Tektronix TDSJIT1 User Manual

User Manual
TDSJIT1 Jitter Analysis Application
071-0527-00
This document applies to software version 1.0.0 and above.
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
Copyright © T ektronix, Inc. All rights reserved. Licensed software products are owned by Tektronix or its suppliers and are protected by United States copyright laws and international treaty provisions.
Use, duplication, or disclosure by the Government is subject to restrictions as set forth in subparagraph (c)(1)(ii) of the Rights in T echnical Data and Computer Software clause at DFARS 252.227-7013, or subparagraphs (c)(1) and (2) of the Commercial Computer Software – Restricted Rights clause at F AR 52.227-19, as applicable.
T ektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
Printed in the U.S.A. T ektronix, Inc., P.O. Box 1000, Wilsonville, OR 97070–1000 TEKTRONIX and TEK are registered trademarks of T ektronix, Inc.
WARRANTY
T ektronix warrants that the media on which this software product is furnished and the encoding of the programs on the media will be free from defects in materials and workmanship for a period of three (3) months from the date of shipment. If a medium or encoding proves defective during the warranty period, T ektronix will provide a replacement in exchange for the defective medium. Except as to the media on which this software product is furnished, this software product is provided “as is” without warranty of any kind, either express or implied. T ektronix does not warrant that the functions contained in this software product will meet Customer’s requirements or that the operation of the programs will be uninterrupted or error-free.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period. If T ektronix is unable to provide a replacement that is free from defects in materials and workmanship within a reasonable time thereafter, Customer may terminate the license for this software product and return this software product and any associated materials for credit or refund.
THIS WARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPLACE DEFECTIVE MEDIA OR REFUND CUSTOMER’S PAYMENT IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT , SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Table of Contents

Getting Started

Operating Basics

Related Documentation vii. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Conventions vii. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contacting T ektronix viii. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Product Description 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Compatibility 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Requirements and Restrictions 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Updates 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Accessories 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installation 3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installing the Application 3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Updates Through a Web Browser 4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Equipment Under T est Connections 4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Functional Overview 5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Starting the Application 5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Local Help 6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using Other Functions 6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T aking a Measurement 7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Viewing the Results 9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Statistics Readout 9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Histogram Graphical Format 9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Profile Graphical Format 10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Exiting the Application 10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tutorial 11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setting Up the Oscilloscope 11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Starting the Application 11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Loading the Reference Waveform File 11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T aking Measurements from One Waveform 12. . . . . . . . . . . . . . . . . . . . . . . . . . . .
T aking Measurements from Two Waveforms 17. . . . . . . . . . . . . . . . . . . . . . . . . . .
Saving a Setup and Exiting the Application 19. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Recalling a Setup 20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Reference

Menu Structure 21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setup Parameters 23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Saving and Recalling Setups 23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Single Waveform Measurements 24. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Histogram Format Setup 26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Profile Format Setup 27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Duo Waveform Measurements 29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Localizing Measurements 31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TDSJIT1 Jitter Analysis Application User Manual
i
Table of Contents

Index

Measurement Descriptions 33. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
W arning Messages 34. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Oscilloscope Setup Guidelines 34. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Methodology 34. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Edge-Time Measurements 35. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Period Measurement 35. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Frequency Measurement 36. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cycle-to-Cycle Measurement 36. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
N-Cycle Measurement 37. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pulse Width Measurement 37. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Duty Cycle Measurement 38. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Skew Measurement 39. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ii
TDSJIT1 Jitter Analysis Application User Manual

List of Figures

Table of Contents
Figure 1: TDSJIT1 Jitter Analysis Application 1. . . . . . . . . . . . . . . . . .
Figure 2: Starting the application 5. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3: Jitter Analysis Application initial display 6. . . . . . . . . . . . . .
Figure 4: Returning to the application 7. . . . . . . . . . . . . . . . . . . . . . . . .
Figure 5: The Control menu 8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6: Setup Period menu (basic setup menu) 12. . . . . . . . . . . . . . . . .
Figure 7: Period lesson: Statistics readout 13. . . . . . . . . . . . . . . . . . . . . .
Figure 8: Histogram format menu 13. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 9: Period lesson: Histogram format 14. . . . . . . . . . . . . . . . . . . . .
Figure 10: Profile format menu 15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 11: Period lesson: Profile format 16. . . . . . . . . . . . . . . . . . . . . . . .
Figure 12: Setup Skew menu 17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 13: Skew lesson: Statistics readout 18. . . . . . . . . . . . . . . . . . . . . .
Figure 14: Skew lesson: Histogram format 18. . . . . . . . . . . . . . . . . . . . . .
Figure 15: Skew lesson: Profile format 19. . . . . . . . . . . . . . . . . . . . . . . . .
Figure 16: Application menu structure 21. . . . . . . . . . . . . . . . . . . . . . . . .
Figure 17: Save Setup and Recall Setup selections in the Utility menu 23
Figure 18: Cycle Start Edge menu 25. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 19: Histogram format menu, part 1 27. . . . . . . . . . . . . . . . . . . . .
Figure 20: Histogram format menu, part 2 27. . . . . . . . . . . . . . . . . . . . .
Figure 21: Profile format menu, part 1 28. . . . . . . . . . . . . . . . . . . . . . . . .
Figure 22: Profile format menu, part 2 28. . . . . . . . . . . . . . . . . . . . . . . . .
Figure 23: Setup Skew menu 30. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 24: Example of a Period measurement and the results 33. . . . . .
TDSJIT1 Jitter Analysis Application User Manual
iii
Table of Contents
iv
TDSJIT1 Jitter Analysis Application User Manual

General Safety Summary

Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read the General Safety Summary in other system manuals for warnings and cautions related to operating the system.
Connect and Disconnect Properly . Connect the probe output to the measurement instrument before connecting the probe to the circuit under test. Disconnect the probe input and the probe ground from the circuit under test before disconnecting the probe from the measurement instrument.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Symbols and Terms
T erms in this Manual. This term may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
TDSJIT1 Jitter Analysis Application User Manual
v
General Safety Summary
vi
TDSJIT1 Jitter Analysis Application User Manual

Preface

Related Documentation

Conventions

This manual contains operating information for the TDSJIT1 Jitter Analysis Application. The manual consists of the following chapters:
H The chapter Getting Started briefly describes the TDSJIT1 Jitter Analysis
Application and provides installation instructions. H The chapter Operating Basics covers basic operating principles of the
application and includes a tutorial. H The chapter Reference describes how to perform specific tasks.
Your oscilloscope user manual provides operating instructions for general oscilloscope features.
This manual uses the following conventions: H This manual refers to the TDSJIT1 Jitter Analysis Application as either the
Jitter Analysis Application or as the application. H When steps require that you make a sequence of selections using front panel
controls and menu buttons, an arrow (
front panel button and a menu, or between menus. Names that are for a main
menu or side menu item are clearly indicated: Press VERTICAL MENU
Coupling (main) DC (side) Bandwidth (main) 250 MHz (side).
) marks each transition between a
TDSJIT1 Jitter Analysis Application User Manual
vii
Preface

Contacting Tektronix

Product Support
Service Support
For other information
To write us Tektronix, Inc.
For application-oriented questions about a Tektronix measure­ment product, call toll free in North America: 1-800-TEK-WIDE (1-800-835-9433 ext. 2400) 6:00 a.m. – 5:00 p.m. Pacific time
Or contact us by e-mail: tm_app_supp@tek.com
For product support outside of North America, contact your local Tektronix distributor or sales office.
Contact your local Tektronix distributor or sales office. Or visit our web site for a listing of worldwide service locations.
http://www.tektronix.com In North America:
1-800-TEK-WIDE (1-800-835-9433) An operator will direct your call.
P.O. Box 1000 Wilsonville, OR 97070-1000
viii
TDSJIT1 Jitter Analysis Application User Manual
Getting Started

Product Description

The TDSJIT1 Jitter Analysis Application is a Java-based application that enhances the jitter analysis capability of Tektronix TDS oscilloscopes. The application offers many of the same timing measurements and two additional measurements: Cycle-to-Cycle and N-Cycle.
Oscilloscopes can take only one measurement in an acquisition waveform. The TDSJIT1 application can take and accumulate measurements throughout an entire waveform.
The application can display the measurement results three ways: as Statistics, in a Histogram graphical format, or in a Profile graphical format (measurement versus measurement index numbers).
Figure 1 shows an example of the Histogram format of a period measurement.
Figure 1: TDSJIT1 Jitter Analysis Application

Compatibility

The Jitter Analysis Application is compatible with the following Tektronix oscilloscopes:
H All TDS 500D and 700D Digital Phosphor Oscilloscopes with Option HD
(hard disk drive) or Option 2M (hard disk drive plus 8 MB record length)
H TDS 600C Digitizing Oscilloscopes with an Option HD (hard disk drive) H TDS 700C Color Digitizing Oscilloscopes serial number B020100 and up,
with Option HD (hard disk drive) or Option 2M (hard disk drive plus 8 MB
record length), and with firmware version 5.2e and up
TDSJIT1 Jitter Analysis Application User Manual
1
Product Description
For a current list of compatible oscilloscopes, check the Tektronix, Inc. web site, http://www.tektronix.com/Measurement/scopes/index.html Drivers category.

Requirements and Restrictions

The TDS Run-Time Environment V1.1 and above must be installed on the oscilloscope to operate the Jitter Analysis Application.
The application does not support control by external GPIB commands.

Updates

You can find information about this and other applications at the Tektronix, Inc. web site, http://www.tektronix.com/Measurement/scopes/index.html Software and Drivers category. Check this site for application updates that you can download and for free applications.
in the Software and
in the

Accessories

There are no standard accessories for this product.
2
TDSJIT1 Jitter Analysis Application User Manual

Installation

This application resides on a floppy disk. You can download updates, if any, from the Tektronix ftp site through a web browser.
NOTE. To operate the Jitter Analysis Application, the TDS Run-Time Environment V1.1 or above must be installed on your oscilloscope.

Installing the Application

To install the application from the floppy disk to your oscilloscope, follow these steps:
1. Power off the oscilloscope.
NOTE. Additional information about the application or installation is located in a Readme.txt file on the floppy disk. You should insert the floppy disk into a DOS-based personal computer and read the Readme.txt file before you continue.
If you are updating the application, the Readme.txt file on the Tektronix ftp site supercedes the Readme.txt file on the floppy disk.
2. Insert the disk in the floppy disk drive and power on the oscilloscope.
NOTE. To verify that the TDS Run-Time Environment V1.1 or above is installed, watch for the name to appear at the top of the display when you power on the oscilloscope. If it does not appear, contact your local Tektronix sales office.
After performing the power-up self-test, the oscilloscope automatically
begins the installation procedure.
As the application loads from the disk, the oscilloscope displays a clock icon
to indicate that it is busy. Also, the floppy disk drive LED is on, indicating
activity. If the clock icon continues to display after the floppy disk LED has
gone out, a problem has occurred with the installation. Repeat the above
procedure. If the problem persists, contact your Tektronix representative.
When the installation is complete, an Installation Complete message
displays.
3. Remove the floppy disk and cycle the power to the oscilloscope.
TDSJIT1 Jitter Analysis Application User Manual
3
Installation

Updates Through a Web Browser

If there is an update for this application, it will be available from the Tektronix ftp site. To install an application update, you will need to download it from the ftp site to a hard disk, copy it to a blank DOS-formatted floppy disk, and then install it on your oscilloscope.
NOTE. More information about changes to the application or installation is in a Readme.txt file on the ftp site. You should read it before you continue.
To copy the application from a web browser, follow these steps:
1. Access the ftp site at ftp://ftp.tek.com/mbd/support/00–index.html#1
2. Scroll through the files to the TDSJIT1 application, select the file, and
download it to your hard disk drive. If necessary, unzip the file.
3. Copy the application from the hard disk to a blank DOS-formatted floppy
disk.
4. Follow the Installing the Application procedure on page 3.

Equipment Under Test Connections

You can use any compatible probe to connect between your EUT (equipment under test) and the oscilloscope. The connection is usually to a clock signal.
The Skew measurement requires two channels or two reference waveforms.
WARNING. To avoid electric shock, you must ensure that power is removed from the EUT before attaching a probe to it. Do not touch exposed conductors except with the properly rated probe tips. Refer to the probe manual for proper use.
Power down the EUT before connecting the probe to it.
.
Table 1 shows the default channel-to-waveform assignments.
T able 1: Channel assignments
Channel or reference Waveform assignment
Ch 1 Main waveform, such as the Clock signal Ch 2 Skewed waveform Ref1 Profile graphical format Ref2 Histogram graphical format
4
TDSJIT1 Jitter Analysis Application User Manual
Operating Basics
Loading...
+ 44 hidden pages