Tektronix TIVM1 2,TIVM1L 2 User Manual

TIVM Series
xx
ZZZ
IsoVu™ Measurement System
User Manual
*P071349501*
071-3495-01
xx
TIVM Series
ZZZ
User Manual
Register now!
Click the following link to protect your product.
www.tek.com/register
www.tek.com
071-3495-01
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
ISOVU is a trademark of Tektronix Inc.
TEKVPI is a registered trademark of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. B o x 5 0 0 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200. Worldwide, visit www.tek.com to nd contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty work may be n the property of Tektronix.
ew or reconditioned to like new performance. All replaced parts, modules and products become
In order to o the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage result b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has b een modied or integrated with other products when the effect of such modication or integration increases the time or difculty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF A NY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
TRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE
TEK AND EXCLUSIVE REMEDY PR OVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
[W2 – 15AUG04]
btain service under this warranty, Customer must notify Tektronix of the defect before the expiration of
ing from attempts by personnel other than Tektronix representatives to install, repair or service the product;
Table of Contents
Important safety information ........... ................................ ................................ ........... v
General safety summary ...................................................................................... v
Service safety summary ............................ ................................ ......................... vii
Terms in this manual ..................... ................................ ................................ ... viii
Symbols and terms on the product......................................................................... viii
Preface .............................................................................................................. ix
Key features ........................... .................................. ................................ ...... ix
Laser certication................. ................................ .................................. .......... ix
Product description ..... ................................ .................................. ..................... x
Models ......................................................................................................... xi
Supported oscilloscopes...................................................................................... xi
Operating information... .................................. ................................ ......................... 1
Accessories..................................................................................................... 1
Operating considerations...................................................................................... 2
Controls and indicators............................... ................................ ......................... 7
Connecting to a circuit ....... ................................ ................................ ................ 10
Self calibration ................................................................................................ 12
AutoZero....................................................................................................... 14
Menu button ............................. ................................ ................................ ...... 14
Offset correction .............................................................................................. 15
1X/2X Range.................... ................................ ................................ .............. 17
Auto Range.................................................................................................... 18
Selecting a sensor tip cable .................................................................................. 19
Output clamping .......... ................................ .................................. .................. 20
Sensor tip loading............................................................................................. 20
Probe compensation .......................................................................................... 21
Deskew......................................................................................................... 21
Input offset ...... .................................. ................................ ............................ 21
Application examples ............... ................................ ................................ .............. 23
Example 1: High-side V
Example 2: High-side drain current measurement ............................ ............................ 25
Example 3: ESD troubleshooting ............................... .................................. .......... 26
Reference information ............................................................................................ 29
Specications...................................... ................................ ............................ 29
Dimensional drawings........................................................................................ 33
IsoVu measurement system block diagram ................ .................................. .............. 36
Tripods........... ................................ ................................ .............................. 37
Installing the probe tip adapters . .... . ..... . ..... . ..... . ..... . ... . . ..... . ..... . ..... . ..... . ... . . . .... . ..... . . 39
Installing the square pins on the circuit board . ..... . .... . . .... . ..... . ..... . ..... . .... . ..... . ..... . ..... . .. 42
measurement................................................................... 23
GS
TIVM Series Measurement System User Manual i
Table of Contents
User service .......... ................................ .................................. ............................ 45
Service offerings ................ .................................. ................................ ............ 45
Preventive maintenance.................. ................................ .................................. .. 45
Performance verication procedures ....................................................................... 46
Propagation delay............................................................................................. 47
Troubleshooting and error conditions .................. .................................. .................. 50
Repack the m
Test record ..................................................................................................... 53
Appendix A: Remote programming ............................................................................. 55
CH<n>:PRObe?............................................................................................... 55
CH<n>:PRObe:AUTOZero EXECute ..................................................................... 55
CH<n>:PRObe:COMMAND “CLAMP”, {“ON” | “OFF”}......................... .................... 55
CH<n>:P
CH<n>:PRObe:FORCEDRange <NR3> .................................................................. 56
CH<n>:PRObe:GAIN? .... ................................ ................................ .................. 56
CH<n>:PRObe:ID {:SERnumber | :TYPe}? ........ ................................ ...................... 56
CH<n>:PRObe:PROPDELay?............ ................................ ................................ .. 57
CH<n>:PRObe:RECDESkew?.............................................................................. 57
CH<n
CH<n>:PRObe:UNIts?........................... .................................. .......................... 57
CH<n>:PROBECOntrol {AUTO | MAN}................................................................. 57
CH<n>:PROBEFunc:EXTAtten <NR3>................................................................... 57
CH<n>:PROBEFunc:EXTDBatten? ....................................................................... 57
CH<n>:PROBEFunc:EXTUnits {“UU” | “None”}... .................................. .................. 58
pendix B: Compliance information .......................................................................... 59
Ap
Safety compliance ............................................................................................ 59
Environmental considerations............................................................................... 61
Index
easurement system for shipment ............................................................ 52
RObe:SET {“CLAMP ON” | “CLAMP OFF”}................................................ 55
>:PRObe:RESistance? ................................................................................ 57
ii TIVM Series Measurement System User Manual
List of Figures
Figure i: TIVM Series IsoVu Measurement System............................................................ x
Figure 1: Maximum safe handling limits for common mode voltages between the sensor head and earth
ground........................................................................................................... 5
Figure 2: RF burn hazard zone around the sensor head............. ................................ ........... 6
Figure 3: Controller indicators and buttons...................................................................... 7
Figure 4: Labels on the sensor head .............................................................................. 9
Figure 5: Sensor tip cable top and botto
Figure 6: Connect the Comp box to the oscilloscope.. . . ..... . ..... . .... . ..... . ..... . .... . ..... . ... . . . .... . ... 11
Figure 7: Connecting the sensor tip cable to the sensor head................................................. 11
Figure 8: Connect the s ensor head to the exible tripod ...... .................................. .............. 12
Figure 9: Probe Setup menu...................................................................................... 14
Figure 10: Digital lter aliasing ................................................................................. 15
Figure 11: Digital lter aberrations (about 2.5% of V
Figure 12: Digital lter aberrations after disabling offset correction........................................ 17
Figure 13: Top sensor tip labels.................................. ................................ ................ 17
Figure 14: Bottom sensor tip labels ............................................................................. 20
Figure 15: Half-bridge circuit showing the gate, source, and drain of the high-side FET.. . .... . ..... . ... 23
Figure 16: High-side turn-on characteristics . ................................ ................................ .. 24
Figure 17: High side current shunt .............................................................................. 25
Figure 18: SMT resistor model .................................................................................. 26
Figure 19: ESD discharge test example....................... .................................. ................ 27
Figure 20: Sensor head dimensions with probe tip cover ................... .................................. 33
Figure 21: Sensor head dimensions without probe tip cover ....... ................................ .......... 34
Figure 22: Controller dimensions................................................................................ 34
Figure 23: Comp box dimensions ............................................................................... 35
Figure 24: Probe tip adapter dimensions.................................... .................................. .. 35
Figure 25: Block diagram .................... .................................. ................................ .. 36
Figure 26: Installing the exible tripod under the DUT. ...................................................... 37
Figure 27: Connecting the sensor head to the top of the DUT with the exible tripod. ................... 37
Figure 28: Connecting to an adapter on the circuit board with the probe tip tripod ....................... 38
Figure 29: Lining up the MMCX-to-0.1-inch (2.54 mm) adapter on the circuit board.................... 39
Figure 30: Lining up the MMCX-to-0.062-inch (1.57 mm) adapter on the circuit board .... . ..... . .... . . 4 0
Figure 31: Pushing the MMCX-to-0.062-inch (1.57 mm) adapter in place................................. 40
Figure 32: Pushing the MMCX-to-0.1-inch (2.54 mm) adapter in place............................ ........ 41
Figure 33: Adapter clearance requirements ...................... ................................ .............. 42
Figure 34: Removing the header from square pins on the circuit board..................................... 43
Figure 35: Using the soldering aide to install the square pins on the circuit board ........................ 44
Figure 36: Propagation delay measurement .................................................................... 50
m labels ............................................................... 9
) ... ................................ .................. 16
p-p
TIVM Series Measurement System User Manual iii
Table of Contents
List of Tables
Table 1: Standard accessories ...... .................................. ................................ ............. 1
Table 2: Opt
Table 3: Input specications...................... .................................. ............................... 3
Table 4: Environmental considerations .......................................................................... 3
Table 5: Controller indicators and buttons.................. ................................ ..................... 7
Table 6: Sensor tip selection table ............................................................................... 19
Table 7: Input offset............................................................................................... 21
Table 8: W
Table 9: Electrical specications .............. ................................ .................................. 29
Table 10: Physical specications ........ .................................. ................................ ...... 33
Table 11: Required equipment for performance verication ................................................. 46
Table 12: Problems and possible solutions ..................................................................... 50
Table 13: Test record........................ .................................. ................................ .... 53
Table
ional accessories ..................................................................................... 2
arranted specications.................................. ................................ .............. 29
14: Sensor tip cables and dynamic ranges ......................... .................................. .... 56
iv TIVM Series Me asurement System User Manual
Important safety information
This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe condition.
To safely perform service on this product, additional information is provided at the end of this section. (See page vii, Service safety summary.)
General safety summary
Use the product only as specied. Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. Carefully read all instructions. Retain these instructions for future reference.
Comply with local and national safety codes.
For correct and safe operation of the product, it is essential that you follow generally accepted safety procedures in addition to the safety precautions specied in this manual.
The product is designed to be used by trained personnel only.
To avoid re or personal
injury
Only qualied perso the cover for repair, maintenance, or adjustment.
Before use, always check the product with a known source to be sure it is operating correctly.
This product is not intended for detection of hazardous voltages.
Use personal protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions related to operating the system.
When incorporating this equipment into a system, the safety of that system is the responsibility of the assembler of the system.
Connect and disconnect properly. Do not connect or disconnect sensor tip cables, test leads, or accessories while they are connected to a voltage source. Use only test leads and accessorie to be suitable for the product.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product. Do not exceed the Measurement Category (CAT) rating and voltage or current rating of the lowest rated individual component of a product or accessory.
nnel who are aware of the hazards involved should remove
s supplied w ith the product, or indicated by Tektronix
TIVM Series Measurement System User Manual v
Important safety information
Do not apply a po
Do not operate without covers. Do not operate this product with covers or panels removed, or with the case open. Hazardous voltage exposure is possible.
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualied service personnel.
Disable the product if it is damaged. Do not use the product if it is damaged or operates incorrectly. If in doubt about safety of the product, turn it off and disconnect it from the instrument. Clearly mark the product to prevent its further operation.
Before use, inspect accessories for mechanical damage and replace when damaged. Do not use them if they are damaged, or if there is exposed metal.
Examine the exterior of the product before you use it. Look for cracks or missing pieces.
Use only specied replacement parts.
Do not operate in wet/damp conditions. Be aware that condensation may occur if a unit is moved from a cold to a warm environment.
tential that exceeds the m aximum rating.
Sensor tip cables
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry. Remove the input signals before you clean
the product.
Provide a safe working environment. Always place the product in a location convenient for viewing the display and indicators.
Be sure your work area meets applicable ergonomic standards. Consult with an ergonomics professional to avoid stress injuries.
Maintain safe clearance from the sensor head and sensor tip cable while connected to the energized circuit as recommended in this manual.
Remove the sensor tip cable and adapters from the test circuit when not in use.
Leave the sensor tip cable connected to the sensor head when not in use.
Use only correct Measurement Category (CAT), voltage, temperature, altitude, and amperage rated sensor tip cables and accessories for any measurement.
Beware of high voltages. Understand the voltage ratings for the product you are using and do not exceed those ratings. It is important to know and understand the maximum measurement voltage rating of the product. The voltage rating depends
vi TIVM Series Measurement System User Manual
Important safety information
on the me asurem Specications section of the manual for more information.
WAR NI NG . To prevent electrical shock, do not exceed the maximum measurement
or maximum voltage category.
Connect and disconnect properly.
CAUTION. To avoid damage to the equipment, de-energize the test circuit before
connecting or disconnecting the sensor tip cable.
Servicesafetysummary
The Serv safely perform service on the p roduct. Only qualied personnel should perform service procedures. Read this Service safety summary and the General safety summary before performing any service procedures.
To avoid electric shock. Do not touch exposed connections.
ent category, the instrument, and your application. Refer to the
icesafetysummarysection contains additional information required to
Do not service alone. Do not perform internal service or adjustments of this product unless another person capable of rendering rst aid and resuscitation is present.
Disconnect power. To avoid electric shock, switch off the product power and
connect the power cord from the mains power before removing any covers or
dis panels, or opening the case for servicing.
Use care when servicing with power on. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
Verify safety after repair. Always recheck ground continuity and mains dielectric strength after performing a repair.
TIVM Series Measurement System User Manual vii
Important safety information
Termsinthismanual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Isolated, electrically oating. The terms isolated, electrically floating,and galvanically isolated are used in this document to indicate a measurement where
there is no direct conduction path to earth ground.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately acc essible as you read the marking.
CAUTION indicates a hazard to property including the product.
this symbol is marked on the product, be sure to consult the manual
When to nd out the nature of the potential hazards and any actions which have to betakentoavoidthem. (Thissymbolmayalsobeusedtorefertheuserto
ings in the manual.)
rat
The following symbol(s) may appear on the product:
viii TIVM Series Measurement System User Manual
Preface
Preface
Key features
Laser certication
This documen Series IsoVu Mea surement System. The measurement system offers a galvanically isolated measurement solution for accurately resolving high bandwidth, low voltage differential signals up to ±50 Vpk in the presence of large common mode voltages with the best in class common mode rejection p erformance across its bandwidth.
New IsoVu technology - galvanically isolated, oating, measurement system
Differential voltages up to ±50 Vpk (depending on sensor tip)
Large common mode voltage range up to 60 kV peak
t provides information for installing and using the Tektronix TIVM
Bandwidth from DC to 1 GHz
> 120 dB (1 million to 1 common mode rejection ratio) from DC to
100 MHz and 80 dB (10,000 to 1) at 1 GHz
This product complies with 21 CFR 1040.10 and 1040.11 except for deviations
suant to Laser Notice No. 50, dated June 24, 2007.
pur
CAUTION. Useofcontrolsoradjustmentsforperformanceofproceduresother
than those specied herein may result in hazardous radiation exposure.
TIVM Series Measurement System User Manual ix
Preface
Product description
The Tektronix TIVM Series IsoVu Measurement System offers a completely galvanically isolated (optical isolation) system. The system consists of a sensor tip cable, a sensor head, a controller, and a TekVPI interface as shown in the following gure. Hazardous voltages in the sensor head are completely isolated from the con
troller and oscilloscope by optical ber cables.
Figure i: TIVM Series IsoVu Measurement System
Comp box
Controller
x TIVM Series Measurement System User Manual
The TekVPI compensation box (Comp box) connects the measurement system to one of the input channels on the oscilloscope. Power is supplied to the
asurement system through the TekVPI interface of the oscilloscope.
me
The controller connects to the oscilloscope through a coaxial cable and the Comp box. Buttons and indicators on the controller provide a means for controlling the
easurement system and indicating the overall status.
m
Preface
Models
Sensor head
Sensor tip cables
The sensor head the controller. It contains an electro-optic converter that converts the electrical signal from the sensor tip cables to an optical signal sent to the controller.
Different se the DUT. The sensor tip cable consists of an SMA connector that connects to the sensor head; two screws to secure the sensor tip cable to the sensor head; the other end of the sensor tip cable connects to the DUT through an MMCX connector and optional adapters.
The TIVM Series IsoVu Measurement System includes the following models:
TIVM1. Tektronix IsoVu 1 GHz Medium Voltage with 3 m cable
TIVM1L. Tektronix IsoVu 1 GHz Medium Voltage with 10 m cable
TIVM05. Tektronix IsoVu 500 MHz Medium Voltage with 3 m cable
TIVM05
TIVM02. Tektronix IsoVu 200 MHz Medium Voltage with 3 m cable
provides an interface between the device-under-test (DUT) and
nsor tip cables are available to connect the measurement system to
L. Tektronix IsoVu 500 MHz Medium Voltage with 10 m cable
Supp
orted oscilloscopes
The measurement system can be used with the following Tektronix oscilloscopes. For oscilloscopes not included in this list, contact your local Tektronix rep
In addition to the above oscilloscopes, the measurement system can also be used with the following oscilloscopes with a TCA-VPI50 adapter.
TIVM02L. Tektronix IsoVu 200 MHz Medium Voltage with 10 m cable
resentative.
MDO3000 series
MSO/DPO4000B serie s
MDO4000B/C series
MSO/DPO5000B serie s
DPO7000C series
MSO/DPO70000C series
MSO/DPO70000DX series
DPO70000SX series
TIVM Series Measurement System User Manual xi
Preface
xii TIVM Series Measurement System User Manual
Operating information
Accessories
This section lists the standard and optional accessories available for the measurement s ystem.
Standard ac
cessories
Table 1: Standard accessories
Tektronix
Accessory
IsoVu product carrying case, soft case
IsoVu accessories carrying case, soft case
Solder aid for 0.062-inch (1.57 mm) pitch square pins (0.016 - 0.018-inch (0.4 - 0.46 mm) square pin installation tool)
5X Sensor tip cable
25X Sensor tip cable
5/16-inch SMA wrench/driver tool
Probe tip adapter (blue), MMCX to 0.1-inch (2.54 mm) square pin (0.025-inch (0.635 mm) square pins)
Probe tip adapter (white), MMCX to 0.062-inch (1.57 mm) square pin (0.016
- 0 .018-inch (0.4 - 0.46 mm) square pins)
DUT Interface pin kit with (qty. 20) 0.018-inch (0.46 mm) round solder-in pins
Flexible tripod with quick release 352-1171-xx
Flexible tripod feet, 3 each
Probe tip tripod support with living h inge, 2 each 352-1170-xx
Certicate of traceable calibration
Data calibration report
part number
016-2108-xx
016-2110-xx
003-1946-xx
IVTIP5X
IVTIP25X
003-1947-xx
131-9717-xx
131-9677-xx
020-3169-xx
344-0693-xx
TIVM Series Measurement System User Manual 1
Operating information
Optional accessories
Additional acc following table lists the optional accessories.
Table 2: Optional accessories
Accessory
1X Sensor tip cable
10X Sensor tip cable
50X Sensor tip cable
Operating considerations
Read this operating requirements and clearance requirements including possible hazardous areas when the measurement system is connected to the DUT.
Measurement system
handling best practices
The mea to avoid damage or degrading the performance due to mishandling. Consider the following precautions when handling the ber-optic cables and sensor tip cables:
essories, such as other sensor tip cables are available. The
Tektronix part number
IVTIP1X
IVTIP10X
IVTIP50X
section before i nstalling your measurement system to be aware of the
surement system consists of quality parts and should be tre ated with care
Do not crush, crimp, or sharply bend the ber-optic cable. Avoid making loops in the ber-optic cable smaller than 5 inches (12.7 cm).
Do not twist the ber-optic cable; twisting the cable will stress the optical bers.
Do not allow kinks or knots to develop in the ber-optic cable.
id putting tension on the ber-optic cable.
Avo
Do not pull or jerk the ber-optic cable, especially when kinks or knots are
esent.
pr
Do not drop the sensor head or controller assembly since damage and
isalignment of the internal optical components can result.
m
Avoid over-bending the sensor tip cables; do not exceed the minimum bend radius of 2.0 inches (5.1 cm).
Avoid crushing the cables by accidentally running over the cable with a chair wheel or by dropping a heavy object onto the cable.
Never support the weight of the sensor head or controller by the ber-optic cable.
Store the measurement system in the supplied carrying case when not in use.
2 TIVM Series Measurement System User Manual
Operating information
Environmental
requirements
The following t
ables describe specications and maximum operating environmental ratings for the measurement system when connected to a DUT and a Tektronix oscilloscope.
Table 3: Input specications
Feature Description
Common mode
Differential mode Sensor tip cable dependent (See the Sensor tip cable voltage
1
This product is designed to be used in the circuit with expected transient overvoltage up to 5000 V peak.
60 kV peak
rating below.)
1
Table 4: Environmental considerations
Feature Description
Temperature
Controller
Operating 0 °C to 40 °C (32 °F to 104 °F)
Non-operating
Sensor head
Operating 0 °C to 70 °C (32 °F to 158 °F)
Non-operating
Sensor tip cables/adapters
Operating and non-operating
Humidity
Controller
Operating 5% to 85% RH (Relative H umidity), at up to 40 °C (104 °F),
Non operating
Sensor head
Operating 5% to 80% RH (Relative Humidity), at up to 40 °C (104 °F)
Non operating
-40°Cto70°C(-40°Fto158°F)
-40°Cto70°C(-40°Fto158°F)
-40°Cto85°C(-40°Fto185°F)
non-condensing
5% to 85% RH (Relative Humidity), at up to 40 °C (104 °F)
5% to 45 % RH above 40 °C (104 °F) up to 70 °C (158 °F), non-condensing
5% to 45 % RH above 40 °C (104 °F) up to 70 °C (158 °F), non-condensing
5% to 85% RH (Relative Humidity), at up to 40 °C (104 °F)
5% to 45 % RH above 40 °C (104 °F) up to 70 °C (158 °F), non-condensing
TIVM Series Measurement System User Manual 3
Operating information
Table 4: Environmental considerations (cont.)
Feature Description
Sensor tip cables/adapters
Operating 5% to 80% RH (Relative Humidity), at up to 40 ° C (104 °F)
5% to 45 % RH above 40 °C (104 °F) up to 85°C (185 °F), non-condensing
Non operating
Altitude
Operating 3000 m (9843 ft.)
Non operating
5% to 85% RH (Relative Humidity), at up to 40 °C (104 °F)
5% to 45 % RH above 40 °C (104 °F) up to 70 °C (158 °F), non-condensing
12,000 m (39370 ft.)
Cleara
nce requirements
The unique common mode voltage range of the measurement system allows it to be used in the presence of high frequency/high voltage common mode signals. It is important to observe all precautions while using this product.
ING. RF burns can occur while using this measurement system. The system
WARN
is intended to isolate the operator from hazardous input voltages (common voltages);the plastic case of the sensor head and the shield on the sensor tip cable do not supply safe isolation.
Maintain the safe clearance from the sensor head and sensor tip cable while the measurement system is connected to the energized circuit as recommended in this document. (See Figure 2 on page 6.) Do not access the RF Burn Hazard Zone while taking measurements on a live circuit.
WARNING. RF burns can occur while using this measurement system. Anyone
making measurements that might come within the RF burn regions indicated in the following gure should be familiar with the hazards of working with signals in these regions and take appropriate action, such as RF shielding for the DUT.
4 TIVM Series Measurement System User Manual
Operating information
While measurin RF burns. Refer to the following derating curve to identify the danger areas. Measuring common mode signals within the light gray area can result in RF burns while directly contacting the sensor head. Measuring common mode signals within the dark gray shaded area can result in RF burns when within 1 m (40 in.) of the sensor head.
Figure 1: Maximum safe handling limits for common mode voltages between the sensor head and earth ground.
g high frequency common mode signals, there is a risk of
TIVM Series Measurement System User Manual 5
Operating information
The following potential RF burn area when working with hazardous voltages.
gure shows the components of the measurement system and the
Figure 2: RF burn hazard zone around the sensor head
6 TIVM Series Measurement System User Manual
Operating information
Controls and i
Controller
ndicators
The following gure shows the indicators and buttons on the controller; their functions are described in the following table.
Figure 3: Controller indicators and buttons
Table 5: Controller indicators and buttons
Item Description
1
2
3
OVERRANGE Indicator. This red LED indicates if the DC/low frequency differential voltage applied to the sensor head or attached sensor tip cable has exceeded the maximum speci ed input voltage levels.
RANGE indicator. The two LEDs indicate the differential input range setting.
RANGE button. Push this button to toggle between one of two differential input voltage ranges.
MENU button. Push this button to view the Probe Control menu on the oscilloscope display.
TIVM Series Measurement System User Manual 7
Operating information
Table 5: C ontroller indicators and buttons (cont.)
Item Description
4
5
6
CLAMPING indicator. This LED indicates whether the output clamping is enabled.
CLAMPING button. Push this button to enable or disable the output clamping feature.
SELF CAL indicator. This LED indicates the self calibration status.
Solid green. The self calibration has passed.
Flashing orange. The self calibration is in progress.
Solid red. The self calibration has failed.
Solid Orange. The self calibration has not been run or is questionable.
SELF CAL button. Push this button to start the self calibration routine.
STATUS indicator. This LED indicates the status of the measurement system:
Solid green. The unit has powered on and passed the power-on self
tests and is in normal operation.
Flashing green. The unit has not completed the power-up sequence. This is typically due to a communication fault between the host oscilloscope and the IsoVu unit. Disconnect the TekVPI comp box and then reattach the comp box.
Solid or ashing red. Error condition that requires the unit to be sent to Tektronix for service.
Flashing red/yellow. The unit has failed the power on self tests. Cycle the power; if the problem persists, the unit needs to be sent to Tektronix for service.
8 TIVM Series Measurement System User Manual
Operating information
Sensor head
Sensor tip cables
Labels on the se the DUT. They also provide reminders of the potential RF burn hazards while connected to the DUT.
Figure 4: Labels on the sensor head
Each sensor tip cable has a set of labels on the top and bottom. The top labels provide reminders of the maximum differential input voltage range for each tip cable. The bottom labels include the name of the tip cable and the differential input resistance and capacitance for the tip cable (differential loading).
nsor head provide h igh-level specications for connecting to
Figure 5: Sensor tip cable top and bottom labels
TIVM Series Measurement System User Manual 9
Operating information
Connecting to
a circuit
WARNING. This measurement system contains laser sources; exposing these
laser sources may cause laser exposure. Except for the sensor tip cables on the sensor head, do not remove any plastic or metallic covers from the sensor head or controller or attempt to disassemble the product.
WARNING. Do not connect the measurement system to an energized circuit to
avoid the risk of shock. Always de-energize the circuit-under-test before installing or removing the tip cable from the circuit-under-test. The plastic case of the sensor head and the shield tip of the sensor cable do not supply the isolation.
WARNING. To avoid the risk of electrical shock or RF burns while the DUT
is energized, do not touch the sensor head or sensor tip cable while taking measur the measurement. (See Figure 2 on page 6.)
Be sure the RF burn hazard zone. (See Figure 1 on page 5.)
ements.Alwayskeepa1m(40in.) clearancefromthesensorheadduring
to check the maximum ratings and derating curve for more information on
ION. To avoid possible damage to the equipment, do not connect the coaxial
CAUT
(common) shield of the sensor tip cable or SMA input to the high impedance portion of a circuit. The additional capacitance can cause circuit damage. Connect the coaxial (common) shield to the low impedance section of the circuit.
NOTE. Touching the sensor head or sensor tip cable when measuring a high
frequency, high voltage common mode signal increases the capacitive coupling and can degrade the common mode loading on the circuit-under-test.
WARNING. To prevent the arc ash caused by a different potential, do not place
the sensor head or sensor tip cable on the circuit that has the different voltage.
10 TIVM Series Measurement System User Manual
Operating information
The following s between a Tektronix oscilloscope and the DUT.
1. Verify the DUT
2. Connect the Comp box to one of the channels on the oscilloscope.
Figure 6: Connect the Comp box to the oscilloscope
3. Refer to the following gure and install the sensor tip cable to the sensor head.
a. Line up the sensor tip cable with the sensor head.
teps describe the process for connecting the measurement system
is not connected to an energized circuit.
Take care to avoid bending or twisting the sensor tip cable assembly during this process.
b. Connect the SMA connector of the sensor tip cable to the sensor head.
Use the accessories wrench to tighten the SMA cable to 4 to 5-in lbs.
Use the adjustment tool that was shipped with your system.
c. Pres
s the sensor tip cable housing against the sensor head and then tighten
the two screws to 3 to 5-in lbs.
the adjustment tool that was shipped with your system.
Use
Figure 7: Connecting the sensor tip cable to the sensor head
TIVM Series Measurement System User Manual 11
Operating information
4. Connect the sen
This support keeps the sensor head steady reducing the potential mechanical stresses at the electrical connection point of the DUT. The support also keeps the sensor head away from surrounding circuits and conductive objects to minimize the parasitic capacitive coupling to these surroundings.
NOTE. Thematingthreadinthese
support, make s ure that mating thread is UNC¼-20.
Figure 8: Connect the sensor head to the exible tripod
sor head to the exible tripod or a similar support.
nsor head is UNC¼-20. If you use a different
Self calibration
NOTE. To obtain the most accurate measurement, allow the measurement system
to warm up for 20 minutes. Then perform the se the tip cable to the DUT and taking the measurement.
5. Connect the MMCX end, of the sensor tip cabl the DUT or to a square pin adapter on the DUT. The adapters connect to square pins with either 0.100-inch (2.54 mm) spacing or 0.062-inch (1.57 mm) spacing. (See Figure 28 on page 38.)
6. Set up the controls on the oscilloscope.
7. Apply power to the DUT to take the measurement.
Press the SELF CAL button on the controller to adjust the operating point of the measurement system for the current range and clamp setting. (This function is not available in the Probe Setup menu of the oscilloscope.)
NOTE. Make sure there is no differential voltage present at the sensor tip cable
when performing the self calibration.
lf calibration before connecting
e to an MMCX connector on
12 TIVM Series Measurement System User Manual
Operating information
Programming
When you power o the controller is orange, indicating that the operating point of the measurement system has not been optimized. As a result, the accuracy of the measurement system might be degraded. The SELF CAL sequence should always be run on the system after it is rst powered on and has warmed up for 20 minutes. After pushing the SELF CAL button, the indicator blinks orange during the self calibratio red when the operation fails.
There are s status indicator turns orange to indicate when this is necessary:
Users wanting to initiate the self calibration through the programmatic interface must congure the measurement system to perform a self calibration every time it receives the AutoZero command. To link these functions, hold the MENU button down and momentarily press the SELF CAL button. The OVERRANGE indicator should blink red twice. This mode is non-volatile, and it also changes the Restore the original operation by repeating the MENU-SELF CAL button pushes. In this case the OVERRANGE indicator will blink once.
n process; it turns solid green when the operation completes or solid
ituations when further self calibration is required. The SELF CAL
The measu
Changes are made to the range (1X|2X) or clamp (ON|OFF) setting.
The temperature in the sensor head changes more than 10 °C.
The internal compensation adjustments have shifted outside their normal operating ranges.
The sensor tip cable is changed.
operation of the AutoZero button in the Probe Setup menu of the oscilloscope.
n the measurement system, the SELF CAL status indicator on
rement system is rst attached to the oscilloscope.
TIVM Series Measurement System User Manual 13
Operating information
AutoZero
Menu button
When the displayed waveform is not centered correctly (for example, due to a small DC offset error), you might need to press the AutoZero button in the Probe Setup menu of the oscilloscope. This might be necessary for the rst time after the self calibration operation has completed. Make sure there is no differential signal pres
ent at the sensor tip cable.
Press the M oscilloscope, similar to the following gure.
Figure 9: Probe Setup menu
ENU button on the controller to view the Probe Setup menu on the
Use the buttons on the oscilloscope to change the probe setups. Some of the functions are the same as pushing the buttons on the controller, such as turning clamping on or off, or setting the input ranges.
14 TIVM Series Measurement System User Manual
Operating information
Offset correc
tion
The measurement system uses state-of-the-art technology that allows the DUT to be totally isolated from the oscilloscope. This results in a very large common mode rejecti be obscured by the h igh common mode interference.
The TIVM Ser any drift in the system due to changes in temperature or ber movement. The offset correction algorithm maintains a constant DC level for the displayed on-screen signal.
One of the components used for offset correction is a digital low-pass lter. Normally, the frequency of the signal is high enough that the operation of the lter is transparent. However, at low frequencies special care must be taken to prevent aberrations from being introduced.
When attempting to capture a signal that has a frequency near 43.5 Hz, 87.0 Hz, or
130.5 Hz, the display might show aliasing. If you cannot avoid these frequencies, the off selected to not alias at 50 Hz or 60 Hz.
on ratio (CMRR) and lets you see small signals that would otherwise
ies products implement a n offset correction algorithm to minimize
set correction should be disabled. The sample rate of the digital lter was
Figure 10: Digital lter aliasing
To disable the offset correction algorithm, hold down the MENU button on the controller and momentarily press the CLAMPING button. The OVERRANGE indicator on the controller should blink two times. To re-enable the correction, repeat the operation; the OVERRANGE indicator should blink once. Turning off the correction is temporary; when the measurement system is removed/re-attached, the offset correction is re-enabled.
TIVM Series Measurement System User Manual 15
Operating information
Another source
of aberrations is introduced when the signal is below the cutoff frequency of the digital lter (5.0 Hz). The following gure shows a close-up view of the front edge of a 0.1 Hz 800 mV
square wave. These aberrations are
p-p
also created by the digital lter.
Figure 11: Digital lter aberrations (about 2.5% of V
p-p
)
The following gure shows the aberrations after the offset correction has been disabled. When the offset correction is disabled, the system no longer corrects for long-term drift due to temperature change and ber movement.
16 TIVM Series Measurement System User Manual
Operating information
1X/2X Range
Figure 12
The top label of each sensor tip cable (IVTIP1X, IVTIP5X, IVTIP10X, IVTIP25X, IVTIP50X) shows the dynamic range in peak-to-peak volts as shown in the following gure.
Figure 13: Top s ensor tip labels
: Digital lter aberrations after disabling offset correction
TIVM Series Measurement System User Manual 17
Operating information
Auto Rang
For example, th
e IVTIP5X tip has a dynamic range of 5 V
when the 1X Range
p-p
is selected. This means that you can display a signal with a ±2.5 V differential voltage. When the 2X Range is selected, the dynamic range increases from 5 V to 10 V
(±5.0 V). Refer to the Linear differential input voltage range in the
p-p
p-p
specications table for more information.
CAUTION. To avoid damaging the probe, do NOT exceed either the RMS OR
Peak Voltag
e ratings. The Maximum Non-Destruct Voltage limits (RMS and Peak Vo ltage) do not increase when the 2X Range is selected. For the IVTIP5X, the 12 V
and ±21.5 Vpklimits are the same for the 1X or 2X Range.
rms
e
By default, the MSO/DPO5000, DPO7000, and MSO/DPO70000 Series Oscilloscopes select the 1X or 2X Range automatically when you change the V/div se However, there are combinations of Range and V/div settings that cannot be reached when Auto Range is selected. For these situations, select Manual Range when full exibility is desired.
tting. This hides the complexity of selecting the range from casual users.
18 TIVM Series Measurement System User Manual
Operating information
Selecting a se
nsor tip cable
CAUTION. Avoid over-voltage conditions that can damage or degrade the sensor
head input termination by selecting the wrong sensor tip cable. The sensor head SMA input is a 50 terminated input. Selecting the correct sensor tip cable attenuation factor is crucial to ensure that the sensor head input termination is not degraded or damaged by an over-voltage condition. Select the sensor tip cable that will p This also provides the highest differential impedance to the circuit-under-test.
When sele following questions:
What is t (for example, under a fault condition)?
What is can tolerate?
How la
What sensitivity do I need (for example, the minimum V/div setting)?
The following table will help you select the correct sensor tip. Start at the top of the table and work down. Choose the rst tip that meets all of your criteria.
rovide the highest attenuation possible for the signal being measured.
cting a sensor tip cable for a particular application, consider the
he maximum RMS/Peak Voltage at the test point being measured
the minimum differential loading (input resistance) that my circuit
rge of a signal do I want to display at one time on the oscilloscope?
le 6: Sensor tip selection table
Tab
Most sensitive
Sensor tip
IVTIP50X
IVTIP25X
IVTIP10X
IVTIP5X
IVTIP1X
1
On 1X range
2
On 2X range
V/div setting
50 mV/div ±50 V (100 V
25mV/div ±25V(50V
10mV/div ±10V(20V
5 m V/div ±5 V (10 V
1 mV/div ±1 V (2 V
Differential input specications
imum
Max
)
non-destruct voltage (RMS)
35 V
rms
25 V
rms
16 V
rms
12 V
rms
3V
rms
1
range
2
p-p
)
p-p
)
p-p
)
p-p
)
p-p
Linear voltage
imum
Max non-destruct voltage (Peak)
200 Vpk
107.5 Vpk
43 Vpk
21.5 Vpk
4.3 Vpk
Differential input resistance
2.5 k
1.25 k
500
250
50
TIVM Series Measurement System User Manual 19
Operating information
Output clampi
ng
Sensor tip loading
The measurement system has a selectable output clamping feature. Push the button on the controller to enable or disable the output clamping feature. When enabled (ind the measurement system into the oscilloscope input. It allows you to increase the vertical sensitivity without over-driving or saturating the oscilloscope input.
Each of the sensor tip cables (IVTIP1X, IVTIP5X, IVTIP10X, IVTIP25X, IVTIP50X) has a differential input resistance listed on the bottom labels as shown in the following gure.
icator is on), the output clamping limits the output voltage swing of
Figure 14: Bottom sensor tip labels
It is important to understand the impact of the tip loading on the DUT. For
mple, the IVTIP5X tip has a differential input resistance of 250 .Ifyou
exa choose the IVTIP25X tip, the differential input resistance increases to 1.25 kΩ. Refer to the Input resistance/capacitance section in the specications table for more information. The sensor tip cables are specially designed to act as a common mode choke that helps reduce the common mode loading.
NOTE. The coaxial (common) shield of the sensor tip cable should always be
onnected to the lowest impedance point (usually a circuit common or power
c supply rail) in the circuit-under-test (relative to the sensor tip cable/center conductor) to obtain the most accurate waveform.
20 TIVM Series Measurement System User Manual
Probe compensation
Deskew
Operating information
The DPO7000 and MSO/DPO70000 Series oscilloscopes have a Compensate Probe feature accessible from the Probe Setup window. Pressing this button always results in a failure because the input resistance of the TIVM sensor tips is too low and cannot be driven by the calibrator output of the oscilloscope. The accuracy of though the Compensate Probe feature fails.
Each oscilloscope family has a unique way of allowing you to adjust the timing relationship between signals acquired between different probes. Consult the user manual or online help o f the oscilloscope for specic directions needed to deskew the probe. The 3 m and 10 m measurement systems have a propagation delay of approximately 35 ns and 68 ns respectively. The actual propagation delay is measur
the TIVM system is still guaranteed on these oscilloscopes even
ed on each measurement system and is stored within each unit.
Input offset
The me This allows you to view a portion of the signal that is off-screen. One of the controls on the oscilloscope can be assigned this function.
The minimum/maximum offset is different for each sensor tip cable. It is the same when the 1X or 2X Range is selected; it is also the same for all V/div settings. Refer the following input offset table.
Table 7: Input offset
Se
IVTIP1X ±2 V
IVTIP5X ±10 V
VTIP10X
I
IVTIP25X ±50 V
IVTIP50X ±100 V
asurement system provides a user-adjustable, input referred offset voltage.
nsor tip cable
put offset voltage
In
20 V
±
TIVM Series Measurement System User Manual 21
Operating information
22 TIVM Series Measurement System User Manual
Application examples
The following examples are provided to help you become familiar with the TIVM Series IsoVu measurement system and to achieve the best performance for your application.
Example 1: H
igh-side V
measurement
GS
Advancements in the components used in switching power supplies have made characterizing the performance of these power supplies increasingly difcult and challengi V
GS
good CMRR is required from the test system. The following gure shows an example of this circuit.
Figure 15: Half-bridge circuit showing the gate, source, and drain of the high-side FET
ng. A particularly challenging measurement is measuring the high-side
in a half bridge. To accurately make this measurement, an exceptionally
In this type of circuit, the gate-source voltage is of interest because the rate at which the device switches is determined by the gate drive characteristics. The reference node for this measurement is the high-side source node, which switches
tween the input supply voltage and the local PCB ground during operation. In a
be measurement system without sufcient CMRR, this rapidly changing common mode voltage results in interference which obscures the measurement. It is important to note that the CMRR for all measurement systems is frequency dependent; however, the frequency that is critical for this measurement is not the switching frequency, but the frequency corresponding to the edge rate. For example, to accurately characteriz e a power supply with a switching frequency of 100 kHz and an edge rate of 1 ns, a system with good CMRR at 350 Mhz is necessary because of the edge speed.
TIVM Series Measurement System User Manual 23
Application examples
In this example
, the gate drive voltage could be about 5 V but usually has some ringing and overshoot which are important to characterize. For this measurement, it is appropriate to use a 10X tip with a 10 V
input (on the 1X range) so that
p-p
the signal is fully resolved and is within the dynamic range of the measurement system.
To get the best CMRR from your TIVM Series IsoVu measurement system, pay careful attention when connecting the measurement system to the DUT. This connection should preserve the signal delity and shield the signal from unwanted interfere
nce. To achieve the best performance from the measurement system, use an MMCX connector as close to the test point as possible. The MMCX connectors are available from a number of vendors and are relatively inexpensive. The key attributes of these connectors, which make them excel for this application, are their compact footprint and solid metal body. The solid metal body and gold contacts provide a w ell-shielded signal path.
The IsoVu input provides a oating differential measurement with differential input resistance from 50 to 2.5 k, depending on the tip attenuation. In the VGS mea
surement example the 10X tip with a 500 input impedance is used. The common mode resistance is extremely high, greater than G , in parallel with a small capacitance from the tip cable shield to ground, typically 2 pF or less. Keep these impedances in mind when determining how to connect the measurement system to the DUT. The source node in the h alf bridge circuit is a very low impedance point, and is the point that should be used to drive the tip
e shield capacitance. The gate driver output, which is also a low impedance
cabl node (but not as low impedance as the source), should be used to drive the center contact of the sensor tip cable. The input impedance of the sensor tip cable for the 10X tip is 500 relative to the shield – not to ground.
An example of a measurement that is possible with IsoVu is measuring the high-side turn-on characteristics shown in the following gure.
Figure 16: High-side turn-on characteristics
24 TIVM Series Measurement System User Manual
Application examples
Exampl
In general ther
e a re three characteristic regions of interest of the turn-on waveform
(See Figure 16.):
The rst regio
nistheC
charge time.
GS
The second region is the Miller Plateau (the time required to charge the gate-drain increases as V
The third re
Miller capacitance (C
increases.
DS
gion occurs when the channel is in conduction and the gate
chargesuptoitsnal value.
Due to the r
apid rise of the voltage on the switch node during the high side turn-on, there can be very high frequency and high amplitude common mode voltage changes during the transition. If this common mode voltage transient is not rejected, then the measurement of the high-side V not possible.
e 2: High-side drain current measurement
Current sensing is a critical measurement in many applications. Using the half bridge circuit again as an example, measuring the high-side drain current, I
y challenging particularly during startup. At startup, there can be common
be ver mode voltage transients due to parasitic inductance in the supply leads in addition to large current swings. Inserting a conventional current probe into the circuit at this point would require adding excessive inductance, which might limit circuit performance. Using a small value resistor as a current shunt makes it possible to take very high frequency current measurements with minimal additional
edance in the drain connection. (See Figure 17.)
imp
), and is VDSdependent. The charge time
GD
during the transition is
GS
,can
D
Figure 17: High side current shunt
TIVM Series Measurement System User Manual 25
Application examples
In a typical app current of 1 A, resulting in a voltage swing of 0.25 V, which can be measured with the measurement system using a 1X or 5X tip. A typical surface mount resistor can have a series inductance of less than 0.2 nH and series capacitance of less than
0.04 pF, resulting in much lower impedance at high frequency than could be obtained with a conventional current probe.
Figure 18: SMT resistor model
Go to http://www.vishay.com/docs/60107/freqresp.pdf for models of different types o
In general, surface mount resistors have fairly low power ratings; care must be take Several vendors make very high power parts that can be used when high power dissipation is required. For example, the RP0402CB-R500FN-2Q from Barry Industries (http://www.barryind.com/)isa1.0W,0.50402 resistor on an AIN substrate, or the RP0402CB-R500FN-2Q from US Microwaves (http://www.usmicrowaves.com/)isa1.5W,1resistor on a BeO substrate.
f surface mount resistors.
n not to exceed these ratings when using them as current shunts.
lication, a 0.25 resistor might be used to measure a transient
Example 3: ESD troubleshooting
Many devices and systems can be negatively affected by ESD discharges.
oubleshooting problems that occur during an ESD discharge can be very
Tr difcult. A piece of test equipment connected to a device undergoing ESD testing has to not only withstand the ESD discharge, but it also has to reject interference caused by the rapid changes in potential on the DUT during an ESD discharge test.
For example, consider using the standard human body model of 100 pF and 1500 . If the DUT is being tested to 4 kV and has a capacitance of 50 pF, the test equipment will be exposed to a voltage change of more than 1 kV in tens of nanoseconds.
26 TIVM Series Measurement System User Manual
Application examples
Figure 19: ESD discharge test example
In this example, if it is suspected that there is a glitch on the communication bus betwee the signals on the bus and to inspect them during the discharge. Because the measurement system uses Galvanic isolation, it rejects any interference from the ESD transient and tolerates the high common mode voltage during the discharge; the communication bus can be monitored throughout the ESD discharge and any irregularities can be investigated without interference from the ESD discharge.
n two devices during the ESD discharge, then it is helpful to connect to
TIVM Series Measurement System User Manual 27
Application examples
28 TIVM Series Measurement System User Manual
Reference information
Specication
Table 8: Warranted specications
Characteristic Description
Propagation delay (Warranted) 3 m ber length: 35 ns ±5 ns (actual propagation delay is measured and stored within each unit)
s
The following tables list the specications for the measurement system. The specications are guaranteed unless otherwise indicated.
The performance limits in this specication are valid with these conditions:
The instrument must be in an environment with temperature, altitude, and humidity within the operating limits described in these specications.
The instrument must have had a warm-up period of at least 20 minutes.
The measurement system is powered from a TekVPI compatible oscilloscope.
Warranted specications describe guaranteed performance with tolerance limits or certain type-tested requirements.
The performance verication procedures for the Propagation delay are listed later in this document. (See page 47, Propagation delay.)
10 m ber length: 68 ns ±7 ns (actual propagation delay is measured and stored within each unit)
le 9: Electrical specications
Tab
Characteristic Description
Controller output termination
Controller output coupling DC coupled
Range attenuation
Terminate the controller output into 50
Sensor tip cable/adapter
Sensor head input SMA 1X (÷1) 2X (÷2)
IVTIP1X, 1X Sensor tip cable 1X (÷1) 2X (÷2)
IVTIP5X, 5X Sensor tip cable 5X (÷5) 10X (÷10)
IVTIP10X, 10X Sensor tip cable 10X (÷10) 20X (÷20)
IVTIP25X, 25X Sensor tip cable 25X (÷25) 50X (÷50)
IVTIP50X, 50X Sensor tip cable 50X (÷50) 100X (÷100)
1X range 2X range
TIVM Series Measurement System User Manual 29
Reference information
Table 9: Electrical specications (cont.)
Characteristic Description
Input resistance/capacitance (attached to sensor head, 50 termination), (Ty pical)
Maximum non-destructive differential input voltage range, (Typical)
CAUTION. To avoid
damaging the measurement system, be aware of the input voltage limits; this specication applies to both the 1X Range and 2X Range.
Linear differential input voltage range, (Typical)
Output clamping range
(Input referred)
(Typ ical)
Output clamping overdrive recovery (Typical)
Sensor tip cable/adapter
Sensor head input SMA 50 ±2%
IVTIP1X, 1X Sensor tip cable 50
IVTIP5X, 5X Sensor tip cable 250
IVTIP10X, 10X Sensor tip cable 500
IVTIP25X, 25X Sensor tip cable 1.25 k
IVTIP50X, 50X Sensor tip cable 2.5 k
Sensor tip cable/adapter
Sensor head input SMA
IVTIP1X, 1X Sensor tip cable
IVTIP5X, 5X Sensor tip cable
IVTIP10X, 10X Sensor tip cable
IVTIP25X, 25X Sensor tip cable
IVTIP50X, 50X Sensor tip cable
Sensor tip cable/adapter
Sensor head input SMA
IVTIP1X, 1X Sensor tip cable
IVTIP5X, 5X Sensor tip cable
IVTIP10X, 10X Sensor tip cable
IVTIP25X, 25X Sensor tip cable
IVTIP50X, 50X Sensor tip cable
Sensor tip cable/adapter Clamping on,
Sensor head input SMA
IVTIP1X, 1X Sensor tip cable
IVTIP5X, 5X Sensor tip cable
IVTIP10X, 10X Sensor tip cable
IVTIP25X, 25X Sensor tip cable
IVTIP50X, 50X Sensor tip cable
<20ns
Resistance
V
rms
3 V 4.3 Vpk
3 V 4.3 Vpk
12 V 21.5 Vpk
16 V 43 Vpk
25 V 107.5 Vpk
35 V 200 Vpk
Clamping Off, 1X Range
±V peak (DC + peak AC)
±0.5 V ±1 V
±0.5 V ±1 V
±2.5 V ±5 V
±5 V ±10 V
±12.5 V ±25 V
±25 V ±50 V
1X Range
± 100 mV ± 200 mV
± 100 mV ± 200 mV
± 500 mV ± 1 V
±1V ±2V
±2.5V ±5V
±5V ±10V
Capacitance
NA
NA
< 1pF
< 1pF
< 1pF
< 1pF
V peak
Clamping Off, 2X Range
±V peak (DC + peak AC)
Clamping on, 2X Range
30 TIVM Series Measurement System User Manual
Table 9: Electrical specications (cont.)
Characteristic Description
System noise
(Typ ical)
DC gain accuracy1, (Input referred) (Typical)
Differential DC gain accuracy
80% to 100% of full scale in 2X Range:
DC offset error voltage
2
(Input referred)
(Typ ical)
Input offset voltage range (Typ ical)
Input offset voltage accuracy (Typical)
Small signal rise time (10% to 90%) (Typical)
(SMA input and with sensor tip cables)
Small signal frequency response
(Typ ical)
(SMA input and with sensor tip cables)
1 GHz system noise (input referred)
Sensor tip cable/adapter
Sensor head input SMA
IVTIP1X, 1X Sensor tip cable
IVTIP5X, 5X Sensor tip cable
IVTIP10X, 10X Sensor tip cable
IVTIP25X, 25X Sensor tip cable
IVTIP50X, 50X Sensor tip cable
±3% ± DC offset error voltage ± input offset accuracy error
±5% ± DC offset error voltage ± input offset accuracy error
Sensor tip cable/adapter
Sensor head input SMA
IVTIP1X, 1X Sensor tip cable
IVTIP5X, 5X Sensor tip cable
IVTIP10X, 10X Sensor tip cable
IVTIP25X, 25X Sensor tip cable
IVTIP50X, 50X Sensor tip cable
Sensor tip cable/adapter Input offset voltage range
Sensor head input SMA
IVTIP1X, 1X Sensor tip cable
IVTIP5X, 5X Sensor tip cable
IVTIP10X, 10X Sensor tip cable
IVTIP25X, 25X Sensor tip cable
IVTIP50X, 50X Sensor tip cable
±5%
TIVM1/TIVM1L
TIVM05/TIVM05L
TIVM02/TIVM02L
TIVM1/TIVM1L DC to 1GHz
TIVM05/TIVM05L DC to 500 MHz
TIVM02/TIVM02L DC to 200 MHz
1X Range 2X Range
<0.8mV
<0.8mV
<4mV
rms
<8mV
rms
<20mV
<40mV
rms
rms
rms
rms
<1.6mV
<1.6mV
<8mV
rms
<16mV
<40mV
<80mV
1X Range 2X Range
±2 mV ±4 mV
±2 mV ±4 mV
±10 m V ±20 mV
±20 m V ±40 mV
±50 m V ±100 mV
±100 mV ±200 mV
±2 V
±2 V
±10 V
±20 V
±50 V
±100 V
Rise time
350 ps
700 ps
1.8 ns
-3 dB bandwidth
Reference information
rms
rms
rms
rms
rms
TIVM Series Measurement System User Manual 31
Reference information
Table 9: Electrical specications (cont.)
Characteristic Description
Common mode rejection ratio (Typical)
Sensor tip cable/adapter DC
IVTIP1X, 1X Sensor tip
> 120 dB 120 dB 110 dB 100 dB 90 dB
cable
IVTIP5X, 5X Sensor tip
> 120 dB 120 dB 110 dB 100 dB 90 dB
cable
IVTIP10X, 10X Sensor
> 120 dB 120 dB 110 dB 100 dB 90 dB
tip cable
IVTIP25X, 25X Sensor
> 120 dB 110 dB 100 dB 100 dB 90 dB
tip cable
IVTIP50X, 50X Sensor
> 120 dB 100 dB 90 dB 90 dB 80 dB
tip cable
MMCX-to 0.1 in
> 120 dB 70 dB 60 dB 40 dB 30 dB (2.54 mm) square pin adapter with sensor tip cable.
MMCX-to 0.062 in
> 120 dB 70 dB 60 dB 40 dB 30 dB (1.57 mm) square pin adapter with sensor tip cable.
Common mode voltage
60 kV peak
3
range
Common mode resistance
N.A. due to Galvanically isolation (ber optic connection)
(Typ ical)
Common mode capacitance
4
~2 pF
(Typ ical)
Overload indicator range
(Typ ical)
Sensor tip cable/adapter Overload indicator On
Sensor head input SMA
IVTIP1X, 1X Sensor tip cable
IVTIP5X, 5X Sensor tip cable
IVTIP10X, 10X Sensor tip cable
IVTIP25X, 25X Sensor tip cable
IVTIP50X, 50X Sensor tip cable
1
The difference between the measured DC gain and the nominal DC gain, divided by the nominal DC gain and expressed as a percent.
2
The input referred offset error voltage when the input is shorted and the probe input offset is set to 0 volts
3
This product is designed to be used in the circuit with expected transient overvoltage up to 5000 V peak.
4
The capacitance between the sensor head and a reference plane. The sensor head is placed six inches (15.25 cm) above the reference plane.
100 MHz 200 MHz 500 MHz
<-3VorVin>+3V
V
in
<-3VorVin>+3V
V
in
<-12VorVin>+12V
V
in
<-16VorVin>+16V
V
in
<-25VorVin>+25V
V
in
V
<-35VorVin>+35V
in
1GHz
32 TIVM Series Measurement System User Manual
Reference information
Table 10: Physi
Characteristic Description
Net weight
Sensor tip cables 0.025 kg (0.055 lb)
Sensor head 0.363 kg (0.8 lb)
Controller box 0.816 kg (1.8 lb)
TekVPI Comp box 0.57 kg (0.125 lb)
Sensor tip cable length 15.24 cm (6.0 in)
Fiber cab
TekVPI cable length
Overal
le length
TIVM1, TIVM02, TIVM05
TIVM1L, TIVM02L, TIVM05L
l length and tolerances
Comp b
Controller to sensor head (TIVM1, TIVM02, TIVM05)
Controller to sensor head
VM1L, TIVM02L, TIVM05L)
(TI
cal specications
ox to controller
(Weight does not include accessories and packaging.)
3m(9.84
10 m (32.
55.88 c
0.558 length.
2.9718 m ±10.2 cm (117 in ±4 in)
9.982 m ±10.2 cm (393 in ±4 in)
ft)
81 ft)
m(22in)
8 m ±3.81 cm (22 in ±1.5 in) shoulder-to-shoulder, Boot area included in the overall
Dimensional drawings
Figure 20: Sensor head dimensions with probe tip cover
TIVM Series Measurement System User Manual 33
Reference information
Figure 21: Sensor head dimensions without probe tip cover
Figure 22: Controller dimensions
34 TIVM Series Measurement System User Manual
Figure 23: Comp box dimensions
Reference information
e 24: Probe tip adapter dimensions
Figur
TIVM Series Measurement System User Manual 35
Reference information
IsoVu measurement system block diagram
The following gure shows a block diagram of the IsoVu measurement system.
Figure 25: Block d iagram
The common mode resistance and capacitance to earth ground is shown in the gure. (See Figure 25.) The common mode resistance is shown as R parasitic and is essentially innite with the IsoVu measurement system since it is galvanically isolated and can be ignored. The common mode coupling capacitance to earth ground and the surrounding circuit is shown as the Parasitic Bridging Capacitance (Cpara sensor head is placed six (6) inches (15.25 cm) above a ground plane.
sitic). This parasitic capacitance will be approximately 2 pF when the
To m in following items:
imize the effects of common mode capacitive loading consider the
ever possible, choose a reference point in the circuit-under-test that is
When static potential with respect to earth ground.
nect the coaxial (common) shield of the sensor tip cable to the lowest
Con impedance point of the circuit.
creasing the physical distance between the sensor head and any conductive
In surface will reduce the parasitic capacitance.
hen using multiple IsoVu systems to measure different points in the circuit
W that do not have the same common mode voltages, keep the sensor heads separated to minimize the capacitive coupling.
36 TIVM Series Measurement System User Manual
Tripods
Reference information
Tektronix provides two tripods as accessories with the measurement system. The Flexible tripod holds the sensor head while connected to the DUT. The Probe tip tripod supports the sensor tip cables while they are connected to adapters on the circuit board.
Flexible tr
ipod
The tripod can be connected to the DUT in different ways. You can secure the tripod to the DUT using optional feet. T he feet can be clamped or attached to the DUT using common screws. This allows you to install the tripod upside down or right side
Figure 26: Installing the exible tripod under the DUT.
up as shown in the following gures.
Figure 27: Connecting the sensor head to the top of the DUT with the exible tripod.
Probe tip tripod
TIVM Series Measurement System User Manual 37
Use the probe tip tripod to connect the sensor tip cables to adapters on the circuit board. This tripod has exible hinges to easily position the probe tip cable above the adapter on the circuit board. Tektronix recommends gluing the tripod in place on the circuit board to provide additional support for the sensor tip cables. The following gure shows an example of connecting to an adapter on the circuit board with the tripod; it reduces stresses on the test point.
Reference information
Figure 28: Connecting to an adapter on the circuit board with the probe tip tripod
38 TIVM Series Measurement System User Manual
Installing the probe tip adapters
Tektronix provides two probe tip adapters to connect the sensor tip cables to pins on the circuit board. The MMCX-to-0.1-inch (2.54 mm) pitch adapter and the MMCX-to-0.062-inch (1.57 mm) pitch adapter.
One end of each adapter has an MMCX socket for connection to an IsoVu tip cable. The other end of the adapter has a center pin socket and four common (shield) sockets around the outside of the adapter. Notches on the adapters can be used to loc the same, the main difference is the spacing of the pins on the circuit board.
ate the shield sockets. The procedure for installing these adapters are
Reference information
To insta l the signal source pin on the circuit board. Use the notch on the adapter to align one of the shield sockets to the common pin on the circuit board. The following gures show examples of lining up the adapters on the circuit board.
To achieve the best electrical performance, especially the CMRR performance and EMI susceptibility, place the probe tip adapter as close as possible to the circuit board.
l the adapters onto the square pins, line up the center of the adapter with
Figure 29: Lining up the MMCX-to-0.1-inch (2.54 mm) adapter on the circuit board
TIVM Series Measurement System User Manual 39
Reference information
Figure 30: Lining up the MMCX-to-0.062-inch (1.57 mm) adapter on the circuit board
After lining up the adapters gently push down on the adapter to seat it in place on the circuit board.
Figure 31: Pushing the MMCX-to-0.062-inch (1.57 mm) adapter in place
40 TIVM Series Measurement System User Manual
Reference information
Figure 32: Pushing the MMCX-to-0.1-inch (2.54 mm) adapter in place
When the adapters are rmly in place on the circuit board connect the sensor tip cable to the top of the adapter while using the probe tip tripod to ease the tension off the probe tip cable and adapter. (See Figure 28 on page 38.)
TIVM Series Measurement System User Manual 41
Reference information
Installing the square pins on the circuit board
The following gure shows the recommended cle arance r equire ments for connecting the adapters to the square pins on the circuit board. The bottoms of the adapters are shown at the top.
Figure 33: Adapter clearance requirements
42 TIVM Series Measurement System User Manual
Reference information
The 0.025-inch board. Some square pins might have headers installed on the circuit board. Tektronix recommends removing the plastic spacer from the square pins to gain closer access to the circuit board as shown in the following gure to achieve the best electrical performance, especially the CMRR. You might need to use a pair of tweezers to remove the spacer as shown in the gure.
(0.635 mm) square pins should already be located on the circuit
34: Removing the header from square pins on the circuit board
Figure
onix provides a set of solder pins (0.018-inch (0.46 mm) diameter) to install
Tektr on the circuit board for use with the MMCX to 0.062-inch (1.57 mm) adapter. Use the soldering aide tool accessory (Tektronix part number, 003-1946-xx) to install these pins on the circuit board.
NOTE. The solder pins are extremely small and can be challenging to handle.
Tektronix recommends using tweezers and a magnifying tool when installing
pins on the circuit board.
the
The solder pins can be installed around a surface mounted component on the
rcuit board, but adequate clearance should be maintained for a good electrical
ci connection for the adapter. (See Figure 33 on page 42.)
OTE. The coaxial (common) shield of the sensor tip cable and tip adapters
N
should always be connected to the lowest impedance point (usually a circuit common or power supply rail) in the circuit-under-test (relative to the sensor tip cable/center conductor) to obtain the most accurate waveform.
TIVM Series Measurement System User Manual 43
Reference information
Use the followi circuit board:
1. Carefully ins following illustration.
ng steps to install the solder pins using the soldering aide on the
ert t he solder pins into the soldering aide as shown in the
Figure 35: Using the soldering aide to install the square pins on the circuit board
2. Use the soldering aide to hold the square pins in place while soldering the square pins to the circuit board.
3. If necessary apply a small amount of adhesive to further strengthen the connection to the c i rcuit board. However, keep the height of the adhesive to a minimum to provide good electrical contact for the adapter. (See Figure 33.)
44 TIVM Series Measurement System User Manual
User service
Service offerings
Tektronix provides service to cover repair under warranty and other services that are designed to meet your specic service needs.
Whether providing warranty repair service or any of the other services listed below, Tektronix service technicians are well equipped to service the IsoVu measurement system. Services are provided at Tektronix Service Centers and on-site at your facility, depending on your location.
Warranty
repair service
Calibration and repair
service
Preventive maintenance
Tektronix warrants this product as described in the warranty statements at the front of this manual. Tektronix technicians provide warranty service at most Tektronix service locations worldwide. The Tektronix Web site provides information on all service
In addition to warranty repair, Tektronix Service offers calibration and other services that provide cost-effective solutions to your service needs and quality stand worldwide by the leading-edge design, manufacturing, and service resources of Tektronix to provide the best possible service.
CAUTION. To prevent damage to the measurement system, do not expose it to
spr head when cleaning the exterior.
locations worldwide.
ards compliance requirements. Tektronix instruments are supported
ays, liquids, or solvents. Avoid getting moisture inside the controller or sensor
TIVM Series Measurement System User Manual 45
User service
Clean the exter
ior surfaces with a dry, lint-free cloth or a soft-bristle brush. If dirt remains, use a soft cloth or swab dampened with a 75% isopropyl alcohol solution. Use only enough solution to dampen the cloth or swab. Do not use abrasive compounds on any part of the instrument.
Performance verication procedures
Use the following procedures to verify the performance of the IsoVu measurement system. Before beginning the procedures, photocopy the test record and use it to record the performance results. (See page 53, Test record.)
Required e
quipment
Table 11: Required equipment for performance verication
Description Minimum requirements Example product
Oscilloscope with TekVPI interface 1GHz,50input support, fully compatible
Pulse generator 1 V
TIVM series 1X sensor tip cable As per description Tektronix IVTIP1X
MMCX Jack (female) to BNC female adapter
The equipment r equired to perform the performance verication procedures are shown in the following table.
Tektronix MDO4104C
with TekVPI interface
, <1ns risetime
p-p
As per description
Tektronix Tek-DPG
Fairview Microwave Product SKU: SM3610
Preparation
Prepare the equipment as follows:
1. Tur
n on the TekVPI oscilloscope.
2. Connect the Tek-DPG Deskew pulse generator into Channel 2 of the TekVPI
cilloscope.
os
3. Allow the test equipment to warm up for 20 minutes at an ambient temperature
f20°Cto30°C.
o
46 TIVM Series Measurement System User Manual
Propagation delay
User service
This procedure veries that the TIVM Series IsoVu measurement system is functioning and meets the warranted propagation delay specication. The propagation delay will be measured by rst applying the output of the pulse generator to the input of the oscilloscope and storing the captured waveform as a referen oscilloscope and the output of the pulse generator is connected to the input of the measurement system. The delay is then measured between the saved reference waveform and the acquired waveform of the measurement system.
NOTE. This procedure is valid for all versions of the TIVM Series IsoVu
measurement systems.
ce waveform. Next, the measurement system is connected to the
Create t he reference
waveform
Complete the following steps to create the reference waveform:
1. Connect the Tek-DPG Deskew Pulse Generator output BNC cable directly to Channel 1 input of the TekVPI oscilloscope.
2. Enable Channel 2 and use the following setups:
Vertical Scale: 500 mV/div.
Vertical Position: -3 divisions
Set Termination: 1MΩ,
Coupling: DC
Bandwidth: FULL
skew: 0 seconds
De
3. Set the Trigger menu using the following setups:
Type: Edge
Source: CH2
Slope: Negative
Level: +1.50 V
Coupling: DC
4. Use the following setups for Channel 1.
Vertical Scale: 200 mV/div
Coupling: DC
Termination: 50
Bandwidth: FULL
TIVM Series Measurement System User Manual 47
User service
Position: 0 (ce
Offset: -500 mV
Deskew: 0 seconds
5. Set the Horizontal menu to the following setups:
Horizontal Scale to 10 ns/div.
Horizontal Position: 40 ns
Acquire: Average 128
6. Set the Tek-DPG to the following setups:
0to-11kHzmode(Mode 1)
Output Enable: ON
7. The Channel 1 waveform should appear approximately on the center of the
oscilloscope display. If the waveform is not vertically centered in the display, adjust the display as necessary to vertically center the waveform.
8. Use the following steps to save the Channel 1 waveform as the reference waveform (R1).
Press MENU on the oscilloscope.
ntered)
Create the TIVM series
waveform
Select Save Waveform.
Source: CH1.
Destination: R1.
Select OK Save to save the Channel 1 waveform as reference waveform
R1. The new reference waveform should now be displayed on the oscilloscope.
9. Disable the output of the Tek-DPG.
10. Disconnect the Tek-DPG Deskew Pulse Generator output BNC cable from
the Channel 1 input of the TekVPI oscilloscope.
Complete the following steps to set up the TIVM series waveform:
1. Connect the Comp box of the TIVM series measurement system to Channel 1 of the TekVPI oscilloscope.
2. Connect a IVTIP1X sensor tip cable to the TIVM series input (tighten the SMA connector from the cable to the sensor head and attach the nose cone.)
3. Allow the measurement system to warm up for 20 minutes at an ambient temperature of 20 ° to 30 °C.
4. Connect an MMCX Jack (female) to BNC female adapter to the output BNC cable of the Tek-DPG.
48 TIVM Series Measurement System User Manual
User service
Measure the propagation
delay
5. Set the TIVM ser
Range: 2X
CLAMPING: Off
Push the SELF CAL button to perform the self calibration (wait for the
SELF CAL Status indicator to turn solid green).
6. On Channel 1 of the oscilloscope set the Vertical Scale to 200 mv/div.
7. Connect the IVTIP1X sensor tip cable to the MMCX Jack (female) to BNC
female adapter.
8. Enable the output of the Tek-DPG.
9. The Channel 1 waveform should appear on the oscilloscope display and be
approximately vertically centered on the d isplay. If not, ONLY adjust the vertical settings as necessary to center the waveform on the display.
Complete the following steps to measure and record the propagation delay.
1. Set up t
a. Select Measure.
he delay measurement on the oscilloscope as follows:
ies to the following:
b. Select Add Measurement.
c. Select Measurement Type: Delay.
d. Select Congure: Delay.
e. Set the Source to R1.
f. Set
g. Select OK Add Measurement.
DelayTotoCH1.
TIVM Series Measurement System User Manual 49
User service
2. The oscillosco
pe display should be similar to the following gure.
Figure 36: Propagation delay measurement
Record the resulting delay measurement on the test record.
3. Disable the output of the Tek-DPG.
Troubleshooting and error conditions
The following table lists possible problems that you might encounter when taking measurements with TIVM Series IsoVu measurement system. Use the table as a quick troubleshooting reference before contacting Tektronix for service.
Table 12: Problems and possible solutions
Problem Remedy
The measurement system will not power on; no indicators are on.
Controller STATUS indicator is ashing green.
Controller STATUS indicator is ashing red and yellow
Verify the TekVPI comp box is securely attached to the oscilloscope. Detach and then reattach the comp box (using a different oscilloscope channel if necessary). If the condition persists, return the system to Tektronix for service.
Detach and then reattach the comp box (using a different oscilloscope channel if necessary). Do not force the comp box into the oscilloscope connector at an angle; connect it to the oscilloscope with a rm, horizontal force. If the condition persists, return the system to Tektronix for service.
This condition indicates a failure in the measurement system. Most often, a failure is detected when the m easurement system is rst connected to the oscilloscope (while executing power-on self tests). Detach and then reattach the comp box to the scope. If the condition persists, return the system to Tektronix for service.
50 TIVM Series Measurement System User Manual
Table 12: Problems and possible solutions (cont.)
Problem Remedy
Controller SELF CAL STATUS indicator turns solid red after pressing SELF CAL button (or never completes).
Waveform distortion when the signal from the D UT is less than 5 Hz or a multiple of 43.5 Hz
Output waveform is clipped or distorted
Frequency response is rolled off Check the bandwidth limit on the oscilloscope.
Remove input signal (differential must be 0.0V).
Allow system to warm up 20 minutes.
Make sure the ber cable is not subjected to dynamic mechanical or thermal stresses.
If the self calibration does not complete in a minute, detach and reattach the comp box to
the oscilloscope – then try again.
Disable the Offset Correction. Refer to the Offset correction section earlier in this document. (See page 15.)
Verify that the Output Clamp is disabled.
Check that the proper sensor tip cable is being used for the measurement. Refer to
Selecting a sensor tip cable section earlier in this document. (See page 19.)
Change the Input Offset to center the signal on-screen.
Make sure the correct Range is selected (1X or 2X).
User service
Check the sensor tip cable continuity and sensor head input resistance (see below).
Output waveform is unstable (low frequency noise and/or DC offset keep changing)
Measurements are inaccurate
Cannot achieve high CMRR Try a different sensor tip cable.
Too much noise and cannot accurately resolve signal
Make sure the ber cable is not subjected to dynamic mechanical and thermal stresses.
Verify that Offset Correction is enabled (detach and then reattach the comp box to make
sure it has been re-enabled).
Run SELF CAL.
Check that the proper sensor tip cable is being used for the measurement. Refer to
Selecting a sensor tip cable section earlier in this document. (See page 19.)
Make sure the correct Range is selected (1X or 2X).
Make sure the signal is on-screen.
Switch to the 1X Range.
Choose a sensor tip cable with less attenuation.
TIVM Series Measurement System User Manual 51
User service
Table 12: Problems and possible solutions (cont.)
Problem Remedy
No signal detected; waveform is a at line
Large DC offset in waveform Run SELF CAL.
Check the sensor tip cable continuity:
Sensor tip cable
IVTIP1X - 1X sensor tip cable
IVTIP5X - 5X sensor tip cable
IVTIP10X - 10X sensor tip cable
IVTIP25X - 25X sensor tip cable
IVTIP50X - 50X sensor tip cable
Measure the SMA input resistance of the Sensor Head using a 4-wire measurement; it
should be between 47.5 and 52.5 . If not, then the Sensor Head has been damaged and needs to be returned to Tektronix for service.
Check the controller STATUS indicator for possible error conditions.
Apply a test signal directly to the SMA input of sensor head to determine if the problem is
the sensor tip cable or the sensor head.
Resistance
0.2
200 ±2%
453 ±2%
1.21 k±2%
2.49 k±2%
Run AutoZero.
Set Input Offset to 0.0 V.
Cannot select between 1X and 2X range (5000/7000/70000 Series oscillocopes)
Probe compensation fails (7000/70000 Series oscilloscopes)
When Auto Range is turned ON, the Range is selected automatically when the V/div setting is changed. You cannot change the Range directly (the button will appear to not function).
This is a normal condition. You can ignore the failure.
Repack the measurement system for shipment
If you need to return the measurement system to Tektronix for repair, use the original packaging. If this is unavailable or not t for use, contact your Tektronix representative to obtain new packaging.
When you return the measurement system to Tektronix, attach a tag showing the following information:
Name of the product owner
Address of the owner
Instrument serial number
A description of problems encountered and/or service required
52 TIVM Series Measurement System User Manual
Test record
Table 13: Test re cord
User service
Photocopy this test record for recording the results of the performance verication procedures.
Model number:
Serial numb
Temperature:
Propagatio
TIVM1(3m
TIVM1L (10
TIVM02 (3
TIVM02L
TIVM05 (
TIVM05
er:
n delay
m ber length)
m ber length)
(10 m ber length)
3mber length)
L(10mber length)
ber length)
Certicate
RH %:
Technician:
Date of cali
Minimum Incoming Outgoing Maximum
30 ns 40 ns
61 ns
30 ns 40 ns
61 ns
30 ns 40 ns
61 ns
number:
bration:
75 ns
75 ns
75 ns
TIVM Series Measurement System User Manual 53
User service
54 TIVM Series Measurement System User Manual
Appendix A: Remote programming
This appendix describes commands and queries that can be sent to the sens or head when attached to a Tektronix oscilloscope. Long-form and short-form keywords are indicated with upper/lower case letters. The commands and queries are supported by most oscilloscopes; differences in s upporting oscilloscopes, if any, are des
For details on the command syntax, refer to the programmer documentation for your oscil
CH<n>:PRObe?
cribed with the commands.
loscope.
Returns p
robe information for channel <n>. Query only.
CH<n>:PRObe:AUTOZero EXECute
This com by the measurement system and then by the oscilloscope. Command only.
The sys calibration before the oscilloscope does its AutoZero.
Refer calibration. (See page 12, Self calibration.)
mand executes the A utoZero function. The operation is rst performed
tem can be congured (with special keystrokes) to perform a full self
to the self calibration procedure for information on performing the self
CH<n>:PRObe:COMMAND “CLAMP”, {“ON” | “OFF”}
Only supported by 3000/4000 Series oscilloscopes.
command enables or disables the clamp circuitry. The parameters enclosed in
The quotes are case sensitive and must be sent as capital letters.
<n>:PRObe:COMMAND? “CLAMP” . The query returns either “ON” or
CH “OFF” enclosed in quotes.
CH<n>:PRObe:SET {“CLAMP ON” | “CLAMP OFF”}
Only supported by 5000/7000/70000 Series oscilloscopes.
The command enables or disables the clamp circuitry. The parameters enclosed in quotes are case insensitive.
CH<n>:PRObe:SET?. The query returns either “CLAMP ON” or “CLAMP OFF” enclosed in quotes. The case is the same as the last command received.
TIVM Series Measurement System User Manual 55
Appendix A: Remote programming
CH<n>:PRObe:
FORCEDRange <NR3>
CH<n>:PRObe:GAIN?
The command selects the dynamic range of the sensor tip in V
and is dependent
p-p
the attached sensor tip cable.
The following table lists the sensor tip cables and the dynamic range.
Table 14: Sensor tip cables and dynamic ranges
Sensor tip cable Range V
1X 1.0 or 2.0
5X 5.0 or 10.0
10X 10.0 or 20.0
25X 25.0 or 50.0
50X 50.0 or 100.0
p-p
CH<n>:PRObe:FORCEDRange? The query returns the dynamic range of the sensor tip in V
p-p
.
Returns the gain factor of the sensor tip (inverse of attenuation). Dependent on the
hed sensor tip cable. Query only.
attac
CH<n>:PRObe:ID {:SERnumber | :TYPe}?
y only. When sending only PRObe:ID?, the query returns the sensor tip type
Quer string followed by the serial number string.
n sending PRObe:SERnumber?, the query returns the serial number string.
Whe
When sending PRObe:TYPe?, the query returns one of the following sensor tip
pe strings (note the trailing spaces within the return strings):
ty
“TIVM1 ”(1GHz, 3m)
“TIVM1L ”(1GHz, 10m)
“TIVM05 ”(500MH z, 3m)
“TIVM05L”(500MHz, 10m)
“TIVM02 ”(200MH z, 3m)
“TIVM02L”(200MHz, 10m)
56 TIVM Series Measurement System User Manual
Appendix A: Remote programming
CH<n>:PRObe:
PROPDELay?
Only supported by 3000/4000 Series oscilloscopes.
Returns the propagation delay value (in seconds). Query only.
CH<n>:PRObe:RECDESkew?
ly supported by 3000/4000 Seriesoscilloscopes.
On
Returns the recommended deskew value (in seconds). Query only.
CH<n>:PRObe:RESistance?
Returns the input resistance (in ohms). It depends on the attached sensor tip cable. Query only.
CH<n>:PRObe:UNIts?
Returns the units of the sensor tip (always “V”). Query only.
CH<n>:PROBECOntrol {AUTO | MAN}
Only supported by 5000/7000/70000 Series oscilloscopes.
The command sets the Auto or Manual sensor tip range control.
CH<n>:PROBE:PROBECOntrol?. The query returns the keyword AUTO or MANUAL.
CH<n>:PROBEFunc:EXTAtten <NR3>
Only supported by 5000/7000/70000 Series oscilloscopes.
The command sets user-dened external attenuation factor.
CH<n>:PROBEFunc:EXTAtten?. The query returns the external attenuation factor.
CH<n>:PROBEFunc:EXTDBatten?
Only supported by 5000/7000/70000 Series oscilloscopes.
Returns the external attenuation factor expressed in dB. Query only.
TIVM Series Measurement System User Manual 57
Appendix A: Remote programming
CH<n>:PROBEF
unc:EXTUnits {“UU” | “None”}
Only supported by 5000/7000/70000 Series oscilloscopes.
The command set the user-dened units. Only two characters are displayed by the oscilloscope. Entering “None” resets the units to default.
CH<n>:PROBEFunc:EXTUnits?. The query returns value has been entered.
“V” unless a user-dened
58 TIVM Series Measurement System User Manual
Appendix B: Compliance information
This section lists the Safety and Environmental standards with which the instrument complies.
Safety compliance
This section lists the safety standards with which the product complies and other safety compliance information.
EU low voltage directive
U.S. nationally recognized
testing laboratory lis ting
Canadian certication
Compliance was demonstrated to the following specication as listed in the Ofcial Journal of the European Union:
Low Voltage Directive 2014/35/EU.
EN 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
EN 61010-031. Particular requirements for handheld probe assemblies for electrical measurement and test eq uipment (partially applicable).
UL 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
UL 61010-031. Particular requirements for handheld probe assemblies for electrical measurement and test eq uipment (partially applicable).
CAN/CSA-C22.2 No. 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
CAN/CSA-C22.2 No. 61010-031. Particular requirements for handheld probe assemblies for electrical measurement and test equipment (partially applicable).
Additional compliances
TIVM Series Measurement System User Manual 59
IEC 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
UL 61010-031. Particular requirements for handheld probe assemblies for electrical measurement and test eq uipment (partially applicable).
EN 60825-1. Safety of Laser Products-Part 1: Equipment Classication a nd Requirements - Edition 2 2007
US 21CFR PT1010 Performance Standard for Electronic Parts 2015.
US 21CFR PT1040 Performance Standards for Light Emitting Products 2015.
Appendix B: Compliance information
Equipment type
Pollution degree
descriptions
Test and measur
A measure of the contaminants that could occur in the environment around and within a product. Typically the internal environment inside a product is considered to be the same as the external. Products should be used only in the environment for which they are rated.
Pollution degree 1. No pollution or only dry, nonconductive pollution occurs. Products in this category are generally encapsulated, hermetically sealed, or located in clean rooms.
Pollution degree 2. Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is caused by condensation must be expect condensation occurs only when the product is out of service.
Polluti that becomes conductive due to condensation. These are sheltered locations where neither temperature nor humidity is controlled. The area is protected from direct sunshine, rain, or direct wind.
Pollution degree 4. Pollution that generates persistent conductivity through conductive dust, rain, or snow. Typical outdoor locations.
ing equipment.
ed.Thislocationisatypicaloffice/home environment. Temporary
on degree 3. Conductive pollution, or dry, nonconductive pollution
Pollution degree rating
IP rating
Measurement and
overvoltage category
scriptions
de
Pollution degree 2 (as dened in IEC 61010-1). Rated for indoor, dry location use only.
0(asdefined in IEC 60529).
IP2
Measurement terminals on this product may be rated for measuring mains voltages from one or more of the following categories (see specic ratings marked on the product and in the manual).
Category I. Circuits not directly connected to a mains supply.
Category II. Circuits directly connected to the building wiring at utilization points (socket outlets and similar points).
Category III. In the building wiring and distribution system.
Category IV. At the source of the electrical supply to the building.
NOTE. Only measurement circuits have a measurement categor y rating. Other
circuits within the product do not have either rating.
60 TIVM Series Measurement System User Manual
Environmental considerations
This section provides information about the environmental impact of the product.
Appendix B: Compliance information
Product end-of-life
handling
Observe the f
ollowing guidelines when recycling an instrument or component:
Equipment recycling. Production of this equipment required the extraction and use of natural resources. The equipment may contain substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. To avoid release of such substances into the environment and to reduce the
use of n atural resources, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
This symbol indicates that this product complies with the applicable European Union re on waste electrical and electronic equipment (WEEE) and batteries. For information about recycling options, check the Tektronix Web site (www.te
quirements according to Directives 2012/19/EU and 2006/66/EC
k.com/productrecycling).
TIVM Series Measurement System User Manual 61
Appendix B: Compliance information
62 TIVM Series Measurement System User Manual
Index
Symbols and Numbers
1X Range, 18 2X Range, 18
A
aberrations, 15, 16 accessories
optional, 2 standard
adapter
clearance requirements, 42
TCA-VPI50, xi aliasing, 15 application examples, 23 Auto Ra AutoZero, 14
,1
nge, 18
B
bend radius
ber-optic cables, 2
sensor tip cables, 2 block diagram, 36
C
ble care, 2
ca cable handling practices, 2 calibration status, 8 clamping button, 8 clamping indicator, 8 cleaning procedures, 45 clearance requirements, 4
adapters, 42 CMRR, 15, 23 comp box, x
dimensions, 34 compensate probe feature
DPO7000,
MSO/DPO70000 series oscilloscopes, 21
compliance
environmental, 61
safety, 59
connecting the measurement
system, 11
controller, x
dimensions, 34
controller
controller indicators
current sensing example, 25 cutof
buttons clamping, 8 menu, 7, 14 range, 7 self cal, 8
clamping over range, 7 range, 7 self cal, 8 status, 8
f frequency, 16
,8
D
derating curve, 4 deskew, 21 differential dynamic range, 17 dimensions
comp box, 34
ntroller, 34
co probe tip adapter, 35 sensor head, 33
E
environmental compliance, 61 environmental considerations, 3 error conditions, 50 ESD discharge example, 26
F
ber-optic cable
safe handling practices, 2
exible tripod, 37
I
input offset, 21
input resistance
sensor tip cables, 20 input specications, 3 IsoVu, ix
K
key featur
es, ix
L
laser certication, ix low-pass lter, 15
M
measurement system status, 8 menu button, 7, 14 MMCX connector, xi, 12, 24
s, xi
model
O
offset correction, 15
disabling, 15 operating requirements, 2 optional accessories, 2 output clamping, 20
r range Indicator, 7
ove
P
packaging, 52 performance verication
procedures, 46
propagation delay, 47
required equipment, 46
test record, 53 probe compensation failures, 21 probe tip adapter
dimensions, 35 probe tip adapters, 39 probe tip tripod, 37 product description, x programmatic interface, 13 propagation delay, 21
TIVM Series Measurement System User Manual 63
Index
R
range button, 7 range indicators, 7 remote programming
CH<n>:PRObe CH<n>:PRObe:
FORCEDRange, 56
CH<n>:PRObe:AUTOZero
EXECute, 55
CH<n>:PRObe:COMMAND
“CLAMP”, 5 CH<n>:PRObe:GAIN?, 56 CH<n>:PRObe:ID?, 56 CH<n>:PRObe:PROPDELay?, 57 CH<n>:PRObe:RECDESkew?, 57 CH<n>:PRObe:RESistance?, 57 CH<n>: CH<n>:PRObe:UNIts?, 57 CH<n>:PROBECOntrol, 57 CH<n>:PROBEFunc:
CH<n>:PROBEFunc:
CH<n>:PROBEFunc:
repackaging, 52 RF burn area, 6 RF burns, 4
PRObe:SET, 55
EXTAtten, 57
EXTD
EXTUnits, 58
?, 55
5
Batten?, 57
S
safety compliance, 59 safety infor safety summary, v self cal
AutoZero, 14 button, 8 indicator, 8, 13 programmi
requirements, 13 self calibration, 12 sensor head, 9
description, xi
dimensions, 33
labels, sensor tip
labels, 9, 17, 20 sensor tip cable
installation, 11
selection considerations, 19
or tip cables, vi, xi, 9
sens
bend radius, 2 service offerings, 45 solder aid, 1 solder aide, 43 solder pin installation, 43
ecications, 29
sp square pin adapters, 12
mation, v
ng, 13
9
standard acces status indicator, 8 supported oscilloscopes, xi
sories, 1
T
TCA-VPI50 adapter, xi test record, 53 tip cables, 9 TIVM02, xi TIVM02L, xi TIVM05, xi TIVM05L, xi TIVM1, xi TIVM1L, xi
,37
tripods troubleshooting, 50
U
user service, 45
W
warranty repair service, 45
64 TIVM Series Measurement System User Manual
Loading...