Tektronix TL725 User manual

TL725 English Instruction Sheet
Page 1
TL725
2x4-Wire Ohms Tweezers Test Leads
Instruction Sheet
The Tektronix TL725 TWZ 2X4-Wire Tweezers Test Leads are designed for use with the Tektronix DMM4020, 4040, and 4050 multimeters. The split contacts of the test lead’s banana plugs are aligned to mate with the split contacts of the DMM4020, 4040, and 4050 HI and LO input connectors.
The split contact design of the meters and the test leads bring both the input and sense terminals of the meters to any 4-wire resistance measurement with only two, instead of four test leads.
PN 3481455
*P071269500*
071269500
TL725 English Instruction Sheet Page 2
Safety Information
 Warning
To avoid possible electric shock or personal injury:
Do not use the test leads to measure voltages above 30 V rms, 42.4 V peak or 60 V dc. Voltages above these levels will cause electric shock.
Do not use the test leads for measurements within Measurement Categories II, III, and IV.
Use the test leads only as specified or the protection provided may be impaired.
Do not use the test leads if they are cracked or have exposed metal.
Do not exceed the test lead’s voltage and current rating.
Make sure test leads are fully mated to the instrument and accessories.
Keep fingers behind the finger guards
Adhere to local and national safety codes.
Individual protective equipment must be used
to prevent shock and injury.
  
International Symbols
gninaeM lobmyS
Earth ground
Hazardous voltage
Risk of danger. Important information. See instruction sheet.
Double Insulated