Tektronix THS730A, THS710A, THS720P, THS720A User Manual

Service Manual

THS710A, THS720A, THS730A & THS720P TekScope
070-9752-01
This document applies to to serial number B010100 and above and firmware version 1.00 and above.
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
Copyright E T ektronix, Inc. All rights reserved. T ektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved. Printed in the U.S.A. T ektronix, Inc., P.O. Box 1000, Wilsonville, OR 97070–1000 TEKTRONIX and TEK are registered trademarks of T ektronix, Inc. T ek Secure is a registered trademark of Tektronix, Inc. T ekTools, TekScope, and IsolatedChannel are trademarks of Tektronix, Inc. Pursuant to DFARS 252.227-7013(e), T ektronix Inc. hereby grants to the Government a nonexclusive, paid-up license
throughout the world of the scope set forth therein for Government purposes for any commercial manuals provided by T ektronix Inc. under this contract.
WARRANTY
T ektronix warrants that this product will be free from defects in materials and workmanship for a period of three (3) years from the date of purchase from an authorized T ektronix distributor. If any such product proves defective during this warranty period, T ektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Batteries are excluded from this warranty.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by T ektronix, with shipping charges prepaid. T ektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the T ektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. T ektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than T ektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; or c) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUST OMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT , SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Service Assurance
If you have not already purchased Service Assurance for this product, you may do so at any time during the product’s warranty period. Service Assurance provides Repair Protection and Calibration Services to meet your needs.
Repair Protection extends priority repair services beyond the product’s warranty period; you may purchase up to three years of Repair Protection.
Calibration Services provide annual calibration of your product, standards compliance and required audit documentation, recall assurance, and reminder notification of scheduled calibration. Coverage begins upon registration; you may purchase up to five years of Calibration Services.
Service Assurance Advantages
H Priced well below the cost of a single repair or calibration H Avoid delays for service by eliminating the need for separate purchase authorizations from your company H Eliminates unexpected service expenses
For Information and Ordering
For more information or to order Service Assurance, contact your local T ektronix representative and provide the informa­tion below. Service Assurance may not be available in locations outside the United States of America.
Name VISA or Master Card number and expiration Company date or purchase order number Address Repair Protection (1,2, or 3 years) City , State, Postal code Calibration Services (1,2,3,4, or 5 years) Country Instrument model and serial number Phone Instrument purchase date
T o learn the location of your nearest Tektronix representative, please call 1-800-TEK-WIDE in North America, 1-503-627-7111 elsewhere, or look us up on the World Wide Web at http://www.tek.com.

Table of Contents

Specifications

Operating Information

Theory of Operation

General Safety Summary v. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service Safety Summary vii. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preface ix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Related Manuals ix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Conventions x. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Specifications 1–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Internal and External Power Sources 2–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Understanding the Front Panel 2–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Connecting and Using the Probes 2–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Main Board 3–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Inverter Board 3–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Display Module 3–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Switch Assembly 3–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Performance Verification

T est Record 4–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Verification Procedures 4–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Adjustment Procedures

Required Equipment 5–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overview of the Adjustment Process 5–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Accessing the Adjustment-Lockout Jumper 5–3. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Oscilloscope Adjustment 5–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Meter Adjustment 5–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing the Adjustment-Lockout Jumper 5–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Maintenance

Preparation 6–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preventing ESD 6–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Inspection and Cleaning 6–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Removal and Installation Procedures 6–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Troubleshooting 6–28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Unpacking and Repacking Instructions 6–40. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
THS710A, THS720A, THS730A & THS720P Service Manual
i
Table of Contents

Options

Electrical Parts List

Diagrams

Mechanical Parts List

Options 7–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Electrical Parts List 8–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Diagrams 9–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Parts Ordering Information 10–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Replaceable Parts List 10–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ii
THS710A, THS720A, THS730A & THS720P Service Manual

List of Figures

Table of Contents
Figure 2–1: Removing and replacing the battery 2–2. . . . . . . . . . . . . . . . .
Figure 2–2: AC adapter attachment 2–2. . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–3: Typical scope-mode display 2–3. . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–4: Typical meter-mode display 2–4. . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–5: Connector locations 2–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–1: Module-level block diagram 3–2. . . . . . . . . . . . . . . . . . . . . . .
Figure 5–1: Location of adjustment-lockout jumper 5–4. . . . . . . . . . . . . .
Figure 5–2: Oscilloscope adjustment setups 5–6. . . . . . . . . . . . . . . . . . . . .
Figure 5–3: Meter adjustment setups 5–9. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 5–4: Relocating the adjustment-lockout jumper 5–11. . . . . . . . . . .
Figure 6–1: Removing the tilt stand 6–8. . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–2: Installing the tilt stand 6–9. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–3: Installing a new front-panel label 6–10. . . . . . . . . . . . . . . . . . .
Figure 6–4: Removing the battery door 6–12. . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–5: Assembling a new front cover 6–13. . . . . . . . . . . . . . . . . . . . . .
Figure 6–6: Installing the gasket 6–14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–7: Installing the handle 6–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–8: Removing the display module 6–17. . . . . . . . . . . . . . . . . . . . . .
Figure 6–9: Opening the display cable connector 6–18. . . . . . . . . . . . . . . .
Figure 6–10: Routing cables to the inverter board 6–19. . . . . . . . . . . . . . .
Figure 6–11: Lifting the chassis 6–20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–12: Foam pad and hole plug locations on the back cover 6–21. .
Figure 6–13: Routing the battery wires 6–22. . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–14: Removing the main board and switch flex-circuit
assembly 6–24. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–15: Foam pad locations on the chassis 6–25. . . . . . . . . . . . . . . . . .
Figure 6–16: Reassembling the chassis 6–27. . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–17: Battery connector location 6–29. . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–18: Bypassing the ON/STBY switch 6–31. . . . . . . . . . . . . . . . . . .
Figure 10–1: Exploded diagram 10–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
THS710A, THS720A, THS730A & THS720P Service Manual
iii
Table of Contents

List of Tables

Table 1–1: Oscilloscope specifications 1–1. . . . . . . . . . . . . . . . . . . . . . . . .
Table 1–2: DMM specifications 1–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 1–3: General specifications 1–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 5–1: Summary of oscilloscope adjustment steps 5–5. . . . . . . . . . . .
Table 5–2: Summary of meter adjustment steps 5–8. . . . . . . . . . . . . . . . .
Table 6–1: Internal inspection check list 6–3. . . . . . . . . . . . . . . . . . . . . . .
Table 6–2: Removal and installation procedures 6–6. . . . . . . . . . . . . . . . .
Table 7–1: TekScope instrument options 7–1. . . . . . . . . . . . . . . . . . . . . . .
Table 10–1: Replaceable standard accessories 10–6. . . . . . . . . . . . . . . . . . .
iv
THS710A, THS720A, THS730A & THS720P Service Manual

General Safety Summary

Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
To Avoid Fire or Personal Injury
Connect and Disconnect Properly . Do not connect or disconnect probes or test
leads while they are connected to a voltage source. Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Replace Batteries Properly. Replace batteries only with the proper type and rating specified.
Recharge Batteries Properly. Recharge batteries for the recommended charge cycle only.
Use Proper AC Adapter. Use only the AC adapter specified for this product. Do Not Operate Without Covers. Do not operate this product with covers or panels
removed. Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present. Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions. Do Not Operate in an Explosive Atmosphere.
THS710A, THS720A, THS730A & THS720P Service Manual
v
General Safety Summary
Symbols and Terms
T erms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
T erms on the Product. These terms may appear on the product: DANGER indicates an injury hazard immediately accessible as you read the
marking. WARNING indicates an injury hazard not immediately accessible as you read the
marking. CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
Battery Recycling
CAUTION
Refer to Manual
WARNING
High Voltage
Double
Insulated
Protective Ground
(Earth) Terminal
This product contains a Nickel Cadmium (NiCd) battery, which must be recycled or disposed of properly. For the location of a local battery recycler in the U.S. or Canada, please contact:
RBRC (800) BATTERY Rechargeable Battery Recycling Corp. (800) 227-7379 P.O. Box 141870 www.rbrc.com Gainesville, Florida 32614
vi
THS710A, THS720A, THS730A & THS720P Service Manual

Service Safety Summary

Only qualified personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary before performing any service
procedures. Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is present.
Use Care When Servicing With Power On. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
To avoid electric shock, do not touch exposed connections.
THS710A, THS720A, THS730A & THS720P Service Manual
vii
Service Safety Summary
viii
THS710A, THS720A, THS730A & THS720P Service Manual

Preface

Related Manuals

This is the Service Manual for the THS710A, THS720A, THS730A, and THS720P TekScope instruments. This manual provides information to trouble­shoot and repair the instrument to the module level.
Additional documentation for the instrument is contained in the related manuals listed below.
User Manual
Language
English 070-9731-XX 070-9741-XX 070-9751-XX French 070-9732-XX 070-9742-XX German 070-9733-XX 070-9743-XX Italian 070-9734-XX 070-9744-XX Korean 070-9735-XX 070-9745-XX Portuguese 070-9736-XX 070-9746-XX Spanish 070-9737-XX 070-9747-XX Simplified Chinese 070-9738-XX 070-9748-XX Traditional
Chinese Japanese 070-9740-XX 070-9799-00
Part Number
070-9739-XX 070-9749-XX
Reference Part Number
Programmer Manual Part Number
THS710A, THS720A, THS730A & THS720P Service Manual
ix
Preface

Conventions

In the Performance Verification, Adjustments, and Maintenance chapters of this manual, TekScope instrument setups are shown in tables similar to the one shown below.
The header of each table contains icons that represent the controls and menu items used to set up the instrument. To make a specific setup, read the table from left to right and then from top to bottom as shown below. The table contains the symbol “—” if no action is required.
1. Choose scope mode or meter mode.
2. Press this but­ton on the front panel.
3. Press this bezel button.
6. 7.
8. 9. 10.
4. Press the bezel button again until this selection is high­lighted.
5. Use the
+
the value for a parameter.
/– rocker to set
x
THS710A, THS720A, THS730A & THS720P Service Manual

Specifications

S
Sequence
e
d
NC
This chapter contains the oscilloscope, DMM, and general specifications for the THS710A, THS720A, THS730A, and THS720P TekScope instruments. All specifications are guaranteed unless noted “typical.” Typical specifications are provided for your convenience but are not guaranteed. Specifications that are marked with the n symbol are checked in the chapter Performance Verification.
All specifications apply to all four instruments unless noted otherwise. All specifications assume horizontal MAG is off, unless noted otherwise. To meet specifications, two conditions must first be met:
H The TekScope instrument must have been operating continuously for ten
minutes within the operating temperature range specified. H You must perform the Compensate Signal Path operation, accessible through
the utility menu, if the ambient temperature changes by more than 5° C.
T able 1–1: Oscilloscope specifications
Acquisition
Acquisition Modes Sample (Normal), Peak detect, Envelope, and Average Acquisition Rate,
typical
ingle
Inputs
Input Coupling DC, AC, or GND Input Impedance,
DC Coupled Maximum Voltag
Between Signal an Common at Input B
Up to 25 waveforms per second (2 channels, sample acquisition mode, MAG on, no measurements)
Acquisition Mode Acquisition Stops After
Sample, Peak Detect Single acquisition, one or two
channels simultaneously
Average, Envelope N acquisitions, one or two chan-
nels simultaneously, N is settable from 2 to 256 or
1 M ±1% in parallel with 25 pF ±2 pF
Overvoltage Category Maximum Voltage
CA T II Environment (refer to page 1–13)
CA T III Environment (refer to page 1–13)
For steady-state sinusoidal waveforms, derate at 20 dB/decade above 100 kHz to 13 V Category description on page 1–13.
at 3 MHz and above. Also, refer to Overvoltage
pk
300 V
150 V
RMS
RMS
THS710A, THS720A, THS730A & THS720P Service Manual
1–1
Specifications
ed
e-
pe d d
Rise Time
T able 1–1: Oscilloscope specifications (Cont.)
Inputs
Maximum Voltage Between Common and Earth Ground at BNC
Maximum Voltage, Channel-to-Channel Commons
Channel-to-Channel Common Mode Rejection, typical
Channel-to-Channel Crosstalk, typical
Common to Chassis Capacitance, typical
Vertical
Number of Channels 2 Digitizers 8 bit resolution, separate digitizers for each channel sample
VOL TS/DIV Range 5 mV/div to 50 V/div at input BNC Polarity Normal and Invert Position Range ±10 divisions n Analog Bandwidth
at BNC, DC Coupl (typical at 5 mV/div, guaranteed at all other settings)
Peak Detect or Env lope Bandwidth, typical (25 MS/s or slower)
600 V
(CAT II) or 300 V
RMS
(CAT III), using rated connectors or
RMS
accessories 30 V
30 V
, 42.4 Vpk, using noninsulated connectors and acessories
RMS
, 42.4 Vpk, using noninsulated connectors and acessories
RMS
100:1 at frequencies 50 MHz, measured on MATH Ch1 – Ch2 waveform, with test signal applied between signal and common of both channels, and with the same VOLTS/DIV and coupling settings on each channel
100:1 at 50 MHz, measured on one channel, with test signal applied between signal and common of the other channel, and with the same VOL TS/DIV and coupling settings on each channel
55 pF
simultaneously
THS710A THS720A THS720P THS730A
60 MHz at input BNC
100 MHz at input BNC
(90 MHz above 35° C)
100 MHz at input BNC
(90 MHz above 35° C)
THS710A THS720A THS720P THS730A
50 MHz
75 MHz 75 MHz 85 MHz
200 MHz at input BNC
(180 MHz above 35° C)
1–2
Analog Bandwidth
Selectable between 20 MHz or full
Limit, typical Lower Frequency
Limit, AC Coupled,
10 Hz at BNC, reduced by a factor of ten when using a 10X passive probe
typical
at BNC,
typical Peak Detect or Enve-
lope Pulse Response,
THS710A THS720A THS720P THS730A
5.8 ns
3.5 ns 3.5 ns 1.75 ns
Captures 50% or greater amplitude of pulses 8 ns wide ( 20 ns wide at 500 ns/div)
typical
THS710A, THS720A, THS730A & THS720P Service Manual
T able 1–1: Oscilloscope specifications (Cont.)
reme
A
e
forms
Sample Rate R
e
SEC/DIV R
e
Type
Type
Vertical
DC Gain Accuracy ±2% for Sample or Average acquisition mode Position Accuracy ±[0.4% × |(position × volts/div)| + (0.1 div × volts/div)] n DC Measu
ccuracy, Averag Acquisition Mode Using 16 Wave-
DC Measurement Accuracy , Sample Acq. Mode, typical
Horizontal
Record Length 2500 samples for each channel
(including MAG)
n Sample Rate and Delay Time Accuracy
Delay Time Range Zero to 50 s
Internal Trigger
n Trigger Sensitivity, Edge Trigger (THS710A, THS720A, and THS720P)
n Trigger Sensitivity, Edge Trigger (THS730A)
ang
ang
nt
Measurement Type Accuracy Absolute voltage measurements ±[2% × |reading +
(position × volts/div)| + (0.1 div × volts/div)]
Delta voltage between any two waveforms acquired under same setup
±[2% × |reading + (position × volts/div)| + (0.15 div × volts/div) +
0.6 mV]
THS710A THS720A THS720P THS730A
5 S/s to 250 MS/s, in a
1.25, 2.5, 5 sequence
THS710A THS720A THS720P THS730A
10 ns/div to 50 s/div
±200 ppm over any 1 ms time interval
Coupling Sensitivity
DC 0.35 div from DC to 50 MHz,
Coupling Sensitivity
DC 0.35 div from DC to 50 MHz,
5 S/s to 500 MS/s, in a
1.25, 2.5, 5 sequence
5 ns/div to 50 s/div
±[2% × |reading| + (0.05 div × volts/div)]
5 S/s to 500 MS/s, in a
1.25, 2.5, 5 sequence
5 ns/div to 50 s/div
increasing to 1 div at 100 MHz (90 MHz above 35° C)
increasing to 1.5 div at 200 MHz (180 MHz above 35° C)
Specifications
5 S/s to 1 GS/s, in a 1.25, 2.5, 5 sequence
2 ns/div to 50 s/div
THS710A, THS720A, THS730A & THS720P Service Manual
1–3
Specifications
Tr
Type
S
d
Tr
Type
T able 1–1: Oscilloscope specifications (Cont.)
Internal Trigger
igger Sensitivity, Edge Trigger typical
Trigger Level Range ±4 divisions from center of screen Motor Trigger Level
Range (THS720P) Trigger Level
Accuracy , typical SET LEVEL TO 50%,
typical Width Range, Pulse
Trigger T ype, typical Width Tolerance
Range, Pulse Trigger Type, typical
Sensitivity, Video Trigger T ype, typical
ignal Formats an
Field Rates, Video
igger
Holdoff Range 495 ns to 10 s
External Trigger
External Trigger , Max­imum Input Voltage
External Trigger Coupling
External Trigger Levels
External Trigger Sensitivity
Coupling Sensitivity
,
NOISE REJ 3.5 times the DC-coupled limits HF REJ
LF REJ 1.5 times the DC-coupled limits
0.1 to 5 divisions from center screen, polarity depends on slope selection
±0.2 divisions, for signals having rise and fall times 20 ns
Operates with input signals 50 Hz
99 ns to 1 s, with resolution of 33 ns or approximately 1% of setting (whichever is greater)
5%, 10%, 15%, or 20%
Composite video signal with negative sync pulse amplitude from 0.6 to 2.5 divisions
Broadcast systems Supports NTSC, PAL, and
Interlaced Trigger on selected line of odd
Non-interlaced T rigger on selected line or any line Line Rates 15 kHz to 65 kHz, in five ranges
600 V
CAT II or 300 V
RMS
CAT III (refer to page 1–13)
RMS
DC only
+0.2 V or +2 V, selectable
500 mV
from DC to 1 MHz, increasing to 1 V
p-p
signal centered at selected trigger level
1.5 times the DC-coupled limit from DC to 30 kHz, attenuates signals above 30 kHz
for frequencies above 1 kHz, attenuates signals below 1 kHz
SECAM
field, selected line of even field, or any line
at 5 MHz, with
p-p
1–4
THS710A, THS720A, THS730A & THS720P Service Manual
Specifications
Amp
de
se
A
y
T able 1–1: Oscilloscope specifications (Cont.)
Measurements
Cursors Voltage difference between cursors (DV)
Time difference between cursors (DT) Reciprocal of DT in Hertz (1/DT) Phase difference between cursors (DDegrees)
Automated Measurements
Voltage and Current Harmonics (THS720P)
Number of Harmonics First 31 harmonics of signal with fundamental frequency between
Harmonics
litu
Accuracy
Harmonics Pha
ccurac
THD-F Measurement Total harmonic distortion relative to the fundamental amplitude
THD-F Accuracy ±4% THD-R Measurement Total harmonic distortion relative to the RMS amplitude
THD-R Accuracy ±4% Frequency Accuracy ±0.2% of reading
Power Measurements (THS720P)
True Power Measurement
Apparent Power Measurement
Reactive Power Measurement
Power Measurements
Accuracy
Amplitude, Burst Width, Cycle Mean, Cycle RMS, Fall Time, Frequency, High, Low, Max, Mean, Min, Negative Duty Cycle, Negative Overshoot, Negative Width, Pk – Pk, Period, Positive Duty Cycle, Positive Overshoot, Positive Width, Rise Time, and RMS
30 Hz and 450 Hz Accuracies below are stated as a percent of the fundamental amplitude
and are valid only if peak-to-peak amplitude is 4 divisions and number of averages 16
Fundamental 2 – 11 12 – 21 22 – 31 ±2.5% ±2.5% ±4% ±4% Fundamental 2 – 11 12 – 21 22 – 31 ±4 ° ±8 ° ±8 °
THD-F
THD-R
1
W
n
V
2
RMS
2
V
RMS
n
V
V
f
V
V
RMS
Vn A
2
f
2
f
n
2
A
RMS
or
2
A
RMS
or
A
A
RMS
A
f
A
2
f
2
f
measured over an integral number of cycles that contain n sample points
VA V
VAR  (VA)2 W
RMS
A
RMS
2
±4% at the BNCs (not including probe uncertainty)
THS710A, THS720A, THS730A & THS720P Service Manual
1–5
Specifications
A
DC Coupled
e
p
T able 1–1: Oscilloscope specifications (Cont.)
Power Measurements (THS720P)
PF Measurement
Power factor (PF)
q Measurement q is the phase difference between the fundamental components of
voltage and current. Positive angle means voltage leads current.
Negative angle means voltage lags current. DPF Measurement Displacement power factor (DPF) = cos Power Factor Mea-
±0.05 surements Accuracy
With P6117 Probe
nalog Bandwidth,
THS710A THS720A THS720P THS730A
60 MHz 100 MHz
true p owe r
apparent power
100 MHz
W
VA
200 MHz
Probe Attenuation 10X Maximum Voltag
Between Probe Ti and Reference Lead
Overvoltage Category Maximum Voltage
CA T II Environment (refer to
page 1–13)
CA T III Environment (refer to
page 1–13)
For steady-state sinusoidal waveforms, derate at 20 dB/decade above
900 kHz to 13 V
Category description on page 1–13. Maximum Voltage
30 V Between Reference Lead and Earth Ground Using P6117 Probe
Maximum Voltage,
30 V Channel-to-Channel Reference Leads Using P6117 Probe
RMS
RMS
, 42.4 V
, 42.4 V
(90 MHz above 35° C)
at 27 MHz and above. Also, refer to Overvoltage
pk
pk
pk
(90 MHz above 35° C)
300 V
RMS
150 V
RMS
(180 MHz above 35° C)
1–6
THS710A, THS720A, THS730A & THS720P Service Manual
T able 1–1: Oscilloscope specifications (Cont.)
A
DC Coupled
e
p
DC Coupled
e
p
Reference Lead
e
Reference
With P5102 Probe
nalog Bandwidth,
THS710A THS720A THS720P THS730A
60 MHz 100 MHz
100 MHz
Specifications
100 MHz
Probe Attenuation 10X Maximum Voltag
Between Probe Ti and Reference Lead,
Overvoltage Category Maximum Voltage
CA T II Environment (refer to page 1–13)
CA T III Environment (refer to page 1–13)
Maximum Voltag Between Probe Ti and AC Coupled
Maximum Voltag Between Lead and Earth Ground
Overvoltage Category Maximum Voltage
CA T II Environment ±1000 V
,
CA T III Environment
Overvoltage Category Maximum Voltage
CA T II Environment 600 V CA T III Environment 300 V
(90 MHz above 35° C)
(90 MHz above 35° C)
1000 V
RMS
600 V
RMS
DC
±600 V
DC
RMS RMS
THS710A, THS720A, THS730A & THS720P Service Manual
1–7
Specifications
e
d
e
r
Gr
d
R
Res
T able 1–2: DMM specifications
General
Resolution 33⁄4 digit, 4000 count full scale reading except as noted Input Resistance, AC
or DC Voltage Input Capacitance,
AC or DC Voltage, typical
Maximum Voltag Between DMM an COM Inputs
Maximum Voltag Between DMM o COM Input and Earth
oun
DC Voltage
anges and
olution
n Accuracy ±(0.5% of reading + 5 counts) Normal Mode
Rejection, typical Common Mode
Rejection, typical
10 M ±10%
100 pF
Overvoltage Category Maximum Voltage
CA T I Environment (refer to
640 V
RMS
(880 VDC)
page 1–13)
CA T II Environment (refer to
600 V
RMS
page 1–13)
CA T III Environment (refer to
300 V
RMS
page 1–13)
Overvoltage Category Maximum Voltage
CA T I Environment (refer to
640 V
RMS
(880 VDC)
page 1–13)
CA T II Environment (refer to
600 V
RMS
page 1–13)
CA T III Environment (refer to
300 V
RMS
page 1–13)
Range Resolution
400.0 mV
0.1 mV
4.000 V 1 mV
40.00 V 10 mV
400.0 V 100 mV
880 V 1 V
Rejects AC signals by >60 dB at 50 Hz or 60 Hz (user selectable)
Rejects AC signals by >100 dB at 50 Hz or 60 Hz (user selectable)
1–8
THS710A, THS720A, THS730A & THS720P Service Manual
Specifications
R
Res
y
500
R
Res
y
S
Resis
e
reme
T able 1–2: DMM specifications (Cont.)
AC Voltage
Conversion Type AC conversions are true RMS. The AC measurement is based on the
AC and DC components of the signal as shown below:
AC Measurement = RMS(AC+DC) – DC
anges and
olution
n Accurac (40 Hz to
Common Mode Rejection, typical
/Resistance
anges and
olution
n Accurac
Bias Voltage for Full
cale
Measu
Hz)
tanc nt, typical
Range Resolution
400.0 mV
4.000 V 1 mV
40.00 V 10 mV
400.0 V 100 mV 640 V 1 V
Input Waveform Maximum Error
Sinusoidal waveforms Nonsinusoidal waveforms with
crest factor up to 3 Rejects AC signals by >60 dB at DC, 50 Hz, and 60 Hz
Range Resolution
400.0
4.000 k 1
40.00 k 10
400.0 k 100
4.000 M 1 k
40.00 M 10 k
Range Maximum Error
All ranges except 40 M ±(0.5% of reading + 2 counts) 40 M ±(2% of reading + 5 counts) for
Range Full Scale Bias Voltage
400.0 350 mV
4.000 k
40.00 k 350 mV
400.0 k 350 mV
4.000 M 400 mV
40.00 M 1.10 V
0.1 mV
±(2% of reading + 5 counts) ±(4% of reading + 5 counts)
0.1
60% relative humidity
200 mV
THS710A, THS720A, THS730A & THS720P Service Manual
1–9
Specifications
Open Cir
T able 1–2: DMM specifications (Cont.)
/Resistance
cuit Voltage,
typical
Continuity Check
Indication An audible tone is generated when measured resistance is below 50 ,
Open Circuit Voltage, typical
Test Current, typical 1 mA
Diode Check
Range Zero to 2 V, measures forward voltage drop of semiconductor junction Voltage Accuracy,
typical Open Circuit Voltage,
typical Test Current, typical 1 mA
Data Logger
Horizontal Scale Range
Vertical Zoom Range 2X, 5X, or 10X
Range Open Circuit Voltage
400.0
All other ranges 1.2 V
typical
4 V
±25%
4 V
30 s/div to 24 hours/div (4 minutes to 8 days, full scale)
4.8 V
1–10
T able 1–3: General specifications
Display
Display Type 4.7 in (120 mm) diagonal liquid crystal Display Resolution 320 horizontal by 240 vertical pixels Display Contrast Adjustable, temperature compensated Backlight Intensity,
typical
35 cd/m
2
THS710A, THS720A, THS730A & THS720P Service Manual
Specifications
ery Char
T able 1–3: General specifications (Cont.)
RS-232 Interface
Device Type DTE. at RJ-45 connector Pinout Signal Pin Number at
9-pin Null Modem Adapter
RTS out 1 8 TXD out 2 6 RXD in 3 5 GND 5 4 DTR out 6 3 CTS in 7 7 RTS out 8 8 DSR in (not used) 4 2 DCD in (not used) 7 1
Probe Compensator Output
Output Voltage, typical
Frequency, typical 1.2 kHz
Power Source
Battery Replaceable Ni-Cd battery pack Battery Life, typical Approximately two hours of continuous use from a full charge Low Battery Indica-
tion, typical Battery Saver Standby Time-out and Backlight Time-out extend battery life. Time-out
Batt Time, typical
External Power 12 VDC nominal, center positive; Operates with input from 10 VDC to
ging
5.0 V into 1 M load
Low battery message first appears approximately ten minutes before the instrument powers off automatically
ranges from 1 minute to 15 minutes, or off. With T ekScope instrument operat-
ing With T ekScope instrument turned
off In external charger 1.5 hours
15 VDC
9 hours
9 hours
Pin Number at RJ-45 Connector
The DC INPUT disconnects itself automatically if >15 VDC is applied. If this occurs, disconnect the overvoltage and then reconnect to a voltage in the proper range.
Memory Retention, typical
Fuse This instrument has no user-replaceable fuses
All memory is retained indefinitely with battery removed and without external power applied.
THS710A, THS720A, THS730A & THS720P Service Manual
1–11
Specifications
Temper
re
y
A
de
Random Size
We
T able 1–3: General specifications (Cont.)
Environmental
atu
Humidit
ltitu
Vibration Operating 2.66 g
Drop Resistance, typical
Moisture Resistance Meets IEC529, IP43 with DC input hole plug, I/O port hole plug, and
Mechanical
ight With battery installed 3.2 lbs (1.5 kg)
Adjustment Interval
The recommended adjustment interval is one year
Operating –10° C to +50° C
Nonoperating –20° C to +60° C
+40° C or below 95% relative humidity
+41° C to +50° C 75% relative humudity
Operating 2,000 m
Nonoperating 15,000 m
from 5 Hz to 500 Hz,
RMS
10 minutes on each axis
Nonoperating 3.48 g
10 minutes on each axis
Survives a 30 in (76 cm) drop onto concrete with only cosmetic damage
battery door installed.
Height 8.53 in (217 mm)
Width 6.95 in (177 mm)
Depth 2.00 in (50.8 mm)
With all standard accessories in
soft carry case
When packaged for domestic
shipment
7.5 lbs (3.4 kg)
9.0 lbs (4.1 kg)
from 5 Hz to 500 Hz,
RMS
1–12
THS710A, THS720A, THS730A & THS720P Service Manual
Specifications
Over
T able 1–3: General specifications (Cont.)
Certifications and Compliances
Certifications Listed UL31 11-1 and CAN/CSA-C22.2 No. 1010.1-92, complies with
EN61010-1/A2
voltage CategoryCategory Examples
CA T III A typical CAT III environment is
the power distribution system within a building or factory. These environments are somewhat protected from lightning strikes, but susceptible to switching transients and other disturbances that may generate high voltage impulses.
CA T II A typical CAT II environment is
the 120/240 V distribution system within a lab or office. These environments are fairly well protected from external high voltage disturbances.
CA T I A typical CAT II environment is
circuitry inside electrical or elec­tronic equipment that is powered by a power supply or a battery.
EC Declaration of Conformity
Meets the intent of Directive 89/336/EEC for Electromagnetic Compatibility and Low-Voltage Directive 73/23/ECC for Product Safety as ammended by 93/68/EEC. Compliance was demonstrated to the following specifications as listed in the official Journal of the European Communities:
EN 55011 Class A: Radiated and Conducted Emissions
13
EN 50081–1 Emissions:
EN 50082–1 Immunity:
EN 61010–1 Safety
1
Tektronix-supplied ferrite bead required on instrument end of RS-232 cable
2
Performance criteria: 5.0 div increase in peak to peak noise (Sample acquisition mode, full bandwidth); otherwise, 1.0 div increase in peak to peak noise
3
Applies to instrument operating with Tektronix-supplied AC adapter
THS710A, THS720A, THS730A & THS720P Service Manual
EN 60555–2 Power Harmonics
IEC 801–2 Electrostatic Discharge IEC 801–3 RF Radiated
2
IEC 801–4 Fast Transients IEC 801–5 Surge
3
1–13
Specifications
1–14
THS710A, THS720A, THS730A & THS720P Service Manual

Operating Information

This chapter provides a quick overview of the following topics:
H Internal and external power sources H Understanding the front panel H Connecting and using the probes
For more detailed information about TekScope instrument operation, refer to the THS710A, THS720A, THS730A, & THS720P User Manual.
For information about unpacking and inspecting the instrument, refer to Unpacking and Repacking Instructions on page 6–40 of this manual.

Internal and External Power Sources

You can power the TekScope instrument from the internal battery pack or from the AC adapter (both are standard accessories).
Using the Battery Pack
Before using the battery for the first time, it must be charged. You can use external power to charge the battery pack while it is in the TekScope instrument. Or you can charge the battery pack with the optional external battery charger. Typical battery charging times are listed below.
Configuration Typical Charging Time
Battery pack in TekScope instrument 9 hours Battery pack in external charger 1.5 hours
You can remove and replace the battery pack without losing any saved information. The current setup, saved setups, saved waveforms, and saved data are stored in nonvolatile memory that does not depend on battery power. To prevent loss of saved information, set the ON/STBY switch to STBY before removing the battery pack. See Figure 2–1 for instruction to remove and replace the battery.
WARNING. To avoid shock hazard, the battery door must remain closed in wet or damp conditions.
THS710A, THS720A, THS730A & THS720P Service Manual
2–1
Operator Information
Battery pack
Figure 2–1: Removing and replacing the battery
Using External Power
For benchtop operation, you can use external power. Attach the AC adapter as shown in Figure 2–2. The DC INPUT disconnects itself if an overvoltage is applied. If this occurs, disconnect and then reconnect the AC adapter to resume operation from external power.
WARNING. To avoid shock hazard, the DC input and I/O port hole plugs must remain closed in wet or damp conditions.
External power to
DC INPUT
2–2
Figure 2–2: AC adapter attachment
THS710A, THS720A, THS730A & THS720P Service Manual
CAUTION. To avoid overheating, do not connect external power while the instrument is in a confined space, such as in the soft case.

Understanding the Front Panel

The SCOPE and METER buttons select the operating mode. The front panel has additional buttons for the functions you use most often and menus to access the more specialized functions.
Operator Information
Using Scope Mode
Press the front-panel SCOPE button to enter scope mode. Then, press AUTO­RANGE to set the vertical, horizontal, and trigger automatically for a usable display. A typical scope-mode display is shown in Figure 2–3.
Status line
Graticule
area
Waveform readout lines
Measurement readout area
Figure 2–3: Typical scope-mode display
THS710A, THS720A, THS730A & THS720P Service Manual
2–3
Operator Information
Using Meter Mode
Press the front-panel METER button to enter meter mode. Press one of the bezel buttons to choose a meter function and then press AUTORANGE to set the range automatically. A typical meter-mode display is shown in Figure 2–4.
Status line
Measurement readout areas
Graticule area
Figure 2–4: Typical meter-mode display
2–4
THS710A, THS720A, THS730A & THS720P Service Manual
Operator Information
Using the Menu System
To use the menu system, follow the steps below:
1. Press a front-panel button to display the menu you want to use.
2. Press a bezel button to choose a menu item. If a pop-up menu appears,
continue to press the bezel button to choose an item in the pop-up menu. You may need to press the Select Page button to access additional menu items.
THS710A, THS720A, THS730A & THS720P Service Manual
2–5
Operator Information
3. After choosing some menu items, you need to set a numerical parameter to
+
complete the setup. Use the
/– rocker to adjust the parameter or press the
TOGGLE button to reset the parameter to its default value.
4. If the OK bezel button is displayed, press it to confirm your choice.
2–6
NOTE. In the Performance Verification, Adjustments, and Maintenance chapters of this manual, TekScope instrument setups are shown in tables. Refer to
Conventions on page x for information about reading these tables.
THS710A, THS720A, THS730A & THS720P Service Manual
Operator Information
Using the Dedicated
Buttons
You can use the dedicated buttons below to take direct actions. These buttons do not require the use of menus.
31
2
4
16 15
14
87910111213
5
6
1. HARD COPY. Initiates a hard copy using the RS-232 port.
2. HOLD. Stops/restarts oscilloscope acquisition or holds/resets meter readout.
3. AUTORANGE. Selects oscilloscope or meter autorange function.
4. CLEAR MENU. Clears menu from display.
5. TRIGGER LEVEL. Adjusts trigger level.
6. SET LEVEL TO 50%. Sets trigger level to midpoint of oscilloscope
waveform.
7. HORIZONTAL POSITION. Adjusts oscilloscope waveform horizontal
position.
8. MAG. Turns 10X horizontal magnification on and off.
9. SEC/DIV. Adjusts waveform horizontal scale factor for oscilloscope or data
logger.
10. VERTICAL POSITION. Adjusts oscilloscope waveform vertical position.
11. WAVEFORM OFF. Removes selected oscilloscope waveform from display.
12. VOLTS/DIV. Adjusts oscilloscope vertical scale factor or meter range.
13. ON/STBY. Selects on or standby. Does not disconnect power from the
instrument.
THS710A, THS720A, THS730A & THS720P Service Manual
2–7
Operator Information
14. CH 1, CH 2, MATH, REF A, REF B. Displays waveform and chooses
selected waveform. In harmonics mode (THS720P), CH 1 and CH 2 also display harmonics of voltage and current waveforms; MATH displays power measurements.
15. SCOPE. Selects scope mode.
16. METER. Selects meter mode.

Connecting and Using the Probes

The oscillscope probes and meter leads attach to the connectors located on the top and side panels of the TekScope instrument as shown in Figure 2–5.
Top panel
Figure 2–5: Connector locations
Side panel
2–8
THS710A, THS720A, THS730A & THS720P Service Manual
Operator Information
Compensating the
Oscilloscope Probes
To maintain signal fidelity, you must compensate each voltage probe for the channel input it is connected to.
1. Connect the oscilloscope probe and then press AUTORANGE.
Ch 1
Probe reference lead to
Probe tip to PROBE COMP
AUTORANGE
2. Check the shape of the displayed waveform.
Overcompensated
Undercompensated
Compensated correctly
3. If necessary, adjust the probe for correct compensation.
P6117 P5102
4. Repeat these steps for the other probe and channel.
THS710A, THS720A, THS730A & THS720P Service Manual
2–9
Operator Information
Attaching the Reference
Leads Correctly
Beware of High Voltages
If you are using both of the oscilloscope channels, you must attach the probe reference lead for each channel directly to your circuit. These attachments are required because the oscilloscope channels are electrically isolated; they do not share a common chassis connection. If you are also using the DMM, you must also attach the DMM common lead to your circuit for the same reason as above.
Understand the voltage ratings for the probes you are using and do not exceed those ratings. Two ratings are important to know and understand:
H The maximum measurement voltage from the probe tip to the probe
reference lead H The maximum floating voltage from the probe reference lead to earth ground These two voltage ratings depend on the probe and your application. Refer to
Specifications beginning on page 1–1 for more information.
WARNING. To prevent electrical shock, do not exceed the measurement or floating voltage ratings for the oscilloscope input BNC connector, probe tip, probe reference lead, DMM input connector, or DMM lead.
2–10
THS710A, THS720A, THS730A & THS720P Service Manual

Theory of Operation

This chapter describes the electrical operation of the TekScope instrument to the module level. It describes the basic operation of each functional circuit block shown in Figure 3–1.

Main Board

The Main board assembly contains the following oscilloscope and meter functions.
Oscilloscope Signal Path
Meter Signal Path
Processor System
Display System
Power Supply
Each of the two oscilloscope input signals enters the main board and passes through an attenuator and buffer amplifier. Then each signal passes through an isolation interface before reaching the trigger circuitry, a sampler, and a digitizer. The digitized signals are written into system memory for transfer to the display system.
The DMM input enters the Main board and passes through a switching network to select the meter function and range. The DMM signal is then sampled and digitized.
The processor system contains a 68331 microprocessor that controls the entire instrument. The processor passes waveforms, meter readings, and text on to the display system. The processor system also contains flash ROM, system RAM, and the RS-232, keyboard, and probe compensation interfaces.
The display system, consisting of a display controller and video memory, processes text and waveforms to refresh the display.
The power supply provides DC power to ciruits on the main board and generates an AC voltage to provide power to the attenuators and buffer amplifiers across the isolation interfaces. It also provides +12 V and –20 V for the display module.

Inverter Board

The inverter board generates an AC voltage for the display backlight.
THS710A, THS720A, THS730A & THS720P Service Manual
3–1
Theory of Operation

Display Module

Switch Assembly

The display module consists of a liquid crystal display (LCD), a display driver, and a flourescent backlight.
The switch assembly is a matrix of switches that form the front-panel buttons. The switch assembly consists of a flex-circuit board and an elastomeric switch mat. Switch closures are read by the processor system.
Battery
Adjustment-
lockout
Not used
12
2
80
EXT
PWR
2
J8
J3
J1
CH 1 CH 2
J9
Main board
J7 J6
PROBE
COMP
8
J5
RS-232
COM
DMM
3–2
5
Inverter boardDisplay assembly Switch assembly
2
Figure 3–1: Module-level block diagram
THS710A, THS720A, THS730A & THS720P Service Manual
17

Performance Verification

Wavetek 9100 U
ers
e
F
5500A
-prod
C
C
This chapter contains performance verification procedures for the specifications marked with the n symbol. The following equipment, or a suitable equivalent, is required to complete these procedures.
Description
DC Voltage Source 60 mV to 800 V, ±0.1% accu-
AC Voltage Source 300 mV to 640 V, ±0.5%
Resistance Standard 360 to 36 M, ±0.1%
Leveled Sine Wave Generator 50 kHz and 100 MHz, ±3%
Time Mark Generator 10 ms period, ±50 ppm accu-
Minimum Requirements
racy
accuracy at 500 Hz
accuracy
amplitude accuracy
racy
Examples
niv
al Calibration System with Oscil­loscope Calibration Modul (Option 250)
luke
Calibrator with Oscilloscope
alibration Option (Option
alibration Option (Option
5500A-SC)
Multi
uct
Banana to Banana Cable (two required)
50 BNC Cable BNC male to BNC male,
50 Feedthrough Termina­tion
Dual Banana to BNC Adapter Banana plugs to BNC female Tektronix part number
Shielded banana jacks on each end
36 in (1 m) long BNC male and female con-
nectors
Tektronix Deluxe Meter Lead Set (ATLDX1)
Tektronix part number 012-0482-00
Tektronix part number 01 1-0049-01
103-0090-00
THS710A, THS720A, THS730A & THS720P Service Manual
4–1
Performance Verification
C
re-
me
y
C
re-
me
y

Test Record

Serial Number
Test Passed Failed
Self Test
Oscilloscope Tests Low Limit Test Result High Limit
hannel 1 DC Measu
nt Accurac
Procedure Performed by Date
5 mV/div 34.05 mV 35.95 mV 500 mV/div 3.405 V 3.595 V 2 V/div 13.62 V 14.38 V 10 V/div 68.1 V 71.9 V
hannel 2 DC Measu
nt Accurac
Channel 1 Bandwidth 425 mV — Channel 2 Bandwidth 425 mV — Sample Rate and Delay Time Accuracy –4 divs +4 divs Channel 1 Edge Trigger Sensitivity Stable trigger — Channel 2 Edge Trigger Sensitivity Stable trigger
5 mV/div 34.05 mV 35.95 mV 500 mV/div 3.405 V 3.595 V 2 V/div 13.62 V 14.38 V 10 V/div 68.1 V 71.9 V
4–2
THS710A, THS720A, THS730A & THS720P Service Manual
Performance Verification
DC
y
AC
y
Resis
y
DMM Tests Low Limit Test Result High Limit
Voltage Accurac
400 mV range, 60 mV input
59.2 mV 60.8 mV
Voltage Accurac
tance Accurac
400 mV range, 360 mV input
4 V range 3.577 V 3.623 V 40 V range 35.77 V 36.23 V 400 V range 357.7 V 362.3 V 880 V range 783 V 801 V 400 mV range 352.3 mV 367.7 mV 4 V range,
600 mV input 4 V range,
3.6 V input 40 V range 35.23 V 36.77 V 400 V range 352.3 V 367.7 V 640 V range 559 V 593 V 400 range 358.0 362.0 4 k range 3.580 k 3.620 k 40 k range,
6 k input 40 k range,
36 k input
357.7 mV 362.3 mV
0.583 V 0.617 V
3.523 V 3.677 V
5.95 k 6.05 k
35.80 k 36.20 k
400 k range 358.0 k 362.0 k 4 M range 3.580 M 3.620 M 40 M range 35.23 M 36.77 M
THS710A, THS720A, THS730A & THS720P Service Manual
4–3
Performance Verification
SCOPEUTILITY

Performance Verification Procedures

In these procedures, TekScope instrument setups are shown in tables. Refer to Conventions on page x for information about how to use these tables.
Before beginning these procedures, two conditions must first be met: H The TekScope instrument must have been operating continuously for ten
minutes within the operating temperature range specified.
H You must perform the Signal Path Compensation operation described below.
If the ambient temperature changes by more than 5° C, you must perform the Signal Path Compensation operation again.
The time required to complete the entire procedure is approximately one hour.
WARNING. Some procedures use hazardous voltages. To prevent electrical shock, always set voltage source outputs to 0 V before making or changing any interconnections.
Signal Path Compensation
Signal path compensation optimizes the oscilloscope accuracy for the current ambient temperature. For maximum accuracy, recompensate the signal path if the ambient temperature changes by 5° C or more.
To compensate the signal path, perform the following steps:
1. Disconnect any probes or cables from the channel 1 and channel 2 input
BNC connectors.
2. Execute the signal path compensation with the following steps:
System Cal Signal Path OK Compensate Signal
Paths
3. When the signal path compensation is complete, press CLEAR MENU to
resume operation of the TekScope instrument.
4–4
THS710A, THS720A, THS730A & THS720P Service Manual
Performance Verification
SCOPEUTILITY
Execute Self Test
This procedure uses internal routines to verify that the TekScope instrument functions and passes its internal self tests. No test equipment or hookups are required. Start the self test with the following setup:
System Diag Loop Once Execute OK Run Test
A dialog box displays the result when the self test completes. Press the CLEAR MENU button to continue operation.
THS710A, THS720A, THS730A & THS720P Service Manual
4–5
Performance Verification
SCOPESA
AS
Check Oscilloscope DC Measurement Accuracy
This test checks the DC measurement accuracy of the oscilloscope.
1. Set the DC voltage source output level to 0 V.
2. Set up the TekScope instrument using the following steps:
VE/
RECALL
CH 1 — VERTICAL
MENU ACQUIRE Mode Average Set to 16 ME
* You may need to press Select Page to display this selection.
Recall Saved Setup
OK Recall Factory
Probe Type Voltage Probe Set to 1X
Select Measrmnt Mean* OK Select
Measrmnt
Recall Factory Setup
3. Connect the TekScope instrument to the DC voltage source as shown below.
DC voltage
source
+
Dual banana to bnc adapter
BNC cable
Ch 1
4–6
THS710A, THS720A, THS730A & THS720P Service Manual
Performance Verification
4. For each VOLTS/DIV setting listed below, perform the following steps: a. Set the DC voltage source output level to the positive voltage listed and
then record the mean measurement as V
pos
.
b. Reverse the polarity of the DC voltage source and then record the mean
measurement as V
neg
.
c. Calculate V
diff
= V
pos
– V
and then compare V
neg
to the accuracy
diff
limits in the table.
DC Voltage Source Output
VOL TS/DIV Setting
5 mV/div +17.5 mV, –17.5 mV 34.05 mV to 35.95 mV 500 mV/div +1.75 V, –1.75 V 3.405 V to 3.595 V 2 V/div +7.00 V, –7.00 V 13.62 V to 14.38 V 10 V/div +35.0 V, –35.0 V 68.1 V to 71.9 V
Levels
Accuracy Limits for V
diff
5. Set DC voltage source output level to 0 V.
6. To check channel 2, repeat step 2 substituting CH 2 for CH 1.
7. Press CH 1 and WAVEFORM OFF to remove the channel 1 waveform
from the display.
8. Repeat steps 3 through 5, substituting CH 2 for CH 1, to complete the check of channel 2.
THS710A, THS720A, THS730A & THS720P Service Manual
4–7
Performance Verification
SCOPESA
AS
Check Oscilloscope
Channel 1 Bandwidth
This test checks the bandwidth of channel 1.
1. Set up the TekScope instrument using the following steps:
VE/
RECALL
ACQUIRE Mode Average Set to 16 TRIGGER Coupling Noise Reject — ME
* You may need to press Select Page to display this selection.
Recall Saved Setup
OK Recall Factory
Gating & High-Low Setup
Select Measrmnt Pk-Pk* OK Select
Measrmnt
Recall Factory Setup
High-Low Method Min/Max
2. Connect the TekScope instrument to the leveled sine wave generator as
shown below.
Leveled sine wave generator
Output
BNC cable
50 feedthrough
terminator
Ch 1
4–8
THS710A, THS720A, THS730A & THS720P Service Manual
Performance Verification
3. Set the TekScope instrument VOLTS/DIV to 100 mV/div.
4. Set the TekScope instrument SEC/DIV to 10 s/div.
5. Set the leveled sine wave generator frequency to 50 kHz.
6. Set the leveled sine wave generator output level so the peak-to-peak measurement is between 599 mV and 601 mV.
7. Set the leveled sine wave generator frequency to 200 MHz if you are
checking a THS730A
, 100 MHz if you are checking a THS720A or a
THS720P, or 60 MHz if you are checking a THS710A.
8. Press MAG.
9. Set the TekScope instrument SEC/DIV to 10 ns/div.
10. Check that the peak-to-peak measurement is 425 mV.
11. Press MAG.
12. Proceed to the next test to check the channel 2 bandwidth.
THS710A, THS720A, THS730A & THS720P Service Manual
4–9
Performance Verification
SCO
———
AS
Check Oscilloscope
Channel 2 Bandwidth
This test checks the bandwidth of channel 2.
1. First check the channel 1 bandwidth using the previous test. Then, perform
these additional steps to check the channel 2 bandwidth:
PE
* You may need to press Select Page to display this selection.
CH 1 WAVEFORM
OFF CH 2 TRIGGER Trigger Source Ch2 ME
Gating & High-Low Setup
Select Measrmnt Pk-Pk* OK Select
Measrmnt
High-Low Method Min/Max
2. Connect the TekScope instrument to the leveled sine wave generator as
shown below.
Leveled sine wave generator
Output
BNC cable
50 feedthrough
terminator
Ch 2
4–10
THS710A, THS720A, THS730A & THS720P Service Manual
Performance Verification
3. Set the TekScope instrument VOLTS/DIV to 100 mV/div.
4. Set the TekScope instrument SEC/DIV to 10 s/div.
5. Set the leveled sine wave generator frequency to 50 kHz.
6. Set the leveled sine wave generator output level so the peak-to-peak measurement is between 599 mV and 601 mV.
7. Set the leveled sine wave generator frequency to 200 MHz if you are
checking a THS730A
, 100 MHz if you are checking a THS720A or a
THS720P, or 60 MHz if you are checking a THS710A.
8. Press MAG.
9. Set the TekScope instrument SEC/DIV to 10 ns/div.
10. Check that the peak-to-peak measurement is 425 mV.
THS710A, THS720A, THS730A & THS720P Service Manual
4–11
Performance Verification
SCOPESA
Check Oscilloscope
Sample Rate and Delay
Time Accuracy
This test checks the oscilloscope time base accuracy.
1. Set up the TekScope instrument using the following steps:
VE/
RECALL
VERTICAL MENU
Recall Saved Setup
OK Recall Factory
Probe Type Voltage Probe Set to 1X
Recall Factory Setup
2. Connect the TekScope instrument to the time mark generator as shown
below.
Time mark
generator
50 feedthrough
terminator
Ch 1
Output
BNC cable
3. Set the time mark generator period to 10 ms.
4. Set the TekScope instrument VOLTS/DIV to 500 mV/div.
5. Use the vertical POSITION rocker to center the test signal on screen.
4–12
THS710A, THS720A, THS730A & THS720P Service Manual
Performance Verification
SCO
SCO
6. Press SET LEVEL TO 50%.
7. Press MAG.
8. Change the TekScope instrument setup using the following steps:
PE
HORIZON­TAL MENU
CLEAR MENU
Time Base Delayed Runs After
Main
9. Set the delayed time base SEC/DIV to 50 ms/div.
10. Change the TekScope instrument setup using the following steps:
PE
HORIZON­TAL MENU
CLEAR MENU
Time Base Delayed Runs After
Main
Set delay time to 10 ms
11. Set the delayed time base SEC/DIV to 500 ns/div.
12. Check that the rising edge of the marker crosses the center horizontal
graticule line within ±4 divisions of center graticule.
NOTE. One division of displacement from graticule center corresponds to a 50 ppm time base error.
THS710A, THS720A, THS730A & THS720P Service Manual
4–13
Performance Verification
SCOPESA
AS
Check Oscilloscope
Channel 1 Edge Trigger
Sensitivity
This test checks the edge trigger sensitivity for channel 1.
1. Set up the TekScope instrument using the following steps:
VE/
RECALL
ACQUIRE Mode Average Set to 16 TRIGGER Mode Normal — ME
* You may need to press Select Page to display this selection.
Recall Saved Setup
OK Recall Factory
Gating & High-Low Setup
Select Measrmnt Ampl* OK Select
Measrmnt
Recall Factory Setup
High-Low Method Min/Max
2. Connect the TekScope instrument to the leveled sine wave generator as
shown below.
Leveled sine wave generator
Output
BNC cable
50 feedthrough
terminator
Ch 1
4–14
THS710A, THS720A, THS730A & THS720P Service Manual
Performance Verification
3. Set the leveled sine wave generator frequency to 200 MHz if you are checking a THS730A or to 100 MHz if you are checking a THS710A, THS720A or THS720P.
4. Set the TekScope instrument VOLTS/DIV to 500 mV/div.
5. Press MAG.
6. Set the TekScope instrument SEC/DIV to 10 ns/div.
7. Set the leveled sine wave generator output level to approximately 50 mV
p-p
so that the measured amplitude is approximately 500 mV. (The measured amplitude can fluctuate around 500 mV.)
8. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and then check that triggering is stable.
9. Change the TekScope instrument setup using the following steps:
SCOPE TRIGGER Slope \ (falling edge)
10. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and then check that triggering is stable.
11. Change the TekScope instrument setup using the following steps:
SCOPE TRIGGER Slope / (rising edge)
12. Proceed to the next test to check the channel 2 edge trigger sensitivity.
THS710A, THS720A, THS730A & THS720P Service Manual
4–15
Performance Verification
SCO
———
AS
Check Oscilloscope
Channel 2 Edge Trigger
Sensitivity
This test checks the edge trigger sensitivity for channel 2.
1. First check the channel 1 edge trigger sensitivity using the previous test.
Then, perform these additional steps to check the channel 2 edge trigger sensitivity:
PE
* You may need to press Select Page to display this selection.
CH 1 WAVEFORM
OFF CH 2 TRIGGER Trigger Source Ch2 ME
Gating & High-Low Setup
Select Measrmnt Ampl* OK Select
Measrmnt
High-Low Method Min/Max
2. Connect the TekScope instrument to the leveled sine wave generator as
shown below.
Leveled sine wave generator
Output
BNC cable
50 feedthrough
terminator
Ch 2
4–16
THS710A, THS720A, THS730A & THS720P Service Manual
Performance Verification
3. Set the leveled sine wave generator frequency to 200 MHz if you are checking a THS730A or to 100 MHz if you are checking a THS710A, THS720A or THS720P.
4. Set the TekScope instrument VOLTS/DIV to 500 mV/div.
5. Set the leveled sine wave generator output level to approximately 50 mV
p-p
so that the measured amplitude is approximately 500 mV. (The measured amplitude can fluctuate around 500 mV.)
6. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and then check that triggering is stable.
7. Change the TekScope instrument setup using the following steps:
SCOPE TRIGGER Slope \ (falling edge)
8. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and then check that triggering is stable.
THS710A, THS720A, THS730A & THS720P Service Manual
4–17
Performance Verification
Check Meter DC Voltage
Accuracy
This test checks the meter DC voltage accuracy.
1. Set the DC voltage source output level to 0 V.
2. Set up the TekScope instrument using the following steps:
METER VDC
3. Connect the TekScope instrument to the DC voltage source as shown below.
DC voltage
source
+
Banana cables
4–18
THS710A, THS720A, THS730A & THS720P Service Manual
Performance Verification
4. For each range, set the DC voltage source output to the level listed below, and then compare the meter reading to the accuracy limits.
DC Voltage Source
Range
400 mV 60 mV 59.2 mV to 60.8 mV 400 mV 360 mV 357.7 mV to 362.3 mV 4 V 3.6 V 3.577 V to 3.623 V 40 V 36 V 35.77 V to 36.23 V 400 V 360 V 357.7 V to 362.3 V 880 V 792 V 783 V to 801 V
Output Level
Accuracy Limits
5. Set the DC voltage source output level to 0 V.
THS710A, THS720A, THS730A & THS720P Service Manual
4–19
Performance Verification
Check Meter AC Voltage
Accuracy
This test checks the meter AC voltage accuracy.
1. Set the AC voltage source output level to 0 V.
2. Set up the TekScope instrument using the following steps:
METER VAC
3. Connect the TekScope instrument to the AC voltage source as shown below.
AC voltage
source
Banana cables
4–20
THS710A, THS720A, THS730A & THS720P Service Manual
Performance Verification
4. Set the AC voltage source output frequency to 500 Hz.
5. For each range, set the AC voltage source output to the level listed below,
and then compare the meter reading to the accuracy limits.
AC Voltage Source
Range
400 mV 360 mV 352.3 mV to 367.7 mV 4 V 600 mV 0.583 V to 0.617 V 4 V 3.6 V 3.523 V to 3.677 V 40 V 36 V 35.23 V to 36.77 V 400 V 360 V 352.3 V to 367.7 V 640 V 576 V 559 V to 593 V
Output Level
Accuracy Limits
6. Set the AC voltage source output level to 0 V.
THS710A, THS720A, THS730A & THS720P Service Manual
4–21
Performance Verification
Check Meter Resistance
Accuracy
This test checks the meter resistance measurement accuracy.
1. Set up the TekScope instrument using the following steps:
METER
2. Connect the TekScope instrument to the resistance standard as shown below.
Resistance
standard
Banana cables
4–22
THS710A, THS720A, THS730A & THS720P Service Manual
Performance Verification
3. For each range, set the resistance standard to the resistance listed below, and then compare the meter reading to the accuracy limits.
Resistance Standard
Range
400 360 358.0 to 362.0 4 k 3.6 k 3.580 kto 3.620 k 40 k 6 k 5.95 kto 6.05 k 40 k 36 k 35.80 kto 36.20 k 400 k 360 k 358.0 kto 362.0 k 4 M 3.6 M 3.580 Mto 3.620 M 40 M* 36 M* 35.23 M to 36.77 M
* These limits apply only when the relative humidity is 60%.
Setting
Accuracy Limits
This completes the performance verification procedure.
THS710A, THS720A, THS730A & THS720P Service Manual
4–23
Performance Verification
4–24
THS710A, THS720A, THS730A & THS720P Service Manual

Adjustment Procedures

Wavetek 9100 U
ers
e
F
5500A
-prod
C
C
This chapter contains adjustment procedures for the TekScope instrument. Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service procedures. Also refer to the chapter Operating Information for information about using the TekScope instrument.
NOTE. The voltage references inside the TekScope instrument are very stable over time and should not require routine adjustment. Before performing any proce­dure in this chapter, first verify that the instrument does not meet specifications. Refer to the chapter Performance Verification for procedures to verify the specifications.

Required Equipment

The following equipment, or a suitable equivalent, is required to complete these procedures.
Minimum
Description
DC Voltage Source 400 mV to 800 V, ±0.1%
AC Voltage Source 400 mV to 600 V, ±0.5%
Resistance Standard 100 W to 10 MW, ±0.1%
Fast-Rise Square Wave Generator
Banana to Banana Cable (two required)
50 W BNC Cable BNC male to BNC male,
50 W BNC Cable BNC male to BNC male,
50 W Feedthrough Termination
Requirements
accuracy
accuracy at 500 Hz
accuracy 300 mV step, 20 ns risetime,
0.5% flattness from 500 ns to 200 ms, 500 Hz or 950 Hz, nominal 50% duty cycle
Shielded banana jacks on each end
36 in (1 m) long
10 in (25 cm) long
BNC male and female con­nectors
Examples
niv
al Calibration System with Oscil­loscope Calibration Modul (Option 250)
luke
Calibrator with Oscilloscope
alibration Option (Option
alibration Option (Option
5500A-SC)
Tektronix Deluxe Meter Lead Set (ATLDX1)
Tektronix part number 012-0482-00
Tektronix part number 012-0208-00
Tektronix part number 01 1-0049-01
Multi
uct
THS710A, THS720A, THS730A & THS720P Service Manual
5–1
Adjustment Procedures
Minimum
Description Examples
Dual Banana to BNC Adapter Banana plugs to BNC female Tektronix part number
Requirements
103-0090-XX
BNC T One male and two female

Overview of the Adjustment Process

The TekScope instrument is protected from inadvertent adjustment by an internal adjustment-lockout jumper. Some disassembly is required to access and remove the jumper before you can adjust the instrument. After adjustment is complete, the jumper should be replaced to protect the calibration of the instrument.
Oscilloscope adjustment and meter adjustment are independent procedures. You can choose to adjust either or both. Each procedure consists of a series of steps; as you move through these steps, the TekScope instrument display provides instructions that describe the specific input signal requirements for each step.
NOTE. During some steps, the instrument may appear to be idle for several minutes while it is processing information internally. Please have patience.
If all steps in the procedure are completed successfully, a “Pass” message is displayed and the new calibration constants take affect. If any step fails, the procedure is aborted and the current calibration is not affected. You can also choose to abort the procedure at any step.
BNC connectors
Tektronix part number 103-0030-XX
5–2
Before performing adjustment procedures, you must warm up the instrument at least ten minutes in an ambient temperature between 20° C and 30° C. Adjust­ments performed prior to warm-up or outside this temperature range may result in poor performance.
In this manual, TekScope instrument setups are shown in tables. Refer to Conventions on page x for information about how to use these tables.
THS710A, THS720A, THS730A & THS720P Service Manual

Accessing the Adjustment-Lockout Jumper

Before adjusting the instrument you must (temporarily) remove the internal adjustment-lockout jumper. You will need a 12 in lb (1.3 N m) torque-limiting screwdriver with Torx T-15 tip and a pair of needle-nose pliers to access and remove the jumper.
WARNING. Before doing this or any other procedure in this manual, read the
General Safety Summary and Service Safety Summary found at the beginning of this manual. Also, to prevent possible injury to service personnel or damage to electrical components, read Preventing ESD on page 6–1.
Use this procedure to access and remove the adjustment-lockout jumper.
1. Disconnect all probes, cables, or meter leads from the instrument.
2. Disconnect the AC adapter, open the battery door, and remove the battery.
3. Place the TekScope instrument face down on a soft surface (such as an
anti-static mat).
Adjustment Procedures
4. Use the torque-limiting Torx T-15 screwdriver to remove the four screws
that hold the case together.
5. Holding the case together, pick it up and place it back side down on a soft
surface.
6. As you lift the front cover off the instrument, press the buttons as necessary
to prevent the switch mat from coming off with the front cover.
7. To remove the switch mat, lift it off the switch flex-circuit assembly. Place it
button-side down on a clean surface.
CAUTION. To avoid contamination of switch contacts, do not touch the carbon contacts on the switch flex-circuit assembly or on the back side of the switch mat.
8. Fold the switch flex-circuit assembly toward you to expose the display
module, inverter board, and adjustment-lockout jumper. The jumper location is shown in Figure 5–1.
9. Use the needle nose pliers to remove the adjustment-lockout jumper from the
instrument. Set the jumper aside for replacement after the adjustment is complete.
10. To reassemble the instrument, fold the switch flex-circuit assembly back
over the display module and inverter board.
THS710A, THS720A, THS730A & THS720P Service Manual
5–3
Adjustment Procedures
Adjustment-lockout
jumper
Figure 5–1: Location of adjustment-lockout jumper
11. To install the switch mat, place it (contact side down) on the switch
flex-circuit assembly. The five rubber guideposts fit into holes in the circuit board.
12. Place the front cover assembly onto the instrument. Taking care not to pinch
the handle, align it into the guides in the front cover. Align the buttons so they all protrude through the holes in the front cover.
13. Holding the case together, pick it up and place it front side down on a soft
surface.
CAUTION. To avoid cross-threading or cutting new threads with the screws, carefully follow the procedure in the next step.
14. To install the four screws, follow these steps: a. Place the screws into their holes in the back cover. b. Using the torque-limiting Torx T-15 screwdriver, slowly turn each
screw backward (counterclockwise) until you feel the thread drop and then gently tighten the screw (turn clockwise) into the existing thread.
5–4
THS710A, THS720A, THS730A & THS720P Service Manual
15. Replace the battery as shown on page 2–2, and then close the battery door.

Oscilloscope Adjustment

Three equipment setups, shown in Figure 5–2, are required to complete this adjustment procedure. For each step in the procedure, the specific input signal required by that step is described in the TekScope instrument display. Refer to these diagrams as necessary during the procedure. Table 5–1 summarizes the steps in the procedure and the signal requirement for each step.
T able 5–1: Summary of oscilloscope adjustment steps
Step Uses Equipment Setup Signal Source Setting
1 Square wave generator setup Fast rise output, 300 mV
Adjustment Procedures
c. When all four all screws are in, hold a corner of the case together firmly
to compress the gasket while tightening its screw until snug. Repeat for the other corners. Do not overtighten the screws (14 in lbs or 1.5 N m maximum torque).
amplitude,
500 Hz or 950 Hz frequency
p-p
2 Zero voltage setup None 5 DC voltage setup +7.000 V 6 DC voltage setup –7.000 V 7 DC voltage setup +700.0 mV 8 DC voltage setup –700.0 mV 9 DC voltage setup +70.00 mV 10 DC voltage setup –70.00 mV
THS710A, THS720A, THS730A & THS720P Service Manual
5–5
Adjustment Procedures
Square wave generator setup
Fast rise square wave generator
Output
36 inch BNC cable
Zero voltage setup
BNC ‘T’ connector
BNC ‘T’ connector
10 inch BNC cable
Ch 1
10 inch BNC cable
Ch 1
50 feedthrough terminator
Ch 2
50 feedthrough terminator
Ch 2
5–6
DC voltage setup
DC voltage
source
BNC ‘T’ connector
+
Dual banana to bnc adapter
36 inch BNC cable
Figure 5–2: Oscilloscope adjustment setups
THS710A, THS720A, THS730A & THS720P Service Manual
10 inch BNC cable
Ch 1
Ch 2
Adjustment Procedures
SCOPEUTILITY
Oscilloscope Adjustment
Procedure
After removing the adjustment-lockout jumper, use this procedure to adjust the oscilloscope voltage references.
1. Turn on the TekScope instrument and let it warm up for at least ten minutes.
2. Connect the fast rise square wave generator to the instrument as shown in the
Square wave generator setup of Figure 5–2.
3. Set the fast rise square wave generator to produce an amplitude between 250 mV
and 300 mV
p-p
, and frequency between 450 Hz and 950 Hz
p-p
(inclusive).
4. Use the TekScope instrument setup below to start the oscilloscope adjust­ment process. For each step, follow the specific instructions displayed by the instrument to connect the correct signal source and then confirm that you are ready to proceed to the next step.
System Cal Factory Scope OK Factory Cal Scope
Completing the
Adjustment Process
NOTE. During some steps, the instrument may appear to be idle for several minutes while it is processing information internally. Please have patience.
If any step in the procedure fails, the procedure terminates. Failure will result if an incorrect signal source is connected. If this happens, you must start the procedure over from the beginning.
If the procedure completes successfully, a “Pass” message is displayed and the new adjustment takes effect.
After adjusting the oscilloscope, continue with one of the following steps: H Proceed to Meter Adjustment on page 5–8 to adjust the meter voltage
references.
H Proceed to Replacing the Adjustment-Lockout Jumper on page 5–11 to
protect the instrument from inadvertent readjustment.
THS710A, THS720A, THS730A & THS720P Service Manual
5–7
Adjustment Procedures

Meter Adjustment

Three equipment setups, shown in Figure 5–3, are required to complete this adjustment procedure. For each step in the procedure, the specific input signal required by that step is described in the TekScope instrument display. Refer to these diagrams as necessary during the procedure. Table 5–2 summarizes the steps in the procedure and the signal requirement for each step.
T able 5–2: Summary of meter adjustment steps
Step Uses Equipment Setup Signal or Resistance Source Setting
1 DC voltage setup +400.0 mV 2 DC voltage setup 4.000 V 3 DC voltage setup 40.00 V 4 DC voltage setup 400.0 V 5 DC voltage setup 850.0 V 6 AC voltage setup 400.0 mV 7 AC voltage setup 4.000 V 8 AC voltage setup 40.00 V 9 AC voltage setup 400.0 V 10 AC voltage setup 600.0 V
RMS RMS RMS RMS RMS
, 500 Hz , 500 Hz , 500 Hz , 500 Hz , 500 Hz
11 Resistance standard setup 100.0 12 Resistance standard setup 1.000 k 13 Resistance standard setup 10.00 k 14 Resistance standard setup 100.0 k 15 Resistance standard setup 1.000 M 16 Resistance standard setup 10.00 M 17 DC voltage setup 18 DC voltage setup
1
In steps 17 and 18, the DC voltage source must be able to sink approximately 1 mA
1 1
500.0 mV
1.000 V
supplied by the TekScope instrument. You must determine whether or not your DC voltage source is capable of sinking current supplied by a device under test. If the
1
DC voltage source cannot sink current, you can connect a 470 ,
W shunt
8
resistor across the output of the DC voltage source during steps 17 and 18 only. The shunt resistor will accommodate the current supplied by the TekScope instrument.
5–8
THS710A, THS720A, THS730A & THS720P Service Manual
DC voltage setup
AC voltage setup
Adjustment Procedures
DC voltage
source
+
Banana cables
AC voltage
source
Banana cables
Resistance standard setup
Banana cables
Figure 5–3: Meter adjustment setups
Resistance
standard
THS710A, THS720A, THS730A & THS720P Service Manual
5–9
Adjustment Procedures
R
UTILITY
Meter Adjustment
Procedure
After removing the adjustment-lockout jumper, use this procedure to adjust the meter voltage references.
1. Turn on the TekScope instrument and let it warm up for at least ten minutes.
2. Connect the DC voltage source to the instrument as shown in the DC voltage
setup of Figure 5–3.
3. Set the DC voltage source amplitude to 400 mV.
4. Use the TekScope instrument setup below to start the meter adjustment
process. For each step, follow the specific instructions displayed by the instrument to connect the correct signal source and confirm that you are ready to proceed to the next step.
METE
System Cal Factory DMM OK Factory Cal DMM
If any step in the procedure fails, the procedure terminates. Failure will result if an incorrect signal source is connected. If this happens, you must start the procedure over from the beginning.
Completing the
Adjustment Process
If the procedure completes successfully, a “Pass” message is displayed and the new adjustment takes effect.
After adjusting the meter, continue with one of the following steps: H Return to Oscilloscope Adjustment on page 5–5 to adjust the oscilloscope
voltage references.
H Proceed to Replacing the Adjustment-Lockout Jumper on page 5–11 to
protect the instrument from inadvertent readjustment.
5–10
THS710A, THS720A, THS730A & THS720P Service Manual

Replacing the Adjustment-Lockout Jumper

After you complete the adjustment process, you should reinstall the adjustment­lockout jumper to prevent inadvertent loss of calibrration. To replace the jumper, repeat the procedure Accessing the Adjustment-Lockout Jumper on page 5–3. However, in place of step 9, reinstall the jumper onto pins 1 and 2 as shown in Figure 5–4.
Adjustment Procedures
Adjustment-lockout
jumper position
Figure 5–4: Relocating the adjustment-lockout jumper
This completes the adjustment procedure.
THS710A, THS720A, THS730A & THS720P Service Manual
5–11
Adjustment Procedures
5–12
THS710A, THS720A, THS730A & THS720P Service Manual

Maintenance

This chapter contains the information needed to do periodic and corrective maintenance on the TekScope instrument. The following sections are included:
H Preparation refers you to important information contained elsewhere in this
manual.
H Preventing ESD provides general information on preventing damage to
internal modules when doing maintenance.
H Inspection and Cleaning provides procedures for inspecting the TekScope
instrument and cleaning its external and internal modules.
H Removal and Installation Procedures provide procedures for the removal and
installation of modules.
H Troubleshooting provides information for isolating failed modules. H Unpacking and Repacking Instructions provides packaging information for
shipment or storage.

Preparation

Preventing ESD

Only qualified personnel should perform service procedures. Read the Service Safety Summary and the General Safety Summary before performing any service
procedures. Also refer to the chapter Operating Information for information about using the TekScope instrument.
When performing any service which requires internal access to the TekScope instrument, adhere to the following precautions to avoid damaging internal modules and their components due to electrostatic discharge (ESD).
CAUTION. Static discharge can damage any semiconductor component in this instrument.
1. Minimize handling of static-sensitive modules.
2. Transport and store static-sensitive modules in their static protected
containers or on a metal rail. Label any package that contains static-sensitive modules.
THS710A, THS720A, THS730A & THS720P Service Manual
6–1
Maintenance

Inspection and Cleaning

3. Discharge the static voltage from your body by wearing a grounded antistatic
wrist strap while handling these modules. Do service of static-sensitive modules only at a static-free work station.
4. Nothing capable of generating or holding a static charge should be allowed
on the work station surface.
5. Handle circuit boards by the edges when possible.
6. Do not slide the modules over any surface.
7. Avoid handling modules in areas that have a floor or work-surface covering
capable of generating a static charge.
This section describes how to inspect for dirt and damage, and how to clean the TekScope instrument. Inspection and cleaning are done as preventive mainte­nance. Preventive maintenance, when done regularly, may prevent malfunctions and enhance reliability.
General Care
Inspection and Cleaning
Procedures
Preventive maintenance consists of visually inspecting and cleaning the TekScope instrument, and using general care when operating it. How often to do maintenance depends on the severity of the environment in which the instrument is used. A proper time to perform preventive maintenance is just before adjustment of the instrument.
Protect the TekScope instrument from adverse weather conditions. The instru­ment is not waterproof.
Do not store or leave the instrument where the LCD display will be exposed to direct sunlight for long periods of time.
The front and back covers help keep dust out of the TekScope instrument and must be in place during normal operation.
CAUTION. To avoid damage to the TekScope instrument, do not expose it to any sprays, liquids, or solvents.
Inspect and clean the TekScope instrument as often as operating conditions require. The collection of dirt on internal components can cause them to overheat and breakdown. Dirt acts as an insulating blanket, preventing efficient heat dissipation. Dirt also provides an electrical conduction path instrument failure, especially under high-humidity conditions.
that could cause an
6–2
THS710A, THS720A, THS730A & THS720P Service Manual
Maintenance
Inspection — Exterior. Inspect the outside of the TekScope instrument for damage, wear, and missing parts. Instruments that appear to have been dropped or otherwise abused should be checked thoroughly to verify correct operation and performance. Immediately repair defects that could cause personal injury or lead to further damage to the instrument.
Cleaning Procedure — Exterior. To clean the TekScope instrument exterior, perform the following steps:
1. Remove loose dust on the outside of the instrument with a lint free cloth.
Use care to avoid scratching the clear plastic display filter.
2. Use a soft cloth or paper towel dampened with water to clean the instrument.
You can use a 75% isopropyl alcohol solution for more efficient cleaning.
CAUTION. To avoid damage to the surface of the TekScope instrument, do not use any abrasive or chemical cleaning agents.
Inspection — Interior. During normal use, the TekScope instrument case seals the internal components against common dust and dirt. However, if it appears that the instrument may have been damaged or has been exposed to an unusually dirty or wet environment, you should inspect and clean the interior.
To inspect and clean the interior, disassemble the instrument according to the procedure Complete Disassembly on page 6–23.
Inspect the internal portions of the instrument for damage and wear using Table 6–1 as a guide. Defects found should be repaired immediately.
T able 6–1: Internal inspection check list
Item Inspect For Repair Action
Front and Back Cover
Hole Plugs, Gasket and Foam Pads
Circuit boards Loose, broken, or corroded solder
Cracks or deformations. Scratched lettering or display filter. Loose connectors or labels.
Misplaced or missing pieces. Relocate or replace defective
connections. Burned circuit boards. Burned, broken, or cracked circuit-run plating.
Repair or replace defective hard­ware.
pieces. Remove failed module and replace
with a new module.
THS710A, THS720A, THS730A & THS720P Service Manual
6–3
Maintenance
T able 6–1: Internal inspection check list (Cont.)
Item Repair ActionInspect For
Solder connections Cold solder or rosin joints. Resolder joint and clean with
isopropyl alcohol.
Wiring and cables Loose plugs or connectors.
Burned, broken, or frayed wiring.
Firmly seat connectors. Repair or replace modules with defective wires or cables.
Cleaning Procedure — Interior. To clean the TekScope instrument interior, perform the following steps:
1. Blow off dust with dry, low-pressure, deionized air (approximately 9 psi).
2. Remove any remaining dust with a lint-free cloth dampened in isopropyl
alcohol (75% solution) and rinse with warm deionized water. (A cotton­tipped applicator is useful for cleaning in narrow spaces and on circuit boards.)
STOP. If, after performing steps 1 and 2, a module is clean upon inspection, skip steps 3 and 4. Proceed with step 5.
If steps 1 and 2 do not remove all the dust or dirt, all modules in the instrument except the Display Module may be spray washed using a solution of 75% isopropyl alcohol. Proceed with step 3.
3. Spray wash dirty parts with the 75% isopropyl alcohol solution and wait
60 seconds for the majority of the alcohol to evaporate.
6–4
4. Use hot (49_ C to 60_ C / 120_ F to 140_ F) deionized water to thoroughly
rinse the parts.
5. Dry all parts with low-pressure, deionized air.
6. Dry all components and assemblies in an oven or drying compartment using
low-temperature (52_ C to 66_ C / 125_ F to 150_ F) circulating air.
Lubrication. There is no periodic lubrication required for the TekScope instru­ment.
THS710A, THS720A, THS730A & THS720P Service Manual

Removal and Installation Procedures

This section describes how to remove and install the major mechanical and electrical modules. It provides the following information:
H List of Modules on page 6–5 describes where to locate a list of replaceable
modules.
H Summary of Procedures on page 6–6 lists the procedures for removal and
installation of modules.
H Tools Required on page 6–7 describes the tools needed to perform the
procedures.
H Tilt Stand, on page 6–8, begins a series of procedures that describe the
removal and installation of modules.
Maintenance
Preparation
List of Modules
Please read the following warning statement. Then read the following general instructions before removing a module.
WARNING. Only qualified personnel should perform service procedures. Before doing this or any other procedure in this manual, read the General Safety Summary and Service Safety Summary found at the beginning of this manual. Also, to prevent possible injury to service personnel or damage to electrical components, read Preventing ESD on page 6–1.
1. Read the Summary of Procedures on page 6–6 to understand which
procedure to use to access a given module. Then read Tools Required on page 6–7 for a list of tools needed to remove and install modules in the TekScope instrument.
2. If you are disassembling the TekScope instrument for cleaning, refer to the
Inspection and Cleaning procedure on page 6–2 for cleaning instructions.
The Mechanical Parts List chapter provides a list of all replaceable modules. Any electrical or mechanical module, assembly, or part listed in the parts list is referred to as a module.
THS710A, THS720A, THS730A & THS720P Service Manual
6–5
Maintenance
Summary of Procedures
Table 6–2 lists the removal and installation procedures in order of increasing complexity.
T able 6–2: Removal and installation procedures
Procedure Modules Accessed Begins on Page
Tilt Stand Tilt stand 6–8 Front-Panel Label Front-panel label 6–10 Front Cover Front cover
Display shield Switch mat Gasket Handle Battery door Screws
Display Module Display module
Inverter board
Back Cover DC power hole plug
6–1 1
6–16
6–19 I/O port hole plug Back cover / back label / battery contact
assembly
Complete Disassembly Chassis / BNC connectors / banana jacks
Main board Switch flex-circuit assembly Display cable Inverter board cable Foam pad set
2
Some pieces in the foam pad set are accessed by the Front Cover and Back Cover procedures.
2
6–23
6–6
THS710A, THS720A, THS730A & THS720P Service Manual
Maintenance
Tools Required
You will need the following tools to remove and replace all the modules in the TekScope instrument.
H Torque-limiting screwdriver, 12 in lb (1.3 N m) range with Torx T-15 tip H Screwdriver with H Screwdriver with
1
@
inch flat blade
4
1
@
inch flat blade
8
H Needle-nose pliers H Sharp knife (X-ACTO knife or equivalent)
THS710A, THS720A, THS730A & THS720P Service Manual
6–7
Maintenance
Tilt Stand
1
@
You will need a
inch flat-bladed screwdriver to remove the tilt stand.
4
STOP. You do not have to remove the tilt stand to access internal modules. In most cases, you do not have to remove the tilt stand to clean the TekScope instrument exterior.
Removal. Use this procedure to remove a defective tilt stand.
1. Place the TekScope instrument face down on a soft surface (such as an
anti-static mat).
2. Insert the
1
@
inch flat-bladed screwdriver into the gap between the tilt stand
4
and instrument case as shown in Figure 6–1.
3. Twist the screwdriver gently to release one side of the tilt stand from the
instrument case. Take care not to damage the case.
6–8
Figure 6–1: Removing the tilt stand
THS710A, THS720A, THS730A & THS720P Service Manual
Maintenance
Installation. Use this procedure to install a new tilt stand.
1. Orient the tilt stand to the back of the instrument case with the hinge pins
toward the top and the side bump toward the handle.
2. Insert one hinge pin into its socket and place the other into the cavity in the
back of the instrument case as shown in Figure 6–2.
3. Rotate the tilt stand to slide the second hinge pin upward and into its socket.
1
Insert hinge pin into the socket.
2
Rotate the tilt stand to slide the second hinge pin into its socket.
Figure 6–2: Installing the tilt stand
THS710A, THS720A, THS730A & THS720P Service Manual
6–9
Maintenance
Front-Panel Label
You will need a sharp knife and needle-nose pliers to remove a defective front-panel label.
CAUTION. Removal of a front-panel label will destroy it. Do not remove it unless it is defective and needs to be replaced.
Removal. Use this procedure to remove a defective front-panel label.
1. Use the sharp knife to pry up a corner of the label.
2. Use the needle-nose pliers to peel the label off.
Installation. Use this procedure to install a new front-panel label in either a new or the existing front cover.
1. Peel the paper off the back side of the new label.
2. Handling the label by its edges only, align it with the recess in the front
cover as shown in Figure 6–3.
3. When properly aligned, press the new label into place.
Front-panel label
Recess in the front cover
Figure 6–3: Installing a new front-panel label
6–10
THS710A, THS720A, THS730A & THS720P Service Manual
Maintenance
Front Cover
You will need a torque-limiting Torx T-15 screwdriver, needle-nose pliers, and sharp knife to remove the front cover and other immediately accessible modules.
CAUTION. Attempting to remove the display shield from the front cover will likely destroy both pieces. If one or the other are defective, both must be replaced.
Removal. Use this procedure to remove the front cover and access the handle, switch mat, battery door, and gasket.
1. Disconnect all probes, cables, or meter leads from the instrument.
2. Disconnect the AC adapter, open the battery door, and remove the battery.
3. Place the TekScope instrument face down on a soft surface (such as an
anti-static mat).
4. Use the Torx T-15 screwdriver to remove the four screws that hold the case
together.
5. Holding the case together, pick it up and place it back side down on a soft
surface.
6. As you lift the front cover off the instrument, press the buttons as necessary
to prevent the switch mat from coming off with the front cover.
7. To remove the handle, slip it off its guides.
8. To remove the switch mat, lift it off the switch flex-circuit assembly. Place it
button-side down on a clean surface.
CAUTION. To avoid contamination of switch contacts, do not touch the carbon contacts on the switch flex-circuit assembly or on the back side of the switch mat.
9. To remove the battery door, use the needle-nose pliers to slip the retaining
tether off of the post inside the instrument case as shown in Figure 6–4.
THS710A, THS720A, THS730A & THS720P Service Manual
6–11
Maintenance
Figure 6–4: Removing the battery door
NOTE. In most cases, the gasket does not have to be removed from the front cover for interior cleaning.
10. To remove the gasket from the front cover, use the needle-nose priers to lift
it out of the groove in the front cover, starting with the accessible tail at the bottom of the front cover.
Installation. Use this procedure to install the battery door, switch mat, handle, gasket, and front cover.
1. Place the TekScope instrument back side down on a soft surface (such as an
anti-static mat).
2. To install the battery door, pass the retaining tether through the battery
opening from the outside and then use the needle-nose pliers to slip the retaining strap onto the post (see Figure 6–4).
3. To install the switch mat, place it (contact side down) on the switch
flex-circuit assembly. The five rubber guideposts fit into holes in the circuit board.
6–12
THS710A, THS720A, THS730A & THS720P Service Manual
Maintenance
4. If you are installing a new front cover, follow these steps to assemble it
before installation. Refer to Figure 6–5.
a. Install a new label using the procedure on page 6–10. b. Install a new display shield into the inside of the new front cover.
Remove the backing paper from the adhesive on the display shield, align it over the two diagonal posts, and then press it into place. The blue border should be even when viewed from the face of the front cover.
c. Install new foam pad as shown in Figure 6–5.
Display shield
Foam pad
Figure 6–5: Assembling a new front cover
d. Install a new gasket into the groove around the edge of the front cover.
As shown in Figure 6–6, start the gasket at the bottom of the cover leaving a short (
1
@
inch) tail through the slot. Carefully press the gasket
8
into place around the rim with the finish the installation, overlap the starting point and then pass the end of the gasket through the remaining slot. Trim the ending tail to length with the sharp knife.
THS710A, THS720A, THS730A & THS720P Service Manual
Front cover
1
@
inch flat-bladed screwdriver. To
8
1
@
8
inch
6–13
Maintenance
Gasket material
Front cover
Trim tails to
Overlap gasket material
1
@
inch
8
Figure 6–6: Installing the gasket
5. Slip the handle into the guides of the back cover as shown in Figure 6–7.
6. Place the front cover assembly onto the instrument. Taking care not to pinch
the handle, align it into the guides in the front cover. Align the buttons so they all protrude through the holes in the front cover.
6–14
7. Holding the case together, pick it up and place it front side down on a soft
surface.
THS710A, THS720A, THS730A & THS720P Service Manual
Velcro strip
Maintenance
Fold end of handle over
Figure 6–7: Installing the handle
THS710A, THS720A, THS730A & THS720P Service Manual
6–15
Maintenance
Display Module
CAUTION. To avoid cross-threading or cutting new threads with the screws, carefully follow the procedure in the next step.
8. To install the four screws, follow these steps: a. Place the screws into their holes in the back cover. b. Using the torque-limiting Torx T-15 screwdriver, slowly turn each
screw backward (counterclockwise) until you feel the thread drop and then gently tighten the screw (turn clockwise) into the existing thread.
c. When all four all screws are in, hold a corner of the case together firmly
to compress the gasket while tightening its screw until snug. Repeat for the other corners. Do not overtighten the screws (12 in lbs or 1.m N m maximum torque).
9. Replace the battery as shown on page 2–2, and then close the battery door.
1
@
You will need a
inch flat-bladed screwdriver to remove the display module
8
and inverter board.
Removal. First remove the front cover and switch mat using the procedure on page 6–11. Then, use the following procedure to remove the display module and inverter board.
1. Fold the switch flex-circuit assembly toward you to expose the display module and inverter board.
CAUTION. To avoid scratching the surface of the display module, do not let it touch any hard object.
The display module surface may scratch easily. To clean it, first try pressurized air. If a soft cloth is required to clean it, use very light pressure.
2. To remove the inverter board, unplug the input and output connectors and then lift the inverter board off the two guideposts. You may want to use the
1
@
inch flat-bladed screwdriver to help release the connectors from their
8
sockets.
3. To remove the display module, follow these steps: a. Disconnect the cable from the display module to the inverter board, if it
is not already disconnected.
6–16
b. Lift the right side of the display module out of the chassis and fold the
display module to the left, as shown in Figure 6–8.
THS710A, THS720A, THS730A & THS720P Service Manual
Display cable connector
Maintenance
Display module
Inverter board
Spring
Switch flex-circuit board
Figure 6–8: Removing the display module
c. Using the
1
@
inch flat-bladed screwdriver, carefully open the display-
8
cable connector as shown in Figure 6–9.
d. When the connector is open, slip the display cable out of the connector
to free the display module.
THS710A, THS720A, THS730A & THS720P Service Manual
6–17
Maintenance
Contact side of cable
Figure 6–9: Opening the display cable connector
Installation. Use this procedure to install the display module and inverter board.
1. Open the display-cable connector on the back of the display module.
2. Hold the display module upright as shown in Figure 6–8.
1
@
3. Insert the display cable into its connector and then use the
inch flat-bladed
8
screwdriver to close the connector. Check that the display cable is fully inserted into the connector.
4. Fold the display module back into the chassis and seat it between the foam pads.
5. Plug the cable from the display module into the inverter board and then route the cable in the channels as shown in Figure 6–10.
6. Place the inverter board into its cavity over the guideposts in the chassis.
7. Reconnect the connector to the bottom of the inverter board and route the
wires as shown in Figure 6–10.
8. If you just installed a new display module, peel off the protective plastic sheet.
9. Fold the switch flex-circuit assembly back over the display module and inverter board.
6–18
10. Install the switch mat and front cover using the procedure on page 6–12.
THS710A, THS720A, THS730A & THS720P Service Manual
Loading...