Getting help and support............................................................................................................................................................. 11
Related documentation........................................................................................................................................................ 11
Required equipment and accessories..................................................................................................................................12
Installing the software.......................................................................................................................................................... 12
Launch the application......................................................................................................................................................... 14
Close the application............................................................................................................................................................14
Help panel............................................................................................................................................................................21
BER Test.......................................................................................................................................................................62
The TBT3/TBT4 Rx application performs the test as per the Gen 2 / Gen 3 USB4 Electrical CTS Revision 1.02.
Welcome
Figure 1: TekRxTest - Thunderbolt4 Gen 2 / Gen 3 Rx application
Key features and benefits
•Calibration and Compliance Testing comes as a part of the receiver solution.
•Automated Calibration procedure is handy along with the ability to reload and rerun older Calibration files.
•Receiver BER, Signal Frequency Variations Training (SFVT), Jitter Tolerance (JTOL), and Sensitivity test for the DUT using the
Electrical Testing Tool (ETT) from USB-IF.
•Jointly with Anritsu BERT MP1900A series, the receiver solution provides the tools and flexibility required to visualize and control the
impairments, observe real-time eye performance for Thunderbolt3/4 devices at 10, 10.3125, 20 and 20.625 Gbps.
•The solution showcases the eye diagram at TP3’ and TP3 endpoints.
•Reliable and accurate results reduce the test execution time and minimize the skillset required to perform calibration and testing.
•Available tests can be run at TP3’ (Case 1) and TP3 (Case 2) test points and come with crosstalk feature.
•Detailed reports are at one’s disposal for all calibration and test modules.
Thunderbolt Receiver Test Application Manual10
Getting help and support
Getting help and support
Related documentation
The following documentation is available as part of the TBT3/4 Gen 2 / Gen 3 Rx receiver test application.
Table 1: Product documentation
ItemPurposeLocation
Application HelpApplication operation and User Interface
details
Technical support
Tektronix values your feedback on our products. To help us serve you better, please send us your suggestions, ideas, or comments on
your application or Real Time Oscilloscope. Contact Tektronix through mail, telephone, or the Web site. See Contacting Tektronix.
When you contact Tektronix Technical Support, please include the following information (be as specific as possible):
General information
Help panel of the application
•All instrument model numbers
•Hardware options, if any
•Modules used
•Your name, company, mailing address, phone number, and FAX number
•Please indicate if you would like to be contacted by Tektronix about your suggestion or comments.
Application specific information
•Software version number
•Description of the problem such that technical support can duplicate the problem
•If possible, save the setup files for all the instruments used and the application.
Thunderbolt Receiver Test Application Manual11
Getting started
Getting started
Required equipment and accessories
This section lists the accessories and test fixtures required to perform the tests.
Table 2: Required equipment and accessories
EquipmentVendorTypeR/OQtyDescription
MP1900AAnritsuEquipment Required1BERT
DPO72304SX or
DPO72304DX or Oscilloscope
of higher bandwidth
CIO – DPOJET pluginTektronixSoftwareRequired1Pre-req option for TBT3/4
DIA-DPOJET Advanced
option
SDLA64TektronixSoftwareRequired1Pre-req option for TBT3/4
SM8852Fairview MicrowaveAccessory Required62.92mm (female) to SMP (female) Cable or
PCIe Gen4 ISI FixturePCI-SIGAccessory Required1This will be replaced when an approved
0.8 m and 2 m USB Type-C
cables
RXSW-NLP-TBT34 orTektronixSoftwareRequired1License; Thunderbolt 3 and 4 Receiver
RXSW-NL1-TBT34 orLicense; Thunderbolt 3 and 4 Receiver
RXSW-FLP-TBT34 orLicense; Thunderbolt 3 and 4 Receiver
RXSW-FL1-TBT34License; Thunderbolt 3 and 4 Receiver
TektronixEquipment Required1Tektronix Real time Oscilloscope Bandwidth
≥21GHz, ≥2-channel oscilloscope
TektronixSoftwareRequired1Pre-req option for TBT3/4
Adapter
version is made available
Any USB-IF approved
cable
Accessory Required1 eachUSB Type C Cables
automation software for TEK scopes and
Anritsu BERT; Perpetual; Node-Locked
automation software for TEK scopes and
Anritsu BERT; 1 year subscription; NodeLocked
automation software for TEK scopes and
Anritsu BERT; Perpetual; Floating
automation software for TEK scopes and
Anritsu BERT; 1 year subscription; Floating
Installing the software
Follow the below steps to download and install the latest TBT3/4 Gen 2 / Gen 3 TekRxTest application.
1. Go to www.tek.com.
Thunderbolt Receiver Test Application Manual12
Getting started
2. Click Downloads. In the Download menu, select DOWNLOAD TYPE as Software and enter TBT3/4 Gen 2 / Gen 3 Rx in the MODEL
OR KEYWORD field and click SEARCH.
3. Select the latest version of the software and follow the instructions to download.
4. Copy the executable file into the instrument to install the software (Real-time oscilloscope or PC).
5. Follow the installation instructions that is available in the website. The software is installed at
C:\ProgramFiles\Tektronix\BERTScope\RxTest60.
6. Click the shortcut icon on the desktop to launch the application.
Note:
•The TBT3/4 Gen 2 / Gen 3 TekRxTest application can be installed on a Tektronix real-time oscilloscope or a PC (Optional).
•You must install the TekRxService application in the real-time oscilloscope to successfully connect the application with the
real-time oscilloscope.
Thunderbolt Receiver Test Application Manual13
Operating basics
Operating basics
Launch the application
To launch the TBT3/4 Gen 2 / Gen 3 TekRxTest application, click the shortcut icon TekRxTest on the desktop and select TBT3 or TBT4
Gen 2 / Gen 3 in the application window.
Close the application
To exit the application, click on the application title bar. Follow on-screen instructions to save the unsaved session or test setup.
Using other methods to exit the application may result in abnormal termination of the application.
Note:
Launch Real-Time Oscilloscope
The TekVISA Socket Server application on the oscilloscope provides the necessary connectivity between the TekRxTest application
and scope. Although it is launched in the background when the scope boots up and the socket is initialized for communication, it is
recommended to verify the status by clicking on the Desktop Tray ―› TekVISA LAN Server Control as shown in the image below. If it is
ready to exchange data, then a wizard would appear as in the below image.
Thunderbolt Receiver Test Application Manual14
Figure 2: Launch Real-Time Oscilloscope
Note: In the unlikely event when the socket is not initialized, the process can be started by clicking on “Start Socket Server” which
gets enabled during such a scenario.
Operating basics
Launch TekRxService
The TBT3/4 Gen 2 / Gen 3 TekRxTest application interfaces with the oscilloscope for data acquisition, analysis, and data retrieval utilizing
TekRxService application. TekRxService should be launched from the oscilloscope before initiating a connection between the oscilloscope
and TekRxTest application.
Thunderbolt Receiver Test Application Manual15
Application panels
Application panels
Application panels overview
The TBT3/4 Gen 2 / Gen 3 receiver test application uses panels to group the configurations and settings. Click on any panel to configure
the associated settings. A panel may have one or more tabs that lists the selections available in that panel. Controls in a tab may change
depending on the settings made in the same tab or another tab.
Figure 3: Application panels overview
Table 3: Application panels overview
ParameterDescription
ConnectionsThis panel displays the Real-Time Oscilloscope and Bit Error Rate Tester (BERT) connection settings.
You can connect to a real-time oscilloscope and BERT by entering the IP address of the instruments.
SettingsThis panel allows configuring various settings for the RT Scope, Analysis Tool, and the Remote Access.
HelpThis panel displays the application help.
CalibrationsThis panel allows you to configure the calibration parameters for TP3', TP3, and SFV and save the
results.
TestsThis panel allows you to configure the JTOL, Sensitivity, BER, and SFVT test settings and view the
results.
Thunderbolt Receiver Test Application Manual16
Application panels
Connections panel
The Connections panel allows you to connect to a real-time oscilloscope and BERT with the TBT3/4 Gen 2 / Gen 3 TekRxTest application.
Enter the IP address of these instruments and click Connect to establish the connection.
Figure 4: Connections panel
Table 4: Connections panel
ConnectionsDescription
BERT
RT Scope
Enter the BERT IP address in the address field and click Connect. When the BERT is connected
successfully, the circle next to BERT in the right end corner turns green.
Note: It is recommended to launch the MP1900A software in the administrator mode to use the
TekRxTest Application.
Enter the RT Scope IP address in the address field and click Connect. When the RT Scope is connected
successfully, the circle next to RT Scope and TekRxService in the right end corner turns green.
Note: Before you click Connect, you must launch the TekRxService in the real-time oscilloscope.
Settings panel
The Settings panel allows you to configure the settings for instruments, analysis tool, and remote access. Click on any tab to configure the
associated settings.
Thunderbolt Receiver Test Application Manual17
Application panels
Figure 5: Settings panel
Table 5: Settings panel configurations
ParameterDescription
Restore DefaultsRestores the application with default settings.
SaveSaves the current test setup.
RecallRecalls the saved test setup.
Basic settings
The basic settings display the parameters for RT Scope, Analysis Tool, and Remote access.
Thunderbolt Receiver Test Application Manual18
RT Scope
Application panels
Figure 6: RT Scope
Table 6: RT Scope
ParameterDescription
Positive ChannelSelect the generator data positive channel from BERT.
Negative ChannelSelect the generator data negative channel from BERT.
Sample RateDisplays the sample rate in GS/s used for the selected channels.
Note: Select appropriate channels (Ch1 and Ch3 or Ch2 and Ch4) which allow 100 GS/s sample rate.
Thunderbolt Receiver Test Application Manual19
Analysis Tool
Application panels
Figure 7: Analysis Tool
Table 7: Analysis Tool
ParameterDescription
DPOJETDisplays the jitter and eye measurements in TP3' and TP3 calibration.
SigTest
Displays the signal quality measurements captured using a real-time oscilloscope.
You need to provide a valid file path for the TBT3/4 Sigtest executable in the Text box File
Path and the name of the executable in the Text box File Name.
Note: TekRxTest application supports the usage of SigTest Analysis for TP3
Calibration.
Thunderbolt Receiver Test Application Manual20
Remote access
Application panels
Figure 8: Remote access
Table 8: Remote access
ParameterDescription
Local IP AddressDisplays the IP address for connecting to the application over socket server.
Listening Port
Time Out
Displays the TCP/IP port number of the port that the socket server is listening through.
Default Value: 4004
Displays the timeout value used when communicating with the socket server.
Default Value: 20 Seconds
Help panel
The help panel launches the TBT3/4 Gen 2 / Gen 3 TekRxTest application help document.
Calibrations panel
Complete TP3', TP3, and SFV calibrations before you start the DUT testing using the TBT3/4 Gen 2 / Gen 3 TekRxTest application. Follow
the instructions in the calibration wizards to automate the calibration for the TP3' and TP3 calibration points. After calibrating these points,
you can save the results.
Thunderbolt Receiver Test Application Manual21
Application panels
TP3' Calibration
The TP3' calibration panel allows you to perform TP3' calibration for Data Rate and save the results.
You must perform TP3' calibration before you start performing the calibration for TP3.
TP3' Calibration Procedure
Click on TP3' under the Calibrations tab to view the previously run calibration reports. At this stage, you can choose the data rate (10 Gbps
or 10.3125 Gbps or 20 Gbps or 20.625 Gbps), the run calibration reports for the selected data rate will be populated in the results table.
When you click on
procedure to perform the calibration.
During this process, the TekRxTest application calibrates the following items:
1. AC-DC Balance: The subsequent calibration modules assume the high and low frequency portions of the waveform to be of equal
amplitude which is accomplished through AC – DC Balance.
2. Eye Height Calibration: A nominal differential amplitude of 700 mV is calibrated to be used in other calibration modules.
3. Preset Calibration: The various predefined Preshoot and De-emphasis levels are required to be calibrated within a tolerance.
4. DDJ Calibration: The preset with the least data dependent jitter is found and used in further modules.
5. ACCM Calibration: The AC Common Mode Interference is required to be calibrated to 100 mVp-p at a frequency of 400 MHz.
6. RJ Calibration: The Random Jitter is to be calibrated to a nominal value of 0.14 UI p-p.
7. PJ Calibration: The periodic jitter at frequencies of 1, 2, 10, 50 and 100 MHz is calibrated to a nominal value of 0.17 UI p-p.
8. TJ Calibration: The Total Jitter at the predefined frequencies of 1, 2, 10, 50 and 100 MHz is calibrated be in the range of 0.3375 –
0.3625 UI p-p (Gen 2 - 10 and 10.3125 Gbps) and 0.355 – 0.405 UI p-p (Gen 3 - 20 and 20.625 Gbps).
9. Input Eye Diagram: In this step of calibration, it is ensured that the signal, with all the calibrated stresses, meets the Eye Height (700
mV for both Gen2 and Gen3) and Eye Width (0.175-0.825 UI for Gen2 and 0.19-0.81 UI for Gen3) requirements for 1, 2, 10, 50 and
100 MHz with the help of a predefined mask, ensuring that there are no mask hits.
, the TP3' wizard will be launched for the selected generation. This wizard will guide you through the sequential
Thunderbolt Receiver Test Application Manual22
Application panels
Figure 9: TP3' Calibration Panel
1.Connection Diagram: This page displays the connection diagram for the TP3' setup.
Figure 10: TP3' Calibration-Connection Diagram
Thunderbolt Receiver Test Application Manual23
Application panels
Click to move to the next step in the calibration process.
2.Initialization: This page displays the description and allows you to initialize the equipment. Click Initialize Equipment to complete
the initialization process.
You can click Automatic Calibration to perform the automatic calibration with the default settings for amplitude, Tx equalization
presets, ACCM, RJ, and PJ parameters without user intervention.
Figure 11: TP3' Calibration-Initialization
Figure 12: SSC Settings
SSC Settings: You can configure the SSC settings at the beginning of TP3’ calibration.
The profile of the SSC is down spread
Note:
Thunderbolt Receiver Test Application Manual24
Table 9: SSC Settings
ParameterDescription
FrequencyEnter the frequency in MHz for SSC to be used for all calibration and test modules.
DeviationEnter the deviation in ppm for SSC to be used.
Click to move to the next step in calibration process.
3.AC-DC Balance: This page displays the graph plots of AC-DC balance.
Application panels
Figure 13: TP3' Calibration-AC-DC Balance
Table 10: TP3' Calibration: AC-DC Balance
ParameterDescription
StartClick Start to run the measurement.
CancelClick Cancel to stop the calibration.
Click to move to the next step in calibration process.
4.EH Calibration: This page displays the graph plots of Eye Height (EH) calibration.
Thunderbolt Receiver Test Application Manual25
Application panels
Figure 14: TP3' Calibration-EH Calibration
Table 11: TP3' Calibration: EH Calibration
ParameterDescription
BERT Amplitude SettingDisplays the calibrated amplitude corresponding to the nominal Eye Height.
StartClick Start to run the measurement.
CancelClick Cancel to stop the calibration.
Click to move to the next step in calibration process.
5.Preset Calibration: This page displays the graph plots of preset calibration.