Tektronix TDSPWR2 Reference manual

Getting Started with TDSPWR2
Y ou can use this side of the Quick Reference to start to take measurements with the Power Measurement and Analysis Software (TDSPWR2). The other side contains a complete menu tree for TDSPWR2 software.
and General Safety information, refer to the Online Help for the application.
The TDSPWR2 Power Measurements and Analysis software transforms a digital oscilloscope into an analysis tool that measures and analyzes power dissipation in power supply switching devices and magnetic components. It then generates detailed test reports in customizable formats.
Performing a Switching Loss Measurement
To measure Switching Loss, follow these steps:
1. Select File> Run Application> TDSPWR2 in the
oscilloscope menu bar.
2. Connect the probes to the device under test. For
example, the test setup for Switching Loss is shown below.
4. Select the Switching Loss option.
5. Configure the options in the common configuration
panel.
6. Select the Source and assign the Voltage and Current
channels.
7. Set the appropriate I-Probe settings.
9. Push the “Running Man” button to start taking
measurements. If the measurement is successful, the application automatically displays the results. You can also view the results by selecting View> Results from the application menu bar.
10. The application displays results for the measurement in
three tabs: T-on Power Loss, T-off Power Loss and Power Loss.
3. In the Power Device tab, select Power Dissipation. Press
Configure.
For up-to-date information on Tektronix oscilloscope solu­tions for Power Measurement, access the www.tektronix.com/Measurement/scopes/
TDSPWR2 Ordering Information
Supports the TDS5000, TDS6000 CSA7000
Applications Software on Windows-Based Oscilloscopes Installation Manual for a complete list of supported
models. The applications CD includes a PDF file of the installation manual.
If you order Option PW2 along with TDS5000/TDS7000/TDS6000/CSA7000 the Power Analysis and Measurements Software will already be installed and enabled.
1
series oscilloscopes; refer to the Optional
1
web page.
, TDS70001,and
8. Select the Deskew option or select Utilities>Deskew to
deskew.
To order an upgrade for an existing oscilloscope:
H Order the TDS5UP Option PW2
H Order the TDS7UP Option PW2
H Order the TDS6UP Option PW2
H Order the CSA7UP Option PW2
Recommended Accessories
Opt. 2M for TDS5000
Opt 2M, 3M for TDS7000 and CSA7000
Deskew Fixture--067-1478-00
Current Probe--Order TCP202, TCP300, TCP400 or AM503B with A63XX probes
Differential Probe--Order P5205, P5210, P5200 and ADA400A
1
Probe adapter: -- TCA--1MEG, while using 1 Mvoltage
probes with TDS6000/CSA7000/TDS7154/ TDS7254/TDS7404
11. T o generate a report, select Report> Report Generator.
12. Select the template in the Generate Report tab.
13. Select the Generate button to post the test data to the
template.
TDSPWR2 Power Measurement and Analysis Reference
www.tektronix.com
*P071108202*
071-1082-02
TDSPWR2 Menu Tree
File
Recall Default Recall Save Recently Recalled Recently Saved Preferences Minimize Exit
Acquisition Alert Message Last Recalled Settings
Tab
Selection
Power Dissipation
BH Analysis
Total Loss
SOA
Dynamic Resistance
Measurements configuration
Hi Power Finder
Switching Loss
I vs V Integral Plot
Magnetic Loss
Source
Source Contd.
Math Destination
SOA Normal SOA Gated SOA DPO
Source
Real Time Math Destination Average Math Destination
Source
Ref Level Hysteresis Waveform Destination
Percentage Absolute
Source
Waveform Destination
Source
Waveform Destination Hysteresis
Switching Loss Magnetic Loss
Switching Loss Magnetic Loss Push Pull Type
Switching Loss
Magnetic Loss Enable Mask Mask Editor option
Source
Math1|Math2|Math3|Math4
Live* Ref
Live* Ref
Live* Ref
Live* Ref
I--Probe Settings
I--Probe Settings
I--Probe Settings
Ref Level | High Hysteresis | Low
Push Pull Type
I--Probe Settings
I--Probe Settings
Live* Ref
Percentage Absolute
Hysteresis
I--Probe Settings
TCP Series
Current Shunt
Custom
AM503S Settings
TCP Series
Current Shunt
Custom
AM503S Settings
Ref Level | High Hysteresis | Low
TCP Series
Current Shunt
Custom
AM503S Settings
Measurements
Select
Configure
Power Device
Line Power
Output Analysis
di / dt dv / dt
Power Quality
Current Harmonics
Total Power Quality
Ripple Line
Ripple Switching
Source
Math Destination
Source
Power Waveform Destination
Energy Waveform Destination
Source
Signal Type
Harmonics
I--Probe
Waveform Destination Energy Waveform
Source
Coupling
Bandwidth
Acquisition Mode
Switching Frequency
Math1|Math2|Math3|Math4
Live* Ref
Math1|Math2|Math3|Math4
Live* Ref
IEC Std MIL Std
Deskew I--Probe Settings
Harmonics Table Configure to IEC Standard
I--Probe Impedance Tables
Power Waveform
Live
AC -- DC
20 MHz, 250 MHz, Full
High Resolution, Pk--Pk
150 KHZ
I--Probe Settings
Deskew I--Probe
Settings
50 Hz 60 Hz
60 Hz 400 Hz
TCP Series
Current Shunt
Custom
AM503S Settings
View
Utilities
Report
Help
* You can also select Math channels with Live signals.
Results
Deskew
Report Generator
Help Topics About TDSPWR 2
Modulation Analysis
Static Real Time
De fin e Test Te mpl ate Define Report LayOut Generate Report
From To
From To
Turn-- On Time
Pulse Width
Duty Cycle
Clock Period
Frequency Modulation
Line Voltage
Output Voltage
Max Turn--On Time
Converter Selection
Source Ref Destination Ref Level Polarity
Source Ref Destination Ref Level Polarity
Edge
Source Ref Destination Ref Level
Edge
Source Ref Level Ref Destination
Edge
Line Voltage Source Max Line Voltage
Output 1 | Output 2 | Output 3
Max Turn--On Time
AC -- DC AC -- DC
Level Percentage
Positive Negative
Rise Fall
Level Percentage
Rise Fall
Level Percentage
Rise Fall
Frequency
Copyright Tektronix, Inc.