAN DYKE AVENUE NEW BRUNSWICK, NJ 08901
TK1373 PATELEC--CEM 10156 TORINO,VAICENTALLO,62/456 ITALY
TK2548 XEROX CORPORATION 7431 EVERGREEN PARKWAY HILLSBORO, OR 97124
TK6121 TUMBLER CORP 4241 BUSINESS CENTER DR FREMONT, CA 94538
Manufacturer Address City, state, zip code
VANCOUVER, WA 98661
COLUMBIA HOUSE BLVD,#120
TMSSM3 700-Pin Socket Support
4--3
Replaceable Parts List
Replaceable parts list
Fig. & index number
4--1--1 672--1613--50 1 CIRCUIT BOARD, W/PGA SOCKET AND CABLES;TMSSM3 80009 672--1613--50
Tektronix part number
--2 211--1187--00 12 SCREW; M3.5 X 8MM,FLT,PHL,STL 18--8;.3C8MXFS 0KB01 .3C8MXFS
071--1387--XX 1 MANUAL,TECH; INSTRUCTION, HARDWARE;TMSSC2,DP TK2548 071--1387--XX
161--0104--00 1 CA ASSY,PWR:3,18 AWG,98 L,250V/10AMP,98
012--1661--50 7 CA ASSY; INTCON,SHLD RIBBON,MCP,72 L 060D9 012--1661--50
016--1947--00 CASE,STORAGE; PLASTIC,W/FOAM,12 1/4 X 8 13/16 X 3
065--0701--00 SHIPPING KIT;TMSSM3 ------ 065--0701--00
--3 020--2550--00 1 ACCESSORY KIT,BLUE;RETENTION BRACKET #4 80009 020--2550--00
--4 020--2551--00 1 ACCESSORY KIT,RED;RETENTION BRACKET #5 80009 020--2551--00
--5 020--2558--00 1 ACCESSORY KIT,GREEN;RETENTION BRACKET #1 80009 020--2558--00
--------------------* 4--5 P6860 PROBE 80009 ORDER BY
161--0104--05 1 CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5
161--0104--06 1 CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5
161--0104--07 1 CA ASSY,PWR:3,1.0MM SQ,240V/10A,2.5
161--0167--00 1 CA ASSY,PWR:3,0.75MM SQ,250V/10A,2.5
Serial no. effective
Serial no. discont’d
Qty Name & description
STANDARD ACCESSORIES
INCH,RTANG,IEC320,RCPT X STR,NEMA 15--5P,W/CORD GRIP
7/8;MONTE
OPTIONAL ACCESSORIES
METER,RTANG,IEC320,RCPT,AUSTRALIA,SAFTEY CONTROLLED
METER,RTANG,IEC320,RCPT,EUROPEAN,SAFTEY CONTROLLED
METER,RTANG,IEC320,RCPT X 13A,FUSED,UK PLUG,(13A FUSE),UK PLUG,(13A FUSE),UNITED KINGDOM,SAFTEY CONTROL
METER,RTANG,IEC320,RCPT,SWISS,NO CORD GRIP,SAFTEY CONTR
Mfr. code
TK6121 ORDER BY DE-
------ 016--1947--00
TK1373 ORDER BY
TK1373 ORDER BY
TK2541 ORDER BY
S3109 ORDER BY
Mfr. part number
SCRIPTION
DESCRIPTION
DESCRIPTION
DESCRIPTION
DESCRIPTION
DESCRIPTION
4--4
TMSSM3 700-Pin Socket Support
Replaceable Parts List
3 4 5
1
2
Figure 4--1: Probe adapter exploded view
TMSSM3 700-Pin Socket Support
4--5
Replaceable Parts List
(Return the preproces­sor unit to the factory if service is required)
4--6
Figure 4--2: Preprocessor unit exploded view
TMSSM3 700-Pin Socket Support
Index

Index

A
About this manual set, ix AC adapter, 2--5 Address, Tektronix, x Address bus, 1--3 Airflow clearance, 2--6 Application, logic analyzer configuration, 1--1
B
BINIT OBS jumper, 1--4
C
Care and maintenance
cleaning, 1--15
external, 1--15 Cetifications and compliance, 2--6 Channel assignment tables, 3--1 Clock rate, 1--3 Connections
LAI Cable, 1--2
P6860, 1-- 2 Contacting Tektronix, x Cooling requirements, 1--2
clock rate, 2--5 power requirements, 2--5 setup time, 2--5 tested clock rate, 2--5
Environmental specifications, 2--6
airflow clearance, 2--6 altitude, 2--6 electrostatic immunity, 2--6 humidity, 2--6 temperature, 2--6
F
Forced, air cooling, 1--2 Fuses, maintenance, 1--15
G
Group definitions, 3--1
I
Installing support software, 1--14
J
D
Data bus, 1--3 Default channel mapping
LAI Cable labels, 1--2
probe labels, 1--2 Definitions
information on basic operations, ix
Master module, ix
Slave module, ix Dimensions, probe adapter, 2--8 Disabling the cache, 1--3 Disassembler
logic analyzer configuration, 1--1
setup, 1--14
E
Electrical specifications, 2--1, 2--5
AC adapter, 2--5
BCKL timing, 2--5
TMSSM3 700-Pin Socket Support
Jumpers, observed, 1--4
L
LAI Cable
configuration, 1--11
connections, 1--2 LAI Cable labels, default channel mapping, 1--2 Loading diagrams, 2--3 Logic analyzer
configuration for disassembler, 1--1
configuration for the application, 1--1
connecting to the target system, 1--5
software compatibility, 1--1
M
Maintenance, fuses, 1--15 Manual
conventions, ix
how to use the set, ix
Index--1
Index
Master module, definition, ix Module connections, 1--2
O
Options, 1--2
P
P6860, 1-- 2 P6860 probe, configuration, 1--11 Phone number, Tektronix, x Power, applying, 1--13 Probe adapter
configuring, 1--4 hardware description, 1--1 jumper positions, 1--4
software compatibility, 1--1 Probe labels, default channel mapping, 1--2 Product support, contact information, x
R
Reference tables
symbol, channel group, group definition, 3--1
viewing, 3--1 Removing hardware, from target system, 1--8, 1--16 Requirements, forced air cooling, 1--2 Reset, 1--3 Restrictions, 1--2
address bus, 1--3
clock rate, 1--3
data bus, 1--3
disabling the cache, 1--3
S
Service support, contact information, x Setups
disassembler, 1--14
support, 1--14 Shipping, probe adapter, assembled, 1--19 Slave module, definition, ix Software compatibility, 2--1 Specifications, 2--1
electrical, 2--1, 2--5
environmental, 2--6
mechanical (dimensions), 2--8 Storage, long--term, 1--17 Support, setup, 1--14 Symbol tables, 3--1
T
Technical support, contact information, x Tektronix, contacting, x Terminology, ix TLA7AX, probes, measurements, 1--2 TLA7AX module, measurements, 1--2
U
URL, Tektronix, x
W
Web site address, Tektronix, x
Index--2
TMSSM3 700-Pin Socket Support