This document applies to TLA System Software
version 5.1.
Warning
These servicing instructions are for use by
qualified personnel only. To avoid personal injury,
do not perform any servicing unless you are
qualified to do so. Refer to all safety summaries
prior to performing service.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price c hange privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
HIn North America, call 1-800-833-9200.
HWorldwide, visit www.tektronix.com to find contacts in your area.
Warranty 6
Tektronix warrants that the parts and modules (“parts”) that it manufactures and sells will be free from defects in
materials and workmanship for a period of three (3) months from the date of shipment. If any such part proves
defective during this warranty period, Tektronix, at its option, either will repair the defective part without charge,
or will provide a replacement in exchange for the defective part. Parts and modules used by Tektronix for
warranty work may be new or reconditioned to like new performance. All replaced parts and m odules become the
property of Tektronix.
In order to obtain service under this warranty, Custom er must notify Tektronix of the defect before the expiration
of the warranty period and make suitable arrangements for the performance of service. Customer shall be
responsible for packaging and shipping the defective part to the service center designated by Tektronix, with
shipping charges prepaid. Tektronix shall pay for the return of the part to Customer if the shipment is to a location
within the country in which the Tektronix service center is located. Customer shall be responsible for payi ng all
shipping charges, duties, taxes, and any other charges for parts returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage
resulting from attempts by personnel other than Tektronix representatives to install, repair or service the part; b) to
repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or
malfunction caused by the use of non-Tektronix supplies; or d) to service a part that has been modified or
integrated with other products or parts when the effect of such modification or integration increases the time or
difficulty of servicing the part.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PART IN LIEU OF ANY OTHER
WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’
RESPONSIBILITY TO REP AIR OR REPLACE DEFECTIVE PARTS IS THE SOLE AND EXCLUSIVE
REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS W ARRANTY. TEKTRONIX AND ITS
VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL
DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADV ANCE NOTICE OF
THE POSSIBILITY OF SUCH DAMAGES.
Warranty 9(b)
Tektronix warrants that the media on which this software product is furnished and the encoding of the programs on
the media will be free from defects in materials and workmanship for a period of three (3) months from the date of
shipment. If any such medium or encoding proves defective during the warranty period, Tektronix will provide a
replacement in exchange for the defective medium. Except as to the media on which this software product is
furnished, this software product is provided “as is” without warranty of any kind, either express or implied.
Tektronix does not warrant that the functions contained in this software product will meet Customer’s
requirements or that the operation of the programs will be uninterrupted or error-free.
In order to obtain service under this warranty, Custom er must notify Tektronix of the defect before the expiration
of the warranty period. If Tektronix is unable to provide a replacement that is free from defects in materials and
workmanship within a reasonable time thereafter, Customer may terminate the license for this software product
and return this software product and any associated materials for credit or refund.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY
OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
TEKTRONIX’ RESPONSIBILITY TO REPLACE DEFECTIVE MEDIA OR REFUND CUSTOMER’S
PA YMENT IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF
THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT,
SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX
OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Table of Contents
TLA5KUP Option 10
General Safety Summaryv...................................
Service Safety Summaryvii....................................
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it.
To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system.
Read the safety sections of the other component manuals for warnings and
cautions related to operating the system.
ToAvoidFireor
Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and
certified for the country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
The inputs are not rated for connection to mains or Category II, III, or IV
circuits.
Connect the probe reference lead to earth ground only.
Power Disconnect. The power cord disconnects the product from the power
source. Do not block the power cord; it must remain accessible to the user at all
times.
Do Not Operate Without Covers. Do not operate this product with covers or panels
removed.
Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
v
General Safety Summary
Provide Proper Ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
Terms in this Manual
Symbols and Terms
on the Product
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
These terms may appear on the product:
HDANGER indicates an injury hazard immediately accessible as you read the
marking.
HWARNING indicates an injury hazard not immediately accessible as you
read the marking.
HCAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
CAUTION
Refer to Manual
Mains Disconnected
OFF (Power)
vi
WARNING
High Voltage
Mains Connected
ON (Power)
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
Protective Ground
(Earth) Terminal
Standby
Earth Terminal
Chassis Ground
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, disconnect the main power by means
of the power cord or, if provided, the power switch.
Use Care When Servicing With Power On. Dangerous voltages or currents may
exist in this product. Disconnect power, remove battery (if applicable), and
disconnect test leads before removing protective panels, soldering, or replacing
components.
To avoid electric shock, do not touch exposed connections.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
vii
Service Safety Summary
viii
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
Preface
This instruction manual contains specific information about the TLA5KUP logic
analyzer field upgrades. The logic analyzer field upgrades consist of software
and hardware options for the TLA5000B and TLA5000 Series Logic Analyzers.
NOTE. Unless otherwise noted all references to TLA5000 Series Logic Analyzers
apply to both TLA5000B and TLA5000 instruments.
Products Eligible for TLA5KUP Upgrades
Table i lists all of the TLA5000 Series Logic Analyzers eligible for upgrades via
the TLA5KUP field upgrade kits. Review Table i to determine if your logic
analyzer is eligible for an upgrade and then use Table ii on page x, to explore
possible upgrade paths.
Table i: TLA5000 Series logic analyzers eligible for upgrades
The TLA5KUP upgrade kits are designed for “One TLA5KUP upgrade kit per
TLA5000 mainframe.” If you want to upgrade multiple mainframes, you must
order one TLA5KUP upgrade kit with the appropriate options for each mainframe.
The latest TLA Application Software has been tested and can be installed and
will be supported by Tektronix only on the Tektronix-supplied version of
Microsoft Windows XP Professional.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
ix
Preface
upgradingtothelatestTLAAp
pgy
L
ASW
V
W
L
A
h
igh
TLA5000StdwithTLASWV4.4
Windows2000
TLA5KUPOption41
p
f
datafromTDSoscilloscope
s
f
SoftwareandMicrosoftWindows
Table ii: TLA5KUP Logic Analyzer Field Upgrade Matrix
New capability desiredCurrent configuration and
TLA software version
Add New Software Features
Add new functionality to existing
TLA5000 logic analyzers by
plication Software.
Microsoft Windows XP Professional operating system is recommendedfor TLASWV5.0or
er.
Add more DRAM
Increase controller memory from
512 MB to 1 GB.
Add more DRAM
Increase controller memory from
512 MB to 1 GB.
Add DVD-ROM/CD-RW Drive
Add DVD-ROM/CD-RW Drive to
back up or transfer user and data
files.
Add New iView External Oscilloscope Capability
Add new iView capability to view
data
rom TDS oscilloscopes
correlated directly on the TLA5000
logic analyzer
Enhance your current TLA5000
logic analyzer by increasing state
speed or record length. Run the
PowerFlex Utility to inform you
what upgrades are available.
Instructions to obtain an upgrade
are provided.
1
.
TLA5000 Std with TLA SW V5.1Microsoft
-
TLA5000 Std with TLA SW V5.0Microsoft
TLA5000 Std with T
or V4.3
Any TLA5000 mainframe--TLA5KUP Option 10
Any TLA5000B mainframe--TLA5KUP Option 11
TLA5000 Std with TLA SW V5.1 or
V5.0
TLA5000 Std with TLA SW V4.4
or V4.3
TLA5000 Std with TLA SW V5.1Microsoft
TLA5000 Std with TLA SW V5.0
TLA5000 Std with TLA SW V4.4
or V4.3
Any TLA5000 series logic
analyzer with TLA SW V4.3 and
above
Add Wheeled Transport CaseAny TLA5000 mainframe--016-1522-xx
1
For a list of supported TDS oscilloscopes, please visit our website at www.tektronix.com/la.
Current configuration and
TLA software version
operating
system
(Same as used with TDS6000 and
TDS7000 instruments)
Service and Upgrade Notes
To prevent personal injury or damage to the instrument, consider the following
requirements before attempting service:
HRead the General Safety Summary and Service Safety Summary found at the
beginning of this manual.
HOnly qualified service personnel should perform the procedures in this
manual.
Be sure to follow all warnings, cautions and notes.
Adjustment and Certification Interval
Generally, you should perform the adjustments and performance verification
procedures described in the TLA5000 Logic Analyzer Series Service Manual
(Tektronix part number, 071-1305-xx) once per year, or following repairs that
may affect adjustment or calibration.
Service Offerings
Tektronix provides service to cover repair under warranty as well as other
services that are designed to meet your specific service needs.
Whether providing warranty repair service or any of the other services listed
below, Tektronix service technicians are equipped to service your logic analyzer.
Services are provided at Tektronix Services Centers.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
xi
Preface
Warranty Repair Service
Calibration and Repair
Service
The warranty for this product is located behind the title page in this manual.
Tektronix technicians provide warranty service at most Tektronix service
locations worldwide. The Tektronix product catalog lists all service locations
worldwide, or you can visit us on our Customer Services World Center Web site
at:
Tektronix.com/Measurement/Service
In addition to warranty repair, Tektronix Service offers calibration and other
services which provide solutions to your service needs and quality standards
compliance requirements.
The following services can be tailored to fit your requirements for calibration
and/or repair of your logic analyzer.
Service Options. Tektronix service options can be selected at the time you
purchase your instrument. You select these options to provide the services that
best meet your service needs.
Service Agreements. If service options are not added to the instrument purchase,
then service agreements are available on an annual basis to provide calibration
services or post-warranty repair coverage. Service agreements may be customized to meet special turn-around time and/or on-site requirements.
Service on Demand. Tektronix offers calibration and repair services on a
“per-incident” basis that is available with standard prices.
Self Service. Tektronix supports repair to the replaceable-part level by providing
for circuit board exchange.
Use this service to reduce down-time for repair by exchanging circuit boards for
remanufactured ones. Tektronix ships updated and tested exchange boards. Each
board comes with a 90-day service warranty.
For More Information. Contact your local Tektronix service center or sales
engineer for more information on any of the Calibration and Repair Services just
described.
xii
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
TLA5KUP Option 10
TLA5KUP Option 10:Memory Upgrade Installation
TLA5KUP option 10 provides the capability of upgrading the memory on your
TLA5000 Series Logic Analyzer to 1 GB.
NOTE. These instructions apply to TLA5000 Series Logic Analyzers only. To
upgrade the memory on the TLA5000B Series Logic Analyzers, please order
TLA5KUP option 11.
Instruments
TLA5000 Series Logic AnalyzersAll Serial Numbers
Parts List
Table 1--1: TLA5KUP Option 10 parts list
QuantityPart numberDescription
1ea156-9442-xx512 MB DIMM
1eaN/ATLA5KUP kit label
Minimum Tool and Equipment List
Table 1--2: Tools required for installing upgrade
Item
no.
1Screwdriver handle
2T-15 Torx tipUsed for removing most the instru-
NameDescription
Accepts TorxR-driver bits
ment’s screws. TorxR-driver bit for
T-15 size screw heads
General tool
number
620-440
640-247
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
1--1
TLA5KUP Option 10: TLA5000 Memory Upgrade
Installation Prerequisites
These instructions assume that you are familiar with servicing the instrument. If
you need further details for disassembling or reassembling the instrument, refer
to the TLA5000 Logic Analyzer Series Service Manual (071-1305-xx). You may
also contact your nearest Tektronix, Inc., Service Center or Tektronix Factory
Service Center for installation assistance.
Be sure to observe the following precautions to avoid damaging the logic
analyzer while preparing to service it.
CAUTION. Many components within the instrument are susceptible to static-discharge damage.
Service only in a static-free environment. Observe standard handling precautions
for static-sensitive devices.
Always wear a grounded wrist strap, grounded foot strap, and static-resistant
apparel while installing this kit.
WARNING. Dangerous voltages may be present. Before performing any procedure
in this subsection, disconnect the power cord from the line voltage source.
Failure to do so could cause serious injury or death.
HDo not handle static-sensitive components on boards.
HTransport and store static-sensitive boards in their original containers or on
conductive foam. Label any package that contains static-sensitive assemblies.
HWear anti-static wrist strap while handling the boards to discharge the static
voltage from your body.
HDo not allow anything capable of holding or generating a static charge on
the work surface.
HAvoid handling boards in areas that have a floor or work surface cover that
is capable of generating a static charge.
Remove the Accessories Pouch
Remove the accessories pouch to access the covers on the instrument.
1--2
1. Turn the logic analyzer off.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
2. Remove the power cord and all probes.
3. Open the pouch and locate the two snaps on the inside front side of the
pouch.
4. Gently pull on each of the two tabs to unsnap the front of the pouch.
5. At the rear of the instrument, peel off the Velcro that holds the pouch to the
instrument.
6. Set the pouch aside.
7. To reinstall the pouch, perform these steps in reverse order. Tighten the T-15
Torx-drive screws to 8-in lbs.
Remove the Trim and Covers
Remove the trim and covers as shown in Figure 1--1 on page 1--4, by following
this procedure.
TLA5KUP Option 10: TLA5000 Memory Upgrade
1. Remove the top cover trim.
a. Remove the two T-15 Torx-drive screws that secure the top cover trim to
the instrument.
b. Remove the snap studs from the top cover.
c. Slide the trim panel toward the rear of the instrument allowing the tabs to
clear the cover openings, then pull out to remove the panel from the
instrument (you may need to loosen the two top black feet on the back of
the instrument if the cover does not come off easily).
2. Remove the right side trim.
NOTE. When you remove the CD Drive trim, notice the notches in the trim that
provide the clearance for the chassis screws located under the trim. When you
replace the trim, make sure that you orient the trim with the notches in the
correct location. Otherwise the trim may not fit properly.
a. Remove the Drive trim by inserting a flat blade screwdriver in the
bottom slot of the trim and gently prying the trim piece out of the side
panel. Pull the trim up and out from the instrument.
b. Remove the two T-15 Torx-drive screws that secure the right side trim to
the bottom of the instrument.
c. Slide the trim panel toward the rear of the instrument allowing the tabs to
clear the cover openings, then pull the panel from the instrument.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
1--3
TLA5KUP Option 10: TLA5000 Memory Upgrade
Accessory
pouch
Top cover trim
Figure 1--1: Trim and covers
Right side
panel
CD-drive trim
T-15 Torx-drive
screws (2)
1--4
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
Remove the Right-Side Cover
Remove the right-side cover to access most of the internal components of the
instrument. Refer to Figure 1--2 while removing the right-side cover.
NOTE. All mounting screw holes are indicated by a star etched around the
mounting hole.
1. Remove the 15 T-15 Torx-drive screws that secure the cover to the top and
right sides of the chassis.
2. Remove the cover.
CAUTION. Take care not to bind or snag the covers on the internal cabling as you
remove or install.
TLA5KUP Option 10: TLA5000 Memory Upgrade
Figure 1--2: Right-side cover removal
Tab
Right side cover
T-15 Torx-drive
screws (14)
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
1--5
TLA5KUP Option 10: TLA5000 Memory Upgrade
Adding Memory
The memory sockets are located on the processor board mounted to the back
panel (refer to Figure 1--3).
1. Install the memory module as follows:
a. For TLA5000 instruments with serial numbers B010000 to B019999,
install the memory module in the single, spare socket.
b. For TLA5000 instruments with serial numbers B020000 or higher, install
the memory module in one of the spare sockets. For optimum performance, leave one socket empty between the original and new DIMMs
(the correct socket for the new memory module is labeled Channel B
DIMM 0).
Memory module (s)
Battery
Verifying Operation
1--6
Figure 1--3: Memory module and battery location
2. Reinstall the trim, covers, and the accessories pouch.
To verify the proper operation of the logic analyzer, follow these steps:
1. Plug the power cord in and turn the logic analyzer on.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
TLA5KUP Option 10: TLA5000 Memory Upgrade
2. Verify that the instrument passes all power-up diagnostics.
3. Exit all applications and close any open windows.
4. Click Start in the Windows tool bar.
5. Select Programs from the Start menu.
6. Select the CheckIt Utilities application from the Programs menu.
7. Run the appropriate verification tests from the application.
8. Turn the logic analyzer off, and then on again before running any application
software including the TLA application software.
Attach the Upgrade Kit Label to the Instrument
After completing all the previous steps, you need to install the upgrade kit label
on the instrument to indicate that the kit is installed.
Locate the big label on the rear of the instrument, and then attach the TLA5KUP
kit label as shown in Figure 1--4.
Place revision
label here.
Kit label
Figure 1--4: Kit label locations
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
1--7
TLA5KUP Option 10: TLA5000 Memory Upgrade
1--8
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
TLA5KUP Option 11
TLA5KUP Option 11:Memory Upgrade Installation
TLA5KUP option 11 provides the capability of upgrading the memory on your
TLA5000B Series Logic Analyzer to 1 GB.
NOTE. These instructions apply to TLA5000B Series Logic Analyzers only. To
upgrade the memory on the TLA5000 Series Logic Analyzers, please order
TLA5KUP option 10.
Instruments
TLA5000B Series Logic AnalyzersAll Serial Numbers
Parts List
Table 2--1: TLA5KUP Option 11 parts list
QuantityPart numberDescription
1ea167-0428-xx512 MB DIMM
1eaN/ATLA5KUP kit label
Minimum Tool and Equipment List
Table 2--2: Tools required for installing upgrade
Item
no.
1Screwdriver handle
2T-15 Torx tipUsed for removing most the instru-
NameDescription
Accepts TorxR-driver bits
ment’s screws. TorxR-driver bit for
T-15 size screw heads
General tool
number
620-440
640-247
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
2--1
TLA5KUP Option 11: TLA5000B Memory Upgrade
Installation Prerequisites
These instructions assume that you are familiar with servicing the instrument. If
you need further details for disassembling or reassembling the instrument, refer
to the TLA5000 Logic Analyzer Series Service Manual (071-1305-xx). You may
also contact your nearest Tektronix, Inc., Service Center or Tektronix Factory
Service Center for installation assistance.
Be sure to observe the following precautions to avoid damaging the logic
analyzer while preparing to service it.
CAUTION. Many components within the instrument are susceptible to static-discharge damage.
Service only in a static-free environment. Observe standard handling precautions
for static-sensitive devices.
Always wear a grounded wrist strap, grounded foot strap, and static-resistant
apparel while installing this kit.
WARNING. Dangerous voltages may be present. Before performing any procedure
in this subsection, disconnect the power cord from the line voltage source.
Failure to do so could cause serious injury or death.
HDo not handle static-sensitive components on boards.
HTransport and store static-sensitive boards in their original containers or on
conductive foam. Label any package that contains static-sensitive assemblies.
HWear anti-static wrist strap while handling the boards to discharge the static
voltage from your body.
HDo not allow anything capable of holding or generating a static charge on
the work surface.
HAvoid handling boards in areas that have a floor or work surface cover that
is capable of generating a static charge.
Remove the Accessories Pouch
Remove the accessories pouch to access the covers on the instrument.
2--2
1. Turn the logic analyzer off.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
2. Remove the power cord and all probes.
3. Open the pouch and locate the two snaps on the inside front side of the
pouch.
4. Gently pull on each of the two tabs to unsnap the front of the pouch.
5. At the rear of the instrument, peel off the Velcro that holds the pouch to the
instrument.
6. Set the pouch aside.
7. To reinstall the pouch, perform these steps in reverse order. Tighten the T-15
Torx-drive screws to 8-in lbs.
Remove the Trim and Covers
Remove the trim and covers as shown in Figure 2--1 on page 2--4, by following
this procedure.
TLA5KUP Option 11: TLA5000B Memory Upgrade
1. Remove the top cover trim.
a. Remove the two T-15 Torx-drive screws that secure the top cover trim to
the instrument.
b. Remove the snap studs from the top cover.
c. Slide the trim panel toward the rear of the instrument allowing the tabs to
clear the cover openings, then pull out to remove the panel from the
instrument (you may need to loosen the two top black feet on the back of
the instrument if the cover does not come off easily).
2. Remove the right side trim.
NOTE. When you remove the CD Drive trim, notice the notches in the trim that
provide the clearance for the chassis screws located under the trim. When you
replace the trim, make sure that you orient the trim with the notches in the
correct location. Otherwise the trim may not fit properly.
a. Remove the Drive trim by inserting a flat blade screwdriver in the
bottom slot of the trim and gently prying the trim piece out of the side
panel. Pull the trim up and out from the instrument.
b. Remove the two T-15 Torx-drive screws that secure the right side trim to
the bottom of the instrument.
c. Slide the trim panel toward the rear of the instrument allowing the tabs to
clear the cover openings, then pull the panel from the instrument.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
2--3
TLA5KUP Option 11: TLA5000B Memory Upgrade
Accessory
pouch
Top cover trim
Figure 2--1: Trim and covers
Right side
panel
CD-drive trim
T-15 Torx-drive
screws (2)
2--4
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
Remove the Right-Side Cover
Remove the right-side cover to access most of the internal components of the
instrument. Refer to Figure 2--2 while removing the right-side cover.
NOTE. All mounting screw holes are indicated by a star etched around the
mounting hole.
1. Remove the 15 T-15 Torx-drive screws that secure the cover to the top and
right sides of the chassis.
2. Remove the cover.
CAUTION. Take care not to bind or snag the covers on the internal cabling as you
remove or install.
TLA5KUP Option 11: TLA5000B Memory Upgrade
Figure 2--2: Right-side cover removal
Tab
Right side cover
T-15 Torx-drive
screws (14)
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
2--5
TLA5KUP Option 11: TLA5000B Memory Upgrade
Adding Memory
The memory sockets are located on the processor board mounted to the back
panel (refer to Figure 2--3).
1. Install the memory module in one of the spare sockets. For optimum
performance, leave one socket empty between the original and new DIMMs
(the correct socket for the new memory module is labeled Channel B
DIMM 0).
Memory module (s)
Battery
Verifying Operation
2--6
Figure 2--3: Memory module and battery location
2. Reinstall the trim, covers, and the accessories pouch.
To verify the proper operation of the logic analyzer, follow these steps:
1. Plug the power cord in and turn the logic analyzer on.
2. Verify that the instrument passes all power-up diagnostics.
3. Exit all applications and close any open windows.
4. Click Start in the Windows tool bar.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
TLA5KUP Option 11: TLA5000B Memory Upgrade
5. Select Programs from the Start menu.
6. Select the CheckIt Utilities application from the Programs menu.
7. Run the appropriate verification tests from the application.
8. Turn the logic analyzer off, and then on again before running any application
software including the TLA application software.
Attach the Upgrade Kit Label to the Instrument
After completing all the previous steps, you need to install the upgrade kit label
on the instrument to indicate that the kit is installed.
Locate the big label on the rear of the instrument, and then attach the TLA5KUP
kit label as shown in Figure 2--4.
TLA5204B
Kit label
Figure 2--4: Kit label locations
Place revision
label here.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
2--7
TLA5KUP Option 11: TLA5000B Memory Upgrade
2--8
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
TLA5KUP option 15 provides the capability of connecting your logic analyzer to
an external Tektronix oscilloscope through the iView cable. Please order
TLA5KUP Option 41 to get the latest TLA application software and the
Windows XP Professional operating system.
Instruments
TLA5000 Series Logic AnalyzersAll Serial Numbers
TLA5000B Series Logic AnalyzersAll Serial Numbers
Requires the latest TLA Application Software, Microsoft Windows XP Professional operating system, and 512 MB minimum mainframe memory. Also
requires the latest version of the NI-488.2 software available on the TLA
Application software CD; refer to the release notes on the CD for software
version history.
Although the instruments will operate with Microsoft Windows 2000 Professional, Tektronix recommends that you upgrade the operating system to Microsoft
Windows XP Professional through TLA5KUP Option 41.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
Online installation instructions are provided within the TLA application through
a wizard. After powering on the instrument, select Add iView External Scope
from the System menu and follow the online instructions.
Attach the Upgrade Kit Label to the Instrument
After completing all the previous steps, you need to install the upgrade kit label
on the instrument to indicate that the kit is installed.
Locate the big label on the rear of the instrument, and then attach the TLA5KUP
kit label as shown in Figure 3--1.
Kit label
Figure 3--1: Kit label locations
Place revision
label here.
3--2
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
TLA5KUP Option 20
TLA5KUP Option 20: CD-RW/DVD-ROM Drive Upgrade
TLA5KUP option 20 allows you to replace the CD-RW drive with a CD-RW/
DVD-ROM drive for your TLA5000 series logic analyzer. This option does not
include any software media and license. Please order TLA5KUP Option 41 or 42
to get the latest software to support your new drive.
NOTE. To play DVD media with your new drive, you must upgrade your software,
with TLA5KUP Option 41. TLA5KUP Option 41 includes the Microsoft Windows
XP Professional Operating System and Media Player V10. You can also add a
third-party DVD decoder.
Instruments
TLA5000 Series Logic AnalyzersAll Serial Numbers
Parts List
Table 4--1: TLA5KUP Option 20 parts list
QuantityPart numberDescription
1ea119-7196-xxCD-RW/DVD-ROM disk drive
1eaN/ATLA5KUP kit label
Minimum Tool and Equipment List
Table 4--2: Tools required for installing upgrade
Item
no.
1Screwdriver handle
2T-15 Torx tipUsed for removing most the instru-
3#1 Phillips screwdriverScrewdriver for removing Phillips
NameDescription
Accepts TorxR-driver bits
ment’s screws. TorxR-driver bit for
T-15 size screw heads
screws, CD-ROM Drive.
General tool
number
620-440
640-247
Standard tool
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
4--1
TLA5KUP Option 20: CD-RW/DVD-ROMDrive Upgrade
4--2
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
The following instructions guide you through the process of replacing the
existing CD-RW ROM drive with a CD-RW/DVD-ROM drive.
Preparation
Be sure to observe the following precautions to avoid damaging the logic
analyzer while preparing to service it. Contact your nearest Tektronix, Inc.,
Service Center or Tektronix Factory Service Center for installation assistance.
CAUTION. Many components within the instrument are susceptible to static-discharge damage.
Service only in a static-free environment. Observe standard handling precautions
for static-sensitive devices.
Always wear a grounded wrist strap, grounded foot strap, and static-resistant
apparel while installing this kit.
WARNING. Dangerous voltages may be present. Before performing any procedure
in this subsection, disconnect the power cord from the line voltage source.
Failure to do so could cause serious injury or death.
HDo not handle static-sensitive components on boards.
HTransport and store static-sensitive boards in their original containers or on
conductive foam. Label any package that contains static-sensitive assemblies.
HWear anti-static wrist strap while handling the boards to discharge the static
voltage from your body.
HDo not allow anything capable of holding or generating a static charge on
the work surface.
HAvoid handling boards in areas that have a floor or work surface cover that
is capable of generating a static charge.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
Remove the accessories pouch to access the covers on the instrument.
1. Turn the logic analyzer off.
2. Remove the power cord and all probes.
3. Open the pouch and locate the two snaps on the inside front side of the
pouch.
4. Gently pull on each of the two tabs to unsnap the front of the pouch.
5. At the rear of the instrument, peel off the Velcro that holds the pouch to the
instrument.
6. Set the pouch aside.
7. To reinstall the pouch, perform these steps in reverse order. Tighten the T-15
Torx-drive screws to 8-in lbs.
Remove the Trim and Covers
Remove the trim and covers as shown in Figure 4--1 on page 4--5, by following
this procedure.
1. Remove the top cover trim.
a. Remove the two T-15 Torx-drive screws that secure the top cover trim to
b. Remove the snap studs from the top cover.
c. Slide the trim panel toward the rear of the instrument allowing the tabs to
2. Remove the right side trim.
NOTE. When you remove the CD Drive trim, notice the notches in the trim that
provide the clearance for the chassis screws located under the trim. When you
replace the trim, make sure that you orient the trim with the notches in the
correct location. Otherwise the trim may not fit properly.
the instrument.
clear the cover openings, then pull out to remove the panel from the
instrument (you may need to loosen the two top black feet on the back of
the instrument if the cover does not come off easily).
4--4
a. Remove the Drive trim by inserting a flat blade screwdriver in the
bottom slot of the trim and gently prying the trim piece out of the side
panel. Pull the trim up and out from the instrument.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
TLA5KUP Option 41
TLA5KUP Option 41: TLA Application Software and
Windows Operating System Upgrade
TLA5KUP option 41 supports the upgrade to TLA Application Software V5.1
and Microsoft Windows XP Professional operating system.
Use this kit to upgrade TLA5000 series logic analyzers with TLA application
software 5.0, or lower, using recovery media. This process overwrites the entire
contents of your hard disk drive.
NOTE. Several third-party software applications are included to use with your
logic analyzer. These applications may include software license agreements. Be
sure to abide by those license agreements.
Instruments
TLA5000 Series Logic AnalyzersAll Serial Numbers
Minimum Tool and Equipment List
No special tools or equipment are required to install this option.
Parts List
Multiple sets of CD or DVDs are included with this kit. Use the set that supports
the serial number range of your instrument.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
5--1
TLA5KUP Option 41: TLA Application Software and Windows Operating System Upgrade
Table 5--1: TLA5KUP Option 41 parts list
QuantityPart numberDescription
1eaN/ATektronix Logic Analyzer Family Application Software V5.1
1eaN/ATektronix Logic Analyzer Family TLA5000 Recovery Media
V5.1 (CD-ROM set for TLA5000 serial numbers B010000 to
B019999)
1eaN/ATektronix Logic Analyzer Family TLA5000 Recovery Media
V5.1 (CD-ROM set for TLA5000 serial numbers B020000 to
B029999)
1eaN/ATektronix Logic Analyzer Family TLA5000 Recovery Medis
V5.1 (DVD-ROM set for TLA5000 serial numbers B030000
to B039999)
1eaN/AMicrosoft Windows XP Professional Certificate of
Authenticity
1eaN/ANero OEM Suite 3 CD
1ea016-1524-xxTLA mouse pad
1 sheetN/ASoftware revision labels
1eaN/ATLA5KUP kit label
1eaN/ARecovery media license
5--2
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
TLA5KUP Option 41: Application Software and Windows
Operating System Instructions
These instructions guide you through the process of upgrading your TLA5000
series logic analyzer software and firmware.
CAUTION. The software installation procedure overwrites the entire contents of
your hard disk. To save any files or software applications on the hard disk, back
them up to another media before continuing with this procedure.
Install Software
This kit includes media that contain the Microsoft Windows XP Professional
operating system and the latest application software. All software required to run
the logic analyzer comes with the media, with the exception of any microprocessor support packages or non-logic analyzer application software.
Back Up Files
NOTE. Use the CD or DVD set that supports the serial number range of your
instrument.
Before loading the recovery media, back up any files, applications, and personal
files. Use the Windows Search utility to locate the files to back up. Table 5--2
lists examples of file suffixes that you may want to back up.
Table 5--2: TLA user file suffixes
SuffixDescription
.tlaTLA setup files and stored data
.tsfTLA symbol files
.tbfTektronix binary format
.tlsTLA script file
.tpgTektronix pattern generator files
.txtTektronix TLA data exchange format
files
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
5--3
TLA5KUP Option 41: Application Software and Windows Operating System Instructions
Table 5--2: TLA user file suffixes (Cont.)
SuffixDescription
.stkStack files
.spzProtocol files
Install Windows XP
Professional and the TLA
Application Software
This procedure automatically reformats the hard disk, installs the Windows XP
Professional Operating system software, and installs the the Tektronix Logic
Analyzer application software. This process may take several minutes.
Change the BIOS Boot Settings and Load the Recovery Media. Before installing the
recovery media you may need to change the BIOS settings to enable booting
from the CD/DVD drive.
1. Restart the logic analyzer and then press function key F2 repeatedly to enter
the BIOS setup.
2. In the BIOS setup, go to the Boot menu.
3. Select Boot Device Priority.
4. Set the Boot devices to boot from the CD/DVD first.
5. Insert the first disc of the reovery media into the drive.
6. Save the settings by pressing function key F10 and confirm that you want to
save the new settings. The instrument will reboot and begin the installation
procedure.
7. Review the license agreement. If you agree to the terms of the license
agreement, proceed with the next step.
5--4
8. Follow the on-screen instructions to complete the installation procedure.
If the recovery process detects errors and prompts you to allow the program
to fix them, click Yes to continue the upgrade process.
9. Reinstall any user files that you backed up previously. Reinstall any software
(such as the microprocessor support packages) that you want to use on the
logic analyzer.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
TLA5KUP Option 41: Application Software and Windows Operating System Instructions
Upgrade the TLA5000 Firmware
Complete the following steps if the TLA startup messages indicate that you need
to upgrade the instrument firmware:
4. Select your instrument from the Firmware Loader -- Connection dialog box
and click Connect.
5. You may see a pop-up window reminding you to cycle the power on the
instrument after completing the upgrade operation. Click Yes to continue.
6. Select the module from the Supported list box near the top of the window.
7. Select Load Firmware from the Execute menu.
8. Click the TLA520x.lod file.
9. Click Open. You will be asked to confirm your action; click Ye s.
10. When the process is complete, exit the firmware loader program.
Upgrade the Mainframe BIOS (TLA5000 Instruments with Serial Numbers
B020000 to B029999 Only)
Use the following steps to determine if you need to upgrade the BIOS on the
logic analyzer mainframe:
1. Restart the logic analyzer and then press function key F2 to enter the BIOS
setup.
2. Check that the BIOS version string ends in P24 (for TLA5000 instruments
with serial numbers from B020000 to B029999).
3. If you have the correct version, skip the remaining steps in this procedure.
Restart the instrument, or press the Escape key, and answer OK to discard
any changes and exit (this allows the instrument to restart).
4. To update the BIOS, complete the following steps:
a. Connect to the internet and go to the following Web site:
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
5--5
TLA5KUP Option 41: Application Software and Windows Operating System Instructions
b. Review the information on the Web site about the Express BIOS update
and then download the file and copy it to your instrument desktop.
c. Close all applications on your instrument and then execute the file that
you just downloaded.
NOTE. Do not power down the instrument until the upgrade is complete.
d. Follow the on-screen instructions.
The logic analyzer will update the BIOS and the screen will be blank.
During the update processes you will hear several beeps and then the
instrument will restart. A message will appear on the screen when the
upgrade is complete.
The upgrade process should not change the BIOS settings of your instrument. If
the screen remains blank, the video BIOS settings might have been reset.
Complete the following steps to recover from any problems:
5. Connect an external monitor to the video port near the RS-232 port.
6. Restart the instrument and then press function key F2 to enter the BIOS
setup.
The BIOS screen should appear on the external monitor.
7. Use the keyboard or front panel keys to change the BIOS settings in
Table 5--3.
Table 5--3: TLA5000 BIOS settings
MenuRecommended settings
Advanced
Peripheral ConfigurationParallel Port Mode
Audio
Diskette ConfigurationDiskette ControllerDisabled
Video ConfigurationPrimary Video AdapterPCI
Boot
Silent BootEnabled
ECP
Disabled
5--6
8. Press function key F10 to save and exit the BIOS setup. The instrument
should restart and the screen should work properly.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
TLA5KUP Option 41: Application Software and Windows Operating System Instructions
Calibrate the Instrument
After verifying proper operation, you need to calibrate the logic analyzer.
1. Allow the instrument to warm up for at least 30 minutes before continuing.
2. In the System window, select Calibration and Diagnostics from the System
menu.
3. Click the Self Calibration tab.
4. Select the Run button and click Yes to any confirmation messages.
NOTE. The calibration process may take several minutes.
The Status column indicates Running while the instrument is being calibrated.
5. Verify that the Status column changes to Calibrated.
6. Close the dialog box.
Installing the TLA Application Software on a PC
You can install the TLA application software on a PC for the following purposes:
HRun the TLA application software in the Offline mode (similar to TLAVu in
earlier software versions).
HControl a logic analyzer with the same TLA Application Software over a
network.
Install Third-Party
Software
To use all of the features of the TLA application software on your PC, you must
install additional third--party software. The following third--party software is
available on the TLA Application Software CD V5.0, or higher:
HThe NI-GPIB software allows you to use the iView software with your PC.
HThe SnagIt software is useful for copying screen shots of logic analyzer data
for use with other applications.
NOTE. Several third-party software applications are included to use with your
logic analyzer. These applications may include software license agreements. Be
sure to abide by those license agreements.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
5--7
TLA5KUP Option 41: Application Software and Windows Operating System Instructions
Install the NI-GPIB Software. Complete the following steps to install the third-party software on your PC:
1. Install the first TLA Application Software CD in the CD drive.
2. Browse to the NI-GPIB-USB folder on the CD and run the Setup.exe
program.
3. Follow the on--screen instructions and note the items below:
a. When prompted, select the Typical installation option.
b. When the Add GPIB wizard appears, select G P I B -- U S B -- H S .
c. After restarting the instrument, the NI--488.2 Getting Started wizard
displays. This is not needed; select Do not show at Windows startup.
d. When you first connect the iView cable, the instrument will detect the
new hardware. Select Install the software automatically (Recommended).
Install the TLA Application
Software
Start the TLA Application
Install the SnagIt Software. Complete the following steps to install the SnagIt
software on your PC:
1. Browse to the SnagIt folder on the CD and run the Setup.exe program.
2. Follow the on--screen instructions.
Browse to the TLA Application software folder on the CD and run the Setup.exe
program. Follow the on--screen instructions to install the TLA application
software.
If you have an older version of the software on the hard disk, the installation
program will detect it and ask if you want to remove it. Follow the on-screen
instructions to remove the software, answering “Yes” to any prompts. Restart the
instrument when prompted and run the Setup.exe program again.
If you have installed the TLA application software on a PC for remote operation,
start the software by double--clicking on the TLA Application icon. The TLA
Connection dialog box displays.
1. Select an instrument in the TLA Connection dialog box and then click
Connect. If your instrument is connected to a network, you can connect to
any unused instrument on your local network.
5--8
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
TLA5KUP Option 41: Application Software and Windows Operating System Instructions
2. Click Offline to start an offline version of the TLA application software. The
TLA Offline software allows you to run the TLA application without
connecting to an instrument. You can analyze previously acquired data from
a logic analyzer, create or modify reference memories, or perform system
tests without being connected to an instrument.
Attach the Upgrade Kit Label to the Instrument
After completing all the previous steps, install the upgrade kit label on the
instrument to indicate that the kit is installed.
Locate the big label on the rear of the instrument and then attach the software
revision label and the TLA5KUP kit label as shown in Figure 5--1.
Kit label
Figure 5--1: Revision and kit label locations
Place revision
label here.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
5--9
TLA5KUP Option 41: Application Software and Windows Operating System Instructions
5--10
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
TLA5KUP Option 42 supports the upgrade to TLA application software
version 5.1. This kit provides instructions to upgrade the TLA application
software and firmware for a TLA5000 Series logic analyzer with TLA application software version 5.0 or lower. This software will run only on TLA instruments with Microsoft Windows XP Professional Operating System or Microsoft
Windows 2000 Professional Operating System purchased from Tektronix. The
software will also run on a PC with Microsoft Windows XP Professional or
Windows 2000 Professional operating systems.
NOTE. To upgrade your TLA instrument to Windows XP Professional operating
system, please order TLA5KUP Option 41.
Instruments
TLA5000 Series Logic AnalyzersAll serial numbers
TLA5000B Series Logic AnalyzersAll serial numbers
Installation Prerequisite
Parts List
Requires TLA Application Software Version 4.3 (or higher), Microsoft Windows XP Professional Operating System or Microsoft Windows 2000 Professional Operationg System.
Table 6--1: TLA5KUP Option 42 parts list
QuantityPart numberDescription
1eaN/ATektronix Logic Analyzer Family Application Software
Version 5.1
1eaN/ANero OEM Suite 3 CD
1ea016-1524-xxTLA mouse pad
1 sheetN/ASoftware revision labels
1eaN/ATLA5KUP kit label
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
The following instructions direct you through the process of upgrading the TLA
application software on your TLA5000 series logic analyzer.
NOTE. Third-party software applications are included to use with your logic
analyzer. These applications may include software license agreements. Be sure to
abide by those license agreements.
Upgrade Instruction Overview
While performing the upgrade procedures, keep the following items in mind:
HThroughout this procedure you will see the term “log on as Administrator.”
The instrument is set up to automatically log on as Administrator (with no
password) so you may not see the log on prompt. If the network setups have
been changed on your instrument, make sure that you log on as Administrator or as a user who has administrator privileges. Failure to do so can prevent
the software upgrade from completing successfully.
HAfter logging on as administrator, quit any applications.
HIf your instrument does not display the file extensions, you need to change
HUse the Windows Add or Remove Programs Control Panel utility to remove
HYou may be asked to restart Windows after you install each software
HIf Windows does not restart normally, press and hold the On/Standby button
Remove the Old Software
Complete the following steps to remove the old software packages from your
logic analyzer. Skip any steps that mention software that isn’t installed on your
logic analyzer.
your folder options to display the file extensions. (From the Control Panel,
double-click Folder Options, select the View tab, and then deselect the “Hide
file extensions for known file types” selection.)
existing software.
package. If you do not restart Windows when prompted, your software may
not install properly and can cause unpredictable behavior.
for five or six seconds to force a shutdown.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
NOTE. Refer to the release notes for a list of software version numbers to
determine which software you may need to upgrade. The release notes are
included on the TLA application software CD.
1. If you have not already done so, power on the logic analyzer and log on as
Administrator.
2. Exit the TLA application.
3. Open the Windows Control Panel.
4. Double-click the Add or Remove Programs icon.
5. For each of the following software packages, select the software package (the
exact names may very depending on the software version), click the
Change/Remove button, and follow the on-screen instructions to remove the
software:
HAdobe Reader
Update the TLA Drivers
HNI-488.2
HNI-Max
HRecordNow MAX
HSnagIt
HTektronix TLA Application
6. Windows will uninstall the selected application. If you are asked for
permission to remove unused shared files or any read-only files, select Yes to
All.
7. After the software has been successfully removed, restart Windows.
Complete the following steps if your instrument has Microsoft Windows 2000
Professional Operating System and had TLA Application Software V4.3 or
V4.4. This procedure removes the old PCI--VXI Bridge driver and installs the
new one required by the TLA driver.
1. After rebooting the instrument, insert the second CD of the TLA Application
software set.
6- 4
2. Open the Windows Control Panel.
3. Open the Device Manager in the Hardware page of the System Properties
dialog box.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
5. Select the module from the Supported list box near the top of the window.
6. Select Load Firmware from the Execute menu.
7. Click the TLA520x.lod file.
8. Click Open. You will be asked to confirm your action; click Yes .
9. When the process is complete, exit the firmware loader program.
Upgrade the Mainframe BIOS (TLA5000 Instruments with Serial Numbers
B020000 to B029999 Only)
Use the following steps to determine if you need to upgrade the BIOS on the
logic analyzer mainframe:
1. Restart the logic analyzer and then press function key F2 to enter the BIOS
setup.
2. Check that the BIOS version string ends in P24 (for TLA5000 instruments
with serial numbers from B020000 to B029999).
3. If you have the correct version, skip the remaining steps in this procedure.
Restart the instrument, or press the Escape key, and answer OK to discard
any changes and exit (this allows the instrument to restart).
4. To update the BIOS, complete the following steps:
a. Connect to the internet and go to the following Web site:
b. Review the information on the Web site about the Express BIOS update
and then download the file and copy it to your instrument desktop.
c.Close all applications on your instrument and then execute the file that
you just downloaded.
NOTE. Do not power down the instrument until the upgrade is complete.
d. Follow the on-screen instructions.
The logic analyzer will update the BIOS and the screen will be blank.
During the update processes you will hear several beeps and then the
instrument will restart. A message will appear on the screen when the
upgrade is complete.
6- 8
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
The upgrade process should not change the BIOS settings of your instrument. If
the screen remains blank, the video BIOS settings might have been reset.
Complete the following steps to recover from any problems:
5. Connect an external monitor to the video port near the RS-232 port.
6. Restart the instrument and then press function key F2 to enter the BIOS
setup.
The BIOS screen should appear on the external monitor.
7. Use the keyboard or front panel keys to change the BIOS settings in
Table 6--2.
Table 6- 2: TLA5000 BIOS settings
MenuRecommended settings
Advanced
8. Press function key F10 to save and exit the BIOS setup. The instrument
Calibrate the Instrument
After verifying proper operation, you need to calibrate the logic analyzer.
1. Allow the instrument to warm up for at least 30 minutes before continuing.
2. In the System window, select Calibration and Diagnostics from the System
3. Click the Self Calibration tab.
Peripheral ConfigurationParallel Port Mode
Audio
Diskette ConfigurationDiskette ControllerDisabled
Video ConfigurationPrimary Video AdapterPCI
Boot
Silent BootEnabled
should restart and the screen should work properly.
menu.
ECP
Disabled
4. Select the Run button and click Yes to any confirmation messages.
NOTE. The calibration process may take several minutes.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
The Status column indicates Running while the instrument is being
calibrated.
5. Verify that the Status column changes to Calibrated.
6. Close the dialog box.
Installing the TLA Application Software on a PC
You can install the TLA application software on a PC for the following purposes:
HTo run the TLA application software in the Offline mode (similar to TLAVu
in earlier software versions).
HTo control a logic analyzer with the same TLA application software version
over a network.
Install Third-Party
Software
To use all of the features of the TLA application software on your PC, you must
install additional third--party software. The following third--party software is
available on the TLA Application Software CD V5.0, or higher:
HThe NI-GPIB software allows you to use the iView software with your PC.
HThe SnagIt software is useful for copying screen shots of logic analyzer data
for use with other applications.
NOTE. Several third-party software applications are included to use with your
logic analyzer. These applications may include software license agreements. Be
sure to abide by those license agreements.
Install the NI-GPIB Software. Complete the following steps to install the third-party software on your PC:
1. Install the TLA Application Software CD in the CD drive.
2. Browse to the NI-GPIB-USB folder on the CD and run the Setup.exe
program.
3. Follow the on--screen instructions and note the items below:
6- 10
a. When prompted, select the Typical installation option.
b. When the Add GPIB wizard appears, select GPIB--USB--HS.
c.After restarting the instrument, the NI--488.2 Getting Started wizard
displays. This is not needed; select Do not show at Windows startup.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual
d. When you first connect the iView cable, the instrument will detect the
new hardware. Select Install the software automatically (Recom-mended).
Install the SnagIt Software. Complete the following steps to install the SnagIt
software on your PC:
1. Browse to the SnagIt folder on the CD and run the Setup.exe program.
2. Follow the on--screen instructions.
Install the TLA Application
Software
Start the TLA Application
Browse to the TLA Application software folder on the CD and run the Setup.exe
program. Follow the on--screen instructions to install the TLA application
software.
If you have an older version of the software on the hard disk, the installation
program will detect it and ask if you want to remove it. Follow the on-screen
instructions to remove the software, answering “Yes” to any prompts. Restart the
instrument when prompted and run the Setup.exe program again.
If you have installed the TLA application software on a PC for remote operation,
start the software by double--clicking on the TLA Application icon. The TLA
Connection dialog box displays.
1. Select an instrument in the TLA Connection dialog box and then click
Connect. If your instrument is connected to a network, you can connect to
any unused instrument on your local network.
2. Click Offline to start an offline version of the TLA application software. The
TLA Offline software allows you to run the TLA application without
connecting to an instrument. You can analyze previously acquired data from
a logic analyzer, create or modify reference memories, or perform system
tests without being connected to an instrument.
TLA5KUP Logic Analyzer Field Upgrade Kit Instruction Manual