Tektronix TIVH02, TIVH02L, TIVH05, TIVH05L, TIVH08 User Manual

...
TIVH Series
xx
ZZZ
IsoVu™ Measurement System
User Manual
*P071355602*
071-3556-02
xx
TIVH Series
ZZZ
IsoVu™ Measurement System
Register now!
Click the following link to protect your product.
www.tek.com/register
www.tek.com
071-3556-02
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
ISOVU is a trademark of Tektronix Inc.
TEKVPI is a registered trademark of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. B o x 5 0 0 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200. Worldwide, visit www.tek.com to nd contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty work may be n the property of Tektronix.
ew or reconditioned to like new performance. All replaced parts, modules and products become
In order to o the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage result b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has b een modied or integrated with other products when the effect of such modication or integration increases the time or difculty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
TRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE
TEK AND EXCLUSIVE REMEDY PR OVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
[W2 – 15AUG04]
btain service under this warranty, Customer must notify Tektronix of the defect before the expiration of
ing from attempts by personnel other than Tektronix representatives to install, repair or service the product;
Table of Contents
Important safety information ............. ................................ .................................. ....... v
General safety summary ...................................................................................... v
Service safety summary.............. ................................ .................................. ..... vii
Terms in this manual ................... ................................ .................................. ... viii
Symbols and terms on the product......................................................................... viii
Preface .............................................................................................................. ix
Key features ........................... ................................ .................................. ...... ix
Laser certication................. ................................ ................................ ............ ix
Product description ... ................................ ................................ ......................... x
Models ......................................................................................................... xi
Supported oscilloscopes..................................................................................... xii
Operating information..................................... .................................. ....................... 1
Accessories..................................................................................................... 1
Operating considerations...................................................................................... 3
Controls and indicators................... .................................. ................................ ... 8
Connecting to a circuit ............... .................................. ................................ ...... 11
Self calibration................................................................................................ 14
AutoZero....................................................................................................... 15
Menu button ........................... ................................ ................................ ........ 15
Offset correction .............................................................................................. 16
1X/2X Range.................. ................................ ................................ ................ 19
Auto Range.................................................................................................... 19
Selecting a sensor tip cable .................................................................................. 20
Output clamping ........ ................................ ................................ ...................... 21
Sensor tip loading............................................................................................. 21
Probe compensation .......................................................................................... 22
Deskew......................................................................................................... 22
Input offset .... .................................. ................................ .............................. 22
Input coupling (AC or DC) .................................................................................. 22
TIV-Series voltage range..................................................................................... 23
Application examples ........... .................................. ................................ ................ 35
Example 1: High-side V Example 2: High-side V
Example 3: High-side drain current measurement .......................... .............................. 38
Example 4: ESD troubleshooting ............. .................................. ............................ 39
Reference information ............................................................................................ 41
Specications.................................... ................................ .............................. 41
Dimensional drawings........................................................................................ 50
IsoVu measurement system block diagram ........ .................................. ...................... 55
and VDSmeasurement......................................................... 35
GS
measurement................................................................... 37
DS
TIVH Series Measurement Syste m User Manual i
Table of Contents
Tripods ......................................................................................................... 56
Connecting the sensor tip cables ............................................................................ 58
Installing the probe tip adapters . .... . ..... . ..... . ..... . ..... . ... . . ..... . ..... . ..... . ..... . ... . . . .... . ..... . . 59
Installing the square pins on the circuit board . .... . ..... . ..... . .... . ..... . ..... . ... . . ..... . ..... . ..... .... 62
User service .......... ................................ ................................ .............................. 65
Service offerings.............. ................................ .................................. .............. 65
Preventive
Performance verication procedures ....................................................................... 66
Propagation delay............................................................................................. 66
Troubleshooting and error conditions .................. ................................ .................... 70
Repack the measurement system for shipment ............................................................ 72
Test record ..................................................................................................... 73
Appendi
CH<n>:PRObe?............................................................................................... 75
CH<n>:PRObe:AUTOZero EXECute ..................................................................... 75
CH<n>:PRObe:COMMAND “CLAMP”, {“ON” | “OFF”} ........................... .................. 75
CH<n>:PRObe:SET {“CLAMP ON” | “CLAMP OFF”}................................................ 75
CH<n>:PRObe:SET {“CLAMPING ON” | “CLAMPING OFF”} ............... ...................... 76
CH<n
CH<n>:PRObe:GAIN? ............ ................................ .................................. ........ 76
CH<n>:PRObe:ID {:SERnumber | :TYPe}? .... ................................ .......................... 77
CH<n>:PRObe:PROPDELay? ................ ................................ .............................. 77
CH<n>:PRObe:RECDESkew?.............................................................................. 77
CH<n>:PRObe:RESistance? ................................................................................ 77
<n>:PRObe:UNIts?..................... ................................ .................................. 77
CH
CH<n>:PROBECOntrol {AUTO | MAN}................................................................. 78
CH<n>:PROBEFunc:EXTAtten <NR3>................................................................... 78
CH<n>:PROBEFunc:EXTDBatten? ....................................................................... 78
Appendix B: Compliance information .......................................................................... 79
Safety compliance ............................................................................................ 79
Environmental considerations ............................................................................... 81
Index
maintenance ...................................................................................... 65
x A: Remote programming............................................................................. 75
>:PRObe:FORCEDRange <NR3> .................................................................. 76
ii TIVH Series Measurement System User Manual
List of Figures
Figure i: TIVH Series IsoVu Measurement System ......................... ................................ ... x
Figure 1: Maximum safe handling limits for common mode voltages between the sensor head and earth
ground. .......................................................................................................... 6
Figure 2: RF burn hazard zone around the sensor head....... ................................ ................. 7
Figure 3: Controller indicators and buttons...................................................................... 8
Figure 4: Labels on the sensor head............................................................................. 10
Figure 5: Sensor tip cable top and bottom labels .............................................................. 10
Figure 6: Connect the Comp box to the oscilloscope.. . . ..... . ..... . .... . ..... . ..... . .... . ..... . ... . . . .... . ... 12
Figure 7: Connecting the sensor tip cable to the sensor head................................................. 12
Figure 8: Connect the s ensor head to the exible tripod ...... ................................ ................ 13
Figure 9: Probe Setup menu...................................................................................... 15
Figure 10: Digital lter aliasing ................................................................................. 17
Figure 11: Digital lter aberrations (about 2.5% of V
Figure 12: Digital lter aberrations after disabling offset correction........................................ 18
Figure 13: Top sensor tip labels.......................................... .................................. ...... 19
Figure 14: Bottom sensor tip labels ............................................................................. 21
Figure 15: System Transfer Function – 1X Range..................... .................................. ...... 24
Figure 16: System Transfer Function – 2X Range..................... .................................. ...... 25
Figure 17: Use of input offset to improve signal delity.. ..... . ..... . ... . . . .... . ..... . ..... ..... . ..... . ... . . . 27
Figure 18: Use of input offset to bring a signal into the measurement system input range ... . . ..... . ... . 28
Figure 19: 1 V Figure 20: 1 V Figure 21: 1 V Figure 22: 1 V Figure 23: 1 V Figure 24: 1 V
Figure 25: Half-bridge circuit showing the gate, source, and drain of the high-side FET.. . .... . ..... . ... 35
Figure 26: High-side turn-on characteristics ..................... .................................. ............ 36
Figure 27: V
Figure 28: High side current shunt .............................................................................. 38
Figure 29: SMT resistor model .................................................................................. 39
Figure 30: ESD discharge test example..... ................................ .................................. .. 40
Figure 31: Typical CMRR values for 1X SMA cable and MMCX series probe tip cables ............... 45
Figure 32: Typical CMRR values for square pin probe tips .................................................. 46
Figure 33: Maximum differential input voltage vs. frequency for MMCX series probe tips . ..... . .... . . 47
Figure 34: Maximum differential input voltage vs. frequency for square pin probe tips ............... .. 47
Figure 35: Differential input impedance vs. frequency chart for the MMCX series probe tips . ..... . ... 48
Figure 36: Differential input impedance vs. frequency chart for the square pin probe tips .... . ..... . .... 49
Figure 37: Sensor head dimensions with probe tip cover ............. ................................ ........ 50
signal (±500 mV) centered at 0 V with no input offset applied.......................... 29
P-P
signal (0 to +1 V) with no input offset applied ............................................ 30
P-P
signal (0 to +1 V) with 500 mV input offset applied...................................... 31
P-P
signal (±500 mV) centered at 0 V with no input offset applied.......................... 32
P-P
signal (0 to +1 V) with no input offset applied ............................................ 33
P-P
signal (0 to +1 V) with 500 mV input offset applied ....................... .............. 34
P-P
and VDSswitching characteristics.............................................................. 38
GS
) ... ................................ .................. 18
p-p
TIVH Series Measurement System User Manual iii
Table of Contents
Figure 38: Sens
Figure 39: Controller dimensions................................................................................ 51
Figure 40: Comp box dimensions . .................................. ................................ ............ 52
Figure 41: Probe tip adapter dimensions........................................................................ 52
Figure 42: SQPIN probe tip dimensions.................. ................................ ...................... 53
Figure 43: WSQPIN probe tip dimensions ..................................................................... 54
Figure 44: Block diagram .................. .................................. ................................ .... 55
Figure 45: Installing the exible tripod under the DUT. ...................................................... 56
Figure 46: Connecting the sensor head to the top of the DUT with the exible tripod.................... 56
Figure 47: Connecting to an adapter on the circuit board with the probe tip tripod ....................... 57
Figure 48: Square pin series sensor tip cables.................................................................. 58
Figure 49: Lining up the MMCX-to-0.1-inch (2.54 mm) adapter on the circuit board.................... 59
Figure 50: Lining up the MMCX-to-0.062-inch (1.57 mm) adapter on the circuit board . . ..... . ... . . .... 60
igure 51: Pushing the MMCX-to-0.062-inch (1.57 mm) adapter in place................................. 60
F
Figure 52: Pushing the MMCX-to-0.1-inch (2.54 mm) adapter in place................................ .... 61
Figure 53: Adapter clearance requirements .................................................................... 62
Figure 54: Removing the header from square pins on the circuit board................................. .... 63
Figure 55: Using the soldering aide to install the square pins on the circuit board .......... .............. 64
Figure 56: Propagation delay measurement .................................................................... 69
or head dimensions without probe tip cover ........... ................................ ...... 51
List of Tables
Table 1: Standard accessories .. ................................ .................................. ................. 1
Table 2: Optional accessories ..................................................................................... 2
Table 3: Input specications...................... ................................ ................................ . 4
Table 4: Environmental considerations .......................................................................... 4
Table 5: Controller indicators and buttons.............. .................................. ....................... 8
Table 6: Offset Correction mode LED indicator............................................................... 16
Table 7: Sensor tip selection table ............................................................................... 20
Table 8: Differential input voltages – 2X Range..... ................................ .......................... 26
Table 9: Warranted specications................................ .................................. .............. 41
Table 10: Electrical specications............................................................................... 41
Table 11: Physical specications ...................... ................................ .......................... 49
Table 12: Required equipment for performance verication ... .................................. ............ 66
Table 13: Problems and possible solutions ..................................................................... 70
Table 14: Test record.................... .................................. ................................ ........ 73
Table 15: Sensor tip cables and dynamic ranges....................... .................................. ...... 76
iv TIVH Series Measurement System User Manual
Important safety information
This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe condition.
To safely perform service on this product, additional information is provided at the end of this section. (See page vii, Service safety summary.)
General safety summary
Use the product only as specied. Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. Carefully read all instructions. Retain these instructions for future reference.
Comply with local and national safety codes.
For correct and safe operation of the product, it is essential that you follow generally accepted safety procedures in addition to the safety precautions specied in this manual.
The product is designed to be used by trained personnel only.
To avoid re or personal
injury
Only qualied perso the cover for repair, maintenance, or adjustment.
Before use, always check the product with a known source to be sure it is operating correctly.
This product is not intended for detection of hazardous voltages.
Use personal protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions related to operating the system.
When incorporating this equipment into a system, the safety of that system is the responsibility of the assembler of the system.
Connect and disconnect properly. Do not connect or disconnect sensor tip cables, test leads, or accessories while they are connected to a voltage source. Use only test leads and accessorie to be suitable for the product.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product. Do not exceed the Measurement Category (CAT) rating and voltage or current rating of the lowest rated individual component of a product or accessory.
nnel who are aware of the hazards involved should remove
s supplied w ith the product, or indicated by Tektronix
TIVH Series Measurement Syste m User Manual v
Important safety information
Do not apply a po
Do not operate without covers. Do not operate this product with covers or panels removed, or with the case open. Hazardous voltage exposure is possible.
Avoid exposed circuitry. Do not touch exposed connections and components when power i
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualied service personnel.
Disable the product if it is damaged. Do not use the product if it is damaged or operates incorrectly. If in doubt about safety of the product, turn it off and disconnect it from the instrument. Clearly mark the product to prevent its further operation.
Before use, inspect accessories for mechanical damage and replace when damaged. Do not use them if they are damaged, or if there is exposed metal.
Examine the exterior of the product before you use it. Look for cracks or missing pieces.
Use only specied replacement parts.
tential that exceeds the m aximum rating.
s present.
Sensor tip cables
t operate in wet/damp conditions. Be aware that condensation may occur if
Do no
a unit is moved from a cold to a warm environment.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry. Remove the input signals before you clean
the product.
Provide a safe working environment. Always place the product in a location convenient for viewing the display and indicators.
Be sure your work area meets applicable ergonomic standards. Consult with an ergonomics professional to avoid stress injuries.
Maintain safe clearance from the sensor head and sensor tip cable while connected to the energized circuit as recommended in this manual.
Remove the sensor tip cable and adapters from the test circuit when not in use.
Leave the sensor tip cable connected to the sensor head when not in use.
Use only correct Measurement Category (CAT), voltage, temperature, altitude, and amperage rated sensor tip cables and accessories for any measurement.
vi TIVH Series Measurement System User Manual
Important safety information
Beware of high v
using and do not exceed those ratings. It is important to know and understand the maximum measurement voltage rating of the product. The voltage rating depends on the measurement category, the instrument, and your application. Refer to the Specications section of the manual for more information.
WAR NI NG . To prevent electrical shock, do not exceed the maximum measurement
or maximum v
Connect and disconnect properly.
CAUTION.
connecting or disconnecting the sensor tip cable.
Servicesafetysummary
The Service safety summary section contains additional information required to safely perform service on the p roduct. Only qualied personnel should perform service procedures. Read this Service safety summary and the General safety
summa
oltages. Understand the voltage ratings for the product you are
oltage category.
To avoid damage to the equipment, de-energize the test circuit before
ry before performing any service procedures.
To avoid electric shock. Do not touch exposed connections.
Do not service alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering rst aid and resuscitation is
sent.
pre
Disconnect power. To avoid electric shock, switch off the product power and disconnect the power cord from the mains power before removing any covers or panels, or opening the case for servicing.
se care when servicing with power on. Dangerous voltages or currents may exist
U
in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
Verify safety after repair. Always recheck ground continuity and mains dielectric strength after performing a repair.
TIVH Series Measurement System User Manual vii
Important safety information
Termsinthismanual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Isolated, electrically oating. The terms isolated, electrically floating,and galvanically isolated are used in this document to indicate a measurement where
there is no direct conduction path to earth ground.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
this symbol is marked on the product, be sure to consult the manual
When to nd out the nature of the potential hazards and any actions which have to betakentoavoidthem.(Thissymbolmayalsobeusedtorefertheuserto
ings in the manual.)
rat
The following symbol(s) may appear on the product:
viii TIVH Series Measurement System User Manual
Preface
Preface
Key features
This documen Series IsoVu Mea surement System. The measurement system offers a galvanically isolated measurement solution for accurately resolving high bandwidth, high voltage differential signals up to 2500 Vpk in the presence of large common mode voltages with the best in class common mode rejection p erformance across its bandwidth.
New IsoVu technology - galvanically isolated, oating, measurement system
Differential voltages up to ±2500 Vpk (sensor tip cable dependent)
Large common mode voltage range up to 60 kV peak
High impedance input up to 40 M(sensor tip cable dependent)
Larg
DC and AC input coupling
t provides information for installing and using the Tektronix TIVH
Bandwidth from DC to 800 MHz (sensor tip cable dependent)
160 dB (100 million to 1 common mode rejection ratio) from DC to1 MHz,
120 dB (1 million to 1) at 100 MHz, 80 dB (10,000 to 1) at 800 MHz
e input offset range up to ±2500 Vpk (sensor tip cable dependent)
Laser certication
This product complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50, dated June 24, 2007.
CAUTION. Useofcontrolsoradjustmentsforperformanceofproceduresother
than those specied herein may result in hazardous radiation exposure.
TIVH Series Measurement Syste m User Manual ix
Preface
Product description
The Tektronix TIVH Series IsoVu Measurement System offers a completely galvanically isolated (optical isolation) system. The system consists of a sensor tip cable, a sensor head, a controller, and a TekVPI interface as shown in the following gure. Hazardous voltages in the sensor head are completely isolated from the con
troller and oscilloscope by optical ber cables.
Figure i: TIVH Series IsoVu Measurement System
Comp box
Controller
x TIVH Series Measurement System User Manual
The TekVPI compensation box (Comp box) connects the measurement system to one of the input channels on the oscilloscope. Power is supplied to the
asurement system through the TekVPI interface of the oscilloscope.
me
The controller connects to the oscilloscope through a coaxial cable and the Comp box. Buttons and indicators on the controller provide a means for controlling the
easurement system and indicating the overall status.
m
Preface
Models
Sensor head
Sensor tip cables
The sensor head the controller. It contains an electro-optic converter that converts the electrical signal from the sensor tip cables to an optical signal sent to the controller.
Several sens DUT:
MMCX10X, 10
MMCX50X, 50X tip cable connector
MMCX250X, 250X tip cable connector
SQPIN100X, 100X 0.1-inch (2.54 mm) square pin connector
SQPIN500X, 500X 0.1-inch (2.54 mm) square pin connector
WSQPIN1000X, 1000X 0.2-inch (5.08 mm) square pin connector
WSQPIN2
The TI
VH Series IsoVu Measurement System includes the following models:
provides an interface between the device-under-test (DUT) and
or tip cables options are available to connect the sensor head to the
X tip cable c onnector
500X, 2500X 0.2-inch (5.08 mm) square pin connector
TIVH08. Tektronix IsoVu 800 MHz High Voltage with 3 m cable
TIVH08L. Tektronix IsoVu 800 MHz High Voltage with 10 m cable
TIVH05. Tektronix IsoVu 500 MHz High Voltage with 3 m cable
TIVH05L. Tektronix IsoVu 500 MHz High Voltage with 10 m cable
TIVH02. Tektronix IsoVu 200 MHz High Voltage with 3 m cable
VH02L. Tektronix IsoVu 200 MHz High Voltage with 10 m cable
TI
TIVH Series Measurement System User Manual xi
Preface
Supported osc
illoscopes
The measurement system can be used with the following Tektronix oscilloscopes. For oscilloscopes not included in this list, contact your local Tektronix representat
WARNING. To
the WSQPIN2500X tip without updating 5 Series MSO rmware to V1.6.X or later. If you use the WSQPIN2500X tip with earlier versions of 5 Series MSO rmware, the oscilloscope rmware will crash.
IfyoumistakenlyconnecttheWSQPIN2500Xtiptoa5SeriesMSOwithanearly version of rmware and need help to recover your probe and/or oscilloscope, please do one of the following:
• View the FAQ at the Tektronix website.
• Contact your local Tektronix Service Center.
To view the FAQ :
1. Cli
2. On the website, click Technical Docs & Downloads.
ive.
5SeriesMSO
prevent the 5 Series MSO oscilloscope from crashing, do not use
ck the following link: www.tek.com/isolated-measurement-systems.
3. Select FAQ in the Sort resources by: list.
4. Click Apply.
MDO3000 Series (WSQPIN tip cables require V1.26 or later oscilloscope rmware)
MDO4000C Series (WSQPIN tip cables require V1.06 or later oscilloscope rmware)
MSO/DPO/MDO4000B Series (WSQPIN tip cables are not compatible)
MSO/DPO5000B Series
DPO7000C Series
In addition to the above oscilloscopes, the measurement system can also be used with the following oscilloscopes with a TCA-VPI50 adapter.
MSO/DPO70000C series
MSO/DPO70000DX series
DPO70000SX series
xii TIVH Series Measurement System User Manual
Operating information
Accessories
This section lists the standard and optional accessories available for the measurement s ystem.
Standard ac
cessories
Table 1: Standard accessories
Tektronix
Accessory
IsoVu product carrying case, soft case
IsoVu accessories carrying case, soft case
50X Sensor tip cable (MMCX-to-SMA) MMCX50X
500X Sensor tip cable (Square pin-to-SMA) SQPIN500X
5/16-inch SMA wrench/driver tool
Probe tip adapter (blue), MMCX to 0.1-inch (2.54 mm) square pin (0.025-inch (0.635 mm) square pins)
Flexible tripod with quick release 352-1171-xx
Flexible tripod feet, 3 each
Probe tip tripod support with living h inge, 2 each 352-1170-xx
User manual (this document)
Certicate of traceable calibration
part number
016-2108-xx
016-2110-xx
003-1951-xx
131-9717-xx
344-0693-xx
071-3556-xx —
TIVH Series Measurement Syste m User Manual 1
Operating information
Optional accessories
Additional acc
essories, such as other sensor tip cables are available. The
following table lists the optional accessories.
Table 2: Optional accessories
Tektronix
Accessory
1X Sensor tip cable (MMCX-to-SMA)
10X Sensor tip c able (MMCX-to-SMA) MMCX10X
250X Sensor tip cable (MMCX-to-SMA) MMCX250X
100X Sensor tip cable (Square pin-to-SMA) SQPIN100X
1000X Sensor tip cable (Square pin-to-SMA) WSQPIN1000X
2500X Sensor tip cable (Square pin-to-SMA) WSQPIN2500X
Solder aid for 0.062-inch (1.57 mm) pitch square pins (0.016 - 0.018-inch
46 mm) square pin installation tool)
(0.4 - 0.
erface pin kit with (qty 20) 0.018-inch (0.46 mm) round solder-in
DUT Int round pins
Probe tip adapter (white), MMCX to 0.062-inch (1.57 mm) square pin (0.016
- 0.018-inch (0.4 - 0.46 mm) square pins)
Lead, MMCX to IC grabber
Lead, 0.100-inch square pin header to IC grabber
Lead, 0.200-inch square pin header to banana jack
1
The electrical performance of this tip will vary widely based on the source impedance of the circuit being measured.
2
ludes alligator clips AC285-FL. To prevent damage to the alligator clip insulation, do not use in high A/m
Inc magnetic elds at high frequencies, which can cause induction heating of the jaws.
2
part number
IVTIP1X
1
003-1946-xx
020-3169–xx
677-xx
131-9
196-3546-xx
196-3547-xx
020-3189-xx
2 TIVH Series Measurement System User Manual
Operating information
Operating con
Measurement system
handling best practices
siderations
Read this section before installing your measurement system to be aware of the operating requirements and clearance requirements including possible hazardous areas when th
The measurement system consists of quality parts and should be treated with care to avoid damage o r degrading the performance due to mishandling. Consider the following
e measurement system is connected to the DUT.
precautions when handling the ber-optic cables and sensor tip cables:
Do not crush, crimp, or sharply bend the ber-optic cable. Avoid making loops in t
Do not twist the ber-optic cable; twisting the cable will stress the optical bers.
Do not allow kinks or knots to develop in the ber-optic cable.
Avoid putting tension on the ber-optic cable.
Donotpullorjerktheber-optic cable, especially when kinks or knots are present.
Do not dr op the sensor head or controller assembly since damage and misalignment of the internal optical components can result.
he ber-optic cable smaller than 5 inches (12.7 cm).
Avoid over-bending the sensor tip cables; do not exceed the minimum bend radius of 2.0 inches (5.1 cm).
Avoid crushing the cables by accidentally running over the cable with a chair wheel or by dropping a heavy object onto the cable.
Never support the weight of the sensor head or controller by the ber-optic cable.
Store the measurement system in the supplied carrying case when not in use.
TIVH Series Measurement Syste m User Manual 3
Operating information
Environmental
requirements
The following t
ables describe specications and maximum operating environmental ratings for the measurement system when connected to a DUT and a Tektronix oscilloscope.
Table 3: Input specications
Feature Description
Common mode
Differential mode Sensor tip cable dependent (See the Sensor tip cable voltage
1
This product is designed to be used in the circuit with expected transient overvoltage up to 5000 V peak.
60 kV peak
rating below.)
1
Table 4: Environmental considerations
Feature Description
Temperature
Controller
Operating 0 °C to 40 °C (32 °F to 104 °F)
Non-operating
Sensor head
Operating 0 °C to 70 °C (32 °F to 158 °F)
Non-operating
Sensor tip cables/adapters
Operating and non-operating
Humidity
Controller
Operating 5% to 85% RH (Relative H umidity), at up to 40 °C (104 °F),
Non operating
-40°Cto70°C(-40°Fto158°F)
-40°Cto70°C(-40°Fto158°F)
-40°Cto85°C(-40°Fto185°F)
non-condensing
5% to 85% RH (Relative Humidity), at up to 40 °C (104 °F)
5% to 45 % RH above 40 °C (104 °F) up to 70 °C (158 °F), non-condensing
4 TIVH Series Measurement System User Manual
Table 4: Environmental considerations (cont.)
Feature Description
Sensor head
Operating 5% to 80% RH (Relative Humidity), at up to 40 °C (104 °F)
5% to 45 % RH above 40 °C (104 °F) up to 70 °C (158 °F), non-condensing
Non operating
Sensor tip cables/adapters
Operating 5% to 80% RH (Relative Humidity), at up to 40 °C (104 °F)
Non operating
Altitude
Operating 3000 m (9843 ft.)
Non operating
5% to 85% RH (Relative Humidity), at up to 40 °C (104 °F)
5% to 45 % RH above 40 °C (104 °F) up to 70 °C (158 °F), non-condensing
5% to 45 % RH above 40 °C (104 °F) up to 85°C (185 °F), non-condensing
5% to 85% RH (Relative Humidity), at up to 40 °C (104 °F)
5% to 45 % RH above 40 °C (104 °F) up to 70 °C (158 °F), non-condensing
12,000 m (39370 ft.)
Operating information
TIVH Series Measurement Syste m User Manual 5
Operating information
Clearance requirements
The unique comm be used in the presence of high frequency/high voltage common mode signals. It is important to observe all precautions while using this product.
WARNING. RF burns can occ ur while using this measurement system. The system
is intended to isolate the operator from hazardous input voltages (common voltages);the plastic case of the sensor head and the shield on the sensor tip cable do not suppl
Maintain the safe clearance from the sensor head and sensor tip cable while the measureme document. (See Figure 2 on page 7.) Do not access the RF Burn Hazard Zone while taking measurements on a live circuit.
WARNING. RF burns can occur while using t his measurement system. Anyone
making measurements that might come within the RF burn regions indicated in the following gure should be familiar with the hazards of working with signals in these regions and take appropriate action, such as RF shielding for the DUT.
While measuring high frequency common mode signals, there is a risk of RF burns. Refer to the following derating curve to identify the danger areas. Measuring common mode signals within the gray shaded area can result in RF burn
nt system is connected to the energized circuit as recommended in this
s within 1 m (40 in.) of the sensor head and earth ground.
on mode voltage range of the measurement system allows it to
y safe isolation.
Figure 1: Maximum safe handling limits for common mode voltages between the sensor head and earth ground.
6 TIVH Series Measurement System User Manual
Operating information
The following potential RF burn area when working with hazardous voltages.
gure shows the components of the measurement system and the
Figure 2: RF burn hazard zone around the sensor head
TIVH Series Measurement Syste m User Manual 7
Operating information
Controls and i
Controller
ndicators
The following gure shows the indicators and buttons on the controller; their functions are described in the following table.
Figure 3: Controller indicators and buttons
Table 5: Controller indicators and buttons
Item Description
1
2
3
OVERRANGE Indicator. This red LED indicates if the DC/low frequency differential voltage applied to the sensor head or attached sensor tip cable has exceeded the maximum specied input voltage levels.
RANGE indicator. The two LEDs indicate the differential input range setting.
RANGE button. Push this button to toggle between one of two differential input voltage ranges.
MENU button. Push this button to view the Probe Control menu on the oscilloscope display.
8 TIVH Series Measurement System User Manual
Table 5: Controller indicators and buttons (cont.)
Item Description
4
5
6
CLAMPING indicator. This LED indicates whether the output clamping is enabled.
CLAMPING button. Push this button to enable or disable the output clamping feature.
SELF CAL indicator. This LED indicates the self calibration status.
Solid green. The self calibration has passed.
Flashing orange. The self calibration is in progress.
Solid red. The self calibration has failed.
Solid Orange. The self calibration has not been run or is questionable.
SELF CAL button. Push this button to start the self calibration routine.
STATUS indicator. This LED indicates the status of the measurement system:
Solid green. The unit has powered on and passed the power-on self
tests and is in normal operation.
Operating information
Sensor head
Flashing green. The unit has not completed the power-up sequence. This is typically due to a communication fault between the host oscilloscope and the IsoVu unit. Disconnect the TekVPI comp box and then reattach the comp box.
Solid or ashing red. Error condition that requires the unit to be sent to Tektronix for service.
Flashing red/yellow. The unit has failed the power on self tests. Cycle the power; if the problem persists, the unit needs to be sent to Tektronix for service.
Labels on the sensor head provide high-level specications for connecting to
e DUT. They also provide reminders of the potential RF burn hazards while
th connected to the DUT.
TIVH Series Measurement Syste m User Manual 9
Operating information
Figure 4: Labels on the sensor head
Sensor tip cables
Each sensor tip cable has a set of labe ls on the top and bottom. The top labels provide reminders of the maximum differential input voltage range for each tip cable. T input resistance and capacitance for the tip cable (differential loading).
he bottom labels include the name of the tip cable and the differential
Figure 5: Sensor tip cable top and bottom labels
10 TIVH Series Measurement System User Manual
Operating information
Connecting to
a circuit
WAR NI NG . This measurement system contains laser sources; exposing these
laser sources may cause laser exposure. Exc ept for the sensor tip cables on the sensor head, do not remove any plastic or metallic covers from the sensor head or controller or attempt to disassemble the product.
WAR NI NG . Do not connect the measurement system to an energized circuit to
avoid the risk of shock. Always de-energize the circuit-under-test before installing or removing the tip cable from the circuit-under-test. The plastic case of the sensor head and the shield tip of the sensor cable do not supply the isolation.
WAR NI NG . To avoid the risk of electrical shock or RF burns while the DUT
is energized, do not touch the sensor head or sensor tip cable while taking measur the measurement. (See Figure 2 on page 7.)
Be sure the RF burn hazard zone. (See Figure 1 on page 6.)
ements. Alwayskeepa1m(40in.)clearancefromthesensorheadduring
to check the maximum ratings and derating curve for m ore information on
ION. To avoid possible damage to the equipment, do not connect the coaxial
CAUT
(common) shield of the sensor tip cable or SMA input to the high impedance portion of a circuit. The additional capacitance can cause circuit damage. Connect the coaxial (common) shield to the low impedance section of the circuit.
NOTE. Touching the sensor head or sensor tip cable when measuring a high
frequency common mode signal increases the capacitive coupling and can degrade the common mode loading on the circuit-under-test.
WAR NI NG . To prevent the arc ash caused by a different potential, do not place
the sensor head or sensor tip cable on the circuit that has the different voltage.
TIVH Series Measurement Syste m User Manual 11
Operating information
The following s between a Tektronix oscilloscope and the DUT.
1. Verify the DUT
2. Connect the Comp box to one of the channels on the oscilloscope.
Figure 6: Connect the Comp box to the oscilloscope
3. Refer to the following gure and install the sensor tip cable to the sensor head.
a. Line up the sensor tip cable with the sensor head.
teps describe the process for connecting the measurement system
is not connected to an energized circuit.
Take car e to avoid bending or twisting the sensor tip cable assembly during this process.
b. Connect the SMA connector of the sensor tip cable to the sensor head.
Use the accessories wrench to tighten the SMA cable to 4 to 5-in lbs.
Use the adjustment tool that was shipped with your system.
c. Pres
s the sensor tip cable housing against the sensor head and then tighten
thetwoscrewsto3to5-inlbs.
the adjustment tool that was shipped with your system.
Use
Figure 7: Connecting the sensor tip c able to the sensor head
12 TIVH Series Measurement System User Manual
Operating information
4. Connect the sen
This support keeps the sensor head steady reducing the potential mechanical stresses at th the sensor head away from surrounding circuits and conductive objects to minimize the parasitic capacitive coupling to these surroundings.
NOTE. The mating thread in the sensor head is UNC¼-20. If you use a different
support, make s ure that mating thread is UNC¼-20.
Figure 8: Connect the sensor head to the exible tripod
sor head to the exible tripod or a similar support.
e electrical connection point of the DUT. The support also keeps
NOTE. To obtain the most accurate measurement, allow the measurement system
m up for 20 minutes. Then perform the self calibration before connecting
to war the tip cable to the DUT and taking the measurement.
5. Conn
6. Set up the controls on the oscilloscope.
7. Apply power to the DUT to take the measurement.
ect the sensor tip cable end to the DUT. If you are using MMCX sensor tip cable, connect it to an MMCX connector on the DUT or to a square pin adapter on the D UT. The adapters connect to square pins with either 0.100-inch (2.54 mm) spacing or 0.062-inch (1.57 mm) spacing. (See Figure 47 on page 57.) If you are using one of the square pin sensor tip cables, connect it directly to the square pins on the DUT. (See page 58, Connecting
e sensor tip cables.)
th
TIVH Series Measurement System User Manual 13
Operating information
Self calibrat
ion
Press the SELF CAL button on the controller to adjust the operating point of the measurement system for the current range and clamp setting. (This function is not available in
NOTE. Make sure there is no differential voltage present at the sensor tip cable
when performing the self calibration.
When you power on the measurement system, the SELF CAL status indicator on the controller is orange, indicating that the operating point of the measurement system has not been optimized. As a result, the accuracy of the measurement system might be degraded. The SELF CAL sequence should always be run on the system after it is rst powered on and has warmed up for 20 minutes. After pu calibration process; it turns solid green when the operation completes or solid red when the operation fails.
There are situations when further self calibration is required. The SELF CAL status indicator turns orange to indicate when this is necessary:
The measurement system is rst attached to the oscilloscope.
the Probe Setup menu of the oscilloscope.)
shing the SELF CAL button, the indicator blinks orange during the self
Programming
ges are made to the range (1X|2X) or clamp (ON|OFF) setting.
Chan
The temperature in the sensor head changes more than 10 °C.
The internal compensation adjustments have shifted outside their normal operating ranges.
The sensor tip cable is changed.
Users wanting to initiate the self calibration through the programmatic interface must congure the measurement system to perform a self calibration every time it receives the AutoZero command. To link these functions, hold the MENU button down and momentarily press the SELF CAL button. The OVERRANGE indicator should blink red twice. This mode is non-volatile, and it also changes the operation of the AutoZero button in the Probe Setup menu of the oscilloscope. Restore the original operation by repeating the MENU-SELF CAL button pushes. In this case the OVERRANGE indicator will blink once.
14 TIVH Series Measurement System User Manual
AutoZero
Menu button
Operating information
When the displayed waveform is not centered correctly (for example, due to a small DC offset error), you might need to press the AutoZero button in the Probe Setup menu of the oscilloscope. This might be necessary for the rst time after the self calibration operation has completed. Make sure there is no differential signal pres
ent at the sensor tip cable.
Press the M oscilloscope, similar to the following gure.
Figure 9: Probe Setup menu
ENU button on the controller to view the Probe Setup menu on the
Use the buttons on the oscilloscope to change the probe setups. Some of the functions are the same as pushing the buttons on the controller, such as turning clamping on or off, or setting the input ranges.
TIVH Series Measurement System User Manual 15
Operating information
Offset correc
tion
The measurement system uses state-of-the-art technology that allows the DUT to be totally isolated from the oscilloscope. This results in a very large common mode rejecti
on ratio (CMRR) and lets you see small signals that would otherwise
be obscured by the high common mode interference.
The TIVH Ser
ies products implement an offset correction algorithm to minimize any drift in the system due to changes in temperature or ber movement. The offset correction algorithm maintains a constant DC level for the displayed on-screen signal.
NOTE. In rmware version 2.02, an optional offset correction a lgorithm was
added that improves DC performance over the original offset correction
hm. In almost all cases, Tektronix recommends using the new, Enhanced
algorit Offset Correction mode.
To chan
ge Offset Correction mode, hold MENU and press CLAMPING. The
Overrange LED will blink to indicate the current Offset Correction mode.
6: Offset Correction mode LED indicator
Table
Overrange LED blink count Offset Correction mode
Once Normal offset correction
e
Twic
ee times
Thr
1
Added in rmware version 2.02.
1
Offset correction disabled
Enhanced offset correction
1
One of the components used for offset correction is a digital low-pass lter. Normally, the frequency of the signal is high enough that the operation of the
ter is transparent. However, at low frequencies special care must be taken to
l prevent aberrations from being introduced.
hen attempting to capture a signal that has a frequency near 43.5 Hz, 87.0 Hz, or
W
130.5 Hz, the display might show aliasing. If you cannot avoid these frequencies, the offset correction should be disabled. The sample rate of the digital lter was selected to not alias at 50 Hz or 60 Hz.
16 TIVH Series Measurement System User Manual
Operating information
Figure 10
To disab
:Digitalfilter aliasing
le the offset correction algorithm, hold down the MENU button on the controller and momentarily press the CLAMPING button. The OVERRANGE indicator on the controller should blink two times. To use Enhanced Offset Correction mode, repeat the MENU and CLAMPING procedure until the Overrange LED blinks three times.
Another source of aberrations is introduced when the signal is below the cutoff frequency of the digital filter (5.0 Hz). The following figure shows a close-up view of the front edge of a 0.1 Hz 800 mV
created by the digital lter.
also
square wave. These aberrations are
p-p
TIVH Series Measurement System User Manual 17
Operating information
Figure 11
The foll
: Digital lter aberrations (about 2.5% of V
owing gure shows the aberrations after the offset correction has been
p-p
)
disabled. When the offset correction is disabled, the system no longer corrects for long-term drift due to temperature change and ber movement.
Figure 12: Digital lter aberrations after disabling offset correction
18 TIVH Series Measurement System User Manual
1X/2X Range
Operating information
The top label of each sensor tip cable (MMCX10X, MMCX50X, MMCX250X, SQPIN100X, SQPIN500X, WSQPIN1000X, WSQPIN2500X) shows the dynamic rang
e as a ±peak-to-peak voltage (with the voltage shown in parenthesis).
Figure 13: Top s ensor tip labels
Auto Range
For example, the MMCX10X tip has a dynamic range of 10 V
when the
p-p
1X Range is selected. This means that you can display a signal with a ±5 V
rential voltage. When the 2X Range is selected, the dynamic range increases
diffe from 10 V
to 20 V
p-p
(±10.0 V). Refer to the Linear differential input voltage
p-p
range in the specications table for more information.
CAUTION. To avoid damaging the probe, do NOT exceed the Peak Voltage rating.
The Maximum Non-Destruct Voltage limit (Peak Voltage) does not increase when the 2X Range is selected. For example, with the MMCX10X, the ±250 V
same for both the 1X or 2X Range.
the
limit is
pk
By default, the MSO/DPO5000, DPO7000, and MSO/DPO70000 Series Oscilloscopes select the 1 X or 2X Range automatically when you change the V/div setting. This hides the complexity of selecting the range from casual users. However, there are combinations of Range and V/div settings that cannot be reached when Auto Range is selected. For these situations, select Manual Range when full exibility is desired.
TIVH Series Measurement System User Manual 19
Operating information
Selecting a se
nsor tip cable
CAUTION. Avoid over-voltage conditions that can damage or degrade the sensor
head input termination by selecting the correct sensor tip cable. Selecting the correct sensor tip cable attenuation factor is crucial to ensure that the sensor head input termination is not degraded or damaged by an over-voltage condition. Select the sensor tip cable that will provide the lowest attenuation possible for the signal bei
When selecting a sensor tip cable for a particular application, consider the followin
What is the maximum RMS/Peak Voltage at the test point being measured (for exa
What is the minimum differential loading (input resistance) that my circuit can tol
How large of a signal do I want to display at one time on the oscilloscope?
What sensitivity do I need (for example, the minimum V/div setting)?
The following table will help you select the correct sensor tip. Start at the top of the table and work down. Choose the rst tip that meets all of your criteria.
ng measured.
g questions:
mple, under a fault condition)?
erate?
Table 7: Sensor tip selection table
Differential input specications
Most sensitive V/div
Sensor tip
WSQPIN2500X
WSQPIN1000X
SQPIN500X
SQPIN100X
MMCX250X
MMCX50X
MMCX10X
1
On 1X Range.
2
On 2X Range.
3
(See page 47, Maximum differential input voltage vs. frequency derating graphs.)
setting
2.5 V
1.0 V
500 mV
100 mV
250 mV
50 mV
10 mV
1
Linear voltage range
±2500 V (5 kV
±1000 V (2 kV
±500 V (1 kV
±100 V (200 V
±250 V (500 V
±50 V (100 V
±10V(20V
P-P
P-P
P-P
P-P
P-P
P-P
P-P
)
Maximum non-destruct voltage
2
(DC + pk AC)
)
)
)
)
)
)
±2500 V
±2500 V
±600 V
±600 V
±250 V
±250 V
3
±250 V
Differential input resistance
40 M
40 M
10 M
10 M
10 M
10 M
10 M
20 TIVH Series Measurement System User Manual
Operating information
Output clampi
ng
Sensor tip loading
The measurement system has a selectable output clamping feature. Push the button on the controller to enable or disable the output clamping feature. When enabled (ind the measurement system into the oscilloscope input. It allows you to increase the vertical sensitivity without over-driving or saturating the oscilloscope input.
Each of the sensor tip cables (MMCX10X, MMCX50X, MMCX250X, SQPIN100X, SQPIN500X, WSQPIN1000X, WSQPIN2500X) has a differential input resistance listed on the bottom labels as shown in the following gure.
icator is on), the output clamping limits the output voltage swing of
re 14: Bottom sensor tip labels
Figu
rtotheInput resistance/capacitance section in the specications table for
Refe more information. The sensor tip cables are specially designed to act as a common mode choke that helps reduce the common mode loading.
NOTE. The coaxial (common) shield of the sensor tip cable should always be
connected to the lowest impedance point (usually a circuit common or power supply rail) in the circuit-under-test (relative to the sensor tip cable/center
onductor) to obtain the most accurate waveform.
c
TIVH Series Measurement System User Manual 21
Operating information
Probe compensation
Deskew
The DPO7000 and MSO/DPO70000 Series oscilloscopes have a Compensate Probe feature accessible from the Probe Setup window. Pressing this button always results in a failure when using these oscilloscopes because of an incompatibility between the oscilloscope and the TIVH system. The accuracy of the TIVH s Compensate Probe feature fails.
Each oscilloscope family has a unique way of allowing you to adjust the timing relationship between signals acquired between different probes. Consult the user manual or online help of the oscilloscope for specic directions needed to deskew the probe. The 3 m and 10 m measurement systems have a propagation delay of approximately 35 ns and 68 ns respectively. The actual propagation delay is measur
ed on each measurement system and is stored within each unit.
ystem is still guaranteed on these oscilloscopes even though the
Input offset
The me This allows you to view a portion of the signal that is off-screen. One of the controls on the oscilloscope can be assigned this function.
To achieve the best performance of the measurement system, ensure that you are properly using the input offset capability of the TIV Series of products. (See page 26, Input offset voltage range.)
Input coupling (AC or DC)
The DC input coupling setting on the oscilloscope menu provides a direct, DC coupled, electrical path in the sensor head; it accepts all types of signals, including unchanging DC voltages, time-varying DC voltages, AC, and combinations of
CandDC.
A
When DC input coupling is enabled on the oscilloscope, the offset range is only dependent on the tip cable attenuation.
The DC Reject input coupling setting on the oscilloscope menu provides an AC-only path in the sensor head, removing DC offset from any mixed signal to view the AC component of the signal. DC reject is useful when you measure small amplitude signals superimposed on a large differential offset component. When DC Reject is enabled, the offset range is not only dependent on the tip cable attenuation, but also depends on the oscilloscope Volts/Div scale factor.
asurement system provides a user-adjustable, input referred offset voltage.
22 TIVH Series Measurement System User Manual
Operating information
TIV-Series vo
ltage range
Common mode
voltage r
Maximum non-destruct
differential voltage range
ange
TheTIV-Seriesisdesignedtoenablecharacterization of high frequency circuits with a wide range of differential voltages in the presence of common mode voltages. Un this section is essential to optimize signal delity and measurement accuracy.
Although th (>60 kV), the differential input range is limited and depends on the tip attenuation, the gain range selected, and the applied offset.
The input voltage conditions are divided into several different input ranges.
Since the TIV-Series sensor head is optically isolated from earth ground, the common m and refers to the signal that can be applied across the probe tip, regardless of the common mode voltage.
The ma voltage that can be applied to the input without damaging the probe. This is a DC +peak AC rating (i.e. no portion of the differential input s ignal should exceed this value). The maximum non-destruct differential voltage varies from ±25 V to ±2500 V depending of the tip cable b eing used. Exceeding this level can cause permanent damage to components of the sensor head.
derstanding the limits for the input operating voltages as d iscussed in
e common mode voltage range of the TIV-Series is very large
ode input range is >60 kV. The differential input range is more limited
ximum non-destruct differential input range is the maximum differential
Linear differential
oltage range
v
Refer to the tip cable label for the maximum non-destruct input range when using each tip cable. For example, for direct input to the sensor head using the SMA
nnector this value is ±25 V, but when using the SQPIN100X tip cable the
co value is ±600 V.
As with any analog measurement system, the sensor used in the TIV series has
limited differential input range. If the differential input signal exceeds this
a range, the gain accuracy is reduced.
In the TIV series there are two factors limiting the linear differential voltage range. The rst is the output voltage range of the amplier and the sec ond is the linear range of the sensor. If the amplier output voltage range is exceeded, the signal displayed on the oscilloscope will be clipped. The effects of exceeding the sensor linear range are less apparent and might lead to inaccurate measurements.
This limited linear range results in a reduction in gain that is gradual and might not always be apparent to the user. To obtain the most accurate measurements, the user must take care to maintain the differential input signals within both the amplier range and the sensor linear range.
TIVH Series Measurement System User Manual 23
Operating information
To ensure that t attenuation value should be selected to keep the peak-to-peak signal within the most accurate range. Check the label on the tip cable to verify that the appropriate tip cable and range have been selected for the signal being applied.
In addition, the TIV series of products all provide input offset capability to assist in maintaining differential input signals in the most accurate range. (See page 26, Input offset voltage range.) The TIV series also includes a self-calibration function that optimizes and centers this range. The self-calibration procedure should alw
System Transfer Function – 1X Range. For the System transfer function in the 1X Range, accuracy is best when the peak-to-peak signal is within the ± 60% window.
he signal to be measured is within the linear range, the tip
ays be executed prior to making any critical measurements.
Figure 15: System Transfer Function – 1X Range
24 TIVH Series Measurement System User Manual
Operating information
System Transfe
Range, accuracy is best when the peak-to-peak signal is within the ± 60% window.
Note that the e
r Function – 2X Range. For the System transfer function in the 2X
rror is higher in the 2X Range than it is in the 1X Range.
Figure 16: System Transfer Function – 2X Range
TIVH Series Measurement System User Manual 25
Operating information
Linear differe
ntial input range. The linear differential input range is divided into
several regions. This range depends on the tip cable attenuation, the gain range (1X / 2X) of the system, the state of the clamping circuit, and the input offset.
Table 8: Differential input voltages – 2X Range
Differential input voltage in
Sensor tip
Sensor Head Input SMA ±1 V (2 V
MMCX10X ±10 V (20 V
MMCX50X ±50 V (100 V
MMCX250X ±250 V (500 V
SQPIN100X ±100 V (200 V
SQPIN500X ±500 V (1 kV
WSQPIN1000X
WSQPIN2500X
1
Peak-to-peak signal should be maintained in this range for best accuracy.
Input offset voltage range
2X range (100% of Full Range)
)
P-P
)
P-P
)
P-P
)
P-P
)
P-P
)
P-P
±1000 V
±2500 V
P-P
P-P
(1)
(1)
To achieve the best performance of the measurement system, it is critical that the user understands the proper use of the input offset capability of the TIV Series of products. Referring to the table above, the input offset capability of the TIV Series extends from ±25 V to ±2500 V depending on the tip used. This offset capability
be used to bring signals that are outside the linear range of the sensor into the
can region with the m ost accurate response.
Input offset adjustment range Range for optimum signal delity
±25 V
±250 V
±250 V
±250 V
±600 V
±600 V
±2500 V
±2500 V
±600 mV (1.2 V
±6 V (12 V
±30V(60V
±150 V (300 V
±60V(120V
±300 V (600 V
P-P
P-P
)
)
P-P
P-P
P-P
±600 V (1000 V
±1500 V (2500 V
)
P-P
)
)
)
)
P-P
)
P-P
1
e minimum/maximum offset is different for each sensor tip cable. It is the same
Th when the 1X or 2X Range is selected; it is also the same for all V/div settings. (See Table 8.)
26 TIVH Series Measurement System User Manual
Operating information
Input offset ex
linear differential input range is specied to be ± 1 V (2 V
ample 1: 2X Range, clamping off, no tip cable. In this state, the
). This means that
P-P
the voltage level at the input connector should be within the range ±1 V with no offset. Without input offset applied, any input voltage outside this ±1 V range will not be displayed with the best accuracy.
However, the input offset control ca n be used to adjust the signal level so that the signal of interest is within this ±1 V window. One way to maintain the signal in the optimum range is to use the scope position control to set the trace (with no input si
gnal applied) to the middle of the screen. Then with the input signal applied, use the offset control to adjust the trace location to display symmetrically around the center of the screen.
For example, if the input signal is a 0 to +1 V triangle wave as shown below, using the 2X Range with no input offset applied will result in a distorted/compressed signal. To make this measurement in this example, the signal should be offset by +500 mV to center it in the most accurate range of the TIV system. In the following gure, the signal compression is exaggerated for illustrative purposes.
Figure 17: Use of input offset to improve signal delity
TIVH Series Measurement System User Manual 27
Operating information
Input offset ex
ample 2: 50 V square wave from +275 V to +325 V. The peak-to-peak
signal of 50 V is within the linear differential input range of several of the tips. However, the peak signal of +325 V exceeds the absolute maximum rating of the MMCX tips.
The SQPIN100X tip can be used to measure this signal since this tip has a ±600 V maximum input voltage rating and a linear differential input range of ±50 V (100 V
) in the 1X Range. To make this measurement in this example, an offset
P-P
of +300 V should be applied to center the signal in the TIV linear range.
Figure 18: Use of input offset to bring a signal into the measurement system input range
28 TIVH Series Measurement System User Manual
Operating information
Input offset ex
ample 3: Output voltage clipping. The TIV Series supports two
different gain ranges (1X and 2X) that are selectable from the controller panel or from the oscilloscope. The 1X Range has better sensitivity while the 2X Range has a larger input voltage range. In the 1X Range, the peak input voltage is limited by the amplier output voltage range a s illustrated in the following gures.
In each of the following gures, the input signal is the same amplitude (1 V
P-P
). The difference is the amount of offset in the input signal and the amount of amplier input offset applied by the measurement system.
Output voltage clipping examples (1X Range). In the following gure, the signal is within the input and output range of the measurement system. The system is in the 1X Range with direct input to the input SMA connector.
gure 19: 1 V
Fi
signal (±500 mV) centered at 0 V with no input offset applied
P-P
TIVH Series Measurement System User Manual 29
Operating information
In the followin
g gure, the s ignal is outside the output range of the measurement system and is causing the amplier output to be clipped. The system is in the 1X Range with direct input to the input SMA connector.
Figure 20: 1 V
signal (0 to +1 V) with no input offset applied
P-P
30 TIVH Series Measurement System User Manual
Operating information
In the followin
g gure, this is the same signal as in the previous gure. With offset applied, this signal is within the input and output range of the measurement system. The system is in the 1X Range with direct input to the input SMA connector.
Figure 21: 1 V
signal (0 to +1 V) with 500 mV input offset applied
P-P
TIVH Series Measurement System User Manual 31
Operating information
Output voltage
clipping examples (2X Range). This is the same signal as in the
1X example. (See Figure 19 on page 29.) This signal is within the input and output range of the measurement system. The system is in the 2X Range with direct input to the input SMA connector.
Figure 22: 1 V
signal (±500 mV) centered at 0 V with no input offset applied
P-P
32 TIVH Series Measurement System User Manual
Operating information
In the followin
g gure, the signal is within the output range of the m easurement system, but the peak level is at the upper limit of the input range and is slightly compressed. The system is in the 2X Range with direct input to the input SMA connector.
Figure 23: 1 V
signal (0 to +1 V) with no input offset applied
P-P
TIVH Series Measurement System User Manual 33
Operating information
In the followin
g gure, with offset applied, the signal is centered within the input and output range of the measurement system, optimizing the measurement accuracy. The system is in the 2X Range with direct input to the input SMA connector.
Figure 24: 1 V
signal (0 to +1 V) with 500 mV input offset applied
P-P
34 TIVH Series Measurement System User Manual
Application examples
The following examples are provided to help you become familiar with the TIVH Series IsoVu measurement system and to achieve the best performance for your application.
Example 1 : H
igh-side V
and VDSmeasurement
GS
Advancements in the components used in switching power supplies have made characterizing the performance of these power supplies increasingly difcult and challe high-side V an exceptionally good CMRR is required from the test system, especially for measuring the high-side V circuit.
Figure 25: Half-bridge circuit showing the gate, source, and drain of the high-side FET
nging. A particularly challenging measurement is measuring the
and VDSin a half bridge. To accurately make this measurement,
GS
. The following gure shows an example of this
GS
In this type of circuit, the gate-source voltage is of interest because the rate at which the device switches is determined by the gate drive characteristics. The reference node for this measurement is the high-side source node, which switches
tween the input supply voltage and the local PCB ground during operation. In a
be measurement system without sufcient CMRR, this rapidly changing common mode voltage results in interference which obscures the measurement. It is important to note that the CMRR for all measurement systems is frequency dependent; however, the frequency that is critical for this measurement is not the switching frequency, but the frequency corresponding to the edge rate. For example, to accurately characteriz e a power supply with a switching frequency of 100 kHz and an edge rate of 1 ns, a system with good CMRR at 350 Mhz is necessary because of the edge speed.
TIVH Series Measurement System User Manual 35
Application examples
In this example
, the gate drive voltage could be about 5 V but usually has some ringing and overshoot which are important to characterize. For this measurement, it is appropriate to use a 10X tip with a 10 V
input (on the 1X Range) so that
p-p
the signal is fully resolved and is within the dynamic range of the measurement system.
To get the best CMRR from your TIVH Series IsoVu measurement system, pay careful attention when connecting the measurement system to the DUT. This connection should preserve the signal delity and shield the signal from unwanted interfere
nce. To achieve the best performance from the measurement system, use an MMCX connector as close to the test point as possible. The MMCX connectors are available from a number of vendors and are relatively inexpensive. The key attributes of these connectors, which make them excel for this application, are their compact footprint and solid metal body. The solid metal body and gold contacts provide a w ell-shielded signal path.
TheTIVHseriesIsoVuinputprovidesaoating differential measurement with differential input resistance from 10 Mto 40 M, depending on the tip
ation. In the VGS measurement example the 10X tip with a 10 Minput
attenu impedance is used. The common mode resistance is extremely high, greater than G , in parallel with a small capacitance from the tip cable shield to ground, typically 2 pF or less. Keep these impedances in mind when determining how to connect the measurement system to the DUT. The source node in the half bridge circuit is a very low impedance point, and is the point that should be used to
e the tip cable shield capacitance. The gate driver output, which is also a low
driv impedance node (but not as low impedance as the source), should be used to drive the center contact of the sensor tip cable. The input impedance of the sensor tip cableforthe10Xtipis10MΩ || 6 pF relative to the shield – not to ground.
An example of a measurement that is possible with IsoVu is measuring the high-side turn-on characteristics shown in the following gure.
Figure 26: High-side turn-on characteristics
36 TIVH Series Measurement System User Manual
Application examples
Exampl
e 2: High-side V
In general ther
e a re three characteristic regions of interest of the turn-on waveform
(See Figure 26 on page 36.):
The rst regio
nistheC
charge time.
GS
The second region is the Miller Plateau (the time required to charge the gate-drain
Miller capacitance (C
increases as V
The third re
gion occurs when the channel is in conduction and the gate
increases.
DS
), and is VDSdependent. The charge time
GD
chargesuptoitsnal value.
Due to the r
apid rise of the voltage on the switch node during the high side turn-on, there can be very high frequency and high amplitude common mode voltage changes during the transition. If this common mode voltage transient is not rejected, then the measurement of the high-side V
during the transition is
GS
not possible.
measurement
DS
Another challenging measurement is the high-side VDSmeasurement in a half bridge. Again, like the V
easurement is the high-side source node, which switches between the input
this m
measurement in Example 1, the reference node for
GS
supply voltage and the local PCB ground during operation. In a measurement system without sufcient CMRR, this rapidly changing common mode voltage results in interference which obscures or distorts the measurement.
To get the best CMRR from your TIVH Series IsoVu measurement system, pay careful attention when connecting the measurement system to the DUT. This connection should preserve the signal delity and shield the signal from unwanted interference. To achieve the best performance from the measurement system, keep
e input leads from the V
th
test points to the IsoVu tip cable as short as possible.
DS
The combination of the high CMRR, high input voltage capabilities, and high
andwidth of the TIVH, enables the ability to measure the actual V
b risetime (tr), and V correlated V
(input) and VDS(output) switching characteristics and interactions.
GS
turn-off falltime (tf), and capture the simultaneous time
DS
turn-on
DS
(See Figure 27 on page 38.)
TIVH Series Measurement System User Manual 37
Application examples
Figure 27: VGSand VDSswitching characteristics
Example 3:
High-side drain current measurement
Current sensing is a critical measurement in many applications. Using the half bridge circuit again as an example, measuring the high-side drain current, I be very c mode voltage transients due to parasitic inductance in the supply leads in addition to large current swings. Inserting a conventional current probe into the circuit at this point would require adding excessive inductance, which might limit circuit performance. Using a small value resistor as a current shunt makes it possible to take very high frequency current measurements with minimal additional imped
hallenging particularly during startup. At startup, there can be common
ance in the drain connection. (See Figure 28.)
,can
D
Figure 28: High side current shunt
In a typical application, a 0.25 resistor might be used to measure a transient current of 1 A, resulting in a voltage swing of 0.25 V, which can be measured with the measurement system using a 10X tip. A typical surface mount resistor can have a series inductance of less than 0.2 nH and series capacitance of less than
0.04 pF, resulting in much lower impedance at high frequency than could be obtained with a conventional current probe.
38 TIVH Series Measurement System User Manual
Application examples
Figure 29: SMT resistor model
Go to http://www.vishay.com/docs/60107/freqresp.pdf for models of different types of surface mount resistors.
In general, surface mount resistors have fairly low power ratings; care must be taken not to exceed these ratings when using them as current shunts. Several vendors make very high power parts that can be used when high power dissipation is required. For example, the RP0402CB-R500FN-2Q from Barry Industries (http://www.barryind.com/)isa1.0W,0.50402 resistor on an AI (http://www.usmicrowaves.com/)isa1.5W,1resistor on a BeO substrate.
N substrate, or the RP0402CB-R500FN-2Q from US Microwaves
Example 4: ESD troubleshooting
Many devices and systems can be negatively affected by ESD discha rges. Troubleshooting problems that occur during an ESD discharge can be very difcult. A piece of test equipment connected to a device undergoing ESD testing has to not only withstand the ESD discharge, but it also has to reject interference caused by the rapid changes in potential on the DUT during an ESD discharge test.
For example, consider using the standard human body model of 100 pF and 1500 . If the DUT is being tested to 4 kV and has a capacitance of 50 pF, the
st equipment will be exposed to a voltage change of more than 1 kV in tens
te of nanoseconds.
TIVH Series Measurement System User Manual 39
Application examples
Figure 30: ESD discharge test example
In this example, if it is suspected that there is a glitch on the communication bus betwee the signals on the bus and to inspect them during the discharge. Because the measurement system uses Galvanic isolation, it rejects any interference from the ESD transient and tolerates the high common mode voltage during the discharge; the communication bus can be monitored throughout the ESD discharge and any irregularities can be investigated without interference from the ESD discharge.
n two devices during the ESD discharge, then it is helpful to connect to
40 TIVH Series Measurement System User Manual
Reference information
Specication
s
The following tables list the specications for the measurement system. The specications are guaranteed unless otherwise indicated.
The performance limits in this specication are valid with these conditions:
The instrument must be in an environment with temperature, altitude, and humidity within the operating limits described in these specications.
The instrument must have had a warm-up period of at least 20 minutes.
The measurement system is powered from a TekVPI compatible oscilloscope.
Warranted specications describe guaranteed performance with tolerance limits or certain type-tested requirements.
The performance verication procedures for the Propagation delay are listed later in this document. (See page 66, Propagation delay.)
Table 9: Warranted specications
Characteristic Description
Propagation delay (Warranted) 3 m ber length: 35 ns ±5 ns (actual propagation delay is measured and stored within each unit)
10 m ber length: 68 ns ±7 ns (actual propagation delay is measured and stored within each unit)
le 10: Electrical specications
Tab
Characteristic Description
Controller output termination
Controller output coupling DC coupled
Range attenuation
Terminate the controller output into 50
Sensor tip cable/adapter
Sensor head input SMA 1X (÷1) 2X (÷2)
MMCX10X, 10X tip cable 10X (÷10) 20X (÷20)
MMCX50X, 50X tip cable 50X (÷50) 100X (÷100)
MMCX250X, 250X tip cable 250X (÷250) 500X (÷500)
SQPIN100X, 100X tip cable 100X (÷100) 200X (÷200)
SQPIN500X, 500X tip cable 500X (÷500) 1000X (÷1000)
WSQPIN1000X, 1000X tip cable 1000X (÷1000) 2000X (÷2000)
WSQPIN2500X, 2500X tip cable 2500X (÷2500) 5000X (÷5000)
1X Range 2X Range
TIVH Series Measurement System User Manual 41
Reference information
Table 10: Electrical specications (cont.)
Characteristic Description
Input resistance/capacitance (attached to sensor head, 1M termination), (Typical)
(See page 48, Differential input impedance graphs.)
Maximum non-destructive differential input voltage range, (Typical)
CAUTION. To avoid
damaging the measurement system, be aware of the input voltage limits; this specication applies to both the 1X Range and 2X Range.
(See page 47, Maximum differential input voltage vs. frequency derating graphs.)
Linear differential input voltage range, (Typical)
Sensor tip cable/adapter
Sensor head input SMA 1 M±2%
MMCX10X, 10X tip cable 10 M
MMCX50X, 50X tip cable 10 M
MMCX250X, 250X tip cable 10 M
SQPIN100X, 100X tip cable 10 M
SQPIN500X, 500X tip cable 10 M
WSQPIN1000X, 1000X tip cable 40 M
WSQPIN2500X, 2500X tip cable 40 M
Sensor tip cable/adapter V (DC + pk AC)
Sensor head input SMA
MMCX10X, 10X tip cable
MMCX50X, 50X tip cable
MMCX250X, 250X tip cable
SQPIN100X, 100X tip cable
SQPIN500X, 500X tip cable
WSQPIN1000X, 1000X tip cable
WSQPIN2500X, 2500X tip cable
Sensor tip cable/adapter
Sensor head input SMA
MMCX10X, 10X tip cable
MMCX50X, 50X tip cable
MMCX250X, 250X tip cable
SQPIN100X, 100X tip cable
SQPIN500X, 500X tip cable
WSQPIN1000X, 1000X tip cable
WSQPIN2500X, 2500X tip cable
Resistance
±25 V
±250 V
±250 V
±250 V
±600 V
±600 V
±2500 V
±2500 V
Clamping Off, 1X Range
±V peak (DC + peak AC)
±0.5 V ±1 V
±5 V ±10 V
±25 V ±50 V
±125 V ±250 V
±50 V ±100 V
±250 V ±500 V
±500 V ±1000 V
±1250 V ±2500 V
Capacitance
20 pF
6pF
3pF
2pF
3.5 pF
3.5 pF
3.5 pF
3.5 pF
Clamping Off, 2X Range
±V peak (DC + peak AC)
42 TIVH Series Measurement System User Manual
Table 10: Electrical specications (cont.)
Characteristic Description
Output clamping range (Input referred) (Typical)
Sensor tip cable/adapter Clamping on,
Sensor head input SMA
MMCX10X, 10X tip cable
MMCX50X, 50X tip cable
MMCX250X, 250X tip cable
SQPIN100X, 100X tip cable
SQPIN500X, 500X tip cable
WSQPIN1000X, 1000X tip cable
WSQPIN2500X, 2500X tip cable
Output clamping overdrive recovery (Ty pical)
System noise (input referred) (Typical)
DC gain accuracy1, (Input referred) (Typical)
Differential DC gain accuracy in 1X Range
Differential DC gain accuracy in 2X Range
60% of ±Full Scale
>60% to 80% of ±Full Scale
>80% to 100% of ±Full Scale
<20ns
Sensor tip cable/adapter
TIVH08/TIVH08L Sensor head input SMA
TIVH05/TIVH05L Sensor head input SMA
TIVH02/TIVH02L Sensor head input SMA
Input referred noise with tip cable
Examples:
±3% ± DC offset error voltage ± input offset accuracy error
±3% ±DC offset error voltage ± input offset accuracy error
0% to –4% ±DC offset error voltage ± input offset accuracy error
0% to –7% ±DC offset error voltage ± input offset accuracy error
1X Range
±100 mV ±200 mV
±1 V ±2 V
±5 V ± 10 V
±25 V ±50 V
±10 V ±20 V
±50V ±100 V
±100 V ±200 V
±250 V ±500 V
1X Range 2X Range
<1.2mV
<0.72mV
<0.61mV
rms
rms
rms
(Sensor head input SMA noise) * (Tip cable attenuation)
TIVH08 1X Range with MMCX10X tip cable: Noise = (1.2 mV =12mV
rms
) * (10)
rms
Reference information
Clamping on, 2X Range
<1.4mV
<0.85mV
<0.75mV
(Sensor head input SMA noise) * (Tip cable attenuation)
TIVH08 2X Range with MMCX10X tip cable: Noise = (1.4 mV =14mV
rms
rms
rms
rms
) * (10)
rms
TIVH Series Measurement System User Manual 43
Reference information
Table 10: Electrical specications (cont.)
Characteristic Description
DC offset error voltage
2
(Input referred)
(Typ ical)
Input offset voltage range (Typ ical)
Input offset voltage accuracy (Typical)
SMA input ±1%
With probe tip
DC Reject (AC coupling) Typical)
Small signal rise time (10% to 90%) (Typical)
(SMA input and with sensor tip cables)
Small signal frequency response
(Typ ical)
(SMA input and with sensor tip cables)
Common mode voltage range
Common mode resistance
(Typ ical)
Common mode capacitance
(Typ ical)
Sensor tip cable/adapter
Sensor head input SMA
MMCX10X, 10X tip cable
MMCX50X, 50X tip cable
MMCX250X, 250X tip cable
SQPIN100X, 100X tip cable
SQPIN500X, 500X tip cable
WSQPIN1000X, 1000X tip cable
WSQPIN2500X, 2500X tip cable
3
Sensor tip cable/adapter Input offset voltage range
Sensor head input SMA
MMCX10X, 10X tip cable
MMCX50X, 50X tip cable
MMCX250X, 250X tip cable
SQPIN100X, 100X tip cable
SQPIN500X, 500X tip cable
WSQPIN1000X, 1000X tip cable
WSQPIN2500X, 2500X tip cable
±2%
< 7 Hz (-3 dB) – low frequency cutoff
TIVH08/TIVH08L 435 ps to 700 ps (depending on the tip)
TIVH05/TIVH05L
TIVH02/TIVH02L
TIVH08/TIVH08L DC to 500 MHz or 800 MHz (depending on the tip)
TIVH05/TIVH05L DC to 500 MHz
TIVH02/TIVH02L DC to 200 MHz
60 kV peak
5
N.A. due to Galvanically isolation (ber optic connection)
~2 pF
1X Range 2X Range
±2 mV ±4 mV
±20 mV ±40 mV
±100 mV ±200 mV
±500 mV ±1 V
±200 mV ±400 mV
±1 V ±2 V
±2 V ±4 V
±5 V ±10 V
±25 V
±250 V
±250 V
±250 V
±600 V
±600 V
±2500 V
±2500 V
Rise time
4
700 ps
1.8 ns
-3 dB bandwidth
4
44 TIVH Series Measurement System User Manual
Reference information
Table 10: Electrical specications (cont.)
Characteristic Description
6
Overload indicator range
(Typ ical)
1
The difference between the measured DC gain and the nominal DC gain, divided by the nominal DC gain and expressed as a percent.
2
The input referred offset error voltage when the input is shorted and the probe input offset is set to 0 volts
3
User adjustable input referenced referred DC Offset Voltage Range. It is independent of range and clamp settings.
4
The 800 MHz bandwidth is achieved with the 50X and greater attenuation tips (MMCX50X, MMCX250X, SQPIN100X, SQPIN500X, WSQPIN1000X, or WSQPIN2500X).
5
This product is designed to be used in the circuit with expected transient overvoltage up to 5000 V peak.
6
The Overload indicator is active when the sensor head input DC/LF voltage exceeds the limits in this table. This overload detection will not detect mid and high frequency voltage transients that can damage the sensor head. This overload detection will not detect an overvoltage condition when the sensor head’sDC Reject (AC coupling) is enabled. The limits are independent of range and clamp settings.
Sensor tip cable/adapter Overload indicator On
<-25VorVin>+25V
Sensor head input SMA
tip cables
V
in
V
(Maximum tip voltage) or Vin (Maximum tip voltage)
in
Example: for MMCX10X:
250VorVin 250 V
V
in
Common mode rejection
ratio graphs
The ability to measure common mode rejection ratios (CMRR) of the IsoVu system below 100 kHz is limited by the dynamic range of test systems. Due to the optical isolation of the IsoVu sensor head, the DC CMRR performance of all tip cables is expected to be greater than 160 dB.
The following gure shows the typical CMRR values for the SMA cable and the MMCX10x, MMCX50X, and MMCX250X probe tip cables.
Figure 31: Typical CMRR values for 1X SMA cable and MMCX series probe tip cables
TIVH Series Measurement System User Manual 45
Reference information
The following SQPIN500X, WSQPIN1000X, and WSQPIN2500X square pin tips.
gure shows the typical CMRR values for the SQPIN100X,
Figure 32: Typical CMRR values for square pin probe tips
46 TIVH Series Measurement System User Manual
Reference information
Maximum differential input
voltage vs. frequency
derating graphs
The following MMC250X probe tips.
Figure 33: Maximum differential input voltage vs. frequency for MMCX series probe tips
gure shows the derating values for MMCX10X, MMCX50X, and
The following gure shows the derating values for SQ WSQPIN1000X, and WSQPIN2500X square pin probe tips.
Figure 34: Maximum differential input voltage vs. frequency for square pin probe tips
PIN100X, SQPIN500X,
TIVH Series Measurement System User Manual 47
Reference information
Differential input
impedance graphs
The following for the MMCX10X, MMCX50X, and MMC250X probe tips.
Figure 35: Differential input impedance vs. frequency chart for the MMCX series probe tips
gure shows the differential input impedance vs. frequency values
48 TIVH Series Measurement System User Manual
Reference information
The following
gure shows the differential input impedance vs. frequency values for the SQPIN100X, SQPIN500X, WSQPIN1000X, and WSQPIN2500X probe tips.
Figure 36: Differential input impedance vs. frequency chart for the square pin probe tips
Table 11: Physical specications
cteristic
Chara
Net weight
or tip cables
Sens
sor head
Sen
troller box
Con
kVPI Comp box
Te
nsor tip cable length
Se
Fiber cable length
TIVH08, TIVH02, TIVH05
TIVH08L, TIVH02L, TIVH05L
TekVPI cable length
Overall length and tolerances
Comp box to controller 0.5588 m ±3.81 cm (22 in ±1.5 in) shoulder-to-shoulder, Boot area included in the overall
iption
Descr
ht does not include accessories and packaging.)
(Weig
5kg(0.055lb)
0.02
63 kg (0.8 lb)
0.3
16 kg (1.8 lb)
0.8
57 kg (0.125 lb)
0.
.24cm(6.0in)
15
3 m (9.84 ft)
10 m (32.81 ft)
55.88 cm (22 in)
length.
TIVH Series Measurement System User Manual 49
Reference information
Table 11: Physical specications (cont.)
Characteristic Description
Controller to sensor head (TIVH08, TIVH02, TIVH05)
Controller to sensor head (TIVH08L, TIVH02L, TIVH05L)
2.9718 m ±10.2 cm (117 in ±4 in)
9.982 m ±10.2 cm (393 in ±4 in)
Dimensional drawings
Figure 37: Sensor head dimensions with probe tip cover
50 TIVH Series Measurement System User Manual
Reference information
Figure 38: Sensor head dimensions without probe tip cover
Figure 39: Controller dimensions
TIVH Series Measurement System User Manual 51
Reference information
Figure 40: Comp box dimensions
Figure 41: Probe tip adapter dimensions
52 TIVH Series Measurement System User Manual
Reference information
Figure 42: SQPIN probe tip dimensions
TIVH Series Measurement System User Manual 53
Reference information
Figure 43: WSQPIN probe tip dimensions
54 TIVH Series Measurement System User Manual
IsoVu measurement system block diagram
The following gure shows a block diagram of the IsoVu measurement system.
Figure 44: Block diagram
The common mode resistance and capacitance to earth ground is shown in the gure. (See Figure 44.) The common mode resistance is shown as R parasitic and is essentially innite with the IsoVu measurement system since it is galvanically isolated and can be ignored. The common mode coupling capacitance to earth ground and the surrounding circuit is shown as the Parasitic Bridging Capacitance (Cpara sensor head is placed six (6) inches (15.25 cm) above a ground plane.
sitic). This parasitic capacitance will be approximately 2 pF when the
Reference information
To m in following items:
imize the effects of common mode capacitive loading consider the
ever possible, choose a reference point in the circuit-under-test that is
When static potential with respect to earth ground.
nect the coaxial (common) shield of the sensor tip cable to the lowest
Con impedance point of the circuit.
creasing the physical distance between the sensor head and any conductive
In surface will reduce the parasitic capacitance.
hen using multiple IsoVu systems to measure different points in the circuit
W that do not have the same common mode voltages, keep the sensor heads separated to minimize the capacitive coupling.
TIVH Series Measurement System User Manual 55
Reference information
Tripods
Tektronix provides two tripods as accessories with the me asurement system. The Flexible tripod holds the sensor head while connected to the DUT. The Probe tip tripod supports the sensor tip cables while they are connected to adapters on the circuit board.
Flexible tr
ipod
The tripod can be connected to the DUT in different ways. You can secure the tripod to the DUT using optional feet. The feet can be clamped or attached to the DUT using common screws. This allows you to install the tripod upside down or right side
Figure 45: Installing the exible tripod under the DUT.
up as shown in the following gures.
Figure 46: Connecting the sensor head to the top of the DUT with the exible tripod.
Probe tip tripod
56 TIVH Series Measurement System User Manual
Use the probe tip tripod to connect the sensor tip cables to adapters on the circuit board. This tripod has exible hinges to easily position the probe tip cable above the adapter on the circuit board. Tektronix recommends gluing the tripod in place on the circuit board to provide additional support for the sensor tip cables. The following gure shows an example of connecting to an adapter on the circuit board with the tripod; it reduces stresses on the test point.
Reference information
Figure 47: Connecting to an adapter on the circuit board with the probe tip tripod
TIVH Series Measurement System User Manual 57
Reference information
Connecting th
e sensor tip cables
Tektronix provides different types of sensor tip cables to connect to the circuit board.
The MMCX sensor tip series connect directly to MMCX connectors on the circuit board or to probe tip adapters installed on the circuit board. (See page 59,
Installing
The square pin series sensor tip cables connect directly to the circuit board as shown in th
the probe tip adapters.)
e following gure.
Figure 48: Square pin series sensor tip cables
The SQPIN series sensor tip cables connect to 0.1-inch (2.54 mm) square pins
he circuit board. The WSQPIN series sensor tip cables connect to 0.2-inch
on t (5.08 mm) square pins on the circuit board.
58 TIVH Series Measurement System User Manual
Installing the probe tip adapters
Tektronix provides two probe tip adapters to connect the MMCX sensor tip cables to pins on the circuit board. The MMCX-to-0.1-inch (2.54 mm) pitch adapter and the MMCX-to-0.062-inch (1.57 mm) pitch adapter.
One end of each adapter has an MMCX socket for connection to an IsoVu MMCX tip cable. The other end of the adapter has a center pin socket and four common (shield) sockets around the outside of the adapter. Notches on the adapters can be used to loc the same, the main difference is the spacing of the pins on the circuit board.
ate the shield sockets. The procedure for installing these adapters are
Reference information
To insta l the signal source pin on the circuit board. Use the notch on the adapter to align one of the shield sockets to the common pin on the circuit board. The following gures show examples of lining up the adapters on the circuit board.
To achieve the best electrical performance, especia lly the CMRR performance and EMI susceptibility, place the probe tip adapter as close as possible to the circuit board.
l the adapters onto the square pins, line up the center of the adapter with
Figure 49: Lining up the MMCX-to-0.1-inch (2.54 mm) adapter on the circuit board
TIVH Series Measurement System User Manual 59
Reference information
Figure 50: Lining up the MMCX-to-0.062-inch (1.57 mm) adapter on the circuit board
After lining up the adapters gently push down on the adapter to seat it in place on the circuit board.
Figure 51: Pushing the MMCX-to-0.062-inch (1.57 mm) adapter in place
60 TIVH Series Measurement System User Manual
Reference information
Figure 52: Pushing the MMCX-to-0.1-inch (2.54 mm) adapter in place
When the adapters are rmly in place on the circuit board connect the sensor tip cable to the top of the adapter while using the probe tip tripod to ease the tension off the probe tip cable and adapter. (See Figure 47 on page 57.)
TIVH Series Measurement System User Manual 61
Reference information
Installing the square pins on the circuit board
The following gure shows the recommended cle arance r equire ments for connecting the adapters to the square pins on the circuit board. The bottoms of the adapters are shown at the top.
Figure 53: Adapter clearance requirements
62 TIVH Series Measurement System User Manual
Reference information
The 0.025-inch board. Some square pins might have headers installed on the circuit board. Tektronix recommends removing the plastic spacer from the square pins to gain closer access to the circuit board as shown in the following gure to achieve the best electrical performance, especially CMRR. You might need to use a pair of tweezers to remove the spacer as shown in the gure.
(0.635 mm) square pins should already be located on the circuit
Figure
Tektr on the circuit board for use with the MMCX to 0.062-inch (1.57 mm) adapter. Use the soldering aide tool accessory (Tektronix part number, 003-1946-xx) to install these pins on the circuit board.
NOTE. The solder pins are extremely small and can be challenging to handle.
Tektronix recommends using tweezers and a magnifying tool when installing the
The solder pins can be installed around a surface mounted component on the ci connection for the adapter. (See Figure 53 on page 62.)
N
should always be connected to the lowest impedance point (usually a circuit common or power supply rail) in the circuit-under-test (relative to the sensor tip cable/center conductor) to obtain the most accurate waveform.
54: Removing the header from square pins on the circuit board
onix provides a set of solder pins (0.018-inch (0.46 mm) diameter) to install
pins on the circuit board.
rcuit board, but adequate clearance should be maintained for a good electrical
OTE. The coaxial (common) shield of the sensor tip cable and tip adapters
TIVH Series Measurement System User Manual 63
Reference information
Use the followi circuit board:
1. Carefully ins following illustration.
ng steps to install the solder pins using the soldering aide on the
ert t he solder pins into the soldering aide as shown in the
Figure 55: Using the soldering aide to install the square pins on the circuit board
2. Use the soldering aide to hold the square pins in place while soldering the square pins to the circuit board.
3. If necessary apply a small amount of adhesive to further strengthen the connection to the c i rcuit board. However, keep the height of the adhesive to a minimum to provide good electrical contact for the adapter. (See Figure 53.)
64 TIVH Series Measurement System User Manual
User service
Service offerings
Tektronix provides service to cover repair under warranty and other services that are designed to meet your specic service needs.
Whether providing warranty repair service or any of the other services listed below, Tektronix service technicians are well equipped to service the IsoVu measurement system. Services are provided at Tektronix Service Centers and on-site at your facility, depending on your location.
Warranty
repair service
Calibration and repair
service
Preventive maintenance
Tektronix warrants this product as described in the warranty statements at the front of this manual. Tektronix technicians provide warranty service at most Tektronix service locations worldwide. The Tektronix Web site provides information on all service
In addition to warranty repair, Tektronix Service offers calibration and other services that provide cost-effective solutions to your service needs and quality stand worldwide by the leading-edge design, manufacturing, and service resources of Tektronix to provide the best possible service.
CAUTION. To prevent damage to the measurement system, do not expose it to
spr head when cleaning the exterior.
Cl dirt remains, use a soft cloth or swab dampened with a 75% isopropyl alcohol solution. Use only enough solution to dampen the cloth or swab. Do not use abrasive compounds on any part of the instrument.
locations worldwide.
ards compliance requirements. Tektronix instruments are supported
ays, liquids, or solvents. Avoid getting moisture inside the controller or sensor
ean the exterior surfaces with a dry, lint-free cloth or a soft-bristle brush. If
TIVH Series Measurement System User Manual 65
User service
Performance v
erication procedures
Use the following procedures to verify the performance of the IsoVu measurement system. Before beginning the procedures, photocopy the test record and use it to
rformance results. (See page 73, Test record.)
Required equipment
record the pe
The equipment r equired to perform the performance verication procedures are shown in the following table.
Table 12: R
Descript
Oscillos
Pulse ge
TIVH ser
MMCX Jack (female) to BNC female adapter
Termination As per description Tektronix 011–0049–xx
equired equipment for performance verication
ion
cope with TekVPI interface
nerator
ies 10X sensor tip cable
Preparation
Minimum r
1GHz,50 with TekVPI interface
1V
As per de
As per d
Prepare the equipment as follows:
1ns risetime
,<
p-p
equirements
input support, fully compatible
scription
escription
Example p
Tektroni
Tektronix Tek-DPG
Tektronix MMCX10X
Fairview Microwave Product SKU: SM3610
roduct
x MDO4104C
Propagation delay
1. Turn on the TekVPI o scilloscope.
2. Con
nect the Tek-DPG Deskew pulse generator into Channel 2 of the TekVPI
oscilloscope.
3. All
ow the test equipment to warm up for 20 minutes at an ambient temperature
of 20 °C to 30 °C.
This procedure veries that the TIVH Series IsoVu measurement system is functioning and me ets the warranted propagation delay specication. The propagation delay will be measured by rst applying the output of the pulse generator to the input of the o scilloscope and storing the captured waveform as a reference waveform. Next, the measurement system is connected to the oscilloscope and the output of the pulse generator is connected to the input of the measurement system. The delay is then measured between the saved reference waveform and the acquired waveform of the measurement system.
NOTE. This procedure is valid for all versions of the TIVH Series IsoVu
measurement systems.
66 TIVH Series Measurement System User Manual
User service
Create t he reference
waveform
Complete the fo
1. Connect the Tek-DPG Deskew Pulse Generator output BNC cable directly to Channel 1 input of the TekVPI oscilloscope.
2. Enable Channel 2 and use the following setups:
Vertical Scale: 500 mV/div.
Vertical Position: -3 divisions
Set Termination: 1MΩ,
Coupling: DC
Bandwidth: FULL
Deskew: 0 seconds
3. Set the Trigger menu using the following setups:
Type: Edge
Source: CH2
Slope: Negative
Level: +1.50 V
llowing steps to create the reference waveform:
Coupling: DC
4. Use the following setups for Channel 1.
Vertical Scale: 200 mV/div
Coupling: DC
Termination: 50
Bandwidth: FULL
Position: 0 (centered)
Offset: -500 mV
Deskew: 0 seconds
5. Set the Horizontal menu to the following setups:
Horizontal Scale to 10 ns/div.
Horizontal Position: 40 ns
Acquire: Average 128
6. Set the Tek-DPG to the following setups:
0to-11kHzmode(Mode 1)
Output Enable: ON
TIVH Series Measurement System User Manual 67
User service
CreatetheTIVHseries
waveform
7. The Channel 1 wa oscilloscope display. If the waveform is not vertically centered in the display, adjust the display as necessary to vertically center the waveform.
8. Use the following steps to save the Channel 1 waveform as the reference waveform (R1).
Press MENU on the oscilloscope.
Select Save
Source: CH1.
Destination: R1.
Select OK Save to save the Channel 1 waveform as reference waveform
R1. The new reference waveform should now be displayed on the oscilloscope.
9. Disable the output of the Tek-DPG.
10. Disconnect the Tek-DPG Deskew Pulse Generator output BNC cable from
the Channel 1 input of the TekVPI oscilloscope.
Complete the following steps to s et up the TIVH series waveform:
1. Connect the Comp box of the TIVH series measurement system to Channel 1 of the TekVPI oscilloscope.
veform should appear approximately on the center of the
Waveform.
2. Connect an MMCX10X sensor tip cable to the TIVH series input (tighten the
SMA connector from the cable to the sensor head and attach the nose cone.)
3. Allow the measurement system to warm up for 20 minutes at an ambient temperature of 20 ° to 30 °C.
4. Connect an MMCX Jack (female) to BNC female adapter to the 50 termination and to the output BNC cable of the Tek-DPG.
5. Set the TIVH series to the following:
ange: 1X
R
CLAMPING: Off
Push the SELF CAL button to perform the self calibration (wait for the
SELF CAL Status indicator to turn solid green).
6. On Channel 1 of the oscilloscope set the Vertical Scale to 200 mv/div.
7. Connect the MMCX10X sensor tip cable to the MMCX Jack (female) to
BNC female adapter.
68 TIVH Series Measurement System User Manual
User service
Measure the propagation
delay
8. Enable the outp
9. The Channel 1 waveform should appear on the oscilloscope display and be
approximatel vertical settings as necessary to center the waveform on the display.
Complete the following steps to measure and record the propagation delay.
1. Set up the delay measurement on the oscilloscope a s follows:
a. Select Measure.
b. Select Add Measurement.
c. Select Measurement Type: Delay.
d. Select C
e. Set the Source to R1.
f. Set the Delay to CH1.
g. Select OK Add Measurement.
2. The oscilloscope display should be similar to the following gure.
ut of the Tek-DPG.
y vertically centered on the display. If not, ONLY adjust the
ongure: Delay.
Figure 56: Propagation delay measurement
Record the resulting delay measurement on the test record.
3. Disable the output of the Tek-DPG.
TIVH Series Measurement System User Manual 69
User service
Troubleshooting and error conditions
The following table lists possible problems that you might encounter when taking measurements with TIVH Series IsoVu measurement system. Use the table as a quick troubleshooting reference before contacting Tektronix for service.
Table 13: Problems and possible solutions
Problem Remedy
The measurement system will not power on; no indicators are on.
Controller STATUS indicator is ashing green.
Controller STATUS indicator is ashing red and yellow
Controller SELF CAL STATUS indicator turns solid red after pressing SELF CAL button (or never completes).
Verify the TekVPI comp box is securely attached to the oscilloscope. Detach and then reattach the comp box (using a different oscilloscope channel if necessary). If the condition persists, return the system to Tektronix for service.
Detach and then reattach the comp box (using a different oscilloscope channel if necessary). Do not force the comp box into the oscilloscope connector at an angle; connect it to the oscilloscope with a rm, horizontal force. If the condition persists, return the system to Tektronix for service.
This condition indicates a failure in the measurement system. Most often, a failure is detected when the m easurement system is rst connected to the oscilloscope (while executing power-on self tests). Detach and then reattach the comp box to the scope. If the condition persists, return the system to Tektronix for service.
Remove input signal (differential must be 0.0V).
Allow system to warm up 20 minutes.
Make sure the ber cable is not subjected to dynamic mechanical or thermal stresses.
If the self calibration does not complete in a minute, detach and reattach the comp box to
the oscilloscope – then try again.
Waveform distortion when the signal from the D UT is less than 5 Hz or a multiple of 43.5 Hz
Output waveform is clipped or distorted
Frequency response is rolled off Check the bandwidth limit on the oscilloscope.
Output waveform is unstable (low frequency noise and/or DC offset keep changing)
Disable the Offset Correction. Refer to the Offset correction section earlier in this document. (See page 16.)
Verify that the Output Clamp is disabled.
Check that the proper sensor tip cable is being used for the measurement. Refer to
Selecting a sensor tip cable section earlier in this document. (See page 20.)
Change the Input Offset to center the signal on-screen.
Make sure the correct Range is selected (1X or 2X).
Check the sensor tip cable continuity and sensor head input resistance (see below).
Make sure the ber cable is not subjected to dynamic mechanical and thermal stresses.
Verify that Offset Correction is enabled (detach and then reattach the comp box to make
sure it has been re-enabled).
70 TIVH Series Measurement System User Manual
Table 13: Problems and possible solutions (cont.)
Problem Remedy
Measurements are inaccurate
Cannot achieve high CMRR Try a different sensor tip cable.
Too much noise and cannot accurately resolve signal
No signal detected; waveform is a at line
Run SELF CAL.
Check that the proper sensor tip cable is being used for the measurement. Refer to
Selecting a sensor tip cable section earlier in this document. (See page 20.)
Make sure the correct Range is selected (1X or 2X).
Make sure the signal is on-screen.
Switch to the 1X Range.
Choose a sensor tip cable with less attenuation.
Check the sensor tip cable continuity:
Sensor tip cable
MMCX10X - 10X sensor tip cable 9 M
MMCX50X - 50X sensor tip cable 10 M
MMCX250X - 250X sensor tip cable 10 M
SQPIN100X - 100X sensor tip cable 10 M
SQPIN250X - 100X sensor tip cable 10 M
WSQPIN1000X - 1000X sensor tip cable 40 M
WSQPIN2500X - 2500X sensor tip cable 40 M
Measure the SMA input resistance of the Sensor Head; it should be between 950 kand
1.05 M . If not, then the Sensor Head has been damaged and needs to be returned to Tektronix for service.
User service
Resistance
Check the controller STATUS indicator for possible error conditions.
Apply a test signal directly to the SMA input of sensor head to determine if the problem is
the sensor tip cable or the sensor head.
Large DC offset in waveform Run SELF CAL.
Run AutoZero.
SetInputOffsetto0.0V.
Cannot select between 1X and 2X Range (5000/7000/70000 Series oscilloscopes)
Probe compensation fails (7000/70000 Series oscilloscopes)
When Auto Range is turned ON, the Range is selected automatically when the V/div setting is changed. You cannot change the Range directly (the button will appear to not function).
This is a normal condition. You c an ignore the failure.
TIVH Series Measurement System User Manual 71
User service
Repack the m ea
surement system for shipment
If you need to return the measurement system to Tektronix for repair, use the original packaging. If this is unavailable or not t for use, contact your Tektronix representat
When you return the measurement system to Tektronix, attach a tag showing the following i
Name of the product owner
Address of the owner
Instrument serial number
A description of problems encountered and/or service required
ive to obtain new packaging.
nformation:
72 TIVH Series Measurement System User Manual
Test record
Table 14: Test re cord
User service
Photocopy this test record for recording the results of the performance verication procedures.
Model number:
Serial numb
Temperature:
Propagatio
TIVH08 (3 m
TIVH08L (1
TIVH02 (3
TIVH02L
TIVH05 (
TIVH05
er:
n delay
0mber length)
m ber length)
(10 m ber length)
3mber length)
L(10mber length)
ber length)
Certicate
RH %:
Technician:
Date of cali
Minimum Incoming Outgoing Maximum
30 ns 40 ns
61 ns
30 ns 40 ns
61 ns
30 ns 40 ns
61 ns
number:
bration:
75 ns
75 ns
75 ns
TIVH Series Measurement System User Manual 73
User service
74 TIVH Series Measurement System User Manual
Appendix A: Remote programming
This appendix describes commands and queries that can be sent to the sens or head when attached to a Tektronix oscilloscope. Long-form and short-form keywords are indicated with upper/lower case letters. The commands and queries are supported by most oscilloscopes; differences in s upporting oscilloscopes, if any, are des
For details on the command syntax, refer to the programmer documentation for your oscil
CH<n>:PRObe?
cribed with the commands.
loscope.
Returns p
robe information for channel <n>. Query only.
CH<n>:PRObe:AUTOZero EXECute
This com by the measurement system and then by the oscilloscope. Command only.
The sys calibration before the oscilloscope does its AutoZero.
Refer calibration. (See page 14, Self calibration.)
mand executes the A utoZero function. The operation is rst performed
tem can be congured (with special keystrokes) to perform a full self
to the self calibration procedure for information on performing the self
CH<n>:PRObe:COMMAND “CLAMP”, {“ON” | “OFF”}
Only supported by 3000/4000 Series oscilloscopes.
command enables or disables the clamp circuitry. The parameters enclosed in
The quotes are case sensitive and must be sent as capital letters.
<n>:PRObe:COMMAND? “CLAMP” . The query returns either “ON” or
CH “OFF” enclosed in quotes.
CH<n>:PRObe:SET {“CLAMP ON” | “CLAMP OFF”}
Only supported by 5000/7000/70000 Series oscilloscopes.
The command enables or disables the clamp circuitry. The parameters enclosed in quotes are case insensitive.
CH<n>:PRObe:SET?. The query returns either “CLAMP ON” or “CLAMP OFF” enclosed in quotes.
TIVH Series Measurement System User Manual 75
Appendix A: Remote programming
CH<n>:PRObe:
CH<n>:PRO
be:FORCEDRange <NR3>
SET {“CLAMPING ON” | “CLAMPING OFF”}
Only supported by MSO 5 Series oscilloscopes.
The command enables or disables the clamp circuitry. The parameters enclosed in quotes are case insensitive.
CH<n>:PRObe:SET?. The query returns either “CLAMPING ON” or “CLAMPING OFF” enclosed in quotes.
The command selects the dynamic range of the sensor tip in V the attached sensor tip cable.
The following table lists the sensor tip cables and the dynamic range.
Table 15: Sensor tip cables and dynamic ranges
Sensor tip cable Dynamic Range V
No tip 1.0 or 2.0
10X 10.0 or 20.0
50X 50.0 or 100.0
100X 100.0 or 200.0
250X 250.0 or 500.0
500X 500.0 or 1000.0
1000X 1000.0 or 2000.0
2500X 2500.0 or 5000.0
and is dependent
p-p
p-p
CH<n>:PRObe:FORCEDRange? The query returns the dynamic range of the sensor tip in V
p-p
.
CH<n>:PRObe:GAIN?
Returns the gain factor of the sensor tip (inverse of attenuation). Dependent on the
ttached sensor tip cable. Query only.
a
76 TIVH Series Measurement System User Manual
Appendix A: Remote programming
CH<n>:PRObe:
CH<n>:
PRObe:PROPDELay?
ID {:SERnumber | :TYPe}?
Query only. When sending only PRObe:ID?, the query returns the sensor tip type string followed by the serial number string.
When sending PRObe:SERnumber?, the query returns the serial number string.
When sending PRObe:TYPe?, the query returns one of the following sensor tip type strings (note the trailing spaces within the return strings):
“TIVH08 ” (800 MHz, 3m)
“TIVH08L” (800 MHz, 10m)
“TIVH05 ” (500MHz, 3m)
“TIVH05
“TIVH02 ” (200MHz, 3m)
“TIVH02L” (200M Hz, 10m)
Only supported by 3000/4000 Series oscilloscopes.
L” (500MHz, 10m)
Returns the propagation delay value (in seconds). Query only.
CH<n>:PRObe:RECDESkew?
Only supported by 3000/4000 Seriesoscilloscopes.
Returns the recommended deskew value (in seconds). Query only.
CH<n>:PRObe:RESistance?
Returns the input resistance (in ohms). It depends on the attached sensor tip cable. Query only.
CH<n>:PRObe:UNIts?
Returns the units of the sensor tip (always “V”). Query only.
TIVH Series Measurement System User Manual 77
Appendix A: Remote programming
CH<n>:PROBEC
Ontrol {AUTO | MAN}
Only supported by 5000/7000/70000/MSO 5 Series oscilloscopes.
The command sets the Auto or Manual sensor tip range control.
CH<n>:PROBE:PROBECOntrol?. The query returns the keyword AUTO or MANUAL.
CH<n>:PROBEFunc:EXTAtten <NR3>
Only supported by 5000/7000/70000/MSO 5 Series oscilloscopes.
The command sets user-dened external attenuation factor.
CH<n>:PROBEFunc:EXTAtten?. The query returns the external attenuation factor.
CH<n>:PROBEFunc:EXTDBatten?
Only supported by 5000/7000/70000/MSO 5 Series oscilloscopes.
Returns the external attenuation factorexpressedindB.Queryonly.
78 TIVH Series Measurement System User Manual
Appendix B: Compliance information
This section lists the Safety and Environmental standards with which the instrument complies. This product is intended for use by professionals and trained personnel on
Questions about the following compliance information may be directed to the following a
Tektronix, Inc. PO Box 500, Beaverton, OR 97077, USA www.tek.com
Safety compliance
This section lists the safety standards with which the product complies and other safety compliance information.
ly; it is not designed for use in households or by children.
ddress:
MS 19-045
EU low voltage directive
U.S. nationally recognized
sting laboratory lis ting
te
Canadian certication
Compliance was demonstrated to the following specication as listed in the Ofcial Journal of the European Union:
Low Voltage Directive 2014/35/EU.
010-1. Safety Requirements for Electrical Equipment for Measurement,
EN 61 Control, and Laboratory Use – Part 1: General Requirements.
1010-031. Particular requirements for handheld probe assemblies for
EN 6 electrical measurement and test eq uipment (partially applicable).
UL 61010-1. Safety Requirements for Electrical Equipment for Measurement,
ntrol, and Labora tory Use – Part 1: General Re quirements.
Co
UL 61010-031. Particular requirements for handheld probe assemblies for
lectrical measurement and test equipment (partially applicable).
e
CAN/CSA-C22.2 No. 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
CAN/CSA-C22.2 No. 61010-031. Particular requirements for handheld probe assemblies for electrical measurement and test equipment (partially applicable).
TIVH Series Measurement System User Manual 79
Appendix B: Compliance information
Additional compliances
Equipmen
ttype
Pollution degree
descriptions
IEC 61010-1. Sa Measurement, Control, and Laborat ory Use – Part 1: General Requireme nts.
UL 61010-031. electrical measurement and test equipment (partially applicable).
EN 60825-1. Requirements - Edition 3 (2014)
IEC 60825-1 and Requirements - Edition 3 (2014)
US 21CFR PT
US 21CFR PT1040 Performance Standards for Light Emitting Products 2015.
Test and measuring equipment.
A measure of the contaminants that could occur in the environment around and within a product. Typically the internal environment inside a product is
eredtobethesameastheexternal. Productsshouldbeusedonlyinthe
consid environment for which they are rated.
tion degree 1. No pollution or only dry, nonconductive pollution occurs.
Pollu Products in this category are generally encapsulated, hermetically sealed, or located in clean rooms.
fety Requirements for Electrical Equipment for
Particular requirements for handheld probe assemblies for
Safety of Laser Products-Part 1: Equipment Classication and
. Safety of Laser Products-Part 1: Equipment Classication
1010 Performance Standard for Electronic Parts 2015.
Pollution degree rating
IP rating
Pollution degree 2. Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is caused by condensation must be expected. This location is a typical ofce/home environment. Temporary condensation occurs only when the product is out of service.
Pollution degree 3. Conductive pollution, or dry, nonconductive pollution that becomes conductive due to condensation. These are sheltered locations where neither temperature nor humidity is controlled. The area is protected
rom direct sunshine, rain, or direct wind.
f
Pollution degree 4. Pollution that generates persistent conductivity through conductive dust, rain, or snow. Typical outdoor locations.
Pollution degree 2 (as dened in IEC 61010-1). Rated for indoor, dry location use only.
IP20 (as dened in IEC 60529).
80 TIVH Series Measurement System User Manual
Appendix B: Compliance information
Measurement and
overvoltage category
Environ
mental considerations
Product end-of-life
descriptions
handling
Measurement te from one or more of the following categories (see specic ratings marked on the product and in the manual).
Category I. Circuits not directly connected to a mains supply.
Category II. Circuits directly connected to the building wiring at utilization points (socket outlets and similar points).
Category III. In the building wiring and distribution system.
Category I
NOTE. Only measurement circuits have a measurement category rating. Other
circuits within the product do not have either rating.
This section provides information about the environmental impact of the product.
Observe the following guidelines when recycling an instrument or component:
Equipment recycling. Production of this equipment required the extraction and use of natural resources. The equipment may contain substances that could be
ful to the environment or human health if improperly handled at the product’s
harm end of life. To avoid release of such substances into the environment and to reduce the use of natural resources, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
rminals on this product may be rated for measuring mains voltages
V. At the source of the electrical supply to the building.
s symbol indicates that this product complies with the applicable European
Thi Union requirements according to Directives 2012/19/EU and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries.
r information about recycling options, check the Tektronix Web site
Fo (www.tek.com/productrecycling).
TIVH Series Measurement System User Manual 81
Appendix B: Compliance information
82 TIVH Series Measurement System User Manual
Loading...