Getting Started with TDSDVI
Side one of this Quick Reference Guide is designed to get
you started making measurements with the DVI Compliance
Test Solution (TDSDVI). Side two contains a complete
menu tree for the TDSDVI software.
NOTE. For complete operating instructions
and General Safety information, refer to the
Online Help for the application.
The TDSDVI software application tests for compliance with
DVI specifications. TDSDVI software is designed to meet
the compliance test requirements of the DVI industry for
physical layer measurements. Supported tests include Eye
Diagram Test, Peak-to-Peak Jitter , Inter-Pair Skew, Intra-Pair
Skew, and Rise and Fall Time testing.
Performing Transmitter Eye-diagram Test
1. From the TDS Windows menu, select File > Run
Application > DVI Compliance Test Solution.
2. Attach the TPA-P test fixture to the device under test. To
calculate Tbit, connect the differential probe to Rxc+
and Rxc-- (transmitter clock). For Eye measurement,
connect another differential probe to Rx0+ and Rx0-(data pair).
3. Connect an SMA cable to the recovered clock (PLL
Clock) output on the fixture.
5. To calculate Tbit, select the Tbit button on the Define
Tbit pane of the application.
6. Select the appropriate channel.
7. Press the “Running Man” icon to run the application.
The Tbit value displays in the V alue field.
8. Select Measurement > Select >Transmitter> Eye
Diagram. (Or, in the Transmitter tab, select the Eye
Diagram measurement.)
9. Press the configure button to configure the application.
The detailed results and the eye diagram display as
shown below.
11. Select Utilities > Report Generator from the TDSDVI
menu to generate a compliance test report.
4. Connect the differential probes and the SMA cable from
the test fixture to your TDS oscilloscope.
For up to date information on Tektronix oscilloscope solutions for DVI testing, go to www.tektronix.com/dvi.
TDSDVI Ordering Information
This application supports TDS7704B, CSA7404B,
TDS7404B, TDS7254B, CSA7404, TDS7404, TDS7254,
TDS6604, and TDS6404 series oscilloscopes.
Order Option DVI (along with an oscilloscope).
Order TDS6/7 UP, TDS7B UP, CSA7B UP, or CSA7 UP,
Option DVI (if ordered as an upgrade to an existing
oscilloscope).
10. Press the “Running Man” icon to run the application.
Recommended Accessories
P7350, P7330, P6330 -- Differential Probes
P7240 -- Single ended probes
AWG510, AWG520, AWG610, AWG710 -- Arbitrary
Waveform Generator
DG2020A -- Data Generator
TDS8000/ 80E04 -- Time Domain Reflectometer
TDSJIT3 -- Jitter analysis package
TDSDVI
Compliance Test Solution
Reference
www.tektronix.com
*P071130001*
071-1300-01

TDSDVI Menu Tree
File
Recall Default
Save . . .
Recall...
Recently Saved
Recently Recalled
Preferences . . .
Minimize
Exit
Select
Yes | No
File browser
Save | Cancel
File browser
Open | Cancel
List of Files
List of Files
General
Advanced
Yes
No
Cancel
Define Tbit
Select Measurement
Use cursors for Eye, RT& FT testing | Prompt for signal connection |
Ref Wfm deletion prompt for Eye | Show report after generation |
Use zoomed eye mask in report
Number of acquisitions
Prompt for result reset
Prompt for Tbit validity
Cable Eye Diagram Test Point
Number of acquisitions
Restore Scope Settings
Calculated
User
Transmitter
Cable
Receiver
Tx Clock
Value
Test at TP2
Test at TP3
Test at TP2 & TP3
Ch1, Ch2, Ch3, Ch4
Eye Diagram | Rise and Fall Time | Pk-Pk Jitter |
Intra-Pair Skew | Inter-Pair Skew
High-Amp Eye Diagram | Low-Amp Eye Diagram |
Pk-Pk Jitter | Intra-Pair Skew | Inter-Pair Skew
High-Amp Eye Diagram | Low-Amp Eye Diagram
Measurements
Results
Configure
Result Details
Report Generator
Transmitter Eye Diagram
Transmitter Rise & Fall Time
Transmitter Pk - Pk Jitter
Transmitter Intra-Pair Skew
Transmitter Inter-Pair Skew
Cable High-Amplitude Eye Diagram
Cable Low-Amplitude Eye Diagram
Cable Pk-Pk Jitter
Cable Intra-Pair Skew
Cable Inter-Pair Skew
Receiver High-Amp Eye Diagram Select Source | Select Pair | Number of Eyes
Receiver Low-Amp Eye Diagram
Details
Statistics
Report Setup
Device Detail
Mode
Clock Frequency
Data | Trigger | Pair | Number of Eyes | Calculate Vswing
Data | Trigger | Pair | Hysteresis | Calculate Vsw ing
Tx Clock | Trigger
Source1 | Source2 | Pair | Hysteresis
Source1 | Source2 | Differential Probe |
Pair1 | Pair2 | Hysteresis | Polarity
Data | Trigger | Pair | Number of Eyes
Data | Trigger | Pair | Number of Eyes
Tx Clock | Trigger
Source1 | Source2 | Pair | Hysteresis
Source1 | Source2 | Differential Probe |
Pair1 | Pair2 | Hysteresis | Polarity
Select Source | Select Pair | Number of Eyes
ID | Description | Prefix
Manual | Automatic
Defined | Resolution | Refresh Rate
Utilities
Help
Compare Results
Topics . . .
About TDSDVI...
Contact Tektronix
Report Name
Select Type
Select File(s) to Compare
Select Destination File
Directory | File Name
Current Vs File | File Vs File
Copyright Tektronix, Inc.