TLP058 FlexChannel
Instruct
ions
®
Logic Probe
Connecting the probe to your circuit
Attach the probe to the circuit using the connectors and adapters shown on the back of these instructions. Select the best method for your needs, and then go to Setting up the Probe.
Setting up the probe
The oscilloscope automatically dete cts and congures channels when you connect a logic probe.
To set and view the digital channel parameters, add the channel with the connected logic probe to the display. Double-tap the Waveform badge for the logic probe channel t o open the conguration menu. Digital c hannel settings include threshold voltage (default is 1.4 V), signal height (for all channels), and channel labels.
Using the probe
See the oscilloscope documentation and Help topics for how to display and trigger on digital logic signals and buses.
Characteristic Description
Maximum input toggle rate
Maximum sample rate
Digital to analog trigger skew
Probe length
500 MHz
6.25 GS/s
5ns
1.0 m (3.28 ft)
Table 2: Environmental specications
Characteristic Description
Temperature
Operating Nonoperating
Humidity Noncondensing, and as limited by a maximum
Operating 5% to 90% relative humidity to +40 °C (104 °F)
Nonoperating
e
Altitud
Operating Nonoperating
0°Cto+50° –40 °C to +71 °C (–40 °F to +160 °F)
Wet-Bulb
5% to 55% relative humidity +40 °C to +50 °C (+104 °F to 122 °F)
5% to 90% relative humidity to +40 °C (+104 °F) 5% to 39% relative humidity +40 °C to +60 °C (+104 °F to 140 °F)
3,000 m (9,842 ft) maximum 12,000 m (39,370 ft) maximum
Equipment Recycling. This symbol indicates that this product complies with the applicable European Union requirements according to Directives 2012/19/EU and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries. For information about recycling options, check the Tektronix Web site.
C(+32°Fto+122°F)
temperature of +39 °C (+102 °F)
1
*P0713
51501*
071-3515-01
Product description
The TLP058 FlexChannel®logic probe connects the Tektronix 5 Series Mixed Signal Oscilloscopes (MSO54, MSO56, MSO58, MSO58LP) to digital buses and signals on your device under test (DUT). The probe contains 8 data channels. You can connect the TLP058 logic probe to any FlexChannel oscilloscope input channel.
All leads include a ground connection at the tip. You can connect the probe leads separately to the target system, or group the leads together using the probe tip holder.
Functional check
Follow the instructions for connecting the probe to the oscilloscope and setting up the probe. Connect your probe leads to active logic signals on your DUT. You should see logic activity immediately on all connected, active channels. If you do not see an active signal, use an analog probe to verify signal activity on the logic signal points.
Typical application
Safety summary
Connect and disconnect properly. Connect the probe output
to the measurement instrument before connecting the probe to the circuit under test. Disconnect the probe input and the probe ground from the circuit under test before disconnecting the probe from the measurem ent instrument.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product. Connect the probe reference lead to earth ground only.
Do not operate without covers.
Avoid exposed circuitry. Do not touch exposed connections
and components when power is present.
Do not o perate with suspected failures. If you suspect there is damage to this product, have it inspected by qualied service personnel.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Safety terms and symbols in this manual.
These t erms may appear in this manual:
WARNING. Warning statements identify conditions or
practices that could result in injury or loss of life.
Connecting the probe to the oscilloscope
1. Move the locking lever to the unlocked position then let go to reset locking lever to the center position.
2. Insert the probe into a FlexChannel channel until fully seated and the lock mechanism clicks.
3. Mov
e the locking lever to the locked position. The status
light should be a solid green.
E. It is normal for the logic probe Status light to ash
NOT
green when the oscilloscope is powering on, and will turn to a
ady green once the oscilloscope is fully running.
ste
NOTE. If the status LED continues ashing green, ashing
, or is a steady red, contact Tektronix Customer Support
red for assistance.
4. To r
emove the probe, move and hold the locking lever at
the unlocked position and pull out the probe.
1. Use the logic probe to view digital signals on a system bus.
2. Use an analog probe to view analog waveform information
of a digital signal.
Accessories
The probe comes standard with a Logic Probe Accessory Kit (Tektronix part number 020-31
70-XX). See the illustration on the following page. See the label in the accessories box lid for information on the individual
accessories.
Specications
Table 1: Electrical and mechanical specications
Characteristic Description
Input channels 8 digital
Input resistance
Input capacitance 3.0 pF
Input signal swing
Minimum 400 mV p-p
Maximum
Maximum nondest– ructive input signal
Threshold voltage ± 40 V
nimum detectable
Mi pulse width
100 k±1.0%
30 V p-p, 200 MHz (centered around the DC threshold voltage) at the probe tip
10 V p-p, 200 MHz (centered around the DC threshold voltage) at the probe tip
30 V p-p, ±42 V peak, ±50 V
s
1n
DC
CAUTION. Caution statements identify conditions or
practices that could result in damage to this product or other property.
Symbols on the product. This symbol may appear on the product:
x
Connecting the probe to the circuit