TCPA300/400 Amplifiers &
TCP300/400 Series AC/DC Current Probes
071-1183-03
This document applies for firmware version 1.0
and above.
Warning
The servicing instructions are for use by qualified
personnel only. To avoid personal injury, do not
perform any servicing unless you are qualified to
do so. Refer to all safety summaries prior to
performing service.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
HIn North America, call 1-800-833-9200.
HWorldwide, visit www.tektronix.com to find contacts in your area.
Warranty 2
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1)
year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its
option, either will repair the defective product without charge for parts and labor, or will provide a replacement in
exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty work
may be new or reconditioned to like new performance. All replaced parts, modules and products become the
property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration
of the warranty period and make suitable arrangements for the performance of service. Customer shall be
responsible for packaging and shipping the defective product to the service center designated by Tektronix, with
shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a
location within the country in which the Tektronix service center is located. Customer shall be responsible for
paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage
resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product;
b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any
damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been
modified or integrated with other products when the effect of such modification or integration increases the time
or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY
OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A P ARTICULAR PURPOSE.
TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND
EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY.
TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL,
OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS
ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Table 8--1: Power cord identification8--5.........................
x
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it. To avoid potential hazards, use this
product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read
the General Safety Summary in other system manuals for warnings and cautions
related to operating the system.
ToAvoidFireor
Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and
certified for the country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Connect and Disconnect Properly. Connect the probe output to the measurement
instrument before connecting the probe to the circuit under test. Disconnect the
probe input and the probe ground from the circuit under test before disconnecting
the probe from the measurement instrument.
Ground the Product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
Connect the ground lead of the probe to earth ground only.
Do Not Operate Without Covers. Do not operate this product with covers or panels
removed.
Use Proper Fuse. Use only the fuse type and rating specified for this product.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
xi
General Safety Summary
Provide Proper Ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
Symbols and Terms
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you read the
marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
CAUTION
Refer to Manual
Do not connect to or
remove from an
uninsulated conductor that
is HAZARDOUS LIVE.
Protective Ground
(Earth) Terminal
Breakable.
Do not drop.
Do not connect
to or discon-
nect from unin-
sulated Haz-
ardous Live
conductors.
WARNING
Hot Surface
Use only on an
insulated wire.
xii
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, switch off the instrument power, then
disconnect the power cord from the mains power.
Use Care When Servicing With Power On. Dangerous voltages or currents may
exist in this product. Disconnect power, remove battery (if applicable), and
disconnect test leads before removing protective panels, soldering, or replacing
components.
To avoid electric shock, do not touch exposed connections.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
xiii
Service Safety Summary
xiv
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Preface
This Instruction Manual supports the operation and basic maintenance of the
TCPA300 and TCPA400 Current Probe Amplifiers, and the TCP300/400 Series
AC/DC current probes that mate with the amplifiers. The current probes covered
in this manual are listed below:
HTCP312 (30 amps, 100 MHz, compatible with TCPA300)
HTCP305 (50 amps, 50 MHz, compatible with TCPA300)
HTCP303 (150 amps, 15 MHz, compatible with TCPA300)
HTCP404XL (500 amps*, 2 MHz, compatible with TCPA400)
*750 amps DC derated with duty cycle
If you are not familiar with these products, please refer to the Getting Started and
Operating Basics chapters of this manual for basic operating information.
If you are an advanced user, the Reference section contains information on
advanced applications as well as user diagnostic and troubleshooting information.
The Performance Verification and Adjustment Procedure sections support the
qualification and calibration of the probes when used with either amplifier.
Manual Conventions
The Maintenance section supports the routine maintenance and repair of
mechanical parts associated with the amplifiers.
The Glossary and Index are provided as quick reference locators for important
information.
The term “amplifier” is used to refer to either the TCPA300 or TCPA400 when
referring to common attributes. If a subject is unique to either amplifier, the
amplifier will be referred to directly by model.
The terms “current probe” and “probe” are used to refer to any of the
TCP300/400 Series current probes when referring to common attributes. If a
subject is unique to a particular probe, the probe will be referred to directly by
model.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
xv
Preface
xvi
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Getting Started
Getting Started
The TCPA300 and TCPA400 current probe amplifiers let you use one probe to
simultaneously measure AC and DC current. The amplifiers convert the sensed
current into a proportional voltage signal that you can measure directly with an
oscilloscope.
The TCPA300 and TCPA400 current probe amplifiers provide better linearity
than other current measurement systems because of a current feedback process
used with the probe. DC measurement capability and high bandwidth allow the
amplifiers to accurately represent square waves and fast-rise signals.
The TCPA300 and TCPA400 and associated probes provide these features:
HSimultaneous DC and AC current measurements up to 750 A peak
HHigh sensitivity
HOne-button autobalancing and probe degaussing
HNo adjustments needed to match a current probe to an individual amplifier
HAC or DC coupling of signal
System Configuration
HDirect scaling and unit readout on compatible TEKPROBE level II
oscilloscopes
A complete current measurement system consists of a current probe amplifier, a
compatible current probe, and an appropriate oscilloscope. Refer to Figure 1--1.
AmplifierTest oscilloscope
Current probe
Input
Output
50 Ω oscilloscope input -- use the
TEKPROBE Interface Cable or use a 50 Ω
cable. (Add 50 Ω termination here if
oscilloscope has only high-impedance input.)
Figure 1- 1: Typical TCPA300/400 current measurement system
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
1- 1
Getting Started
TCPA300 and TCPA400
Current Probe Amplifiers
Current Probes
The amplifier amplifies the current sensed by the probe and converts the current
to a proportional voltage that is displayed on an oscilloscope or other similar
measuring device.
The following Tektronix current probes are compatible with the TCPA300
Amplifier:
HTCP312 (30 amps, 100 MHz)
HTCP305 (50 amps, 50 MHz)
HTCP303 (150 amps, 15 MHz)
The following Tektronix current probes are compatible with the TCPA400
Amplifier:
HTCP404XL (750 amps*, 2 MHz)
* 500 amps continuous, 750 amps DC derated with duty cycle
You can also use the CT-4 High-Current Transformer with the TCP305 and
TCP312 current probes to extend the AC current measurement range to
20,000 peak amps.
Oscilloscope
An oscilloscope displays the output from the current measuring system. A 50 Ω
cable is included to connect the amplifier to the oscilloscope input channel. A
TEKPROBE-to-TEKPROBE interface cable is also included for connecting to
TEKPROBE level II oscilloscopes.
If the oscilloscope does not have an input that can be set to 50 Ω impedance, you
need a feedthrough 50 Ω termination. This termination is included as a standard
accessory with your TCPA300 and TCPA400 Current Probe Amplifiers.
1- 2
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Options
Getting Started
Table 1--1 lists options that are available for the TCPA300 and TCPA400
amplifiers.
Table 1- 1: Amplifier options
OptionDescription
A1Universal Euro power cord
A2United Kingdom power cord
A3Australia power cord
A5Switzerland power cord
A6Japan power cord
ACChina power cord
A99No power cord
L5Japanese Instruction Manual
Table 1--2 lists the Tektronix service options you can order for your amplifiers
and probes. Designed to support tracking of calibration to requirements of
ISO9000 and to provide for extended repair coverage, these options help fix your
long-term maintenance costs and eliminate unplanned expenditures. Tektronix
Service Options are available at the time you order your instrument. Contact
your local Tektronix Sales Office for more information.
Table 1- 2: Service options
OptionDescription
D1Provides the initial Test Data Report from the factory on delivery.
C3Provides factory calibration certification on delivery, plus two more years of
calibration coverage. Throughout the coverage period, the instrument will be
calibrated according to its Recommended Calibration Interval.
D3Provides test data on delivery plus a Test Data Report for every calibration
performed during three years of coverage (requires Option C3).
R3Extends product repair warranty to a total of three years.
C5Provides factory calibration certification on delivery, plus four more years of
calibration coverage. Throughout the coverage period, the instrument will be
calibrated according to its Recommended Calibration Interval.
D5Provides test data on delivery plus a Test Data Report for every calibration
performed during five years of coverage (requires Option C5).
R5Extends product repair warranty to a total of five years.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
1- 3
Getting Started
Standard Accessories
The following accessories are shipped with the amplifiers and probes. Refer to
the Replaceable Parts List beginning on page 8--1 for Tektronix part numbers to
use in ordering accessories.
Amplifiers
Probes
The following accessories are shipped with the TCPA300 and TCPA400
amplifiers.
HPower Cord (customer-chosen option)
HBNC Cable
HTermination, 50 Ω, 2W
HTEKPROBE Interconnect Cable
HInstruction Manual (English or Japanese; customer-chosen language option)
HCertificate of Traceable Calibration
When you order a current probe, you will receive these accessories:
HProbe cover
HProbe ground lead, 6 inch length (TCP305 and TCP312 only)
HInstruction Sheet
HCertificate of Traceable Calibration
1- 4
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Optional Accessories
Getting Started
You can order the following optional accessories for the amplifiers and probes.
Refer to the Replaceable Parts List beginning on page 8--1 for Tektronix part
numbers to use in ordering accessories.
HOne-turn 50 Ω current loop. The current loop is used in the performance
verification procedure for checking the performance of the TCPA300
Amplifier and the compatible probes.
HCT-4 High-Current Transformer. If you need to measure high-amplitude AC
currents, consider using the CT-4 with the TCP303 and TCP312 probes. The
CT-4 provides step-down ratios of 20:1 or 1000:1. For more information
about the CT-4, consult your Tektronix sales representative.
HTCPA Calibration Adapter. Use the TCPA Calibration Adapter to verify the
amplifier(s) performance independent of the current probes.
HTravel Case. The travel case includes room to store one amplifier and two
current probes, along with related cables and adapters.
HDeskew Fixture. This fixture converts the PROBE COMPENSATION output
or TRIGGER OUTPUT of the TDS5000 or TDS7000 into a set of test point
connections that allow you a convenient way to compensate for timing
differences between voltage and current probes.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
1- 5
Getting Started
Probe Covers
The TCP300/400 Series Current Probes come with a probe cover that stores the
probe when not in use. Use the probe cover to hold your probe in a convenient
place at your bench or workstation when you are not using it. You can attach the
probe cover to the side of the bench to keep the probe off of your work surface.
See Figure 1--2.
TCP305/312
TCP303/TCP404XL
Figure 1- 2: Using the probe covers
1- 6
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Travel Case
Getting Started
The travel case is a recommended accessory for the TCP A300/400 Amplifiers.
The travel case includes room to store one amplifier and two TCP300/400 Series
Current Probes, one of each size. (For example, you can store a TCP305 and a
TCP303 probe.) A compartment is included to store associated cables and
terminations. See Figure 1--3 for the proper location of the equipment.
Instruction
manual
Large current
probe
Probe holders
Small current
probe
Figure 1- 3: Equipment locations in the travel case
Cables &
terminations
Amplifier
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
1- 7
Getting Started
Connecting the Amplifier to an Oscilloscope
You will need an oscilloscope to display the TCPA300 and TCPA400 measurement output. To use the full dynamic range of the probe/amplifier combination,
the oscilloscope must be capable of displaying a vertical scale factor of 1 mV/div
to 1V/div.
If you are using a TEKPROBE II-compatible oscilloscope, use the TEKPROBEto-TEKPROBE interface cable. Otherwise, use the supplied 50 Ω BNC cable to
connect the amplifier OUTPUT connector to your oscilloscope (see Figure 1--1
on page 1--1).
The input impedance of the oscilloscope channel must be 50 Ω, or you will
encounter slowed pulse response, increased aberrations, or incorrect DC
measurement amplitudes. If your oscilloscope provides only 1 MΩ inputs, you
need to attach a 50 Ω feed-through termination between the oscilloscope input
and the BNC cable. Do not install this termination at the amplifier end of the
BNC cable.
To utilize the full bandwidth capability of the TCPA300 and TCPA400 and
attached current probe, the oscilloscope bandwidth must be approximately five
times that of the current probe. For example, when using a TCP312 Current
Probe, the oscilloscope bandwidth must be at least 500 MHz. When using a
TCP305 Current Probe, the oscilloscope bandwidth must be at least 250 MHz.
Power on the Amplifier
After you have connected the amplifier to the oscilloscope, allow the equipment
to warm up to a stable temperature; usually 20 minutes is required.
Connect the power cord to the power input connector on the rear of the amplifier,
and then connect the power cord to your local mains supply (100 VAC to 240
VAC, 50 Hz to 400 Hz). To allow for proper ventilation, place the rear panel of
the amplifier at least 2 inches away from any obstructions. Set the amplifier on
the bottom rubber feet, and keep papers and other items away from the bottom of
the amplifier which could restrict airflow and cause overheating.
Power on the amplifier by pressing the ON/STANDBY button at the lower-left
corner of the front panel. The amplifier goes through a self-test and cycles the
front-panel LEDs.
NOTE. The amplifier stores the power state it is in when the power cord is
unplugged. If you do not put the amplifier into STANDBY mode before unplugging it, the amplifier will power on immediately when you plug it in again.
When you connect a probe to the amplifier, the amplifier uses detection circuitry
to indicate probe conditions such as noncompatible probe type and probe open.
1- 8
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual