Tektronix TDS820 Data Sheet

Page 1
TDS 820
Features
Specs
Ordering Information
Digital Sampling Oscilloscopes
TDS 820
This product is discontinued. Characteristics
SIGNAL Acquisition System
TDS 820 TDS 820 Opt. 1D
Channels 2 2 Rise time 58.3 ps 43.8 ps Bandwidth
(0.35/rise time)
6 GHz 8 GHz
Max operating input voltage
2 V
p-p
; ±3 V DC 1 V
p-p
; ±1.5 V DC
Sensitivity 2 m V/div to
200 mV/div
1 mV/div to 100 mV/div
Random noise 1.2 mV max, 600
µV typical
600 µV max, 300
µV typical
DC Gain Accuracy - ±0.7% after user-initiated automatic vertical calibration.
Vertical Resolution - 14-Bits (Approx. 16,384 levels over
10.24 vertical divisions). Input Impedance - 50 Ohm.
Acquisition Modes
Normal - One sample acquired with each trigger event. Envelope - Max/min values acquired over one or more
acquisitions. Average - Waveform averages selectable from 2 to 10,000.
Time Base System
Time Bases - Main, Delayed.
Page 2
Time/Division Range - 20 ps/division to 5 ms/division in 1-2-5
steps or settable from the numeric keypad in 5 ps steps.
Delta time measurement accuracy -
Time interval Accuracy
Ti >1 ns 0.1% x interval +15ps 100 ps 5 ps 10 ps 2 ps
Channel Deskew - Up to 100 ns (each channel). Record Length - 500, 1000, 2000, and 5000, and 15,000
samples per channel. Pre -Trigger View Time - 1.5 ns.
Triggering System
Trigger Sources - External input, internal rate generator, CH 1, CH 2.
External Trigger Sensitivity - 40 mV
p-p
from DC to 200 MHz,
increasing linearly to 200 mV
p-p
at 2 GHz.
External Trigger Minimum Pulse Width - 0.25 ns. Internal Trigger Sensitivity - 80 mV
p-p
from DC to 200 MHz,
increasing linearly to 400 mV
p-p
at 1 GHz.
Trigger Delay Jitter - 3 ps RMS + 30 ppm of time base delay. Holdoff Range - 15 µs to 2 s.
Display
Waveform Style - Dots or vectors. Infinite and variable persistence from 250 ms to 10 s.
Gray Scaling - With variable persistence selected, waveform points gradually decay through 16 levels of intensity, providing "z-axis" information about rapidly changing waveforms.
Graticules - Full, grid, cross hair, frame. Format - YT and XY. CRT Type - 7 in. diagonal, magnetic deflection. Horizontal
raster-scan. P4 White phosphor.
Page 3
CRT Resolution - 640 horizontal by 480 vertical displayed pixels.
Measurement System
Automatic waveform measurements -
Period Frequency High Low + Width - Width Maximum Minimum Rise Fall Peak to Peak Amplitude + Duty Cycle - Duty Cycle + Overshoot - Overshoot Propagation Delay Burst Width Mean Cycle Mean RMS Cycle RMS Phase Cycle Area + Cross - Cross Area
Continuous update of up to four measurements on any combination of waveforms. Snapshot mode shows all measurements on the selected waveforms.
Thresholds - Settable in percentage or voltage. Cursor Measurements - Absolute, Delta; Volts, Time,
Frequency. Cursor Types - Horizontal bars (volts); Vertical bars (time);
Paired (volts and time).
Waveform Processing
Waveform Functions - Interpolate-selectable sin(x)/x or linear, Average, FFT, integrate, and differentiate.
Arithmetic Operators - Add, Subtract, Multiply, Invert. Autosetup - Single button, automatic setup on selected input
signal for vertical, horizontal and trigger systems.
Computer Interface
Page 4
GPIB (IEEE -488.2) Programmability - Full talk/listen modes. Control of all modes, settings, and measurements.
Hard Copy
Formats - HP ThinkJet, Epson, Postscript, Interleaf, DeskJet, LaserJet, EPS Monochrome, EPS Color, TIFF, PCX, BMP, HPGL.
Optional Hardcopy Interface - Centronics and RS-232C.
Storage
Waveforms - Up to 15 K points. Setups - 10 front-panel setups.
Power Requirements
Line Voltage Range - 90 to 250 V RMS. Line Frequency - 47 to 63 Hz. Power Consumption - 250 W max.
Environment and Safety
Temperature - Operating: 0 to +50°C. non-operating: -40 to +75°C.
Humidity - Operating and non-operating: Up to 95% relative humidity at or below +40°C; to 75% relative humidity from +41°C to 50°C.
Altitude - Operating: 15,000 ft. non-operating: 40,000 ft. Electromagnetic Compatibility - Meets MIL-STD-461C,
CE-03, Part 4, Curve # 1, RE-02, Part 7; meets VDE 0871, Category B, FCC rules and regulations, Part 15, Subpart B, Class A.
Safety - UL3111-1, CSA1010.1, EN61010-1, IEC61010-1.
Physical Characteristics
Dimensions mm in.
Height 193 7.6 with acc. pouch 236 9.3 Width 445 17.5
Page 5
Depth with front cover 432 17
Weight kg lbs.
Net 13.2 29.1 Shipping 23.6 52
Top of Page
Features
Specs
Ordering Information
The TDS Series complies with IEEE Standard
488.2-1987, and with Tektronix Standard Codes and Formats.
Tektronix Measurement products are manufactured in ISO registered facilities.
49A-10733-3p88, 06/1997, 09/01/1999
© Copyright Tektronix, Inc | webmaster@tek.com | Privacy Policy